K Number
K173374

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-12-13

(47 days)

Product Code
Regulation Number
872.3630
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TSV™ BellaTek® Encode® Healing Abutments are intended for use as an accessory to endosseous dental implants during endosseous and gingival healing to prepare gingival tissue for acceptance of a final abutment and restoration.

Device Description

The purpose of this submission is to obtain 510(k) premarket clearance for the TSVIM BellaTek® Encode® Healing Abutments compatible with Tapered Screw-Vent® and Trabecular Metal implant systems. Abutments sizing is based on implant platform diameter.

TSV™ BellaTek® Encode® Healing Abutments are designed to aid in soft tissue contouring during the healing period after implant placement, creating an emergence profile for the final prosthesis. They have the added design feature of machined markings for identification when taking an abutment level impression or an intraoral scan/digital impression. The occlusal surface of the device include markings that provide information about the mating implant's position and orientation.

The principal of operation and Encode "Coding Scheme of the same as the primary predicate device. The pattern of the markings for the subject device is specific to the Tapered Screw-Vent", and Trabecular Metal implant lines.

AI/ML Overview

The provided text does not contain information about acceptance criteria or a study proving that the device meets specific performance criteria. Instead, it is a 510(k) summary for a medical device (TSV™ BellaTek® Encode® Healing Abutments) demonstrating substantial equivalence to a predicate device.

The document explicitly states: "No clinical data were included in this submission." and relies on non-clinical testing data and equivalence to a predicate device.

Therefore, I cannot provide the requested information about acceptance criteria, device performance, sample sizes, ground truth, expert opinions, or MRMC studies, as this information is not present in the provided text.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

December 13, 2017

Biomet 3i % Linda Schulz Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130

Re: K173374

Trade/Device Name: TSV™ BellaTek® Encode® Healing Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: October 26, 2017 Received: October 27, 2017

Dear Linda Schulz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Mary S. Runner -S

For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K173374

Device Name

TSVTM BellaTek® Encode® Healing Abutments

Indications for Use (Describe)

The TSV™ BellaTek® Encode® Healing Abutments are intended for use as an accessory to endosseous dental implants during endosseous and gingival healing to prepare gingival tissue for acceptance of a final abutment and restoration.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments reqarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

PSC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

510(k) Summary K173374 Biomet 3i TSV™ BellaTek® Encode® Healing Abutments

October 26, 2017

ADMINISTRATIVE INFORMATION

Manufacturer NameBiomet 3i4555 Riverside DrivePalm Beach Gardens, FL 33410Telephone: +1 (561) 776-6700Fax: +1 (561) 776-1272
Official ContactNichole O'Neil, Regulatory Affairs Specialist
Representative/ConsultantLinda K. Schulz, BSDH, RDHFloyd G. Larson, MS, MBAPaxMed International, LLC12264 El Camino Real, Suite 400San Diego, CA 92130Telephone: +1 (858) 792-1235Fax: +1 (858) 792-1236Email: lschulz@paxmed.comflarson@paxmed.com

DEVICE NAME AND CLASSIFICATION

Trade/Proprietary NameTSVTM BellaTek® Encode® Healing Abutments
Common NameDental Implant Abutment
Classification NameEndosseous dental implant abutment
Classification Regulations21 CFR 872.3630, Class II
Product CodeNHA
Classification Panel:Dental Products Panel
Reviewing Branch:Dental Devices Branch

PREDICATE DEVICE INFORMATION

Primary Predicate:
K170013Eztetic™ BellaTek® Encode® Healing AbutmentsBiomet 3i
Reference Predicates for Compatibility:
K061410Zimmer Dental Implant SystemZimmer Dental, Inc.
K133339Zimmer Dental Tapered Screw-Vent® T Implant, HA CoatedZimmer Dental, Inc.
Zimmer Dental Tapered Screw-Vent® M Implant, HA Coated
K113753Tapered Screw-Vent® X ImplantZimmer Dental, Inc.
K132258Zimmer Dental Trabecular Metal Implant SystemZimmer Dental, Inc.

{4}------------------------------------------------

510(k) Summary Page 2 of 4

INDICATIONS FOR USE

The TSV™ BellaTek® Encode® Healing Abutments are intended for use as an accessory to endosseous dental implants during endosseous and gingival healing to prepare gingival tissue for acceptance of a final abutment and restoration.

DEVICE DESCRIPTION

The purpose of this submission is to obtain 510(k) premarket clearance for the TSVIM BellaTek® Encode® Healing Abutments compatible with Tapered Screw-Vent® and Trabecular Metal implant systems. Abutments sizing is based on implant platform diameter.

TSV™ BellaTek® Encode® Healing Abutments are designed to aid in soft tissue contouring during the healing period after implant placement, creating an emergence profile for the final prosthesis. They have the added design feature of machined markings for identification when taking an abutment level impression or an intraoral scan/digital impression. The occlusal surface of the device include markings that provide information about the mating implant's position and orientation.

The principal of operation and Encode "Coding Scheme of the same as the primary predicate device. The pattern of the markings for the subject device is specific to the Tapered Screw-Vent", and Trabecular Metal implant lines.

Compatible Implant Sizes
Implant SystemImplant Diameter(mmD)Implant Platform(mmD)Implant Length(mmL)
Tapered Screw -Vent®3.73.58, 10, 11.5, 13,16
Tapered Screw -Vent®4.13.58, 10, 11.5, 13,16
Tapered Screw -Vent®4.74.58, 10, 11.5, 13,16
Tapered Screw -Vent®6.05.78, 10, 11.5, 13,16
Screw-Vent®3.33.58,10,13,16
Screw-Vent®3.73.58,10,13,16
Screw-Vent®4.74.58,10,13,16
Trabecular Metal™3.73.510, 11.5, 13
Trabecular Metal™4.13.510, 11.5, 13
Trabecular Metal™4.74.510, 11.5, 13
Trabecular Metal™6.05.710, 11.5, 13

Compatible Implant Sizes

PERFORMANCE DATA

Non-clinical testing data submitted or relied upon to demonstrate substantial equivalence included radiation sterilization validation according to ISO 11137-1 and 11137-2, demonstrating assurance level (SAL) of 10 °, Limulus amebocyte lysate (LAL) testing according to AANSI/AAMI ST 72, biological evaluation according to ISO 10993-1 by reference to K170013, demonstrating acceptable biocompatibility, accelerated and real time aging studies by reference to K170013, demonstrating a shelf life of five years, and MR evaluation by reference to K170013. The subject devices are labeled as MR conditional.

No clinical data were included in this submission.

{5}------------------------------------------------

EQUIVALENCE TO MARKETED DEVICE

Biomet 3i submits the information in this Premarket Notification to demonstrate that, for the purposes of FDA's regulation of medical devices, the subject device, TSV™ BellaTek® Encode® Healing Abutments, is substantially equivalent in indications and design principles to the legally marketed primary predicate device.

ComparisonSubject DevicePrimary Predicate Device
TSVTM BellaTek® Encode® HealingAbutmentsK170013The EzteticTM BellaTek® Encode®Healing Abutments
Biomet 3iBiomet 3i
Indications for UseThe TSVTM BellaTek® Encode®Healing Abutments are intended for useas an accessory to endosseous dentalimplants during endosseous andgingival healing to prepare gingivaltissue for acceptance of a finalabutment and restoration.The EzteticTM BellaTek® Encode®Healing Abutments are intended for useas an accessory to endosseous dentalimplants during endosseous and gingivalhealing to prepare gingival tissue foracceptance of a final abutment andrestoration.
Design
Scanning IdentificationEncode® Coding Scheme(Machined Markings)Encode® Coding Scheme(Machined Markings)
Emergence Profile (mm)3.8, 5.0, 5.6, 6.0, 6.83.8, 5.0
Platform Diameter (mm)3.5, 4.5, 5.72.9
Gingival Height (mm)3, 5, 73, 4, 6, 8
Implant/AbutmentConnectionInternalInternal
Material
AbutmentTi-6Al-4V ELITi-6Al-4V ELI
ScrewTi-6Al-4V ELITi-6Al-4V ELI

The intended use of the subject device as a healing abutment machined markings for identification is the same as the primary predicate K170013. The Indications for the subject device is the same as the primary predicate Eztetic™ BellaTek® Encode® Healing Abutment, K170013, changing only the name.

The principal of operation and basic design of the Encode Coding Scheme of the subject device is the same as the primary predicate device, Eztetic™ BellaTek® Encode® Healing Abutment, K170013. The pattern of the markings for the subject device is specific to the Tapered Screw-Vent, and Trabecular Metal implant lines.

All of the subject device components are manufactured from the same facilities, using the same manufacturing processes as used for the previously cleared primary predicate device in K170013.

{6}------------------------------------------------

CONCLUSION

The subject device and the primary predicate device have intended use, have same technological characteristics, and are made of the same materials. The subject device and predicate devices encompass the same design principals, including the Encode® Coding Scheme. The subject and predicate devices are packaged in the same materials and are to be sterilized using the same methods.

The data included in this submission demonstrate substantial equivalence to the primary predicate device listed above.

§ 872.3630 Endosseous dental implant abutment.

(a)
Identification. An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)