K Number
K160389

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-03-11

(29 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NovaCross™ Microcatheter is intended to be used in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature and for guidewire exchange.

Device Description

The NovaCross™ Microcatheter is a sterile, single-use, single lumen, over-the-wire, disposable percutaneous support catheter designed for use in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature for guidewire exchange.

The NovaCross™ Microcatheter consists of telescopic shaft, over tube, and a proximal handle that allows for manual device manipulation and a means for flushing the catheter lumen. A key element of the device is a temporarily deployable and retractable distal Nitinol Scaffold, which is visible through fluoroscopy when deployed by the user, and expands to the width of the artery to provide anchoring to aid the user in establishing greater support near the treatment site.

AI/ML Overview

This document is a 510(k) summary for the Nitiloop NovaCross™ Microcatheter, indicating it is a modified version of a previously cleared device (K143608). As such, the focus is on demonstrating substantial equivalence to the predicate device rather than presenting a comprehensive study proving novel acceptance criteria.

The acceptance criteria are implicitly defined by maintaining key performance characteristics to be "as safe and effective" as the predicate device, despite minor modifications. The study performed is a series of engineering tests designed to ensure these characteristics are maintained.

Here's a breakdown of the requested information based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

Given that this is a special 510(k) for a modified device, the "acceptance criteria" are implied to be maintaining the performance of the predicate device. The reported device performance is that it met these implied criteria.

Acceptance Criteria (Implied: Performance equivalent to predicate device K143608)Reported Device Performance (NovaCross™ Microcatheter K160389)
Simulated Use: Maintain performance after assembly, sterilization, transportation, and accelerated aging.Test performed at baseline, Time 0 (after sterilization/transportation), and Time 1 (after 1 year accelerated aging/transportation). The document states performance data demonstrates the modified device is "as safe and effective as the predicate."
Tensile Strength: Maintain strength after assembly, sterilization, transportation, and accelerated aging.Test performed at baseline, Time 0, and Time 1. The document states performance data demonstrates the modified device is "as safe and effective as the predicate."
Kink Resistance: Maintain resistance to kinking after assembly, sterilization, transportation, and accelerated aging.Test performed at baseline, Time 0, and Time 1. The document states performance data demonstrates the modified device is "as safe and effective as the predicate."
Tip Flexibility: Maintain flexibility after assembly, sterilization, transportation, and accelerated aging.Test performed at baseline, Time 0, and Time 1. The document states performance data demonstrates the modified device is "as safe and effective as the predicate."

Note: The document explicitly states, "The following tests are the same as those submitted in the original submission," implying that the methods and success criteria for these tests were already established and met by the predicate device. The current submission confirms these tests were repeated and the modified device performed equivalently.


2. Sample Size Used for the Test Set and the Data Provenance

  • Sample Size for Test Set: The document does not specify the exact sample sizes (N values) for each of the mechanical performance tests (Simulated use, Tensile Strength, Kink Resistance, Tip Flexibility). It only states that the tests were performed.
  • Data Provenance: The tests would have been performed by Nitiloop Ltd. or a contracted testing facility. The data provenance is retrospective in the sense that it's comparing a modified device's performance to an already-cleared predicate, but the specific testing conducted for the K160389 submission is prospective with respect to that submission. The country of origin of the data is implied to be Israel, given the submitter's address, or wherever the contract testing was conducted.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

This type of information is not applicable to this submission. This is a submission for a physical medical device (microcatheter) and the "tests" refer to engineering and mechanical performance assessments, not clinical studies or AI algorithm validation that would require expert-established ground truth. The "ground truth" here is the physical performance measurement itself, compared against pre-defined engineering specifications.


4. Adjudication Method for the Test Set

Not applicable. As noted above, these are objective engineering tests, not clinical assessments requiring adjudication. The results are quantitative measurements against established specifications.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No. This is a clearance for a physical medical device, not an AI algorithm, and therefore an MRMC study is not relevant or included in this document.


6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

Not Applicable. This device is a microcatheter, a physical instrument. There is no AI algorithm involved in the function or clearance of this device.


7. The Type of Ground Truth Used

For the engineering tests (Simulated use, Tensile Strength, Kink Resistance, Tip Flexibility), the "ground truth" is based on pre-defined engineering specifications and performance characteristics established for the predicate device. The modified device's performance is measured and compared to confirm it meets these specifications, ensuring it is "as safe and effective" as the predicate. This is based on objective measurement data rather than expert consensus, pathology, or outcomes data, which are more relevant for diagnostic or therapeutic efficacy claims.


8. The Sample Size for the Training Set

Not applicable. This device is a physical microcatheter, not an AI model requiring a training set.


9. How the Ground Truth for the Training Set was Established

Not applicable. As there is no AI model or training set, this question is irrelevant.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing right, suggesting a focus on people and health.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 11, 2016

Nitiloop Ltd. % Michael Daniel President Daniel & Daniel Consulting, LLC 340 Jones Lane Gardnerville, Nevada 89460

Re: K160389

Trade/Device Name: Novacross Microcatheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: February 10, 2016 Received: February 11, 2016

Dear Michael Daniel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use Statement

DEPARTMENT OF HEALTH AND HUMAN SERVICESForm Approved: OMB No. 0910-0120
Food and Drug AdministrationExpiration Date: January 31, 2017
Indications for UseSee PRA Statement on last page

510(k) Number (if known) K160389

Device Name NovaCross™ Microcatheter

Indications for Use (Describe)

The NovaCross™ Microcatheter is intended to be used in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature and for guidewire exchange.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number. "

FORM FDA 3881 (1/14)

Page 1 of 1 FDA

PSC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

Attachment 4

Image /page/3/Picture/3 description: The image shows the logo for "Nitiloop CROSSING THE LIMITS". The logo consists of a red square with a white curved line inside, followed by the word "Nitiloop" in black, and the phrase "CROSSING THE LIMITS" in smaller black letters below the word "Nitiloop".

510(k) SUMMARY NitiLoop's NovaCross™ Microcatheter

Submitter:

  • Nitiloop Ltd. 4 Haomanut Street P.O. Box 12892 Netanya, 4250438 Israel Tel: +972-54-664-5494 Fax: +972-72-2501631

Date Prepared: February 3, 2016

Contact Person:

Michael A. Daniel Daniel & Daniel Consulting, LLC 340 Jones Lane Gardnerville, NV 89460 Tel: 775.392.2970 Fax: 610.545.0799 madaniel(@clinregconsult.com

Back-up Company Contact:

Orly Maor 25 Sirkin Street Kfar Saba 44421 Israel Tel: +972-9-7453607 oram.ma@gmail.com

Device:

Name of Device: Common or Usual Name: Classification Name: Regulatory Class: Product Code:

NovaCross™ Microcatheter NovaCross™ Microcatheter Percutaneous catheter (21 CFR 870.1250) II DQY

Predicate Device:

{4}------------------------------------------------

The predicate device is the NovaCross™ Microcatheter, cleared under K143608.

Purpose of the Special 510(k) notice:

The NovaCross™ Microcatheter is a modification to NovaCross™ Microcatheter.

Intended Use:

The NovaCross™ Microcatheter is intended to be used in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature and for guidewire exchange.

Technological Characteristics:

The NovaCross™ Microcatheter is a sterile, single-use, single lumen, over-the-wire, disposable percutaneous support catheter designed for use in conjunction with a steerable guidewire to access discrete regions of the coronary and peripheral vasculature for guidewire exchange.

The NovaCross™ Microcatheter consists of telescopic shaft, over tube, and a proximal handle that allows for manual device manipulation and a means for flushing the catheter lumen. A key element of the device is a temporarily deployable and retractable distal Nitinol Scaffold, which is visible through fluoroscopy when deployed by the user, and expands to the width of the artery to provide anchoring to aid the user in establishing greater support near the treatment site.

The NovaCross™ Microcatheter is similar in its design and it achieves its intended use by means of the same mechanisms as the predicate device.

Performance Data:

The following tests are the same as those submitted in the original submission:

  • Simulated use -the test was performed at baseline (immediately following assembly, prior to sterilization), Time 0 (after sterilization and simulated transportation), and Time 1 (after accelerated aging of 1 year and simulated transportation).
  • -Tensile Strength- The test was performed at baseline (immediately following assembly, prior to sterilization), Time 0 (after sterilization and simulated transportation), and Time 1 (after accelerated aging of 1 year and simulated transportation).

{5}------------------------------------------------

  • -Kink resistance- The test was performed at baseline (immediately following assembly, prior to sterilization), Time 0 (after sterilization and simulated transportation), and Time 1 (after accelerated aging of 1 year and simulated transportation).
  • -Tip Flexibility- the test was performed at baseline (immediately following assembly, prior to sterilization), Time 0 (after sterilization and simulated transportation), and Time 1 (after accelerated aging of 1 year and simulated transportation).

Substantial Equivalence:

The modified NovaCross™ Microcatheter has the same intended use and indications, principles of operation, and technological characteristics as the cleared NovaCrossTM Microcatheter. The minor differences in the handle design and in the PEBAX material do not raise any new or different questions of safety or effectiveness. Performance data demonstrates that the modified NovaCross™ Microcatheter is as safe and effective as the predicate NovaCross™ Microcatheter. Thus, the NovaCross™ Microcatheter is substantially equivalent to its predicate device.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).