K Number
K172993

Validate with FDA (Live)

Date Cleared
2017-12-05

(69 days)

Product Code
Regulation Number
892.1550
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Digital Color Doppler Ultrasound System is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic (neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Muscular, Musculo-skeletal (Conventional and Superficial), Cardiac (pediatric and adult), Trans-esoph.(Cardiac), Laparoscopic, OB/Gyn and Urology.

Device Description

This SonoScape P20 series Digital Color Doppler Ultrasound System is an integrated preprogrammed color ultrasound imaging system, capable of producing high detail resolution intended for clinical diagnostic imaging applications. The basic principle is that system transmits ultrasonic energy into patient body and implements post processing of received echoes to generate onscreen display of anatomic structures and fluid flow within the body. This system is a Track 3 device that employs a wide array of probes that include linear array, convex array and phased array with a frequency range of 1.0 MHz to 25.0 MHz. This system consists of a mobile console with touch screen and keyboard control panel, power supply module, color LCD monitor and optional probes. This system is a mobile, general purpose, software controlled, color diagnostic ultrasound system. Its basic function is to acquire ultrasound data and to display the image in B-Mode (including Tissue Harmonic Image), M-Mode, TDI, Color-Flow Doppler, Pulsed Wave Doppler, Continued Wave Doppler and Power Doppler, or the combination of these modes, Elastography, 3D/4D.

AI/ML Overview

The provided text does not contain information about acceptance criteria or specific studies that prove the device meets these criteria in the typical sense of a clinical trial evaluating diagnostic performance (e.g., sensitivity, specificity, accuracy).

Instead, the document is a 510(k) summary for a SonoScape P20 series Digital Color Doppler Ultrasound System, focusing on demonstrating substantial equivalence to a predicate device (SonoScape P50 Digital Color Doppler Ultrasound System K170999).

The "studies" described are primarily non-clinical tests to ensure safety and performance according to established standards, rather than clinical efficacy or diagnostic accuracy studies.

Here's a breakdown of the requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state quantitative acceptance criteria for imaging performance (like sensitivity or specificity) for clinical applications. Instead, it states that the device was evaluated for its ability to meet design specifications and relevant safety and performance standards.

Acceptance Criteria (Implicit)Reported Device Performance
Compliance with IEC 60601-1:2005+A1:2012 (General safety)The P20 system met all design specifications and conformed to this standard.
Compliance with IEC 60601-1-2:2007 (EMC)The P20 system met all design specifications and conformed to this standard.
Compliance with IEC 60601-2-37:2007 (Ultrasound safety)The P20 system met all design specifications and conformed to this standard.
Compliance with ISO 10993-5:2009 (In vitro cytotoxicity)The P20 system met all design specifications and conformed to this standard.
Compliance with ISO 10993-10:2010 (Irritation & skin sensitization)The P20 system met all design specifications and conformed to this standard.
Compliance with AIUM/NEMA UD 2:2004 (R2009) (Acoustic output measurement)The P20 system met all design specifications and conformed to this standard.
Compliance with AIUM/NEMA UD 3:2004 (R2009) (Real-time acoustic output display)The P20 system met all design specifications and conformed to this standard.
Functional Equivalence to Predicate Device (SonoScape P50 K170999)The P20 Series has the same intended use and comparable basic and key technical features (design, operation controls, display modes, operation modes, measurement items, Cine Loop, Operating and Storage Condition) as the predicate device. New probes and biopsy brackets were compared to and found to be similar or previously cleared with other predicate devices. This demonstrates substantial equivalence in safety and effectiveness for diagnostic ultrasound imaging and fluid flow analysis. No specific performance metrics (e.g., image quality ratings, diagnostic accuracy) are provided in a quantitative format for clinical applications.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document states: "No clinical testing was required." The "test set" in this context refers to the non-clinical evaluations to confirm adherence to engineering and safety standards. There is no mention of a patient-based test set or its provenance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable, as no clinical testing was performed to establish diagnostic ground truth. The "ground truth" for the non-clinical tests would be the specifications of the standards (e.g., a certain level of electrical safety, a specific acoustic output).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable, as no clinical testing was performed requiring adjudication of diagnostic findings.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a diagnostic ultrasound system, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a diagnostic ultrasound system.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the non-clinical tests conducted, the "ground truth" or reference was the technical specifications of the device and the requirements of the referenced international and national standards (e.g., IEC 60601-1, ISO 10993-5, AIUM/NEMA UD 2).

8. The sample size for the training set

Not applicable. This document describes a traditional medical device (ultrasound system), not a machine learning algorithm that requires a training set of data.

9. How the ground truth for the training set was established

Not applicable. No training set was used in the context of machine learning. The "truth" for device design and verification is established through engineering principles and compliance with recognized standards.

{0}------------------------------------------------

December 5, 2017

Image /page/0/Picture/1 description: The image contains two logos. On the left is the Department of Health & Human Services logo, which features a stylized caduceus. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Sonoscape Medical Corp. % Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building, Yuquan Road, Nanshan Shenzhen 518051, GuangDong CHINA

Re: K172993

Trade/Device Name: P20 Series Digital Color Doppler Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: November 16, 2017 Received: November 24, 2017

Dear Toki Wu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael D. O'Hara For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K172993

Device Name

P20 Series Digital Color Doppler Ultrasound System

Type of Use (Select one or both, as applicable)

Indications for Use (Describe)

The Digital Color Doppler Ultrasound System is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic (neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Muscular, Musculo-skeletal (Conventional and Superficial), Cardiac (pediatric and adult), Trans-esoph.(Cardiac), Laparoscopic, OB/Gyn and Urology.

Over-The-Counter Use (21 CFR 801 Subpart C) X Prescription Use (Part 21 CFR 801 Subpart D)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including thetime to review instructions, search existing data sources, gather and maintain the data needed and completeand review the collection of information. Send comments regarding this burden estimate or any other aspectof this information collection, including suggestions for reducing this burden, to:
Department of Health and Human ServicesFood and Druq AdministrationOffice of Chief Information OfficerPaperwork Reduction Act (PRA) Staff

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

PRAStaff@fda.hhs.gov

EF

PSC Publishing Services (301) 443-6740

{3}------------------------------------------------

SonoScape P20, P20Pro, P15, P22, P25, P10 Plus System: Diagnostic Ultrasound Pulsed Echo System

Diagnostic Ultrasound Pulsed Doppler Imaging System

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalPPPPPNote 1Notes 2,4,5
AbdominalPPPPPNote 1Notes 2,4,5
Intra-operative Specify
Intra-operative Neuro
LaparoscopicPPPPPNote 1Notes 2,4
PediatricPPPPPNote 1Notes 2,4
Small Organ (specify)PPPPPNote 1Notes2,4,5,6,7
Neonatal CephalicPPPPPPNote 1Notes 2,3,4
FetalImaging&OtherAdult CephalicPPPPPPNote 1Notes 2,3,4
Trans-rectalPPPPPNote 1Notes 2,4,5
Trans-vaginalPPPPPNote 1Notes 2,4,5
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2,4
Musculo-skeletal(Superficial)PPPPPNote 1Notes 2,4
Intravascular
Other (Ob/GYN)PPPPPNote 1Notes 2,4,5
Other (Urology)PPPPPNote 1Notes 2,4,5
Cardiac AdultPPPPPPNote 1Notes 2,3,4
Cardiac PediatricPPPPPPNote 1Notes 2,3,4
CardiacIntravascular(Cardiac)
Trans-esoph.(Cardiac)PPPPPPNote 1Notes 2,3,4
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vesselPPPPPNote 1Notes 2,4
Cerebral vascularP

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

{4}------------------------------------------------

Transducer: C322 Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalImaging&OtherFetalPPPPPNote 1Notes 2,4
AbdominalPPPPPNote 1Notes 2,4
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)PPPPPNote 1Notes 2,4
Other (Urology)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
PeripheralVesselOther (specify)
Peripheral vessel
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

{5}------------------------------------------------

Transducer: 3C-A Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationSpecific(TRACKS 1 & 3)Mode of Operation
General(TRACK 1ONLY)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalPPPPPNote 1Notes 2,4
AbdominalPPPPPNote 1Notes 2,4
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)PPPPPNote 1Notes 2,4
Other (Urology)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D

Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

{6}------------------------------------------------

Transducer: L752 Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)NNNNNNNote 1Notes2,4,6,7
Neonatal Cephalic
FetalAdult Cephalic
lmaging&Trans-rectal
OtherTrans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)NNNNNNNote 1Notes 2,4
Musculo-skeletal(Superficial)NNNNNNNote 1Notes 2,4
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vesselNNNNNNNote 1Notes 2,4
VesselCerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{7}------------------------------------------------

Transducer: C1-6 Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalPPPPPNote 1Notes 2,4
AbdominalPPPPPNote 1Notes 2,4
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)PPPPPNote 1Notes 2,4
Other (Urology)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

N = new indication; P = previously cleared by FDA;

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{8}------------------------------------------------

Transducer: VC6-2 Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalPPPPPNote 1Notes 2,4,5
AbdominalPPPPPNote 1Notes 2,4,5
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)PPPPPNote 1Notes 2,4,5
Other (Urology)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

{9}------------------------------------------------

Transducer: C613 Micro-curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther* CombinedOther* Specify
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)Doppler
OphthalmicOphthalmic
Fetal
AbdominalPPPPPNote 1Notes 2,4
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
PediatricPPPPPNote 1Notes 2,4
Small Organ (specify)
Neonatal CephalicPPPPPPNote 1Notes 2,3,4
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
CardiacCardiac PediatricPPPPPPNote 1Notes 2,3,4
Intravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{10}------------------------------------------------

Transducer: 8P1 Phased Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
PediatricPPPPPNote 1Notes 2,4
Small Organ (specify)
FetalImaging&OtherNeonatal CephalicPPPPPPNote 1Notes 2,3,4
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
CardiacCardiac PediatricPPPPPPNote 1Notes 2,3,4
Intravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{11}------------------------------------------------

Transducer: S1-5 Phased Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
AbdominalPPPPPNote 1Notes 2,4
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal CephalicPPPPPPNote 1Notes 2,3,4
FetalAdult CephalicPPPPPPNote 1Notes 2,3,4
Imaging&Trans-rectal
OtherTrans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac AdultPPPPPPNote 1Notes 2,3,4
Cardiac PediatricPPPPPPNote 1Notes 2,3,4
Intravascular(Cardiac)
CardiacTrans-esoph. (Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 5: 4D

Note 3: TDI Note 4: 3D Note 6: Small Organ: breast, thyroid, testes

{12}------------------------------------------------

Transducer: 7P-B Phased Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
PediatricNNNNNNote 1Notes 2
Small Organ (specify)
Neonatal CephalicNNNNNNNote 1Notes 2
FetalAdult Cephalic
Imaging&Trans-rectal
OtherTrans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac PediatricNNNNNNNote 1Notes 2
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{13}------------------------------------------------

Transducer: L741 Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)NNNNNNote 1Notes 2,6
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card)
Musculo-skeletal(Conventional)NNNNNNote 1Notes 2
Musculo-skeletal(Superficial)NNNNNNote 1Notes 2
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vesselNNNNNNote 1Notes 2
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{14}------------------------------------------------

Transducer: L742 Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)PPPPPNote 1Notes 2,4,6
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2,4
Musculo-skeletal(Superficial)PPPPPNote 1Notes 2,4
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vesselPPPPPNote 1Notes 2,4
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{15}------------------------------------------------

Transducer: 1012 Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther*
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerCombinedSpecify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)PPPPPNote 1Notes 2,4,6
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2,4
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vesselPPPPPNote 1Notes 2,4
Cerebral vascular

P = previously cleared by FDA; E = added under this appendix N = new indication;

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{16}------------------------------------------------

Transducer: 9L-A Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)PPPPPNote 1Notes 2,4,6
Neonatal Cephalic
FetalAdult Cephalic
Imaging&Trans-rectal
OtherTrans-vaginal
Trans-urethral
Trans-esoph. (non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2,4
Musculo-skeletal(Superficial)PPPPPNote 1Notes 2,4
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vesselNNNNNNote 1Notes 2,4
VesselCerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{17}------------------------------------------------

Transducer: 12L-A Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)PPPPPNote 1Notes 2,4,6
Neonatal Cephalic
FetalAdult Cephalic
Imaging&Trans-rectal
OtherTrans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2,4
Musculo-skeletal(Superficial)PPPPPNote 1Notes 2,4
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vesselPPPPPNote 1Notes 2,4
VesselCerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{18}------------------------------------------------

Transducer: 6V1 Micro-curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectalPPPPPNote 1Notes 2,4
Trans-vaginalPPPPPNote 1Notes 2,4
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)PPPPPNote 1Notes 2,4
Cardiac Adult
Cardiac Pediatric
CardiacIntravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

{19}------------------------------------------------

Transducer: 6V3 Micro-curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectalPPPPPNote 1Notes 2,4
Trans-vaginalPPPPPNote 1Notes 2,4
Trans-urethral
Trans-esoph. (non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)PPPPPNote 1Notes 2,4
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

{20}------------------------------------------------

Transducer: 6V7 Micro-curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther*
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerCombinedSpecify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectalPPPPPNote 1Notes 2,4
Trans-vaginalPPPPPNote 1Notes 2,4
Trans-urethral
Trans-esoph. (non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)PPPPPNote 1Notes 2,4
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

{21}------------------------------------------------

Transducer: EC9-5 Micro-curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectalPPPPPNote 1Notes 2,4
Trans-vaginalPPPPPNote 1Notes 2,4
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)PPPPPNote 1Notes 2,4
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D

Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

{22}------------------------------------------------

Transducer: BCC9-5 Micro-curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectalPPPPPNote 1Notes 2,4
Trans-vaginalPPPPPNote 1Notes 2,4
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)PPPPPNote 1Notes 2,4
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{23}------------------------------------------------

Transducer: BCL10-5 Biplane Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectalPPPPPNote 1Notes 2,4
Trans-vaginalPPPPPNote 1Notes 2,4
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)PPPPPNote 1Notes 2,4
Cardiac Adult
Cardiac Pediatric
CardiacIntravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{24}------------------------------------------------

Transducer: VE9-5 Micro-curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
Adult Cephalic
FetalImaging&OtherTrans-rectalPPPPPNote 1Notes2,4,5
Trans-vaginalPPPPPNote 1Notes2,4,5
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)PPPPPNote 1Notes2,4,5
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

{25}------------------------------------------------

Transducer: MPTEE Phased Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalImaging&OtherFetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)PPPPPPNote 1Notes2,3,4
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

{26}------------------------------------------------

Transducer: MPTEE mini Phased Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalAdult Cephalic
Imaging&Trans-rectal
OtherTrans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)PPPPPPNote 1Notes2,3,4
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

{27}------------------------------------------------

Transducer: LAP7 Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
LaparoscopicPPPPPNote 1Notes 2,4
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalAdult Cephalic
Imaging&Trans-rectal
OtherTrans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{28}------------------------------------------------

Transducer: PWD2.0

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalAdult CephalicP
Imaging&Trans-rectal
OtherTrans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascularP

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{29}------------------------------------------------

Transducer: 12L-B Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)PPPPPNote 1Notes 2,4,6
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2,4
Musculo-skeletal(Superficial)PPPPPNote 1Notes 2,4
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
Peripheral vesselPPPPPNote 1Notes 2,4
PeripheralVesselCerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{30}------------------------------------------------

Transducer: 13L-A Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)PPPPPNote 1Notes 2,4,6
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2,4
Musculo-skeletal(Superficial)PPPPPNote 1Notes 2,4
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vesselPPPPPNote 1Notes 2,4
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI

Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

{31}------------------------------------------------

Transducer: 18L-A Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)PPPPPNote 1Notes 2,4,6
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2,4
Musculo-skeletal(Superficial)PPPPPNote 1Notes 2,4
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
Peripheral vesselPPPPPNote 1Notes 2,4
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI

Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

{32}------------------------------------------------

Transducer: 10L-I Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)PPPPPNote 1Notes 2,4,6
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2,4
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vesselPPPPPNote 1Notes 2,4
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

{33}------------------------------------------------

Transducer: VC2-9 Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalPPPPPNote 1Notes 2,4,5
AbdominalPPPPPNote 1Notes 2,4,5
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)PPPPPNote 1Notes 2,4,5
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{34}------------------------------------------------

Transducer: 6V3A Micro-curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginalPPPPPNote 1Notes 2,4
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{35}------------------------------------------------

Transducer: 12C-ER Micro-curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectalPPPPPNote 1Notes 2,4
Trans-vaginalPPPPPNote 1Notes 2,4
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)PPPPPNote 1Notes 2,4
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{36}------------------------------------------------

Transducer: 12LT-A Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalImaging&OtherFetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)PPPPPNote 1Notes 2,4,6
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2,4
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vesselPPPPPNote 1Notes 2,4
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{37}------------------------------------------------

Transducer: 12LI-A Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)PPPPPNote 1Notes 2,4,6
Neonatal Cephalic
FetalAdult Cephalic
Imaging&Trans-rectal
OtherTrans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2,4
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vesselPPPPPNote 1Notes 2,4
VesselCerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{38}------------------------------------------------

Transducer: 6CI-A Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalImaging&OtherFetalPPPPPNote 1Notes 2,4
AbdominalPPPPPNote 1Notes 2,4
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)PPPPPNote 1Notes 2,4
Other (Urology)
Cardiac Adult
CardiacCardiac Pediatric
Intravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{39}------------------------------------------------

Transducer: 6CT-A Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalPPPPPNote 1Notes 2,4
AbdominalPPPPPNote 1Notes 2,4
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)PPPPPNote 1Notes 2,4
Other (Urology)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

{40}------------------------------------------------

Transducer: ML3-18 Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)PPPPPNote 1Notes 2,4,6
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2,4
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vesselPPPPPNote 1Notes 2,4
VesselCerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{41}------------------------------------------------

Transducer: CWD5.0

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac AdultP
Cardiac PediatricP
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{42}------------------------------------------------

Transducer: 3P-A Phased Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
AbdominalNNNNNNote 1Notes 2,4
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal CephalicNNNNNNNote 1Notes 2,3,4
FetalAdult CephalicNNNNNNNote 1Notes 2,3,4
Imaging&Trans-rectal
OtherTrans-vaginal
Trans-urethral
Trans-esoph. (non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac AdultNNNNNNNote 1Notes 2,3,4
Cardiac PediatricNNNNNNNote 1Notes 2,3,4
Intravascular(Cardiac)
CardiacTrans-esoph. (Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{43}------------------------------------------------

510(k) Summary

1. Submitter [21 CFR807.92 (a) (1)]

Submitter:SONOSCAPE MEDICAL CORP.
Address:4/f, 5/f, 8/f, 9/f & 10/f, Yizhe Building, Yuquan Road,
Nanshan, Shenzhen 518051, Guangdong, China
Contact Person:Toki Wu
Tel:+86 755 26722890
Fax:+86 755 26722850
Email:ra@sonoscape.net
Date PreparedSep 11, 2017

2. Device [21 CFR807.92 (a) (2)]

Trade Name:P20 Series Digital Color Doppler Ultrasound System
Common Name: Diagnostic Ultrasound System and Transducers

Classification Regulatory:

FR NumberProduct Code
Ultrasonic Pulsed Doppler Imaging System892.155090-IYN
Ultrasonic Pulsed Echo Imaging System892.156090-IYO
Diagnostic Ultrasound Transducer892.157090-ITX

Classification Panel: Radiology Device Class: ll

3. Predicate Device(s) [21 CFR 807.92(a) (3)]

The identified predicate device within this submission is as follows: SonoScape P50 Digital Color Doppler Ultrasound System K170999

4. Device Description [21 CFR 807.92(a) (4)]

This SonoScape P20 series Digital Color Doppler Ultrasound System is an integrated preprogrammed color ultrasound imaging system, capable of producing high detail resolution intended for clinical diagnostic imaging applications.

The basic principle is that system transmits ultrasonic energy into patient body and implements post processing of received echoes to generate onscreen display of

{44}------------------------------------------------

anatomic structures and fluid flow within the body.

This system is a Track 3 device that employs a wide array of probes that include linear array, convex array and phased array with a frequency range of 1.0 MHz to 25.0 MHz.

This system consists of a mobile console with touch screen and keyboard control panel, power supply module, color LCD monitor and optional probes.

This system is a mobile, general purpose, software controlled, color diagnostic ultrasound system. Its basic function is to acquire ultrasound data and to display the image in B-Mode (including Tissue Harmonic Image), M-Mode, TDI, Color-Flow Doppler, Pulsed Wave Doppler, Continued Wave Doppler and Power Doppler, or the combination of these modes, Elastography, 3D/4D.

5. Intended Use [21 CFR 807.92(a) (5)]

The SonoScape P20 system is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic (neonatal and adult), Transrectal, Trans-vaginal, Peripheral Vascular, Cerebral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (pediatric and adult), Trans-esoph.(Cardiac), Laparoscopic, OB/Gyn and Urology.

6. Comparison with the Predicate device [21 CFR 807.92(a) (6)]

P20 Series Digital Color Doppler Ultrasound System is comparable with and substantially equivalent to the predicate device:

SonoScape P50 Digital Color Doppler Ultrasound System K170999

Intended Use Comparison:

Compared with the predicate device SonoScape P50 (K170999), the Subject Device P20 Series has the same intended use.

Technical Characteristics Comparison:

The basic and key technical features of the Subject Device P20 Series are the same as the predicate device SonoScape P50 (K170999), including Design, Operation Controls, Display Modes, Operation Modes, Measurement Items, Cine Loop, Operating and Storage Condition.

The detailed technical features can be found in General Device Descriptions of this submission.

Probes Comparison:

510(k) Summary

{45}------------------------------------------------

Subject device P50 Series has the similar probes as the predicate device SonoScape S45 (K160283).

ComparisonltemsPredicate deviceSonoScape P50 Exp/P50Pro/P50S/P50/P47/P40Exp/P40Pro/P40S/P40/P37/P30Exp/P30 Pro/P30S/P30/P27Subject deviceSonoScape P20, P20Pro, P15,P22, P25, P10 PlusRemark
C322 Micro-curved Array3C-A Curved ArrayC1-6 Curved ArrayC1-6A Curved Array6CT-A Curved Array6CI-A Curved ArrayC2-9 Curved ArrayC322 Micro-curved Array3C-A Curved ArrayC1-6 Curved Array6CT-A Curved Array6CI-A Curved Array
VC6-2 Curved ArrayC613 Micro-curved ArrayVC2-9A Curved ArrayVC2-9 Curved ArrayVC6-2 Curved ArrayC613 Micro-curved ArrayVC2-9 Curved Array
Probes8P1 Phased Array4P-A Phased ArrayS1-5 Phased Array7P-A Phased Array8P1 Phased ArrayS1-5 Phased Array7P-B Phased Array3P-A Phased ArraySEAnalysis1
L742 Linear Array1012 Linear Array9L-A Linear Array12L-A Linear ArrayVL12-5 Linear Array12L-B Linear Array13L-A Linear Array18L-A Linear Array10L-I Linear ArrayML3-18 Linear Array12LT-A Linear Array12LI-A Linear ArrayL742 Linear ArrayL752 Linear ArrayL741 Linear Array1012 Linear Array9L-A Linear Array12L-A Linear Array12L-B Linear Array13L-A Linear Array18L-A Linear Array10L-I Linear ArrayML3-18 Linear Array12LT-A Linear Array12LI-A Linear ArraySEAnalysis1

Table 1 Probes Comparison

{46}------------------------------------------------

6V1 Micro-curved Array6V1 Micro-curved Array
6V3 Micro-curved Array6V3 Micro-curved Array
6V3A Micro-curved Array6V3A Micro-curved Array
6V7 Micro-curved Array6V7 Micro-curved Array
EC9-5 Micro-curved ArrayEC9-5 Micro-curved Array
BCC9-5 Micro-curved ArrayBCC9-5 Micro-curved Array
BCL10-5 Biplane (Micro-curvedBCL10-5 Biplane (Micro-curved
+ Linear Array)+ Linear Array)
VE9-5 Micro-curved ArrayVE9-5 Micro-curved Array
12C-ER Micro-curved Array12C-ER Micro-curved Array
C3-10V Micro-curved Array
VE3-10 Micro-curved Array
PWD2.0 DopplerPWD2.0 Doppler
CWD5.0 CWCWD5.0 CW
CWD8.0 CWMPTEE Phased Array
CWD2.0 CWMPTEE mini Phased Array
MPTEE Phased ArrayLAP7 Linear Array
MPTEE mini Phased Array
LAP7 Linear Array

SE Analysis 1 :

Compared with the predicate device, there are four new Curved Array probes as followed:

(L752), which is cleared with predicate device SonoScape S45 (K160283). The new probe (L741, 3P-A, 7P-B) which is cleared with predicate device SonoScape X5 (K160258).

Biopsy Bracket Comparison:

Subject device P20 Series has the similar Biopsy Brackets as the predicate device SonoScape P50 (K170999).

Table 2 Biopsy Bracket Comparison
-----------------------------------------
ComparisonItemsSubject DeviceSonoScape P20, P20 Pro, P15,P22, P25, P10 PlusPredicate DeviceSonoScape P50 Exp/P50Pro/P50S/P50/P47/P40Exp/P40Pro/P40S/P40/P37/P30Exp/P30Pro/P30S/P30/P27(K170999)Remark
BiopsyBracketNGBC613NGB3C-ANGBC322NGBC613NGB3C-ANGBC322SEAnalysis 2

510(k) Summary

Page 4 of 7

{47}------------------------------------------------

ComparisonItemsSubject DeviceSonoScape P20, P20 Pro, P15,P22, P25, P10 PlusPredicate DeviceSonoScape P50 Exp/P50Pro/P50S/P50/P47/P40Exp/P40Pro/P40S/P40/P37/P30Exp/P30Pro/P30S/P30/P27(K170999)Remark
NGBC1-6NGBC1-6
NGBL752NGBC1-6A
NGBVC6-2NGBC2-9
NGBVC2-9NGBVC6-2
NGBVC2-9
NGBVC2-9A
NGBL742-2NGBL742-2
NGBL741NGB12L-A
NGB12L-ANGB9L-A
NGB9L-ANGB12L-B
NGB12L-BNGB13L-A
NGB13L-ANGB18L-A
NGB18L-ANGB10L-I
NGB10L-INGBML3-18
NGBML3-18NGBVL12-5
NGB4P-ANGB4P-A
NGBS1-5NGBS1-5
NGB7P-BNGB7P-A
NGB6V1NGB6V1
NGB6V3-2NGB6V3-2
NGB6V7NGB6V7
NGBBCL10-5NGBBCL10-5
NGBVE9-5NGBVE9-5
NGBBCC9-5NGBVE3-10
NGBEC9-5NGBC3-10V
NGBBCC9-5
NGBEC9-5

SE Analysis 2:

Compared with predicate device, there are three new biopsy brackets, NGBL752, NGBL741, NGB7P-B, the three biopsy brackets are not cleared with predicate device (P50 series), but they have been cleared with the device which is cleared by K150045. The detail information of all the seven biopsy brackets are described in table 6 as followed

{48}------------------------------------------------

BiopsybracketsModelAvailableprobemodelPatientContactingMaterialTypeApplicationKnumberDevice
NGBL741L741Stainlesssteel(SUS 304)ReusableBodysurfaceK150045ട് 30
NGB7P-B7P-BStainlesssteel(SUS 304)ReusableBodysurfaceK150045S30
NGBL752L752Stainlesssteel(SUS 304)ReusableBodysurfaceK150045S40
Table 3 Technical specifications of new Biopsy brackets
-----------------------------------------------------------------

7. Non-Clinical Tests [21 CFR 807.92(b) (1)]

The P20 Series Digital Color Doppler Ultrasound System has been evaluated for electrical, mechanical, thermal and electromagnetic compatibility safety,

biocompatibility and acoustic output.

Laboratory tests (including Phantom tests) were conducted to verify that the P20 system met all design specifications and the P20 system conformed to applicable medical device standards.

The P20 Series system has been designed and manufactured to meet the following standards:

IEC 60601-1:2005+A1:2012, Medical Electrical Equipment- Part 1: General requirements for basic safety and essential performance [08/20/2012];

IEC 60601-1-2:2007, Medical Electrical Equipment -Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility-Requirements and tests [03/30/2007];

IEC 60601-2-37:2007, Medical Electrical Equipment-Part 2-37: Particular requirements for basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment [08/09/2007];

ISO 10993-5:2009, Biological Evaluation of Medical Devices, Part 5-Tests for in vitro cytotoxicity [06/01/2009];

ISO 10993-10:2010, Biological Evaluation of Medical Devices- Part 10: Tests for irritation and skin sensitization [08/01/2010];

AIUM/NEMA UD 2:2004 (R2009), Acoustic output measurement standard for diagnostic

{49}------------------------------------------------

ultrasound equipment [08/21/2009]; and

AIUM/NEMA UD 3:2004 (R2009), Standard for real-time display of thermal and mechanical acoustic output indices on diagnostic ultrasound equipment [08/13/2010].

8. Clinical Test [21 CFR 807.92(b) (2)]

No clinical testing was required.

9. Substantially Equivalent Conclusions [21 CFR 807.92(b) (3)]

In accordance with the 21 CFR Part 807 and based on the information provided in this premarket notification, SONOSCAPE MEDICAL CORP. concludes that P20 Series Digital Color Doppler Ultrasound System is substantially equivalent to the predicate device with regard to safety and effectiveness.

§ 892.1550 Ultrasonic pulsed doppler imaging system.

(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.