K Number
K172297

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-10-18

(79 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NovaCross™ Microcatheter is intended to be used in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature and for guidewire exchange.

Device Description

The NovaCross™ Microcatheter is a sterile, single-use, single lumen, over-the-wire, disposable percutaneous support catheter designed for use in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature for guidewire exchange.

The NovaCross™ Microcatheter consists of telescopic shaft, over tube, and a proximal handle that allows for manual device manipulation and a means for flushing the catheter lumen. A key element of the device is a temporarily deployable and retractable distal Nitinol Scaffold, which is visible through fluoroscopy when deployed by the user, and expands to the width of the artery to provide an anchoring to aid the user in establishing greater support near the treatment site.

AI/ML Overview

The provided text is a 510(k) summary for the NovaCross™ Microcatheter, focusing on demonstrating substantial equivalence to a predicate device. It explicitly states that the submission is a "Special 510(k) submission" to implement modifications to an already cleared device. This type of submission relies on performance testing to assure that the modified device remains as safe and effective as the predicate.

Here's an analysis of the acceptance criteria and study described:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not provide a formal table of explicit acceptance criteria with numerical targets. Instead, it describes various performance tests conducted to ensure the modified device functions similarly to the predicate. The "reported device performance" is generally stated as the tests being successfully completed and demonstrating similarity to the predicate.

Test PerformedAcceptance Criteria (Implied)Reported Device Performance
Simulated Use TestingDevice performs as intended during simulated use, effectively guiding, tracking, and exchanging guidewires."successfully completed" (implied, as it supports substantial equivalence)
BiocompatibilityMeets ISO 10993 standards for contact with blood and tissue (cytotoxicity, hemolysis, etc.).Passed based on biological risk assessment, cytotoxicity, and hemolysis testing.
Mechanical Tests:
- Bond StrengthSufficient bond strength to ensure device integrity during use."successfully completed" (implied)
- Visibility TestRadiopaque markers/features provide adequate visibility under fluoroscopy."successfully completed" (implied)
- Radial Force TestScaffold provides appropriate radial force for anchoring without causing damage."successfully completed" (implied)
- Axial Force TestDevice maintains structural integrity and pushability/trackability."successfully completed" (implied)
GLP Animal StudyDevice functions safely and effectively, demonstrating adequate scaffold deployment, anchoring, and maneuverability without adverse tissue reaction."successfully completed" (implied, to assess new scaffold and device characteristics)

2. Sample Size Used for the Test Set and the Data Provenance

The document does not specify the sample sizes used for any of the individual tests (simulated use, mechanical, biocompatibility, or the GLP animal study).

The data provenance is within the manufacturer's testing facilities or accredited labs, as it pertains to engineering and preclinical testing of a medical device intended for regulatory submission. There is no mention of country of origin for the data or whether it's retrospective or prospective in the context of human clinical data, as this is a device modification submission based on non-clinical performance.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

For this type of device modification submission, which relies on engineering and preclinical testing, the concept of "ground truth" established by human experts in a clinical context is not directly applicable. The "ground truth" for these tests is defined by engineering specifications, recognized standards (like ISO for biocompatibility), and animal model observation by veterinary and scientific staff.

Therefore, the document does not mention a number of experts or their qualifications for establishing ground truth in the way it would for an AI-based diagnostic device. The expertise would be in the form of engineers, toxicologists, veterinarians, and study directors overseeing the performance and animal studies.

4. Adjudication Method for the Test Set

Adjudication methods (like 2+1, 3+1) are typically used in clinical studies where multiple human readers or evaluators assess patient data (e.g., medical images) to establish a consensus ground truth. This is not applicable to the engineering and preclinical tests described for the NovaCross™ Microcatheter. The test results are objective measurements or observations defined by the test protocols.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This type of study assesses human reader performance, often with and without AI assistance, and is relevant for AI-powered diagnostic devices. The NovaCross™ Microcatheter is a physical interventional medical device, not a diagnostic imaging or AI tool.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

No, a standalone algorithm-only performance study was not done. This is relevant for AI algorithms. The NovaCross™ Microcatheter is a physical device that functions in conjunction with a steerable guidewire and is manipulated by a human operator (physician).

7. The Type of Ground Truth Used

The "ground truth" for the various performance tests is based on:

  • Engineering Specifications: For mechanical tests, the ground truth is adherence to design specifications and established performance parameters.
  • Industry Standards: For biocompatibility, the ground truth is compliance with recognized standards such as ISO 10993.
  • Physiological Response/Observation: For the GLP animal study, the ground truth is the objective observation of device performance in a living system and the histological assessment of tissue reaction.
  • Intended Use Definition: For simulated use, the ground truth is the successful demonstration of the device's ability to achieve its intended function (e.g., guidewire exchange, access discrete regions).

8. The Sample Size for the Training Set

Not applicable. The NovaCross™ Microcatheter is a physical medical device, not an AI or machine learning algorithm. Therefore, there is no "training set" in the context of AI model development.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below.

October 18, 2017

Nitiloop Ltd. % Mr. Michael Daniel President Daniel & Daniel Consulting, LLC 340 Jones Lane Gardnerville, Nevada 89460

Re: K172297

Trade/Device Name: NovaCross Microcatheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DOY Dated: September 14, 2017 Received: September 18, 2017

Dear Mr. Michael Daniel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.qov

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172297

Device Name NovaCross™ Microcatheter

Indications for Use (Describe)

The NovaCross™ Microcatheter is intended to be used in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature and for guidewire exchange.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for "Nitiloop". The logo consists of a red square with a white curved line inside, followed by the word "Nitiloop" in black bold letters. Below the word "Nitiloop" is the phrase "CROSSING THE LIMITS" in smaller, black letters.

510(k) SUMMARY NitiLoop NovaCross™ Microcatheter

Submitter:

Nitiloop Ltd. 4 Ha'Omanut Street P.O. Box 12892 Netanya, 4250438 Israel Tel: +972-54-664-5494 Fax: +972-72-2501631

Date Prepared: July 25, 2017

Contact Person:

Michael A. Daniel Daniel & Daniel Consulting, LLC 340 Jones Lane Gardnerville, NV 89460 Tel: 775.392.2970 Fax: 610.545.0799 madaniel@clinregconsult.com

or Orly Maor 25 Sirkin Street Kfar Saba 44421 Israel Tel: +972-9-7453607 oram.ma@gmail.com

Device:

Name of Device: Common or Usual Name: Classification Name: Regulatory Class: Product Code:

NovaCross™ Microcatheter NovaCross™ Microcatheter Percutaneous catheter (21 CFR 870.1250) II DQY

Predicate Device:

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for "Nitiloop". The logo consists of a red square with a white curved line inside, followed by the word "Nitiloop" in black, bold letters. Below the word "Nitiloop" is the phrase "CROSSING THE LIMITS" in smaller, black letters.

The predicate device is the NovaCross™ Microcatheter, cleared under K143608 and K160389.

Purpose of the Special 510(k) submission:

The NovaCross™ Microcatheter Special 510(k) was submitted to implement modifications to the NovaCross™ Microcatheter.

Intended Use:

The NovaCross™ Microcatheter is intended to be used in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature and for guidewire exchange.

Technological Characteristics:

The NovaCross™ Microcatheter is a sterile, single-use, single lumen, over-the-wire, disposable percutaneous support catheter designed for use in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature for guidewire exchange.

The NovaCross™ Microcatheter consists of telescopic shaft, over tube, and a proximal handle that allows for manual device manipulation and a means for flushing the catheter lumen. A key element of the device is a temporarily deployable and retractable distal Nitinol Scaffold, which is visible through fluoroscopy when deployed by the user, and expands to the width of the artery to provide an anchoring to aid the user in establishing greater support near the treatment site.

The NovaCross™ Microcatheter is similar in its design and it achieves its intended use by means of the same mechanisms as the predicate device. Minor modifications were implemented including the addition of a middle tube, a shorter handle, increasing the scaffold ID and OD, and increasing the device outer diameter.

Performance Data:

To evaluate the modifications to the NovaCross Microcatheter, the following tests were performed:

  • Simulated use testing
  • · Biocompatibility testing based upon a biological risk assessment, cytotoxicity testing and hemolysis testing

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for "Nitiloop". The logo consists of a red square with rounded corners and a white design inside, followed by the word "Nitiloop" in bold, black letters. Below the word "Nitiloop" is the phrase "CROSSING THE LIMITS" in smaller, black letters.

  • • Mechanical tests, including a bond strength test, visibility test, radial force test, and axial force test.
  • · GLP animal study test to assess the new scaffold and device characteristics

Substantial Equivalence:

The modified NovaCross™ Microcatheter has the same intended use and similar indications, principles of operation, and technological characteristics as the cleared NovaCross™M Microcatheter. The minor differences in the handle design and in the materials do not raise any new questions of safety or effectiveness. Performance data demonstrates that the modified NovaCross™ Microcatheter is similar to the cleared NovaCross™ Microcatheter. Thus, the modified NovaCross™ Microcatheter is substantially equivalent to its predicate device.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).