K Number
K171543

Validate with FDA (Live)

Date Cleared
2017-11-16

(174 days)

Product Code
Regulation Number
888.3358
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Renovis Surgical Hip Replacement System is indicated for patients suffering from:

  1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
  2. Rheumatoid arthritis;
  3. Correction of functional deformity;
  4. Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques; and
  5. Revision procedures where other treatment or devices have failed.
    The Renovis A400 Hip System is intended for cementless applications unless used with the Renovis Cemented Hip Stem.
    The Renovis Surgical Porous Coated Acetabular Shell System is intended for cementless applications.
    The Cemented Hip Stem is intended for cemented applications.
    The Renovis Tesera Trabecular Technologies (T3) Acetabular Shell System is intended for cementless applications.
    The Bipolar Head is for use in conjunction with Renovis femoral heads and femoral stems.
    Bipolar outer heads are not for use with acetabular shells and liners.
Device Description

The subject of this application is Renovis Extended Lateral Offset Femoral Stems which are an expansion of the FDA cleared Renovis Surgical Hip Replacement System. Extended Lateral Offset Femoral Stems are provided in a variety of sizes to fit patient need. This application also includes a new device specific neck trial and X-ray template, and additional implants and instruments.
The Extended Lateral Offset Femoral Stems are manufactured from the same material as other FDA cleared Renovis femoral stems. The instruments are manufactured from the same material as other instruments included in FDA cleared submissions.

AI/ML Overview

This document is a 510(k) summary for a medical device (Renovis Surgical Hip Replacement System), not a clinical study report. Therefore, it does not contain the information required to answer your questions regarding acceptance criteria and a study proving device performance in the context of an AI/human-in-the-loop system.

The document discusses the substantial equivalence of an expansion to an existing device (Extended Lateral Offset Femoral Stems) to previously cleared predicate devices. It relies on bench performance testing and material equivalence, rather than a clinical study with human subjects or AI-assisted diagnostic performance.

Here's why the requested information cannot be extracted from this document:

  • AI/Machine Learning Context: The document is about a physical medical device (hip prosthesis components), not a software or AI-driven diagnostic/decision support system. Therefore, concepts like "acceptance criteria" for AI performance, "test set," "training set," "ground truth," "experts," or "MRMC studies" are not applicable in this context.
  • Study Type: The "study" mentioned is "Performance Testing - Bench," which refers to laboratory tests of the device's physical properties and durability, not clinical trials or comparative effectiveness studies involving human readers or AI.
  • Device Performance: The "reported device performance" refers to successful completion of specific ISO and ANSI/AAMI standards for physical and sterilisation properties, not diagnostic accuracy metrics (e.g., sensitivity, specificity, AUC).

Summary of what can be extracted, and why the rest cannot:

  1. A table of acceptance criteria and the reported device performance:

    • Acceptance Criteria (implicit): Compliance with ISO 7206-4, ISO 7206-6, ANSI/AAMI/ISO 11137-1, ANSI/AAMI/ISO 11137-2, ANSI/AAMI ST72:2011(R)2016 for bacterial endotoxin testing, and material standards like ASTM F136-13, ASTM A564-13, ASTM A276-17.
    • Reported Device Performance: "Renovis successfully completed the following performance tests." (Lists the standards). "endotoxin limits are in compliance with ANSI/AAMI ST72:2011(R)2016." The results indicate the device met the requirements of these standards.
    • Why partial: These are engineering and material performance criteria for a physical implant, not diagnostic performance metrics for an AI system.
  2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Cannot be provided. This refers to bench testing of physical components, not a clinical study or AI model evaluation. There is no "test set" of patient data in the sense you're asking about.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Cannot be provided. This is not relevant to a physical device's bench testing. "Ground truth" in this context would be physical measurements and material properties confirmed by engineering and quality control experts, not clinical diagnosis.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Cannot be provided. This is for clinical or AI performance evaluation, not bench testing.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Cannot be provided. This type of study is for AI-assisted diagnostic tools, not physical implants.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Cannot be provided. This is about an algorithm's performance, which is not applicable to this device.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Cannot be provided (in the AI/clinical sense). For bench testing, the "ground truth" is adherence to established engineering standards (e.g., fatigue limits, material composition specifications).
  8. The sample size for the training set:

    • Cannot be provided. There is no "training set" for a physical device in this context, as there is no AI model being trained.
  9. How the ground truth for the training set was established:

    • Cannot be provided. Not applicable.

In conclusion, the provided FDA document is for a traditional medical device (hip prosthesis) and focuses on engineering, material, and sterilization compliance, not AI algorithm performance or clinical diagnostic accuracy.

{0}------------------------------------------------

November 16, 2017

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services - USA logo. To the right of that is the FDA logo in blue. Underneath the FDA logo is the word "ADMINISTRATION".

Renovis Surgical Technologies % Sharyn Orton Senior Consultant MEDIcept, Inc 200 Homer Ave Ashland, Massachusetts 01721

Re: K171543

Trade/Device Name: Renovis Surgical Hip Replacement System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint Metal/Polymer/Metal Semi-Constrained Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: OQG, OQI, LPH, LZO Dated: October 12, 2017 Received: October 13, 2017

Dear Sharyn Orton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good

{1}------------------------------------------------

manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Katherine D. Kavlock -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K171543

Device Name Renovis Surgical Hip Replacement System Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

Indications for Use (Describe)

The Renovis Surgical Hip Replacement System is indicated for patients suffering from:

    1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis:
  1. Rheumatoid arthritis;

  2. Correction of functional deformity;

  3. Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement. unmanageable using other techniques: and

  4. Revision procedures where other treatment or devices have failed.

The Renovis A400 Hip System is intended for cementless applications unless used with the Renovis Cemented Hip Stem. The Renovis Surgical Porous Coated Acetabular Shell System is intended for cementless applications.

The Cemented Hip Stem is intended for cemented applications.

The Renovis Tesera Trabecular Technologies (T3) Acetabular Shell System is intended for cementless applications.

The Bipolar Head is for use in conjunction with Renovis femoral heads and femoral stems.

Bipolar outer heads are not for use with acetabular shells and liners.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden. to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

ORENOVIS

Traditional 510(k) Summary as required by 21 CFR 807.92(a) K171543

A ) Submitted by:Renovis Surgical Technologies1901 W. Lugonia Ave, Ste 340Redlands, CA 92374Phone: 909-557-2360Fax: 909-307-8571
Official Contact:Anthony DeBenedictisVice President of Quality Assurance
Consultant:Sharyn Orton, Ph.D.MEDIcept, Inc.200 Homer AveAshland, MA 01721
B) Classification Name:Prosthesis, Hip Revision System
Proprietary Name:Renovis Surgical Hip Replacement System
Device Class:Class II
Regulations21 CFR 888.3358 Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis21 CFR 888.3353 Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis
and Product Codes:OQG, OQI, LPH, LZO
Classification panel:Orthopedic
C) Predicate:K112897 Renovis A400 Surgical Hip Joint ReplacementProsthesis (Tapered Lateral Offset Femoral Stems)K141676 Renovis Surgical Porous Acetabular Cup SystemK143647 Renovis Surgical Hip Replacement System
D) Date Prepared:November 13, 2017

{4}------------------------------------------------

E) Device Description:

The subject of this application is Renovis Extended Lateral Offset Femoral Stems which are an expansion of the FDA cleared Renovis Surgical Hip Replacement System. Extended Lateral Offset Femoral Stems are provided in a variety of sizes to fit patient need. This application also includes a new device specific neck trial and X-ray template, and additional implants and instruments.

The Extended Lateral Offset Femoral Stems are manufactured from the same material as other FDA cleared Renovis femoral stems. The instruments are manufactured from the same material as other instruments included in FDA cleared submissions.

Renovis implants and instruments comply with the following material standards:

  • ASTM F136-13 Standard Specification for Wrought Titanium-6Aluminum-. 4 Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications
  • ASTM A564-13 Standard Specification for Hot-Rolled and Cold-Finished Age-● Hardening Stainless Steel Bars and Shapes
  • ASTM A276-17 Standard Specification for Stainless Steel Bars and Shapes ●

E) Intended Use/Indications For Use:

The Renovis Surgical Hip Replacement System is indicated for patients suffering from: 1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis;

    1. Rheumatoid arthritis:
    1. Correction of functional deformity;
  1. Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques; and
    1. Revision procedures where other treatment or devices have failed.
      The Renovis A400 Hip System is intended for cementless applications unless used with the Renovis Cemented Hip Stem.

The Renovis Surgical Porous Coated Acetabular Shell System is intended for cementless applications.

The Cemented Hip Stem is intended for cemented applications.

The Renovis Tesera Trabecular Technologies (T3) Acetabular Shell System is intended for cementless applications.

The Bipolar Head is for use in conjunction with Renovis femoral heads and femoral stems.

Bipolar outer heads are not for use with acetabular shells and liners.

{5}------------------------------------------------

F) Substantial Equivalence Comparison and Discussion

Additional predicate devices include Renovis FDA cleared Tapered Lateral Offset Femoral Stems in K141676 and K143647. The Extended Lateral Offset Femoral Stems are manufactured in a similar way, from the same material; have the same sizes; same sterilization; and same shelf-life as the Tapered Lateral Offset Femoral Stems in FDA cleared K112897, K141676 and K143647.

The difference between the predicate and new femoral stems is the offering of an extended lateral offset. The results of performance testing support that the Extended Lateral Offset Femoral Stems do not raise new issues of safety and effectiveness and are substantially equivalent to the predicate devices.

G) Performance Testing - Bench

Renovis successfully completed the following performance tests:

  • ISO 7206-4: 2010 Implants for surgery - Partial and total hip-joint prostheses - Part 4: Determination of endurance properties and performance of stemmed femoral components
  • ISO 7206-6: 2013 Implants for surgery - Partial and total hip joint prostheses - Part 6: Determination of endurance properties of head and neck region of stemmed femoral components
  • ANSI/AAMI/ISO 11137-1:2006 (R2010) and A1:2013 Sterilization of health care ● products - Radiation and Sterilization of health care products
  • ANSI/AAMI/ISO 11137-2:2013 Radiation Part 2: Establishing the sterilization dose ●

Pyrogenicity Testing

Bacterial endotoxin testing was conducted in compliance with ANSI/AAMI ST72:2011(R)2016 Bacterial endotoxins - Test methods, routine monitoring, and alternatives to batch testing, and endotoxin limits are in compliance with ANSVAAMI ST72:2011(R)2016.

Biocompatibility, sterilization, and shelf life

There is no change in material, sterilization method or shelf-life from the predicate devices.

H) Additional Consensus Standards and Guidance

  • ASTM F565-04 (reapproved 2013) Standard Practice for Care and Handling of ● Orthopedic Implants and Instruments
  • ASTM F983-86 (reapproved 2013) Standard Practice for Permanent Marking of Orthopaedic Implant Components

{6}------------------------------------------------

  • Class II Special Controls Guidance Document: Hip Joint Metal/Polymer Constrained ● Cemented or Uncemented Prosthesis; guidance for Industry and FDA" dated April 30, 2002
  • Non-clinical Information for Femoral Stem Prostheses" dated September 17, 2007. ●
  • Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile Guidance for Industry and Food and Drug Administration Staff," January 2016, updated March 16, 2016

Conclusion

The Renovis Extended Lateral Offset Femoral Stems are substantially equivalent to the FDA cleared Renovis Tapered Lateral Offset Femoral Stems.

§ 888.3358 Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis.

(a)
Identification. A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.(b)
Classification. Class II.