K Number
K171131

Validate with FDA (Live)

Date Cleared
2017-11-08

(205 days)

Product Code
Regulation Number
880.5860
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MHC Standard and NRFit™ Tip Syringes are intended to inject fluids into or withdraw fluids from the body.

Device Description

The MHC Standard and NRFit™ Tip Syringes are provided sterile or in bulk non-sterile for further processing (e.g., sterilization). They are single use devices consisting of rigid polycarbonate barrels and plungers with a synthetic rubber gasket. The syringe barrels with neuraxial tips are ISO 80369 - 6 compliant and allow for connections of neuraxial specific applications while reducing the likelihood of misconnections to non-neuraxial devices. The syringe barrels are printed with graduated markings in cc (millliters) indicating the volume of liquid inside the syringe barrel.

AI/ML Overview
  1. Table of Acceptance Criteria and Reported Device Performance:
TestAcceptance Criteria (ISO 80369-6:2016 & ISO 80369-20:2016)Reported Device Performance
Visual Inspection (burrs, hooks, cracks, foreign contamination, missing components)No visual defects adversely affecting performance or safety.Device passed visual inspection.
Stress Cracking (ISO 80369-6, sec. 6.3 & ISO 80369-20, Annex E)No evidence of stress cracking after testing.Device passed stress cracking test.
Fluid Leakage by Pressure Decay (ISO 80369-20, Annex B)Meet specified leakage limits.Device met fluid leakage requirements.
Subatmospheric Pressure Air Leakage (ISO 80369-20, Annex D)Meet specified leakage limits.Device met subatmospheric pressure air leakage requirements.
Resistance to Separation from Axial Load (ISO 80369-20, Annex F)Maintain connection under specified axial load.Device maintained connection under specified axial load.
BiocompatibilityMeet guidelines in FDA guidance document "Use of International Standard ISO10993-1."Device met biocompatibility guidelines.
  1. Sample Size Used for the Test Set and Data Provenance:

    • The document does not explicitly state the specific sample size used for each non-clinical test. It generally refers to "verification and validation testing" performed with the subject neuraxial (NRFit) Tip Syringes.
    • Data Provenance: The tests are non-clinical, implying they were conducted in a laboratory setting. The country of origin of the data is not specified, but the testing was performed to international standards (ISO 80369-6 and ISO 80369-20) and FDA guidance for biocompatibility. The data is prospective, generated specifically for this submission.
  2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • This information is not applicable. The "ground truth" for the test set is established by the specified acceptance criteria within the ISO standards and FDA guidance rather than expert consensus on medical images or diagnoses. The "experts" would be the personnel conducting the laboratory tests according to the standard operating procedures of the testing facility.
  3. Adjudication Method for the Test Set:

    • This information is not applicable. Adjudication methods like 2+1 or 3+1 typically refer to the resolution of discrepancies between multiple human readers in diagnostic studies. For non-clinical device performance testing against defined standards, results are objective measurements compared against predetermined thresholds.
  4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    • No, an MRMC comparative effectiveness study was not done. The submission explicitly states: "Clinical tests were not required to demonstrate performance of between the subject and predicate devices. Product functionality has been adequately assessed by non-clinical tests."
  5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • This information is not applicable as the device is a medical syringe, not an AI algorithm.
  6. The Type of Ground Truth Used:

    • The ground truth used for the acceptance criteria and testing is based on established international standards (ISO 80369-6:2016 and ISO 80369-20:2016) and FDA guidance documents (ISO10993-1 for biocompatibility). These standards define objective performance thresholds and test methods to ensure the safety and efficacy of the device.
  7. The Sample Size for the Training Set:

    • This information is not applicable. The device is a physical medical device (syringe) and does not involve AI or machine learning algorithms that would require a "training set."
  8. How the Ground Truth for the Training Set was Established:

    • This information is not applicable as there is no training set for this type of device submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 8, 2017

Mazza Healthcare, LLC. % Jon Gilbert Regulatory Consultant to Mazza Healthcare, LLC Jon Gilbert 1641 Jeurissen Lane Chanhassen, Minnesota 55317

Re: K171131

Trade/Device Name: MHC Standard and NRFit™ Tip Syringes Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF Dated: September 29, 2017 Received: October 2, 2017

Dear Jon Gilbert:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Tina
Kiang-S

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K171131

Device Name

MHC Standard and NRFit™ Tip Syringes

Indications for Use (Describe)

The MHC Standard and NRFitM Tip Syringes are intended to inject fluids into or withdraw fluids from the body.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5. 510(k) Summary

Submitter

Sponsor:Mazza Healthcare, LLC2101 Waukegan Road, Suite 208Bannockburn, IL 60015Jon Gilbert, Consultant+01 (906) 361-3237igilb.raca@gmail.com
Contact:
Date of Preparation:September 29, 2017
Subject Device:Common Name:Trade Name:Requlation Number:Regulation Name:Review Panel:Regulatory Class:Product Code:Standard and Neuraxial Tip SyringesMHC Standard and NRFit™ Tip Syringes21 CFR 880.5860Piston syringeGeneral HospitalClass IIFMF
Primary Predicate:510(k):Vesco Medical NRFit™ Tip SyringesK170218 cleared February 24, 2017
Predicate Device:510(k):DMC Medical Single-Use Polycarbonate SyringeK103736 cleared June 10, 2011

Device Description:

The MHC Standard and NRFit™ Tip Syringes are provided sterile or in bulk non-sterile for further processing (e.g., sterilization). They are single use devices consisting of rigid polycarbonate barrels and plungers with a synthetic rubber gasket. The syringe barrels with neuraxial tips are ISO 80369 - 6 compliant and allow for connections of neuraxial specific applications while reducing the likelihood of misconnections to non-neuraxial devices. The syringe barrels are printed with graduated markings in cc (millliters) indicating the volume of liquid inside the syringe barrel. The proposed models for the standard and neuraxial devices are listed below in Table 1.

MHC Model #Item Description
09-050-WH5cc Polycarbonate Luer Slip Tip Syringe W/Silicone
09-060-WH5cc Polycarbonate Luer Lock Tip Syringe W/Silicone
09-055-WHN5cc Polycarbonate Neuraxial Slip Tip Syringe W/Silicone
09-010-WH10cc Polycarbonate Luer Slip Tip Syringe w/Silicone
09-120-WH10cc Polycarbonate Luer Lock Tip Syringe w/Silicone
09-015-WHN10cc Polycarbonate Neuraxial Slip Tip Syringe w/Silicone

Table 1. MCH Models and Description

{4}------------------------------------------------

A difference between the predicate syringe (K103736) without a neuraxial tip and the subject syringe with a neuraxial tip is the design of the needle s hub. While the predicate devices (K103736) are designed with a luer slip tip (taper) or luer lock connector the subject neuraxial devices are equipped with a NRFit™ neuraxial connection as standardized in ISO 80369-6. Additionally, the NRFit™ Tip Syringe piston is yellow.

The clinical technique, indications for use, technical specifications, materials used, and sterility status (validation and sterility assurance level) are identical between the subject and predicate Standard syringes. As stated above, the NRFit™ tip syringes provides connections of neuraxial specific applications while reducing the likelihood of misconnections to non-neuraxial devices

Indications for Use:

The MHC Standard and NRFit™ Tip Syringes are intended to inject fluids into or withdraw fluids from the body.

Performance Testing:

The performance of the subject MHC Standard Tip Syringes is identical to the predicate devices as no material (with the exception of the yellow piston), manufacturing or technological changes have occurred. The performance of the subject MHC NRFit™ Tip Syringes is similar to the predicate neuraxial syringes (K170218) as both designs are based upon ISO 80369-6, Small bore connectors for liquids and gases in healthcare applications - Part 6: Connectors for neuraxial applications.

Non-Clinical Tests

Verification and validation testing was performed with the subject neuraxial (NRFit) Tip Syringes. It was found that subject neuraxial (NRFit) Tip Syringes are in compliance with design and performance requirements according to ISO 80369-6:2016 "Small-bore connectors for liquids and gases in healthcare applications - Part 6: Connectors for neuraxial applications," and ISO 80369-20:2016 "Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common Test Methods, including Visual Inspection (burrs, hooks, cracks, foreign contamination, missing components), Stress Cracking per ISO 80369-6 (sec. 6.3) and ISO 80369-20 (Annex E), Fluid Leakage by Pressure Decay per ISO 80369-20 (Annex B), Subatmospheric Pressue Air Leakage per ISO 80369-20 (Annex D), and, Resistance to Separation from Axial Load per ISO 80369-20 (Annex F). Biocompatibility testing has demonstrated the subject devices meet the guidelines as described in the FDA guidance document entitled, "Use of International Standard ISO10993-1, " Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".

Clinical and Animal Tests

Clinical tests were not required to demonstrate performance of between the subject and predicate devices. Product functionality has been adequately assessed by non-clinical tests.

{5}------------------------------------------------

Component:Vesco MedicalK170218DMC Medical,Ltd, K103736MHC Standardand Neuraxial*
Syringe Barrel:PolypropylenePolycarbonatePolycarbonate
Plunger Tip:Synthetic RubberElastomerElastomer
PlungerPolypropyleneABS polymerABS polymer
Syringe Piston LubeCoatingPolydimethylsiloxane,Silanol TerminatedMedical GradeOilMedical GradeOil
Calibrated BarrelVolume:YesYesYes
RecommendedSterilization method:EtOEtOEtO
510(k) ApprovalK170218K103736K171131

Table 2 Technological Characteristics & Substantial Equivalence

*Subiect device Standard and NRFit components are comprised of same materials and manufactured using same manufacturing methods. K103736 best represents predicate materials and components while K170218 and ISO 80369-6 represents requisite neuraxial tip dimensional configurations and technological characteristics and standards.

The operational characteristics, when compared to the predicate device, are identical. The user connects the syringe via a standard slip tip or threaded luer lock, then manually advances or withdraws the plunger internal to the barrel to express or withdraw fluids. Fluids are measured via the printed external housing of the barrel: measurements are indicated in cc (milliliters). Operation is similar for most all piston svringes whether fitted with a threaded luer end, a slip-fit tip or neuraxial tip.

The subject device standard syringes are similar in materials and performance to the Sponsor's currently distributed standard svringes cleared in K103736. The technological characteristics of the subject device neuraxial syringes are consistent with ISO standard and are equivalent to the predicate device cleared in K170218.

Conclusion:

The MHC Standard and NRFit™ Tip Syringes are intended to inject fluids into or
withdraw fluids from the body. The materials, intended use, technological and operational characteristics of the subject devices are substantially equivalent to the predicate Standard Syringes of K103736 and the neuraxial syringes cleared in K170218.

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).