K Number
K173492

Validate with FDA (Live)

Date Cleared
2017-12-22

(39 days)

Product Code
Regulation Number
866.6060
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MolecularMD MRDx BCR-ABL Test is an in vitro diagnostic test for the quantitative detection of BCR-ABL 1 transcripts (e13a2/b2a2 and/or e14a2/63a2) and the ABL1 endogenous control mRNA in peripheral blood specimens from patients previously diagnosed with t(9:22) positive chronic myeloid leukemia (CML). The ratio of BCR-ABL1 is calculated and reported on the WHO International Scale. The test utilizes quantitative, real-time reverse transcription polymerase chain reaction performed on the Applied Biosystems 7500 Fast Dx instrument.

The MolecularMD MRDx BCR-ABL Test is intended to measure BCR-ABL mRNA transcript levels in patients diagnosed with t(9;22) positive CML during monitoring of treatment with Tyrosine Kinase Inhibitors (TKIs).

The device is also intended to be used in the serial monitoring for BCR-ABL mRNA transcript levels as an aid in identifying CML patients in the chronic phase being treated with nilotinib who may be candidates for treatment discontinuation and for monitoring of treatment-free remission.

The test does not differentiate between e13a2 or e14a2 fusion transcripts and does not monitor other rare fusion franscripts resulting from t(9:22). The test is not intended for the diagnosis of CML.

Device Description

The MolecularMD MRDx® BCR-ABL Test is a quantitative, real-time polymerase chain reaction test that provides quantitation of BCR-ABL1 (hereafter BCR-ABL) transcript e13a2/b2a2 or e14a2/b3a2 and ABL1 (hereafter ABL) transcript levels in RNA extracted from peripheral blood samples collected from CML patients. Peripheral blood is collected in either EDTA or PAXgene Blood RNA Tubes. Each collection tube type requires a specified RNA extraction method and RNA input amount. Total RNA is extracted from peripheral blood and serves as the template for RT-qPCR. The test is performed using a onestep RT-qPCR protocol wherein the reverse transcription and quantitative, real-time PCR reactions are performed in the same well. BCR-ABL and ABL amplicons are generated and detected in real-time using TaqMan® MGB probes. The MolecularMD MRDx BCR-ABL Test is performed on the Applied Biosystems 7500 Fast Dx Real-Time PCR Instrument. The instrument integrates a thermal cycler, a fluorometer and application specific software. The ABI 7500 Fast Dx instrument software v1.4.1 calculates the BCR-ABL and ABL copy numbers using a standard curve generated with calibrators on each individual plate. The data are exported as a CSV file for further analysis in the MRDx® BCR-ABL Test Software. The BCR-ABL/ABL ratio is calculated and converted to the International Scale by the MRDx BCR-ABL Test Software. The MRDx BCR-ABL Test Software is used to analyze all test results. This software, provided with the MRDx BCR-ABL Test, is used to calculate the BCR-ABL/ABL % IS and MR value for patient sample using the conversion factor for the MRDx BCR-ABL Test after validating each MRDx BCR-ABL Test result against the run acceptance criteria and sample acceptance criteria.

AI/ML Overview

The MolecularMD MRDx BCR-ABL Test is an in vitro diagnostic test for the quantitative detection of BCR-ABL1 transcripts and ABL1 endogenous control mRNA in peripheral blood specimens from patients previously diagnosed with t(9;22) positive chronic myeloid leukemia (CML). It calculates the ratio of BCR-ABL1 to ABL1 and reports it on the WHO International Scale. The device is intended to measure BCR-ABL mRNA transcript levels in CML patients during monitoring of treatment with Tyrosine Kinase Inhibitors (TKIs). It is also intended for serial monitoring of BCR-ABL mRNA transcript levels to aid in identifying CML patients in the chronic phase being treated with nilotinib who may be candidates for treatment discontinuation and for monitoring of treatment-free remission.

Here's an overview of the acceptance criteria and study data:


1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a consolidated table of acceptance criteria for all performance characteristics. However, individual acceptance criteria are mentioned within the description of each study. Below is a summary of acceptance criteria and the device's reported performance, extracted from the document:

Performance CharacteristicAcceptance CriteriaReported Device Performance
Traceability to WHO ISHigh traceability to WHO reference standards, consistent conversion factor across multiple lots.Deming regression analyses showed a conversion factor of 1.1 for all three lots (A, B, C), demonstrating consistency. Data from all three lots showed high traceability to WHO reference standards. Slope was approximately 1.0 (0.98-1.0) and correlation was 1.0 for all lots.
Accuracy (Correlation to ddPCR)High concordance between MRDx BCR-ABL Test results and the reference ddPCR test results, with acceptable predicted systematic differences at clinical decision points. (Implicit: Deming regression slope close to 1, y-intercept close to 0, high Pearson coefficient).EDTA: Deming Slope: 0.96 (95% CI [0.94, 0.99]), Deming Y-Intercept: 0.077 (95% CI [0.0027, 0.17]), Pearson Coefficient: 0.984. Predicted difference at MR3: -0.031 (95% CI [-0.053, -0.0049]), MR4.0: -0.067 (95% CI [-0.10, -0.031]), MR4.5: -0.085 (95% CI [-0.13, -0.039]).PAXgene: Deming Slope: 0.98 (95% CI [0.95, 1.0]), Deming Y-Intercept: -0.021 (95% CI [-0.11, 0.056]), Pearson Coefficient: 0.985. Predicted difference at MR3: -0.088 (95% CI [-0.11, -0.062]), MR4.0: -0.11 (95% CI [-0.15, -0.071]), MR4.5: -0.12 (95% CI [-0.17, -0.071]). Conclusion: High concordance observed for both tube types.
Limit of Blank (LoB)Not explicitly stated as a numerical criterion, but implies no measurable BCR-ABL values in negative samples.Out of 180 replicates from 30 BCR-ABL negative samples, 177 had no measurable BCR-ABL values. Three (1.7%) had measurements well below the LoD and were reported as undetected.
Limit of Detection (LoD)Not explicitly stated as a numerical criterion, but the calculated LoD values should be acceptable for clinical use.EDTA: Estimated LoD was 0.00029% IS (MR5.5) for both e13a2 and e14a2 transcripts. PAXgene: Estimated LoD was 0.00039% IS (MR5.4) for e13a2 and 0.00029% IS (MR5.5) for e14a2. The final LoD was determined to be 0.00029% IS (MR5.5) for EDTA and 0.00039% IS (MR5.4) for PAXgene.
Limit of Quantitation (LoQ)Total Error (TE) ≤ 0.5 log10. LoQ to be set at or below a clinically relevant threshold (e.g., MR4.5 for reporting).EDTA: LoQ was 0.0016% IS (MR4.8). PAXgene: LoQ was 0.0025% IS (MR4.6). The MRDx BCR-ABL Test Software limits the reported LoQ and quantitated results to 0.0032% IS (MR4.5) for both blood tube types.
Analytical Specificity (Interfering Substances)No clinically significant difference in results when compared to matrix control samples.Endogenous: Largest difference between endogenous substance spike and matrix control for PAXgene was -0.080 log10 (95% CI: -0.32 to 0.16) at MR4.5. For EDTA, it was 0.060 log10 (95% CI: -0.042 to 0.16) at MR4.5. No clinically significant difference observed. Exogenous: Largest difference between exogenous substance spike and matrix control for PAXgene was 0.040 log10 (95% CI: -0.010 to 0.090) at MR3.0. For EDTA, it was 0.010 log10 (95% CI: -0.022 to 0.22) at MR4.5. No clinically significant difference observed. Conclusion: None of the potential interferents had significant interference.
Analytical Specificity (Primer Specificity)Primers and probes amplify intended targets, no cross-reactivity with ABL2 IVT RNA. Cross-reactivity with e19a2 BCR-ABL is expected but mitigated by labeling.Sequencing results of PCR products for BCR-ABL and ABL matched Genbank sequences, demonstrating amplification of intended targets. No cross-reactivity with ABL2 IVT RNA observed. Cross-reactivity with e19a2 BCR-ABL transcript was observed as expected, and addressed in labeling with a precaution statement.
Analytical Specificity (Specimen Cross Contamination)No negative samples (0% IS) should have a BCR-ABL/ABL % IS value detectable above the LoD of the assay when juxtaposed with high positive samples.No negative samples for either extraction method had a detectable BCR-ABL/ABL % IS value above the LoD. Conclusion: No significant carryover between wells.
Precision (Repeatability)SD log10 ≤ 0.25.PAXgene: Highest SD was 0.10 for e14a2 MR4.5. All levels for both transcripts passed (SD ≤ 0.25 log10). EDTA: Highest SD was 0.087 for e14a2 MR4.5. All levels for both transcripts passed (SD ≤ 0.25 log10).
Precision (Reproducibility)Total %CV for MR values should be within acceptable limits (typically < 10-15% for molecular response assays, though not explicitly stated as a single numerical criterion, the presented data suggest these are acceptable).EDTA (Total SD/CV): Ranged from 0.051 MR (2.5% CV) for e14a2 L2 to 0.13 MR (3.0% CV) for e14a2 L5, and 0.060 MR (4.9% CV) for e13a2 L1 to 0.096 MR (2.5% CV) for e13a2 L4. PAXgene (Total Within-Laboratory SD/CV): Ranged from 0.057 MR (7.1% CV) for e13a2 L1 to 0.24 MR (5.3% CV) for e14a2 L5. All values appear to meet implicit precision requirements.
Linear Range (Assay Range)Linear regression analyses should show linearity (non-linear coefficients from second and third order polynomials insignificant (p > 0.05) or acceptable Degree of Nonlinearity). Assay should be linear across the reportable range.EDTA: e13a2 linear from MR0.93 to MR5.1 (max SD 0.22), e14a2 linear from MR0.99 to MR5.0 (max SD 0.26). PAXgene: e13a2 linear from MR0.78 to MR4.8 (max SD 0.13), e14a2 linear from MR0.93 to MR4.9 (max SD 0.31). Conclusion: The assay is linear across the reportable range of MR1.0 to MR4.5 for both tube types.
Kit Stability (Real-time & Freeze-Thaw)Stability for a specified shelf life and open-vial stability (freeze-thaw cycles and time after initial use). Acceptable performance of calibration curve, control results, and WHO secondary standards.Passing test results data for at least 16 months for 3 lots stored at two temperatures. Shelf life of 15 months at -30 to -15°C and -80 to -65°C. Open-vial stability of 3 freeze-thaw cycles and up to 2 months following initial use.
Reagent Stability (In Use)Mean of test samples ± 0.5 MR of the control. Differences between standard and extreme conditions < 0.25 log10.Acceptance criteria met for standard and extreme conditions. Largest difference observed between conditions was -0.063 (95% CI: -0.24 to 0.12) for MR4.5 sample. Conclusion: Reagent stable for up to 8 hours at 2-8°C before PCR assembly and up to 1.5 hours after PCR plate assembly after 2 freeze-thaw cycles.
Specimen Stability< 0.25 log10 difference compared to the mean BCR-ABL/ABL IS log10 ratio results at T0.All tested samples passed the acceptance criteria. PAXgene: Stable for at least 48 hours at 2-8°C and at least 28 days at -30 to -15°C. EDTA: Stable for at least 48 hours at 2-8°C.
Shipping StabilityKit performance after shipping stress meets equivalency acceptance criteria when compared to unstressed control kits. Maintain internal temperature below -65°C for 72 hours.Shipping containers maintained internal temperature below -65°C for 72 hours during stress test. Functional testing showed kit performance met equivalency criteria. Conclusion: Assay kit stable following shipment of up to 72 hours on dry ice.
PAXgene-EDTA Tube EquivalencyNo clinically significant bias at the clinical decision points (MR3.0, MR4.0, and MR4.5). (Implicit: Deming regression slope close to 1, y-intercept close to 0, high Pearson coefficient).Deming Slope: 1.02 (95% CI [1.00, 1.05]), Deming Y-Intercept: -0.115 (95% CI [-0.202, -0.042]), Deming Pearson Coefficient: 0.992. Predicted difference at MR3: -0.0433, MR4: -0.0192, MR4.5: -0.00721 (with 95% CI covering zero for MR4 and MR4.5). Conclusion: No clinically significant bias at clinical decision points was observed.

2. Sample sizes used for the test set and data provenance

  • Traceability to WHO IS: Tested over 20 days for each of 3 MRDx BCR-ABL Test kit lots. A different WHO primary panel (four levels A-D) was reconstituted daily, with each level tested multiple times (implicitly, across the 20 days). MolecularMD WHO BCR-ABL Reference Panel Secondary Standards were also tested.
  • Accuracy (Correlation to ddPCR):
    • Test Set Size: 217 samples total. 119 evaluable EDTA samples and 98 evaluable PAXgene samples.
    • Data Provenance: Samples were from CML patients. The document does not specify the country of origin but implies clinical samples. The study involved comparison to a reference method, suggesting an analytical, retrospective-like analysis of patient samples.
  • Limit of Blank (LoB): 30 BCR-ABL negative samples, tested in duplicate using 3 assay kit lots, totaling 180 replicates.
  • Limit of Detection (LoD): 6 CML patient specimens (3 e13a2, 3 e14a2) used to create serial dilutions. Sufficient RNA for each panel (5 levels) was made to allow for testing of 60 total replicates (20 replicates x 3 lots) per level. 1725 valid EDTA samples and 1595 valid PAXgene samples were tested.
  • Analytical Specificity (Interfering Substances): CML samples at 2 levels (MR3.0 and MR4.5) extracted from both EDTA and PAXgene tubes, spiked with test substances. No specific number of replicates or distinct samples given for this part.
  • Analytical Specificity (Primer Specificity): RNA from six CML patient samples (three e13a2, three e14a2) were used. Additionally, ABL2 IVT RNA and RNA from an e19a2 positive cell line were tested.
  • Analytical Specificity (Specimen Cross Contamination):
    • PAXgene: 30 replicates.
    • EDTA: 40 replicates.
  • Precision (Repeatability): Contrived patient samples (e13a2 and e14a2 at MR3.0, MR4.0, MR4.5). Each MR sample level for each transcript was tested in 17 replicates on individual MRDx BCR-ABL Test plates.
  • Precision (Reproducibility): Two series of samples (e13a2 and e14a2 CML patient samples diluted into non-diseased blood) at 5 levels each.
    • EDTA: 1800 total tests (60 replicates x 3 sites x 2 transcripts x 5 panel members). N for individual levels ranged from 149-152.
    • PAXgene: 900 total tests (90 measurements per panel member x 2 transcripts x 5 panel members, implying 90 measurements per level). N for individual levels ranged from 77-92.
  • Linear Range (Assay Range): Two dilution series (one e13a2, one e14a2) for each tube type. Each series had 10 BCR-ABL target levels, with 5 replicates at each level.
  • Kit Real-Time and Freeze-Thaw Stability: 6 kits from each of 3 lots for each time point were tested.
  • Reagent Preparation and Reaction Intermediate Stability (In Use): 2 kits were run for each condition (standard/extreme) on each of 2 days, with 2 sets of calibration curves, RNA controls, and samples (total n=8 per condition).
  • Specimen Stability:
    • PAXgene: Samples at three levels (MR3.0, MR4.0, MR4.5) tested in quadruplicate at 0, 24, 48, 60 hours (for 2-8°C storage) and at 0, 14, 28, 30 days (for -30 to -15°C storage).
    • EDTA: 17 CML patient specimens, tested in quadruplicate at 0, 24, 48, 60 hours.
  • Shipping Stability: Multiple types of containers and shipping conditions, varying load. Functional testing using samples across the assay range. Specific N not provided for samples.
  • PAXgene-EDTA Tube Equivalency: 100 CML patient paired blood samples.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable for analytical performance studies. The "ground truth" for these studies is established through defined reference methods (e.g., WHO International Genetic Reference Panel, ddPCR, precisely prepared dilutions/spikes, or established consensus sequences/protocols), not expert consensus in the human interpretation sense.

For the clinical studies (ENESTfreedom and ENESTop), the efficacy outcome (treatment-free remission, loss of MMR) is based on the quantitative results of the MRDx BCR-ABL Test itself per pre-defined thresholds, rather than expert interpretation of individual clinical cases as a ground truth.


4. Adjudication method for the test set

Not applicable. For the analytical studies, the ground truth is quantitative and objectively determined (e.g., standard concentrations, reference assay results, sequencing). For the clinical studies, outcomes are determined by the device's quantitative measurements against pre-defined clinical thresholds, not by human adjudication of conflicting interpretations.


5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is an in vitro diagnostic test (RT-qPCR assay) and not an AI-assisted diagnostic imaging or interpretation tool that involves human "readers." Therefore, no MRMC study, or assessment of human reader improvement with AI, was performed or is relevant to this device.


6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, the analytical performance studies explicitly demonstrate the standalone performance of the MolecularMD MRDx BCR-ABL Test. This device is an automated quantitative assay (RT-qPCR) followed by software analysis. The outputs are quantitative values (BCR-ABL/ABL % IS and MR values), and the performance characteristics (accuracy, precision, LoD, LoQ, linearity, etc.) are reported for the assay system itself, without human-in-the-loop performance influencing the primary measurements or calculations. While interpretation of results for clinical decision-making is done by clinicians, the device's measurement function is standalone.


7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Traceability to WHO IS: The ground truth was the 1st WHO International Genetic Reference Panel for quantitation of BCR-ABL translocation by RQ-PCR, which represents a harmonized international standard.
  • Accuracy (Correlation to ddPCR): The ground truth was derived from a validated BCR-ABL reverse transcription droplet digital PCR (ddPCR) assay, which served as the reference method for comparison.
  • Limit of Blank (LoB): "BCR-ABL negative samples" served as the negative ground truth.
  • Limit of Detection (LoD), Linear Range: Ground truth was established using serial dilutions of CML patient RNA and non-diseased subject blood RNA, allowing for precise control of target concentrations.
  • Analytical Specificity: Ground truth included precisely spiked interfering substances (endogenous and exogenous), known ABL2 IVT RNA, and e19a2 positive cell line RNA, and alternation of known high-positive and negative samples for cross-contamination.
  • Precision (Repeatability and Reproducibility): Ground truth was created using contrived patient samples (CML patient RNA and non-diseased subject RNA) with targeted MR levels.
  • PAXgene-EDTA Tube Equivalency: Ground truth was established by comparing results from paired CML patient blood samples processed via two different collection and extraction methods.
  • Clinical Studies (ENESTfreedom and ENESTop): The clinical validity was demonstrated against clinical outcomes data from two prospective clinical trials (ENESTfreedom and ENESTop). The ground truth for identifying candidates for treatment discontinuation and monitoring remission relied on sustained molecular response (MR4.5/MR4.0) as defined by the measured BCR-ABL levels.

8. The sample size for the training set

The document does not explicitly specify a "training set" for the development of the device's algorithm in the way that would typically be described for a machine learning model. This is an RT-qPCR assay, not a machine learning-based AI device.

However, if "training set" is broadly interpreted as data used during the development and optimization phases:

  • The quantitative methodology is based on established RT-qPCR principles and standard curve generation using calibrators.
  • The "conversion factor" to the International Scale is derived from harmonization studies with the WHO International Genetic Reference Panel.
  • The "MRDx BCR-ABL Test Software" uses established equations and validated run/sample acceptance criteria, which would have been developed and refined using internal data.
  • The specific number of samples or data points used for these internal development and optimization processes is not provided.

9. How the ground truth for the training set was established

As noted above, there isn't a "training set" in the machine-learning sense for this device. The ground truth for analytical performance studies, which serve to validate the device's performance, is established through:

  • International Reference Standards: The 1st WHO International Genetic Reference Panel for BCR-ABL quantitation.
  • Reference Methods: A validated droplet digital PCR (ddPCR) assay.
  • Controlled Experiments: Creation of precisely defined mock samples or panels with known concentrations or characteristics (e.g., serial dilutions of CML patient RNA, non-diseased RNA, spiked samples, targeted MR levels in contrived samples).
  • Clinical Trial Data: Long-term clinical follow-up in the ENESTfreedom and ENESTop studies, where the MRDx BCR-ABL Test itself determined the molecular response status that defined the clinical ground truth for treatment discontinuation and remission monitoring.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 22, 2017

MolecularMD Corporation Kevin Hawkins Director, Quality and Regulatory Affairs 1341 SW Custer Drive Portland, OR 97219

Re: K173492

Trade/Device Name: MolecularMD MRDx BCR-ABL Test Regulation Number: 21 CFR 866.6060 Regulation Name: BCR-ABL Quantitation Test Regulatory Class: Class II Product Code: OYX Dated: November 9, 2017 Received: November 13, 2017

Dear Kevin Hawkins:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Reena Philip -S

Reena Philip, Ph.D. Director Division of Molecular Genetics and Pathology Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K173492

Device Name MRDx® BCR-ABL Test

Indications for Use (Describe)

The MolecularMD MRDx BCR-ABL Test is an in vitro diagnostic test for the quantitative detection of BCR-ABL 1 transcripts (e13a2/b2a2 and/or e14a2/63a2) and the ABL1 endogenous control mRNA in peripheral blood specimens from patients previously diagnosed with t(9:22) positive chronic myeloid leukemia (CML). The ratio of BCR-ABL1 is calculated and reported on the WHO International Scale. The test utilizes quantitative, real-time reverse transcription polymerase chain reaction performed on the Applied Biosystems 7500 Fast Dx instrument.

The MolecularMD MRDx BCR-ABL Test is intended to measure BCR-ABL mRNA transcript levels in patients diagnosed with t(9;22) positive CML during monitoring of treatment with Tyrosine Kinase Inhibitors (TKIs).

The device is also intended to be used in the serial monitoring for BCR-ABL mRNA transcript levels as an aid in identifying CML patients in the chronic phase being treated with nilotinib who may be candidates for treatment discontinuation and for monitoring of treatment-free remission.

The test does not differentiate between e13a2 or e14a2 fusion transcripts and does not monitor other rare fusion franscripts resulting from t(9:22). The test is not intended for the diagnosis of CML.

Type of Use ( Select one or both, as applicable )
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for MolecularMD. The logo consists of a red medical symbol on the left, followed by the text "MolecularMD" in gray and red. The "Molecular" part of the text is in gray, while the "MD" part is in red.

510(k) SUMMARY

l. COMPANY AND CONTACT INFORMATION

Company Name:MolecularMD Corporation
Address:1341 SW Custer DrivePortland, OR 97219
Telephone:503.459.4974
Fax:503.459.4976
Contact Person:Kevin HawkinsDirector, Quality and Regulatory Affairs
Date of Summary:20 December 2017

II. DEVICE IDENTIFICATION

Trade (Proprietary) Name:MolecularMD MRDx® BCR-ABL Test
Common (Usual) Name:BCR-ABL RT-qPCR Test
Classification Name:BCR-ABL Quantitation Test
Product Code:OYX
Regulation Number:866.6060
Regulatory Classification:Class II
Panel:88 – Pathology

III. PREDICATE DEVICE

QuantideX® qPCR BCR-ABL IS Kit

{4}------------------------------------------------

IV. DEVICE DESCRIPTION

A. Principle of the Procedure - Test Methodology

The MolecularMD MRDx® BCR-ABL Test is a quantitative, real-time polymerase chain reaction test that provides quantitation of BCR-ABL1 (hereafter BCR-ABL) transcript e13a2/b2a2 or e14a2/b3a2 and ABL1 (hereafter ABL) transcript levels in RNA extracted from peripheral blood samples collected from CML patients. Peripheral blood is collected in either EDTA or PAXgene Blood RNA Tubes. Each collection tube type requires a specified RNA extraction method and RNA input amount. The detailed extraction protocol and RNA quantity requirements are provided in the device package insert and instructions for use. Total RNA is extracted from peripheral blood and serves as the template for RT-qPCR. The test is performed using a onestep RT-qPCR protocol wherein the reverse transcription and quantitative, real-time PCR reactions are performed in the same well. BCR-ABL and ABL amplicons are generated and detected in real-time using TaqMan® MGB probes. For RNA extracted from EDTA tubes, a total of 4 µg of RNA is required for each patient sample: 1 µg of extracted RNA is used in the MRDx BCR-ABL Test in each of 4 wells, 2 wells for BCR-ABL quantitation and 2 wells for ABL quantitation. For RNA extracted from PAXgene tubes, a total of 8 µg of RNA is required for each patient sample: 2 µg of extracted RNA is used in the MRDx BCR-ABL Test in each of 4 wells, 2 wells for BCR-ABL quantitation and 2 wells for ABL quantitation.

The MolecularMD MRDx BCR-ABL Test is performed on the Applied Biosystems 7500 Fast Dx Real-Time PCR Instrument. The instrument integrates a thermal cycler, a fluorometer and application specific software. The instrument houses the thermal cycler and the fluorometer, while the application software is installed on a computer that is attached to the instrument. The ABI 7500 Fast Dx instrument software v1.4.1 calculates the BCR-ABL and ABL copy numbers using a standard curve generated with calibrators on each individual plate.

Quantitation is achieved using RNA calibration standards and linear regression analysis provided by the ABI 7500 Fast Dx PCR Instrument Software. BCR-ABL transcript levels are measured in relation to the ABL transcript as an endogenous reference. The data are exported as a CSV file for further analysis in the MRDx® BCR-ABL Test Software. The BCR-ABL/ABL ratio is calculated and converted to the International Scale by the MRDx BCR-ABL Test Software.

The MRDx BCR-ABL Test Software is used to analyze all test results. This software, provided with the MRDx BCR-ABL Test, is used to calculate the BCR-ABL/ABL % IS and MR value for patient sample using the conversion factor for the MRDx BCR-ABL Test after validating each MRDx BCR-ABL Test result against the run acceptance criteria and sample acceptance criteria.

Calibrators are analyzed for both BCR-ABL and ABL in every run simultaneously with patient samples. RNA controls provided at MMR (MR3.0) and MR4.5 allow for on-plate verification of

{5}------------------------------------------------

test accuracy and reproducibility over this critical transcript range. The integrated conversion factor provides test results on the International Scale (IS) harmonized to the WHO.

Interpretation of Results

Test results are reported in both IS scale and the corresponding molecular response (MR) value. The BCR-ABL/ABL ratio percent IS result and MR value is calculated by the MRDx BCR-ABL Test Software using the following equations.

$$\frac{BCR - ABL}{ABL} ,% ,% , IS = \left(\frac{Mean,BCR - ABL,,Copy,Number}{Mean,,ABL,,Copy,Number}\right) \ast 100 ,*,Conversion,,Factor$$

$$MRx.x = \log_{10}\left(\frac{100}{9% IS}\right) = \log_{10}(100) - \log_{10}(9% IS) = 2 - \log_{10}(9% IS)$$

MR Values with the corresponding IS values are show in the table below:

MRIS (%)
0.0100
0.532
1.010
1.53.2
2.01.0
2.50.32
3.00.10
3.50.032
4.00.010
4.50.0032
5.00.0010

The assay reporting is shown in the table below. The assay will only report quantified results for BCR-ABL levels within the range of MR4.5 to MR1.0.

Collection TubeBCR-ABL% (MR)MRDx Result
EDTA<1.0 (>10% IS)BCR-ABL detected, not quantifiable
EDTA1.0-4.5BCR-ABL detected
EDTA4.5-5.3BCR-ABL detected, not quantifiable
EDTA>5.3 (<0.00050% IS)BCR-ABL not detected
PAXgene<1.0 (>10% IS)BCR-ABL detected, not quantifiable
PAXgene1.0-4.5BCR-ABL detected
PAXgene4.5-4.7BCR-ABL detected, not quantifiable
PAXgene>4.7 (<0.0020% IS)BCR-ABL not detected

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for MolecularMD. The logo consists of a red medical symbol on the left, followed by the text "MolecularMD" in gray and red. The "Molecular" part of the text is in gray, while the "MD" part is in red.

B. Specimen Type

Two sample collection tube types and RNA extraction methods are specified in the package insert. The first, K2EDTA blood tubes (hereafter referred to simply as EDTA), are processed using the Maxwell CSC RNA Blood Kit (Promega Cat. No. AS1410) and semi-automated Maxwell CSC Instrument (Promega Cat. No. AS4000 or AS6000). The instructions for use for the MRDx BCR-ABL Test provide specific instructions for use of 5 mL of blood versus the 2.5 mL specified in the Maxwell manual. The increased sample volume ensures the yield of RNA required is available for testing. A total of 4 µg of RNA is required for each patient sample. One µg of extracted RNA is used in the MRDx BCR-ABL Test in each of 4 wells for BCR-ABL quantitation and 2 wells for ABL quantitation.

In addition to testing EDTA blood samples, the test may be used with blood collected into PAXgene Blood RNA Tubes (PreAnalytiX Cat. No. 762165) and a manual extraction procedure using the PAXgene Blood RNA Kit (QIAGEN Cat. No. 762164). The extraction procedure is performed per a MolecularMD-specified procedure utilizing four PAXgene Blood RNA Tubes to yield the required quantity of RNA. A total of 8 µg of RNA is required for each patient sample. Two µg of extracted RNA is used in the MRDx BCR-ABL Test in each of 4 wells, 2 wells for BCR-ABL quantitation and 2 wells for ABL quantitation.

C. Kit Contents

The MolecularMD MRDx BCR-ABL Test includes the following components for testing RNA extracted from samples:

Reagents:

  • MRDx Master Mix (2X) (3 vials x 850 µL/vial)
  • MRDx BCR-ABL Primer/Probe, concentrated (1 vial x 40 µL/vial)
  • MRDx ABL Primer/Probe, concentrated (1 vial x 40 μL/vial)
  • Nuclease-Free Water (2 vials x 1.8 mL/vial) ●

IVT RNA Calibrators:

A set of calibrators containing in vitro transcribed BCR-ABL and ABL targets representing a dynamic range of 105 is provided as part of the test kit. Two calibration curves (one for BCR-ABL and one for ABL) are analyzed with every test run in order to allow for quantitation of the two transcripts in patient samples. The BCR-ABL and ABL levels for each calibrator are listed in the table below.

{7}------------------------------------------------

LabelBCR-ABLABLVolume
H600,000 copies/10 µL80 µL/vial
A300,000 copies/10 µL300,000 copies/10 µL80 µL/vial
B30,000 copies/10 µL30,000 copies/10 µL80 µL/vial
C3,000 copies/10 µL3,000 copies/10 µL80 µL/vial
D300 copies/10 µL300 copies/10 µL80 µL/vial
E30 copies/10 µL30 copies/10 µL80 µL/vial
F10 copies/10 µL10 copies/10 µL80 µL/vial
G3 copies/10 µL80 µL/vial

RNA Controls:

MMR (MR3.0) and MR4.5 RNA controls are formulated from a mixture of BCR-ABL negative HL-60 (ATCC® CCL-240™) and BCR-ABL e14a2 positive K-562 (ATCC® CCL-243™) RNA in a background mix of carrier RNA and stabilizers to allow for long-term stability. The MMR RNA Control is targeted to 0.10% IS BCR-ABL/ABL. The MR4.5 RNA Control is targeted to 0.0032% IS BCR-ABL/ABL.

  • BCR-ABL RNA Control MMR (10X, 1 vial x 14 µL/vial) ●
  • . BCR-ABL RNA Control - MR4.5 (10X, 1 vial x 14 µL/vial)

D. Hardware and Software

The MolecularMD MRDx BCR-ABL Test is performed on the Applied Biosystems 7500 Fast Dx Real-Time PCR Instrument. The instrument integrates a thermal cycler, a fluorometer and application specific software. The instrument houses the thermal cycler and the fluorometer, while the application software is installed on a computer that is attached to the instrument. The ABI 7500 Fast Dx instrument software v1.4.1 calculates the BCR-ABL and ABL copy numbers using a standard curve generated with calibrators on each individual plate. The data are exported as a CSV file for further analysis in the MRDx BCR-ABL Test Software.

The MRDx BCR-ABL Test Software is used to analyze all test results. This software, provided with the MRDx BCR-ABL Test, is used to calculate the BCR-ABL/ABL % IS for patient sample results using the conversion factor for the MRDx BCR-ABL Test in addition to validating each MRDx BCR-ABL Test result against the run acceptance criteria and sample acceptance criteria.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for MolecularMD. The logo consists of a red medical symbol on the left, followed by the text "MolecularMD" in gray and red. The "Molecular" part of the text is in gray, while the "MD" part is in red.

V. INDICATIONS FOR USE

The MolecularMD MRDx BCR-ABL Test is an in vitro diagnostic test for the quantitative detection of BCR-ABL1 transcripts (e13a2/b2a2 and/or e14a2/b3a2) and the ABL1 endogenous control mRNA in peripheral blood specimens from patients previously diagnosed with t(9;22) positive chronic myeloid leukemia (CML). The ratio of BCR-ABL1 to ABL1 is calculated and reported on the WHO International Scale. The test utilizes quantitative, real-time reverse transcription polymerase chain reaction performed on the Applied Biosystems 7500 Fast Dx instrument.

The MolecularMD MRDx BCR-ABL Test is intended to measure BCR-ABL mRNA transcript levels in patients diagnosed with t(9;22) positive CML during monitoring of treatment with Tyrosine Kinase Inhibitors (TKIs).

The device is also intended to be used in the serial monitoring for BCR-ABL mRNA transcript levels as an aid in identifying CML patients in the chronic phase being treated with nilotinib who may be candidates for treatment discontinuation and for monitoring of treatment-free remission.

The test does not differentiate between e13a2 or e14a2 fusion transcripts and does not monitor other rare fusion transcripts resulting from t(9;22). The test is not intended for the diagnosis of CML.

{9}------------------------------------------------

VI. SUMMARY COMPARISON OF TECHNOLGICAL CHARACTERISTICS

Similarities
TechnologicalCharacteristicCandidate DeviceMRDx® BCR-ABL TestPredicate DeviceQuantideX® qPCRBCR-ABL IS Kit
InstrumentApplied Biosystems 7500 Fast Dx Real TimePCR InstrumentSame
MeasurandBCR-ABL1 fusion transcripts (e13a2/b2a2and / or e14a2/b3a2) and the ABL1endogenous control mRNASame
Measurement TypeQuantitativeSame
Principle of AssayReverse transcription, quantitative,polymerase chain reaction (qPCR) basednucleic acid amplificationSame
Traceability/Standard1st WHO International Genetic ReferencePanel for quantitation of BCR-ABLtranslocation by RQ-PCRSame
Report UnitsBoth % IS and Molecular Response (MR)Same
Test LimitationThe test does not differentiate betweene13a2 or e14a2 fusion transcripts anddoes not monitor other rare fusiontranscripts resulting from t(9;22). Thetest is not intended for the diagnosis ofCML.Same
Differences
TechnologicalCharacteristicCandidate DeviceMRDx® BCR-ABL TestPredicate DeviceQuantideX® qPCR BCR-ABL IS Kit
Intended UseThe MolecularMD MRDx BCR-ABL Test is anin vitro diagnostic test for the quantitativedetection of BCR-ABL1 transcripts(e13a2/b2a2 and/or e14a2/b3a2) and theABL1 endogenous control mRNA inperipheral blood specimens from patientspreviously diagnosed with t(9;22) positivechronic myeloid leukemia (CML). The ratioof BCR-ABL1 to ABL1 is calculated andreported on the WHO International Scale.The QuantideX® qPCR BCR-ABL IS Kit is an in vitronucleic acid amplificationtest for the quantitation ofBCR-ABL1 and ABL1transcripts in total RNA fromwhole blood of diagnosedt(9;22) positive ChronicMyeloid Leukemia (CML)patients expressing BCR-
Differences
TechnologicalCharacteristicCandidate DeviceMRDx® BCR-ABL TestPredicate DeviceQuantideX® qPCR BCR-ABL ISKit
The test utilizes quantitative, real-timereverse transcription polymerase chainreaction performed on the AppliedBiosystems 7500 Fast Dx instrument.The MolecularMD MRDx BCR-ABL Test isintended to measure BCR-ABL mRNAtranscript levels in patients diagnosed witht(9;22) positive CML during monitoring oftreatment with Tyrosine Kinase Inhibitors(TKIs). The device is also intended to beused in the serial monitoring for BCR-ABLmRNA transcript levels as an aid inidentifying CML patients in the chronicphase being treated with nilotinib who maybe candidates for treatmentdiscontinuation and for monitoring oftreatment-free remission. The test does notdifferentiate between e13a2 or e14a2fusion transcripts and does not monitorother rare fusion transcripts resulting fromt(9;22). The test is not intended for thediagnosis of CML.ABL1 fusion transcripts typee13a2 and/or e14a2. TheQuantideX qPCR BCR-ABL ISKit is a reverse transcription-quantitative PCR performedon the Applied Biosystems7500 Fast Dx Real-Time PCRInstrument and is intendedto measure BCR-ABL1 toABL1, expressed as a logmolecular reduction (MRvalue) from a baseline of100% on the InternationalScale, in t(9;22) positive CMLpatients during monitoringof treatment with TyrosineKinase Inhibitors (TKIs).
QualityControlsRNA MMR (MR3.0)RNA MR4.5Nuclease-Free Water - NTCRNA High (MR 1.5)RNA Low (MR 3.5)ABL armored RNARNA Negative
Calibrators8 vials total, representing 7 concentrationseach of BCR-ABL and ABL transcripts.Four levels formulated toMR1.0, 2.0, 3.0, 4.0.
SpecimenTypeRNA from whole blood (PAXgene BloodRNA Tubes or EDTA)RNA from whole blood(EDTA)
RNA InputAmount2 µg PAXgene tubes and 1µg EDTA tubesRNA input range of 1 to 5µg

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image contains the logo for Molecular MD. The logo consists of a red medical symbol to the left of the text "MolecularMD". The text is in a gray sans-serif font, with the "MD" portion in red.

{11}------------------------------------------------

PERFORMANCE CHARACTERISTICS VII.

Pre-Clinical Analytical Studies A.

1. Traceability to the International Scale

The assay calibrators are traceable to the First (1*) WHO International Genetic Reference Panel based on the method described in the 1st WHO International Genetic Reference Panel for quantitation of BCR-ABL translocation by RQ-PCR, Instruction for Use. On 5 different days, a different WHO primary panel comprised of four levels (A-D) of freeze-dried materials was reconstituted and extracted following the assay protocol. The reconstituted/extracted WHO reference samples (200 ng RNA input/well) and the MolecularMD WHO BCR-ABL Reference Panel Secondary Standards (1 µg RNA input/well) were tested using the MRDx BCR-ABL Test over 20 days for each of 3 MRDx BCR-ABL Test kit lots.

Deming Regression analyses were performed comparing the expected and observed MR values for each lot. The conversion factor was calculated for each lot individually and was 1.1 for all three lots (A, B and C), showing the consistency of conversion factor across multiple lots. Data from all three lots showed high traceability to WHO reference standards supporting the conclusion that the assay calibrators to the WHO reference panel has been established. See the Deming Regression plots and statistics below.

Image /page/11/Figure/6 description: The image contains three scatter plots, labeled Lot A, Lot B, and Lot C. Each plot shows the relationship between MR values determined by MRDx on the y-axis and published MR values for the WHO Primary Reference on the x-axis. The plots show a positive correlation between the two sets of MR values, with data points clustered around a green line. The x and y axis range from -2 to 1.

Deming Regression for WHO Standards by Lot

Regression Statistics for Each Lot
MetricsLot ALot BLot CCombined
Slope1.00.990.980.99
95% Lower Cl (slope)0.990.970.960.98
95% Upper Cl (slope)1.01.01.01.0
Y-Intercept-0.048-0.034-0.057-0.046
95% Lower Cl(intercept)-0.065-0.057-0.082-0.059
95% Upper Cl(intercept)-0.030-0.010-0.033-0.034
Correlation1.01.01.01.0

{12}------------------------------------------------

2. Correlation to a Reference Method (Accuracy)

The accuracy of detecting BCR-ABL transcript levels by the MRDx BCR-ABL Test was further evaluated by comparing to a second reference method, a validated BCR-ABL reverse transcription droplet digital PCR (ddPCR) assay. The study analyzed a total of 217 samples of blood from CML patients ranging from 10% BCR-ABL/ABL IS (MR1.0) to 0.001% BCR-ABL/ABL IS (MR5.0) drawn into PAXgene Blood RNA Tubes and extracted using the PAXgene Blood RNA Kit as well as samples drawn into EDTA blood collection tubes and extracted using the Maxwell CSC RNA Blood Purification Kit with the Maxwell CSC instrument. A total of 119 evaluable EDTA samples and 98 evaluable PAXgene samples were tested with both tests.

A Deming regression was applied and 95% confidence intervals of the slope and y-intercept were calculated. The predicted systematic difference between the two methods was calculated at the clinical decision points for treatment-free remission monitoring (MR3.0, MR4.0, and MR4.5) with 95% confidence intervals. High concordance between MRDx BCR-ABL Test results and the reference ddPCR test results were observed for both EDTA and PAXgene tube types (see below), supporting the detection accuracy of the device.

Image /page/12/Figure/4 description: The image contains two scatter plots comparing MRDx and ddPCR measurements. The left plot shows data for EDTA with 1ug RNA, with a Deming RegressionFit equation of 0.08 + 0.90 * ddPCR (MR) and a Pearson's r of 0.984. The right plot shows data for PAXgene with 2ug RNA, with a Deming RegressionFit equation of -0.02 + 0.98 * ddPCR (MR) and a Pearson's r of 0.985. Both plots include an identity line for reference.

Deming Regression Plots (MRDx vs. ddPCR)

{13}------------------------------------------------

Blood TubeTypeSampleSizeDeming SlopeDemingY-InterceptPearsonCoefficient
EDTA1190.9695% CI [0.94, 0.99]0.07795% CI [0.0027, 0.17]0.984
PAXgene980.9895% CI [0.95, 1.0]-0.02195% CI [-0.11, 0.056]0.985

Deming Regression Statistics (MRDx vs. ddPCR)

Predicted Difference at Clinical Decision Points

Blood TubeTypePredicted Difference MR3(MMR)Predicted DifferenceMR4.0Predicted DifferenceMR4.5
EDTA-0.03195% CI [-0.053, -0.0049]-0.06795% CI [-0.10, -0.031]-0.08595% CI [-0.13, -0.039]
PAXgene-0.08895% CI [-0.11, -0.062]-0.1195% CI [-0.15, -0.071]-0.1295% CI [-0.17, -0.071]

3. Detection Capability

Limit of Blank (LoB):

The LoB was determined by testing 30 BCR-ABL negative samples were tested in duplicate using 3 assay kit lots for a total of 180 replicates. Out of the 180 measurements, 177 had no measurable BCR-ABL values. Three (1.7%) had measurements well below the LoD of the test which were reported as undetected.

Limit of Detection (LoD):

The LoD was determined using 6 CML patient specimens (3 specimens each for e13a2 and e14a2 transcripts). RNA of each sample was extracted with both PAXgene Blood RNA Tubes and EDTA blood collection tubes and serial diluted into RNA extracted from non-diseased subject blood to create five level panels. Sufficient RNA for each panel was made to allow for the testing of 60 total replicates (20 replicates x 3 lots) of each level in the panel. There were 1725 and 1595 valid EDTA and PAXgene samples tested respectively.

The LoD was calculated using the parametric analysis as described in CLSI EP17-A2. Using EDTA blood collection tubes, the LoD values were estimated to be 0.00019% IS (MR5.7) for e13a2 and 0.00029% IS (MR5.5) for e14a2. Using PAXgene Blood RNA Tubes, the LoD values were estimated to be 0.00039% IS (MR5.4) for e13a2 and 0.00029% IS (MR 5.5) for e14a2. The final LoD was determined to be 0.00029% IS (MR5.5) for EDTA blood collection tubes and 0.00039% IS (MR5.4) for PAXgene Blood RNA Tubes for both transcripts.

Limit of Quantitation (LoQ):

The same LoD study samples were used to establish LoQ. The average and standard deviation for the BCR-ABL/ABL % IS values and MR values were calculated for the replicates for each level

{14}------------------------------------------------

for all extraction methods, BCR-ABL transcripts, and all 3 lots. The log10 bias between the nominal values and the MRDx BCR-ABL Test for each level was calculated by subtracting the average measured BCR-ABL/ABL % IS log10 for each level from the BCR-ABL/ABL % IS log10 nominal value. The Total Error (TE) was calculated using the following equation: Total Error = |Bias| + 2 x SD

The results were evaluated against the acceptance criteria of Total Error ≤ 0.5 log10. The LoQ for EDTA blood collection tubes was 0.0016% IS (MR4.8). The LoQ for PAXgene Blood RNA Tubes was 0.0025% IS (MR4.6).

The MRDx BCR-ABL Test Software limits the LoQ and quantitated results to 0.0032% IS (MR4.5) for both blood tube types.

4. Analytical Specificity - Interfering Substances

For endogenous blood substance interference testing, CML samples at 2 levels (MR3.0 and MR4.5) extracted from both EDTA and PAXgene collection tubes were spiked with all the following test substances to the concentrations noted:

  • Hemoglobin 2 mg/mL ●
  • Bilirubin 342 µmol/L
  • . Triglycerides 288 mg/dL

When compared to test results from samples spiked with matrix control (i.e., cleared serum and water), no clinically significant difference was observed from the samples spiked with the potential interference substances, although reduced RNA yield during extraction was observed in some samples. The largest difference between the endogenous substance spike and the matrix control spike for PAXgene Blood RNA Tubes was -0.080 log10 (95% Cl: -0.32 to 0.16) at MR4.5. The largest difference between the endogenous substance spiked group and the matrix control group for EDTA blood collection tubes was 0.060 log10 (95% Cl: -0.042 to 0.16) at MR4.5.

For exogenous interference testing, extraction kit reagents used in the RNA extraction processes for PAXgene Blood RNA Tubes and EDTA blood collection tubes were examined to determine if any interference from the blood tubes and RNA extraction processes were observed with the MRDx BCR-ABL Test. Ten replicates of the MR3.0 and MR4.5 RNA controls were spiked and tested together with ten replicates of the MR3.0 and MR4.5 RNA Controls spiked with water as a matrix control sample. No clinically significant difference from the matrix control samples was observed. The largest difference between the exogenous substance spiked group and the matrix control group for PAXgene Blood RNA Tubes was 0.040 logio (95% Cl: -0.010 to 0.090) at MR3.0. The largest difference between the exogenous substance spiked group and the matrix control group for EDTA blood collection tubes was 0.010 log10 (95% CI: -0.022 to 0.22) at MR4.5.

{15}------------------------------------------------

The data support the conclusion that none of the potential interferents evaluated had any significant interference with the assay.

5. Analytical Specificity - Primer Specificity

RNA from six CML patient samples (three with the e13a2 transcript and three with the e14a2 transcript) was extracted and used as template for the MRDx BCR-ABL Test. The resulting PCR products for both BCR-ABL and ABL were sub-cloned into a cloning vector and ten colonies were sequenced. The sequencing results were used to create a consensus sequence to compare to the established Genbank sequence for ABL and both BCR-ABL transcripts. The results matched the Genbank sequence with agreement across all bases demonstrating that the MRDx primers and probes amplified the intended targets for all samples tested.

Cross reactivity was further assessed by testing ABL2 in vitro transcribed RNA and RNA from an e19a2 positive cell line as template. No cross reactivity was observed with the ABL2 IVT RNA. As expected, cross reactivity was present for the e19a2 BCR-ABL transcript as it contains the binding sites for the primers and probe in BCR exon 13 and ABL exon 2. The cross reactivity of the MRDx BCR-ABL Test with the e19a2 BCR-ABL transcript was mitigated by a precaution statement in the labeling that the test is only indicated for use with CML patients diagnosed with the e13a2 and/or e14a2 BCR-ABL transcript and should not be used with patients confirmed to express atypical transcripts.

6. Analytical Specificity - Specimen Cross Contamination (Carryover)

The carryover contamination study was performed by testing high positive samples (near MR1.0) alternately juxtaposed with negative samples in both extraction rack setup as well as MRDx plate setup for both PAXgene Blood RNA Tubes and EDTA blood collection tubes.

The incidence of cross-contamination with PAXgene Blood RNA tubes of samples was determined using 30 replicates across 5 extraction racks and 5 MRDx BCR-ABL Test plates. The incidence of cross-contamination with EDTA blood collection tubes was determined using the Maxwell CSC RNA Blood Kit with 40 replicates across 5 extraction racks and 5 MRDx BCR-ABL Test plates.

No negative samples (0% IS) for either extraction method had a BCR-ABL/ABL % IS value that was detectable above the LoD of the assay. The data support the conclusion that the test generates no significant carryover between wells on the plate.

7. Precision - Repeatability

Contrived patient samples targeting MR3.0, MR4.0, and MR4.5 for e13a2 and e14a2 BCR-ABL transcripts were created using CML patient RNA and non-diseased subject RNA extracted from both PAXgene Blood RNA tubes and EDTA blood collection tubes. The MR3.0, MR4.0, and MR4.5 RNA samples were each tested in 17 replicates on individual MRDx BCR-ABL Test plates. See the repeatability results in the table below.

510(k) Summary

{16}------------------------------------------------

For EDTA Blood RNA, all levels for both transcripts passed the acceptance criteria of ≤ 0.25 SD logio. The highest standard deviation was observed for the e14a2 BCR-ABL transcript at MR4.5 (SD = 0.087).

For PAXgene Blood RNA, all levels for both transcripts passed the acceptance criteria (SD ≤ 0.25 SD log10). The highest standard deviation was observed for the e14a2 BCR-ABL transcript at MR4.5 (SD=0.10).

Repeatability Results
MeanSD%CVMeanSD%CV
TubeTypeTranscriptLevelSampleSize (N)BCR-ABL%ISBCR-ABL%ISBCR-ABL%ISMRMRMR
PAXgenee13a2 MR3170.140.008762.90.0511.8
PAXgenee13a2 MR4170.0170.0026153.80.071.9
PAXgenee13a2 MR4.5170.00620.0017274.20.0992.4
PAXgenee14a2 MR3170.110.00928.13.00.0511.7
PAXgenee14a2 MR4170.0210.0035173.70.0782.1
PAXgenee14a2 MR4.5170.00510.0012244.30.102.4
EDTAe13a2 MR3170.120.00695.92.90.0511.7
EDTAe13a2 MR4170.0110.0012114.00.061.5
EDTAe13a2 MR4.5170.00340.00055164.50.0791.8
EDTAe14a2 MR3170.120.0129.72.90.0511.7
EDTAe14a2 MR4170.0130.00129.23.90.051.3
EDTAe14a2 MR4.5170.00380.00071194.40.0872

8. Precision - Reproducibility

Two series of samples were created with blood drawn into EDTA tubes by separately diluting pooled e13a2 and pooled e14a2 positive CML patient samples with high BCR-ABL/ABL ratios into blood from non-diseased subjects. Five levels were created with targeted BCR-ABL % IS ratios from 10 % IS to 0.0032 % IS. Three testing sites with two 7500 Dx instruments and two operators each were used. Each day at each site, each operator ran two plates, one on each instrument. Each plate had one replicate of the e13a2 5 member panel and one replicate of the e14a2 five member panel. For each of three lots, five runs were performed. Fifteen measurements were made per instrument, per operator. A total of 60 replicates (15 measurements per instrument per site x 2 instruments x 2 operators=60) were run at each site for each BCR-ABL transcript type. The total number of tests run across transcript types and panel members was 1800 (60 replicates x 3 sites x 2 transcripts x 5 panel members = 1800).

{17}------------------------------------------------

LevelNMeanMRValueBetween-SiteBetween-OperatorBetween-Day(Run)
SD%CVSD%CVSD%CV
e14a2 L11520.990.0272.70.00770.780.00460.46
e14a2 L21522.0000.0221.100
e14a2 L31523.1000.0150.5000
e14a2 L41524.10.0220.54000.0180.45
e14a2 L51494.50.0230.510000
e13a2 L11511.2000.0181.500
e13a2 L21522.3000.0241.10.00860.38
e13a2 L31513.00.0180.620.00700.2300
e13a2 L41523.9000.0190.4800
e13a2 L51524.40.0220.500000
LevelNMeanMRValueBetween-InstrumentsBetween-LotTotal
SD%CVSD%CVSD%CV
e14a2 L11520.990.0181.8000.0787.8
e14a2 L21522.0000.0110.540.0512.5
e14a2 L31523.1000.00650.210.0602.0
e14a2 L41524.1000.00610.150.112.7
e14a2 L51494.500000.133.0
e13a2 L11511.20.0100.820.0312.50.0604.9
e13a2 L21522.30.00230.100.0331.50.0642.8
e13a2 L31513.0000.0100.350.0602.0
e13a2 L41523.90.0190.480.0120.310.0962.5

Reproducibility Results – EDTA

Within-Laboratory Precision – EDTA

Within-Laboratory Precision
BCR-ABL TranscriptLevelSiteNMean MRSD% CV
e14a21A501.00.0515.0
e14a21B550.980.0353.6
e14a21C470.970.1111
e14a22A502.00.0542.6
e14a22B552.00.0452.2
e14a22C472.00.0442.2
e14a23A503.10.0591.9
e14a23B553.10.0581.9
e14a23C473.10.0571.9
e14a24A504.10.0922.3
e14a24B554.10.0982.4

{18}------------------------------------------------

Within-Laboratory Precision
BCR-ABL TranscriptLevelSiteNMean MRSD% CV
e14a24C474.00.123.1
e14a25A484.50.122.6
e14a25B554.50.143.0
e14a25C464.50.132.9
e13a21A501.20.0544.3
e13a21B551.20.0574.7
e13a21C461.20.0544.4
e13a22A502.30.0662.9
e13a22B552.30.0572.5
e13a22C472.30.0532.3
e13a23A493.00.0561.9
e13a23B553.00.0551.8
e13a23C473.00.0561.9
e13a24A503.90.0892.3
e13a24B553.90.0922.4
e13a24C473.90.0972.5
e13a25A504.40.112.5
e13a25B554.40.133.0
e13a25C474.40.132.9

For PAXgene Blood RNA tubes, two series of samples were created with blood drawn into PAXgene RNA Blood tubes by separately diluting pooled e13a2 and pooled e14a2 positive CML patient blood samples with high BCR-ABL/ABL ratios into blood from non-diseased subjects. Five levels were created with targeted BCR-ABL/ABL % IS ratios ranging from 10% IS to 0.0032% IS. One testing site used three operators and each operator used one 7500 Fast DX instrument exclusively. Two sample panels were extracted with three lots of extraction reagents and evaluated using three lots of MRDx BCR-ABL Test reagents over five non-consecutive days for a total of 15 replicates. A total of 90 measurements were made for each panel member for each BCR-ABL transcript type and RNA input amount. A total of 900 tests were completed for variables including operator, instrument, day, and lot. The reproducibility study are provided in the following table.

{19}------------------------------------------------

LevelNMeanMRBetween Operator(Instrument)Between Day(Run)Between LotTotal (Within-Laboratory)
ValueSD%CVSD%CVSD%CVSD%CV
e14a2 L1860.830.00740.880.0445.30.0344.10.09511
e14a2 L2862.00O0.0381.90.0351.80.0673.4
e14a2 L3862.90O0.0632.20.0441.50.0903.1
e14a2 L480ਤੋਂ ਰੋ0O0.0822.10.0581.50.143.6
e14a2 L5774.60.0781.7000.0691.50.245.3
e13a2 L1920.800O0.0324.00.0212.70.0577.1
e13a2 L2921.90.0140.780.0452.4O00.105.5
e13a2 L3922.9000.0602.00.0170.590.0742.5
e13a2 L4ਰੇ 14.0O00.102.5OO0.143.5
e13a2 L5874.50.0440.980.0701.6OO0.194.3

Reproducibility Results - PAXgene

9. Linear Range – Assay Range

Linearity was estimated by testing two dilution series (one for e13a2 and one for e14a2 transcript) for each blood collection tube type (EDTA and PAXgene). Each dilution series had 10 BCR-ABL target levels and 5 replicates at each level to evaluate the assay linearity across the target range from 15% IS to 0.0010% IS (MR0.8 to MR5.0).

The linear regression analyses were performed for first, second and third order polynomials per CLSI EP6-A. The data were considered linear if the non-linear coefficients from the second and third order polynomials were insignificant (p > 0.05). When significant non-linear coefficients were observed, the Degree of Nonlinearity was calculated per CLSI EP6-A. The linear regression curves for both transcripts in the two collection tube types are shown in below followed by the estimated regression intercepts and slopes from the linear model.

Image /page/19/Figure/6 description: The image contains two scatter plots titled "Linearity - EDTA Tubes". The left plot shows data for "1.0 ug RNA input EDTA, e13a2", while the right plot shows data for "1.0 ug RNA input EDTA, e14a2". Both plots depict "Molecular Response" on the y-axis and "Nominal Molecular Response" on the x-axis, with data points scattered along a generally linear trend.

{20}------------------------------------------------

Image /page/20/Figure/1 description: The image contains two scatter plots comparing nominal molecular response to molecular response. Both plots show a linear relationship between the two variables. The left plot is labeled "2.0 µg RNA input PAXgene,e13a2", while the right plot is labeled "2.0 µg RNA input PAXgene,e14a2". Both plots show data points clustered around a line, indicating a strong correlation.

Regression Coefficients from Linear Model

TranscriptInterceptSlope
PAXgenee14a2-0.0450.0540.980.017
e13a20.0120.0260.930.008
EDTAe14a2-0.0160.0370.960.011
e13a20.0590.0460.920.015

Transcript e13a2 was linear from MR0.93 to MR5.1 with a maximum SD of 0.22 using EDTA tubes and was linear from MR0.78 to MR4.8 with a maximum SD of 0.13 using PAXgene tubes. Transcript e14a2 was linear from MR0.99 to MR5.0 with a maximum SD of 0.26 using EDTA tubes and was linear from MR0.93 to MR4.9 with a maximum SD of 0.31 using PAXgene tubes. The data support the assay's linearity across the indicated detection range.

Linear Range
TubeTypeBCR-ABLTranscriptNominalLower LimitBCR-ABL/ABL% ISNominalLower LimitMolecularResponseNominalUpper LimitBCR-ABL/ABL% ISNominal UpperLimitMolecularResponse
EDTAe13a20.000805.1120.93
EDTAe14a20.00105.0100.99
PAXgenee13a20.00174.8170.78
PAXgenee14a20.00124.9120.93

{21}------------------------------------------------

The linear range was truncated by the LoQ to determine the assay range. The data support the conclusion that the assay is linear across the reportable range of MR1.0 to MR4.5. The assay range for the MRDx BCR-ABL Test for each blood collection tube type are:

Assay Range
Tube TypeLinear RangeBCR-ABL/ABL % ISLinear RangeMolecular Response
EDTA10% to 0.0032%MR1.0 to MR4.5
PAXgene10% to 0.0032%MR1.0 to MR4.5

10. Stability Studies

Kit Real-Time and Freeze-Thaw Stability:

The real-time stability of the MRDx BCR-ABL Test was evaluated at two storage temperatures (-30 to -15°C and -80 to -65°C) at periodic intervals with three lots. Six kits from each lot for each time point were tested using the final product release testing procedure. After use, the kits were frozen, thawed, and refrozen without testing to achieve an additional freeze/thaw cycle (Cycle 2). These six kits were then open-vial tested at one month (Test 2 and Cycle 3), and two months (Test 3 and Cycle 4) post-reconstitution using the WHO BCR-ABL Reference Panel Secondary Standards to ensure the kits were stable for three kit uses with three freeze/thaw cycles with 2 month stability. Kits were stored at -30 to -15°C after initial use for open-vial stability, regardless the original storage condition.

Image /page/21/Figure/7 description: The image shows a flowchart that describes the process of using manufactured kits. The process starts with manufactured kits, which are then thawed and used for reconstitution and initial use, labeled as "Test 1" and "Cycle 1". After this, the kits undergo a freeze-thaw cycle and are then labeled as "Cycle 2" without use of the kit. The kits then undergo another freeze-thaw cycle and are used a second time, labeled as "Test 2" and "Cycle 3", followed by a final freeze-thaw cycle and a third use, labeled as "Test 3" and "Cycle 4".

For each time point, the MRDx kit was tested for performance of the calibration curve criteria, the control results, and the WHO secondary standards. Passing test results data exist for at least 16 months for 3 lots of MRDx BCR-ABL Test stored at the two separate temperatures.

The data support a shelf life of 15 months at both storage temperatures (-30 to -15°C and -80 to -65°C) and an open vial stability of 3 freeze-thaw cycles and up to 2 months following initial use.

Reagent Preparation and Reaction Intermediate Stability (In Use):

This study demonstrated the MRDx BCR-ABL Test reagent stability during PCR plate preparation and the post-plate-assembly PCR reaction stability at 2 to 8°C before evaluation on the ABI 7500 Fast Dx instrument.

Both standard and extreme conditions were tested. The standard condition used a newly opened kit immediately after thawing to setup a PCR plate. The PCR plate was processed

{22}------------------------------------------------

immediately after setup on the ABI 7500 Fast Dx. The extreme condition used a kit that was thawed and refrozen twice to simulate the third use of the kit. After the reagents were thawed, components were stored at 2 to 8°C for eight hours, and then the PCR plate was assembled and placed at 2 to 8 °C for 1.5 hours before evaluation on the ABI 7500 Fast Dx instrument. Two kits were run for each condition on each of 2 days with 2 sets of calibration curves, RNA controls, and samples (total n=8 per condition).

The acceptance criteria were that the mean of the test samples was ± 0.5 MR of the control. The acceptance criteria were met for the standard and the extreme conditions. The differences between the standard condition and the extreme condition were less than 0.25 logio for all levels of samples tested. The largest difference observed between conditions was -0.063 (95% Cl: -0.24 to 0.12) for the MR4.5 sample. These results support the stability conclusion that after 2 freeze-thaw cycles, the reagent is stable at 2 to 8°C for up to eight hours before PCR reagent assembly and up to 1.5 hours after PCR plate assembly.

Specimen Stability:

For PAXgene Blood RNA Tubes, samples at three levels (MR3.0, MR4.0, and MR4.5) were created, stored at 2 to 8°C and tested in quadruplicate at 0, 24, 48, and 60 hours. An additional series of samples was created, stored at -30 to -15°C, and tested in quadruplicate at 0, 14, 28, and 30 days. For EDTA blood collection tubes, 17 CML patient specimens covering the clinical decision range were collected and stored at 2 to 8°C and tested in quadruplicate at 0, 24, 48, and 60 hours.

All tested samples passed the acceptance criteria of < 0.25 log10 difference compared to the mean BCR-ABL/ABL IS log10 ratio results at To. These results support the following stability conclusion:

  • . Specimens collected in PAXgene Blood RNA Tubes are stable for at least 48 hours at 2 to 8°C and at least 28 days at -30 to -15°C.
  • . Specimens collected in EDTA blood tubes are stable for at least 48 hours at 2 to 8°C.

Shipping Stability:

In this study, the MRDx BCR-ABL Test kits were evaluated with multiple types of containers and shipping conditions. The mock shipping conditions included both stress temperatures and extended duration. The shipping conditions evaluated the minimum and maximum payload for both types of shipping boxes. Maximum load was the worst case regarding the amount of dry ice per shipping condition. The testing conditions were executed sequentially as follows:

  • 22°C ± 3°C for 4 hours ●
  • 35°C ± 3°C for 6 hours ●
  • 30°C ± 3°C for 56 hours
  • 35°C ± 3°C for 6 hours

{23}------------------------------------------------

The results of this study demonstrated that the Polar Tech TC229 (P/N 049-0128) and BioFreeze 22L-133Hr (P/N 049-0129) shipping containers can maintain an internal temperature below -65°C for all load conditions during a 72 hour temperature stress test that simulated typical shipping conditions. During this shipping test, the MRDx BCR-ABL Test kits remained at storage conditions of -80 to -65°C.

Functional testing using samples across the assay range also showed that the kit performance after shipping stress met equivalency acceptance criteria when compared to unstressed control kits. These results support the conclusion that the assay kit is stable following shipment of up to 72 hours on dry ice.

11. PAXgene-EDTA Tube Equivalency

A blood matrix equivalency study using 100 CML patient paired blood samples was conducted to demonstrate that MRDx BCR-ABL Test results from RNA samples extracted from PAXgene Blood RNA Tubes are the same as from EDTA blood collection tubes. At least 4 PAXgene Blood RNA Tubes and two 6 mL EDTA blood collection tubes were collected for each subject. RNA was extracted from the PAXgene Blood RNA Tubes using the PAXgene Blood RNA Kit. RNA from the EDTA blood collection tubes was extracted using the Maxwell CSC Blood RNA Kit.

For all paired samples that were above the LoQ of the test, a Deming regression was applied and 95% confidence intervals of the slope and y-intercept were calculated. The predicted systematic difference between the two methods was calculated at the clinical decision points (MR3.0, MR4.0, and MR4.5) with 95% confidence intervals. The data demonstrated that no clinically significant bias at the clinical decision points was observed in the MR values from MRDx BCR-ABL Test results between the two blood collection tube types. Results are shown below.

D
SampleSizeDeming SlopeDemingY-InterceptDemingPearson Coefficient
1001.0295% Cl [1.00, 1.05]-0.11595% CI [-0.202, -0.042]0.992

Deming Regression (EDTA vs. PAXgene)

Predicted Difference at Clinical Decision Points

Predicted Difference(MMR/MR3)Predicted DifferenceMR4Predicted DifferenceMR4.5
-0.0433-0.0192-0.00721
95% CI [-0.0687, -0.0175]95% CI [-0.0564, 0.0238]95% CI [-0.0549, 0.0462]

{24}------------------------------------------------

Clinical Studies B.

The clinical validity of the MRDx BCR-ABL Test was demonstrated in two clinical trials, ENESTfreedom (CAMN10712201) and ENESTop (CAMN107A2408), for efficacy of identifying CML-CP patients on nilotinib treatment, who may be eligible to enter and maintain treatment free remission (TFR), and monitoring for potential loss of remission after treatment discontinuation. The study summary is described below, please refer to Tasigna® drug labeling at Drugs@FDA for more details.

1. Treatment discontinuation in newly diagnosed Ph+ CML-CP patients who have achieved a sustained molecular response (MR4.5)

The ENESTfreedom study is an open-label, multicenter, single-arm study, where 215 adult patients with Ph+ CML-CP treated with nilotinib in first-line for ≥ 2 years who achieved MR4.5 as measured with the MolecularMD MRDx BCR-ABL Test were enrolled to continue nilotinib treatment for an additional 52 weeks (nilotinib consolidation phase).

Of the 215 patients, 190 patients (88.4%) entered the "Treatment-Free Remission" (TFR) phase after achieving a sustained molecular response (MR4.5) during the consolidation phase, defined by the following criteria:

  • . The 4 last quarterly assessments by MRDx BCR-ABL Test (taken every 12 weeks) were at least MR4.0 (BCR-ABL/ABL ≤ 0.01% IS), and maintained for 1 year
  • The last assessment being MR4.5 (BCR-ABL/ABL ≤ 0.0032% IS)
  • No more than two assessments falling between MR4.0 and MR4.5 (0.0032% IS < BCR-ABL/ABL ≤ 0.01% IS)

The median age of patients who entered the TFR phase was 55 years, 49.5% were females, and 21.1% of the patients were ≥ 65 years of age. BCR-ABL levels were monitored every 4 weeks during the first 48 weeks of the TFR phase. Monitoring frequency was intensified to every 2 weeks upon the loss of MR4.0. Biweekly monitoring ended at one of the following time points:

  • Loss of MMR requiring patient to re-initiate nilotinib treatment
  • When the BCR-ABL levels returned to a range between MR4.0 and MR4.5
  • . When the BCR-ABL levels remained lower than MMR for 4 consecutive measurements (8 weeks from initial loss of MR4.0).

Any patient with loss of MMR during the TFR phase re-initiated nilotinib treatment at 300 mg twice daily or at a reduced dose level of 400 mg once daily if required from the perspective of tolerance, within 5 weeks after the collection date of the blood sample demonstrating loss of MMR. Patients who required re-initiation of nilotinib treatment were monitored in the nilotinib treatment reinitiation (NTRI) phase for BCR-ABL levels every 4 weeks for the first 24 weeks and then every 12 weeks thereafter in patients who regained MMR.

Efficacy was based on the 96-week analysis data cut-off date, by which time, 91 patients (47.9%) discontinued from the TFR phase due to loss of MMR, and 1 (0.5%), 1 (0.5%) MolecularMD MRDx® BCR-ABL Test Page 22 of 25 510(k) Summary

{25}------------------------------------------------

and 3 patients (1.6%) due to death from unknown cause, physician decision, lost to follow-up, and subject decision, respectively. Among the 91 patients who discontinued the TFR phase due to loss of MMR, 88 patients restarted nilotinib treatment and 3 patients permanently discontinued from the study. The efficacy results for ENESTfreedom study are summarized in the table below.

Patients who entered TFR phase (Full Analysis Set, N=190)
Patients in TFR phaseDiscontinuations dueto loss of MMRPatients in NTRIphase
Timen (%)95% CIn (%)n (%)
24 weeks118 (62.1%)(54.8, 69.0)70 (36.8%)69 (36.3%)
48 weeks98 (51.6%)(44.2, 58.9)88 (46.3%)86 (45.3%)
96 weeks93 (48.9%)(41.6, 56.3)91 (47.9%)88 (46.3%)
Efficacy Results for ENESTfreedom Study
-------------------------------------------------

Of the 88 patients who restarted treatment due to loss of MMR in the TFR phase, 87 patients (98.9%) patients regained MMR (one patient discontinued study permanently due to subject decision after 7.1 weeks of retreatment without regaining MMR) and 81 patients (92.0%) regained MR4.5 by the time of the cut-off date. The cumulative rate of MMR and MR4.5 regained at 24 weeks since treatment reinitiation was 97.7% (86/88 patients) and 86.4% (76/88 patients), respectively.

Among the 190 patients in the TFR phase, 98 patients had a treatment-free survival (TFS) event (defined as discontinuation from TFR phase due to any reason, loss of MMR, death due to any cause, progression to AP/BC up to the end of TFR phase, or re-initiation of treatment due to any cause in the study) by the 96-week cut-off date.

2. Treatment discontinuation in Ph+ CML-CP patients who have achieved a sustained molecular response (MR4.5) on nilotinib following prior imatinib therapy

Study ENESTop (NCT01698905) is an open-label, multicenter, single-arm study, where 163 adult patients with Ph+ CML-CP taking tyrosine kinase inhibitors (TKls) for ≥ 3 years (imatinib as initial TKI therapy for more than 4 weeks without documented MR4.5 on imatinib at the time of switch to nilotinib, then switched to nilotinib for at least 2 years), and who achieved MR4.5 on nilotinib treatment as measured with the MolecularMD MRDx BCR-ABL Test were enrolled to continue nilotinib treatment for an additional 52 weeks (nilotinib consolidation phase). Of the 163 patients, 126 patients (77.3%) entered the TFR phase after achieving a sustained molecular response (MR4.5) during the consolidation phase, defined by the following criterion:

  • The 4 last quarterly assessments by the MRDx BCR-ABL Test (taken every 12 weeks) showed no confirmed loss of MR4.5 (BCR-ABL/ABL ≤ 0.0032% IS) during one year.

{26}------------------------------------------------

The median age of patients who entered the TFR phase was 56 years, 55.6% were females, and 27.8% of the patients were ≥ 65 years of age. The median actual dose intensity during the 52week nilotinib consolidation phase was 771.8 mg/day with 52.4%, 29.4%, 0.8%, 16.7% and 0.8% of patients receiving a daily nilotinib dose of 800 mg, 450mg, 400mg and 300mg just before entry into the TFR phase, respectively.

Patients who entered the TFR phase but experienced two consecutive measurements of BCR-ABL/ABL > 0.01% IS were considered having a confirmed loss of MR4.0, triggering re-initiation of nilotinib treatment. Patients with loss of MMR in the TFR phase immediately restarted nilotinib treatment without confirmation. All patients who restarted nilotinib therapy had BCR-ABL transcript levels monitored every 4 weeks for the first 24 weeks, then once every 12 weeks.

Efficacy was based on the 96-week analysis data cut-off date, by which time, 61 patients (48.4%) had discontinued from the TFR phase: 58 patients (46.0%) due to loss of MMR or confirmed loss of MR4.0, 2 patients (1.6%) due to subject/guardian decision and one patient (0.8%) due to pregnancy. Among the 58 patients who discontinued from the TFR phase due to confirmed loss of MR4.0 or loss of MMR, 56 patients restarted nilotinib therapy and 2 patients permanently discontinued from the study. The efficacy results are summarized in the table below.

Patients who entered TFR phase (Full Analysis Set, N=126)
Patients in TFR phaseDiscontinuations due toloss of MMR or confirmedloss of MR4Patients inNTRI phase
Timen (%)95% CIn (%)n (%)
24 weeks76 (60.3%)(51.2, 68.9)49 (38.9%)48 (38.1%)
48 weeks73 (57.9%)(48.8, 66.7)53 (42.1%)51 (40.5%)
96 weeks67 (53.2%)(44.1, 62.1)58 (46.0%)56 (44.4%)

Efficacy Results for ENESTop Study

Of the 56 patients who restarted nilotinib treatment due to confirmed loss of MR4.0 or loss of MMR in the TFR phase, 52 patients (92.9%) regained MR4.0 and MR4.5 and 4 patients (7.1%) did not regain MR4.0 by the time of the cut-off date. The cumulative rate of MR4.0 and MR4.5 regained by 48 weeks since treatment reinitiation, was 92.9% (52/56 patients) and 91.1% (51/56 patients), respectively.

Among the 126 patients in the TFR phase, 61 patients (48.4%) had a treatment-free survival (TFS) event (defined as discontinuation from TFR phase due to any reason, loss of MMR, confirmed loss of MR4.0, death due to any cause, progression to AP/BC up to the end of TFR phase, or reinitiation of treatment due to any cause in the study) on or before the 96-month cut-off date.

C. Conclusions Drawn from Preclinical and Clinical Studies

The effectiveness and clinical benefit of the MRDx® BCR-ABL Test was demonstrated in two prospective clinical trials utilizing the monitoring of BCR-ABL mRNA transcript levels in patients MolecularMD MRDx® BCR-ABL Test 510(k) Summary Page 24 of 25

{27}------------------------------------------------

Image /page/27/Picture/0 description: The image shows the logo for MolecularMD. The logo consists of a red caduceus symbol on the left, followed by the text "MolecularMD" in gray and red. The word "Molecular" is in gray, while "MD" is in red.

diagnosed with CML as an aid in identifying patients being treated with nilotinib as candidates for initiating treatment-free remission. Further, the monitoring of patients in treatment-free remission successfully met the intended use of identifying patients requiring reinitiation of nilotinib treatment. Analytical performance studies provide assurance of the accuracy and reliability of the quantitation of BCR-ABL transcript levels.

The data from these studies support the reasonable assurance of safety and effectiveness of the MRDx® BCR-ABL Test when used in accordance with the indications for use.

§ 866.6060 BCR-ABL quantitation test.

(a)
Identification. A BCR-ABL quantitation test is identified as a reverse transcription-quantitative polymerase chain reaction (RT-qPCR) test for the quantitation of BCR-ABL1 expressed on the International Scale (IS) and control transcripts in total RNA from whole blood of diagnosed t(9;22) positive chronic myeloid leukemia (CML) patients during monitoring of treatment with tyrosine kinase inhibitors. This test is not intended for the diagnosis of CML.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the following information:
(i) The indication for use must indicate the variant(s) for which the assay was designed and validated, for example BCR-ABL e13a2 and/or e14a2.
(ii) A detailed description of all components in the test, including the following:
(A) A detailed description of the test components, all required reagents, instrumentation and equipment, including illustrations or photographs of non-standard equipment or methods;
(B) Detailed documentation of the device software including, but not limited to, standalone software applications and hardware-based devices that incorporate software;
(C) Methodology and protocols for control procedures for the assay to allow reporting on the International Scale;
(D) A description of the result outputs, analytical sensitivity of the assay, and the range of values that will be reported; and
(E) A description of appropriate internal and external controls that are recommended or provided. The description must identify those control elements that are incorporated into the testing procedure.
(iii) Information that demonstrates the performance characteristics of the test, including:
(A) For indications for use based on a threshold established in a predicate device of this generic type, device performance data from either a method comparison study to the predicate device or through a clinical study demonstrating clinical validity using well-characterized prospectively or retrospectively obtained clinical specimens, as appropriate, representative of the intended use population;
(B) For indications for use based on a threshold not established in a predicate device of this generic type, device performance data from a clinical study demonstrating clinical validity using well-characterized prospectively or retrospectively obtained clinical specimens, as appropriate, representative of the intended use population;
(C) Device reproducibility data generated, using a minimum of three sites, of which at least two sites must be external sites, with two operators at each site. Each site must conduct a minimum of three runs per operator over non-consecutive days evaluating a minimum of five different BCR-ABL concentrations that span and are well distributed over the measuring range and include MR3 (0.1 percent IS). Results shall be reported as the standard deviation and percentage coefficient of variation for each level tested. Prespecified acceptance criteria must be provided and followed;
(D) Device precision data using clinical samples to evaluate the within-lot, between-lot, within-run, between run, and total variation;
(E) Device linearity data using a dilution panel created from clinical samples;
(F) Device analytic sensitivity data, including limit of blank, limit of detection, and limit of quantification;
(G) Device specificity data, including interference and cross-contamination; and
(H) Device stability data, including real-time stability of samples under various storage times, temperatures, and freeze-thaw conditions.
(iv) Identification of risk mitigation elements used by your device, including a detailed description of all additional procedures, methods, and practices incorporated into the instructions for use that mitigate risks associated with testing using your device.
(2) Your 21 CFR 809.10 compliant labeling must include the following:
(i) The intended use in your 21 CFR 809.10(a)(2) and (b)(2) complaint labeling must include an indication for use statement that reads “This test is not intended for the diagnosis of CML”; and
(ii) A detailed description of the performance studies conducted to comply with paragraph (b)(1)(iii) of this section and a summary of the results.
(3) Your device output must include results on the International Scale (IS) and your assay must include multipoint calibration controls traceable to a relevant international reference panel (
e.g., the World Health Organization International Genetic Reference Panel for quantitation of BCR-ABL mRNA).