K Number
K173155

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-06-07

(251 days)

Product Code
Regulation Number
868.5150
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Pakter Curved Needle Set is intended for discography, diagnostic sampling, aspiration, and injection.

Device Description

The Pakter Curved Needle Set is manufactured as a coaxial needle set and includes a preshaped curved needle and an introducer needle. The curved needle consists of a nitinol cannula connected to a polycarbonate hub at the proximal end and is characterized with a distal curve and a 30° bevel distal tip. It also includes a stainless steel stylet that is designed for a transitional fit. The curved needle with stylet is 25 gage in diameter and has a length of 20 or 15 centimeters. The introducer needle consists of a stainless steel cannula connected to an acrylic hub at the proximal end. The introducer needle is also designed with a stainless steel stylet. The introducer needle with stylet is 21 gage in diameter and has a length of 15 or 10 centimeters. When used as intended, the introducer needle and stylet are first inserted percutaneously and advanced adjacent to the target anatomy. After removing the introducer needle stylet, the curved needle is then inserted through the introducer needle cannula and slowly advanced to the target anatomy. The Pakter Curved Needle Set is sterilized by ethylene oxide and intended for one-time use.

AI/ML Overview

This document describes the premarket notification for the Pakter Curved Needle Set (K173155), a Class II medical device intended for discography, diagnostic sampling, aspiration, and injection. The submission aims to demonstrate substantial equivalence to a predicate device, the Atraumatic Standard Sprotte® Needle (K911260).

Here's an analysis of the provided information regarding acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance:

Test NameAcceptance CriteriaReported Device Performance
Tensile Strength - Curved Needle Cannula Hub-to-ShaftPeak tensile load ≥ 22 N (in accordance with BS EN ISO 7864:2016 and BS EN ISO 11070:2014)The acceptance criterion was met.
Tensile Strength - Curved Needle Stylet Hub-to-ShaftPeak tensile load ≥ 11 N (in accordance with BS EN ISO 7864:2016 and BS EN ISO 11070:2014)The acceptance criterion was met.
Tensile Strength - Introducer Needle Cannula Hub-to-ShaftPeak tensile load ≥ 44 N (in accordance with BS EN ISO 7864:2016 and BS EN ISO 11070:2014)The acceptance criterion was met.
Tensile Strength - Introducer Needle Stylet Hub-to-ShaftPeak tensile load ≥ 22 N (in accordance with BS EN ISO 7864:2016 and BS EN ISO 11070:2014)The acceptance criterion was met.
Stiffness - Introducer Needle Cannula and StyletDeflection < 0.45 mm when a force of 9.6 N is applied (in accordance with BS EN ISO 9626:2016)The acceptance criterion was met.
Resistance to Breakage - Curved Needle CannulaNo visible breakage after cyclic testing (in accordance with BS EN ISO 9626:2016)The acceptance criterion was met.
Resistance to Breakage - Introducer Needle CannulaNo visible breakage after cyclic testing (in accordance with BS EN ISO 9626:2016)The acceptance criterion was met.
Resistance to CorrosionNo visual signs of corrosion after testing (in accordance with Annex B of BS EN ISO 11070)The acceptance criterion was met.
Tensile Strength Post-Corrosion - Curved Needle Cannula Hub-to-Shaft(Same as pre-corrosion tensile strength) Peak tensile load ≥ 22 N (in accordance with BS EN ISO 7864:2016 and BS EN ISO 11070:2014)The acceptance criterion was met.
Tensile Strength Post-Corrosion - Curved Needle Stylet Hub-to-Shaft(Same as pre-corrosion tensile strength) Peak tensile load ≥ 11 N (in accordance with BS EN ISO 7864:2016 and BS EN ISO 11070:2014)The acceptance criterion was met.
Tensile Strength Post-Corrosion - Introducer Needle Cannula Hub-to-Shaft(Same as pre-corrosion tensile strength) Peak tensile load ≥ 44 N (in accordance with BS EN ISO 7864:2016 and BS EN ISO 11070:2014)The acceptance criterion was met.
Tensile Strength Post-Corrosion - Introducer Needle Stylet Hub-to-Shaft(Same as pre-corrosion tensile strength) Peak tensile load ≥ 22 N (in accordance with BS EN ISO 7864:2016 and BS EN ISO 11070:2014)The acceptance criterion was met.
Liquid Leakage Under Positive PressureMaintain a positive pressure of 300 kPa for 30 seconds without leakage (in accordance with BS EN ISO 11070:2014)The acceptance criterion was met.
Patency of LumenA stainless steel stylet with a diameter of 0.009 inches (0.23 mm) shall pass through the lumen of the curved needle cannula.The acceptance criterion was met.
Dimensional Analysis and CompatibilityCurved needle and introducer needle dimensions within specified tolerances and components are compatible.Testing conducted demonstrated the criteria were met.
Biocompatibility TestingDemonstrated biocompatibility (tests for cytotoxicity, sensitization, intracutaneous irritation, acute systemic toxicity, material mediated pyrogen, and hemolysis) per ISO 10993-1 and FDA guidance.Testing demonstrated the biocompatibility of the device.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

The document extensively lists bench testing conducted on the Pakter Curved Needle Set. However, it does not specify the sample size (i.e., number of units tested) for any of the bench tests described. The testing includes "time-zero and five-year accelerated aged testing," indicating a prospective evaluation of device performance over time. The document does not provide information on the country of origin of the data, but based on the FDA submission and reference to BS EN ISO standards, the testing was likely conducted in accordance with international quality standards, possibly in the US or a country adhering to these standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This section is not applicable to the provided document. The studies described are bench tests evaluating the physical and material properties of the device, not clinical performance or diagnostic accuracy that would require expert-established ground truth.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

This section is not applicable as the studies are bench tests, not clinical studies requiring adjudication of outcomes or interpretations.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This section is not applicable. The document describes a traditional 510(k) premarket notification for a physical medical device (needle set), not an AI-powered diagnostic or assistive technology. Therefore, an MRMC study related to AI performance is not relevant.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

This section is not applicable, as the device is not an algorithm or AI system.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for the bench tests described is based on established engineering and material science standards and specifications. For instance, tensile strength is measured against a quantitative N value, breakage is determined by visual inspection, and corrosion by visual signs after standardized exposure. These are objective physical measurements rather than clinical or diagnostic "ground truth."

8. The sample size for the training set

This section is not applicable. The device is a physical medical instrument, not a machine learning model, so there is no "training set."

9. How the ground truth for the training set was established

This section is not applicable. There is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

June 7, 2018

Cook Incorporated Johnathan Liu Regulatory Affairs Specialist 750 Daniels Way Bloomington, Indiana 47404

Re: K173155

Trade/Device Name: Pakter Curved Needle Set Regulation Number: 21 CFR 868.5150 Regulation Name: Anesthesia Conduction Needle Regulatory Class: Class II Product Code: BSP Dated: May 7, 2018 Received: May 8, 2018

Dear Johnathan Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Michael J. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K173155

Device Name Pakter Curved Needle Set

Indications for Use (Describe)

The Pakter Curved Needle Set is intended for discography, diagnostic sampling, aspiration, and injection.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is in large, white, sans-serif font, and the word "MEDICAL" is in a smaller, white, sans-serif font below the word "COOK".

2.0 510(k) SUMMARY

Pakter Curved Needle Set 21 CFR §807.92(c) Date Prepared: May 2, 2018

Submitted By:

Submission:Traditional 510(k) Premarket Notification
510(k) Number:K173155
Applicant:Cook Incorporated
Applicant Address:Cook Incorporated
750 Daniels Way
Bloomington, IN 47404
Contact:Johnathan Liu
Email:regsubmissions@cookmedical.com
Contact Phone Number:(812) 335-3575 x104509
Contact Fax Number:(812) 332-0281

Device Information:

Trade Name:Pakter Curved Needle Set
Common Name:Needle, Conduction, Anesthetic (W/Wo Introducer)
Classification Name:Anesthesia conduction needle
Regulation:21 CFR §868.5150
Product Code:BSP
Device Class:II
Classification Panel:Anesthesiology

Predicate Device:

The Pakter Curved Needle Set is considered substantially equivalent to the following device: Atraumatic Standard Sprotte® Needle (K911260, Pajunk GMBH) cleared on October 21, 1991.

{4}------------------------------------------------

Device Description:

The Pakter Curved Needle Set is manufactured as a coaxial needle set and includes a preshaped curved needle and an introducer needle. The curved needle consists of a nitinol cannula connected to a polycarbonate hub at the proximal end and is characterized with a distal curve and a 30° bevel distal tip. It also includes a stainless steel stylet that is designed for a transitional fit. The curved needle with stylet is 25 gage in diameter and has a length of 20 or 15 centimeters. The introducer needle consists of a stainless steel cannula connected to an acrylic hub at the proximal end. The introducer needle is also designed with a stainless steel stylet. The introducer needle with stylet is 21 gage in diameter and has a length of 15 or 10 centimeters. When used as intended, the introducer needle and stylet are first inserted percutaneously and advanced adjacent to the target anatomy. After removing the introducer needle stylet, the curved needle is then inserted through the introducer needle cannula and slowly advanced to the target anatomy. The Pakter Curved Needle Set is sterilized by ethylene oxide and intended for one-time use.

Indications for Use:

The Pakter Curved Needle Set is intended for discography, diagnostic sampling, aspiration, and injection.

Comparison to Predicate Device:

The Pakter Curved Needle Set and the predicate device, Atraumatic Standard Sprotte® Needle (K911260), are considered substantially equivalent in that these devices are identical in principles of operation and fundamental technologies. Additionally, the subject device has similar indications for use, coaxial design, materials, and dimensions as the predicate device.

The differences in technological characteristics between the subject device and the predicate device include the following:

  • . Materials - The subject device needle cannula is made of Nitinol, while the predicate device's cannula is made of stainless steel. The subject device's introducer needle cannula hub is acrylic, while the predicate device's hub is made of polycarbonate.
  • . Dimensions - The subject device's needle cannula and introducer needle cannula are longer in length than those of the predicate device: the subject device's needle cannula gage size is smaller than that of the predicate device; the needle cannula tip of the subject device is curved with a 30-degree bevel, while that of the predicate

{5}------------------------------------------------

device is straight with a pencil point bevel; and the introducer needle cannula gage size of the subject device is smaller than that of the predicate device.

  • . Additionally, the subject device set includes a needle cannula stylet, while the predicate device set does not.
    These differences in technological characteristics between the subject and predicate devices do not raise any different questions of safety and/or effectiveness.
PREDICATE DEVICESUBJECT DEVICE
Atraumatic Standard Sprotte®Needle(K911260)Pakter Curved Needle Set
Regulation21 CFR § 868.5150Identical
Product CodeBSPIdentical
ClassificationIIIdentical
Indications for UseTo gain entry into or puncture thespinal cavity permittinginjection/withdrawal of fluids forpurposes of diagnostic lumbarpuncture, myelography/discographyand chemonuceolysis procedures.The Pakter Curved Needle Set isintended for discography, diagnosticsampling, aspiration, and injection.
Device is for One-timeUseYesIdentical
Needle *CannulaMaterialStainless SteelNitinol
Cannula HubMaterialPolycarbonateIdentical
Cannula Length7 to 15 cm15, 20 cm
Cannula Size19.5 to 24 gage25 gage
Cannula TipStraight, Pencil pointCurved, 30° bevel
Stylet MaterialStainless SteelIdentical
Introducer NeedleCannulaMaterialStainless SteelIdentical
Cannula HubMaterialPolycarbonateAcrylic
Cannula Size0.7 to 1.0 mm21 gage (0.82 mm)
Cannula Length3, 4 cm10, 15 cm
Stylet MaterialNo styletStainless Steel
PackagingPouchPeel Pouch
Sterilization MethodEtOIdentical
Sterility AssuranceLevel10-6Identical

*Subject device needle is curved. Predicate device needle is not curved.

Indications for Use: The indications for use of the subject device is similar to that of the predicate device. Both the subject device and predicate device are indicated for discography. Although the wording is different, the subject device indications for

{6}------------------------------------------------

"diagnostic sampling, aspiration, and injection" are synonymous with the predicate device indications for "injection/withdrawal of fluids." The only difference between the subject device and predicate device intended use statements is that the subject device is not indicated for myelography and chemonucleolysis. The wording differences and exclusion of myelography and chemonucleolysis in the indications for use statement of the subject device do not raise different questions of safety and/or effectiveness.

Technological Characteristics:

The subject device. Pakter Curved Needle Set, underwent the following applicable testing to ensure reliable design and performance under the specified testing parameters:

Bench Testing (including time-zero and five-year accelerated aged testing)

  • Tensile Strength .
    • Curved Needle Cannula Hub-to-Shaft The peak tensile load for the hub-to-. shaft section of the curved needle cannula shall be greater than or equal to 22 N in accordance with BS EN ISO 7864:2016 and BS EN ISO 11070:2014. The acceptance criterion was met.
    • . Curved Needle Stylet Hub-to-Shaft – The peak tensile load for the hub-to-shaft section of the curved needle stylet shall be greater than or equal to 11 N in accordance with BS EN ISO 7864:2016 and BS EN ISO 11070:2014. The acceptance criterion was met.
    • Introducer Needle Cannula Hub-to-Shaft The peak tensile load for the hub-to-. shaft section of the introducer needle cannula shall be greater than or equal to 44 N in accordance with BS EN ISO 7864:2016 and BS EN ISO 11070:2014. The acceptance criterion was met.
    • Introducer Needle Stylet Hub-to-Shaft The peak tensile load for the hub-to-. shaft section of the introducer needle stylet shall be greater than or equal to 22 N in accordance with BS EN ISO 7864:2016 and BS EN ISO 11070:2014. The acceptance criterion was met.
  • . Stiffness - Introducer Needle Cannula and Stylet: The introducer needle cannula and stylet shall show a deflection of less than 0.45 mm when a force of 9.6 N is applied in accordance with BS EN ISO 9626:2016. The acceptance criterion was met.

{7}------------------------------------------------

  • . Resistance to Breakage
    • . Curved Needle Cannula - The curved needle cannula shall show no visible breakage after cyclic testing in accordance with BS EN ISO 9626:2016. The acceptance criterion was met.
    • . Introducer Needle Cannula – The introducer needle cannula shall show no visible breakage after cyclic testing in accordance with BS EN ISO 9626:2016. The acceptance criterion was met.
  • . Resistance to Corrosion - Following testing in accordance with the method given in Annex B of BS EN ISO 11070. the curved needle and introducer needle will be examined for visual signs of corrosion and the results will be reported. The acceptance criterion was met.
  • . Tensile Strength Post-Corrosion
    • Curved Needle Cannula Hub-to-Shaft The peak tensile load for the hub-to-. shaft section of the curved needle cannula shall be greater than or equal to 22 N in accordance with BS EN ISO 7864:2016 and BS EN ISO 11070:2014. The acceptance criterion was met.
    • . Curved Needle Stylet Hub-to-Shaft – The peak tensile load for the hub-to-shaft section of the curved needle stylet shall be greater than or equal to 11 N in accordance with BS EN ISO 7864:2016 and BS EN ISO 11070:2014. The acceptance criterion was met.
    • Introducer Needle Cannula Hub-to-Shaft The peak tensile load for the hub-to-. shaft section of the introducer needle cannula shall be greater than or equal to 44 N in accordance with BS EN ISO 7864:2016 and BS EN ISO 11070:2014. The acceptance criterion was met.
    • Introducer Needle Stylet Hub-to-Shaft The peak tensile load for the hub-to-. shaft section of the introducer needle stylet shall be greater than or equal to 22 N in accordance with BS EN ISO 7864:2016 and BS EN ISO 11070:2014. The acceptance criterion was met.
  • Liquid Leakage Under Positive Pressure The curved needle cannula shall maintain a . positive pressure of 300 kPa for 30 seconds without leakage in accordance with BS EN ISO 11070:2014. The acceptance criterion was met.
  • . Patency of Lumen – A stainless steel stylet with a diameter of 0.009 inches (0.23 mm) shall pass through the lumen of the curved needle cannula. The acceptance criterion was met.

{8}------------------------------------------------

  • . Dimensional Analysis and Compatibility - Testing conducted demonstrated that the curved needle and introducer needle dimensions are within the specified tolerances and the components are compatible.
    Biocompatibility Testing:

  • . Per ISO 10993-1 and FDA guidance, testing for cytotoxicity, sensitization, intracutaneous irritation, acute systemic toxicity, material mediated pyrogen, and hemolysis study demonstrated the biocompatibility of the subject device.

Conclusion:

The results of these tests confirm that the Pakter Curved Needle Set meets the design input requirements based on the intended use and support the conclusion that this device does not raise different questions of safety and/or effectiveness and is substantially equivalent to the predicate device, Atraumatic Standard Sprotte® Needle (K911260).

§ 868.5150 Anesthesia conduction needle.

(a)
Identification. An anesthesia conduction needle is a device used to inject local anesthetics into a patient to provide regional anesthesia.(b)
Classification. Class II (performance standards).