K Number
K102531

Validate with FDA (Live)

Device Name
MUCOGRAFT
Date Cleared
2010-11-17

(75 days)

Product Code
Regulation Number
872.3930
Panel
Dental
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use
  • covering of implants placed in immediate or delayed extraction sockets;
  • localized gingival augmentation to increase keratinized tissue (KT) around teeth and implants;
  • alveolar ridge reconstruction for prosthetic treatment;
  • guided tissue regeneration procedures in recession defects for root coverage
Device Description

MUCOGRAFT® is a pure collagen membrane obtained by a standardized controlled manufacturing process. The membrane is made of collagen type I and type III without further cross-linking or chemical treatment. The collagen is extracted from veterinary certified pigs and is carefully purified to avoid antigenic reactions. MUCOGRAFT® is sterilized in double blisters by gamma irradiation. MUCOGRAFT® has a bilayer structure with one smooth, non-permeable layer and one porous. The "outer," smooth side has a smooth surface which is cell occlusive and prevents cell adhesion and acts as a barrier. It allows tissue adherence favoring wound healing. This side is turned towards the soft tissue. The smooth texture has appropriate elastic properties to accommodate i suturing to the host mucosal margins and to protect the graft material from oral trauma during biode radation and healing. The "inner" porous layer consists of collagen fibers in a loose, porous arrangement to enable cell invasion. This porous layer is made from pig skin. This side is turned toward the bone defect and/or soft tissue to encourage boneforming cells and tissue growth and to stabilize the blood clot.

AI/ML Overview

The provided text describes the MUCOGRAFT® Collagen Matrix, a pure collagen membrane for dental grafting procedures. It establishes substantial equivalence to previous versions and other collagen membranes, rather than presenting a study to prove acceptance criteria for a new AI-powered medical device. Therefore, much of the requested information regarding AI device testing is not applicable.

Here's a breakdown of the available information based on your request:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state acceptance criteria in terms of quantitative metrics for a device's performance that would typically be seen in an AI device study (e.g., sensitivity, specificity, AUC). Instead, it relies on demonstrating substantial equivalence to predicate devices and evidence of safety and efficacy through preclinical and clinical studies for specific indications.

Acceptance Criteria (Implied for Substantial Equivalence):

CriterionReported Device Performance (MUCOGRAFT®)
Similar Material CompositionPure collagen type I and type III, derived from veterinary certified pigs.
Similar Bilayer StructureSmooth, non-permeable layer (cell occlusive) and porous layer (cell invasion).
Similar Manufacturing ProcessStandardized controlled manufacturing, gamma irradiation sterilization.
Similar Intended Use & IndicationsRevised indications are similar to existing MUCOGRAFT® and DYNAMATRIX® products.
Safety and Efficacy in Preclinical StudiesUneventful healing, normal-appearing tissue, similar revascularization and epithelialization to autogenous graft (in animal models).
Safety and Efficacy in Clinical StudiesSafe and effective for gingival augmentation (increase KT/tissue thickness) and root coverage, with results similar to DynaMatrix.

2. Sample Size Used for the Test Set and Data Provenance

  • Preclinical Studies:
    • Wehrhan et al. 2010: Porcine model (sample size not specified).
    • Herford and Boyne 2002: Primate model (sample size not specified).
  • Clinical Studies:
    • Multiple clinical studies are mentioned, but specific sample sizes for each are not provided in this document. The studies are cited as "clinical studies have been conducted" and are not detailed within this summary.
  • Data Provenance: The document implies an international scope given the sponsor's location (Switzerland) and the cited authors. Some cited studies (e.g., Herford and Boyne, McGuire and Scheyer) are from the US. The studies are assumed to be prospective clinical trials, but this is not explicitly stated for all.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

N/A. This information is typically relevant for studies using expert-labeled data as ground truth for AI algorithms. For this medical device (a collagen matrix), the "ground truth" is established through clinical and histological outcomes observed by researchers and clinicians in the preclinical and clinical studies, rather than a separate expert review panel for a test set.

4. Adjudication Method for the Test Set

N/A. This is relevant for studies where multiple annotators or reviewers might disagree on ground truth labels, necessitating an adjudication process. This is not applicable to the preclinical and clinical studies evaluating a physical implantable device.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

N/A. This concept is specific to evaluating AI's impact on human performance (e.g., radiologists interpreting images with and without AI assistance). The MUCOGRAFT® device is a biomaterial, not an AI diagnostic/interpretive tool, so an MRMC study is not relevant.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study Done

N/A. As MUCOGRAFT® is not an algorithm, this question is not applicable.

7. Type of Ground Truth Used

For the preclinical and clinical studies, the "ground truth" is based on:

  • Histological and pathological examination: For outcomes like uneventful healing, normal-appearing tissue, revascularization, and epithelialization (e.g., in animal models).
  • Clinical observation and measurement: For outcomes like gingival augmentation (increase in keratinized tissue width/thickness) and root coverage, assessed by clinicians based on established dental criteria.

8. Sample Size for the Training Set

N/A. The concept of a "training set" applies to machine learning algorithms. This document describes a medical device tested through traditional scientific studies (preclinical and clinical), not an AI system.

9. How the Ground Truth for the Training Set Was Established

N/A. As there is no training set mentioned for an AI algorithm, this question is not applicable.

{0}------------------------------------------------

102531

NOV 1 7 2010

510(k) Summary

MUCOGRAFT®

l, SPONSOR

Ed. Geistlich Soehne Ag für Chemische Industrie Geistlich Pharma Ag Bahnhofstrasse 40 CH-6110 Wolhusen SWITZERLAND

Contact Person: Peter S. Reichertz, (202) 772-5333 Date Prepared: October 29, 2010

DEVICE NAME 2.

Proprietary Name Common/Usual Name: . Classification Name:

MUCOGRAFT® Collagen Matrix Barrier, Animal Source, Dental

PREDICATE DEVICES 3.

MUCOGRAFT® (K012423, K061244 and K073711) DYNAMATRIX K010952

. INTENDED USE

MUCOGRAFT® Collagen Matrix currently is indicated for:

  • Simultaneous use of GBR-membrane (MUCOGRAFT) and implants; -
  • Covering of implants placed in immediate extraction sockets; -
  • Covering of implants placed in delayed extraction sockets; -
  • Localized ridge augmentation for later implantation; -
  • Alveolar ridge reconstruction for prosthetic treatment; -
  • Covering of bone defects after root resection, cystectomy, removal of retained teeth;
  • Guided bone regeneration in dehiscence defects; and
  • Guided tissue regeneration procedures in periodontal and recession defects.

{1}------------------------------------------------

K1025

DEVICE DESCRIPTION

ડ.

MUCOGRAFT® is a pure collagen membrane obtained by a standardized controlled manufacturing process. The membrane is made of collagen type I and type III without further cross-linking or chemical treatment. The collagen is extracted from veterinary certified pigs and is carefully purified to avoid antigenic reactions. MUCOGRAFT® is sterilized in double blisters by gamma irradiation. MUCOGRAFT® has a bilayer structure with one smooth, non-permeable layer and one porous. The "outer," smooth side has a smooth surface which is cell occlusive and prevents cell adhesion and acts as a barrier. It allows tissue adherence favoring wound healing. This side is turned towards the soft tissue. The smooth texture has appropriate elastic properties to accommodate i suturing to the host mucosal margins and to protect the graft material from oral trauma during biode radation and healing. The "inner" porous layer consists of collagen fibers in a loose, porous arrangement to enable cell invasion. This porous layer is made from pig skin. This side is turned toward the bone defect and/or soft tissue to encourage boneforming cells and tissue growth and to stabilize the blood clot.

BASIS FOR SUBSTANTIAL EQUIVALENCE 6.

MUCOGRAFT® collagen matrix is substantially equivalent to Geistlich's existing products MUCOGRAFT® (K012423, K061244 and K073711) and DYNAMATRIX® (K082058 and K010952). The current 510(k) requests clearance of MUCOGRAFT® for a · revised set of indications regarding guided tissue regeneration, as follows:

  • covering of implants placed in immediate or delayed extraction sockets; ●
  • localized gingival augmentation to increase keratinized tissue (KT) around teeth and implants:
  • alveolar ridge reconstruction for prosthetic treatment; .
  • recession defects for root coverage . ●

These indications are similar to those accepted for MUCOGRAFT® and also for DYNAMATRIX®, which is cleared for gingival augmentation. MUCOGRAFT®, like DYNAMATRIX®, is a collagen membrane used in dental grafting procedures.

Preclinical and clinical studies support the safety and efficacy of the product for these indications for use.

{2}------------------------------------------------

Two preclinical studies have been conducted in two animal models: these studies provide evidence that Mucograft can be used successfully for dermal grafting (Wehrhan et al. 2010) and oral soft tissue augmentation (Herford and Boyne 2002). Wehrhan et al. chose a porcine model for their study, and Herford and Boyne used a primate model.

The two preclinical studies provide evidence of the following outcomes with Mucograft:

Uneventful healing (Herford and Boyne 2002, Wehrhan et al. 2010); and

  • Normal-appearing tissue: (1) Normal-appearing mucosa, complete re-epithelialization ● and generation of attached mucosa/ keratinized tissue (Herford and Boyne 2002); and (2) Similar revascularization and epithelialization to split thickness autogenous graft (Wehrhan et al. 2010).
    Clinical studies have been conducted to evaluate the use of (1) Mucograft, as well as its predicate DynaMatrix, for gingival augmentation to increase keratinized tissue (KT, also referred to as "keratinized gingival" or "KG") and/or tissue thickness and (2) Mucograft for root coverage. These clinical studies provide evidence for the safe and effective use of Mucograft when placed (1) on top of a surgically-prepared wound bed or underneath a surgical flap, and (2) around teeth or dental implants.

The clinical studies (and their respective manuscripts, publications, or other source documents) provide clinical evidence of the device's use for gingival augmentation and root coverage procedures and demonstrates that it can be used safely and effectively on top of a surgicallycreated wound bed, underneath a surgical flap, around teeth, and around dental implants. In addition, the results with MUCOGRAFT® are similar to the published results reported for DynaMatrix.

This body of valid scientific evidence substantiates the proposed, more specific wording changes to the indications for use statement (changes are in bold):

  • Localized gingival augmentation to increase keratinized tissue (KT) around teeth or implants:
  • Localized gingival augmentation to increase tissue thickness around teeth or . implants; and
  • Recession defects for root coverage.

-3-

{3}------------------------------------------------

Supporting Published Literature

Charles A. A practice-based clinical research evaluation of the effectiveness of a collagen matrix used to augment keratinized tissue around dental implants. Manuscript in preparation.

Herford AS, Akin A, Cicciu M, Maiorana C, Boyne PJ. Use of a porcine collagen matrix as an alternative to autogenous tissue for grafting oral soft tissue defects. J Oral Maxillofac Surg 68:1463-1470, 2010.

Herford AS and Boyne PJ. Evaluation of a special collagen implant material as a substitute for free mucosal or skin grafts in oral soft tissue surgery. Proceedings of the American Institute of Oral Biology, Palm Springs, CA, October 2002 (pp 103-109), McGuire MK and Scheyer ET. Xenogeneic collagen matrix with coronally advanced flap compared to connective tissue with coronally advanced flap for the treatment of dehiscence-type recession defects. J. Periodontol 81:1108-1117, 2010.

Nevins M, Nevins ML, Camelo M, Borges Camelo JM, Schupbach P, Kim DM. The clinical efficacy of DynaMatrix Extracellular Membrane in augmenting keratinized tissue. Int J Periodontics Restorative Dent. 30:151-161, 2010.

Panaite D. Practice-based clinical research evaluation of the effectiveness of Mucograft® Collagen Matrix to increase keratinized tissues around dental implants. Abstract presented at the 24th annual meeting of the Academy of Osseointegration (San Diego, CA, February 26-28, 2009).

Sanz M. Lorenzo R. Aranda JJ, Martin C, Orsini M. Clinical evaluation of a new, collagen matrix (Mucograft® prototype) to enhance the width of keratinized tissue in patients with fixed prosthetic restorations: A randomized prospective clinical trial. J Clin Periodontol. 36:868-876, 2009. Epub 2009 Aug 12

Wehrhan F, Nkenke E, Melnychenko I, Amann K, Schlegel KA, Goerlach C, Zimmermann W-H. Schultze-Mosgau S. 'Skin repair using a porcine collagen VIII membrane - Vascularization and epithelialization properties. Dermatol Surg 36:919-930, 2010.

-4-

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three bars representing its wings, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the eagle. The logo is black and white.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

NOV 1 7 2010

Ed. Geisilich Sochne Ag Fur Chemische Industrie C/O Mr. Peter S. Reichertz Sheppard Mullin Richter & Hampton LLP 1300 I Street N.W., 11th Floor East Washington, DC 20005

Re: K102531

Trade/Device Name: MUCOGRAFT® Collagen Matrix Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: II Product Code: NPL Dated: October 29, 2010 Received: November 1, 2010

Dear Mr. Reichertz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2- Mr. Reichertz

. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRI-'s Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

h for

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and

Radiological Health

Enclosure

{6}------------------------------------------------

K!02531

Indications for Use

510(k) Number (if known):

Device Name:

MUCOGRAFT® Collagen Matrix.

Indications for Use:

.

  • D covering of implants placed in immediate or delayed extraction sockets;
  • localized gingival augmentation to increase keratinized tissue (KT) around teeth and implants;
  • alveolar ridge reconstruction for prosthetic treatment; U
  • guided tissue regeneration procedures in recession defects for root coverage

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use X (Part 21 CFR 801 Subpart D) OR

Over-the-Counter Use (21 CFR 801 Subpart C) ·

Susan Rurrei

(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital
Division of Anesthesiology, General Devices Division of Antonios of Antonio Devices

(10953

510(k) Number: _

  • | -

W02-EAST:9PSR1\200322663.1

§ 872.3930 Bone grafting material.

(a)
Identification. Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.(b)
Classification. (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.