K Number
K173164

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-06-29

(273 days)

Product Code
Regulation Number
876.5665
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ECOLAB POU Water Filters are intended to operate on EPA quality drinking water sources as a microbial retention filter. ECOLAB POU Water Filters are suitable for the control of bacteria equal to or greater in size than Brevundimons diminuta. The POU Filters are suitable for general point of use infection control for procedures such as superficial wound cleansing, cleaning of equipment, washing of surgeon's hands and bathing where the reduction of such microorganisms in the water is desired.

The POU Filters are not intended for use in the production of USP sterile water for use in infusion, injection of fluids for use in dialysis treatments.

Device Description

The Ecolab POU Water Filter is a disposable Polyethersulfone (PES) membrane filter sealed in a white polypropylene housing designed to be installed by facility personnel at point of use locations where the control of bacteria in EPA quality drinking water is desired for hand- washing, bathing, showering and instrument cleaning applications. These filters are designed to be installed as attachments to faucets or as a handheld shower head attached to a water outlet.

AI/ML Overview

The request is about a water filter, not an AI/ML medical device. Therefore, questions 2, 3, 4, 5, 8, and 9 are not applicable to this document. The document describes a traditional medical device (water filter) and its performance characteristics, not an algorithm, so terms like "ground truth," "training set," "test set," and "human-in-the-loop" performance are not relevant.

However, I can extract information related to acceptance criteria and performance studies for the Ecolab POU Water Filter based on the provided text.

Here's the information derived from the provided document, addressing the relevant parts of your request:

Device: Ecolab POU Water Filter (K173164)

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria / Performance MetricTarget (Acceptance Criteria)Reported Device Performance
Microbial Retention (against Brevundimonas diminuta)Suitable for control of bacteria equal to or greater in size than Brevundimonas diminuta.All Ecolab POU Water Filters retained the test organism Brevundimonas diminuta challenge of ≥ 10^7^ organisms/cm² and prevented bacterial grow-through over the service life of each model. This meets the predicate device's > 10^7^ bacteria reduction.
Filter Flow Rate (at 3 bar inlet pressure)Not explicitly stated as an "acceptance criterion" with a specific numerical target, but flow rates are a performance characteristic.7-Day tap filter: 6.3 L/min14-Day tap filter: 6.5 L/min31-Day tap filters: 6 L/min
Maintenance of Microbial Retention Effectiveness after Cyclic Pulse Testing (Tap Filters)Maintain microbial retention effectiveness after continuous use repeated shocks of 89.9 psi (6.2 bar) at 151°F (66°C).Maintained microbial retention effectiveness (as measured by diffusional flow integrity test) after continuous use repeated shocks of 89.9 psi (6.2 bar) at 151°F (66°C).
Maintenance of Microbial Retention Effectiveness after Cyclic Pulse Testing (Shower Wand Filters)Maintain microbial retention effectiveness after continuous use repeated shocks of 79.8 psi (5.5 bar) at 122°F (50°C).Maintained microbial retention effectiveness (as measured by diffusional flow integrity test) after continuous use repeated shocks of 79.8 psi (5.5 bar) at 122°F (50°C).
SterilityAssure product sterility.Verified that the gamma irradiation process assures product sterility (according to ISO 11137-1 and ISO 11137-2).
Package Integrity (to maintain sterile barrier)Capable of maintaining a sterile barrier for the labeled shelf life.Demonstrated that the packaging is capable of maintaining a sterile barrier for the labeled 36-month shelf life (according to ISO 11607-1 and 11607-2).
Retention of Functional Characteristics (through shelf life)Retain functional characteristics through shelf life.Verified that the filters retain their functional characteristics through 36 months.
BiocompatibilitySatisfy biocompatibility requirements for the intended use (cytotoxicity, irritation, sensitization, acute systemic toxicity, material-mediated pyrogenicity).All biocompatibility testing was conducted in accordance with ISO 10993-5 (Cytotoxicity), ISO 10993-10 (Irritation, Sensitization), and ISO 10993-11 (Acute Systemic Toxicity, Material Mediated Pyrogenicity) under GLP 21 CFR Part 58 Regulations. The document states "Biocompatibility of the Ecolab POU Water Filters was verified," implying successful completion and meeting of criteria for all listed tests.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable as the device is a physical water filter, not an algorithm. The "performance" assessment is based on the filter's physical and biological filtration capabilities.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For this physical device, "ground truth" equates to laboratory testing against established standards and known microbial challenges.

  • Microbial Retention: The "ground truth" was the Brevundimonas diminuta challenge ≥ 10^7^ organisms/cm², with the effectiveness judged by the absence of bacterial grow-through. This is a direct measurement of the device's capability.
  • Sterility: The "ground truth" for sterility was verification against ISO 11137-1 and ISO 11137-2 standards for sterilization processes.
  • Package Integrity: The "ground truth" for package integrity was verification against ISO 11607-1 and 11607-2 standards.
  • Biocompatibility: The "ground truth" for biocompatibility was verification against ISO 10993 series standards and GLP 21 CFR Part 58 Regulations.

There are no concepts of "training sets" or "ground truth establishment" for a training set as would apply to machine learning models. The tests were performed to directly demonstrate the specified physical and biological performance characteristics of the manufactured device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

June 29, 2018

Ecolab Inc. Jennifer Willner, RAC Sr. Director Regulatory Affairs One Ecolab Place St. Paul, MN 55102

Re: K173164

Trade/Device Name: Ecolab POU Water Filter Regulation Number: 21 CFR 876.5665 Regulation Name: Water purification system for hemodialysis Regulatory Class: Class II Product Code: NHV Dated: May 29, 2018 Received: May 30, 2018

Dear Jennifer Willner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K173164

Device Name Ecolab POU Water Filter

Indications for Use (Describe)

The ECOLAB POU Water Filters are intended to operate on EPA quality drinking water sources as a microbial retention filter. ECOLAB POU Water Filters are suitable for the control of bacteria equal to or greater in size than Brevundimons diminuta. The POU Filters are suitable for general point of use infection control for procedures such as superficial wound cleansing, cleaning of equipment, washing of surgeon's hands and bathing where the reduction of such microorganisms in the water is desired.

The POU Filters are not intended for use in the production of USP sterile water for use in infusion, injection of fluids for use in dialysis treatments.

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Ecolab. The logo is in blue and features the company name in a bold, sans-serif font. The "O" in Ecolab is replaced with a stylized image of a star or asterisk inside of a circle. There is a registered trademark symbol next to the "B" in Ecolab.

510(k) Summary

Manufacturer:Ecolab Inc.
One Ecolab Place
St Paul, MN 55102 USA
Establishment Registration Number: 9009921
Contact:Jennifer Willner, RAC
Sr. Director Regulatory Affairs
Ecolab Inc.
One Ecolab Place, St Paul, MN 55102 USA
Phone: 651 250 4348
Date PreparedMay 24, 2018
Trade NameEcolab POU Water Filter
ClassificationClass II
Regulation NameWater purification system for hemodialysis
Regulation Number21 CFR 876.5665
Product CodeNHV
Predicate DeviceEvoqua Water Technologies, Nosogard Filters, K153784
Device DescriptionThe Ecolab POU Water Filter is a disposable Polyethersulfone (PES)
membrane filter sealed in a white polypropylene housing designed to
be installed by facility personnel at point of use locations where the
control of bacteria in EPA quality drinking water is desired for hand-
washing, bathing, showering and instrument cleaning applications.
These filters are designed to be installed as attachments to faucets or
as a handheld shower head attached to a water outlet.
Indication for UseThe ECOLAB POU Water Filters are intended to operate on EPA quality
drinking water sources as a microbial retention filter. ECOLAB POU
Water Filters are suitable for the control of bacteria equal to or greater
in size than Brevundimons diminuta. The POU Filters are suitable for
general point of use infection control for procedures such as superficial
wound cleansing, cleaning of equipment, washing of surgeon's hands
and bathing where the reduction of such microorganisms in the water
is desired.
The POU Filters are not intended for use in the production of USP sterile
water for use in infusion, injection or production of fluids for use in
dialysis treatments.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the Ecolab logo in blue. The logo consists of the word "ECOLAB" in a bold, sans-serif font. The "O" in Ecolab is a circle with a star-like design inside. There is a registered trademark symbol to the right of the logo.

Substantial Equivalence Comparison Table
Subject DevicePredicate Device
510(k) NumberK173164K153784
Trade NameEcolab POU Water FilterNosogard Filters
ManufacturerEcolab Inc.Evoqua Water Technologies, LLC.
Indications forUseThe ECOLAB POU Water Filters areintended to operate on EPA qualitydrinking water sources as a microbialretention filter. ECOLAB POU WaterFilters are suitable for the control ofbacteria equal to or greater in sizethan Brevundimonas diminuta. ThePOU Filters are suitable for generalpoint of use infection control forprocedures such as superficialwound cleansing, cleaning ofequipment, washing of surgeon'shands and bathing where thereduction of such microorganisms inthe water is desired.The POU Filters are not intended foruse in the production of USP sterilewater for use in infusion, injection orproduction of fluids for use in dialysistreatmentsThe NOSOGARD line of filters areintended to operate on EPA qualitydrinking water sources as a microbialretention filter. NOSOGARD filters areavailable in models suitable for thecontrol of bacteria equal to or greaterin size than Brevundimonas diminutaand models for microbial removalequal to or greater than Legionellapneumophila. The NOSOGARD issuitable for general point of useinfection control for procedures suchas cleaning, rinsing of equipment,hand washing and bathing where thereduction of such microorganisms inthe water is desired.The Nosogard is not intended for usein the production of USP sterile waterfor use in infusion, injection orproduction of fluids for use in dialysistreatments.
DeviceDescriptionA disposable Polyethersulfone (PES)membrane filter sealed in a whitepolypropylene housing designed tobe installed by facility personnel atpoint of use locations where thecontrol of bacteria in EPA qualitydrinking water is desired for hand-washing, bathing, showering andinstrument cleaning applications.These filters are designed to beinstalled as attachments to faucetsor as a handheld shower headattached to a water outlet.A disposable Polyethersulfone (PES)membrane filters sealed in apolypropylene housing designed to beinstalled by facility personnel at pointof use locations where the control ofbacteria in the drinking water supply isdesired for device cleaning,handwashing and bathingapplications. These filters are designedto be installed as attachments tofaucets or as a handheld shower head.
Feed WaterSourceSAMEDrinking water plumbing at point ofuse.
Feed WaterQualitySAMEPotable Water as defined by the EPANational Primary DrinkingWaterRegulations (EPA Quality Water)
Substantial Equivalence Comparison Table
Subject DevicePredicate Device
Useful LifeUp to 7 days and 31 days depending on modelUp to 7 days or 31 days or 62 days depending on model
Maximum InletPressure89.9 psi (6.2 bar): tap filters79.8 psi (5.5 bar): shower wands89.9 psi (pounds / square Inch) for all models
Sterile DeviceSAMEShipped in sterile condition and package
CasingSAMEPolypropylene
FilterSAMEPolyethersulfone (PES) pleated membrane
Element(s)SAME
BacteriaReduction*SAME> 107

{5}------------------------------------------------

Device Description & Summary of Technological Characteristics:

The ECOLAB POU Water Filter is a disposable polyethersulfone (PES) membrane filter sealed in a white polypropylene housing designed to be installed by facility personnel at point of use locations where the control of bacteria in EPA quality drinking water is desired for hand- washing, showering and instrument cleaning applications. These filters are designed to be installed as attachments to faucets or as a handheld shower head attached to a water outlet. The POU water filters are supplied sterile for single use and are not intended to be resterilized.

Model(REF)DescriptionMax. UseDaysTypeInletOutlet
60658457-Day Sterile Tap Filterw/Spray Outlet7TapQuickconnectSpray
60658467-Day Sterile Tap Filterw/Nozzle Outlet7TapQuickconnectNozzle
606570231-Day Sterile Tap Filterw/Spray Outlet31TapQuickconnectSpray
606570331-Day Sterile Tap Filterw/Nozzle Outlet31TapQuickconnectNozzle
606570431-Day Sterile Shower WandFilter31Showerwand½" NPTShower
Ecolab POU Filters Models and Maximum Service Life Once Installed

The ECOLAB POU Water Filters have the same intended use as the predicate device. The ECOLAB POU Water Filters and the predicate device are constructed of the same materials; both incorporate a 1µm polypropylene prefilter and a 0.2µm polyethersulfone pleated filter membrane. Both products are available in tap filter and shower wand configurations and have the same range of maximum service life from 7 days to 30 days. Both devices are supplied sterile in packaging capable of maintaining a sterile barrier for the labeled shelf life of three years.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the Ecolab logo in blue. The logo consists of the word "ECOLAB" in a bold, sans-serif font. The "O" in "ECOLAB" is replaced by a circular graphic with a starburst pattern inside. There is a registered trademark symbol to the right of the word "ECOLAB".

The predicate devices have models suitable for the control of bacteria equal to or greater insize than Legionella peumophila and models suitable for retention of the smaller Brevundimons diminuta bacteria. All ECOLAB POU Water Filters are suitable for retention of the smaller Brevundimons diminuta over the maximum service life of the model.

Non-Clinical Performance Testing

All ECOLAB POU Water Filters retained the test organism Brevundimonas diminuta challenge of ≥ 10′ organisms/cm² and prevented bacterial grow-through over the service life of each model when tested according to ASTM F838-15 Standard Test Method for Deterial Reterial Retention of Membrane Filters.

Filter flow rates at 3 bar inlet pressure are 7-Day tap filter – 6.3 L/min; 14-Day tap filter – 6.5 L/min; 31-Day tap filters - 6 L/min.

Simulated Use Cyclic Pulse Testing demonstrates that the tap filters maintain microbial retention effectiveness, as measured by diffusional flow integrity test, after continuous use repeated shocks of 89.9 psi (6.2 bar) at 151°F (66°C). The shower wand filters maintain microbial retention effectiveness, as measured by diffusional flow integrity test, after continuous use repeated shocks of 79.8 psi (5.5 bar) at 122°F (50°C).

The Ecolab POU Water Filters are provided sterile in sterile packaging. Testing according to ISO 11137-1 and ISO 11137-2, sterilization of health care products, verifies that the gamma irradiation process assures product sterility.

Package testing according to ISO 11607-1 and 11607-2, Packaging For Terminally Sterilized Medical Devices, has demonstrated that the packaging is capable of maintaining a sterile barrier for the labeled 36 month shelf life. Testing also verified that the filters retain their functional characteristics through 36 months.

Biocompatibility Testing Results

Biocompatibility of the Ecolab POU Water Filters was verified according the biocompatibility tests performed on devices after exposure to all manufacturing conditions and gamma sterilization. All biocompatibility testing was conducted in accordance to ISO 10993-5 and under GLP 21 CFR Part 58 Regulations:

Biological EffectApplicable EN ISO 10993 Standard
CytotoxicityPart 5: Tests for Cytotoxicity – In Vitro Methods
IrritationPart 10: Tests for Sensitization and Irritation
SensitizationPart 10: Tests for Sensitization and Irritation
Acute Systemic ToxicityPart 11: Tests for Systemic Toxicity
Material Mediated PyrogenicityPart 11: Tests for Systemic Toxicity

§ 876.5665 Water purification system for hemodialysis.

(a)
Identification. A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.(b)
Classification. Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.