K Number
K171720

Validate with FDA (Live)

Device Name
encevis
Date Cleared
2018-04-19

(311 days)

Product Code
Regulation Number
882.1400
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use
  1. encevis is intended for the review, monitoring and analysis of EEG recordings made by electroencephalogram (EEG) devices using scalp electrodes and to aid neurologists in the assessment of EG. This device is intended to be used by qualified medical practitioners who will exercise professional judgment in using the information.

  2. The seizure detection component of encevis is intended to mark previously acquired sections of adult (greater than or equal to 18 years) EEG recordings that may correspond to electrographic seizures, in order to assist qualified clinical practitioners in the assessment of EEG traces. EEG recordings should be obtained with a full scalp montage according to the standard 10/20-system.

  3. The spike detection component of encevis is intended to mark previously acquired sections of the patient's EEG recordings that may correspond to spikes, in order to assist qualified clinical practitioners in the assessment of EEG traces. The Spike Detection component is intended to be used in adult patients greater than or equal to 18 years. encevis Spike Detection performance has not been assessed for intracranial recordings.

  4. encevis includes the calculation and display of a set of quantitative measures intended to monitor and analyze the EEG waveform. These include frequency bands, rhythmic and periodic patterns and burst suppression. These quantitative EEG measures should always be interpreted in conjunction with review of the original EEG waveforms.

  5. The aEEG functionality included in encevis is intended to monitor the state of the brain.

  6. encevis provides notifications on an on-screen display for seizure detection, spike detection, quantitative EEG and a EG that can be used when processing a record during acquisition. Delays of up to several minutes can occur between the beginning of a seizure, the occurrence of a spike or detection of quantitative EEG features and when the encevis notifications will be shown to a user. encevis notifications cannot be used as a substitute for real time monitoring of the underlying EEG by a trained expert.

  7. encevis PureEEG (Artifact Reduction) is intended to reduce artifacts in a standard 10-20 EEG recording. PureEEG does not remove the entiract signal, and is not effective for other types of artifacts. PureEEG may modify portions of waveforms representing cerebral activity. Waveforms must still be read by a qualified medical practitioner trained in recognizing artifact, and any interpretation or diagnosis must be made with reference to the original waveforms.

  8. This device does not provide any diagnostic conclusion about the patient's condition to the user.

Device Description

encevis combines several modalities for viewing and analyzing EEG data in one integrated software package. Encevis consists of the following modalities: encevis EEG-Viewer, encevis artifact reduction (PureEEG), encevis seizure detection (EpiScan), encevis spike detection (EpiSpike), encevis rhythmic and periodic patterns, encevis aEEG, encevis frequency bands, encevis Burst Suppression.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the studies performed to prove the device meets these criteria, based on the provided document. The information is organized according to your requested points.

encevis Device Performance Study Summary

1. Table of Acceptance Criteria and Reported Device Performance

The document presents performance metrics for various components of the encevis device, often comparing them to a predicate device (Persyst). The acceptance criteria are implicitly defined by the non-inferiority margins used in statistical testing.

Feature (Component)Acceptance Criterion (Implicit)Reported Device Performance (encevis)Comparison to Predicate (Persyst)
Seizure Detection (EpiScan)Non-inferiority to predicate device (Persyst) with margins: - PPA: 10% - NDR: 1 (false detections/24 hours)Average PPA: 75.40% (95% CI=[64.5, 86.3]) Average NDR: 7.01 false detections in 24 hours (95% CI=[5.9, 8.2])PPA is non-inferior (Predicate PPA: 75.94% (95% CI=[65.5, 86.4])) NDR is non-inferior (Predicate NDR: 10.61 false detections in 24 hours (95% CI=[6.8, 14.5]))
Spike Detection (EpiSpike)Non-inferiority to predicate device (Persyst) with margins: - PPA: 3% - NPA: 3% - PLPA: 3%Average PPA: 61.02% (95% CI=[47.9-74.1]) Average NPA: 98.58% (95% CI=[97.9-99.3]) Average PLPA: 96.09% (95% CI=[88.8-103.4])PPA is non-inferior (Predicate PPA: 8.7% (95% CI=[4.4-13.0])) NPA is non-inferior (Predicate NPA: 99.69% (95% CI=[99.4-99.9])) PLPA is non-inferior (Predicate PLPA: 93.97% (95% CI=[83.6-104.3]))
Artifact Reduction (PureEEG)Non-inferiority to predicate device (Persyst) with margins: - Relative suppression of true EEG: 1 dB (delta CI = [-0.09, -0.04]) - Signal-to-noise ratio (SNR) after artifact removal: 0.10 (delta CI = [4.78, 6.60])Relative suppression of clean EEG lower (better) than predicate in 82/111 cases. SNR after artifact removal higher (better) than predicate in 75/80 cases.Both parameters found non-inferior.
Rhythmic and Periodic Patterns (NeuroTrend)Sensitivity and Specificity for pattern detection: (No explicit acceptance criteria stated as thresholds, but performance shown relative to expert annotations)ANY (all patterns): Sensitivity: 77.59% (95% CI=[75.5-79.7]) Specificity: 86.50% (95% CI=[85.8-87.2]) PD: Sensitivity: 63.37% (95% CI=[60.7-66.0]) Specificity: 96.57% (95% CI=[96.2-96.9]) ARA (RTA, RAA, SW): Sensitivity: 92.72% (95% CI=[88.2-97.2]) Specificity: 94.76% (95% CI=[94.4-95.2]) RDA: Sensitivity: 92.56% (95% CI=[87.5-97.7]) Specificity: 90.44% (95% CI=[89.9-91.0])Not directly compared to a predicate for these metrics. Inter-reader agreement for localization was moderate (Kappa=0.45, CI=0.43-0.48).
aEEG (NeuroTrend)Functional equivalence to proposed method (Zhang and Ding, 2013) and good accordance with predicate (Persyst)Frequency response very similar to published version; both hemispheres show same characteristic; stop band suppression -30dB or higher; slope in pass band approx. -12dB/decade. Good accordance with Persyst and raw EEG values.Demonstrated good accordance with predicate (Persyst)
Frequency Bands (NeuroTrend)Correct assignment to frequency bands (Delta, Theta, Alpha, Beta) with measurement error for amplitudes below 5%; correctly identifies globally dominant background frequency.Correctly assigns test signals to corresponding frequency bands; measurement error for amplitudes below 5%. Relative proportion corresponding to true frequency band > 50%.Not directly compared to a predicate.
Burst Suppression (NeuroTrend)Validated by sensitivity, specificity, PPV, NPV relative to expert consensus.Sensitivity: 87.28% (95% CI=[90.8, 92.3]) Specificity: 92.16% (95% CI=[91.3, 92.9]) PPV: 61.09% (95% CI=[57.9, 64.2]) NPV: 98.09% (95% CI=[97.6, 98.5]) ACC: 91.56% (95% CI=[90.8,92.3])Not directly compared to a predicate.

2. Sample Size Used for the Test Set and Data Provenance

  • Seizure Detection:

    • Sample Size: 55 subjects. 50 patients with seizure events, 5 subjects without epilepsy.
    • Data Provenance: Scalp-EEG recordings from video-EEG monitoring in an epilepsy monitoring unit. Patients were 18 years of age or older. The specific country of origin is not stated but context suggests a European context (AIT Austrian Institute of Technology). The data is retrospective.
    • Total EEG data reviewed: 1619 hours of EEG, with a maximum of 30 hours per subject.
  • Spike Detection:

    • Sample Size: 23 patients. 18 subjects (>=18 years) with spike events, 5 subjects (>=18 years) diagnosed with no epilepsy.
    • Data Provenance: Scalp-EEG recordings from video-EEG monitoring in an epilepsy monitoring unit. The specific country of origin is not stated but context suggests a European context. The data is retrospective.
  • Artifact Reduction:

    • Sample Size: 128 EEG data records. 60 records from epilepsy monitoring units (31 seizure segments from 31 subjects, 33 spike segments from 6 subjects) and 65 from ICU patients (65 subjects).
    • Data Provenance: From different patient groups, covering adult patients in epilepsy monitoring and in critical care. The specific country of origin is not stated but context suggests a European context. The data is retrospective. Each record was 10 seconds of data.
  • Rhythmic and Periodic Patterns:

    • Sample Size: 83 long-term EEGs.
    • Data Provenance: Prospectively recorded from ICU-patients at two different centers. The specific country of origin is not stated but context suggests a European context.
    • Total Annotation Segments: 11935 common annotation segments (first minute of each hour, split into three 20-second segments).
  • aEEG (NeuroTrend):

    • Sample Size: Not explicitly stated for real EEG data comparison with Persyst, beyond "real EEG data were used."
    • Data Provenance: Not explicitly stated.
  • Frequency Bands (NeuroTrend):

    • Sample Size: Not explicitly stated for manually selected EEGs from epilepsy or ICU patients, beyond "Each of these EEG samples is representative..."
    • Data Provenance: Manually selected from epilepsy or ICU patients.
  • Burst Suppression (NeuroTrend):

    • Sample Size: 83 long-term EEGs.
    • Data Provenance: Recorded from intensive care patients from two different centers. The specific country of origin is not stated but context suggests a European context.
    • Total Annotation Segments: 3978 valid annotation segments (first minute of each hour).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Seizure Detection: 3 independent neurologists for blinded review. Qualifications not explicitly detailed (e.g., years of experience).
  • Spike Detection: 3 independent neurologists for blinded review. Qualifications not explicitly detailed.
  • Artifact Reduction: 3 independent epileptologists or neurologists for blinded review. Qualifications not explicitly detailed.
  • Rhythmic and Periodic Patterns: 2 clinical neurophysiologists who were naive to these EEGs. Qualifications not explicitly detailed.
  • Burst Suppression: 2 clinical neurophysiologists who were naive to these EEGs. Qualifications not explicitly detailed.

4. Adjudication Method for the Test Set

  • Seizure Detection: An event was considered a "true seizure" if the time interval of two out of three reviewers overlapped by at least 1 second. The seizure epoch was defined as the overlapping time range of these two reviewers. This is a 2+1 consensus method.
  • Spike Detection: An event was considered a "true spike" only if the time interval of two out of three reviewers overlapped. For localization, the 3D-coordinates of the electrode next to the spike maximum, averaged over reviewers, were used. This implies a 2+1 consensus for detection and a form of consensus/averaging for localization.
  • Artifact Reduction: The document states that "annotations of clean EEG recordings without any artifacts, and moreover annotations of artifacts that can be superimposed to the clean recordings" were needed, and "three independent epileptologists or neurologists for blinded review of the EEG data" were engaged. It doesn't explicitly state the adjudication method (e.g., 2+1, 3+1) for these annotations, but likely implies a consensus approach to establish ground truth about "clean" vs "artifact" segments.
  • Rhythmic and Periodic Patterns: Annotations had to be consistent between both reviewers to be used in the sensitivity and specificity measurement. This implies 100% agreement between 2 reviewers.
  • Burst Suppression: The performance was analyzed for consensus annotations of the two reviewers, meaning annotation segments where both reviewers showed the same decision. This implies 100% agreement between 2 reviewers.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC study comparing human readers with AI assistance versus without AI assistance was not explicitly conducted or reported. The studies focused on the standalone performance of the AI components and their non-inferiority to a predicate device (another software), not on the improvement of human readers with AI assistance.

6. If a Standalone Performance (Algorithm Only Without Human-in-the-Loop Performance) Was Done

Yes, standalone performance was assessed for all tested components. The clinical performance data presented (seizure detection, spike detection, artifact reduction, rhythmic and periodic patterns, aEEG, frequency bands, burst suppression) evaluated the algorithm's output against established ground truth without a human in the loop interpreting the AI's results. For instance, seizure detection compared the algorithm's detected seizure time points to the expert consensus.

7. The Type of Ground Truth Used

  • Seizure Detection: Expert consensus (2 out of 3 neurologists).
  • Spike Detection: Expert consensus (2 out of 3 neurologists).
  • Artifact Reduction: Expert annotations (epileptologists/neurologists) for clean EEG and artifact patterns.
  • Rhythmic and Periodic Patterns: Expert consensus (2 clinical neurophysiologists with 100% agreement).
  • aEEG: Based on the proposed method of Zhang and Ding (2013) for frequency response, and comparison to another FDA-approved software (Persyst) for real EEG data.
  • Frequency Bands: Based on predefined frequency borders (Delta, Theta, Alpha, Beta) and manual selection of EEG recordings representative of specific background-EEG-frequency bands by experts.
  • Burst Suppression: Expert consensus (2 clinical neurophysiologists with 100% agreement) for burst suppression patterns.

8. The Sample Size for the Training Set

The document does not explicitly provide the sample size for the training set for any of the algorithms. It often refers to "large amount of EEG data from different centers" for bench testing, but does not distinguish between training and test data or specify training set sizes.

9. How the Ground Truth for the Training Set Was Established

Similarly, the document does not explicitly describe how ground truth for the training set was established, as it does not detail the training process or the data used for it. The focus is on the validation/test set and its ground truth establishment.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA logo, which is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 1, 2018

AIT Austrian Institute of Technology GmbH % Arne Briest CEO VISAMED GmbH Kastellstr, 8 Karlsruhe, D-76227 De

Re: K171720

Trade/Device Name: encevis Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OMB, OLT, OMA Dated: July 24, 2017 Received: July 26, 2017

Dear Arne Briest:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely, Michael J. Hoffmann -S

for

Carlos L. Peña. PhD. MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K171720

Device Name encevis

Indications for Use (Describe)

  1. encevis is intended for the review, monitoring and analysis of EEG recordings made by electroencephalogram (EEG) devices using scalp electrodes and to aid neurologists in the assessment of EG. This device is intended to be used by qualified medical practitioners who will exercise professional judgment in using the information.

  2. The seizure detection component of encevis is intended to mark previously acquired sections of adult (greater than or equal to 18 years) EEG recordings that may correspond to electrographic seizures, in order to assist qualified clinical practitioners in the assessment of EEG traces. EEG recordings should be obtained with a full scalp montage according to the standard 10/20-system.

  3. The spike detection component of encevis is intended to mark previously acquired sections of the patient's EEG recordings that may correspond to spikes, in order to assist qualified clinical practitioners in the assessment of EEG traces. The Spike Detection component is intended to be used in adult patients greater than or equal to 18 years. encevis Spike Detection performance has not been assessed for intracranial recordings.

  4. encevis includes the calculation and display of a set of quantitative measures intended to monitor and analyze the EEG waveform. These include frequency bands, rhythmic and periodic patterns and burst suppression. These quantitative EEG measures should always be interpreted in conjunction with review of the original EEG waveforms.

  5. The aEEG functionality included in encevis is intended to monitor the state of the brain.

  6. encevis provides notifications on an on-screen display for seizure detection, spike detection, quantitative EEG and a EG that can be used when processing a record during acquisition. Delays of up to several minutes can occur between the beginning of a seizure, the occurrence of a spike or detection of quantitative EEG features and when the encevis notifications will be shown to a user. encevis notifications cannot be used as a substitute for real time monitoring of the underlying EEG by a trained expert.

  7. encevis PureEEG (Artifact Reduction) is intended to reduce artifacts in a standard 10-20 EEG recording. PureEEG does not remove the entiract signal, and is not effective for other types of artifacts. PureEEG may modify portions of waveforms representing cerebral activity. Waveforms must still be read by a qualified medical practitioner trained in recognizing artifact, and any interpretation or diagnosis must be made with reference to the original waveforms.

  8. This device does not provide any diagnostic conclusion about the patient's condition to the user.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/2 description: The image shows the logo for the Austrian Institute of Technology. The logo consists of the letters "AIT" in a stylized font, with the words "AUSTRIAN INSTITUTE OF TECHNOLOGY" in smaller letters to the right of the letters. The letters and words are in a dark red color. The logo is simple and modern, and it is likely used on the organization's website, marketing materials, and other communications.

510k Summary encevis

1. Submission Sponsor and Application Correspondent

A. Submission Sponsor

AIT Austrian Institute of Technology GmbH Donau-City-Straße 1 1220 Vienna – Austria Phone: +43 50550-4203 Fax: +43 50550-4125

B. Submission Correspondent

VISAMED GmbH Arne Briest CEO Phone:+49 721 476 4847 Cell: +49 170 4888 498 Fax: +49 721 476 4849 Email: arne.briest@visamed.com

2. Date Prepared

April 18th, 2018

{5}------------------------------------------------

Image /page/5/Picture/2 description: The image shows the logo for the Austrian Institute of Technology (AIT). The logo consists of the letters "AIT" in a gray, sans-serif font. To the right of the letters, the words "AUSTRIAN INSTITUTE OF TECHNOLOGY" are written in a smaller, red, sans-serif font. The logo is simple and modern, and the colors are muted.

3. Device Identification

Trade/Proprietary Name:encevis
Common Name:Electroencephalograph
Classification Regulation:21CFR882.1400
Product Code:OMB, OLT, OMA
Class:II
Panel:Neurology

4. Legally Marketed Predicate Devices

Primary Predicate:K132306 Persyst 12 EEG Review and Analysis Software
Additional Predicate:K151929 Persyst 13 EEG Review and Analysis Software

{6}------------------------------------------------

Image /page/6/Picture/2 description: The image shows the logo for the Austrian Institute of Technology (AIT). The logo consists of the letters "AIT" in a stylized, geometric font in gray. To the right of the letters, the words "AUSTRIAN INSTITUTE OF TECHNOLOGY" are written in a smaller, sans-serif font in a dark red color. The logo is simple and modern, and the colors are muted and professional.

5. Device Description

encevis combines several modalities for viewing and analyzing EEG data in one integrated software package. Encevis consists of the following modalities:

encevis EEG-Viewer

The encevis EEG-viewer is intended for the review and the analysis of EEG-recordings that were recorded with an electroencephalography device using scalp electrodes. It shall aid the user in the examination of EEG-recordings. This includes frequency filtering of data, scaling of data in x and y direction and visualization in different montages. In addition, the encevis EEG-viewer can start several modules that automatically analyze the EEG and present the results in form of markers or in the form of modified EEG-curves. All included modules are intended for the user in the examination and monitoring of EEG-recordings.

encevis artifact reduction (PureEEG)

The artefact reduction encevis PureEEG is an analysis module that automatically recognizes and reduces artefacts in the EEG-data that come from EMG or electrode artefacts.

encevis seizure detection (EpiScan)

The seizure detection encevis EpiScan is a module for the automatic marking of areas in the EEG that could correspond to epileptic seizures with electrographic correlate. EpiScan makes the results available to the user in form of marker in a marker list. This analysis can take place in parallel to the recording (ad-hoc) or after the recording finished (post-hoc).

encevis spike detection (EpiSpike)

The spike detection encevis EpiSpike is a module for the automatic marking of areas in the EEG that could correspond to spikes or spike-waves. A graphical user interface presents the reser. The user interface contains a time line per channel, a list of spike clusters that contain spikes and a list of spikes contained in a selected cluster. In addition, either the EEG or the averaged EEG 0.5 seconds before the spike maximum to 0,5 seconds after spike maximum for all spikes in a selected cluster is shown. This post-hoc analysis can take place in parallel to the recording finished.

encevis rhythmic and periodic patterns

encevis rhythmic and periodic patterns is a feature for the analysis of EEG-recordings. It automatically detects EEG-patterns defined in the Standardized Critical Care EEG Terminology of the American Clinical Neurophysiology Society (Hirsch, L.J., et al., 2013. American Clinical Neurophysiology Society's Standardized Critical Care EEG Terminology: 2012 version. J. Clin. Neurophysiol. 30. 1-27) and graphically presents the results to the user. Additionally, it detects and visualizes rhythmic patterns with frequencies of up to 12Hz. It serves as a support during the examination of EEG-recordings in the ICU. This post-hoc analysis can take place in parallel to the recording or after the recording finished.

encevis aEEG

Encevis aEEG calculate and visualize a continuous measure that describes the EEG by showing the aEEG as defined in "Zhang, D., Ding, H., 2013. Calculation of compact amplitude-integrated EEG tracing and upper and lower margins using raw EEG data. Health (N. Y.) 05, 885-891. doi:10.4236/health.2013.55116".

{7}------------------------------------------------

Image /page/7/Picture/2 description: The image shows the logo for the Austrian Institute of Technology (AIT). The logo consists of the letters "AIT" in a stylized, sans-serif font in gray. To the right of the letters, the words "AUSTRIAN INSTITUTE OF TECHNOLOGY" are written in a smaller, sans-serif font in a dark red color. The logo is simple and modern, and the colors are muted and professional.

encevis frequency bands

Encevis frequency bands calculate and visualize a continuous measure that describe the EEG by showing the frequency distribution of the EEG. The relative proportions of the four frequency bands Delta, Theta, Alpha, and Beta are shown. The intensity of the colors corresponds to the amplitudes in these four frequency bands.

encevis Burst Suppression

Encevis burst suppression feature calculates and visualizes suppression rate and suppression time in percent. It calculates periods of burst suppression in the EEG and marks them by vertical bars with red color. The definition of burst suppression patterns follows the guidelines of the American Clinical Neurophysiology Society ICU EEG Terminology (Hirsch, L.J., LaRoche, S.M., Gaspard, N., Gerard, E., Svoronos, A., Herman, S.T., Mani, R., Arif, H., Jette, N., Minazad, Y., Kerrigan, J.F., Vespa, P., Hantus, S., Claassen, J., Young, G.B., So. E., Kaplan, P.W., Nuwer, M.R., Fountain, N.B., Drislane, F.W., 2013, American Clinical Neurophysiology Society's Standardized Critical Care EEG Terminology: 2012 version. J. Clin. Neurophysiol. 30, 1-27. doi:10.1097/WNP.0b013e3182784729).

{8}------------------------------------------------

Image /page/8/Picture/2 description: The image shows the logo for the Austrian Institute of Technology. The logo consists of the letters "AIT" in a stylized, sans-serif font. To the right of the letters is the text "AUSTRIAN INSTITUTE OF TECHNOLOGY" in a smaller, sans-serif font. The letters and text are in a dark red color.

Indication for Use Statement 6.

  1. encevis is intended for the review, monitoring and analysis of EEG recordings made by electroencephalogram (EEG) devices using scalp electrodes and to aid neurologists in the assessment of EEG. This device is intended to be used by qualified medical practitioners who will exercise professional judqment in usinq the information.

  2. The seizure detection component of encevis is intended to mark previously acquired sections of adult (greater than or equal to 18 years) EEG recordings that may correspond to electrographic seizures, in order to assist qualified clinical practitioners in the assessment of EEG traces. EEG recordings should be obtained with a full scalp montage according to the standard 10/20-system.

  3. The spike detection component of encevis is intended to mark previously acquired sections of the patient's EEG recordings that may correspond to spikes, in order to assist qualified clinical practitioners in the assessment of EEG traces. The Spike Detection component is intended to be used in adult patients greater than or equal to 18 years. encevis Spike Detection performance has not been assessed for intracranial recordings.

  4. encevis includes the calculation and display of a set of quantitative measures intended to monitor and analyze the EEG waveform. These include frequency bands, rhythmic and periodic patterns and burst suppression. These quantitative EEG measures should always be interpreted in conjunction with review of the original EEG waveforms.

  5. The aEEG functionality included in encevis is intended to monitor the state of the brain.

  6. encevis provides notifications on an on-screen display for seizure detection, spike detection, quantitative EEG and aEEG that can be used when processing a record during acquisition. Delays of up to several minutes can occur between the beginning of a seizure, the occurrence of a spike or detection of quantitative EEG features and when the encevis notifications will be shown to a user. encevis notifications cannot be used as a substitute for real time monitoring of the underlying EEG by a trained expert.

  7. encevis PureEEG (Artifact Reduction) is intended to reduce EMG and electrode artifacts in a standard 10-20 EEG recording. PureEEG does not remove the entifact signal, and is not effective for other types of artifacts. PureEG may modify portions of waveforms representing cerebral activity. Waveforms must still be read by a qualified medical practitioner trained in recognizing artifact, and any interpretation or diagnosis must be made with reference to the original waveforms.

  8. This device does not provide any diagnostic conclusion about the patient's condition to the user.

{9}------------------------------------------------

Image /page/9/Picture/2 description: The image shows the logo for the Austrian Institute of Technology (AIT). The logo consists of the letters "AIT" in a stylized, geometric font. The letters are in gray. To the right of the letters, the words "AUSTRIAN INSTITUTE OF TECHNOLOGY" are written in a smaller, sans-serif font. The words are in a dark red color.

7. Substantial Equivalence Discussion

The following table compares the encevis to the predicate device with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.

encevisPersyst 12Persyst 13
(primary predicate)(additional predicate)
510k Reference(subject device)K132306K151929
Product CodeOMBOMBOMB
Additional CodesOLT, OMAOLT, OMAOLT, OMA
ClassIIIIII
Regulation Number21CFR882.140021CFR882.140021CFR882.1400
Regulation NameElectroencephalographElectroencephalographElectroencephalograph
ManufacturerAIT Austrian Institute ofPersyst DevelopmentPersyst Development
Technology GmbHCorporationCorporation
General Device DescriptionEEG Review and AnalysisEEG Review and AnalysisEEG Review and Analysis
SoftwareSoftwareSoftware
Shows EEGYESYESYES
Has Artefact reductionYESYESYES
Identifies seizuresYESYESYES
Identifies spikesYESYESYES
Calculates quantitative EEGYESYESYES
measures
Calculated EEG measuresYESYESYES
displayed
Online interface to MobergYESNONO
monitor
Type of EEG-AnalysisPost-hoc analysisPost-hoc analysisPost-hoc analysis
Type of EEGScalp EEGScalp EEGScalp EEG
Population ageAdults (age > 18)Adults (age >18);Spike detection for ages > 1month.
UserThis device is intended to beThis device is intended to beThis device is intended to be
used by qualified medicalused by qualified medicalused by qualified medical
practitioners who will exercisepractitioners who will exercisepractitioners who will exercise
professional judgment in usingprofessional judgment in usingprofessional judgment in using
the information.the information.the information.
Input FilesDisplay and calculation basedDisplay and calculation isDisplay and calculation is
on EEG data recorded bybased on EEG data recordedbased on EEG data recorded
external EEG systems. Theyby external EEG systems.by external EEG systems.
are either read from the EEG-They are read from the EEG-They are read from the EEG-
file provided by the EEGfile provided by the EEGfile provided by the EEG
system or can be send tosystemsystem
encevis using the interface
provided by AIT(AITInterfaceDLL)
No standard data format
Complianceavailable in the industryNo standard data formatavailable in the industryNo standard data formatavailable in the industry
Output FilesResults are stored in aResults are stored inResults are stored in
database and/or is send overadditional files in the fileadditional files in the file
the interface AITInterfaceDLLsystem placed in the samesystem placed in the same
to an external EEG system.folder as the EEG file. Userfolder as the EEG file. User
User output is given byoutput is given by graphicaloutput is given by graphical
graphical user interfacesuser interfacesuser interfaces
Compatible Equipment andSoftwareEncevis can read and process EEG data from several EEG vendors. A list of compatible EEG systems can be found onEncevis can read and process EEG data from several EEG vendors. A list of compatible EEG systems can be found onEncevis can read and process EEG data from several EEG vendors. A list of compatible EEG systems can be found on
http://www.encevis.comhttp://www.persyst.com/support/supported-formats/http://www.persyst.com/support/supported-formats/

{10}------------------------------------------------

Image /page/10/Picture/2 description: The image shows the logo for the Austrian Institute of Technology (AIT). The logo consists of the letters "AIT" in a stylized, geometric font in gray. To the right of the letters, the words "AUSTRIAN INSTITUTE OF TECHNOLOGY" are written in a smaller, sans-serif font in a dark red color. The logo is simple and modern, and the colors are muted and professional.

Table 1: Comparison between encevis and the predicate device

8. Non-Clinical Performance Data

Software verification and validation testing was conducted and documentation provided as recommended by the FDA Guidance for Industry and FDA Staff, Guidance for the Content of Software Contained in Medical Devices. Traceability has been documented between all system specifications to validation test protocols. Verification and validation testing includes:

    1. Code inspections
    1. Unit level testing
    1. Integration level testing
    1. System level testing

In addition, tests according to "IEC 62366-1:2015, Medical Devices Part 1—Application of usability engineering to medical devices" have been performed.

The software for this device is determined as a "moderate" level of concern because a failure or latent flaw could lead to a minor injury to the patient through incorrect information or through the action of the care provider.

Verification and validation activities established the safety and performance characteristics of the subject device with respect to the predicate device. The following performance data have been provided in support of the substantial equivalence determination.

FeatureTesti əməlາດ ປ່ວງ ການ ອາ ໄວinin cinsinə aid bitki növü. İstinadlar Respublikasının fəsiləsinə filmləri fəsiləsinə cinsinə aid bitki növü. İstinadlar Respublikasının fəsiləsinə filmlərindəin fəsiləsinə cinsinə aid bitki növü. İstinadlar Respublikası Respublikası Respublikası Respublikası Respublikası Respublikası Respublikası Respublikasıoanaspueq (suənbəyanissendone tempəsinə aid bitki növü. İstinadlar Quedd
Direct comparison with predicatedevice×××××
Bench test on large amount ofEEG data×××××××
Software test (System,Integration, Unit)×××××××××
Table 2: Type of performance test per feature

For bench tests, detection results of the modules were compared to annotations set by clinical EEG experts using large amount of EEG data from different centers. Where possible, the results of encevis

{11}------------------------------------------------

Image /page/11/Picture/2 description: The image shows the logo for the Austrian Institute of Technology. The logo consists of the letters "AIT" in a gray sans-serif font. To the right of the letters is the text "AUSTRIAN INSTITUTE OF TECHNOLOGY" in a smaller, maroon sans-serif font. The logo is simple and modern, and the colors are muted and professional.

were directly compared with the results of the predicate device. Suitable statistical measures like sensitivity and specificity were calculated.

The encevis (stand-alone software) meets all the stated requirements for overall design, performance. biocompatibility and electrical safety and passed all the testing noted above.

9. Clinical Performance Data

Seizure detection performance testing:

For performance evaluation of the encevis seizure detection device we measured positive percentage agreement (detection sensitivity based on the reference standard) and negative disagreement rate (false detections per 24 hours based on reference standard) by comparing seizure detections to consensus annotations from three independent reviewers. Second, to define the acceptable performance level of the encevis seizure detection device we also measured positive percentage agreement and neqative disagreement rate of the predicate device Persyst using the same study population and the same gold standard annotations. A statistical test is then used to show that the encevis seizure detection performance is non-inferior to the performance of the predicate device.

Study population

We included scalp-EEG recordings of 55 subjects that underwent video-EEG monitoring in an epilepsy monitoring unit for the purpose of differential diagnosis or pre-surgical evaluation. All patients where 18 years of age or older. 50 patients where included that showed seizure events during recording and were diagnosed of having epilepsy. Further, we included the five subjects that were diagnosed of not having epilepsy (Subject-ID 30-34). No further selection of subjects was made.

Reference Standard

To define the reference standard, a total of 1619 hours of EEG from these 55 subjects were presented to three independent neurologist for blinded review. The review sessions was to identify the start and end times of epileptic seizures to define "true seizure" epochs for later performance evaluation of the automatic seizure detection algorithm. The 1619 hours of EEG consisted of a maximum of 30 hours of continuous EEG data from each subject. For subjects without epilepsy the first available 30 hours of recording were included. The EEG experts were asked to mark the time positions of the seizure onset and end. An event was considered as "true seizure" only if the time interval of two out of three reviewers overlapped by at least 1 second. A seizure epoch was then defined as the overlapping time range of two reviewers.

The following tables show the number of EEG hours and seizures per subject.

{12}------------------------------------------------

Image /page/12/Picture/2 description: The image shows the logo for the Austrian Institute of Technology (AIT). The logo consists of the letters "AIT" in a stylized, geometric font in gray. To the right of the letters, the words "AUSTRIAN INSTITUTE OF TECHNOLOGY" are written in a smaller, sans-serif font in a dark red color. The logo is simple and modern, and the colors are muted and professional.

Image /page/12/Figure/3 description: The image contains two bar charts. The chart on the left is titled "Number of EEG hours per subject" and shows the length of EEG test data in hours for different subjects. The chart on the right is titled "Number of seizures per patient" and shows the number of seizures for different subjects. The number of seizures ranges from 0 to 29, while the length of EEG test data ranges from 0 to 30 hours.

Figure 1: Number of EEG hours and seizures per subject

{13}------------------------------------------------

Image /page/13/Picture/2 description: The image shows the logo for the Austrian Institute of Technology (AIT). The logo consists of the letters "AIT" in a gray, sans-serif font. To the right of the letters is the text "AUSTRIAN INSTITUTE OF TECHNOLOGY" in a smaller, red, sans-serif font. The logo is simple and modern, and it is likely used on the organization's website, marketing materials, and other official documents.

Inter-rater agreement

Inter-rater agreement was measured between any pair of reviewers using Cohen's kappa (x). The agreement values were к=0.69 (95% СI=[0.62, 0.77]) between reviewer 2, к=0.76 (95% Cl=[0.69, 0.83]) between reviewer 1 and reviewer 3, and κ=0.79 (95% Cl=[0.72, 0.85]) between reviewer 2 and reviewer 3. The average agreement resulted in «=0.74. According to the qualitative classification of Landis and Koch (Landis JR, Koch GG, Biometrics. 1977 Mar;33(1):159-74) the average к value of 0.74 can be interpreted as substantial agreement.

Detection Performance

To define positive percentage agreement (PPA) and negative disagreement rate (NDR, given as false detections in 24 hours) for each patient the seizure epochs defined by consensus annotations of two out of three reviewers were compared to automatically calculated seizure time points of the encevis seizure detection device and the predicate device Persyst. The encevis seizure detection device results in a single time point for each detection that is used in this validation. The predicate device Persyst was used with default settings (perception score = 0.5) and the given start time point was used in this validation. The logical variables true positive (FP), and false negative (FN) are defined as follows: seizure epochs are counted as TP if at least one detection occurred within the epoch time range. Detections outside of seizure epochs were defined as false positives (FP). Seizure epochs without a matching detection were defined as false negative (FN).

Results

The average positive percentage agreement of the subjects with at least one "true seizure" event resulted in 75.40% (95% Cl=[64.5. 86.3]) for encevis seizure detection and in 75.94% (95% Cl=[65.5. 86.4]) for the predicate device Persyst.

The average negative disagreement rate (NDR) was 7.01 false detections in 24 hours (95% Cl=[5.9, 8.21) for the encevis seizure detection and 10.61 false detections in 24 hours (95% Cl=16.8. 14.5)) for predicate device Persyst. The following two figures show the average results including confidence intervals.

Image /page/13/Figure/10 description: The image is a plot comparing the average positive percentage agreement between two different systems, encevis and Persyst. The x-axis represents the average positive percentage agreement, ranging from 0 to 100. The plot includes 95% confidence intervals for each system. The average positive percentage agreement for encevis is approximately 75%, while the average positive percentage agreement for Persyst is approximately 72%.

Figure 2: Average and the confidence interval of the positive agreement performance between encevis and Persyst.

{14}------------------------------------------------

Image /page/14/Picture/2 description: The image shows the logo for the Austrian Institute of Technology. The logo consists of the letters AIT in a stylized, sans-serif font in gray. To the right of the letters, the words "AUSTRIAN INSTITUTE OF TECHNOLOGY" are written in a smaller, sans-serif font in maroon. The logo is simple and modern, and the colors are muted and professional.

Image /page/14/Figure/3 description: The image is a plot comparing the average negative disagreement rate and 95% confidence intervals for two systems, encevis and Persyst. The x-axis represents the average negative disagreement rate, measured in false detections per 24 hours, ranging from 0 to 16. Encevis has a lower average negative disagreement rate, with its confidence interval centered around 7, while Persyst's confidence interval is centered around 11.

Figure 3: Average and the confidence interval of the negative disagreement performance between encevis and Persyst. Low numbers are better.

A Two One-Sided Test (TOST) procedure for paired samples (Walker E, Nowacki AS, J Gen Intern Med. 2011 Feb:26(2):192-6) was used to test the non-inferiority of the encevis seizure detection device to the predicate device. For statistical comparison, a type I error of 0.05 and non-inferiority margins of 10% for positive percentage agreement (PPA) and 1 for negative disagreement rate (NDR) were used. The two performance parameters PPA and NDR were tested independently to measure the noninferiority of both device parameters.

Both device parameters PPA and NDR of the encevis seizure detection are found to be non-inferior to the parameters of predicate device Persyst.

encevis spike detection performance testing

The clinical truth was determined based on the results of blinded review sessions from three neurologists. The "true spike" events (clinical truth) were then compared to automatically calculated spike time points of the encevis spike detection device to define true positives (TP), false positives (FP), false negatives (FN), and true negatives (TN) for each patient. With these values the positive percentage agreement (PPA) and negative percentage agreement (NPA) for each patient are calculated. In addition, "true spike" events were compared to the automatic detections of Persyst resulting in PPA and NPA values for the predicate device. Furthermore, the localization performance of both systems EpiSpike and Persyst was evaluated based on the localization information given by the detection systems and the spatial information provided by the reviewer (clinical truth). We define a positive localization percentage agreement (PLPA) which is calculated for each patient.

Study population

To prove the validity of the spike detection system, EpiSpike was tested with the EEG of 23 patients. For clinical validation, we included scalp-EEG recordings of 23 subjects that underwent video-EEG monitoring in an epilepsy monitoring unit for differential diagnosis or pre-surgical evaluation. 18 subjects of 18 years of age or older that showed spike events during recording based on initial clinical

{15}------------------------------------------------

Image /page/15/Picture/2 description: The image shows the logo for the Austrian Institute of Technology. The logo consists of the letters "AIT" in a stylized, gray font on the left. To the right of the letters, the words "AUSTRIAN INSTITUTE OF TECHNOLOGY" are written in a smaller, red font. The logo is simple and modern, and the colors are muted and professional.

information where included. In addition, five subjects of 18 years of age or older that were diagnosed of not having epilepsy were included (Subject-ID 9-13). No further selection of subjects were made.

The statistical parameters PPA. NPA and PLPA were used in a two one-sided test (TOST. (Walker E et. al.) using paired samples in order to show the non-inferiority of EpiSpike device compared to the predicate of Persyst.

Reference standard

To define the clinical truth the EEG from all subjects were presented to three independent Neurologists for blinded review. The goal of the review sessions was to identify all "true focal spikes" for later performance evaluation of the automatic spike detection algorithm. The EEG experts were asked to mark the time positions at the beginning and the end of the spike. Furthermore, the reviewers were asked to specify the electrode which is next to the spike maximum (phase reversal).

An event was considered as "true spike" only if the time interval of two out of three reviewers overlapped. For the determination of the localization performance. the 3D-coordinates of the electrode which is next to the spike maximum averaged over reviewers was used. The determined average position is considered as the clinical truth with respect to the localization and is used to evaluate the localization performance of EpiSpike and the predicate Persyst.

Performance evaluation

Data of all 23 subjects was processed with EpiSpike. In order to compare the obtained results of EpiSpike with the predicate Persyst, the same data was processed with the spike detector of Persyst 12. The detection systems were evaluated by means of suitable performance measures like positive percentage agreement (PPA) and negative percentage agreement (NPA). For measuring the localization performance, we defined a positive localization percentage agreement (PLPA). The basis for the performance evaluation are the annotations of the EEG experts which were placed at the onset and the end of the spike. Comparison between the time instances of the annotations and the time instances of automatic detections allows assessing the performance. The detection resolution of both systems, EpiSpike and the Persyst spike detection was one microsecond.

Results of the performance measures

The average positive percentage agreement of the 15 subjects with at least one "true spike" event resulted in 61.02% (95% Cl=[47.9-74.1]) for encevis spike detection and in 8.7% (95% Cl=[4.4-13.0]) for the predicate device Persyst.

The average negative percentage agreement of all 23 subjects was 98.58 (95% Cl=[97.9-99.3]) for the encevis spike detection and 99.69 (95% Cl=[99.4-99.9]) for predicate device Persyst.

The average positive localization percentage agreement of the 12 subjects with at least one "true positive" event was 96.09 (95% CI=[88.8.-103.4]) for the encevis spike detection and 93.97 (95% Cl=[83.6-104.3]) for predicate device Persyst.

In the following figures the averages and the confidence intervals for the PPA, NPA and PLPA are visualized.

{16}------------------------------------------------

Image /page/16/Figure/0 description: This image shows a plot comparing the average positive percentage agreement between two entities, encevis and Persyst. The x-axis represents the average positive percentage agreement, ranging from 0 to 100. Persyst has an average positive percentage agreement of around 10%, while encevis has an average positive percentage agreement of around 60%.

Figure 4: This figure compares the average and the confidence interval of the positive percentage agreement performance between encevis and Persyst.

Image /page/16/Figure/2 description: The image is a bar chart that shows the average negative percentage agreement, including 95% confidence intervals. The x-axis is labeled "Average negative percentage agreement (%)" and ranges from 50 to 100. The y-axis shows the names of two items: "encevis" and "Persyst". The average negative percentage agreement for "encevis" is approximately 98%, while the average negative percentage agreement for "Persyst" is approximately 99%.

Figure 5: This figure compares the average and the confidence interval of the negative percentage agreement performance between encevis and Persyst.

{17}------------------------------------------------

Image /page/17/Figure/0 description: The image is a graph comparing the average positive localization percentage agreement between encevis and Persyst, including 95% confidence intervals. The x-axis represents the average positive localization percentage agreement, ranging from 0 to 100. The graph shows that both encevis and Persyst have high average positive localization percentage agreement, with encevis slightly higher than Persyst. The confidence intervals for both are relatively narrow.

Figure 6: This figure compares the average and the confidence interval of the positive localization percentage agreement performance between encevis and Persyst.

A Two One-Sided Test (TOST) for paired samples (Walker et al) was used to test the non-inferiority of the encevis spike detection device to the predicate device. For statistical comparison, a type I error of 0.05 and non-inferiority margins of 3% for positive percentage agreement (PPA), the negative percentage agreement (NPA) and the positive localization percentage agreement (PLPA) are used. The three performance measures PPA, NPA and PLPA were tested independently to measure the noninferiority of all device parameters separately.

encevis artifact reduction performance testing:

The quality of an artifact removal algorithm is determined by two aspects.

    1. The method should not significantly modify true, clean EEG pattern that are not disturbed by artifacts. To quantify the performance of the algorithms with regard to this aspect, changes in clean EEG patterns due to the algorithms are evaluated.
    1. The method should suppress artifacts that are superimposed on the true EEG as far as possible, revealing the underlying, pure EEG patterns. To quantify the ability of the algorithms to remove artifacts, signal-to-noise ratios will be measured before and after artifact removal.

For these measurements we need clean, pure EEG patterns and artifacts of different types. In order to identify these patterns, three EEG experts Neurologists are engaged as independent reviewers.

Validation data

For the validation study. 128 EEG data records from different patient groups are used, covering all intended use populations of encevis, i.e., adult patients in epilepsy monitoring and in critical care. Each record consisted of 10 seconds of data to be evaluated. The datasets include 60 patients from epilepsy monitoring units and 65 from ICU patients. These data were selected as follows:

Epilepsy monitoring – seizure EEGs: We include 31 EEG segments from 31 subjects of 18 years of age or older that underwent video-EEG monitoring in an epilepsy monitoring unit for the purpose of

{18}------------------------------------------------

Image /page/18/Picture/2 description: The image shows the logo for the Austrian Institute of Technology. The logo consists of the letters "AIT" in a gray, sans-serif font. To the right of the letters is the text "AUSTRIAN INSTITUTE OF TECHNOLOGY" in a smaller, red, sans-serif font. The logo is simple and modern, and it is likely used on the organization's website, marketing materials, and other official documents.

differential diagnosis or pre-surgical evaluation and that showed seizure events during recording and were diagnosed of having epilepsy.

Epilepsy monitoring - spikes: We include 33 EEG segments from 6 subjects of 18 vears of age or older that underwent video-EEG monitoring in an epilepsy monitoring unit for the purpose of differential diagnosis or pre-surgical evaluation that showed spikes during recording.

Intensive care unit: We include 65 EEG segments from 65 subjects of 18 years of age or older that have been admitted to an intensive care unit due to severe neurological disorders (cerebral hypoxia, cerebral ischemia, cerebral hemorrhage of different genesis, cerebral tumors, status epilepticus, infections, toxidromes, encephalopathies of different genesis, cerebral malformations and craniocerebral traumas) on a systemic or localized basis. The random selection includes 9 segments with seizures, 10 segments with rhythmic activity, 11 segments with periodic discharges, 17 segments with burst-suppression and 18 segments without any pattern.

Expert review

For this validation study we need annotations of clean EEG recordings without any artifacts, and moreover annotations of artifacts that can be superimposed to the clean recordings. We enqage three independent epileptologists or neurologists for blinded review of the EEG data from EMU and ICU.

Statistical testing

A Two One-Sided Test (TOST) procedure for paired samples (Walker E, Nowacki AS, J Gen Intern Med. 2011 Feb:26(2):192-6) is used to test the non-inferiority of the encevis artifact reduction compared to the predicate device. For statistical comparison, a type I error of 0.05 and non-inferiority margins of 1dB.

The hypothesis to test non-inferiority of the relative suppression of true EEG in dB is defined as:

  • . H0: The relative suppression of true EEG in dB of the encevis artifact removal is higher than the suppression of true EEG in dB of the predicate device.
  • . H1: The relative suppression of true EEG in dB of the encevis artifact removal is lower than or equal to the suppression of true EEG in dB of the predicate device.

The hypothesis to test the signal-to-noise ratio after artifact removal is defined as:

  • . H0: The signal-to-noise ratio after artifact removal by encevis is lower than the signal-to-noise ratio after artifact removal by of the predicate device.
  • H1: The signal-to-noise ratio after artifact removal by encevis is higher than or equal to the ● signal-to-noise ratio after artifact removal by of the predicate device.

The results of the evaluation of relative suppression of clean EEG are summarized in the following table in %. This number means, that the variance of the clean EEG activity has been suppressed by this relative value, i.e., low values are desired. Due to technical reasons, only 111 out of 145 test cases could be evaluated: in the remaining 34 cases, Persyst produced zero lines in all channels.

{19}------------------------------------------------

Traditional 510(k)

510(k) Premarket Notification

Image /page/19/Picture/2 description: The image shows the logo for the Austrian Institute of Technology. The logo consists of the letters "AIT" in a stylized, geometric font in gray. To the right of the letters, the words "AUSTRIAN INSTITUTE OF TECHNOLOGY" are written in a smaller, sans-serif font in maroon. The logo is simple and modern, and the colors are muted and professional.

Image /page/19/Figure/4 description: The image is a bar graph comparing the relative suppression of clean EEG between Encevis and Persyst. The x-axis represents the percentage of relative suppression of clean EEG, ranging from 0.00% to 70.00%. The y-axis lists different data points, numbered from 1 to 129. The graph uses blue bars to represent Encevis and red bars to represent Persyst, showing the relative suppression percentage for each data point.

Figure 7: Relative suppression of clean EEG by encevis and Persyst

{20}------------------------------------------------

Traditional 510(k)

The results of the evaluation SNR prior and post artifact removal are summarized in the following table in dB. This numbers show the signal-to-noise ratio (noise=artifacts) that has been achieved after artifact removal, i.e., high values are desired. Due to technical reasons, only 80 out of 104 test cases could be evaluated: in the remaining 24 cases, Persyst produced zero lines in all channels.

Image /page/20/Figure/4 description: The image is a bar graph titled "SNR prior and post artifact removal [dB]". The graph compares the initial SNR with the SNR after artifact removal using Encevis and Persyst. The x-axis represents the SNR in decibels, ranging from -60 to 20, and the y-axis represents different data points labeled with numbers from 1 to 103. The graph shows the SNR values for each data point before and after artifact removal using the two methods.

Figure 8: SNR prior and post artifact removal by encevis and Persyst

{21}------------------------------------------------

Image /page/21/Picture/2 description: The image shows the logo for the Austrian Institute of Technology (AIT). The logo consists of two parts: a stylized, gray-colored graphic on the left and the text "AUSTRIAN INSTITUTE OF TECHNOLOGY" on the right. The text is in a dark red color and is stacked in two lines.

Results of Statistical testing

The results of the Two One-Sided Test for relative Suppression of clean EEG (Test-Control) are (95% delta CI=[-0.09, -0.04], margin = 0.01):

-0.04 < 0.01 -> H1

The results of the Two One-Sided Test for signal-to-noise ratios after artifact removal are (95% delta Cl=[4.78,6.60], marqin = 0.10):

4.78 > -0.10 -> H1

Both device parameters, "relative Suppression of clean EEG" and "signal-to-noise ratifact removal" of the encevis artifact reduction are therefore non-inferior to the parameters of predicate device Persyst.

In the statistical evaluation of both device parameters, "relative Suppression of clean EEG" and "signalto-noise ratios after artifact removal" of the encevis artifact reduction are shown to be non-inferior to the parameters of predicate device Persyst. Moreover in 82 out of 111 test cases, the suppression of clean EEG by encevis was lower compared to Persyst. And in 75 out of 80 test cases, the SNR after artifact removal by encevis was higher compared to Persyst. It can be concluded that the encevis artifact reduction "PureEEG" does not perform worse that the artifact reduction by the predicate device.

encevis rhvthmic and periodic patterns performance testing:

The detection of rhythmic and periodic patterns in NeuroTrend is used to visually mark EEG segments with rhythmic or periodic signal content. The definition of rhythmic and periodic patterns follow the quidelines of the ACNS (American Clinical Neurophysiology Society) ICU EEG Terminology (Hirsch et al., 2013). NeuroTrend displays all detected rhythmic and periodic patterns in plots called "Pattern Localization" and "Pattern Frequency".

For the validation we compared and statistically analyze annotations of two human EEG-readers with the detections of NeuroTrend. We showed that the detected patterns have a high sensitivity and specificity compared to manual annotated EEG segments. We prospectively recorded 83 long term EEGs from ICU-patients at two different centers using the international 10-20 electrode system with a sampling rate of 256Hz.

EEGs were annotated by two clinical neurophysiologists that were naive to these EEGs. The annotation procedure included the first minute of each hour, were each minute was split into three independent segments of 20 seconds resulting in 11935 common annotation segments. Several non-overlapping annotations were allowed in each annotation segment. Each annotation may have an arbitrary start and end position but has to be fully included in the annotation segment. For each annotation, the reviewer was allowed to choose between one of the following pattern types:

    1. PD: periodic pattern
    1. RDA: rhythmic delta activity
    1. RTA: rhythmic theta activity
    1. RAA: rhythmic alpha activity
    1. SW: rhythmic spike-and-wave activity
    1. BS: burst suppression pattern
    1. No annotation (short NOPA).

{22}------------------------------------------------

Image /page/22/Picture/2 description: The image shows the logo for the Austrian Institute of Technology (AIT). The logo consists of the letters AIT in a stylized, geometric font. The letters are in gray, and to the right of the letters is the text "AUSTRIAN INSTITUTE OF TECHNOLOGY" in a smaller, red font. The logo is simple and modern, and it is likely used to represent the organization's brand.

In addition to the type of the pattern the localization property had to be set by the human reviewers. This property was defined in (Hirsch et al., 2013) as main term 1:

    1. G: generalized pattern
    1. L: lateralized pattern

The annotations from the two reviewers were then used as gold standard condition to test sensitivity and specificity of the rhythmic and periodic pattern detection of NeuroTrend. Annotations had to be consistent between both reviewers to be used in the sensitivity and specificity measurement.

The detection performance was defined by assigning one of four possible test conditions to each of the 1 minute annotation segments: true positive (TP), false positive (FP), true negative (TN), and false negative (FN). An annotation segment was counted as TP if a detection and an annotation was present. An annotation segment with a gold standard annotation but without any detection will be counted as FN. An annotation segment with detections but without annotations will be counted as FP. An annotation segment without gold standard annotation and without detections will be counted as TN.

The sensitivity is defined as:

$$\mathbf{SE} \text{ [%]} = #\text{TP} / (#\text{TP} + #\text{FN}) \text{ * } \text{100}$$

The specificity is defined as:

SP [%] = #TN/(#TN+#FP) * 100

The # symbol stands for "number of". The symbol "#TP" represents therefore the number of true positive annotation segments.

The localization information will be validated by comparing the concise annotations of the two human reviewers for all correctly detected markers (the TP detections).

The result of the manual annotation procedure was evaluated using the Cohens' kappa statistic. This statistic measures the level of agreement between two reviewers. A kappa value of 0.66 was measured between reviewer 1 and reviewer 2.

REVIEWER 2
NOPATPDRAARDARTASW
REVIEWER 1NOPAT107573111478234
PD5881290063140
RAA1161100
RDA1355011914
RTA50251231070
SW10030020
Cohens Kappa:0.66 (CI=0.64-0.67)Substantial agreement

Table 3: Cohens' kappa statistic for the evaluation of the pattern detection

The overall detection performance measures the sensitivity of the NeuroTrend detections without evaluating the pattern type. The result is marked with the label "ANY" in the result file. This result proofs the ability of NeuroTrend to detect any relevant pattern and ignores pattern type mismatches. The result of the periodic pattern group is labeled as "PD". This result shows the sensitivity and specificity of the periodic pattern detections. The result of the rhythmic delta activity

{23}------------------------------------------------

Image /page/23/Picture/2 description: The image shows the logo for the Austrian Institute of Technology (AIT). The logo consists of the letters "AIT" in a stylized, geometric font in gray. To the right of the letters, the words "AUSTRIAN INSTITUTE OF TECHNOLOGY" are written in a smaller, sans-serif font in a dark red color. The logo is simple and modern, and the colors are muted and professional.

pattern detections is labeled as "RDA". The result of the ARA group shows the result of aggressive rhythmic activity, including the pattern types RTA, RAA, and SW.

Pattern TypeSensitivity(%)Specificity(%)
ANY77.59% (95% CI=[75.5-79.7])86.50% (95% CI=[85.8-87.2])
PD63.37% (95% CI=[60.7-66.0])96.57% (95% CI=[96.2-96.9])
ARA (including RTA, RAA, SW)92.72% (95% CI=[88.2-97.2])94.76% (95% CI=[94.4-95.2])
RDA92.56% (95% CI=[87.5-97.7])90.44% (95% CI=[89.9-91.0])

Table 4: Sensitivity and specify for encevis pattern detection

The inter reader agreement table of the localization information (ACNS Main Term 1) compares the consistent annotations of two EEG experts to the localization shown in NeuroTrend. The result is shown in the following table:

NeuroTrend
REVEWER 1+2G
G683ત્ત્વે ઉત્પત્તર તે ઉત્પત્તર તે જીવન જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામમાં પ્રાથમિક શાળા, પંચાયતઘર, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામન
113168
Cohens Kappa:0.45, Cl=0.43-0.48(Moderate agreement)

Table 5: Inter-reader agreement between reviewers and NeuroTrend for encevis pattern localization

encevis aEEG performance testing:

Amplitude-integrated EEG (aEEG) is a popular method for monitoring cerebral function by displaying the amplitude trend of brain activity. It is the boundary of the EEG waveform (i.e. the envelope) and not the EEG itself (i.e. the carrier) that characterizes the tendency of amplitude changes (Zhang and Ding, 2013).

The Background-AEEG module of NeuroTrend estimates and visualizes the temporal evolution (trend) of the EEG amplitude. The implementation is oriented on the proposed method of (Zhang and Ding. 2013)

In the first step the frequency response of the module is checked for equality with the proposed method of (Zhang and Dinq, 2013). This test only considers the correct slope (dB loss per decade) not the correct filter gain factor. In this test, sinusoidal one-channel test data with increasing from 0.5Hz to 32Hz and amplitude of 40uV are generated, one test case for each hemisphere. With the results of the module the frequency response is determined and checked if the dB loss per decade within the band pass (cut-off frequencies of 2 and 15Hz) is -12db/dec and the maximum gain factor in the stop band is not greater than -30dB. This step validates the correct implementation of the filters and its characteristics (expect the gain factor) within the module.

In the second step the results of the module are compared with the aEEG results of Persyst (CE certified and FDA approved software; http://www.persyst.com/) using real EEG data. The configuration of Persyst is set in a way to allow an adequate comparison.

After successful validation according to the description above we will have shown that the Background-AEEG module correctly determines the averaged EEG amplitude of the left and right hemisphere according to the proposes method of (Zhang and Ding, 2013)

{24}------------------------------------------------

Figure 9: Frequency response of the left (left side) and the right hemisphere (right side).

In the first validation step the frequency response of the aEEG module is checked for equality with the pro-posed method of (Zhang and Ding, 2013). The resulting frequency response of the Background-AEEG module is shown in Figure 9 .The following conclusions were drawn from the results:

  • . The determined characteristic is very similar to the published version in (Zhang and Ding, 2013). Only the absolute shift of the complete frequency response is different but because only changes in aEEG values are of clinical relevance this detail is irrelevant.
  • . Both hemispheres show the same characteristic
  • . In the stop band there is a suppression of -30dB and higher
  • . The slope in the pass band is approximately -12dB/decade

In the second step the results of the aEEG module are compared with the aEEG results of Persyst. For this test, real EEG data were used. The aEEG of the same EEG segment using either Persyst or the Background-AEEG module of encevis NeuroTrend were compared. The test cases showed that the aEEG values of the Background-AEEG module of encevis NeuroTrend and Persyst are in good accordance. Furthermore, the aEEG values are in good accordance with the corresponding raw EEG.

encevis frequency bands performance testing:

The background-frequency module of NeuroTrend estimates and visualizes the temporal evolution (trend) of relative proportions of dominant EEG-waveform-frequencies. The result is graphically presented using a plot (cf. Figure 10), where the x-axis represents the time-axis, and four stacked areas in different colors and widths represent the relative proportions of the four frequency bands Delta, Theta, Alpha, and Beta for subsequent time windows with lengths of 15 seconds. The intensity of the colors furthermore corresponds to the amplitudes in these four frequency bands. This representation allows the user to identify time epochs that are dominated by a specific frequency band. E.g., EEGslowing or, in other words, an epoch with dominant delta- or theta-wave can be recognized in the graphical representation by broad stretches of the corresponding areas.

Image /page/24/Figure/10 description: The image shows a stacked proportion chart of frequency bands. The y-axis ranges from 0% to 100%, and the x-axis represents the time or data points. The chart displays the proportions of different frequency bands, including beta, alpha, theta, and delta, with corresponding colors. The x-axis is labeled 0-200 uV.

Figure 10: Graphical representation of the Background-EEG-Frequency evaluation results.

{25}------------------------------------------------

Image /page/25/Picture/2 description: The image shows the logo for the Austrian Institute of Technology. The logo consists of the letters "AIT" in a stylized, geometric font in gray. To the right of the letters, the words "AUSTRIAN INSTITUTE OF TECHNOLOGY" are written in a smaller, sans-serif font in a dark red color. The logo is simple and modern, reflecting the institute's focus on technology and innovation.

In order to proof the validity of the Background-EEG-frequency module we followed a twostep approach. In the first step it was shown that the assignment of sinusoidal test data to frequency bands (Delta, Theta, Alpha, or Beta) is correct according to the above definitions of frequency borders. In this test, sinusoidal test data with frequencies across all four bands and amplitudes ranging from 2 uV to 200 µV were generated. Then it was verified, that the algorithm correctly assigns each test signal to the corresponding frequency band, and that the measurement error for amplitudes are below 5 %. This validates the correct assignments of single, 3-second EEG epochs to a frequency band and amplitude.

In a seconds step it is shown that the globally dominant background frequency within a 15-seconds window is correctly identified. This is done using manually selected EEG recordings from epilepsy- or ICU patients. Each of these EEG samples is representative for a specific background-EEG-frequency band, i.e., it is mainly dominated by delta-, theta-, alpha-, or beta-waves, For these samples the background-EEG-frequency module calculates the proportional composition of frequency bands. The one frequency band with the largest proportion can be seen as the globally dominant background frequency, if this proportion is particularly high. Thus it is verified for each of these representative examples that the relative proportion corresponding to the true frequency band is greater than 50 %.

encevis Burst Suppression performance testing:

The detection of burst suppression patterns and quantitative measure for the discontinuity of the EEG shown in NeuroTrend was validated using the following approach:

    1. The time point of the detected burst suppression patterns will be compared to annotations defined by two clinical EEG experts using EEG data from a multicenter study. Sensitivity will be calculated.
    1. The quantitative measure of the amplitude loss of the suppression and the suppression time in percent will be validated using an artificial EEG. The EEG file includes a set burst suppression patterns with different values for suppression time and suppression amplitude loss. The calculation results of the quantitative burst suppression plots shown in NeuroTrend will be compared to precalculated values.

We recorded 83 long term EEGs from intensive care patients from two different centers using the international 10-20 electrode system with a sampling rate of 256Hz. EEGs were annotated by two clinical neurophysiologists that were naive to these EEGs. The annotation procedure included the first minute of each hour resulting in 3978 valid annotation segments. The reviewers were allowed to assign two categories for each annotation segment:

    1. EEG with burst suppression patterns (BS)
    1. EEG without burst suppression patterns (BS)

Statistical analysis of the detection performance was done by defining the annotations of the reviewers as gold standard and by comparing these annotations to the detection results of the computational method. Each one minute EEG segments annotated as "EEG with burst suppression" with an overlapping burst suppression detection segment of 15 seconds was defined as true positive (TP) event. One minute EEG segments annotated as "EEG with burst suppression patterns" without any overlapping burst suppression result were defined as false negatives (FN). One minute segments annotated as "EEG without burst suppression patterns" and with an overlapping burst suppression detection result are defined as false positives (FP), all other segments are defined as true negatives (TN).

{26}------------------------------------------------

Image /page/26/Picture/2 description: The image shows the logo for the Austrian Institute of Technology. The logo consists of the letters "AIT" in a stylized, geometric font in gray. To the right of the letters, the words "AUSTRIAN INSTITUTE OF TECHNOLOGY" are written in a smaller, sans-serif font in a dark red color. The logo is simple and modern, and the use of gray and red gives it a professional look.

The following table shows the evaluation results of the automatic burst suppression detection method in NeuroTrend using 3978 segments annotated by two reviewers. The results of the automatic burst suppression detection method were compared to the manual annotations of the reviewers. The detection performance was analyzed for consensus annotations of the two reviewers. The consensus annotations only include annotation segments where both reviewers showed the same decision about Burst Suppression pattern. The measured values for sensitivity (SE), specificity (SP), positive predictive value (PPV), and negative predictive value (NPV) prove the validly of the detection algorithm. The very large sample size does imply a statistically high confidence.

Rev.(n)Rev. IDsSE (%)SP (%)PPV (%)NPV (%)ACC (%)
2JH+JK87.28(95% CI=[90.8, 92.3])92.16(95% CI=[91.3, 92.9])61.09(95% CI=[57.9, 64.2])98.09(95% CI=[97.6, 98.5])91.56(95% CI=[90.8,92.3])

Table 6: Performance of the automatic burst suppression detection method

10. Statement of Substantial Equivalence

encevis is substantially equivalent in design and intended use to the predicate device. Any differences between the subject and predicate device have no significant influence on safety or effectiveness as established through performance testing. Therefore, the encevis raises no new issues of safety or effectiveness when compared to the predicate device.

§ 882.1400 Electroencephalograph.

(a)
Identification. An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.(b)
Classification. Class II (performance standards).