K Number
K170686

Validate with FDA (Live)

Date Cleared
2017-07-14

(129 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

BRIGHTWAY BRAND NITRILE EXAMINATION GLOVES, POWDER FREE [PURPLE] TESTED FOR USE WITH CHEMOTHERAPY DRUGS is a disposable device intended for medical purpose worn on the examiners hand or finger to prevent contamination between patient and examiner.

The tested chemotherapy drugs and their breakthrough detection times are as follows :

Test Chemotherapy Drug Name and ConcentrationMinimum Breakthrough Detection Time
· Blenoxane (15mg/ml),(15,000 ppm)>240 Minutes
· Busulfan(6mg/ml),(6,000 ppm)>240 Minutes
· Carmustine (BCNU) (3.3mg/ml),(3,300 ppm)3.6 Minutes
· Cisplatin(1.0mg/ml), (1.000 ppm)>240 Minutes
· Cyclophosphamide/Cytoxan(20mg/ml), (20,000 ppm)>240 Minutes
· Cytarabine(100mg/ml), (100,000 ppm)>240 Minutes
· Dacarbazine(DTIC) 10mg/ml, (10,000 ppm)>240 Minutes
· Daunorubicin(5mg/ml), (5,000 ppm)>240 Minutes
· Docetaxel(10mg/ml), (10,000 ppm)>240 Minutes
· Doxorubicin HCL(2mg/ml),(2,000 ppm)>240 Minutes
· Ellence (2mg/ml), (2,000 ppm)>240 Minutes
· Etoposide /Toposar (20mg/ml), (20,000 ppm)>240 Minutes
· Fludarabine (25mg/ml), (25,000 ppm)>240 Minutes
· Fluorouracil(50mg/ml), (50,000 ppm)>240 Minutes
· Gemcitabine (38mg/ml), (38,000 ppm)>240 Minutes
· Idarubicin(1.0mg/ml), (1,000 ppm)>240 Minutes
· Ifosfamide (50mg/ml), (50,000 ppm)>240 Minutes
· Irinotecan(20mg/ml), (20,000 ppm)>240 Minutes
· Mechlorethamine HCI(1.0mg/ml), (1,000 ppm)>240 Minutes
· Melphalan(5mg/ml), (5,000 ppm)>240 Minutes
· Methotrexate(25mg/ml),(25,000 ppm)>240 Minutes
· Mitomycin C(0.5 mg/ml), (500 ppm)>240 Minutes
· Mitoxantrone(2mg/ml),(2,000 ppm)>240 Minutes
· Paclitaxel(6.0mg/ml),(6,000 ppm)>240 Minutes
· Paraplatin (10mg/ml),(10,000 ppm)>240 Minutes
· Rituximab(10mg/ml),(10,000 ppm)>240 Minutes
· Thiotepa (10mg/ml),(10,000 ppm)15.9 Minutes
· Trisenox(0.1mg/ml), (100 ppm)>240 Minutes
· Vincristine Sulfate(1.0mg/ml),(1,000 ppm)>240 Minutes

Please note that Carmustine and Thiotepa have extremely low permeation times of 3,6 minutes, respectively

WARNING: Do Not Use With : Carmustine, ThioTEPA

Device Description

The subject device in this 510(k) Notification is a Purple Nitrile Examination glove tested for use with Chemotherapy drugs.

The subject device is a patient examination glove made from nitrile compound, Purple color, powder free and non-sterile (as per 21 CFR 880.6250, class I).

The principle operation of the medical device to provide single use barrier protection for the wearer and the device meets all the requirement specifications for Barrier Protection, tensile properties as defined in ASTM D6319-10, Standard specification for Nitrile Examination Gloves.

This device also complies with requirements for Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs as per ASTM D6978-05.

AI/ML Overview

The provided document pertains to a 510(k) premarket notification for "BRIGHTWAY BRAND NITRILE EXAMINATION GLOVES, POWDER FREE [PURPLE] TESTED FOR USE WITH CHEMOTHERAPY DRUGS." This is a Class I medical device, and the submission aims to demonstrate substantial equivalence to a legally marketed predicate device (K992162 Safeskin Purple Powder-Free Nitrile Examination Gloves for Use with Chemotherapeutic Drugs Labeling Claim).

The document does not describe an AI/ML-based device or a study involving human readers and AI assistance for medical image analysis. Instead, it details the physical and chemical properties of examination gloves, their resistance to chemotherapy drugs, and their biocompatibility, as per standard testing protocols for such devices.

Therefore, many of the requested elements for an AI/ML device study (e.g., sample size for AI test and training sets, number of experts for ground truth, MRMC studies, standalone algorithm performance, AI effect size) are not applicable to this document.

However, I can extract the acceptance criteria and performance data relevant to this specific device (examination gloves):

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are generally based on meeting established ASTM (American Society for Testing and Materials) and ISO (International Organization for Standardization) standards as outlined in the document, and demonstrating performance comparable to the predicate device. The reported performance indicates compliance with these standards.

Device CharacteristicsTest StandardAcceptance Criteria (General)Reported Device Performance
Product: Nitrile Examination Gloves, Powder Free (Purple) Tested For Use with Chemotherapy Drugs
Primary Function: Barrier protection to prevent contamination between patient and examiner. Tested for resistance to permeation by chemotherapy drugs.
Chemical Permeation:ASTM D6978-05 (Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs)Minimum breakthrough detection times for various chemotherapy drugs, typically >240 minutes for most tested drugs. Special note for drugs with low permeation times.Met or exceeded >240 minutes for most tested drugs. Carmustine (3.3mg/ml): 3.6 minutes; Thiotepa (10mg/ml): 15.9 minutes. (These values are declared and warned against, indicating they meet the disclosure requirement but are below the general >240 min goal, hence the warning). The device meets the requirement by truthfully reporting its performance for each chemical.
Physical Attributes (General):ASTM D6319-10 (Standard Specification for Nitrile Examination Gloves)Compliance with all requirements in ASTM D6319-10, including dimensions, physical properties (tensile strength, elongation).Complies with ASTM D6319-10 both before and after accelerated aging. All sizes comply with length, width, and thickness dimensions.
Freedom from Holes:ASTM D6319-10 by D5151-06Must meet specified AQL (Acceptable Quality Level) for freedom from holes.Pass (Implies meeting or exceeding the AQL).
Powder Free Designation:ASTM D6319-10 by D6124-06Less than 2.0 mg powder per glove.Less than 2.0 mg per glove; Pass.
Biocompatibility:ISO 10993-10:2010 (Reapproved 2014)Not a skin irritant; Not a skin sensitizer.Under conditions of the test, not a skin irritant; Under conditions of the test, not a skin sensitizer.
ISO 10993-11:2006 (Reapproved 2010)Does not cause systemic toxicity.Under conditions of the test, does not cause systemic toxicity.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document does not explicitly state the specific number of gloves (sample size) for each particular test. However, it references standards like ASTM D6319-10 and ISO 2859 (Sampling Procedures and Tables for Inspection by Attributes), which would dictate the appropriate sample sizes for the various tests (e.g., tensile strength, dimensions, pinholes, chemotherapy drug permeation). For example, ASTM D6978-05 for chemotherapy drug permeation typically involves testing a specified number of samples (often 3 replicates) per drug, which are then averaged or used to determine breakthrough time.
  • Data Provenance: The testing was "non-clinical performance data" (Section 7). The manufacturing company, Brightway Holdings Sdn. Bhd., is located in Malaysia (Section 3). The testing was conducted according to international standards (ASTM, ISO), meaning the data provenance is from laboratory tests, not from human retrospective or prospective clinical data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

  • Not Applicable. This is a physical device being tested against established ASTM and ISO standards in a laboratory setting. There is no concept of "ground truth" derived from human expert interpretation (like medical images) for this type of product. The "ground truth" is defined by the objective measurements and protocols outlined in the referenced standards.

4. Adjudication Method for the Test Set

  • Not Applicable. As this involves laboratory testing against objective standards (e.g., measuring breakthrough time, tensile strength), there is no human adjudication process involved in establishing "ground truth" or test results in the way it would be for interpreting medical images. The measurements are taken by lab technicians and reported against the standard's pass/fail criteria or quantitative thresholds.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable. This document is for examination gloves, not an AI/ML device. Therefore, no MRMC study with human readers or AI assistance was conducted.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Not Applicable. This is for examination gloves, not an AI/ML algorithm.

7. The Type of Ground Truth Used

  • Ground Truth: The "ground truth" for this device is defined by objective, standardized measurements and laboratory test results as per recognized ASTM and ISO international standards.
    • For chemotherapy drug permeation: Breakthrough time measured according to ASTM D6978-05.
    • For physical properties (tensile strength, elongation, dimensions, holes): Measured according to ASTM D6319-10 and its referenced sub-standards (e.g., ASTM D412, D5151).
    • For biocompatibility: Biological responses (irritation, sensitization, systemic toxicity) measured according to ISO 10993-10 and ISO 10993-11.

8. The Sample Size for the Training Set

  • Not Applicable. As this is not an AI/ML device, there are no training sets in the context of machine learning. The "training" for manufacturing occurs through process control and quality management systems to ensure consistent product performance.

9. How the Ground Truth for the Training Set was Established

  • Not Applicable. No AI/ML training set is involved.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol of medicine, with three human profiles incorporated into the design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 14, 2017

Brightway Holdings Sdn. Bhd. % Dr. Wava Truscott Consultant Truscott Medsci Associates, LLC 180 Burkemeade Ct Roswell. Georgia 30075

Re: K170686

Trade/Device Name: Nitrile Examination Gloves, Powder Free (Purple) Tested For Use with Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZA, LZC Dated: June 25, 2017 Received: June 26, 2017

Dear Dr. Truscott:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tara A. Ryan -S

for

Lori Wiggins Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

K170686

Device Name

BRIGHTWAY BRAND NITRILE EXAMINATION GLOVES, POWDER FREE [PURPLE] TESTED FOR USE WITH CHEMOTHERAPY DRUG5

Indications for Use (Describe)

BRIGHTWAY BRAND NITRILE EXAMINATION GLOVES, POWDER FREE [PURPLE] TESTED FOR USE WITH CHEMOTHERAPY DRUGS is a disposable device intended for medical purpose worn on the examiners hand or finger to prevent contamination between patient and examiner.

The tested chemotherapy drugs and their breakthrough detection times are as follows :

Test Chemotherapy Drug Name and ConcentrationMinimum Breakthrough Detection Time
· Blenoxane (15mg/ml),(15,000 ppm)>240 Minutes
· Busulfan(6mg/ml),(6,000 ppm)>240 Minutes
· Carmustine (BCNU) (3.3mg/ml),(3,300 ppm)3.6 Minutes
· Cisplatin(1.0mg/ml), (1.000 ppm)>240 Minutes
· Cyclophosphamide/Cytoxan(20mg/ml), (20,000 ppm)>240 Minutes
· Cytarabine(100mg/ml), (100,000 ppm)>240 Minutes
· Dacarbazine(DTIC) 10mg/ml, (10,000 ppm)>240 Minutes
· Daunorubicin(5mg/ml), (5,000 ppm)>240 Minutes
· Docetaxel(10mg/ml), (10,000 ppm)>240 Minutes
· Doxorubicin HCL(2mg/ml),(2,000 ppm)>240 Minutes
· Ellence (2mg/ml), (2,000 ppm)>240 Minutes
· Etoposide /Toposar (20mg/ml), (20,000 ppm)>240 Minutes
· Fludarabine (25mg/ml), (25,000 ppm)>240 Minutes
· Fluorouracil(50mg/ml), (50,000 ppm)>240 Minutes
· Gemcitabine (38mg/ml), (38,000 ppm)>240 Minutes
· Idarubicin(1.0mg/ml), (1,000 ppm)>240 Minutes
· Ifosfamide (50mg/ml), (50,000 ppm)>240 Minutes
· Irinotecan(20mg/ml), (20,000 ppm)>240 Minutes
· Mechlorethamine HCI(1.0mg/ml), (1,000 ppm)>240 Minutes
· Melphalan(5mg/ml), (5,000 ppm)>240 Minutes
· Methotrexate(25mg/ml),(25,000 ppm)>240 Minutes
· Mitomycin C(0.5 mg/ml), (500 ppm)>240 Minutes
· Mitoxantrone(2mg/ml),(2,000 ppm)>240 Minutes
· Paclitaxel(6.0mg/ml),(6,000 ppm)>240 Minutes
· Paraplatin (10mg/ml),(10,000 ppm)>240 Minutes
· Rituximab(10mg/ml),(10,000 ppm)>240 Minutes
· Thiotepa (10mg/ml),(10,000 ppm)15.9 Minutes
· Trisenox(0.1mg/ml), (100 ppm)>240 Minutes
· Vincristine Sulfate(1.0mg/ml),(1,000 ppm)>240 Minutes

Please note that Carmustine and Thiotepa have extremely low permeation times of 3,6 minutes, respectively

WARNING: Do Not Use With : Carmustine, ThioTEPA

Type of Use (Select one or both, as applicable)Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

FORM FDA 3881 (8/14)

Page 1 of 2

PSC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

K170686 510(K) SUMMARY

13

BRIGHTWAY BRAND NITRILE EXAMINATION GLOVES, POWDER FREE, [PURPLE] TESTED FOR USE WITH CHEMOTHERAPY DRUGS

    1. Submitter :
Company Name :BRIGHTWAY HOLDINGS SDN. BHD.
Street Address :Lot 1559, Jalan Istimewa,
Batu Belah, 42100 Klang
Selangor Darul Ehsan.
Country :Malaysia
Phone No. :603-3343 1007 & 603-3343 1094.
Fax No. :603-3341 4800
E-mail Address :brightway@brightway919.com
Contact Person :Mr. G. Baskaran (Group Managing Director)
baskar@brightway919.com
Mr. Felix Darrel (Group Marketing Manager)
felix.marketing@brightway919.com
  1. Preparation Date : July 13, 2017

  2. Name of the Device :

Device trade or proprietary name: BRIGHTWAY BRAND NITRILE EXAMINATION GLOVES, POWDER FREE[PURPLE] TESTED FOR USE WITH CHEMOTHERAPY DRUGS

{4}------------------------------------------------

Device Classification Name: Polymer Patient Examination Glove (21 CFR 88-6250)

Device common or usual name: Powder-Free Nitrile Examination Gloves Tested For Use With Chemotherapy Drugs

FDA Device Class : Class 1

Product Code : LZC, LZA

    1. ldentification of the Device :
      This Class 1 medical device; product code LZA, LZC with the Trade Name: BRIGHTWAY BRAND NITRILE EXAMINATION GLOVES, POWDER FREE (Purple) TESTED FOR USE WITH CHEMOTHERAPY DRUGS, was tested for Chemotherapy drug penetration resistance per ASTM D6978-05 and meets all the requirements of ASTM D6319-10 and FDA 21 CFR 880.6250.

Predicate device:

Legally marked device to which substantial equivalence is claimed:

    1. K992162 Safeskin Purple Powder-Free Nitrile Examination Gloves for Use with Chemotherapeutic Drugs Labeling Claim
    1. Device Description :

The subject device in this 510(k) Notification is a Purple Nitrile Examination glove tested for use with Chemotherapy drugs.

The subject device is a patient examination glove made from nitrile compound, Purple color, powder free and non-sterile (as per 21 CFR 880.6250, class I).

The principle operation of the medical device to provide single use barrier protection for the wearer and the device meets all the requirement specifications for Barrier Protection, tensile properties as defined in ASTM D6319-10, Standard specification for Nitrile Examination Gloves.

This device also complies with requirements for Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs as per ASTM D6978-05.

{5}------------------------------------------------

6. Intended use of the Device / Indication of Use

ੇ ਮੁੱਤਰ

BRIGHTWAY BRAND EXAMINATION GLOVES, POWDER FREE, [PURPLE] TESTED FOR USE WITH CHEMOTHERAPY DRUGS is a disposable device intended for medical purpose worn on the examiners hand or finger to prevent contamination between patient and examiner.

This device has been tested with chemotherapy drugs and their breakthrough detection times are as follows:

Test Chemotherapy Drug Name and Concentration

Test Chemotherapy Drug Name and ConcentrationMinimum Breakthrough Detection Time
Blenoxane (15mg/ml), (15,000 ppm)>240 Minutes
Busulfan(6mg/ml),(6,000 ppm)>240 Minutes
Carmustine (BCNU) (3.3mg/ml),(3,300 ppm)3.6 Minutes
Cisplatin(1.0mg/ml), (1,000 ppm)>240 Minutes
Cyclophosphamide/Cytoxan(20mg/ml), (20,000 ppm)>240 Minutes
Cytarabine(100mg/ml), (100,000 ppm)>240 Minutes
Dacarbazine (DTIC) 10mg/ml, (10,000 ppm)>240 Minutes
Daunorubicin(5mg/ml), (5,000 ppm)>240 Minutes
Docetaxel(10mg/ml), (10,000 ppm)>240 Minutes
Doxorubicin HCL(2mg/ml),(2,000 ppm)>240 Minutes
Ellence (2mg/ml), (2,000 ppm)>240 Minutes
Etoposide/Toposar (20mg/ml), (20,000 ppm)>240 Minutes
Fludarabine (25mg/ml), (25,000 ppm)>240 Minutes
Fluorouracil(50mg/ml), (50,000 ppm)>240 Minutes
Gemcitabine (38mg/ml), (38,000 ppm)>240 Minutes
Idarubicin (1.0mg/ml), (1,000 ppm)>240 Minutes
Ifosfamide (50mg/ml), (50,000 ppm)>240 Minutes
Irinotecan(20mg/ml), (20,000 ppm)>240 Minutes
Mechlorethamine HCI(1.0mg/ml), (1,000 ppm)>240 Minutes
Melphalan(5mg/ml), (5,000 ppm)>240 Minutes
Methotrexate(25mg/ml), (25,000 ppm)>240 Minutes
Mitomycin C(0.5 mg/ml), (500 ppm)>240 Minutes
Mitoxantrone(2mg/ml), (2,000 ppm)>240 Minutes
Paclitaxel(6.0mg/ml),(6,000 ppm)>240 Minutes
Paraplatin (10mg/ml), (10,000 ppm)>240 Minutes
Rituximab(10mg/ml), (10,000 ppm)>240 Minutes
Thiotepa (10mg/ml), (10,000 ppm)15.9 Minutes
Trisenox(0.1mg/ml), (100 ppm)>240 Minutes
Vincristine Sulfate(1.0mg/ml), (1,000 ppm)>240 Minutes

Please Note: Carmustine and ThioTEPA have extremely low permeation times with breakthrough detected in less than 30 minutes: Carmustine (3.3mg/mL): 3.6 minutes; ThioTEPA (10mg/mL): 15.9 minutes WARNING: Not for Use with: Carmustine, ThioTEPA

{6}------------------------------------------------

7. Substantial Equivalence Based on Assessment of Non-Clinical Performance Data

a.) Testing was performed per ASTM D6319: Standard Specification for Nitrile Examination Gloves for Medical Application, utilizing the currently re-approved versions of the test methods there-in referenced:

  • ASTM D412 Test Methods for Vulcanized Rubber and Thermoplastic Elastomers— ● Tension
  • ASTM D573 Test Method for Rubber—Deterioration in an Air Oven
  • ASTM D3578 Specification for Rubber Examination Gloves
  • ASTM D3767 Practice for Rubber—Measurement of Dimensions
  • o ASTM D5151 Test Method for Detection of Holes in Medical Gloves
  • ASTM D6124 Test Method for Residual Powder on Medical Gloves
  • ISO 2859 Sampling Procedures and Tables for Inspection by Attributes

Test results show that under the conditions of the testing, there is no difference in physical attributes between the proposed device and the predicate device.

b.) Both the proposed device and the predicate were tested for use with chemotherapy drugs per:

  • o ASTM D6978-05 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs
    c.) Biocompatibility testing was performed utilizing:

  • . ISO 10993-10: Biological evaluation of medical devices – Part 10: Tests for Irritation and Sensitization. Both dermal irritation and sensitization (Magnusson & Kingman) were performed.

  • . ISO 10993-11: Biological evaluation of medical devices – Part 11: Tests for Systemic toxicity were conducted as had been performed on the predicate device.

Both the proposed device and the predicate device demonstrated that they were nonirritating and non-sensitizing, and that they did not demonstrate systemic toxicity under conditions of the studies performed.

All testing performed demonstrated that under the conditions of the test or study conducted, that the proposed medical device and the predicate device performed the same or similarly.

8. A summary, side by side Comparison Table is presented:

{7}------------------------------------------------

510(k) Summary Table

Comparison of Proposed Device to Predicate Device (pg. 1 of 2)

DeviceCharacteristicsTestStandardProposed DevicePredicate DeviceK992162ComparisonAnalysis
Trade NameN/ABrightway Brand NitrileExamination Gloves, PowderFree, [Purple] Tested for Usewith Chemotherapy DrugsSafeskin Purple Powder-FreeNitrile Examination Gloves Testedfor use with Chemotherapy DrugsSimilar
510(k)ReferenceN/AK992162N/A
Common NameN/APatient Examination GlovePatient Examination GloveSame
Product CodeN/ALZA, LZCLZA, LZCSame
Intended useN/AA powder-free patientexamination glove is adisposable device intended formedical purposes that is wornon the examiners hand or fingerto prevent contaminationbetween patient and examiner.Additionally, the gloves weretested for use withchemotherapy drugs inaccordance with ASTM D6978-05 Standard Practice forAssessment of Medical Glove toPermeation by chemotherapyDrugs.A powder-free patientexamination glove is adisposable device intended formedical purposes that is wornon the examiners hand or fingerto prevent contaminationbetween patient and examiner.Additionally, the gloves weretested for use withchemotherapy drugs inaccordance with ASTM D6978-05 Standard Practice forAssessment of Medical Glove toPermeation by chemotherapyDrugs.Same
LabelingN/AThere are no special labelingclaims. Does not claimhypoallergenic on labelThere are no special labelingclaims. Does not claimhypoallergenic on labelSame
RegulationNumberN/A21 CFR 880.625021 CFR 880.6250Same
MaterialsN/ANitrileNitrileSame
ColorN/APurplePurpleSame
Textures fingersN/AYesYesSame
DesignConfigurationsASTMD6319-10Extra-SmallSmallMediumLargeExtra-LargeExtra-SmallSmallMediumLargeExtra-LargeSame
DimensionsLengthASTMD6319-10All sizes comply with lengthdimensionsAll sizes comply with lengthdimensionsSame
DimensionsWidthASTMD6319-10All sizes comply with widthdimensionsAll sizes comply with widthdimensionsSame
DimensionsThicknessASTMD6319-10All sizes comply with thicknessdimensionsAll sizes comply with thicknessdimensionsSame
DeviceCharacteristicsTestStandardProposed DevicePredicate DeviceK992162ComparisonAnalysis
PhysicalProperties -Tensile StrengthASTMD6319-10Complies with ASTMD6319-10 both before andafter accelerated agingComplies with ASTMD6319-10 both before andafter accelerated agingSame
PhysicalProperties -ElongationASTMD6319-10Complies with ASTMD6319-10 both before andafter accelerated agingComplies with ASTMD6319-10 both before andafter accelerated agingSame
Freedom fromHolesASTMD6319-10by D5151-06PassPassSame
Powder FreeDesignationASTMD6319-10byD6124-06Less than 2.0 mg per glove;PassLess than 2.0 mg per glove;PassSame
BiocompatibilityISO 10993-10:2010Reapproved2014Under conditions of the test,Not a skin irritantUnder conditions of the testNot a skin irritantSame
ISO 10993-10:2010Reapproved2014Under conditions of the test,Not a skin sensitizerUnder conditions of the test,Not a skin sensitizerSame
ISO 10993-11:2006Reapproved2010Under conditions of the testDoes not causesystemic toxicityUnder conditions of the testDoes not causesystemic toxicitySame
Prescription vs.ОТСN/AОТСОТСSame
Sterile vs.Non-sterileN/ANon-SterileNon-SterileSame
Single UseN/AYesYesSame
Tested for usewithChemotherapyDrugsASTMD6978-05Please note: Carmustine andThioTEPA have extremely lowbreakthrough times of lessthan 30 minutesPlease note: Carmustine andThioTEPA have extremely lowbreakthrough times of lessthan 30 minutesSame
WarningN/AWARNING: Not for Use with:Carmustine, ThioTEPAWARNING: Not for Use with:Carmustine, ThioTEPASame

{8}------------------------------------------------

Part 10. Executive Summary

Comparison of Proposed Device to Predicate Device (pg. 2 of 2)

{9}------------------------------------------------

9.0 Conclusion

Based on intended uses, technological characteristics and non-clinical performance data, the subject device BRIGHTWAY BRAND NITRILE EXAMINATION GLOVES, POWDER FREE, [PURPLE] TESTED FOR USE WITH CHEMOTHERAPY DRUGS is substantially equivalent to the predicate device K992162 - Safeskin Purple Powder-Free Nitrile Examination Gloves for Use with Chemotherapeutic Drugs Labeling Claim.

The subject device meets the requirements of ASTM D 6319- 10 standards as well as applicable 21 CFR and meets FDA recognized, physical properties requirements, pinhole requirements, powder free requirements, resistance to permeation by chemotherapy drugs testing, biocompatibility studies and labeling requirements presented in this document.

There are no safety or efficacy issues or new claims different from the "substantially equivalent" predicate.

Based on the complete list of non-clinical tests, biological safety studies, labeling, intended use, materials, and processes of manufacture, the subject device herein proposed is as safe, as effective, and performs as well as the legally marketed predicate device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.