K Number
K150045

Validate with FDA (Live)

Date Cleared
2015-04-28

(106 days)

Product Code
Regulation Number
892.1550
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SonoScape S30/S40 system is a general-purpose ultrasonic instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, thyroid), Cephalic (neonatal and adult), Transrectal, Trans-vaginal, Peripheral Vascular, Muscular, Musculo-skeletal (Conventional and Superficial), Cardiac (neonatal and adult), Trans-esoph.(Cardiac), Laparoscopic, OB/Gyn and Urology.

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Fetal, Abdominal, Pediatric, Small Organ (specify: breast, thyroid, testes), Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Musculo-skeletal (Conventional), Musculo-skeletal (Superficial), Other (Ob/GYN), Other (Urology), Cardiac Adult, Cardiac Pediatric, Trans-esoph.(Cardiac), Peripheral vessel, Cerebral vascular, Laparoscopic.

Device Description

This SonoScape S30/S40 Digital Color Doppler Ultrasound System is an integrated preprogrammed color ultrasound imaging system, capable of producing high detail resolution intended for clinical diagnostic imaging applications. The basic principle is that system transmits ultrasonic energy into patient body and implements post processing of received echoes to generate onscreen display of anatomic structures and fluid flow within the body. This system is a Track 3 device that employs a wide array of probes that include linear array, convex array and phased array with a frequency range of 2.0 MHz to 15.0 MHz. This system consists of a mobile console with touch screen and keyboard control panel, power supply module, color LCD monitor and optional probes. This system is a mobile, general purpose, software controlled, color diagnostic ultrasound system. Its basic function is to acquire ultrasound data and to display the image in B-Mode (including Tissue Harmonic Image), M-Mode, TDI, Color-Flow Doppler, Pulsed Wave Doppler, Continued Wave Doppler and Power Doppler, or the combination of these modes, Elastography, 3D/4D. The subject of this submission is the addition of new indications, new probes and special function to original SonoScape S30 and the addition of S40 model.

AI/ML Overview

This submission describes the SonoScape S30/S40 Digital Color Doppler Ultrasound System and its substantial equivalence to previously cleared predicate devices. The document does not contain information about acceptance criteria or a specific study proving the device meets those criteria, as one would typically find for a new, non-substantially equivalent device or an AI/ML product requiring performance evaluation. Instead, the submission focuses on demonstrating that the S30/S40 system is substantially equivalent to existing devices.

Therefore, many of the requested categories cannot be filled as they pertain to a performance study that was not conducted or detailed in this 510(k) summary. The submission relies on the concept of "substantial equivalence" to previously cleared devices.

Here's a breakdown of the information that can be extracted or is explicitly stated as not applicable from the provided text:

1. Table of Acceptance Criteria and Reported Device Performance
This information is not provided in the document. The submission is a 510(k) for substantial equivalence, not a performance study against predefined metrics. The "performance" is implicitly deemed equivalent to the predicate devices.

2. Sample Size Used for the Test Set and Data Provenance
Not applicable. No specific test set or data provenance for a performance study is described. The submission focuses on technical comparisons and compliance with standards.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
Not applicable. Ground truth for a performance study is not established or reported.

4. Adjudication Method for the Test Set
Not applicable. No performance study is detailed that would require an adjudication method.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
Not applicable. This submission does not describe an MRMC study or any AI components that would suggest one was performed. The device is an ultrasound system with various imaging modes, not an AI-powered diagnostic tool. The "AI" concept isn't mentioned in the context of improving human reader performance.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
Not applicable. This document describes an ultrasound system, not an AI algorithm.

7. Type of Ground Truth Used
Not applicable. As no clinical performance study for new capabilities is presented, no ground truth is discussed. The "clearance" is based on substantial equivalence to existing devices with established clinical use.

8. Sample Size for the Training Set
Not applicable. This document describes an ultrasound system, not an AI algorithm with a training set.

9. How the Ground Truth for the Training Set Was Established
Not applicable. This document describes an ultrasound system, not an AI algorithm with a training set.


Information that is provided in the document related to device evaluation:

The study that "proves the device meets the acceptance criteria" (in the context of a 510(k) for substantial equivalence) is the Non-Clinical Tests section, which confirms compliance with recognized standards and similarity to predicate devices.

Non-Clinical Tests (from Section 7):

  • Evaluation Focus: Electrical, mechanical, thermal and electromagnetic compatibility safety, biocompatibility, and acoustic output.
  • Strain Elastography Function: Laboratory tests were conducted to verify that the S30/S40 system and its strain Elastography function met all design specifications.
  • Standards Conformance: The S30/S40 system was designed and manufactured to meet the following standards:
    • IEC 60601-1
    • IEC 60601-1-2
    • IEC 60601-2-37
    • ISO 10993-5
    • ISO 10993-10
    • UD2
    • UD3

Clinical Test (from Section 8):

  • Requirement: "No clinical testing was required." This indicates that the FDA considered the non-clinical tests and comparison to predicate devices sufficient to demonstrate substantial equivalence, and no new clinical performance claims were being made that would necessitate a clinical study.

Substantial Equivalence Conclusion (from Section 9):

  • The manufacturer concludes that the S30/S40 Digital Color Doppler Ultrasound System is substantially equivalent to the predicate devices (SonoScape S30 K132527, SonoScape S40 K131213, and SonoScape S9 K142710) with regard to safety and effectiveness, based on the non-clinical tests and comparisons.

Summary of Device Performance (Implied by Substantial Equivalence):
The device is expected to perform comparably to the predicate devices mentioned (SonoScape S30 K132527, S40 K131213, and S9 K142710) for its specified indications for use across various probes and modes. The addition of new probes (C345, C613, 2P2, 3P1, 5P2, 8P1, 6V1A, CWD2.0, CWD5.0, PWD2.0, and LAP7) and the special function of Elastography were evaluated through non-clinical testing and also found to be substantially equivalent or not to raise new risks. The new indications for Cerebral Vascular and Laparoscopic use were also evaluated within this framework.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The caduceus is rendered in black, and the text is also in black.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 28, 2015

SonoScape Company Limited % Toki Wu Yizhe Building, Yuquan Road Nanshan, Shenzhen, 518051 P.R.CHINA

Re: K150045

Trade/Device Name: S30/S40 Digital Color Doppler Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: March 3, 2015 Received: March 30, 2015

Dear Toki Wu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert A Ochs

Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K150045

Device Name

S30/S40 Digital Color Doppler Ultrasound System

Indications for Use (Describe)

The SonoScape S30/S40 system is a general-purpose ultrasonic instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, thyroid), Cephalic (neonatal and adult), Transrectal, Trans-vaginal, Peripheral Vascular, Muscular, Musculo-skeletal (Conventional and Superficial), Cardiac (neonatal and adult), Trans-esoph.(Cardiac), Laparoscopic, OB/Gyn and Urology.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including thetime to review instructions, search existing data sources, gather and maintain the data needed and completeand review the collection of information. Send comments regarding this burden estimate or any other aspectof this information collection, including suggestions for reducing this burden, to:
Department of Health and Human ServicesFood and Druq AdministrationOffice of Chief Information OfficerPaperwork Reduction Act (PRA) StaffPRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

SonoScape S30/S40 System:

Diagnostic Ultrasound Pulsed Echo System

Diagnostic Ultrasound Pulsed Doppler Imaging System

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmicPPPPPNote 1Notes 2,4,5
FetalPPPPPNote 1Notes 2,4,5
AbdominalPPPPPNote 1Notes 2,4,5
Intra-operative Specify
Intra-operative Neuro
LaparoscopicNNNNNNote 1Notes 2,4
PediatricPPPPPNote 1Notes 2,4
Small Organ (specify)PPPPPNote 1Notes 2,4,6,7
Neonatal CephalicPPPPPPNote 1Notes 2,3,4
FetalImaging&OtherAdult CephalicPPPPPPNote 1Notes 2,3,4
Trans-rectalPPPPPNote 1Notes 2,4
Trans-vaginalPPPPPNote 1Notes 2,4
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2,4
Musculo-skeletal(Superficial)PPPPPNote 1Notes 2,4
Intravascular
Other (Ob/GYN)PPPPPNote 1Notes 2,4,5
Other (Urology)PPPPPNote 1Notes 2,4
Cardiac AdultPPPPPPNote 1Notes 2,3,4
Cardiac PediatricPPPPPPNote 1Notes 2,3,4
CardiacIntravascular(Cardiac)
Trans-esoph. (Cardiac)PPPPPPNote 1Notes 2,3,4
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vesselPPPNPPNote 1Notes 2,4
Cerebral vascularN

P = previously cleared by FDA; E = added under this appendix N = new indication;

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

Note 5: 4D

{4}------------------------------------------------

Transducer: C322 Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalPPPPPNote 1Notes 2,4
AbdominalPPPPPNote 1Notes 2,4
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)PPPPPNote 1Notes 2,4
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

{5}------------------------------------------------

Transducer: C344 Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalPPPPPNote 1Notes 2,4
AbdominalPPPPPNote 1Notes 2,4
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)PPPPPNote 1Notes 2,4
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

Note 4: 3D

Note 7: Elastography

{6}------------------------------------------------

Transducer: C345 Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalPPPPPNote 1Notes 2,4
AbdominalPPPPPNote 1Notes 2,4
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)PPPPPNote 1Notes 2,4
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

Note 4: 3D

Note 7: Elastography

{7}------------------------------------------------

Transducer: C353 Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*Combined
OphthalmicOphthalmic
FetalPPPPPNote 1Notes 2,4
AbdominalPPPPPNote 1Notes 2,4
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)PPPPPNote 1Notes 2,4
Other (Urology)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 4: 3D

Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{8}------------------------------------------------

Transducer: C542 Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
AbdominalPPPPPNote 1Notes 2,4
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
PediatricPPPPPNote 1Notes 2,4
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

Note 4: 3D

Note 7: Elastography

{9}------------------------------------------------

Transducer: VC6-2 Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalPPPPPNote 1Notes 2,4,5
AbdominalPPPPPNote 1Notes 2,4,5
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)PPPPPNote 1Notes 2,4,5
Other (Urology)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

{10}------------------------------------------------

Transducer: C613 Micro-curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
AbdominalNNNNNNote 1Notes 2,4
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
PediatricNNNNNNote 1Notes 2,4
Small Organ (specify)
Neonatal CephalicNNNNNNNote 1Notes 2,3,4
FetalAdult Cephalic
Imaging&Trans-rectal
OtherTrans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac PediatricNNNNNNNote 1Notes 2,3,4
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 4: 3D

Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{11}------------------------------------------------

Transducer: 2P1 Phase Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
AbdominalPPPPPNote 1Notes 2,4
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal CephalicPPPPPPNote 1Notes 2,3,4
FetalAdult CephalicPPPPPPNote 1Notes 2,3,4
Imaging&Trans-rectal
OtherTrans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac AdultPPPPPPNote 1Notes 2,3,4
Cardiac PediatricPPPPPPNote 1Notes 2,3,4
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{12}------------------------------------------------

Transducer: 2P2 Phase Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
AbdominalPPPPPNote 1Notes 2,4
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal CephalicPPPPPPNote 1Notes 2,3,4
FetalAdult CephalicPPPPPPNote 1Notes 2,3,4
Imaging&Trans-rectal
OtherTrans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac AdultPPPPPPNote 1Notes 2,3,4
Cardiac PediatricPPPPPPNote 1Notes 2,3,4
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{13}------------------------------------------------

Transducer: 3P1 Phase Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
AbdominalPPPPPNote 1Notes 2,4
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal CephalicPPPPPPNote 1Notes 2,3,4
FetalAdult CephalicPPPPPPNote 1Notes 2,3,4
Imaging&Trans-rectal
OtherTrans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac AdultPPPPPPNote 1Notes 2,3,4
Cardiac PediatricPPPPPPNote 1Notes 2,3,4
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{14}------------------------------------------------

Transducer: 5P1 Phase Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
PediatricPPPPPNote 1Notes 2,4
Small Organ (specify)
Neonatal CephalicPPPPPPNote 1Notes 2,3,4
FetalAdult Cephalic
Imaging&Trans-rectal
OtherTrans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac PediatricPPPPPPNote 1Notes 2,3,4
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 4: 3D

Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{15}------------------------------------------------

Transducer: 5P2 Phase Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
PediatricPPPPPNote 1Notes 2,4
Small Organ (specify)
Neonatal CephalicPPPPPPNote 1Notes 2,3,4
FetalAdult Cephalic
Imaging&Trans-rectal
OtherTrans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac PediatricPPPPPPNote 1Notes 2,3,4
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{16}------------------------------------------------

Transducer: 8P1 Phase Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
PediatricPPPPPNote 1Notes 2,4
Small Organ (specify)
Neonatal CephalicPPPPPPNote 1Notes 2,3,4
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac PediatricPPPPPPNote 1Notes 2,3,4
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

Note 4: 3D

Note 7: Elastography

{17}------------------------------------------------

Transducer: L741 Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)PPPPPNote 1Notes 2,4,6,7
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2,4
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph. (Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vesselPPPPPNote 1Notes 2,4
Cerebral vascular

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

Note 4: 3D

Note 7: Elastography

{18}------------------------------------------------

Transducer: L742 Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalImaging&OtherFetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)PPPPPNote 1Notes 2,4,6
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2,4
Musculo-skeletal(Superficial)PPPPPNote 1Notes 2,4
Intravascular
Other (Ob/GYN)
Other (Urology)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vesselPPPPPNote 1Notes 2,4
Cerebral vascular

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

Note 4: 3D

Note 7: Elastography

{19}------------------------------------------------

Transducer: L743 Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)PPPPPNote 1Notes 2,4,6
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2,4
Musculo-skeletal(Superficial)PPPPPNote 1Notes 2,4
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vesselPPPPPNote 1Notes 2,4
Cerebral vascular

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 4: 3D

Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{20}------------------------------------------------

Transducer: L752 Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)PPPPPNote 1Notes 2,4,6,7
Neonatal Cephalic
FetalAdult Cephalic
Imaging&Trans-rectal
OtherTrans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2,4
Musculo-skeletal(Superficial)PPPPPNote 1Notes 2,4
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vesselPPPPPNote 1Notes 2,4
VesselCerebral vascular

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

Note 4: 3D

Note 7: Elastography

{21}------------------------------------------------

Transducer: 10L1 Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)PPPPPNote 1Notes 2,4,6
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2,4
Musculo-skeletal(Superficial)PPPPPNote 1Notes 2,4
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vesselPPPPPNote 1Notes 2,4
PeripheralVesselCerebral vascular

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

Note 4: 3D

Note 7: Elastography

{22}------------------------------------------------

Transducer: 6V1 Micro-curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectalPPPPPNote 1Notes 2,4
Trans-vaginalPPPPPNote 1Notes 2,4
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)PPPPPNote 1Notes 2,4
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

Note 4: 3D

Note 7: Elastography

{23}------------------------------------------------

Transducer: 6V3 Micro-curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectalPPPPPNote 1Notes 2,4
Trans-vaginalPPPPPNote 1Notes 2,4
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)PPPPPNote 1Notes 2,4
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

Note 4: 3D

Note 7: Elastography

{24}------------------------------------------------

Transducer: 6V1A Micro-curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalAdult Cephalic
Imaging&Trans-rectal
OtherTrans-vaginalPPPPPNote 1Notes 2,4
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

Note 4: 3D

Note 7: Elastography

{25}------------------------------------------------

Transducer: EC9-5 Micro-curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectalPPPPPNote 1Notes 2,4
Trans-vaginalPPPPPNote 1Notes 2,4
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)PPPPPNote 1Notes 2,4
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 4: 3D

Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{26}------------------------------------------------

Transducer: BCC9-5 Micro-curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectalPPPPPNote 1Notes 2,4
Trans-vaginalPPPPPNote 1Notes 2,4
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)PPPPPNote 1Notes 2,4
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{27}------------------------------------------------

Transducer: BCL10-5 Biplane Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalImaging&OtherAdult Cephalic
Trans-rectalPPPPPNote 1Notes 2,4
Trans-vaginalPPPPPNote 1Notes 2,4
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)PPPPPNote 1Notes 2,4
Cardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Cerebral vascular

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 4: 3D

Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{28}------------------------------------------------

Transducer: MPTEE Phased Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalAdult Cephalic
Imaging&Trans-rectal
OtherTrans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
CardiacIntravascular(Cardiac)
Trans-esoph. (Cardiac)PPPPPPNote 1Notes 2,3,4
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 4: 3D

Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{29}------------------------------------------------

Transducer: MPTEE mini Phased Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalAdult Cephalic
Imaging&Trans-rectal
OtherTrans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac Pediatric
CardiacIntravascular(Cardiac)
Trans-esoph.(Cardiac)PPPPPPNote 1Notes 2,3,4
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 4: 3D

Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{30}------------------------------------------------

Transducer: CWD2.0

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalAdult Cephalic
Imaging&Trans-rectal
OtherTrans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac AdultN
Cardiac PediatricN
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

Note 4: 3D

Note 7: Elastography

{31}------------------------------------------------

Transducer: CWD5.0

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalAdult Cephalic
Imaging&Trans-rectal
OtherTrans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
Cardiac Adult
Cardiac PediatricN
Intravascular(Cardiac)
CardiacTrans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vesselN
VesselCerebral vascular

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 4: 3D

Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{32}------------------------------------------------

Transducer: PWD2.0

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalAdult CephalicN
Imaging&Trans-rectal
OtherTrans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascularN

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{33}------------------------------------------------

Transducer: LAP7 Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
LaparoscopicNNNNNNote 1Notes 2,4
Pediatric
Small Organ (specify)
Neonatal Cephalic
FetalAdult Cephalic
Imaging&Trans-rectal
OtherTrans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
Other (Urology)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselCerebral vascular

P = previously cleared by FDA; N = new indication; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 4: 3D

Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

{34}------------------------------------------------

510(k) Summary

1. Submitter [21 CFR807.92 (a) (1)]

Submitter:SonoScape Company Limited
Address:Yizhe Building, Yuquan Road, Nanshan, Shenzhen518051, P.R.China
Contact Person:Toki Wu
Tel:+86 755 26722890
Fax:+86 755 26722850
Email:wusq@sonoscape.net
Date PreparedJanuary 9, 2015

2. Device [21 CFR807.92 (a) (2)]

Trade Name:S30/S40 Digital Color Doppler Ultrasound System
Common Name:Diagnostic Ultrasound System and Transducers

Classification Regulatory:

FR NumberProduct Code
Ultrasonic Pulsed Doppler Imaging System892.155090-IYN
Ultrasonic Pulsed Echo Imaging System892.156090-IYO
Diagnostic Ultrasound Transducer892.157090-ITX

Classification Panel: Radiology Device Class: ll

3. Predicate Device(s) [21 CFR 807.92(a) (3)]

The identified predicate devices within this submission are as follows:
SonoScape S30 Digital Color Doppler Ultrasound SystemK132527
SonoScape S40 Digital Color Doppler Ultrasound SystemK131213
SonoScape S9 Portable Digital Color Doppler Ultrasound SystemK142710

4. Device Description [21 CFR 807.92(a) (4)]

This SonoScape S30/S40 Digital Color Doppler Ultrasound System is an integrated preprogrammed color ultrasound imaging system, capable of producing high detail resolution intended for clinical diagnostic imaging applications.

{35}------------------------------------------------

The basic principle is that system transmits ultrasonic energy into patient body and implements post processing of received echoes to generate onscreen display of anatomic structures and fluid flow within the body.

This system is a Track 3 device that employs a wide array of probes that include linear array, convex array and phased array with a frequency range of 2.0 MHz to 15.0 MHz.

This system consists of a mobile console with touch screen and keyboard control panel, power supply module, color LCD monitor and optional probes.

This system is a mobile, general purpose, software controlled, color diagnostic ultrasound system. Its basic function is to acquire ultrasound data and to display the image in B-Mode (including Tissue Harmonic Image), M-Mode, TDI, Color-Flow Doppler, Pulsed Wave Doppler, Continued Wave Doppler and Power Doppler, or the combination of these modes, Elastography, 3D/4D.

The subject of this submission is the addition of new indications, new probes and special function to original SonoScape S30 and the addition of S40 model.

New indications Cerebral Vascular and Laparoscopic.

Add C345, C613, 2P2, 3P1, 5P2, 8P1, 6V1A, CWD2.0, CWD5.0, New probes PWD2.0 and LAP7 probes.

Special function Elastography.

Added Model S40

5. Intended Use [21 CFR 807.92(a) (5)]

The SonoScape S30/S40 system is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic (neonatal and adult), Transrectal, Trans-vaginal, Peripheral Vascular, Cerebral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (neonatal and adult), Trans-esoph.(Cardiac), Laparoscopic, OB/Gyn and Urology.

6. Comparison with the Predicate Devices [21 CFR 807.92(a) (6)]

S30/S40 Digital Color Doppler Ultrasound System is comparable with and substantially equivalent to the predicate devices:

SonoScape S30 Digital Color Doppler Ultrasound SystemK132527
SonoScape S40 Digital Color Doppler Ultrasound SystemK131213
SonoScape S9 Portable Digital Color Doppler Ultrasound SystemK142710

{36}------------------------------------------------

Intended Use Comparison:

Compared with SonoScape S9 (K142710), the Subject Device S30/S40 has the same intended uses.

Technical Characteristics Comparison:

The basic and main technical features of the Subject Device S30/S40 are similar to SonoScape S9 (K142710) and the same as the Predicated Device original SonoScape S30 (K132527) and S40 (K131213) separately, including Design, Operation Controls, Display Modes, Measurement Items, Cine Loop and Operating and Storage Condition. For Operation Modes, Elastography is a special Operation Mode for the Subject Device S30/S40, but already employed by many marketed devices and same as the Predicate

Device SonoScape S9 (K142710).

The detailed technical features comparison and analysis can be found in Substantial Equivalence Comparison and the detailed technical features can be found in General Device Descriptions of the submission.

Probes Comparison:

Subject device S30/S40 has the similar probes as the predicated device SonoScape S9 (K142710) and original SonoScape S30 (K132527).

Subject devicePredicate DeviceRemark
SonoScape S30/S40SonoScape S9
C322 Micro-curved ArrayC322 Micro-curved Array
C344 Curved ArrayC344 Curved ArraySEAnalysis1
C345 Curved ArrayC353 Curved Array
C353 Curved ArrayC542 Curved Array
C542 Curved Array
VC6-2 Curved ArrayVC6-2 Curved ArraySame
C613 Micro-curved ArrayC613 Micro-curved ArraySame
2P2 Phased Array2P2 Phased Array
3P1 Phased Array3P1 Phased ArraySame
5P2 Phased Array5P2 Phased Array
8P1 Phased Array8P1 Phased Array
L741 Linear ArrayL741 Linear Array
L742 Linear ArrayL742 Linear Array
L743 Linear ArrayL743 Linear ArraySame
L752 Linear ArrayL752 Linear Array
10L1 Linear Array10L1 Linear Array
6V1 Micro-curved Array6V1 Micro-curved ArraySE
6V1A Micro-curved ArrayAnalysis1

Table 1 Probes Comparison

510(k) Summary

{37}------------------------------------------------

6V3 Micro-curved Array6V3 Micro-curved Array
EC9-5 Micro-curved ArrayEC9-5 Micro-curved Array
BCC9-5 Micro-curved ArrayBCC9-5 Micro-curved ArraySame
BCL10-5 Biplane (Micro-curvedBCL10-5 Biplane (Micro-curved +
+ Linear Array)Linear Array)
MPTEE Phased ArrayMPTEE Phased Array
(Multi-plane)(Multi-plane)Same
MPTEE mini Phased ArrayMPTEE mini Phased Array
(Multi-plane)(Multi-plane)
CWD2.0, 2.0 MHzCWD2.0, 2.0 MHz
CWD5.0, 5.0 MHzCWD5.0, 5.0 MHzSame
PWD2.0, 2.0 MHzPWD2.0, 2.0 MHz
LAP7, Linear ArrayLAP7, Linear ArraySame
Subject devicePredicate DeviceRemark
SonoScape S30/S40Original SonoScape S30
2P1 Phased Array2P1 Phased ArraySame
5P1 Phased Array5P1 Phased Array

SE Analysis 1:

Compared with the predicate device SonoScape S9 and original SonoScape S30, there are only two new probes (C345 and 6V1A) for the subject device, but no new intended use. And the frequency and performance of C345 and 6V1A probes is the same as C344 and 6V1 probes respectively for predicate device.

Moreover, compared with predicated devices, the subject device (S30/S40) complies with the same regulation and safety standards and has the consistent acoustic output levels. Therefore they can be considered Substantially Equivalent in safety and effectiveness, and no new risk is raised, so the SE is not affected.

7. Non-Clinical Tests [21 CFR 807.92(b) (1)]

The S30/S40 Digital Color Doppler Ultrasound System has been evaluated for electrical, mechanical, thermal and electromagnetic compatibility safety, biocompatibility and acoustic output.

Laboratory tests (including strain Elastography function) were conducted to verify that the S30/S40 system met all design specifications and the S30/S40 system conformed to applicable medical device standards. The S30/S40 system has been designed and manufactured to meet the following standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-5, ISO10993-10, UD2, and UD3.

{38}------------------------------------------------

8. Clinical Test [21 CFR 807.92(b) (2)]

No clinical testing was required.

9. Substantially Equivalent Conclusions [21 CFR 807.92(b) (3)]

In accordance with the 21 CFR Part 807 and based on the information provided in this premarket notification, SonoScape Company Limited concludes that S30/S40 Digital Color Doppler Ultrasound System is substantially equivalent to the predicate devices with regard to safety and effectiveness.

§ 892.1550 Ultrasonic pulsed doppler imaging system.

(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.