K Number
K171050

Validate with FDA (Live)

Date Cleared
2017-11-09

(216 days)

Product Code
Regulation Number
872.3930
Panel
Dental
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Geistlich Fibro-Gide® is intended for soft tissue augmentation. It is indicated for:

  • Localized gingival augmentation to increase keratinized tissue (KT) around teeth and implants A
  • Alveolar ridge reconstruction for prosthetic treatment A
  • Recession defects for root coverage A
Device Description

Geistlich Fibro-Gide® is a fully resorbable, porous, collagen matrix of porcine origin of a spongious consistency. The collagen is extracted from veterinary certified pigs and is carefully purified to avoid antigenic reactions. The collagen scaffold is weakly crosslinked. Geistlich Fibro-Gide® is sterilized in double packaging by Gamma-irradiation.

Geistlich Fibro-Gide® is an implantable device intended for use in soft tissue augmentation procedures. As described in more detail below, the device is indicated specifically for insufficient tissue volume at the alveolar ridge and for soft tissue recession. It has mechanical properties appropriate to withstand the mechanical stresses that occur after wound closure in soft tissue augmentation procedures, i.e., it has good volume stability and it withstands early resorption to allow the formation of new soft tissue and degrades over time. In addition, the matrix is designed with an appropriate thickness to provide sufficient space for the ingrowth of new soft tissue. Due to its good wettability, suturability and biological properties, the device becomes well integrated into the surrounding soft tissue.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device called Geistlich Fibro-Gide®. This document focuses on demonstrating substantial equivalence to a predicate device (Geistlich Mucograft®) rather than establishing novel safety and effectiveness criteria. Therefore, the information provided does not contain traditional "acceptance criteria" for a new device's absolute performance against a set standard, but rather criteria for demonstrating similarity to an already approved device.

Here's an analysis based on the categories you provided:

1. A table of acceptance criteria and the reported device performance:

Since this is a substantial equivalence submission, "acceptance criteria" are implicitly defined by the characteristics of the predicate device. The performance shown is a comparison to that predicate.

CharacteristicPredicate (Geistlich Mucograft®) Performance (Implicit Acceptance Criteria)Geistlich Fibro-Gide® Reported Performance
Intended UseSoft Tissue AugmentationSoft Tissue Augmentation
Indications for Use- Covering of implants- Localized gingival augmentation (KT) - Alveolar ridge reconstruction- Recession defects for root coverage- Localized gingival augmentation to increase keratinized tissue (KT) around teeth and implants - Alveolar ridge reconstruction for prosthetic treatment - Recession defects for root coverage
Dimensions15x20 mm, 20x30 mm (K102531), 30x40 mm (K073711, K012423) (thickness app. 2.5 - 5 mm)15x20 mm, 20x40mm (thickness app. 6mm)
FormSponge-like matrixSponge-like matrix
ColorAlmost whiteWhite to almost white
PorosityVery porous materials with average pore values of 90% or more. The average surface area and the bulk density are almost identical.Very porous materials with average pore values of 90% or more. The average surface area and the bulk density of Geistlich Fibro-Gide® is almost identical to that of Geistlich Mucograft.
Capillarity and wettabilitySpontaneously wettable and completely soaked in aqueous solution in less than one minuteSpontaneously wettable and completely soaked in aqueous solution in less than one minute
Cross-linkingNot chemically cross-linkedChemically cross-linked with EDC and NHS
Easy to trimCan be cut with surgical instrumentsCan be cut with surgical instruments
Can be suturedCan be fixed with sutures, as demonstrated by sutureability testingCan be fixed with sutures, as demonstrated by sutureability testing
Raw MaterialPorcine connective tissue, Porcine skin tissuePorcine connective tissue, Porcine skin tissue
CompositionPorcine collagen: Product is produced from the same intermediate collagenous productsPorcine collagen: Product is produced from the same intermediate collagenous products
Collagen / ElastinMajor protein components Collagen I and Collagen III (no Collagen II) and ElastinMajor protein components Collagen I and Collagen III (no Collagen II) and Elastin
Amino Acid CompositionVery similar with equal amounts of amino acidsVery similar with equal amounts of amino acids
FatTrace (less than 0.5%)Trace (less than 0.5%)
GlycosaminoglycansTrace (less than 0.5%)Trace (less than 0.5%)
Other proteinsNone (>0.5%)None
pH3-73-7
Source of raw materialIdentical porcine tissueIdentical porcine tissue
ManufactureMultistage validated, SOP controlled purification processMultistage validated, SOP controlled purification process
Manufacturing conditionsQuality systems regulation (CFR Part 820)Quality systems regulation (CFR Part 820)
PackagingISO 11607, Parts 1 and 2 compliant. Sterile double layer packaging in aluminum pouch.Conforms to ISO 11607, Parts 1 and 2. Sterile double layer packaging, including aluminum layer to protect against vapor penetration.
SterilizationGamma irradiation SAL 10-6; sterile, single useGamma irradiation SAL 10-6; Device provided sterile, for single use only
UserRestricted to licensed dentistsRestricted to licensed dentists
Biocompatibility(Implicitly passed the same ISO 10993 tests as the subject device)Passed all tests: Cytotoxicity, Irritation, Sensitization, Acute systemic toxicity, Pyrogenicity, Genotoxicity (Bacterial reverse mutagenicity, Chromosomal aberration, Micronucleus study), Local tissue response after implantation (4, 12, 26-week subcutaneous), Subchronic systemic toxicity, Chronic systemic toxicity, Leachables.
Animal Study Effectiveness(Implied safe and effective based on predicate's approval)Demonstrated tissue integration, continuous resorption, and comparable degradation rate to predicate. Acceptable safety profile for distal organs, hematologic parameters, and clinical chemistries. Demonstrated substantial equivalence in soft tissue augmentation and local tissue effects.
Clinical Data (Effectiveness/Safety)(Implied safe and effective based on predicate's approval)Demonstrated product safety and effectiveness in the product's indications for use.

Key finding regarding acceptance criteria: The acceptance criteria for Geistlich Fibro-Gide® are defined by its ability to demonstrate substantial equivalence to Geistlich Mucograft® across various characteristics, including physical properties, composition, manufacturing, and performance in biocompatibility, animal, and clinical studies. The "performance" is that it "passes" these comparative tests.


2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

  • Bench Testing: No specific sample sizes are provided for the bench tests. The data provenance is not explicitly stated as country of origin, but the sponsor and manufacturing/sterilization locations are in Switzerland. The testing is assumed to be prospective, undertaken for this submission.
  • Biocompatibility Testing: The specific sample sizes for each biocompatibility test (e.g., number of L929 fibroblast cultures, rabbits, guinea pigs, mice, rats) are not explicitly stated in this summary. The tests adhere to ISO 10993 standards, suggesting standard methods and sample sizes for these types of studies. The provenance of these animal studies is generally considered prospective, conducted for regulatory submission.
  • Animal Studies (Rat and Dog): "a rat and in a dog study" – exact number of animals is not provided. These were conducted "compliant with 21 CFR Part 58" (Good Laboratory Practice for Nonclinical Laboratory Studies), indicating prospective and controlled studies. The country is not specified, but given the sponsor and manufacturing locations, it is likely Europe (Switzerland) or a contracted lab.
  • Clinical Data: "The clinical data from a controlled, parallel and randomized study in patients presenting insufficient soft tissue volume." The sample size (number of patients) is not provided. This describes a prospective study design. The country of origin for the clinical data is not specified.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

This document does not describe the establishment of a "ground truth" by experts in the context of diagnostic accuracy, which is common for AI/imaging devices. Instead, the ground truth for this medical device is:

  • Bench testing: Established by laboratory measurements and adherence to scientific standards.
  • Biocompatibility: Established by adherence to ISO 10993 standards and generally accepted biological endpoints.
  • Animal studies: Established by histological examination, physiological measurements, and comparison to the predicate, likely reviewed by veterinary pathologists or relevant scientists.
  • Clinical study: The "ground truth" for safety and effectiveness is derived from the clinical outcomes of the study itself, assessed by the investigators and clinicians involved in the trial.

There is no mention of a panel of experts specifically establishing a "ground truth" for the test sets in the way an imaging device might rely on expert consensus.


4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

No adjudication method for disagreements among experts is described, as the studies primarily involve objective measurements (benchmarks, lab results) or clinical trial outcomes rather than subjective interpretations requiring adjudication.


5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This device is a collagen matrix for soft tissue augmentation, not an imaging or diagnostic AI device. Therefore, MRMC studies and "human readers improve with AI vs without AI assistance" are irrelevant to this submission.


6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not applicable. This device is a physical implantable medical device, not an algorithm or AI.


7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The "ground truth" for this submission consists of:

  • Bench testing results: Objective measurements of physical, chemical, and functional properties.
  • Histopathology and physiological data from animal studies: Pathological findings and biological responses in animal models.
  • Clinical outcomes data: Measurements and assessments of patient safety and effectiveness endpoints from a clinical trial.

8. The sample size for the training set:

Not applicable. This device is a physical product, not an AI/ML model that requires a training set.


9. How the ground truth for the training set was established:

Not applicable. (See #8)

{0}------------------------------------------------

November 9, 2017

Image /page/0/Picture/1 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

Geistlich Pharma AG % Janice M. Hogan Regulatory Counsel Hogan Lovells US LLP 1835 Market Street, 29th Floor Philadelphia, Pennsylvania 19103

Re: K171050

Trade/Device Name: Geistlich Fibro-Gide Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: Class II Product Code: NPL Dated: October 12, 2017 Received: October 12, 2017

Dear Janice Hogan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

Page 2 - Janice Hogan

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mary S. Runner -S

for

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K171050

Device Name

Indications for Use (Describe)

Geistlich Fibro-Gide® is intended for soft tissue augmentation. It is indicated for:

  • Localized gingival augmentation to increase keratinized tissue (KT) around teeth and implants A
  • Alveolar ridge reconstruction for prosthetic treatment A
  • Recession defects for root coverage A

Type of Use (Select one or both, as applicable)

Z Prescription Use (Part 21 CFR 801 Subpart D)

□ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

GEISTLICH FIBRO-GIDE®

SPONSOR

Geistlich Pharma AG Bahnhofstrasse 40 CH-6110 Wolhusen Switzerland

Contact Person:Marco Steiner
Company:Geistlich Pharma AG
Phone Number:Direct 011 41 41 492 67 64
Company 011 41 492 55 55
E-mail:marco.steiner@geistlich.ch
Date Prepared:November 9, 2017

DEVICE NAME

Proprietary Name:Geistlich Fibro-Gide®
Common/Usual Names:Porcine Collagen Matrix
Classification Name:Barrier, animal source, intraoral (NPL)

PREDICATE DEVICES

Geistlich Mucograft® (K102531) (Primary predicate device)

Geistlich Mucograft® and Geistlich Mucograft® Seal (K140518) (Reference device)

Geistlich Mucograft® (K073711, K012423) (Reference devices)

GENOSS Collagen Membrane (K102307) (Reference device)

DEVICE DESCRIPTION

Geistlich Fibro-Gide® is a fully resorbable, porous, collagen matrix of porcine origin of a spongious consistency. The collagen is extracted from veterinary certified pigs and is carefully purified to avoid antigenic reactions. The collagen scaffold is weakly crosslinked. Geistlich Fibro-Gide® is sterilized in double packaging by Gamma-irradiation.

Geistlich Fibro-Gide® is an implantable device intended for use in soft tissue augmentation procedures. As described in more detail below, the device is indicated

{4}------------------------------------------------

specifically for insufficient tissue volume at the alveolar ridge and for soft tissue recession. It has mechanical properties appropriate to withstand the mechanical stresses that occur after wound closure in soft tissue augmentation procedures, i.e., it has good volume stability and it withstands early resorption to allow the formation of new soft tissue and degrades over time. In addition, the matrix is designed with an appropriate thickness to provide sufficient space for the ingrowth of new soft tissue. Due to its good wettability, suturability and biological properties, the device becomes well integrated into the surrounding soft tissue.

INTENDED USE AND INDICATIONS FOR USE

Geistlich Fibro-Gide® is intended for soft tissue augmentation. It is indicated for:

  • Localized gingival augmentation to increase keratinized tissue (KT) around teeth and implants

  • Alveolar ridge reconstruction for prosthetic treatment A
  • Recession defects for root coverage.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

Geistlich Fibro-Gide® and its primary predicate device Geistlich Mucograft® are implantable, collagen matrices (scaffolds) used in dental soft tissue augmentation procedures. Both devices consist of collagen fibers in a loose, porous arrangement to enable cell invasion and settlement, providing an environment to allow for soft tissue growth by using the patient's own healing capacities. Both devices are fully resorbable and do not require a second intervention for removal. Like its reference device, Geistlich Fibro-Gide® is cross-linked. The device dimensions are similar and both devices require adaptation to fit the defect size.

Bench testing was conducted to compare Geistlich Fibro-Gide® against its predicate device in terms of its appearance, porosity, amino acid and protein composition, capillarity and wettability, as well as trimming, suturing and degradation capabilities.

CharacteristicGeistlich Fibro-Gide®Geistlich Mucograft(K102531 – Predicatedevice; K140518, K073711,K012423 – Referencedevices)
Intended Use
Intended UseSoft Tissue AugmentationSoft Tissue Augmentation
Indications for UseGeistlich Fibro-Gide® is
CharacteristicGeistlich Fibro-Gide®Geistlich Mucograft(K102531 - Predicatedevice; K140518, K073711,K012423 - Referencedevices)
intended for soft tissueaugmentation. It is indicatedfor:Localized gingivalaugmentation toincrease keratinizedtissue (KT) aroundteeth and implants Alveolar ridgereconstruction forprosthetic treatment Recession defectsfor root coverageCovering of implantsplaced in immediateor delayed extractionsockets Localized gingivalaugmentation toincrease keratinizedtissue (KT) aroundteeth and implants Alveolar ridgereconstruction forprosthetic treatment Recession defectsfor root coverage
Physical Characteristics
Dimensions15x20 mm, 20x40mm(thickness app. 6mm)15x20 mm, 20x30 mm(K102531)30x40 mm (K073711,K012423)(thickness app. 2.5 - 5 mm)
FormSponge-like matrixSponge-like matrix
ColorWhite to almost whiteAlmost white
PorosityVery porous materials withaverage pore values of 90%or more. The averagesurface area and the bulkdensity of Geistlich Fibro-Gide® is almost identical tothat of Geistlich Mucograft.Very porous materials withaverage pore values of 90%or more. The averagesurface area and the bulkdensity of GeistlichMucograft is almost identicalto that of Geistlich Fibro-Gide.
CharacteristicGeistlich Fibro-Gide®Geistlich Mucograft(K102531 - Predicatedevice; K140518, K073711,K012423 - Referencedevices)
Capillarity and wettabilitySpontaneously wettable andcompletely soaked inaqueous solution in lessthan one minuteSpontaneously wettable andcompletely soaked inaqueous solution in lessthan one minute
Cross-linkingChemically cross-linked withEDC and NHSNot chemically cross-linked
Easy to trim with surgicalinstruments (to fit atemplate)Can be cut with surgicalinstrumentsCan be cut with surgicalinstruments
Can be suturedCan be fixed with sutures,as demonstrated bysutureability testingCan be fixed with sutures,as demonstrated bysutureability testing
Composition Materials
Raw MaterialPorcine connective tissue,Porcine connective tissue,
Porcine skin tissuePorcine skin tissue
CompositionPorcine collagen: Product isproduced from the sameintermediate collagenousproductsPorcine collagen: Product isproduced from the sameintermediate collagenousproducts
Collagen / ElastinMajor protein componentsCollagen I and Collagen III(no Collagen II) and ElastinMajor protein componentsCollagen I and Collagen III(no Collagen II) and Elastin
Amino Acid CompositionVery similar with equalamounts of amino acidsVery similar with equalamounts of amino acids
FatTrace (less than 0.5%)Trace (less than 0.5%)
GlycosaminoglycansTrace (less than 0.5%)Trace (less than 0.5%)
Other proteins > 0.5%NoneNone
pH3-73-7
CharacteristicGeistlich Fibro-Gide®Geistlich Mucograft(K102531 - Predicatedevice; K140518, K073711,K012423 - Referencedevices)
Other Characteristics
Source of raw materialIdentical porcine tissueIdentical porcine tissue
ManufactureMultistage validated, SOPcontrolled purificationprocessMultistage validated, SOPcontrolled purificationprocess
Manufacturing conditionsQuality systems regulation(CFR Part 820)Quality systems regulation(CFR Part 820)
PackagingConforms to ISO 11607,Parts 1 and 2.Sterile double layerpackaging, includingaluminum layer to protectagainst vapor penetration.Conforms to ISO 11607,Parts 1 and 2.Sterile double layerpackaging, in aluminumpouch to protect againstvapor penetration.
Manufacture / PackaginglocationGeistlich Pharma AG,Wolhusen,SwitzerlandGeistlich Pharma AG,Wolhusen,Switzerland
Sterilization locationSynergy Health, Daeniken,SwitzerlandSynergy Health, Daeniken,Switzerland
SterilityGamma irradiationSAL 10-6; Device providedsterile, for single use onlyGamma irradiationSAL 10-6; Device providedsterile, for single use only
UserRestricted to licenseddentistsRestricted to licenseddentists
Principles of Operation
Principles of OperationImplantable resorbablecollagen matrix (scaffold)consisting of collagen fibersin a loose, porousImplantable resorbablecollagen matrix (scaffold)consisting of collagen fibersin a loose, porous
CharacteristicGeistlich Fibro-Gide®Geistlich Mucograft(K102531 - Predicatedevice; K140518, K073711,K012423 - Referencedevices)
arrangement to enable cellinvasionarrangement to enable cellinvasion
Performance Standards
Conformity to standards- ISO 10993-1(Biocompatibility)- 10993-2 (Animal Welfare)- 10993-3 (Genotoxicity)- 10993-6 (Local Effects)- 10993-10 (Irritation /Sensitization)- 10993-11 (Systemic Tox.)- 10993-12 (Sample Prep.)- ISO 11137-1 (SterilizationVal)- ISO 11137-2 (SterilizationDose)- ISO 11607-1 (Sterilization)- ISO 11607-2 (Sterilization)- ISO 11737-1 (Sterilization)- ISO 11737-2 (Sterilization)- ISO 11737-3 (Sterilization)- ISO 11607 (Packaging)- ISO 14698-1(Cleanrooms)- ISO 14971 (RiskManagement)- ISO 22441-1 (AnimalTissues)- ISO 22442-2 (AnimalTissues)- ISO 22442-3 (ViralClearance)- USP 39 NF34 <151>:Pyrogen test- USP 39 NF34 85:Endotoxin test- ASTM F1980 (AcceleratedAging)- ASTM F2450-10 (TissueEngineering)- ISO 10993-1(Biocompatibility)- 10993-2 (Animal Welfare)- 10993-3 (Genotoxicity)- 10993-6 (Local Effects)- 10993-10 (Irritation /Sensitization)- 10993-11 (Systemic Tox.)- 10993-12 (Sample Prep.)- ISO 11137-1 (SterilizationVal)- ISO 11137-2 (SterilizationDose)- ISO 11607-1 (Sterilization)- ISO 11607-2 (Sterilization)- ISO 11737-1 (Sterilization)- ISO 11737-2 (Sterilization)- ISO 11737-3 (Sterilization)- ISO 11607 (Packaging)- ISO 14698-1(Cleanrooms)- ISO 14971 (RiskManagement)- ISO 22441-1 (AnimalTissues)- ISO 22442-2 (AnimalTissues)- ISO 22442-3 (ViralClearance)- USP 39 NF34 <151>:Pyrogen test- USP 39 NF34 85:Endotoxin test- ASTM F1980 (AcceleratedAging)- ASTM F2450-10 (TissueEngineering)

Comparison Table

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

PERFORMANCE DATA

All relevant biocompatibility tests were conducted as required according to Guidance for Industry and Food and Drug Administration Staff Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. Testing included Cytotoxicity, Irritation, Sensitization, Acute System Toxicity, Genotoxicity, Subchronic System Toxicity, Chronic System Toxicity, Implantation, Pyrogenicity, and Leachables. Other testing included viral clearance studies and residual chemical testing and toxicological assessment. Results indicate that the device is biocompatible.

Test (Standard)Test method/ modelResults
Cytotoxicity (ISO 10993-5)In vitro mouse L929 fibroblast cellculture assayPass
Irritation (ISO 10993-10)Intracutaneous reactivity in the rabbitPass
Sensitization (ISO 10993-10)Guinea pig maximization testPass
Acute systemic toxicity (ISO 10993-11)Acute systemic toxicity test in themousePass
Pyrogenicity (USP <151>)Rabbit pyrogen testPass
Genotoxicity (ISO 10993-3)Bacterial reverse mutagenicity assayin Salmonella typhimurium andEscherichia coli (Ames test)In vitro chromosomal aberration studyin human lymphocytesMouse peripheral blood micronucleusstudyPass
Local tissue response afterimplantation (ISO 10993-6)4-week subcutaneous implantation inratsPass

{10}------------------------------------------------

12-week subcutaneous implantation in ratsPass
26-week subcutaneous implantation in ratsPass
Subchronic systemic toxicity (ISO 10993-11)4-week subcutaneous implantation in rats, with systemic toxicity endpointPass
Chronic systemic toxicity (ISO 10993-11)26-week subcutaneous implantation in rats, with systemic toxicity endpointPass
Leachables (ISO 10993-17)GC/MS fingerprint and ICP analysisPass

Geistlich Fibro-Gide® was tested in a rat and in a dog study compliant with 21 CFR Part 58 to demonstrate substantial equivalence, safety and performance to its predicate device. The studies confirmed tissue integration, continuous resorption and comparable degradation rate of Geistlich Fibro-Gide® and the predicate devices. Further, the investigation of distal organs, hematologic parameters and clinical chemistries confirmed that the safety profile is acceptable. The dog study also investigated soft tissue augmentation and local tissue effects after several soft tissue augmentation periods in Geistlich Fibro-Gide® and the predicate device Geistlich Mucograft®. In all instances, the subject device was demonstrated to be substantially equivalent to the predicate device.

The clinical data from a controlled, parallel and randomized study in patients presenting insufficient soft tissue volume demonstrated product safety and effectiveness in the product's indications for use.

The biocompatibility testing, in combination with the bench testing, animal studies, and published clinical data included in this submission, demonstrates the substantial equivalence of Geistlich Fibro-Gide® to its predicate device, Geistlich Mucograft®.

CONCLUSION

Based on the data provided within this 510(k) submission as summarized above, it can be concluded that Geistlich Fibro-Gide® is substantially equivalent to the predicate device Geistlich Mucograft® with regard to intended use and indications for use, technological characteristics, including principles of operation, and performance characteristics as shown in a series of biocompatibility, bench, animal, and clinical testing. Thus, the subject device is substantially equivalent.

§ 872.3930 Bone grafting material.

(a)
Identification. Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.(b)
Classification. (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.