Intended Use

The WS80A Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.

The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Small Organ, Neonatal Cephalic. Adult Cephalic. Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.

Device Description

The WS80A is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as B mode, M mode, Color Doppler imaging, Power Doppler imaging (including Directional Power Doppler mode; S-Flow), PW Spectral Doppler mode, CW Spectral Doppler mode, Harmonic imaging, Tissue Doppler imaging, Tissue Doppler Wave, 3D imaging mode (real time 4D imaging mode), Elastoscan Mode or as a combination of these modes. The WS80A also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The WS80A has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

AI/ML Overview

I'm sorry but there is no acceptance criteria or study that describes how the device meets the acceptance criteria from the context you provided.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is a stylized image of three overlapping human profiles, which are connected to three wavy lines below.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 2, 2016

Samsung Medison Co., Ltd. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street, NW BUFFALO MN 55313

Re: K153529

Trade/Device Name: WS80A Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: January 18, 2016 Received: January 19, 2016

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert Ocks

Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K153529

Device Name

WS80A Diagnostic Ultrasound System

Indications for Use (Describe)

The WS80A Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.

The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Small Organ, Neonatal Cephalic. Adult Cephalic. Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) No.:

Device Name: WS80A Diagnostic Ultrasound System Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal/Obstetrics (See Note 3)PPPPPNote 1Notes 2, 3,4, 5, 6, 7, 8, 9, 11
Abdominal (See Note 10)PPPPPNote 1Notes 2, 4, 5, 6, 7, 8, 9, 10, 11, 12
Intra-operative (See Note 6)NNNNNNote 1Note 7, 9, 11
Intra-operative (Neuro.)NNNNNNote 1Note 7, 9, 11
Laparoscopic
& OtherPediatricPPPPPNote 1Notes 2, 4, 6, 7, 8, 9, 11
Small Organ (See Note 5)PPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 12
Neonatal CephalicPPPPPNote 1Notes 7
Adult CephalicPPPPPNote 1Notes 7
Trans-rectalPPPPPNote 1Notes 2, 7, 8, 9, 11, 12
Trans-vaginalPPPPPNote 1Notes 2, 7, 8, 9, 11, 12
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Musculo-skel. (Superfic.)PPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Intra-luminal
Other (spec.) (See Note 13)PPPPPNote 1Notes 2, 7,8, 9, 11, 12
CardiacCardiac AdultPPPPPNote 1Notes 4, 7
Cardiac PediatricPPPPPNote 1Notes 4, 7
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Other (spec.)

N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PD+PW, B+PPI+PW, B+ID+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Concurrence of Center for Devices and Radiological Health (CDRH) Prescription Use (Per 21 CFR 801.109)

{4}------------------------------------------------

510(k) No.:

Device Name: SC1-6 for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal/Obstetrics (See Note 3)PPPPPNote 1Notes 2, 4, 7, 8, 9, 11
Abdominal (See Note 10)PPPPPNote 1Notes 2, 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.) (See Note 13)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PPF+W, B+PPI+PW, B+ID+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

  • Note 3: Includes infertility monitoring of follicle development
  • Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
  • Note 6: Abdominal organs and peripheral vessel
  • Note 7: Tissue Harmonic Imaging (THI)
  • Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 4: Color M-mode

{5}------------------------------------------------

510(k) No.:

Device Name: C2-6 for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Notes 2, 4, 7, 8, 9, 11
Abdominal (See Note 10)PPPPPNote 1Notes 2, 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.) (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PD+PW, B+PPI+PW, B+ID+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

  • Note 3: Includes infertility monitoring of follicle development
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 4: Color M-mode

Note 8: 3D imaging

{6}------------------------------------------------

510(k) No.:

Device Name: E3-12A for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Notes 2, 8, 7, 9
Abdominal (See Note 10)PPPPPNote 1Notes 2, 7, 8, 9, 12
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPNote 1Notes 2, 7, 8, 9, 12
Trans-vaginalPPPPPNote 1Notes 2, 7, 8, 9, 12
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.) (See Note 13)PPPPPNote 1Notes 2, 7, 8, 9, 12
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PPF+W, B+PPH+W, B+ID+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

  • Note 3: Includes infertility monitoring of follicle development
  • Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
  • Note 6: Abdominal organs and peripheral vessel
  • Note 7: Tissue Harmonic Imaging (THI)
  • Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 4: Color M-mode

{7}------------------------------------------------

510(k) No.:

Device Name: VR5-9 for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Note 2, 7, 8, 9
Abdominal (See Note 10)PPPPPNote 1Note 2, 7, 8, 9, 12
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPNote 1Note 2, 7, 8, 9, 12
Trans-vaginalPPPPPNote 1Note 2, 7, 8, 9, 12
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.) (See Note 13)PPPPPNote 1Note 2, 7, 8, 9, 12
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PPF+W, B+PPH+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

{8}------------------------------------------------

510(k) No.:

Device Name: L3-12A for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Note 2, 5, 6, 7, 9, 11
Abdominal (See Note 10)PPPPNote 1Notes 2, 4, 6, 7, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote 1Notes 2, 4, 6, 7, 9, 11
Small Organ (See Note 5)PPPPNote 1Note 2, 5, 6, 7, 9, 11, 12
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2, 5, 6, 7, 9, 11
Musculo-skel. (Superfic.)PPPPNote 1Note 2, 5, 6, 7, 9, 11
Intra-luminal
Other (spec.) (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Note 2, 5, 6, 7, 9, 11
Other (spec.)

N= new indication; P= previously cleared by FDA K152962; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PPF+W, B+PPH+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

{9}------------------------------------------------

510(k) No.:

Device Name: L5-13 for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)PPPPNote 1Note 2, 5, 6, 7, 9, 11, 12
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2, 5, 6, 7, 9, 11
Musculo-skel. (Superfic.)PPPPNote 1Note 2, 5, 6, 7, 9, 11
Intra-luminal
Other (spec.) (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Note 2, 5, 6, 7, 9, 11
Other (spec.)

N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PPF+PW, B+PPI+PW, B+ID+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

{10}------------------------------------------------

510(k) No.:

Device Name: V4-8 for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal/Obstetrics (See Note 3)PPPPPNote 1Note 2, 4, 7, 8, 9, 11
Abdominal (See Note 10)PPPPPNote 1Note 2, 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.) (See Note 13)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PPF+W, B+PPH+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

  • Note 3: Includes infertility monitoring of follicle development
  • Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
  • Note 6: Abdominal organs and peripheral vessel
  • Note 7: Tissue Harmonic Imaging (THI)
  • Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 4: Color M-mode

{11}------------------------------------------------

510(k) No.:

Device Name: V5-9 for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Note 2, 7, 8, 9, 11
Abdominal (See Note 10)PPPPPNote 1Note 2, 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPNote 1Note 2, 7, 8, 9, 11
Trans-vaginalPPPPPNote 1Note 2, 7, 8, 9, 11
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.) (See Note 13)PPPPPNote 1Notes 2, 7, 8, 9, 11
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PPF+PW, B+PPI+PW, B+ID+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

{12}------------------------------------------------

510(k) No.:

Device Name: CA1-7A for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
General(Track I only)Specific(Tracks I & III)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPNote 1Notes 2, 4, 7, 8, 9, 11
Abdominal (See Note 10)PPPPNote 1Notes 2, 7, 8, 9, 11, 13
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote 1Notes 2, 7, 8, 9, 11
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Notes 2, 7, 8, 9, 11
Musculo-skel. (Superfic.)PPPPNote 1Notes 2, 7, 8, 9, 11
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K152962; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PPF+W, B+PPH+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

  • Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
    Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 4: Color M-mode

{13}------------------------------------------------

510(k) No.:

Device Name: CA2-8A for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Notes 2, 4, 7, 8, 9, 11
Abdominal (See Note 10)PPPPPNote 1Notes 2, 4, 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PPF+PW, B+PPI+PW, B+ID+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

  • Note 3: Includes infertility monitoring of follicle development
  • Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
  • Note 6: Abdominal organs and peripheral vessel
  • Note 7: Tissue Harmonic Imaging (THI)
  • Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging
    Note 12: ElastoScan

Note 4: Color M-mode

Note 13: Includes Urology/Prostate

{14}------------------------------------------------

510(k) No.:

Device Name: CF4-9 for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPNote 1Notes 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
PediatricPPPPNote 1Notes 8, 9, 11
Small Organ (See Note 5)
Neonatal CephalicPPPPNote 1Notes 8, 9, 11
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Notes 8, 9, 11
Other (spec.)

N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PD+PW, B+PPI+PW, B+ID+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

  • Note 7: Tissue Harmonic Imaging (THI)
    Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 8: 3D imaging

{15}------------------------------------------------

510(k) No.:

Device Name: LA3-16A for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
(Track I only)(Tracks I & III)
OphthalmicOphthalmic
Fetal (See Note 3)
Abdominal (See Note 10)
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)PPPPNote 1Note 2, 5, 6, 7, 9, 11, 12
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2, 5, 6, 7, 9, 11
Musculo-skel. (Superfic.)PPPPNote 1Note 2, 5, 6, 7, 9, 11
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Note 2, 5, 6, 7, 9, 11
Other (spec.)

N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PD+PW, B+PPI+PW, B+ID+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

{16}------------------------------------------------

510(k) No.:

Device Name: PE2-4 for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPPNote 1Note 7
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal CephalicPPPPPNote 1Note 7
Adult CephalicPPPPPNote 1Note 7
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac AdultPPPPPNote 1Note 4, 7
CardiacCardiac PediatricPPPPPNote 1Note 4, 7
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PD+PW, B+PPI+PW, B+ID+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

{17}------------------------------------------------

510(k) No.:

Device Name: CV1-8A for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPNote 1Note 2, 4, 7, 8, 9, 11
Abdominal (See Note 10)PPPPNote 1Note 2, 7, 8, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.) (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PD+PW, B+PPI+PW, B+ID+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

  • Note 3: Includes infertility monitoring of follicle development
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging
    Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Note 4: Color M-mode

{18}------------------------------------------------

510(k) No.:

Device Name: LV3-14A for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)PPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Musculo-skel. (Superfic.)PPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Other (spec.)

N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PD+PW, B+PPI+PW, B+ID+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

  • Note 3: Includes infertility monitoring of follicle development
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 4: Color M-mode

Note 8: 3D imaging

Concurrence of Center for Devices and Radiological Health (CDRH) Prescription Use (Per 21 CFR 801.109)

{19}------------------------------------------------

510(k) No.:

Device Name: CA2-9A for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)NNNNNNote 1Notes 2, 4, 7, 8, 9, 11
Abdominal (See Note 10)NNNNNNote 1Notes 2, 4, 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PPF+W, B+PPH+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

{20}------------------------------------------------

510(k) No.:

Device Name: CA3-10A for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal/Obstetrics (See Note 3)NNNNNNote 1Notes 2, 3, 7, 8, 9, 11
Abdominal (See Note 10)NNNNNNote 1Notes 2, 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
PediatricNNNNNNote 1Notes 2, 7, 8, 9, 11
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselNNNNNNote 1Notes 2, 7, 8, 9, 11
Other (spec.)

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PD+PW, B+PPI+PW, B+ID+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 7: Tissue Harmonic Imaging (THI)
  • Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 4: Color M-mode

Note 6: Abdominal organs and peripheral vessel

{21}------------------------------------------------

510(k) No.:

Device Name: EA2-11B for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Note 2, 7, 8, 9
Abdominal (See Note 10)PPPPPNote 1Note 2, 7, 8, 9, 12
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPNote 1Note 2, 7, 8, 9, 12
Trans-vaginalPPPPPNote 1Note 2, 7, 8, 9, 12
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)PPPPPNote 1Note 2, 7, 8, 9, 12
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vessel
VesselOther (spec.)

N= new indication; P= previously cleared by FDA K150478; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PP1+PW, B+PPI+PW, B+ID+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

{22}------------------------------------------------

510(k) No.:

Device Name: LA2-9A for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
& OtherPediatric
Small Organ (See Note 5)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Musculo-skel. (Superfic.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Other (spec.)

N= new indication; P= previously cleared by FDA K151663; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PD+PW, B+PPI+PW, B+ID+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

{23}------------------------------------------------

510(k) No.:

Device Name: LA3-16AI for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
(Track I only)(Tracks I & III)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)
Intra-operative (See Note 6)PPPPNote 1Note 7, 9, 11
Intra-operative (Neuro.)PPPPNote 1Note 7, 9, 11
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 7, 9, 11
Musculo-skel. (Superfic.)PPPPNote 1Note 7, 9, 11
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Note 7, 9, 11
Other (spec.)

N= new indication; P= previously cleared by FDA K152962; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PD+PW, B+PPI+PW, B+ID+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

  • Note 3: Includes infertility monitoring of follicle development
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)
  • Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 4: Color M-mode

{24}------------------------------------------------

510(k) No.:

Device Name: LA4-18B for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
(Track I only)(Tracks I & III)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal ImagingLaparoscopic
&OtherPediatric
Small Organ (See Note 5)PPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Musculo-skel. (Superfic.)PPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vesselPPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
VesselOther (spec.)

N= new indication; P= previously cleared by FDA K151663; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PD+PW, B+PPI+PW, B+ID+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

  • Note 3: Includes infertility monitoring of follicle development
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)
    Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 4: Color M-mode

Note 8: 3D imaging

{25}------------------------------------------------

510(k) No.:

Device Name: PA3-8B for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPPNote 1Note 7
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
PediatricPPPPPNote 1Note 7
Small Organ (See Note 5)
Neonatal CephalicPPPPPNote 1Note 7
Adult CephalicPPPPPNote 1Note 7
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
CardiacCardiac AdultPPPPPNote 1Note 4, 7
Cardiac PediatricPPPPPNote 1Note 4, 7
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K152962; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CW, B+C+PW, B+PD+PW, B+PD+PW, B+PPI+PW, B+ID+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

  • Note 2: Includes imaging for guidance of biopsy

  • Note 3: Includes infertility monitoring of follicle development

  • Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging
    Note 12: ElastoScan

Note 4: Color M-mode

{26}------------------------------------------------

510(k) No.:

Device Name: PM1-6A for use with WS80A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPPNote 1Note 7
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult CephalicPPPPPNote 1Note 7
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac AdultPPPPPNote 1Note 4, 7
CardiacCardiac PediatricPPPPPNote 1Note 4, 7
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K151663; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+DPI, B+CW, B+C+PW, B+PD+PW, B+PD+PW, B+PPI+PW, B+TD+PW, B+C+M, Dual/Quad, , B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

  • Note 3: Includes infertility monitoring of follicle development
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 7: Tissue Harmonic Imaging (THI)
    Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 4: Color M-mode

Note 6: Abdominal organs and peripheral vessel

Note 8: 3D imaging

{27}------------------------------------------------

510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92.

1. Submitter's Information: 21 CFR 807.92(a)(1)

SAMSUNG MEDISON CO., LTD. 42, Teheran-ro 108-gil, Gangnam-gu, Seoul, Korea

Contact Person:

Kyeong-Mi, Park Regulatory Affairs Manager

Telephone:82.2.2194.1373
Facsimile:82.2.556.3974

Data Prepared: November 5, 2015

2. Name of the device:

Common/Usual Name:
Diagnostic Ultrasound System and Accessories
Proprietary Name:
WS80A Diagnostic Ultrasound System
Classification Names:FR NumberProduct Code
Ultrasonic Pulsed Doppler Imaging System892.1550IYN
Ultrasound Pulsed Echo Imaging System892.1560IYO
Diagnostic Ultrasound Transducer892.1570ITX

3. Identification of the predicate or legally marketed device:

  • WS80A Diagnostic Ultrasound System (K143089) -
  • -HS70A Diagnostic Ultrasound System (K152962)
  • -RS80A Diagnostic Ultrasound System (K151663)
  • A35 Diagnostic Ultrasound System (K150478) -
  • -Voluson E8 Diagnostic Ultrasound System (K122327)
  • -LOGIQ E9 Diagnostic Ultrasound System (K142160)

{28}------------------------------------------------

4. Device Description:

The WS80A is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as B mode, M mode, Color Doppler imaging, Power Doppler imaging (including Directional Power Doppler mode; S-Flow), PW Spectral Doppler mode, CW Spectral Doppler mode, Harmonic imaging, Tissue Doppler imaging, Tissue Doppler Wave, 3D imaging mode (real time 4D imaging mode), Elastoscan Mode or as a combination of these modes. The WS80A also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The WS80A has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

5. Intended Uses:

The WS80A Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.

The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.

6. Technological Characteristics:

The WS80A is substantially equivalent with respect to safety, effectiveness, and functionality to the WS80A Diagnostic Ultrasound System (K143089), HS70A Diagnostic Ultrasound System (K152962), RS80A Diagnostic Ultrasound System (K151663), A35 Diagnostic Ultrasound System (K150478), Voluson E8 Diagnostic Ultrasound System (K122327) and LOGIQ E9 Diagnostic Ultrasound System (K142160).

It is substantially equivalent with respect to safety, effectiveness, and functionality to the [5D LB]] of WS80A (K143089) and [Fractional Limb2] of Voluson E8 (K113758) in regards to the device with [5D Limb Vol].

It is substantially equivalent with respect to safety, effectiveness, and functionality to the [Realistic Vue] and [Natural Vue3] and of RS80A (K151663) in regards to the device with [Crystal Vue]. It is substantially equivalent with respect to safety, effectiveness, and functionality to the [5D CNS] of

WS80A (K143089) in regards to the device with [5D CNS+]. It is substantially equivalent with respect to safety, effectiveness, and functionality to the [5D Heart] of

WS80A (K143089) in regards to the device with [5D Heart Color].

All systems transmit ultrasonic energy into patients, then perform post processing of received echoes to generate on-screen display of anatomic structures and fluid flow within the body. All system allow for specialized measurements of structures and flow, and calculations.

These are described in detail in the technological characteristics comparison table as below.

{29}------------------------------------------------

WS80AV3.00Comment
Clinicalapplication• Addition of clinical applications: Intra-operative
Operation ofmodes• No addition
Appliedtransducers• Addition of 8 transducers: CA2-9A, CA3-10A, EA2-11B,LA2-9A, LA3-16AI, LA4-18B,PA3-8B and PM1-6A.Transducers The previouslycleared transducers CA2-9A None CA3-10A None EA2-11B K150478 LA2-9A K151663 LA3-16AI K152962 LA4-18B K151663 PA3-8B K151663 PM1-6A K152962
SW Features• Adding the new S/W features: 5D Limb Vol and Crystal Vue.• Adding the predicate S/W features: S-Detect, E-Thyroid, E-Breast andNeedle Mate.• Improving the S/W feature that waspreviously cleared in WS80A(K143089): 5D CNS+, 5D Heart Color, DICOM,EZ Exam+, Setup&Preference.• Changing the S/W feature's name- EZ Exam → EZ Exam+Software improvements migrated frompredicate device as below.<Description of S/W Features>• 5D Limb Vol: This feature allows you to detectfractional limb contour and to measure semi-automatically the volume of fractional limb.• Crystal Vue: This feature is the volumerendering technology that visualizing interiorand exterior structures of 3D volume data.• 5D CNS+: This feature is an expanded version of5D CNS, providing the additional six referenceviews for fetal brain diagnostics.• 5D Heart (including 5D Heart Color): Thisfeature assists with diagnosis of fetal heartconditions by using the STIC data of the fetus.5D Heart Color is an expanded version of 5DHeart, providing the color data for the diagnosisof fetal heart blood flow.• S-Detect: This module analyzes the selectedlesion from the ultrasound image and shows theanalyzed data.• E-Thyroid: This function assists with thediagnosis of thyroid lesions. ROI is used to showthe contrast of an elastogram.• E-Breast: This is the function for breast lesionexam. Calculates the strain between the area ofthe suspected lesion (ROI) and normal breast fatand displays the results.• Needle Mate: It provides the Needle Matefunction to display the Needle position moreclearly in B Image by emphasizing the Needleimage
SW Features The previouslycleared SWFeatures S-Detect K151663 E-Thyroid K151663 E-Breast K151663 Needle Mate K151663
HW Features• No addition

{30}------------------------------------------------

Feature / CharacteristicsThe subject deviceWS80AWS80A(K143089)HS70A(K152962)RS80A(K151663)A35(K150478)Voluson E8(K122327)LOGIQ E9(K142160)
Indication for Use
- Fetal/Obstetrics
- Abdominal
- Gynecology
- Pediatric
- Small Organ
- Neonatal Cephalic
- Adult Cephalic
- Trans-rectal
- Trans-vaginal
- Musculo-skeletal(Conventional)
- Musculo-skeletal(Superficial)
- Urology
- Cardiac Adult
- Cardiac Pediatric
- Peripheral vessel
- Intra-operative
Scanhead Types
- Linear Array
- Curved Linear Array
- Endocavity
- Phased Array
- Static Probes
Scanhead Frequency
1.0 ~ 20.0 MHz
Modes of Operation
- B-mode
- M-mode
- Pulsed wave (PW) Doppler
- Continuous wave (CW) Doppler
- Color Doppler
- Power Amplitude Doppler
- Tissue Harmonic Imaging
- 3D/4D imaging mode
- Combined modes
Safety & EMC
Compliance
- IEC 60601-1
- UL 60601-1
Feature / CharacteristicsThe subject deviceWS80AWS80A(K143089)HS70A(K152962)RS80A(K151663)A35(K150478)Voluson E8(K122327)LOGIQ E9(K142160)
- CSA C22.2 No.601.1
- IEC 60601-2-37
- IEC 60601-1-2
Acoustic Output
Display Standard
Track 3
Patient ContactMaterials
Tested to ISO 10993-1
Functionality
- Quick Scan (Q Scan)
- ClearVision(SMDR/SMDR evo)
- MultiVision(Spatial CompoundImaging)
- Panoramic
- Needle Mate
- Elastoscan
- E-Thyroid
- E-Breast
- 3D Imaging(Volume DataAcquisition)
- 3D Imagingpresentation
3D Cine/4D Cine
5D Cine
- 3D RenderingMPR(Multi Planer Render)
- 3D XIMSV(Multi Slice View)Oblique View
- 3D MXIVolume Slice,Mirror View
- XI Volume CT(Volume CT)
- 3D MagiCut
- Volume Calculation(VOCAL, XIVOCAL)
- XI STIC
- HDVI
- 5D LB
- 5D CNS / 5D CNS+
Feature / CharacteristicsThe subjectdeviceWS80AWS80A(K143089)HS70A(K152962)RS80A(K151663)A35(K150478)Voluson E8(K122327)LOGIQ E9(K142160)
- 5D Heart / 5D HeartColor
- 5D Limb Vol√ 1)√ 2)
- Realistic Vue (FRV)
- Crystal Vue√ 3)
- Volume NT/IT
- 5D NT
- 2D NT
- Auto IMT+(Auto IMT)
- MPI
- S-Detect
- ADVR

510(k) Summary / Statement Certification

{31}------------------------------------------------

510(k) Summary / Statement Certification

{32}------------------------------------------------

7. A brief discussion of the bench and non-clinical tests conducted on the subject device

The device has been evaluated for acoustic output, biocompatibility effectiveness as well as thermal, electrical, electromagnetic and mechanical safety and has been found to conform to applicable medical device safety standards.

The WS80A and its application comply with voluntary standards as below:

  • UL 60601-1, Safety requirements for Medical Equipment
  • CSA C22.2 No. 601.1, Safety requirements for Medical Equipment
  • IEC60601-2-37, Diagnostic Ultrasound Safety Standards
  • EN/IEC60601-1. Safety requirements for Medical Equipment
  • EN/IEC60601-1-2, EMC requirements for Medical Equipment
  • NEMA UD-2, Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment
  • NEMA UD-3, Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment
  • ISO10993-1, Biocompatibility
  • ISO14971, Application of risk management to medical devices

Summary of Clinical Tests:

Not applicable. The subject of this submission, WS80A, did not require clinical studies to support substantial equivalence.

8. Conclusion

Intended uses and other key features are consistent with traditional clinical practices and FDA guidelines. The design, development and quality process of the manufacturer confirms with 21 CFR 820 and ISO 13485. The device is designed to conform to applicable medical device safety standards and compliance. Therefore, SAMSUNG MEDISON CO., LTD. considers the WS80A to be as safe, as effective, and performance is substantially equivalent to the predicate devices.

END of 510(K) Summary

§ 892.1550 Ultrasonic pulsed doppler imaging system.

(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.