K Number
K171566

Validate with FDA (Live)

Date Cleared
2018-07-19

(415 days)

Product Code
Regulation Number
862.1215
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Atellica IM High-Sensitivity Troponin I (TnIH) assay is for in the quantitative measurement of cardiac troponin I in human serum or plasma (lithium heparin) using the Atellica IM Analyzer. The assay can be used to aid in the diagnosis of acute myocardial infarction (AMI).

Device Description

The Atellica IM TnIH assay components include the Atellica IM TnIH Primary Reagent ReadyPack (containing Lite Reagent and Solid Phase Reagent) and the Atellica IM TnIH Calibrator (containing High Calibrator (Cal H) and Low Calibrator (Cal L)). The Lite Reagent contains Bovine serum albumin (BSA) conjugated to a recombinant monoclonal (sheep) Fab anti-human cTnl labeled with acridinium ester in HEPES buffer; stabilizers; preservatives. The Solid Phase Reagent contains Streptavidin-coated magnetic latex particles with 2 biotinylated (mouse and sheep) monoclonal anti-troponin I antibodies in buffer; stabilizers; preservatives. The High Calibrator contains Human serum; human cTnl; preservatives. The Low Calibrator contains HEPES buffer; bovine serum albumin (BSA); surfactants; preservatives. The methodology is Chemiluminescence and the assay protocol is a Sandwich immunoassay.

AI/ML Overview

Here's an analysis of the provided text, focusing on acceptance criteria and study details:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state formal acceptance criteria in a structured table. However, implied acceptance criteria are based on CLSI (Clinical and Laboratory Standards Institute) guidelines and the reported performance of the device. The "High-Sensitivity Designation" section (10.10) presents criteria for classification as a high-sensitivity troponin test.

CategoryCriterion (Implied Acceptance)Reported Device Performance (Atellica IM TnIH)
High Sensitivity Designation (as per IFCC Task Force)
%CV at 99th percentile≤10%Meets Criterion: The assay meets this criterion (explicitly stated: "The Atellica IM TnIH assay meets both of these criteria."). Specific %CV values not explicitly provided for the 99th percentile in the summary.
Measurable Conc. for 50%+ Healthy IndividualsAttainable above LoD for at least 50% of healthy individualsMeets Criterion: The assay meets this criterion (explicitly stated: "The Atellica IM TnIH assay meets both of these criteria."). Specific data not provided in the summary.
Precision (CLSI EP5-A3)Demonstrated acceptable repeatability and within-lab precision for various concentrations. (Specific numerical thresholds not stated as AC)
Serum Samples:
12.72 pg/mLRepeatability %CV: 4.3; Within-Lab %CV: 4.7
127.93 pg/mLRepeatability %CV: 1.8; Within-Lab %CV: 2.4
1334.97 pg/mLRepeatability %CV: 1.7; Within-Lab %CV: 2.1
13815.89 pg/mLRepeatability %CV: 1.4; Within-Lab %CV: 1.9
Lithium Heparin Plasma Samples:
12.03 pg/mLRepeatability %CV: 4.1; Within-Lab %CV: 5.3
131.21 pg/mLRepeatability %CV: 1.7; Within-Lab %CV: 2.1
1363.38 pg/mLRepeatability %CV: 2.0; Within-Lab %CV: 2.3
12862.97 pg/mLRepeatability %CV: 1.7; Within-Lab %CV: 2.3
Linearity (CLSI EP06-A)Demonstrated acceptable deviation from linear fit. (Specific numerical thresholds not stated as AC)Deviations from linear fit generally low, with largest reported as 7.47% (Serum Full Range, Sample I).
Dilution RecoveryIndividual sample recoveries within 20% (implied)Meets Implied Criterion: All individual sample recoveries were within 20%. Mean of 1:2 dilutions: 99.6%; Mean of 1:5 dilutions: 91.6%.
Hook EffectNo hook effect up to a specified concentration. (Implied AC)Meets Implied Criterion: No hook effect up to 500,000 pg/mL.
Detection Limit (CLSI EP17-A2)Defined LoB, LoD, and LoQ. (Implied AC)LoB: 0.50 pg/mL; LoD: 1.60 pg/mL; LoQ: 2.50 pg/mL (at 20% total CV).
Endogenous Interference (CLSI EP07-A2)Typically <10% interference from specified substances. (Implied AC)All tested endogenous substances showed <10% interference, with the highest being 5.43% for Cholesterol (Serum).
Drug Interference (CLSI EP07-A2)<10% interference from tested therapeutic drugs. (Implied AC)Meets Implied Criterion: All tested therapeutic drugs caused <10% interference.
Cross-ReactivityNegligible or 0% cross-reactivity with specified substances. (Implied AC)Meets Implied Criterion: All tested potential cross-reacting substances showed 0.00% cross-reactivity.
Heterophile InterferenceNo interference with HAMA or RF. (Implied AC)Meets Implied Criterion: No interference with HAMA or RF was observed.
Clinical Performance (Diagnostic Accuracy for AMI)High sensitivity, specificity, PPV, and NPV across various timepoints. (Specific thresholds not explicitly stated as AC for submission)Reported ranges for Sensitivity (e.g., 75.0-97.0%), Specificity (e.g., 83.4-93.6%), PPV (e.g., 44.6-67.7%), and NPV (e.g., 94.0-99.5%) depending on matrix, time, and 99th percentile cutoff used.

2. Sample Size Used for the Test Set and Data Provenance

The main test set can be considered the clinical performance study.

  • Sample Size: The number of patients varies depending on the timepoint and matrix. For example, for Lithium Heparin Plasma at 0-1.5hr, N=45 for AMI cases and N=401 for non-AMI cases (female/male specific 99th percentiles). When considering overall 99th percentile for both genders, the sample sizes are much larger (e.g., Li Hep Plasma 0-1.5hr, N=145 AMI, N=964 non-AMI).
  • Data Provenance:
    • Country of Origin: Specimens were collected at 29 sites from different regions across the United States.
    • Retrospective or Prospective: The study enrolled patients who presented to the emergency department, or ambulatory care center equivalent, with signs or symptoms suspicious for a possible acute coronary syndrome (ACS) event. This indicates a prospective data collection.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Number of Experts: An "independent adjudication committee" was used. The exact number of individuals on this committee is not specified, but it's referred to in the plural ("cardiologists").
  • Qualifications of Experts: The ground truth for AMI diagnosis was established by cardiologists. No specific years of experience are provided, but their specialty (cardiologist) is a key qualification.

4. Adjudication Method for the Test Set

The adjudication method was performed by an independent committee of cardiologists, based on the Third universal definition of myocardial infarction consensus guideline endorsed by the European Society of Cardiology (ESC), the American College of Cardiology Foundation (ACCF), the American Heart Association (AHA), and the World Heart Federation (WHF). The specific methods of consensus (e.g., 2+1, majority vote) within this committee are not detailed in the summary.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

  • No MRMC study was done. This device is an in vitro diagnostic (IVD) assay for quantitative measurement of a biomarker (cardiac troponin I), not an AI-based image analysis or decision support system for human readers. Therefore, the concept of human readers improving with or without AI assistance does not apply to this type of device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Yes, a standalone study was done. The performance characteristics (precision, linearity, detection limits, interference, cross-reactivity) and the clinical performance (sensitivity, specificity, PPV, NPV) are all measures of the device's performance as an assay system (Atellica IM TnIH assay on the Atellica IM Analyzer) producing quantitative results. This is inherently a standalone performance evaluation by the device itself, without a human interpretation loop required for the direct output (the quantitative troponin I value). Human clinicians then use this value in conjunction with other clinical findings to make a diagnosis.

7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

  • The ground truth for the clinical performance study was the adjudicated diagnosis of Acute Myocardial Infarction (AMI). This adjudication was performed by an independent committee of cardiologists based on the "Third universal definition of myocardial infarction consensus guideline." This is a form of expert consensus based on established clinical guidelines and outcomes data.

8. The Sample Size for the Training Set

  • The document does not provide information on a training set sample size. This is expected as the Atellica IM TnIH is a traditional IVD assay (immunoassay), not an AI/ML device that requires a separate training phase. The development of such assays involves method development, optimization, and internal validation, but not "training sets" in the context of machine learning.

9. How the Ground Truth for the Training Set Was Established

  • As there is no mention of a "training set" in the context of an AI/ML algorithm, this question is not applicable to this device. The assay development would typically involve purified analytes, known concentrations, and reference methods used for calibration and validation during the research and development phase to ensure accurate measurement of cardiac troponin I. The "traceability and value assignment" (Section 10.14) mentions standardization to an internal standard manufactured using human heart homogenate.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 19, 2018

Siemens Healthcare Diagnostics Inc. Matthew Gee Senior Manager, Regulatory Affairs 511 Benedict Avenue Tarrytown, NY 10591

Re: K171566

Trade/Device Name: Atellica IM High-Sensitivity Troponin I (TnIH) Regulation Number: 21 CFR 862.1215 Regulation Name: Creatine phosphokinase/creatine kinase or isoenzymes test system Regulatory Class: Class II Product Code: MMI Dated: June 26, 2018 Received: June 27, 2018

Dear Matthew Gee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR

{1}------------------------------------------------

Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Kellie B. Kelm -S

for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K171566

Device Name

Atellica IM High-Sensitivity Troponin I (TnIH)

Indications for Use (Describe)

The Atellica IM High-Sensitivity Troponin I (TnIH) assay is for in the quantitative measurement of cardiac troponin I in human serum or plasma (lithium heparin) using the Atellica IM Analyzer. The assay can be used to aid in the diagnosis of acute myocardial infarction (AMI).

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92 and the Safe Medical Device Act of 1990.

The assigned 510(k) Number is: K171566

Date Prepared 1.

July 13, 2018

Applicant Information 2.

Contact:Matthew Gee, M.Sc.Senior Manager, Regulatory Affairs
Address:Siemens Healthcare Diagnostics Inc511 Benedict AvenueTarrytown, NY 10591-5097
Phone:914-255-8782
Fax:914-524-3579

Email: matthew.gee@siemens.com

3. Regulatory Information

Table 1. Regulatory Information for Atellica IM High-Sensitivity Troponin I (TnIH)

Trade NameAtellica® IM High-Sensitivity Troponin I (TnIH)
Model Numbers10997840 (1-pack); 10997841 (5-pack)
Common NameImmunoassay Method, Troponin Subunit
Regulation Number862.1215
Regulation DescriptionCreatine phosphokinase /creatine kinase or isoenzymes test system
Product CodeMMI
FDA ClassificationClass II
Review PanelClinical Chemistry (75)

Predicate Device Information 4.

Predicate Device Name: Elecsys Troponin T Gen 5 STAT Immunoassay

510(k) Number: K162895

Intended Use / Indications for Use 5.

The Atellica® IM High-Sensitivity Troponin I (TnIH) assay is for in vitro diagnostic use in the quantitative measurement of cardiac troponin I in human serum or plasma (lithium heparin) using the Atellica® IM Analyzer. The assay can be used to aid in the diagnosis of acute myocardial infarction (AMI).

Device Description 6.

Table 2. Summary of Ingredients of the Atellica IM TnlH Assay Components

{4}------------------------------------------------

510(k) Summary of Safety and Effectiveness

ComponentVolumeIngredients
Atellica IM TnIH Primary Reagent ReadyPack (included in assay kit)
Atellica IM TnIHLite Reagent8.0 mL/packBovine serum albumin (BSA) conjugated to arecombinant monoclonal (sheep) Fab anti-human cTnl(~0.2-0.4 µg/mL) labeled with acridinium ester inHEPES buffer; stabilizers; preservatives
Atellica IM TnIHSolid Phase Reagent13.0 mL/packStreptavidin-coated magnetic latex particles (0.45mg/mL) with 2 biotinylated (mouse and sheep)monoclonal anti-troponin I antibodies in buffer;stabilizers; preservatives
Atellica IM TnIH Calibrator (included in assay kit)
Atellica IM TnIHHigh Calibrator (Cal H)1.0 mL/vial(lyophilized)Human serum; human cTnl; preservatives
Atellica IM TnIHLow Calibrator (Cal L)1.0 mL/vialHEPES buffer; bovine serum albumin (BSA);surfactants; preservatives

Purpose of the Submission 7.

The purpose of this premarket notification is to submit a new device (Atellica IM TnIH) to FDA for consideration for clearance.

8. Comparison of Predicate Device and Modified Device

ItemAtellica IM TnIH(Candidate Device)Elecsys Troponin T Gen 5 STATImmunoassay(Predicate Device)
Intended UseThe Atellica® IM High-SensitivityTroponin I (TnIH) assay is for in vitrodiagnostic use in the quantitativemeasurement of cardiac troponin Iin human serum or plasma using theAtellica® IM Analyzer. The assay canbe used to aid in the diagnosis ofacute myocardial infarction (AMI).Immunoassay for the in vitroquantitative determination ofcardiac troponin T (cTnT) inlithium heparin plasma. Theimmunoassay is intended to aid inthe diagnosis of myocardialinfarction. Theelectrochemiluminescenceimmunoassay "ECLIA" is intendedfor use on the cobas systemanalyzers.
Indicationsfor UseThe assay can be used to aid in thediagnosis of acute myocardial infarction(AMI).The immunoassay is intended toaid in the diagnosis of myocardialinfarction.
MethodologyChemiluminescenceElectrochemiluminescence
Assay ProtocolSandwich immunoassaySame
AnalyteCardiac troponin ICardiac troponin T
Specimen TypeLithium heparin plasma and serumLithium heparin plasma
Lower Limit ofMeasuring RangeLoQSame

Table 3. Comparison of Atellica IM TnIH Assay to Predicate

{5}------------------------------------------------

ItemAtellica IM TnIH(Candidate Device)Elecsys Troponin T Gen 5 STATImmunoassay(Predicate Device)
Measuring Range2.50-25,000 pg/mL (ng/L)6.0-10,000 pg/mL (ng/L)
Upper 99thPercentile CutoffFemale-Lithium Heparin: 34.11 pg/mLMale-Lithium Heparin: 53.48 pg/mLCombined-Lithium Heparin: 45.20 pg/mLFemale-Serum: 38.64 pg/mLMale-Serum: 53.53 pg/mLCombined-Serum: 45.43 pg/mLOverall: 45.20 pg/mLFemale: 14 pg/mL (ng/L)Male: 22 pg/mL (ng/L)Combined: 19 pg/mL (ng/L)
Calibration2-point calibrationSame

Table 3. Comparison of Atellica IM TnIH Assay to Predicate

9. Standard/Guidance Document References

The following recognized standards from Clinical Laboratory Standards Institute (CLSI) were used as a basis of the study procedures described in this submission:

  • Evaluation of Precision Performance of Quantitative Measurement Methods; ■ Approved Guideline – Third Edition (CLSI EP05-A3, 2014; Recognition No. 7-251)
  • I Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline (CLSI EP06-A, 2003; Recognition No. 7-193)
  • l Interference Testing in Clinical Chemistry; Approved Guideline - Second Edition (CLSI EP07-A2, 2005; Recognition No. 7-127)
  • I Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline - Second Edition (CLSI EP17-A2, 2012; Recognition No. 7-233)
  • I Defining. Establishing and Verifying Reference Intervals in the Clinical Laboratory: Approved Guideline - Third Edition (CLSI EP28-A3c - formerly C28-A3c, 2010; Recognition No. 7-224)
  • l Medical devices – Application of risk management to medical devices (ANSI/AAMI/ISO 14971:2007/(R)2010; Recognition No. 5-70)

{6}------------------------------------------------

10. Performance Characteristics

10.1 Precision

A 20-day precision study was performed according to CLSI EP5-A3. Samples included eight (8) samples (4 serum; 4 lithium heparin plasma) from AMI patients. These samples were diluted with native serum or lithium heparin plasma from healthy subjects. Samples were assayed twice a day in replicates of 2, for 20 days (n = 80 replicates per sample). Testing was performed on 2 instruments. The following are representative of the results obtained:

SampleMeanRepeatabilityWithin-Lab
(pg/mL)SD(pg/mL)%CVSD(pg/mL)%CV
Serum 112.720.554.30.594.7
Serum 2127.932.301.83.092.4
Serum 31334.9722.281.727.482.1
Serum 413815.89192.051.4266.911.9
Lithium Heparin Plasma 112.030.494.10.645.3
Lithium Heparin Plasma 2131.212.231.72.752.1
Lithium Heparin Plasma 31363.3827.112.032.002.3
Lithium Heparin Plasma 412862.97212.911.7291.002.3

10.2 Linearity

Two linearity studies were performed according to CLSI EP06-A, each using 9 samples prepared by mixing a high-spiked cTnl sample with a low cTnl sample. The first study spanned the assay range and the second study ranged to ~150 pg/mL. Each study was tested with lithium heparin plasma and serum. The mean was taken from each sample tested in duplicate. The following are representative of the results obtained:

SampleDeviation from Linear Fit (% or pg/mL)
Li HepFull RangeSerumFull RangeLi Hep~150 pg/mLSerum~150 pg/mL
A-0.53%0.27%2.23%0.73%
B3.26%-6.46%-0.07%-0.04%
C2.08%-4.50%-0.11%-0.06%
D0.78%-2.95%-0.09%-0.05%
E-0.01%-0.27%-0.04%-0.03%
F-1.40%1.20%0.01%-0.01%
G-2.49%3.24%0.07%0.01%
H-3.50%5.43%0.13%0.03%
I-4.58%7.47%0.19%0.06%

{7}------------------------------------------------

10.3 Dilution Recovery

Eight (8) native AMI samples (4 lithium heparin plasma and 4 serum) measuring above the analytical measuring range (i.e. >25000 pg/mL) were diluted to 1:2 and 1:5 with Multi-Diluent 11. Recoveries for individual samples were all within 20%. The mean of all 1:2 dilutions was 99.6%. The mean of all 1:5 dilutions was 91.6%.

10.4 Hook Effect

A study was performed to evaluate hook effect. There is no hook effect with the Atellica IM TnIH assay up to 500,000 pg/mL.

10.5 Detection Limit

The limit of blank (LoB), limit of detection (LoD), and the limit of quantitation (LoQ) were determined as described in CLSI protocol EP17-A2. The Atellica IM TnIH assay has an LoB of 0.50 pg/mL, an LoD of 1.60 pg/mL, and an LoQ of 2.50 pg/mL.

The LoB is defined as the highest measurement result that is likely to be observed for a blank sample. The LoD is defined as the lowest concentration of cardiac troponin I that can be detected with 95% probability. The LoQ is defined as the lowest concentration of cardiac troponin I that can be detected at a total CV of 20%.

10.6 Endogenous Interference

Endogenous interference studies were performed according to CLSI EP07-A2. Sample pools for each matrix (~60 pg/mL cTnl) were spiked with potential interferents. Control samples were prepared by spiking sample pools with the appropriate diluent at the same volume as the interfering substance stock. For substances spiked at doses that caused >10% interference, serial measurements were taken and analyzed by linear regression. Results are presented below.

Endogenous SubstanceMatrixControlDose(pg/mL)TestDose(pg/mL)% Interference
Bilirubin (Conjugated)40 mg/dLLi Hep58.0857.14-1.61%
Serum59.2457.95-2.17%
Bilirubin (Unconjugated)60 mg/dLLi Hep58.7858.75-0.06%
Serum58.8758.26-1.04%
Biotin3500 ng/mLLi Hep59.1658.86-0.51%
Serum59.4059.570.28%
Cholesterol500 mg/dLLi Hep50.3351.903.12%
Serum51.4551.470.04%
Hemoglobin500 mg/dLLi Hep57.2956.55-1.29%
Serum58.1055.31-4.81%
Protein (Albumin)6 g/dLLi Hep57.5756.18-2.40%
Serum57.6458.361.25%
Protein (Gamma Globulin)2.5 g/dLLi Hep53.4453.28-0.30%
Serum52.6854.783.99%
Protein (Total)12 g/dLLi Hep54.8957.354.48%
Serum57.0456.33-1.25%
Triglycerides2000 mg/dLLi Hep51.8352.050.43%
Serum52.2051.84-0.70%

{8}------------------------------------------------

10.7 Drug Interference

Therapeutic drug interference studies were performed according to CLSI EP07-A2. Sample pools (~60 pq/mL cTnl) for each matrix were tested. Control samples were prepared by spiking sample pools with the appropriate diluent at the same volume as the interfering substance stock. At the tested concentrations, all drugs caused <10% interference on the Atellica IM TnIH assay.

DrugLow ConcentrationHigh Concentration
Abciximab5 µg/mL40 µg/mL
Acetaminophen20 µg/mL200 µg/mL
Acetylsalicylic Acid261 µg/mL652 µg/mL
Allopurinol13 µg/mL40 µg/mL
Amiodarone1.8 µg/mL6.1 µg/mL
Ampicilin10 µg/mL53 µg/mL
Ascorbic Acid12 µg/mL60 µg/mL
Atenolol1.1 µg/mL10 µg/mL
Caffeine12 µg/mL60 µg/mL
Captopril1.0 µg/mL5.0 µg/mL
Cefoxitin120 µg/mL660 µg/mL
Cinnarizine200 ng/mL400 ng/mL
Clopidogrel37.5 µg/mL75 µg/mL
Cocaine0.1 µg/mL10 µg/mL
Digoxin1.4 ng/mL6.1 ng/mL
Digitoxin30 ng/mL60 ng/mL
Diltiazem0.2 µg/mL6.2 µg/mL
Disopyramide3.5 µg/mL10 µg/mL
Dopamine0.3 µg/mL0.9 µg/mL
Doxycycline10.0 µg/mL30 µg/mL
Erythromycin11 µg/mL60 µg/mL
Furosemide20 µg/mL60 µg/mL
Ibuprofen40 µg/mL500 µg/mL
Isosorbide Dinitrate50 ng/mL150 ng/mL
Lisinopril0.10 µg/mL0.30 µg/mL
Lovastatin40 ng/mL80 ng/mL
Low MW Heparin6.75 U/mL30 U/mL
Methotrexate546 µg/mL910 µg/mL
Methyldopa4.2 µg/mL15 µg/mL
MethylprednisoloneN/A40 µg/mL
Mexiletine1.3 µg/mL4.0 µg/mL
Nicotine37 ng/mL1000 ng/mL
Nifedipine125 ng/mL400 ng/mL
Nitrofurantoin2.0 µg/mL4.0 µg/mL
Nitroglycerine7.5 ng/mL160 ng/mL
Phenobarbital24 µg/mL97 µg/mL
Phenytoin12 µg/mL50 µg/mL
Primidone10.5 µg/mL40 µg/mL
Propanolol0.50 µg/mL2.0 µg/mL
DrugLow ConcentrationHigh Concentration
Quinidine3.7 µg/mL12 µg/mL
Simvastatin16 µg/mL32 µg/mL
Theophylline12 µg/mL40 µg/mL
Thyroxine0.08 µg/mL1.01 µg/mL
Tissue Plasminogen Activator1.15 µg/mL2.3 µg/mL
Trimethoprim12 µg/mL40 µg/mL
Verapamil0.33 µg/mL2.0 µg/mL
Warfarin2.0 µg/mL10 µg/mL

{9}------------------------------------------------

510(k) Summary of Safety and Effectiveness

Cross-Reactivity 10.8

Cross-reactivity studies were performed using two sample pools per matrix of approximately 0 pg/mL and 60 pg/mL cTnl. These sample pools were spiked with potential cross-reactants. Control samples were prepared by spiking sample pools with the appropriate diluent at the same volume as the interfering substance stock. Results are presented below.

PotentialCross-ReactingSubstanceMatrixcTnlLevelControlDose(pg/mL)TestDose(pg/mL)% Cross-Reactivity
ActinLi HepZero1.471.220.00%
Li HepLow56.2855.180.00%
SerumZero1.041.220.00%
SerumLow56.3155.290.00%
Cardiac Troponin TLi HepZero1.4021.710.00%
Li HepLow55.3875.920.00%
SerumZero0.9322.260.00%
SerumLow55.5678.490.00%
CK-MBLi HepZero1.571.900.00%
Li HepLow55.9256.650.00%
SerumZero1.181.320.00%
SerumLow54.1255.370.00%
MyoglobinLi HepZero1.161.390.00%
Li HepLow55.4355.050.00%
SerumZero1.281.030.00%
SerumLow55.6154.650.00%
Myosin Light ChainLi HepZero1.361.230.00%
Li HepLow55.3955.790.00%
SerumZero0.931.090.00%
SerumLow53.9053.300.00%
Skeletal Troponin ILi HepZero1.590.960.00%
Li HepLow55.5253.550.00%
SerumZero1.070.480.00%
SerumLow53.9054.760.00%
TropomyosinLi HepZero1.471.040.00%
Li HepLow56.2256.310.00%
SerumZero0.941.010.00%
SerumLow54.1554.560.00%

{10}------------------------------------------------

510(k) Summary of Safety and Effectiveness

PotentialCross-ReactingSubstanceMatrixcTnlLevelControlDose(pg/mL)TestDose(pg/mL)% Cross-Reactivity
Troponin CLi HepZero1.6413.510.00%
Troponin CLi HepLow56.2669.650.00%
Troponin CSerumZero1.1215.380.00%
Troponin CSerumLow55.4569.160.00%

10.9 Heterophile Interference

Heterophile interference studies were performed using lithium heparin plasma and serum spiked with high RF samples, and serum spiked with high HAMA samples. Control samples were prepared by spiking sample pools with the appropriate diluent at the same volume as the interfering substance stock. No interference with HAMA or RF was observed.

10.10 High-Sensitivity Designation

The International Federation of Clinical Chemistry (IFCC) Task Force on Clinical Applications of Cardiac Bio-Markers defines a high-sensitivity troponin test as one that meets the following analytical criteria:1

    1. % CV at the 99th percentile value should be ≤10%
    1. Measurable concentrations should be attainable at a concentration above the LoD for at least 50% of healthy individuals

The Atellica IM TnIH assay meets both of these criteria.

Method Comparison with Predicate Device 10.11

Not applicable.

Matrix Comparison 10.12

Not applicable. All performance studies were performed in all applicable matrices (lithium heparin plasma and serum).

1 Apple FS, Sandoval Y, Jaffe AS, et al. Cardiac troponin assays: Guide to understanding analytical characteristics and their impact on clinical care. Clin Biochem 2017;63:73-81.

{11}------------------------------------------------

10.13 Clinical Studies

A clinical performance study was conducted to evaluate the diagnostic accuracy of the Atellica IM TnIH assay in terms of the clinical concordance between the 90th percentile cutoff and the presence or absence of an adjudicated acute myocardial infarction (AMI) diagnosis. Specimens were collected at 29 sites from different regions across the United States. Testing of specimens was performed at 3 sites.

In this study, the sites enrolled all patients who presented to the emergency department, or ambulatory care center equivalent, with signs or symptoms suspicious for a possible acute coronary syndrome (ACS) event. The diagnosis of AMI was performed by an independent adjudication committee which included cardiologists. The adjudication was based on the Third universal definition of myocardial infarction consensus guideline endorsed by the European Society of Cardiology (ESC), the American College of Cardiology Foundation (ACCF), the American Heart Association (AHA), and the World Heart Federation (WHF).

The clinical concordance study evaluated clinical sensitivity, clinical specificity, positive predictive value (PPV) and negative predictive value (NPV) of the Atellica IM TnlH assay in terms of its correlation to the diagnosis of AMI.

Results were analyzed according to time from presentation to the emergency department.

SensitivitySpecificityPPVNPV
MatrixTimepointNEstimate95% CINEstimate95% CINEstimate95% CINEstimate95% CI
Li HepPlasma0-1.5hr4584.4%71.2-92.340193.5%90.7-95.56459.4%47.1-70.538298.2%96.3-99.1
≥1.5-2.5 hr7989.9%81.3-94.872091.8%89.6-93.613054.6%46.0-62.966998.8%97.7-99.4
≥2.5-3.5 hr7394.5%86.7-97.862191.6%89.2-93.612157.0%48.1-65.557399.3%98.2-99.7
≥3.5-4.5 hr5094.0%83.8-97.948789.5%86.5-91.99848.0%38.3-57.743999.3%98.0-99.8
≥4.5-6 hr2696.2%81.1-99.323887.0%82.1-90.75644.6%32.4-57.620899.5%97.3-99.9
≥6-9 hr6994.2%86.0-97.737488.0%84.3-90.911059.1%49.7-67.833398.8%97.0-99.5
≥9-24 hr7494.6%86.9-97.934287.4%83.5-90.511361.9%52.7-70.430398.7%96.7-99.5
≥24 hr2796.3%81.7-99.311080.9%72.6-87.24755.3%41.2-68.69098.9%94.0-99.8
Serum0-1.5hr4281.0%66.7-90.040793.6%90.8-95.66056.7%44.1-68.438997.9%96.0-99.0
≥1.5-2.5 hr7789.6%80.8-94.672192.0%89.7-93.712754.3%45.7-62.767198.8%97.7-99.4
≥2.5-3.5 hr6792.5%83.7-96.861992.4%90.0-94.210956.9%47.5-65.857799.1%98.0-99.6
≥3.5-4.5 hr4893.8%83.2-97.948490.3%87.3-92.69248.9%38.9-59.044099.3%98.0-99.8
≥4.5-6 hr2696.2%81.1-99.323687.7%82.9-91.35446.3%33.7-59.420899.5%97.3-99.9
≥6-9 hr6395.2%86.9-98.437888.6%85.0-91.410358.3%48.6-67.333899.1%97.4-99.7
≥9-24 hr7394.5%86.7-97.834289.5%85.8-92.310565.7%56.2-74.131098.7%96.7-99.5
≥24 hr2696.2%81.1-99.311182.9%74.8-88.84456.8%42.2-70.39398.9%94.2-99.8

Using the female-specific 99th percentiles for lithium heparin plasma (34.11 pg/mL) and serum (38.64 pg/mL), the following results were obtained.

{12}------------------------------------------------

Using the male-specific 99th percentiles for lithium heparin plasma (53.48 pg/mL) and serum
(53.53 pg/mL), the following results were obtained.
SensitivitySpecificityPPVNPV
MatrixTimepointNEstimate95% CINEstimate95% CINEstimate95% CINEstimate95% CI
Li HepPlasma0-1.5hr10075.0%65.7-82.556391.5%88.9-93.512361.0%52.1-69.154095.4%93.3-96.8
≥1.5-2.5 hr16187.6%81.6-91.890591.2%89.1-92.822163.8%57.3-69.984597.6%96.4-98.5
≥2.5-3.5 hr12889.8%83.4-94.074890.0%87.6-91.919060.5%53.4-67.268698.1%96.8-98.9
≥3.5-4.5 hr9990.9%83.6-95.159392.2%89.8-94.113666.2%57.9-73.655698.4%97.0-99.1
≥4.5-6 hr4092.5%80.1-97.422390.6%86.0-93.85863.8%50.9-74.920598.5%95.8-99.5
≥6-9 hr12491.1%84.8-95.053188.5%85.5-91.017464.9%57.6-71.648197.7%96.0-98.7
≥9-24 hr14193.6%88.3-96.649385.6%82.2-88.420365.0%58.2-71.343197.9%96.1-98.9
≥24 hr3591.4%77.6-97.014388.1%81.8-92.44965.3%51.3-77.112997.7%93.4-99.2
Serum0-1.5hr9877.6%68.3-84.757391.6%89.1-93.612461.3%52.5-69.454796.0%94.0-97.3
≥1.5-2.5 hr16286.4%80.3-90.992591.5%89.5-93.121963.9%57.4-70.086897.5%96.2-98.3
≥2.5-3.5 hr12887.5%80.7-92.276589.9%87.6-91.918959.3%52.1-66.070497.7%96.3-98.6
≥3.5-4.5 hr9988.9%81.2-93.759992.5%90.1-94.313366.2%57.8-73.756598.1%96.5-98.9
≥4.5-6 hr3894.7%82.7-98.521890.8%86.3-94.05664.3%51.2-75.520099.0%96.4-99.7
≥6-9 hr12291.0%84.6-94.952689.9%87.1-92.216467.7%60.2-74.448497.7%96.0-98.7
≥9-24 hr14093.6%88.2-96.649985.6%82.2-88.420364.5%57.7-70.843697.9%96.1-98.9
≥24 hr3786.5%72.0-94.114488.9%82.7-93.04866.7%52.5-78.313396.2%91.5-98.4

Using the overall 99th Percentile (45.20 pg/mL), the following results were obtained for both genders combined.

SensitivitySpecificityPPVNPV
MatrixTimepointNEstimate95% CINEstimate95% CINEstimate95% CINEstimate95% CI
Li HepPlasma0-1.5hr14577.2%69.8-83.396491.9%90.0-93.519058.9%51.8-65.791996.4%95.0-97.4
≥1.5-2.5 hr24090.0%85.6-93.2162590.6%89.1-92.036858.7%53.6-63.6149798.4%97.6-98.9
≥2.5-3.5 hr20192.0%87.5-95.0136990.6%88.9-92.031458.9%53.4-64.2125698.7%97.9-99.2
≥3.5-4.5 hr14992.6%87.3-95.8108090.8%89.0-92.423758.2%51.9-64.399298.9%98.0-99.4
≥4.5-6 hr6697.0%89.6-99.246189.2%86.0-91.711456.1%47.0-64.941399.5%98.3-99.9
≥6-9 hr19392.7%88.2-95.690587.6%85.3-89.629161.5%55.8-66.980798.3%97.1-99.0
≥9-24 hr21594.0%89.9-96.483586.6%84.1-88.731464.3%58.9-69.473698.2%97.0-99.0
≥24 hr6293.5%84.6-97.525383.4%78.3-87.510058.0%48.2-67.221598.1%95.3-99.3
Serum0-1.5hr14078.6%71.1-84.698092.2%90.4-93.818659.1%52.0-65.993496.8%95.5-97.7
≥1.5-2.5 hr23987.9%83.1-91.4164691.0%89.5-92.335858.7%53.5-63.6152798.1%97.3-98.7
≥2.5-3.5 hr19590.8%85.9-94.1138490.8%89.1-92.230558.0%52.4-63.4127498.6%97.8-99.1
≥3.5-4.5 hr14791.2%85.5-94.8108391.0%89.1-92.523257.8%51.3-63.999898.7%97.8-99.2
≥4.5-6 hr6496.9%89.3-99.145489.6%86.5-92.110956.9%47.5-65.840999.5%98.2-99.9
≥6-9 hr18593.5%89.0-96.390488.7%86.5-90.627562.9%57.1-68.481498.5%97.4-99.2
≥9-24 hr21393.9%89.8-96.484186.7%84.2-88.831264.1%58.6-69.274298.2%97.0-99.0
≥24 hr6390.5%80.7-95.625585.5%80.6-89.39460.6%50.5-69.922497.3%94.3-98.8

{13}------------------------------------------------

10.14 Traceability and Value Assignment

The Atellica IM TnIH assay is standardized to an internal standard manufactured using human heart homogenate. Assigned values for calibrators are traceable to this standardization.

10.15 Stability

The Atellica IM TnIH reagents and calibrators are stable until the date printed on the box label when stored at 2-8°C.

The onboard stability of the Atellica IM TnIH reagents is 28 days with a pack calibration interval of 31 days, and a lot calibration interval of 47 days.

Conclusions 11.

The Atellica IM High-Sensitivity Troponin I (TNIH) assay (Reagents and Calibrators) is substantially equivalent in principle and performance to the currently-marketed predicate device, the Elecsys Troponin T Gen 5 STAT Immunoassay, cleared under 510(k) K162895.

§ 862.1215 Creatine phosphokinase/creatine kinase or isoenzymes test system.

(a)
Identification. A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.(b)
Classification. Class II.