K Number
K171070

Validate with FDA (Live)

Date Cleared
2017-08-09

(121 days)

Product Code
Regulation Number
892.1550
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The WS80A Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.

The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.

Device Description

The WS80A is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as B mode, M mode, Color Doppler imaging, Power Doppler imaging (including Directional Power Doppler mode; S-Flow), PW Spectral Doppler mode, CW Spectral Doppler mode, Harmonic imaging, Tissue Doppler imaging, Tissue Doppler Wave, 3D imaging mode (real time 4D imaging mode), Elastoscan Mode or as a combination of these modes. The WS80A also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The WS80A has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

AI/ML Overview

The provided document is a 510(k) Premarket Notification for the SAMSUNG MEDISON WS80A Diagnostic Ultrasound System. It details an ultrasound system and its transducers, along with several software features.

Here's an analysis based on your requested points:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state acceptance criteria in a quantitative format (e.g., "accuracy must be > 90%"). Instead, it states that "Results of all conducted testing and assessments were found acceptable and did not raise any new issues of safety or effective." This implies the acceptance criteria were met by demonstrating functionality and safety equivalent to predicate devices.

The reported device performance is described qualitatively as "found acceptable" for the functionalities listed.

Feature/TestAcceptance Criteria (Implied)Reported Device Performance
IOTA-ADNEX FunctionalityFunctionality equivalent to predicate device (Voluson E8)Functionality assessed and found acceptable.
BiometryAssist FunctionalityFunctionality equivalent to predicate device (Voluson E8)Functionality assessed and found acceptable.
CrystalVue Flow Image PerformanceImage quality equivalent to predicate device (WS80A K153529)Image performance assessed and found acceptable.
Needle Mate Image QualityImage quality equivalent to predicate device (WS80A K153529)Image quality improved from previous version and found acceptable.
Acoustic Output, Biocompatibility, Thermal, Electrical, Electromagnetic, Mechanical SafetyConformance to applicable medical device safety standardsEvaluated and found to conform to applicable medical device safety standards.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample size used for the test set for the software features (IOTA-ADNEX, BiometryAssist, CrystalVue Flow, Needle Mate).

The data provenance is not explicitly mentioned as country of origin nor as retrospective or prospective data. The nature of the tests (bench and non-clinical) suggests controlled testing environment rather than clinical patient data for performance evaluation of new features in this submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not provide information on the number of experts used to establish ground truth for the test set, nor their specific qualifications. For features like IOTA-ADNEX and BiometryAssist, which provide statistical information or aid in measurements, the "ground truth" would likely be established through comparison with established clinical methods or manual expert measurements. However, these details are not present.

4. Adjudication Method for the Test Set

The document does not mention any specific adjudication method (e.g., 2+1, 3+1). The nature of "performance tests" and "image performance assessments" indicates an internal validation process, but details on how discrepancies or uncertainties were resolved are not provided.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

The document states: "Not applicable. The subject of this submission, WS80A, did not require clinical studies to support substantial equivalence." This indicates that no MRMC comparative effectiveness study was done as part of this 510(k) submission. The evaluation focused on equivalence to predicate devices and adherence to safety standards.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

The document describes "performance tests" for the functionality of "IOTA-ADNEX" and "BiometryAssist" and "image performance assessments" for "CrystalVue Flow" and "Needle Mate". While these tests evaluate the software features themselves, the document does not explicitly describe them as standalone algorithm-only performance studies independent of human interaction or interpretation steps. The phrasing "helps the user measure fetal growth indices" for BiometryAssist suggests a human-in-the-loop function. Given the lack of clinical study described, these were likely technical validations rather than full standalone clinical performance evaluations.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

The document does not specify the type of ground truth used for the performance evaluation of the software features. For "IOTA-ADNEX" and "BiometryAssist," ground truth would typically refer to validated clinical diagnoses, pathological findings, or precise manual measurements, but these details are not provided. For "CrystalVue Flow" and "Needle Mate," ground truth would relate to objective image quality metrics or expert assessment of image features, also not detailed.

8. The Sample Size for the Training Set

The document does not provide any information regarding the sample size for the training set. This is expected, as the document describes a 510(k) submission for an ultrasound system and its features, not a particular AI model that would typically have a distinct training set. The features listed (IOTA-ADNEX, BiometryAssist, CrystalVue Flow, Needle Mate) are likely algorithms or advanced processing techniques, but details about their development (e.g., training data) are not included in this regulatory document.

9. How the Ground Truth for the Training Set Was Established

As no information is provided regarding a training set, the document does not describe how ground truth for a training set was established.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features the HHS logo, which consists of a stylized caduceus with three human profiles facing right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circle around the logo.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 9, 2017

SAMSUNG MEDISON CO., LTD. % Emily Woo Regulatory Affairs Manager 42. Teheran-ro 108-gil Gangnam-gu. Seoul 06176 KOREA

Re: K171070

Trade/Device Name: WS80A Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: July 12, 2017 Received: July 25, 2017

Dear Emily Woo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Michael D. O'HaraFor

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

K171070

Device Name WS80A Diagnostic Ultrasound System

Indications for Use (Describe)

The WS80A Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.

The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) No.:

Device Name: WS80A Diagnostic Ultrasound System Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Notes 2, 3,4, 5, 6, 7, 8, 9, 11, 12
Abdominal (See Note 10)PPPPPNote 1Notes 2, 4, 5, 6, 7, 8, 9,10, 11, 12
Intra-operative (See Note 6)PPPPNote 1Note 7, 9, 11
Intra-operative (Neuro.)PPPPNote 1Note 7, 9, 11
Fetal Imaging& OtherLaparoscopic
PediatricPPPPPNote 1Notes 2, 4, 6, 7, 8, 9, 11
Small Organ (See Note 5)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 12
Neonatal CephalicPPPPPNote 1Notes 7
Adult CephalicPPPPPNote 1Notes 7
Trans-rectalPPPPNote 1Notes 2, 7, 8, 9, 11, 12
Trans-vaginalPPPPNote 1Notes 2, 7, 8, 9, 11, 12
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Musculo-skel. (Superfic.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Intra-luminal
Other (spec.) (See Note 13)PPPPNote 1Notes 2, 7,8, 9, 11, 12
Cardiac AdultPPPPPNote 1Notes 4, 7
CardiacCardiac PediatricPPPPPNote 1Notes 4, 7
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Other (spec.)

N= new indication; P= previously cleared by FDA K153529; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CW, B+C+PW, B+DD+PW, B+PPI+PW, B+PPI+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging
    Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

{4}------------------------------------------------

510(k) No.:

Device Name: SC1-6 for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPNote 1Notes 2, 4, 7, 8, 9, 11
Abdominal (See Note 10)PPPPNote 1Notes 2, 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.) (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K153529; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CW, B+C+PW, B+DD+PW, B+PD+PW, B+PPI+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging
    Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

{5}------------------------------------------------

510(k) No.:

Device Name: C2-6 for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal/Obstetrics (See Note 3)PPPPPNote 1Notes 2, 4, 7, 8, 9, 11
Abdominal (See Note 10)PPPPPNote 1Notes 2, 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.) (See Note 13)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K153529; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CW, B+C+PW, B+DD+PW, B+PD+PW, B+PPI+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging
    Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

{6}------------------------------------------------

510(k) No.:

Device Name: E3-12A for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Notes 2, 8, 7, 9
Abdominal (See Note 10)PPPPPNote 1Notes 2, 7, 8, 9, 12
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPNote 1Notes 2, 7, 8, 9, 12
Trans-vaginalPPPPPNote 1Notes 2, 7, 8, 9, 12
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.) (See Note 13)PPPPPNote 1Notes 2, 7, 8, 9, 12
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K153529; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging
    Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

{7}------------------------------------------------

510(k) No.:

Device Name: VR5-9 for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Note 2, 7, 8, 9
Abdominal (See Note 10)PPPPPNote 1Note 2, 7, 8, 9, 12
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPNote 1Note 2, 7, 8, 9, 12
Trans-vaginalPPPPNote 1Note 2, 7, 8, 9, 12
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.) (See Note 13)PPPPNote 1Note 2, 7, 8, 9, 12
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K153529; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

{8}------------------------------------------------

510(k) No.:

Device Name: L3-12A for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPNote 1Note 2, 5, 6, 7, 9, 11
Abdominal (See Note 10)PPPPNote 1Notes 2, 4, 6, 7, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal ImagingLaparoscopic
&OtherPediatricPPPPNote 1Notes 2, 4, 6, 7, 9, 11
Small Organ (See Note 5)PPPPNote 1Note 2, 5, 6, 7, 9, 11, 12
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2, 5, 6, 7, 9, 11
Musculo-skel. (Superfic.)PPPPNote 1Note 2, 5, 6, 7, 9, 11
Intra-luminal
Other (spec.) (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vesselPPPPNote 1Note 2, 5, 6, 7, 9, 11
VesselOther (spec.)

N= new indication; P= previously cleared by FDA153529; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging
    Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

{9}------------------------------------------------

510(k) No.:

Device Name: L5-13 for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)PPPPNote 1Note 2, 5, 6, 7, 9, 11, 12
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2, 5, 6, 7, 9, 11
Musculo-skel. (Superfic.)PPPPNote 1Note 2, 5, 6, 7, 9, 11
Intra-luminal
Other (spec.) (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Note 2, 5, 6, 7, 9, 11
Other (spec.)

N= new indication; P= previously cleared by FDA K153529; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

{10}------------------------------------------------

510(k) No.:

Device Name: V4-8 for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal/Obstetrics (See Note 3)PPPPPNote 1Note 2, 4, 7, 8, 9, 11
Abdominal (See Note 10)PPPPPNote 1Note 2, 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.) (See Note 13)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K153529; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

{11}------------------------------------------------

510(k) No.:

Device Name: V5-9 for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPNote 1Note 2, 7, 8, 9, 11
Abdominal (See Note 10)PPPPNote 1Note 2, 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPNote 1Note 2, 7, 8, 9, 11
Trans-vaginalPPPPNote 1Note 2, 7, 8, 9, 11
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.) (See Note 13)PPPPNote 1Notes 2, 7, 8, 9, 11
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vessel
VesselOther (spec.)

N= new indication; P= previously cleared by FDA K153529; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

{12}------------------------------------------------

510(k) No.:

Device Name: CA1-7A for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal/Obstetrics (See Note 3)PPPPPNote 1Notes 2, 4, 7, 8, 9, 11
Abdominal (See Note 10)PPPPPNote 1Notes 2, 7, 8, 9, 11, 13
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
PediatricPPPPPNote 1Notes 2, 7, 8, 9, 11
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPPNote 1Notes 2, 7, 8, 9, 11
Musculo-skel. (Superfic.)PPPPPNote 1Notes 2, 7, 8, 9, 11
Intra-luminal
Other (See Note 13)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K153529; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

{13}------------------------------------------------

510(k) No.:

Device Name: CA2-8A for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Notes 2, 4, 7, 8, 9, 11
Abdominal (See Note 10)PPPPPNote 1Notes 2, 4, 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vessel
VesselOther (spec.)

N= new indication; P= previously cleared by FDA K153529; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging
    Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

{14}------------------------------------------------

510(k) No.:

Device Name: CF4-9 for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPNote 1Notes 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
PediatricPPPPNote 1Notes 8, 9, 11
Small Organ (See Note 5)
Neonatal CephalicPPPPNote 1Notes 8, 9, 11
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Notes 8, 9, 11
Other (spec.)

N= new indication; P= previously cleared by FDA K153529; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+TD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPF+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

{15}------------------------------------------------

510(k) No.:

Device Name: LA3-16A for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
(Track I only)(Tracks I & III)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal (See Note 3)
Abdominal (See Note 10)
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
Pediatric
Small Organ (See Note 5)PPPPPNote 1Note 2, 5, 6, 7, 9, 11, 12
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPPNote 1Note 2, 5, 6, 7, 9, 11
Musculo-skel. (Superfic.)PPPPPNote 1Note 2, 5, 6, 7, 9, 11
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPPNote 1Note 2, 5, 6, 7, 9, 11
Other (spec.)

N= new indication; P= previously cleared by FDA K153529; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+TD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPF+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

  • Note 3: Includes infertility monitoring of follicle development
  • Note 6: Abdominal organs and peripheral vessel
  • Note 7: Tissue Harmonic Imaging (THI)
  • Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging
    Note 12: ElastoScan

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

{16}------------------------------------------------

510(k) No.:

Device Name: PE2-4 for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPPNote 1Note 7
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal CephalicPPPPPNote 1Note 7
Adult CephalicPPPPPNote 1Note 7
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac AdultPPPPPNote 1Note 4, 7
CardiacCardiac PediatricPPPPPNote 1Note 4, 7
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vessel
VesselOther (spec.)

N= new indication; P= previously cleared by FDA K153529; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+DD+PW, B+DPD+PW, B+PPF+PW, B+TD+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

{17}------------------------------------------------

510(k) No.:

Device Name: CV1-8A for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPNote 1Note 2, 4, 7, 8, 9, 11
Abdominal (See Note 10)PPPPNote 1Note 2, 7, 8, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.) (See Note 13)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K153529; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+TD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPF+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

  • Note 3: Includes infertility monitoring of follicle development
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)
  • Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 4: Color M-mode

{18}------------------------------------------------

510(k) No.:

Device Name: LV3-14A for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)PPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Musculo-skel. (Superfic.)PPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Other (spec.)

N= new indication; P= previously cleared by FDA K153529; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+TD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPF+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

{19}------------------------------------------------

510(k) No.:

Device Name: CA2-9A for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Notes 2, 4, 7, 8, 9, 11
Abdominal (See Note 10)PPPPPNote 1Notes 2, 4, 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K153529; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging
    Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

{20}------------------------------------------------

510(k) No.:

Device Name: CA3-10A for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPNote 1Notes 2, 3, 7, 8, 9, 11
Abdominal (See Note 10)PPPPNote 1Notes 2, 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote 1Notes 2, 7, 8, 9, 11
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Notes 2, 7, 8, 9, 11
Other (spec.)

N= new indication; P= previously cleared by FDA K153529; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+TD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPF+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

  • Note 3: Includes infertility monitoring of follicle development
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)
  • Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 4: Color M-mode

{21}------------------------------------------------

510(k) No.:

Device Name: EA2-11B for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPNote 1Note 2, 7, 8, 9
Abdominal (See Note 10)PPPPNote 1Note 2, 7, 8, 9, 12
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal ImagingLaparoscopic
&OtherPediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPNote 1Note 2, 7, 8, 9, 12
Trans-vaginalPPPPNote 1Note 2, 7, 8, 9, 12
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)PPPPNote 1Note 2, 7, 8, 9, 12
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vessel
VesselOther (spec.)

N= new indication; P= previously cleared by FDA K153529; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

{22}------------------------------------------------

510(k) No.:

Device Name: LA2-9A for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)PPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Musculo-skel. (Superfic.)PPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Other (spec.)

N= new indication; P= previously cleared by FDA K153529; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+TD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPF+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

  • Note 3: Includes infertility monitoring of follicle development
  • Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
  • Note 6: Abdominal organs and peripheral vessel
  • Note 7: Tissue Harmonic Imaging (THI)
  • Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 4: Color M-mode

{23}------------------------------------------------

510(k) No.:

Device Name: LA3-16AI for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)
Intra-operative (See Note 6)PPPPNote 1Note 7, 9, 11
Intra-operative (Neuro.)PPPPNote 1Note 7, 9, 11
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 7, 9, 11
Musculo-skel. (Superfic.)PPPPNote 1Note 7, 9, 11
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Note 7, 9, 11
Other (spec.)

N= new indication; P= previously cleared by FDA K153529; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+TD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPF+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

{24}------------------------------------------------

510(k) No.:

Device Name: LA4-18B for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal/Obstetrics (See Note 3)
Abdominal (See Note 10)
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
Pediatric
Small Organ (See Note 5)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Musculo-skel. (Superfic.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Intra-luminal
Other (See Note 13)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Other (spec.)

N= new indication; P= previously cleared by FDA K153529; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+TD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPF+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

  • Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
  • Note 6: Abdominal organs and peripheral vessel
  • Note 7: Tissue Harmonic Imaging (THI)
  • Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging
    Note 12: ElastoScan

Note 4: Color M-mode

{25}------------------------------------------------

510(k) No.:

Device Name: PA3-8B for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPPNote 1Note 7
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPPNote 1Note 7
Small Organ (See Note 5)
Neonatal CephalicPPPPPNote 1Note 7
Adult CephalicPPPPPNote 1Note 7
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac AdultPPPPPNote 1Note 4, 7
CardiacCardiac PediatricPPPPPNote 1Note 4, 7
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K153529; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+TD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPF+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

{26}------------------------------------------------

510(k) No.:

Device Name: PM1-6A for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPPNote 1Note 7
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult CephalicPPPPPNote 1Note 7
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac AdultPPPPPNote 1Note 4, 7
CardiacCardiac PediatricPPPPPNote 1Note 4, 7
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K153529; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+TD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPF+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

  • Note 3: Includes infertility monitoring of follicle development
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging
    Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 4: Color M-mode

{27}------------------------------------------------

510(k) No.:

Device Name: EV3-10B for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Note 2, 7, 8, 9, 11, 12
Abdominal (See Note 10)PPPPPNote 1Note 2, 7, 8, 9, 11, 12
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal ImagingLaparoscopic
&OtherPediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPNote 1Note 2, 7, 8, 9, 11, 12
Trans-vaginalPPPPPNote 1Note 2, 7, 8, 9, 11, 12
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)PPPPPNote 1Note 2, 7, 8, 9, 11, 12
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vessel
VesselOther (spec.)

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPF+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

  • Note 3: Includes infertility monitoring of follicle development
  • Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
  • Note 6: Abdominal organs and peripheral vessel
  • Note 7: Tissue Harmonic Imaging (THI)
  • Note 8: 3D imaging
  • Note 9: MultiVision (Spatial Compound Imaging)
  • Note 10: Includes Renal, Gynecology/Pelvis
  • Note 11: Panoramic imaging
  • Note 12: ElastoScan
  • Note 13: Includes Urology/Prostate

Note 4: Color M-mode

{28}------------------------------------------------

510(k) No.:

Device Name: LM4-15B for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 12
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Musculo-skel. (Superfic.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Other (spec.)

N= new indication; P= previously cleared by FDA K162094; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPF+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

  • Note 2: Includes imaging for guidance of biopsy
  • Note 3: Includes infertility monitoring of follicle development
  • Note 4: Color M-mode
  • Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
  • Note 6: Abdominal organs and peripheral vessel
  • Note 7: Tissue Harmonic Imaging (THI)
  • Note 8: 3D imaging
  • Note 9: MultiVision (Spatial Compound Imaging)
  • Note 10: Includes Renal, Gynecology/Pelvis
  • Note 11: Panoramic imaging
  • Note 12: ElastoScan
  • Note 13: Includes Urology/Prostate

{29}------------------------------------------------

510(k) No.:

Device Name: PA4-12B for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
(Track I only)(Tracks I & III)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPNote 1
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote 1
Small Organ (See Note 5)
Neonatal CephalicPPPPNote 1
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac PediatricPPPPPNote 1Note 4, 7
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K162094; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPF+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

  • Note 3: Includes infertility monitoring of follicle development
    Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

Note 4: Color M-mode

{30}------------------------------------------------

510(k) Premarket Notification - Traditional

Image /page/30/Picture/2 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white, set against a blue, oval-shaped background. The oval is tilted slightly, giving the logo a dynamic appearance.

510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92.

1. Submitter's Information: 21 CFR 807.92(a)(1)

SAMSUNG MEDISON CO., LTD. 42, Teheran-ro 108-gil, Gangnam-gu, Seoul, Korea

Contact Person:

Emily Woo Regulatory Affairs Manager

Telephone:82.2.2194.7120
Facsimile:82.2.556.3974

Data Prepared: July 12, 2017

2. Device Name and Classification:

Common/Usual Name:Diagnostic Ultrasound System and Accessories
Proprietary Name:WS80A Diagnostic Ultrasound System
Device Class:Class II
Classification Names/FR Number/Product Code:Ultrasonic Pulsed Doppler Imaging SystemUltrasound Pulsed Echo Imaging SystemDiagnostic Ultrasound Transducer892.1550892.1560892.1570IYNIYOITX

3. Identification of the predicate or legally marketed device:

  • WS80A Diagnostic Ultrasound System (K153529) -
  • -RS80A Diagnostic Ultrasound System (K162094)
  • -Voluson E8 Diagnostic Ultrasound System (K142472)

{31}------------------------------------------------

510(k) Premarket Notification - Traditional

Image /page/31/Picture/2 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white, set against a blue oval background. The oval is tilted slightly, giving the logo a dynamic appearance.

4. Device Description:

The WS80A is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as B mode, M mode, Color Doppler imaging, Power Doppler imaging (including Directional Power Doppler mode; S-Flow), PW Spectral Doppler mode, CW Spectral Doppler mode, Harmonic imaging, Tissue Doppler imaging, Tissue Doppler Wave, 3D imaging mode (real time 4D imaging mode), Elastoscan Mode or as a combination of these modes. The WS80A also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The WS80A has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

5. Intended Use:

The WS80A Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.

The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.

6. Substantial Equivalence:

The WS80A is substantially equivalent with respect to safety, effectiveness, and functionality to the WS80A Diagnostic Ultrasound System (K153529), RS80A Diagnostic Ultrasound System (K162094), and Voluson E8 Diagnostic Ultrasound System (K142472).

It is substantially equivalent with respect to safety, effectiveness, and functionality to the [LR21] of Voluson E8 (K142472) in regards to the device with [IOTA-ADNEX]. It is substantially equivalent with respect to safety, effectiveness, and functionality to the [Sono Biometry2] of Voluson E8 (K142472) in regards to the device with [Biometry Assist]. It is substantially equivalent with respect to safety, effectiveness, and functionality to the [Crystal Vue2] of WS80A (K1535529) in regards to the device with [CrystalVue Flow].

All systems transmit ultrasonic energy into patients, then perform post processing of received echoes to generate on-screen display of anatomic structures and fluid flow within the body. All system allow for specialized measurements of structures and flow, and calculations. These are described in detail in the technological characteristics comparison table as below.

{32}------------------------------------------------

510(k) Premarket Notification – Traditional

Image /page/32/Picture/2 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white, set against a blue oval background. The font is sans-serif and bold, and the oval is tilted slightly.

WS80AV4.00Comment
Clinicalapplication• No addition
Operation ofmodes• No addition
Appliedtransducers• Addition of 3 transducers: EV3-10B, LM4-15B and PA4-12BTransducers The previously cleared transducers EV3-10B None (but upgraded from V5-9 in K153529) LM4-15B K162094 PA4-12B K162094
SW Features• Adding the new S/W features: IOTA-ADNEX, BiometryAssist.• Improved S/W feature which waspreviously cleared in WS80A (K153529): CrystalVue Flow (from Crystal Vue)• Improved the image quality of the S/Wwhich was cleared previously (K153529).: Needle MateDescription of S/W Features• IOTA-ADNEX: This feature provides statisticalinformation according to the ADNEX Risk Model forovary ultrasound exams.• BiometryAssist: This function helps the user measurefetal growth indices.• CrystalVue Flow: This feature is to display renderedimages in Crystal Vue(K153529) with Color Doppler.
HW Features• No addition

{33}------------------------------------------------

Image /page/33/Picture/1 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The word is set against a blue, oval-shaped background.

510(k) Premarket Notification – Traditional

Feature / CharacteristicsThe subject deviceWS80AWS80A(K153529)RS80A(K162094)Voluson E8(K142472)
Indication for Use
- Fetal/Obstetrics
- Abdominal
- Gynecology
- Pediatric
- Small Organ
- Neonatal Cephalic
- Adult Cephalic
- Trans-rectal
- Trans-vaginal
- Musculo-skeletal(Conventional)
- Musculo-skeletal(Superficial)
- Urology
- Cardiac Adult
- Cardiac Pediatric
- Peripheral vessel
- Intra-operative
Scanhead Types
- Linear Array
- Curved Linear Array
- Endocavity
- Phased Array
- Static Probes
Scanhead Frequency
1.0 ~ 20.0 MHz
Modes of Operation
- B-mode
- M-mode
- Pulsed wave (PW) Doppler
- Continuous wave (CW) Doppler
- Color Doppler
- Power Amplitude Doppler
- Tissue Harmonic Imaging
- 3D/4D imaging mode
- Combined modes
Safety & EMC Compliance
- IEC 60601-1
- UL 60601-1
Feature / CharacteristicsThe subject deviceWS80APredicate devices
WS80A(K153529)RS80A(K162094)Voluson E8(K142472)
- CSA C22.2 No.601.1
- IEC 60601-2-37
- IEC 60601-1-2
Acoustic Output
Display Standard
Track 3
Patient ContactMaterials
Tested to ISO 10993-1
Functionality
- Quick Scan (Q Scan)
- ClearVision(SMDR/SMDR evo)
- MultiVision(Spatial CompoundImaging)
- Panoramic
- Needle Mate
- Elastoscan
- E-Thyroid
- E-Breast
- 3D Imaging(Volume DataAcquisition)
- 3D Imagingpresentation
3D Cine/4D Cine
5D Cine
- 3D RenderingMPR(Multi Planer Render)
- 3D XIMSV(Multi Slice View)Oblique View
- 3D MXIVolume Slice,Mirror View
- XI Volume CT(Volume CT)
- 3D MagiCut
- Volume Calculation(VOCAL, XIVOCAL)
- XI STIC
- HDVI
- 5D LB
- 5D CNS / 5D CNS+
The subject devicePredicate devices
Feature / CharacteristicsWS80AWS80ARS80AVoluson E8
(K153529)(K162094)(K142472)
- 5D Follicle
- 5D Heart / 5D Heart
Color
- 5D Limb Vol
- Realistic Vue (FRV)
- Crystal Vue /√(3)
CrystalVue Flow
- Volume NT/IT
- 5D NT
- 2D NT
- Auto IMT+
(Auto IMT)
- MPI
- ADVR
- IOTA-ADNEX√(1)
- BiometryAssist√(2)

{34}------------------------------------------------

Image /page/34/Picture/1 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The word is placed inside of a blue, oval-shaped background.

510(k) Premarket Notification – Traditional

{35}------------------------------------------------

Image /page/35/Picture/1 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The word is placed inside a blue, oval-shaped background.

510(k) Premarket Notification – Traditional

  1. LR2

  2. Sono Biometry

  3. Crystal Vue

{36}------------------------------------------------

510(k) Premarket Notification - Traditional

Image /page/36/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The word is placed inside of a blue, oval shape.

7. A brief discussion of the bench and non-clinical tests conducted on the subject device

The device has been evaluated for acoustic output, biocompatibility effectiveness as well as thermal, electrical, electromagnetic and mechanical safety and has been found to conform to applicable medical device safety standards.

The WS80A and its application comply with voluntary standards as below:
Reference No.Title
IEC 60601-1AAMI ANSI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 andA2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: Generalrequirements for basic safety and essential performance (IEC 60601-1:2005, MOD)
IEC 60601-1-2AAMI / ANSI / IEC 60601-1-2:2007(R)2012, Medical electrical equipment - Part 1-2:General requirements for basic safety and essential performance – Collateral standard:Electromagnetic compatibility - Requirements and tests (Edition 3)
IEC 60601-2-37IEC 60601-2-37:2007 Edition 2.0 2007-08, Medical electrical equipment - Part 2-37:Particular requirements for the basic safety and essential performance of ultrasonicmedical diagnostic and monitoring equipment
ISO10993-1AAMI / ANSI / ISO 10993-1:2009/(R)2013, Biological evaluation of medical devices– Part 1: Evaluation and testing within a risk management process
ISO14971ISO 14971:2007, Medical devices - Application of risk management to medical devices
NEMA UD 2-2004NEMA UD 2-2004 (R2009)Acoustic Output Measurement Standard for Diagnostic Ultrasound EquipmentRevision 3
NEMA UD 3-2004NEMA UD 3-2004 (R2009)Standard for Real-Time Display of Thermal and Mechanical Acoustic Output Indiceson Diagnostic Ultrasound Equipment, Revision 2

The WS80A and its application comply with voluntary standards as below:

Summary of the performance testing data:

Performance tests were conducted to test the functionality of the "IOTA-ADNEX" and "BiometryAssist". Image performance assessments were conducted to the "CrystalVue Flow" and Needle Mate". These tests have been performed to assess the functionality of the subject device. Results of all conducted testing and assessments were found acceptable and did not raise any new issues of safety or effective.

Summary of Clinical Tests:

Not applicable. The subject of this submission, WS80A, did not require clinical studies to support substantial equivalence.

8. Conclusion

Intended uses and other key features are consistent with traditional clinical practices and FDA guidelines. The design, development and quality process of the manufacturer confirms with 21 CFR 820 and ISO 13485. The device is designed to conform to applicable medical device safety standards and compliance. Therefore, SAMSUNG MEDISON CO., LTD. considers the WS80A to be as safe, as effective, and performance is substantially equivalent to the predicate devices.

END of 510(K) Summary