K Number
K170938

Validate with FDA (Live)

Device Name
CalSet IGF-1
Manufacturer
Date Cleared
2017-04-26

(27 days)

Product Code
Regulation Number
862.1150
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

CalSet IGF-1 is used for calibrating the quantitative Elecsys IGF-1 assay on the Elecsys and cobas e immunoassay analyzers.

Device Description

CalSet IGF-1 is used for calibrating the quantitative Elecsys IGF-1 assay on the Elecsys and cobas e immunoassay analyzers. CalSet IGF-1 is a two concentration level set of lyophilized human serum matrix that is traceable to the WHO IS 02/254 Reference Material. The CalSet includes:

  • IGF-1 Call: 2 bottles, each for 1.0 mL of calibrator 1 .
  • IGF-1 Cal2: 2 bottles, each for 1.0 mL of calibrator 2 .
    IGF-1 in two concentration ranges (approximately 20 ng/mL and approximately 500 ng/mL) in a human serum matrix.
AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the CalSet IGF-1 device:

Acceptance Criteria and Device Performance Study for CalSet IGF-1

The CalSet IGF-1 is a calibrator used for the quantitative Elecsys IGF-1 assay. Its performance is primarily evaluated through studies demonstrating the stability and accurate value assignment of the calibrator itself.

1. Table of Acceptance Criteria and Reported Device Performance

FeatureAcceptance CriteriaReported Device Performance
Value AssignmentAssigned values must fall within specified acceptable ranges for each lot.The assigned value of each calibrator is defined as the median value obtained over at least six (6) runs on at least three (3) modules of the respective calibrator. Values must fall within specified acceptable ranges for each lot. (Implied met, as the overall conclusion states the device performs as well or better than the predicate).
Reconstituted Stability at -20°CStable for at least 31 days at -20 ± 5°C.The reconstituted CalSet IGF-1 is stable for at least 31 days at -20 ± 5℃.
Reconstituted Stability at 2-8°CStable for at least 72 hours at 2-8°C.The reconstituted CalSet IGF-1 is stable for at least 72 hours at 2-8°C.
On-board Stability (20-25°C)Stable for at least 6 hours on-board the analyzers at 20-25°C.The reconstituted CalSet IGF-1 is stable for at least 6 hours on-board the analyzers at 20-25°C.
Accelerated Stability (35°C)The accelerated stability model employed met the specification (specific quantitative criteria not provided in the text).The accelerated stability model employed met the specification.
Real-time Stability (Lyophilized)Stable up to the stated expiration date (e.g., 12 months).The lyophilized CalSet IGF-1 is stable up to the stated expiration date.

2. Sample Size Used for the Test Set and Data Provenance

The provided document describes performance validation studies for the calibrator itself, rather than a diagnostic device tested on patient samples. Therefore, the concept of a "test set" in the context of patient data with country of origin and retrospective/prospective nature doesn't directly apply here.

  • Value Assignment Test Set: For value assignment, calibrators are run in duplicate on four (4) cobas e 411 modules and four (4) modules of the MODULAR ANALYTICS E170/cobas e 602 with all Elecsys IGF-1 reagent lots available. The assigned value is the median over at least six (6) runs on at least three (3) modules.
  • Stability Studies Test Set: For stability studies, "on-test and reference materials" (which are CalSet IGF-1 samples under different storage conditions) were tested in duplicate on the cobas e 411 immunoassay analyzer. A "freshly reconstituted set of CalSet IGF-1 (stored at 2 - 8°C)" served as the reference material for accelerated stability.
  • Data Provenance: The studies were performed by Roche Diagnostics, a global company with facilities in Germany and the United States (as per the establishment registration numbers). The specific location where these studies were conducted is not explicitly stated, but it's internal company data for product validation. The data is prospective as it describes studies performed to validate the manufactured calibrator.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable as the device is a calibrator, not a diagnostic tool requiring expert interpretation of patient data. The "ground truth" for the calibrator's value assignment is based on its traceability to a defined international standard (WHO IS 02/254 Reference Material) and the analytical performance of the Elecsys IGF-1 assay on specific immunoassay analyzers.

4. Adjudication Method for the Test Set

This is not applicable as the device is a calibrator. The value assignment process involves calculating a median value from multiple runs on multiple modules, which is an objective statistical calculation, not a subjective adjudication by experts.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that require human interpretation (e.g., radiologists reading images) where the AI assists the human reader. CalSet IGF-1 is a calibrator for an automated immunoassay, and human interpretation is not part of its intended use or evaluation.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

The device itself is a calibrator, which functions without "human-in-the-loop" once it's used in the automated immunoassay system. The performance studies described (value assignment, stability) are essentially "standalone" evaluations of the calibrator's characteristics and its interaction with the specified immunoassay system. There isn't an "algorithm" in the typical sense; rather, it's a precisely defined chemical reference material.

7. The Type of Ground Truth Used

The ground truth for the CalSet IGF-1 is the WHO IS 02/254 Reference Material. This is an internationally recognized standard for IGF-1 quantification, providing the basis for traceability and accurate value assignment to the calibrator.

8. The Sample Size for the Training Set

This information is not applicable. The CalSet IGF-1 is a calibrator, not an AI/ML algorithm. Therefore, there is no "training set" in the context of machine learning. The calibrator's formulation and manufacturing processes are established based on scientific principles and analytical chemistry, not data training.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable. As stated above, there is no training set for this device. The ground truth for the calibrator's values is established through traceability to the WHO IS 02/254 Reference Material.

{0}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/0/Picture/1 description: The image shows the logo for the United States Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 26, 2017

ROCHE DIAGNOSTICS LISA KLINEDINST REGULATORY AFFAIRS PRINCIPAL 9115 HAGUE ROAD INDIANAPOLIS IN 46250

Re: K170938

Trade/Device Name: CalSet IGF-1 Regulation Number: 21 CFR 862.1150 Regulation Name: Calibrator Regulatory Class: II Product Code: JIT Dated: March 29, 2017 Received: March 30, 2017

Dear Lisa Klinedinst:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kellie B. Kelm -S

for Courtney H. Lias,, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K170938

Device Name CalSet IGF-1

Indications for Use (Describe)

CalSet IGF-1 is used for calibrating the quantitative Elecsys IGF-1 assay on the Elecsys and cobas e immunoassay analyzers.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

CalSet IGF-1: 510(k) Summary for K170938

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

In accordance with 21 CFR 807.87, Roche Diagnostics hereby submits official notification as required by Section 510(k) of the Federal Food, Drug and Cosmetics Act of our intention to market the devices described in this Premarket Notification 510(k).

The purpose of this Traditional 510(k) Premarket Notification is to obtain FDA review and clearance for CalSet IGF-1.

{4}------------------------------------------------

Submitter NameRoche Diagnostics
Address9115 Hague RoadP.O. Box 50416Indianapolis, IN 46250-0457
ContactLisa K. KlinedinstPhone: (317) 521-1942FAX: (317) 521-2324Email: lisa.klinedinst@roche.com
Date PreparedApril 26, 2017
510(k)K170938
Proprietary NameCalSet IGF-1
Common NameCalSet IGF-1
Classification NameCalibrator, Secondary
Product Codes,Regulation NumbersJIT, 862.1150
Predicate DevicesCalSet Vitamin D total II (K162840)
Establishment RegistrationThe establishment registration numbers:Roche Diagnostics GmbH in Mannheim, Germany is 9610126,Roche Diagnostics Penzberg, Germany is 9610529Roche Diagnostics in the United States is 1823260

{5}------------------------------------------------

DEVICE DESCRIPTION: CALSET IGF-1 1.

CalSet IGF-1 is used for calibrating the quantitative Elecsys IGF-1 assay on the Elecsys and cobas e immunoassay analyzers. CalSet IGF-1 is a two concentration level set of lyophilized human serum matrix that is traceable to the WHO IS 02/254 Reference Material. The CalSet includes:

  • IGF-1 Call: 2 bottles, each for 1.0 mL of calibrator 1 .
  • IGF-1 Cal2: 2 bottles, each for 1.0 mL of calibrator 2 .

IGF-1 in two concentration ranges (approximately 20 ng/mL and approximately 500 ng/mL) in a human serum matrix.

INTENDED USE FOR CALSET IGF-1 2.

CalSet IGF-1 is used for calibrating the quantitative Elecsys IGF-1 assay on the Elecsys and cobas e immunoassay analyzers.

3. TECHNOLOGICAL CHARACTERISTICS

The following table compares the CalSet IGF-1 with its predicate device.

{6}------------------------------------------------

FeaturePredicate Device(CalSet Vitamin D total II)Candidate Device
Intended Use/Indications for UseCalSet Vitamin D total II is used forcalibrating the quantitative ElecsysVitamin D total II assay on the cobas e 411immunoassay analyzer.CalSet IGF-1 is used for calibrating thequantitative Elecsys IGF-1 assay on theElecsys and cobas e immunoassay analyzers.
Analyte25-hydroxyvitamin D3IGF-1
MatrixHuman serum matrix with added25-hydroxyvitamin D3Human serum matrix with added IGF-1
LevelsTwoSame
Target RangesCal 1: approximately 2 ng/mLCal 2: approximately 45 ng/mLCal 1: approximately 20 ng/mLCal 2: approximately 500 ng/mL
FormatLyophilizedSame
HandlingCarefully dissolve the contents of onebottle by adding exactly 1.0 mL of distilledor deionized water and allow to standclosed for 15 minutes to reconstitute. Mixcarefully, avoiding the foam formation.Same
Storage and StabilityReconstituted calibrators:Reconstituted calibrators:
At 2-8°C – 72 hoursAt 2-8°C - 72 hours
At -20°C - 12 weeks (freeze only once)on the cobas e 411 analyzer at 20-25°C- upto 5 hourson MODULAR ANALYTICS E170, cobas ee 601 and cobas e 602 analyzers at 20-25°CAt -20° ± 5°C - 31 dayson the cobas e 411 analyzer at 20-25°C- up to6 hourson MODULAR ANALYTICS E170, cobas e601 and cobas e 602 analyzers at 20-25°C -

Calibrator Comparison Table 1:

Non-Clinical Performance Evaluation

Non-clinical performance evaluation for CalSet IGF-1 is briefly summarized below.

Standardization and Traceability 3.1.

The Elecsys IGF-1 assay is traceable to the WHO IS 02/254 Reference Material. The Reference Material (WHO IS 02/254) is used for Reference Standardization to assign values to the Working Calibrator consisting of a native human serum sample panel (single donors, pools and spiked samples covering the entire measuring range). These Working Calibrators are for routine use in subsequent standardizations. The assigned values for CalSet IGF-1 are read from the Working Calibrator Curve.

{7}------------------------------------------------

Value Assignment 3.2.

The CalSet IGF-1 assigned values are determined with the Elecsys IGF-1 assay. A Reference Material (WHO IS 02/254) is used for Reference Standardization to assign values to the Working Calibrator consisting of a native human serum sample panel (single donors, pools and spiked samples covering the entire measuring range). These Working Calibrators are for routine use in subsequent standardizations. The assigned values for CalSet IGF-1 are read from the Working Calibrator Curve.

For each CalSet IGF-1 manufactured, the calibrators are run in duplicate on four (4) cobas e 411 modules and four (4) modules of the MODULAR ANALYTICS E170/cobas e 602 with all Elecsys IGF-1 reagent lots available. The assigned value of each calibrator is defined as the median value obtained over at least six (6) runs on at least three (3) modules of the respective calibrator. The assigned values for CalSet IGF-1 are read from the Working Calibrator Curve. The values must fall within specified acceptable ranges for each lot.

Calibrator Stability 3.3.

Five studies were performed in order to verify the stability claims for CalSet IGF-1.

  • Study 1: Stability at -20°C after reconstitution combined study .
  • Study 2: Stability at 2-8°C after reconstitution combined study .
  • Study 3: On-board stability at 20-25°C after reconstitution combined study .
  • Study 4: Accelerated stability at 35°C combined study .
  • Study 5: Real-time stability .
  • 3.3.1. Study 1: Stability at -20°C after reconstitution - combined study

The on-test and reference materials were tested in duplicate on the cobas e 411 immunoassay analyzer. The on-test material was reconstituted and stored in closed vials for 3 weeks at 35℃ + 32 days at -20 ± 5°C + 7 hours at 20-25°C. The on-test signal recovery was calculated as percent of the reference value.

The reconstituted CalSet IGF-1 is stable for at least 31 days at -20 ± 5℃.

{8}------------------------------------------------

3.3.2. Study 2: Stability at 2-8°C after reconstitution - combined study

The on-test and reference materials were tested in duplicate on the cobas e 411 immunoassay analyzer. The on-test material was reconstituted and stored in closed vials for 3 weeks at 35°C + 73 hours at 2-8°C + 7 hours at 20-25°C. The on-test signal recovery was calculated as percent of the reference value.

The reconstituted CalSet IGF-1 is stable for at least 72 hours at 2-8°C.

3.3.3. Study 3: On-board stability at 20-25°C after reconstitution - combined study

The on-test and reference materials were tested in duplicate on the cobas e 411 immunoassav analyzer. The on-test material was reconstituted and stored in closed vials for 3 weeks at 35°C + 32 days at -20 ± 5°C + 7 hours at 20-25°C. The on-test signal recovery was calculated as percent of the reference value.

The reconstituted CalSet IGF-1 is stable for at least 6 hours on-board the analyzers at 20-25°C.

Study 4: Accelerated stability at 35°C – combined study 3.3.4.

The on-test and reference materials were tested in duplicate on the cobas e 411 immunoassay analyzer. The on-test material was stored lyophilized (as supplied to the user) at 35°C for 3 weeks and afterwards reconstituted and stored for 73 hours at 2-8°C + 7 hours at 20-25°C. The reference material was a freshly reconstituted set of CalSet IGF-1 (stored at 2 - 8°C). After 3 weeks, the on-test and reference materials were tested in duplicate. The on-test signal recovery was calculated as a percent of the reference value.

The accelerated stability model employed met the specification.

3.3.5. Study 5: Real-time stability

For the on-going real-time stability study, the CalSet IGF-1 test material is stored at 2-8℃. The CalSets are tested in triplicate at specified intervals over the shelf life of the device up to the planned shelf life plus one month (13 months).

Real-time stability is calculated based on the recovery of signal of stressed calibrator (stored at 2-8°C) vs. unstressed calibrator (stored at -80°C). At the specified intervals over the shelf life,

{9}------------------------------------------------

the mean value of the stressed calibrator was calculated as a percent recovery of the unstressed value (each set is tested in duplicates at the same time point).

The lyophilized CalSet IGF-1 is stable up to the stated expiration date.

4. CONCLUSIONS

The information provided in this 510(k) Premarket Notification will support a determination of substantial equivalence for CalSet IGF-1. The data supports safe, effective devices which perform as well or better than the predicate device.

§ 862.1150 Calibrator.

(a)
Identification. A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.