(56 days)
The device is intended for use by a qualified physician for ultrasound evaluation of Fetal/Obstetrics; Abdominal; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transvaginal; Transesophageal and Intraoperative (abdominal, thoracic, vascular).
The LOGIQ S8 is a full featured, general purpose diagnostic ultrasound system which consists of a mobile console approximately 520 mm wide (keyboard), 865 mm deep and 1760 mm high that provides digital acquisition, processing and display capability. The user interface includes a computer keyboard, specialized controls, 10.1-inch LCD touch screen and color widescreen monitor.
The provided text describes a 510(k) premarket notification for the GE LOGIQ S8 ultrasound system. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving novel claims or establishing new acceptance criteria through extensive clinical studies.
Therefore, the document does not provide specific acceptance criteria in terms of analytical or clinical performance metrics (e.g., sensitivity, specificity, accuracy for a particular disease or measurement), nor does it detail a study performed to prove such metrics against acceptance criteria. Instead, it demonstrates substantial equivalence based on technological characteristics, intended use, and compliance with recognized safety and performance standards.
Here's an breakdown of the information that can be extracted based on your request, as well as indications of what is not present in this type of regulatory filing:
1. A table of acceptance criteria and the reported device performance
This information is not provided in the document as it is a 510(k) summary for showing substantial equivalence, not a clinical trial report establishing new performance criteria. The "performance" described relates primarily to equivalence with predicate devices and compliance with general safety and performance standards.
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not provided. The document explicitly states: "The subject of this premarket submission, LOGIQ S8, did not require clinical studies to support substantial equivalence." The testing mentioned is for acoustic output, biocompatibility, cleaning/disinfection, and safety standards.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not provided as no clinical studies are reported to have been conducted for this submission.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not provided as no clinical studies are reported to have been conducted for this submission.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not provided. The device is an ultrasound imaging system, and there's no mention of AI assistance or MRMC studies in this document.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This information is not provided. The device is a diagnostic ultrasound imaging system, not an AI algorithm for standalone interpretation.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
This information is not provided as no clinical studies are reported to have been conducted for this submission. The "ground truth" for the device's claims are the safety and performance characteristics of its predicate devices and compliance with recognized standards.
8. The sample size for the training set
This information is not provided as no machine learning or AI training is mentioned as being part of this submission's substantial equivalence argument.
9. How the ground truth for the training set was established
This information is not provided as no machine learning or AI training is mentioned.
Summary of what the document does provide regarding device performance and compliance:
The document focuses on demonstrating that the GE LOGIQ S8 is substantially equivalent to legally marketed predicate devices. This is achieved by:
- Intended Use Equivalence: Stating that the LOGIQ S8 has the same clinical intended use as the predicate LOGIQ S8 (K152195).
- Technological Equivalence: Declaring that it employs the same fundamental scientific technology as its predicate device(s).
- Imaging Modes: Noting that it has the same imaging modes as the predicate LOGIQ S8, with additions of Fibroscan elastography and CAP (Controlled Attenuation Parameter), which are equivalent to features found on the reference predicate FibroScan (K160524).
- Safety and Standards Compliance:
- Manufactured with materials evaluated and found safe.
- Acoustic power levels are below applicable FDA limits.
- Designed in compliance with approved electrical and physical safety standards.
- Compliance with voluntary standards:
- AAMI/ANSI ES60601-1 (Medical Electrical Equipment - General Requirements for Safety)
- IEC60601-1-2 (Electromagnetic Compatibility)
- IEC60601-2-37 (Ultrasonic Medical Diagnostic and Monitoring Equipment Safety)
- ISO10993-1 (Biological Evaluation of Medical Devices)
- ISO14971 (Application of risk management to medical devices)
- NEMA UD 2 (Acoustic Output Measurement Standard)
- NEMA UD 3 (Real Time Display of Thermal and Mechanical Acoustic Output Indices)
- NEMA, Digital Imaging and Communications in Medicine (DICOM) Set.
- Quality assurance measures applied to development (Risk analysis, Requirements Reviews, Design Reviews, Testing on unit level, Integration testing, Final Acceptance Testing, Performance testing, Safety testing).
- Transducer and patient contact materials are biocompatible.
The document lists various transducers cleared with the system and for which clinical applications and modes of operation they are "previously cleared" (P), "new" (N), or "added under Appendix E" (E), primarily based on equivalence to other GE Ultrasound systems (e.g., LOGIQ E9, VIVID E90/E95).
In conclusion, for this specific 510(k) submission, the "acceptance criteria" are compliance with established medical device safety and performance standards and demonstration of substantial equivalence to predicate devices, rather than predefined performance thresholds from new clinical trials.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is a stylized symbol consisting of three overlapping human profiles facing to the right, with flowing lines above them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 11, 2017
GE Healthcare % Mr. Bryan Behn Regulatory Affairs Director 9900 Innovation Drive WAUWATOSA WI 53226
Re: K170445 Trade/Device Name: LOGIO S8 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: February 13, 2017 Received: February 14, 2017
Dear Mr. Behn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael O'Hara
For
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K170445
Device Name LOGIO 88
Indications for Use (Describe)
The device is intended for use by a qualified physician for ultrasound evaluation of Fetal/Obstetrics; Abdominal; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transvaginal; Transesophageal and Intraoperative (abdominal, thoracic, vascular).
Type of Use (Select one or both, as applicable)
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a blue circle. The circle has a decorative border with swirling patterns, giving it a classic and recognizable appearance.
GE Healthcare 510(k) Premarket Notification Submission
Indications for Use Forms
The following forms represent indications with clinical applications and exam types along with the modes of operation for the LOGIQ S8 system and for all of its probe/mode combinations. Combinations identified by "P" represent those previously cleared with the unmodified LOGIQ S8. "N" is new in this submission and "E" have been added by Appendix E of the FDA Ultrasound Guidance.
The subject modification does not alter the previously cleared system level indications and clinical applications.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized, cursive font. The letters are enclosed within a blue circle, and there are decorative white swirls around the perimeter of the circle.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S8 Ultrasound System
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes* | HarmonicImaging | CodedPulse | Other[Notes] |
| Ophthalmic | |||||||||||
| Fetal / Obstetrics[7] | P | P | P | P | P | P | P | P | P | P | 5,6,9 |
| Abdominal[1] | P | P | P | P | P | P | P | P | P | P | 3,5,6,9,10,11,12, 13 |
| Pediatric | P | P | P | P | P | P | P | P | P | P | 3,5,6,9,10, 11, 13 |
| Small Organ[2] | P | P | P | P | P | P | P | P | P | P | 3,5,6,9,10 |
| Neonatal Cephalic | P | P | P | P | P | P | P | P | P | P | 3,5,6,9 |
| Adult Cephalic | P | P | P | P | P | P | P | P | P | P | 3,5,6,9 |
| Cardiac | P | P | P | P | P | P | P | P | P | P | |
| Peripheral Vascular | P | P | P | P | P | P | P | P | P | P | 3,5,6,9,10 |
| Musculo-skeletal Conventional | P | P | P | P | P | P | P | P | P | P | 3,5,6,9,10 |
| Musculo-skeletal Superficial | P | P | P | P | P | P | P | P | P | P | 3,5,6,9,10 |
| Other[4] | P | P | P | P | P | P | P | P | P | P | 3,5,6,9,10 |
| Exam Type, Means of Access | |||||||||||
| Transesophageal | P | P | P | P | P | P | P | P | P | P | |
| Transrectal | P | P | P | P | P | P | P | P | P | P | 3,5,6,9 |
| Transvaginal | P | P | P | P | P | P | P | P | P | P | 3,5,6,7,9 |
| Transuretheral | |||||||||||
| Intraoperative[8] | P | P | P | P | P | P | P | P | P | P | 3,5,6,9 |
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[13] A Mode
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular border. The border has decorative swirls or flourishes around the letters. The logo is presented in a blue color.
GE LOGIQ S8 with C1-5-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes* | HarmonicImaging | CodedPulse | Other[Notes] |
| Ophthalmic | |||||||||||
| Fetal / Obstetrics[7] | P | P | P | P | P | P | P | P | P | P | 5,6,9 |
| Abdominal[1] | P | P | P | P | P | P | P | P | P | P | 3,5,6,9 |
| Pediatric | P | P | P | P | P | P | P | P | P | P | 3,5,6,9 |
| Small Organ[2] | |||||||||||
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | |||||||||||
| Cardiac | |||||||||||
| Peripheral Vascular | P | P | P | P | P | P | P | P | P | P | 3,5,6,9 |
| Musculo-skeletal Conventional | |||||||||||
| Musculo-skeletal Superficial | |||||||||||
| Other[4] | P | P | P | P | P | P | P | P | P | P | 3,5,6,9 |
| Exam Type, Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal | |||||||||||
| Transvaginal | |||||||||||
| Transuretheral | |||||||||||
| Intraoperative[8] | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
[1] Abdominal includes Renal, GYN/Pelvic. Notes:
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaqing mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized, cursive font. The letters are enclosed within a circular border, and there are decorative swirls or flourishes surrounding the circle, giving it a classic and recognizable appearance.
GE LOGIQ S8 with 9L-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes* | HarmonicImaging | CodedPulse | Other(Notes) |
| Ophthalmic | |||||||||||
| Fetal / Obstetrics[7] | P | P | P | P | P | P | P | P | 5,6,9 | ||
| Abdominal[1] | P | P | P | P | P | P | P | P | P | 3,5,6,9,10 | |
| Pediatric | P | P | P | P | P | P | P | P | P | 3,5,6,9,10 | |
| Small Organ[2] | P | P | P | P | P | P | P | P | P | 3,5,6,9,10 | |
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | |||||||||||
| Cardiac | |||||||||||
| Peripheral Vascular | P | P | P | P | P | P | P | P | P | 3,5,6,9,10 | |
| Musculo-skeletal Conventional | P | P | P | P | P | P | P | P | P | 3,5,6,9,10 | |
| Musculo-skeletal Superficial | P | P | P | P | P | P | P | P | P | 3,5,6,9,10 | |
| Other[4] | |||||||||||
| Exam Type, Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal[8] | |||||||||||
| Transvaginal | |||||||||||
| Transuretheral | |||||||||||
| Intraoperative[8] | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
[1] Abdominal includes Renal, GYN/Pelvic. Notes:
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular border. The color of the logo is a light blue. The logo is simple and recognizable, representing the General Electric brand.
GE LOGIQ S8 with ML6-15-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes* | HarmonicImaging | CodedPulse | Other[Notes] |
|---|---|---|---|---|---|---|---|---|---|---|---|
| Ophthalmic | |||||||||||
| Fetal / Obstetrics[7] | |||||||||||
| Abdominal[1] | |||||||||||
| Pediatric | P | P | P | P | P | P | P | P | P | 3,5,6,9 | |
| Small Organ[2] | P | P | P | P | P | P | P | P | P | 3,5,6,9 | |
| Neonatal Cephalic | P | P | P | P | P | P | P | P | P | 9 | |
| Adult Cephalic | |||||||||||
| Cardiac | |||||||||||
| Peripheral Vascular | P | P | P | P | P | P | P | P | P | 3,5,6,9 | |
| Musculo-skeletal Conventional | P | P | P | P | P | P | P | P | P | 3,5,6,9 | |
| Musculo-skeletal Superficial | P | P | P | P | P | P | P | P | P | 3,5,6,9 | |
| Other[4] | |||||||||||
| Exam Type, Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal[8] | |||||||||||
| Transvaginal | |||||||||||
| Transuretheral | |||||||||||
| Intraoperative[8] | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized, cursive font. The letters are enclosed within a blue circle, and there are decorative white swirls around the perimeter of the circle.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S8 with M5S-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical Application | B | M | PW | CW | Color | Color M | Power | Combined | Harmonic | Coded | Other | |
| Anatomy/Region of Interest | Doppler | Doppler | Doppler | Doppler | Doppler | Modes | Imaging | Pulse | [Notes] | |||
| Ophthalmic | ||||||||||||
| Fetal / Obstetrics[7] | ||||||||||||
| Abdominal[1] | P | P | P | P | P | P | P | P | P | P | 5,9 | |
| Pediatric | ||||||||||||
| Small Organ[2] | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | P | P | P | P | P | P | P | P | P | P | 5,9 | |
| Cardiac | P | P | P | P | P | P | P | P | P | P | ||
| Peripheral Vascular | ||||||||||||
| Musculo-skeletal Conventional | ||||||||||||
| Musculo-skeletal Superficial | ||||||||||||
| Other[4] | ||||||||||||
| Exam Type, Means of Access | ||||||||||||
| Transesophageal | ||||||||||||
| Transrectal[8] | ||||||||||||
| Transvaginal | ||||||||||||
| Transuretheral | ||||||||||||
| Intraoperative[8] | ||||||||||||
| Intraoperative Neurological | ||||||||||||
| Intravascular | ||||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
[1] Abdominal includes Renal, GYN/Pelvic. Notes:
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular border. The border has decorative swirls or flourishes around the letters. The logo is presented in a blue color.
GE LOGIQ S8 with IC5-9-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes* | HarmonicImaging | CodedPulse | Other[Notes] |
| Ophthalmic | |||||||||||
| Fetal / Obstetrics[7] | P | P | P | P | P | P | P | P | P | 5,6,9 | |
| Abdominal[1] | P | P | P | P | P | P | P | P | P | 3,5,9 | |
| Pediatric | |||||||||||
| Small Organ[2] | |||||||||||
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | |||||||||||
| Cardiac | |||||||||||
| Peripheral Vascular | |||||||||||
| Musculo-skeletal Conventional | |||||||||||
| Musculo-skeletal Superficial | |||||||||||
| Other[4] | P | P | P | P | P | P | P | P | P | 3,5,6,9 | |
| Exam Type, Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal[8] | P | P | P | P | P | P | P | P | P | 3,5,6,9 | |
| Transvaginal | P | P | P | P | P | P | P | P | P | 3,5,6,7,9 | |
| Transuretheral | |||||||||||
| Intraoperative[8] | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular border. The border has decorative swirls or flourishes around the letters. The logo is presented in a blue color.
GE LOGIQ S8 with 3CRF-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes* | HarmonicImaging | CodedPulse | Other[Notes] |
| Ophthalmic | |||||||||||
| Fetal / Obstetrics[7] | |||||||||||
| Abdominal[1] | P | P | P | P | P | P | P | P | P | 5,6,9 | |
| Pediatric | P | P | P | P | P | P | P | P | P | 5,6,9 | |
| Small Organ[2] | |||||||||||
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | |||||||||||
| Cardiac | |||||||||||
| Peripheral Vascular | |||||||||||
| Musculo-skeletal Conventional | |||||||||||
| Musculo-skeletal Superficial | |||||||||||
| Other[4] | |||||||||||
| Exam Type, Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal[8] | |||||||||||
| Transvaginal | |||||||||||
| Transuretheral | |||||||||||
| Intraoperative[8] | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular border. The color of the logo is a light blue. The logo is simple and recognizable, representing the General Electric brand.
GE LOGIQ S8 with L8-18i-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes* | HarmonicImaging | CodedPulse | Other[Notes) | |
| Ophthalmic | ||||||||||||
| Fetal / Obstetrics[7] | ||||||||||||
| Abdominal[1] | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ[2] | P | P | P | P | P | P | P | P | P | 5,9 | ||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Cardiac | ||||||||||||
| Peripheral Vascular | P | P | P | P | P | P | P | P | P | 5,9 | ||
| Musculo-skeletal Conventional | P | P | P | P | P | P | P | P | P | 5,9 | ||
| Musculo-skeletal Superficial | P | P | P | P | P | P | P | P | P | 5,9 | ||
| Other[4] | ||||||||||||
| Exam Type, Means of Access | ||||||||||||
| Transesophageal | ||||||||||||
| Transrectal[8] | ||||||||||||
| Transvaginal | ||||||||||||
| Transuretheral | ||||||||||||
| Intraoperative[8] | P | P | P | P | P | P | P | P | P | 5,9 | ||
| Intraoperative Neurological | ||||||||||||
| Intravascular | ||||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
[1] Abdominal includes Renal, GYN/Pelvic. Notes:
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaqing mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular border. The color of the logo is a light blue. The logo is simple and recognizable, representing the General Electric brand.
GE LOGIQ S8 with S4-10-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes* | HarmonicImaging | CodedPulse | Other[Notes] | |
| Ophthalmic | ||||||||||||
| Fetal / Obstetrics[7] | ||||||||||||
| Abdominal[1] | ||||||||||||
| Pediatric | P | P | P | P | P | P | P | P | P | P | 5,9 | |
| Small Organ[2] | ||||||||||||
| Neonatal Cephalic | P | P | P | P | P | P | P | P | P | P | 5,9 | |
| Adult Cephalic | ||||||||||||
| Cardiac | P | P | P | P | P | P | P | P | P | P | ||
| Peripheral Vascular | ||||||||||||
| Musculo-skeletal Conventional | ||||||||||||
| Musculo-skeletal Superficial | ||||||||||||
| Other[4] | ||||||||||||
| Exam Type, Means of Access | ||||||||||||
| Transesophageal | ||||||||||||
| Transrectal[8] | ||||||||||||
| Transvaginal | ||||||||||||
| Transuretheral | ||||||||||||
| Intraoperative[8] | ||||||||||||
| Intraoperative Neurological | ||||||||||||
| Intravascular | ||||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular border. The color of the logo is a light blue. The logo is simple and recognizable, representing the General Electric brand.
GE LOGIQ S8 with P2D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes* | HarmonicImaging | CodedPulse | Other[Notes] |
| Ophthalmic | |||||||||||
| Fetal / Obstetrics[7] | |||||||||||
| Abdominal[1] | |||||||||||
| Pediatric | |||||||||||
| Small Organ[2] | |||||||||||
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | P | ||||||||||
| Cardiac | P | ||||||||||
| Peripheral Vascular | P | ||||||||||
| Musculo-skeletal Conventional | |||||||||||
| Musculo-skeletal Superficial | |||||||||||
| Other[4] | |||||||||||
| Exam Type, Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal[8] | |||||||||||
| Transvaginal | |||||||||||
| Transuretheral | |||||||||||
| Intraoperative[8] | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
[1] Abdominal includes Renal, GYN/Pelvic. Notes:
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaqing mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular border. The color of the logo is a light blue. The logo is simple and recognizable, representing the General Electric brand.
GE LOGIQ S8 with P6D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes* | HarmonicImaging | CodedPulse | Other[Notes] |
| Ophthalmic | |||||||||||
| Fetal / Obstetrics[7] | |||||||||||
| Abdominal[1] | |||||||||||
| Pediatric | |||||||||||
| Small Organ[2] | |||||||||||
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | P | ||||||||||
| Cardiac | P | ||||||||||
| Peripheral Vascular | P | ||||||||||
| Musculo-skeletal Conventional | |||||||||||
| Musculo-skeletal Superficial | |||||||||||
| Other[4] | |||||||||||
| Exam Type, Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal[8] | |||||||||||
| Transvaginal | |||||||||||
| Transuretheral | |||||||||||
| Intraoperative[8] | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
[1] Abdominal includes Renal, GYN/Pelvic. Notes:
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaqing mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular border. The color of the logo is a light blue. The logo is simple and recognizable, representing the General Electric brand.
GE LOGIQ S8 with C2-9-D Transducer
Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes* | HarmonicImaging | CodedPulse® | Other[Notes] |
| Ophthalmic | |||||||||||
| Fetal/Obstetrics[7] | P | P | P | P | P | P | P | P | P | [5,6,9] | |
| Abdominal[1] | P | P | P | P | P | P | P | P | P | [3,5,6,9] | |
| Pediatric | P | P | P | P | P | P | P | P | P | [3,5,6,9] | |
| Small Organ[2] | |||||||||||
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | |||||||||||
| Cardiac Adult | |||||||||||
| Cardiac Pediatric | |||||||||||
| Peripheral Vascular | P | P | P | P | P | P | P | P | P | [3,5,6,9] | |
| Musculo-skeletal Conventional | |||||||||||
| Musculo-skeletal Superficial | |||||||||||
| Other[4] | |||||||||||
| Exam Type, Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal | |||||||||||
| Transvaginal | |||||||||||
| Transurethral | |||||||||||
| Intraoperative[8] | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
[1] Abdominal includes Renal, GYN/Pelvic. Notes:
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle quidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
System provides real-time 3D and 4D acquisition when used with special 4D probes
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a blue circle. There are small, curved lines around the outer edge of the circle, giving it a textured appearance.
GE LOGIQ S8 with 11L-D Transducer
Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Anatomy/Region of Interest | B | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes* | Harmonic Imaging | Coded Pulse® | Other [Notes] |
| Ophthalmic | |||||||||||
| Fetal/Obstetrics[7] | |||||||||||
| Abdominal[1] | |||||||||||
| Pediatric | P | P | P | P | P | P | P | P | P | [3,5,6] | |
| Small Organ[2] | P | P | P | P | P | P | P | P | P | [3,5,6] | |
| Neonatal Cephalic | P | P | P | P | P | P | P | P | P | [5,6] | |
| Adult Cephalic | |||||||||||
| Cardiac Adult | |||||||||||
| Cardiac Pediatric | |||||||||||
| Peripheral Vascular | P | P | P | P | P | P | P | P | P | [3,5,6] | |
| Musculo-skeletal Conventional | P | P | P | P | P | P | P | P | P | [3,5,6] | |
| Musculo-skeletal Superficial | P | P | P | P | P | P | P | P | P | [3,5,6] | |
| Other[4] | |||||||||||
| Exam Type, Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal | |||||||||||
| Transvaginal | |||||||||||
| Transurethral | |||||||||||
| Intraoperative[8] | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
[1] Abdominal includes Renal, GYN/Pelvic. Notes:
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle quidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
System provides real-time 3D and 4D acquisition when used with special 4D probes
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{17}------------------------------------------------
Image /page/17/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined within a blue circle. The letters are stylized with curved lines, and the circle has decorative elements around its perimeter.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S8 with 6S-D Transducer
Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical Application | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes* | HarmonicImaging | CodedPulse♦ | Other[Notes] | |
| Anatomy/Region of Interest | ||||||||||||
| Ophthalmic | ||||||||||||
| Fetal/Obstetrics[7] | P | P | P | P | P | P | P | P | P | P | 5 | |
| Abdominal[1] | ||||||||||||
| Pediatric | P | P | P | P | P | P | P | P | P | P | 5 | |
| Small Organ[2] | ||||||||||||
| Neonatal Cephalic | P | P | P | P | P | P | P | P | P | P | 5 | |
| Adult Cephalic | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | P | P | P | P | P | P | P | P | P | P | 5 | |
| Peripheral Vascular | ||||||||||||
| Musculo-skeletal Conventional | ||||||||||||
| Musculo-skeletal Superficial | ||||||||||||
| Other[4] | ||||||||||||
| Exam Type, Means of Access | ||||||||||||
| Transesophageal | ||||||||||||
| Transrectal | ||||||||||||
| Transvaginal | ||||||||||||
| Transurethral | ||||||||||||
| Intraoperative[8] | ||||||||||||
| Intraoperative Neurological | ||||||||||||
| Intravascular | ||||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle quidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
System provides real-time 3D and 4D acquisition when used with special 4D probes
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{18}------------------------------------------------
Image /page/18/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. The circle has a decorative, swirling pattern around the letters, giving it a classic and recognizable appearance.
GE LOGIQ S8 with S1-5-D Transducer
Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes* | HarmonicImaging | CodedPulse® | Other[Notes] | |
| Ophthalmic | ||||||||||||
| Fetal/Obstetrics[7] | P | P | P | P | P | P | P | P | P | P | [5,6,9] | |
| Abdominal[1] | P | P | P | P | P | P | P | P | P | P | [5,6,9] | |
| Pediatric | ||||||||||||
| Small Organ[2] | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Peripheral Vascular | P | P | P | P | P | P | P | P | P | P | [5,6,9] | |
| Musculo-skeletal Conventional | ||||||||||||
| Musculo-skeletal Superficial | ||||||||||||
| Other[4] | ||||||||||||
| Exam Type, Means of Access | ||||||||||||
| Transesophageal | ||||||||||||
| Transrectal | ||||||||||||
| Transvaginal | ||||||||||||
| Transurethral | ||||||||||||
| Intraoperative[8] | ||||||||||||
| Intraoperative Neurological | ||||||||||||
| Intravascular | ||||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
[1] Abdominal includes Renal, GYN/Pelvic. Notes:
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle quidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
System provides real-time 3D and 4D acquisition when used with special 4D probes
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{19}------------------------------------------------
Image /page/19/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. The circle has a decorative, swirling pattern around the letters, giving it a classic and recognizable appearance.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S8 with RAB6-D Transducer
Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | Other[Notes] | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Anatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes* | HarmonicImaging | CodedPulse | ||
| Ophthalmic | ||||||||||||
| Fetal/Obstetrics | P | P | P | P | P | P | P | P | P | [5,6] | ||
| Abdominal | P | P | P | P | P | P | P | P | P | [5,6] | ||
| Pediatric | P | P | P | P | P | P | P | P | P | [5,6] | ||
| Small Organ | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Peripheral Vascular | ||||||||||||
| Musculo-skeletal Conventional | ||||||||||||
| Musculo-skeletal Superficial | ||||||||||||
| Other | ||||||||||||
| Exam Type, Means of Access | ||||||||||||
| Transesophageal | ||||||||||||
| Transrectal | ||||||||||||
| Transvaginal | ||||||||||||
| Transurethral | ||||||||||||
| Intraoperative | ||||||||||||
| Intraoperative Neurological | ||||||||||||
| Intravascular | ||||||||||||
| Laparoscopic |
N = new indication; P* = previously cleared by FDA; E = added under Appendix E
[1] Abdominal includes Renal, GYN/Pelvic. Notes:
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle quidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
System provides real-time 3D and 4D acquisition when used with special 4D probes
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{20}------------------------------------------------
Image /page/20/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a blue circle. There are small white accents around the perimeter of the circle, giving it a slightly three-dimensional appearance.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S8 with RIC5-9-D Transducer
Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical Application | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes* | HarmonicImaging | CodedPulse° | Other[Notes] | |
| Anatomy/Region of Interest | ||||||||||||
| Ophthalmic | ||||||||||||
| Fetal/Obstetrics[7] | P | P | P | P | P | P | P | P | P | P | [5,6] | |
| Abdominal[1] | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ[2] | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Peripheral Vascular | ||||||||||||
| Musculo-skeletal Conventional | ||||||||||||
| Musculo-skeletal Superficial | ||||||||||||
| Other[4] | P | P | P | P | P | P | P | P | P | P | [3,5,6] | |
| Exam Type, Means of Access | ||||||||||||
| Transesophageal | ||||||||||||
| Transrectal | P | P | P | P | P | P | P | P | P | P | [3,5,6] | |
| Transvaginal | P | P | P | P | P | P | P | P | P | P | [3,5,6] | |
| Transurethral | ||||||||||||
| Intraoperative[8] | ||||||||||||
| Intraoperative Neurological | ||||||||||||
| Intravascular | ||||||||||||
| Laparoscopic |
N = new indication; P* = previously cleared by FDA; E = added under Appendix E
[1] Abdominal includes Renal, GYN/Pelvic. Notes:
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle quidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
System provides real-time 3D and 4D acquisition when used with special 4D probes
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{21}------------------------------------------------
Image /page/21/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a blue circle. The circle has decorative swirls around its perimeter, giving it a classic and recognizable appearance.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S8 with 6Tc-RS Transducer
Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical Application | B | M | PW | CW | Color | Color M | Power | Combined | Harmonic | Coded | Other | ||
| Anatomy/Region of Interest | Doppler | Doppler | Doppler | Doppler | Doppler | Modes* | Imaging | Pulse® | [Notes] | ||||
| Ophthalmic | |||||||||||||
| Fetal/Obstetrics[7] | |||||||||||||
| Abdominal[1] | |||||||||||||
| Pediatric | |||||||||||||
| Small Organ[2] | |||||||||||||
| Neonatal Cephalic | |||||||||||||
| Adult Cephalic | |||||||||||||
| Cardiac Adult | P | P | P | P | P | P | P | P | P | P | |||
| Cardiac Pediatric | |||||||||||||
| Peripheral Vascular | |||||||||||||
| Musculo-skeletal Conventional | |||||||||||||
| Musculo-skeletal Superficial | |||||||||||||
| Other[4] | |||||||||||||
| Exam Type, Means of Access | |||||||||||||
| Transesophageal | P | P | P | P | P | P | P | P | P | P | |||
| Transrectal | |||||||||||||
| Transvaginal | |||||||||||||
| Transurethral | |||||||||||||
| Intraoperative[8] | |||||||||||||
| Intraoperative Neurological | |||||||||||||
| Intravascular | |||||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
[1] Abdominal includes Renal, GYN/Pelvic. Notes:
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle quidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
System provides real-time 3D and 4D acquisition when used with special 4D probes
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{22}------------------------------------------------
Image /page/22/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular border. The border has decorative swirls or flourishes around the letters. The logo is presented in a blue color.
GE LOGIQ S8 with 3Sp-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes* | HarmonicImaging | CodedPulse | Other[Notes] | |
| Ophthalmic | ||||||||||||
| Fetal / Obstetrics[7] | ||||||||||||
| Abdominal[1] | P | P | P | P | P | P | P | P | P | P | 5 | |
| Pediatric | ||||||||||||
| Small Organ[2] | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | P | P | P | P | P | P | P | P | P | P | 5 | |
| Cardiac Adult & Pediatric | P | P | P | P | P | P | P | P | P | P | 5 | |
| Peripheral Vascular | ||||||||||||
| Musculo-skeletal Conventional | ||||||||||||
| Musculo-skeletal Superficial | ||||||||||||
| Other[4] | ||||||||||||
| Exam Type, Means of Access | ||||||||||||
| Transesophageal | ||||||||||||
| Transrectal[8] | ||||||||||||
| Transvaginal | ||||||||||||
| Transuretheral | ||||||||||||
| Intraoperative[8] | ||||||||||||
| Intraoperative Neurological | ||||||||||||
| Intravascular | ||||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
System provides real-time 3D and 4D acquisition when used with special 4D probes
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{23}------------------------------------------------
Image /page/23/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular border. The border has decorative swirls or flourishes around the letters. The logo is presented in a blue color.
GE LOGIQ S8 with C2-6b-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes* | HarmonicImaging | CodedPulse | Other[Notes] | |
| Ophthalmic | ||||||||||||
| Fetal / Obstetrics[7] | ||||||||||||
| Abdominal[1] | P | P | P | P | P | P | P | P | 5,6,9 | |||
| Pediatric | P | P | P | P | P | P | P | P | 5,6,9 | |||
| Small Organ[2] | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Cardiac Adult & Pediatric | ||||||||||||
| Peripheral Vascular | ||||||||||||
| Musculo-skeletal Conventional | ||||||||||||
| Musculo-skeletal Superficial | ||||||||||||
| Other[4] | ||||||||||||
| Exam Type, Means of Access | ||||||||||||
| Transesophageal | ||||||||||||
| Transrectal[8] | ||||||||||||
| Transvaginal | ||||||||||||
| Transuretheral | ||||||||||||
| Intraoperative[8] | ||||||||||||
| Intraoperative Neurological | ||||||||||||
| Intravascular | ||||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
System provides real-time 3D and 4D acquisition when used with special 4D probes
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{24}------------------------------------------------
Image /page/24/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular border. The color of the logo is a light blue. The logo is simple and recognizable, representing the General Electric brand.
GE LOGIQ S8 with 10C-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes | Harmonic Imaging | Coded Pulse | Other [Notes] |
| Ophthalmic | |||||||||||
| Fetal / Obstetrics[7] | |||||||||||
| Abdominal[1] | |||||||||||
| Pediatric | P | P | P | P | P | P | P | P | 5 | ||
| Small Organ[2] | |||||||||||
| Neonatal Cephalic | P | P | P | P | P | P | P | P | 5 | ||
| Adult Cephalic | |||||||||||
| Cardiac Adult & Pediatric | |||||||||||
| Peripheral Vascular | P | P | P | P | P | P | P | P | 5 | ||
| Musculo-skeletal Conventional | |||||||||||
| Musculo-skeletal Superficial | P | P | P | P | P | P | P | P | 5,6 | ||
| Other[4] | |||||||||||
| Exam Type, Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal[8] | |||||||||||
| Transvaginal | |||||||||||
| Transuretheral | |||||||||||
| Intraoperative[8] | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
System provides real-time 3D and 4D acquisition when used with special 4D probes
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{25}------------------------------------------------
Image /page/25/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a circular border. The logo is colored in a light blue hue.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S8 with RSP6-16-D Transducer
Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical Application | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes* | HarmonicImaging | CodedPulse® | Other[Notes] |
| Anatomy/Region of Interest | |||||||||||
| Ophthalmic | |||||||||||
| Fetal/Obstetrics[7] | |||||||||||
| Abdominal[1] | |||||||||||
| Pediatric | |||||||||||
| Small Organ[2] | P | P | P | P | P | P | P | P | P | [5,6] | |
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | |||||||||||
| Cardiac Adult | |||||||||||
| Cardiac Pediatric | |||||||||||
| Peripheral Vascular | P | P | P | P | P | P | P | P | P | [5,6] | |
| Musculo-skeletal Conventional | P | P | P | P | P | P | P | P | P | [5,6] | |
| Musculo-skeletal Superficial | P | P | P | P | P | P | P | P | P | [5,6] | |
| Other [4] | |||||||||||
| Exam Type, Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal | |||||||||||
| Transvaginal | |||||||||||
| Transurethral | |||||||||||
| Intraoperative[8] | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added via Appendix E
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
System provides real-time 3D and 4D acquisition when used with special 4D probes
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{26}------------------------------------------------
Image /page/26/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized monogram. The monogram is enclosed within a circular frame, and the frame has decorative swirls or flourishes around it. The logo is colored in a shade of blue.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S8 with C1-6-D Transducer
Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical ApplicationAnatomy/Region of Interest | Mode of Operation | Other[Notes] | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes* | HarmonicImaging | CodedPulse® | |||
| Ophthalmic | ||||||||||||
| Fetal/Obstetrics[7] | P | P | P | P | P | P | P | P | [5,6,9] | |||
| Abdominal[1] | P | P | P | P | P | P | P | P | [3,5,6,9,10] | |||
| Pediatric | ||||||||||||
| Small Organ[2] | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Peripheral Vascular | P | P | P | P | P | P | P | P | [3,5,6,9,10] | |||
| Musculo-skeletal Conventional | ||||||||||||
| Musculo-skeletal Superficial | ||||||||||||
| Other[4] | P | P | P | P | P | P | P | P | [3,5,6,9,10] | |||
| Exam Type, Means of Access | ||||||||||||
| Transesophageal | ||||||||||||
| Transrectal | ||||||||||||
| Transvaginal | ||||||||||||
| Transurethral | ||||||||||||
| Intraoperative[8] | ||||||||||||
| Intraoperative Neurological | ||||||||||||
| Intravascular | ||||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added via Appendix E
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaqing mode
[6] Needle quidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
System provides real-time 3D and 4D acquisition when used with special 4D probes
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{27}------------------------------------------------
Image /page/27/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular border. The border has decorative swirls or flourishes around the letters. The logo is presented in a blue color.
GE LOGIQ S8 with L3-12-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes* | HarmonicImaging | CodedPulse | Other[Notes] | ||
| Ophthalmic | |||||||||||||
| Fetal / Obstetrics[7] | P | P | P | P | P | P | P | P | P | P | 5,6 | ||
| Abdominal[1] | P | P | P | P | P | P | P | P | P | P | 3,5,6 | ||
| Pediatric | P | P | P | P | P | P | P | P | P | P | 3,5,6 | ||
| Small Organ[2] | P | P | P | P | P | P | P | P | P | P | 3,5,6 | ||
| Neonatal Cephalic | |||||||||||||
| Adult Cephalic | |||||||||||||
| Cardiac Adult & Pediatric | |||||||||||||
| Peripheral Vascular | P | P | P | P | P | P | P | P | P | P | 3,5,6 | ||
| Musculo-skeletal Conventional | P | P | P | P | P | P | P | P | P | P | 3,5,6 | ||
| Musculo-skeletal Superficial | P | P | P | P | P | P | P | P | P | P | 3,5,6 | ||
| Other[4] | |||||||||||||
| Exam Type, Means of Access | |||||||||||||
| Transesophageal | |||||||||||||
| Transrectal | |||||||||||||
| Transvaginal | |||||||||||||
| Transuretheral | |||||||||||||
| Intraoperative[8] | |||||||||||||
| Intraoperative Neurological | |||||||||||||
| Intravascular | |||||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added via Appendix E
[1] Abdominal includes Renal, GYN/Pelvic. Notes:
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
System provides real-time 3D and 4D acquisition when used with special 4D probes
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{28}------------------------------------------------
Image /page/28/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular frame. The frame has decorative swirls or flourishes around the perimeter, giving it a classic or vintage appearance. The logo is presented in a blue color.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S8 with P8D Transducer
Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical Application | B | M | PW | CW | Color | Color M | Power | Combined | Harmonic | Coded | Other | |
| Anatomy/Region of Interest | Doppler | Doppler | Doppler | Doppler | Doppler | Modes* | Imaging | Pulse® | [Notes] | |||
| Ophthalmic | ||||||||||||
| Fetal/Obstetrics[7] | ||||||||||||
| Abdominal[1] | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ[2] | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Peripheral Vascular | P | |||||||||||
| Musculo-skeletal Conventional | ||||||||||||
| Musculo-skeletal Superficial | ||||||||||||
| Other[4] | ||||||||||||
| Exam Type, Means of Access | ||||||||||||
| Transesophageal | ||||||||||||
| Transrectal | ||||||||||||
| Transvaginal | ||||||||||||
| Transurethral | ||||||||||||
| Intraoperative[8] | ||||||||||||
| Intraoperative Neurological | ||||||||||||
| Intravascular | ||||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added via Appendix E
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
System provides real-time 3D and 4D acquisition when used with special 4D probes
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{29}------------------------------------------------
Image /page/29/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular border. The border has decorative swirls or flourishes around the letters. The logo is presented in a blue color.
GE LOGIQ S8 with BE9CS-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes* | HarmonicImaging | CodedPulse | Other[Notes] |
|---|---|---|---|---|---|---|---|---|---|---|---|
| Ophthalmic | |||||||||||
| Fetal / Obstetrics[7] | |||||||||||
| Abdominal[1] | |||||||||||
| Pediatric | |||||||||||
| Small Organ[2] | |||||||||||
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | |||||||||||
| Cardiac Adult & Pediatric | |||||||||||
| Peripheral Vascular | |||||||||||
| Musculo-skeletal Conventional | |||||||||||
| Musculo-skeletal Superficial | |||||||||||
| Other[4] | P | P | P | P | P | P | P | P | 3, 5, 6 | ||
| Exam Type, Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal[8] | P | P | P | P | P | P | P | P | 3, 5, 6 | ||
| Transvaginal | P | P | P | P | P | P | P | P | 3, 5, 6 | ||
| Transuretheral | |||||||||||
| Intraoperative[8] | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA, E = added via Appendix E
[1] Abdominal includes Renal, GYN/Pelvic. Notes:
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaqing mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{30}------------------------------------------------
Image /page/30/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular frame. The frame has decorative swirls or flourishes around the perimeter, giving it a classic or vintage appearance. The logo is presented in a blue color.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S8 with C2-7-D Transducer
Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Anatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes* | HarmonicImaging | CodedPulse® | Other[Notes] | |
| Ophthalmic | ||||||||||||
| Fetal/Obstetrics[7] | ||||||||||||
| Abdominal[1] | P | P | P | P | P | P | P | P | 5,6,9 | |||
| Pediatric | ||||||||||||
| Small Organ[2] | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Peripheral Vascular | ||||||||||||
| Musculo-skeletal Conventional | ||||||||||||
| Musculo-skeletal Superficial | ||||||||||||
| Other[4] | ||||||||||||
| Exam Type, Means of Access | ||||||||||||
| Transesophageal | ||||||||||||
| Transrectal | ||||||||||||
| Transvaginal | ||||||||||||
| Transurethral | ||||||||||||
| Intraoperative[8] | ||||||||||||
| Intraoperative Neurological | ||||||||||||
| Intravascular | ||||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
System provides real-time 3D and 4D acquisition when used with special 4D probes
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{31}------------------------------------------------
Image /page/31/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized, cursive font. The letters are enclosed within a blue circle, and there are decorative swirls or flourishes around the perimeter of the circle, also in blue. The logo is simple, recognizable, and has been associated with the General Electric company for many years.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S8 with L3-9i-D Transducer
Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes* | HarmonicImaging | CodedPulse* | Other[Notes] | |
| Ophthalmic | ||||||||||||
| Fetal/Obstetrics[7] | ||||||||||||
| Abdominal[1] | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ[2] | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Peripheral Vascular | ||||||||||||
| Musculo-skeletal Conventional | ||||||||||||
| Musculo-skeletal Superficial | ||||||||||||
| Other[4] | ||||||||||||
| Exam Type, Means of Access | ||||||||||||
| Transesophageal | ||||||||||||
| Transrectal | ||||||||||||
| Transvaginal | ||||||||||||
| Transurethral | ||||||||||||
| Intraoperative[8] | P | P | P | P | P | P | P | P | 5 | |||
| Intraoperative Neurological | ||||||||||||
| Intravascular | ||||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
System provides real-time 3D and 4D acquisition when used with special 4D probes
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{32}------------------------------------------------
Image /page/32/Picture/0 description: The image shows the General Electric (GE) logo, which is a blue circle with the letters 'g' and 'e' intertwined in the center. The letters are stylized and connected, creating a unique and recognizable design. The logo is simple yet iconic, representing the company's brand identity.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S8 with C3-10-D Transducer
Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical Application | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes* | HarmonicImaging | CodedPulse* | Other[Notes] |
| Anatomy/Region of Interest | |||||||||||
| Ophthalmic | |||||||||||
| Fetal/Obstetrics[7] | |||||||||||
| Abdominal[1] | |||||||||||
| Pediatric | P | P | P | P | P | P | P | P | 5,9 | ||
| Small Organ[2] | P | P | P | ||||||||
| Neonatal Cephalic | P | P | P | P | P | P | P | P | 5,9 | ||
| Adult Cephalic | |||||||||||
| Cardiac Adult | |||||||||||
| Cardiac Pediatric | |||||||||||
| Peripheral Vascular | P | P | P | P | P | P | P | P | 5,9 | ||
| Musculo-skeletal Conventional | |||||||||||
| Musculo-skeletal Superficial | P | P | P | P | P | P | P | P | 5,6,9 | ||
| Other[4] | |||||||||||
| Exam Type, Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal | |||||||||||
| Transvaginal | |||||||||||
| Transurethral | |||||||||||
| Intraoperative[8] | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
System provides real-time 3D and 4D acquisition when used with special 4D probes
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{33}------------------------------------------------
Image /page/33/Picture/0 description: The image shows the General Electric (GE) logo, which is a blue circle with the letters 'g' and 'e' intertwined in the center. The letters are stylized and connected, creating a unique and recognizable design. The logo is simple yet iconic, representing the company's brand identity.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S8 with C1-6VN-D Transducer
Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes* | HarmonicImaging | CodedPulse* | Other[Notes] |
| Ophthalmic | |||||||||||
| Fetal/Obstetrics[7] | E | E | E | E | E | E | E | E | [5,6,9] | ||
| Abdominal[1] | E | E | E | E | E | E | E | E | [3,5,6,9,10] | ||
| Pediatric | |||||||||||
| Small Organ[2] | |||||||||||
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | |||||||||||
| Cardiac Adult | |||||||||||
| Cardiac Pediatric | |||||||||||
| Peripheral Vascular | E | E | E | E | E | E | E | E | [3,5,6,9] | ||
| Musculo-skeletal Conventional | |||||||||||
| Musculo-skeletal Superficial | |||||||||||
| Other[4] | E | E | E | E | E | E | E | E | [3,5,6,9] | ||
| Exam Type, Means of Access | |||||||||||
| Transesophageal | |||||||||||
| Transrectal | |||||||||||
| Transvaginal | |||||||||||
| Transurethral | |||||||||||
| Intraoperative[8] | |||||||||||
| Intraoperative Neurological | |||||||||||
| Intravascular | |||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E (previously cleared with LOGIQ E9 K163077)
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
System provides real-time 3D and 4D acquisition when used with special 4D probes
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{34}------------------------------------------------
Image /page/34/Picture/0 description: The image shows the General Electric (GE) logo, which is a blue circle with the letters 'g' and 'e' intertwined in the center. The letters are stylized and connected, creating a unique and recognizable design. The logo is simple, clean, and has been used by GE for many years.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S8 with C2-9VN-D Transducer
Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical ApplicationAnatomy/Region of Interest | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes* | HarmonicImaging | CodedPulse° | Other[Notes] | |
| Ophthalmic | ||||||||||||
| Fetal/Obstetrics[7] | E | E | E | E | E | E | E | E | [5,6,9] | |||
| Abdominal[1] | E | E | E | E | E | E | E | E | [3,5,6,9] | |||
| Pediatric | E | E | E | E | E | E | E | E | [3,5,6,9] | |||
| Small Organ[2] | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Peripheral Vascular | E | E | E | E | E | E | E | E | [3,5,6,9] | |||
| Musculo-skeletal Conventional | ||||||||||||
| Musculo-skeletal Superficial | ||||||||||||
| Other[4] | ||||||||||||
| Exam Type, Means of Access | ||||||||||||
| Transesophageal | ||||||||||||
| Transrectal | ||||||||||||
| Transvaginal | ||||||||||||
| Transurethral | ||||||||||||
| Intraoperative[8] | ||||||||||||
| Intraoperative Neurological | ||||||||||||
| Intravascular | ||||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E (previously cleared with LOGIQ E9 K163077)
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
System provides real-time 3D and 4D acquisition when used with special 4D probes
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{35}------------------------------------------------
Image /page/35/Picture/0 description: The image shows the General Electric (GE) logo, which is a blue circle with the letters 'g' and 'e' intertwined in the center. The letters are stylized and connected, creating a unique and recognizable design. The logo is simple yet iconic, representing the company's brand identity.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S8 with C2-7VN-D Transducer
Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical Application | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes* | HarmonicImaging | CodedPulse® | Other[Notes] | |
| Anatomy/Region of Interest | ||||||||||||
| Ophthalmic | ||||||||||||
| Fetal/Obstetrics[7] | ||||||||||||
| Abdominal[1] | E | E | E | E | E | E | E | E | [5,6,9] | |||
| Pediatric | ||||||||||||
| Small Organ[2] | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Peripheral Vascular | ||||||||||||
| Musculo-skeletal Conventional | ||||||||||||
| Musculo-skeletal Superficial | ||||||||||||
| Other[4] | ||||||||||||
| Exam Type, Means of Access | ||||||||||||
| Transesophageal | ||||||||||||
| Transrectal | ||||||||||||
| Transvaginal | ||||||||||||
| Transurethral | ||||||||||||
| Intraoperative[8] | ||||||||||||
| Intraoperative Neurological | ||||||||||||
| Intravascular | ||||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E (previously cleared with LOGIQ E9 K163077)
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
System provides real-time 3D and 4D acquisition when used with special 4D probes
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{36}------------------------------------------------
Image /page/36/Picture/0 description: The image shows the General Electric (GE) logo, which is a blue circle with the letters 'g' and 'e' intertwined in the center. The letters are stylized and connected, creating a unique and recognizable design. The logo is simple yet iconic, representing the company's brand identity.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S8 with M5Sc-D Transducer
Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical Application | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes* | HarmonicImaging | CodedPulse* | Other[Notes] | |
| Anatomy/Region of Interest | ||||||||||||
| Ophthalmic | ||||||||||||
| Fetal/Obstetrics[7] | E | E | E | E | E | E | E | E | E | E | [5,9] | |
| Abdominal[1] | E | E | E | E | E | E | E | E | E | E | [5,9] | |
| Pediatric | ||||||||||||
| Small Organ[2] | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | E | E | E | E | E | E | E | E | E | E | [5,9] | |
| Cardiac Adult | E | E | E | E | E | E | E | E | E | E | [5,9] | |
| Cardiac Pediatric | E | E | E | E | E | E | E | E | E | E | [5,9] | |
| Peripheral Vascular | ||||||||||||
| Musculo-skeletal Conventional | ||||||||||||
| Musculo-skeletal Superficial | ||||||||||||
| Other[4] | ||||||||||||
| Exam Type, Means of Access | ||||||||||||
| Transesophageal | ||||||||||||
| Transrectal | ||||||||||||
| Transvaginal | ||||||||||||
| Transurethral | ||||||||||||
| Intraoperative[8] | ||||||||||||
| Intraoperative Neurological | ||||||||||||
| Intravascular | ||||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E (previously cleared with VIVID E90/E95 K150087)
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
System provides real-time 3D and 4D acquisition when used with special 4D probes
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{37}------------------------------------------------
Image /page/37/Picture/0 description: The image shows the General Electric (GE) logo, which is a blue circle with the letters 'g' and 'e' intertwined in the center. The letters are stylized and connected, creating a unique and recognizable design. The logo is simple, clean, and has been used by GE for many years.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S8 with FibroScan M+ Transducer
Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical Application | B | M | PW | CW | Color | Color M | Power | Combined | Harmonic | Coded | Other | |
| Anatomy/Region of Interest | Doppler | Doppler | Doppler | Doppler | Doppler | Modes* | Imaging | Pulse® | [Notes] | |||
| Ophthalmic | ||||||||||||
| Fetal/Obstetrics[7] | ||||||||||||
| Abdominal[1] | N | [11,12, 13] | ||||||||||
| Pediatric | N | [11, 13] | ||||||||||
| Small Organ[2] | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Peripheral Vascular | ||||||||||||
| Musculo-skeletal Conventional | ||||||||||||
| Musculo-skeletal Superficial | ||||||||||||
| Other[4] | ||||||||||||
| Exam Type, Means of Access | ||||||||||||
| Transesophageal | ||||||||||||
| Transrectal | ||||||||||||
| Transvaginal | ||||||||||||
| Transurethral | ||||||||||||
| Intraoperative[8] | ||||||||||||
| Intraoperative Neurological | ||||||||||||
| Intravascular | ||||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[13] A Mode
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
System provides real-time 3D and 4D acquisition when used with special 4D probes
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{38}------------------------------------------------
Image /page/38/Picture/0 description: The image shows the General Electric (GE) logo, which is a blue circle with the letters 'g' and 'e' intertwined in the center. The letters are stylized and connected, creating a unique and recognizable design. The logo is simple yet iconic, representing the company's brand identity.
Diagnostic Ultrasound Indications for Use Form GE LOGIQ S8 with FibroScan XL+ Transducer
Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Clinical Application | B | M | PWDoppler | CWDoppler | ColorDoppler | Color MDoppler | PowerDoppler | CombinedModes* | HarmonicImaging | CodedPulse® | Other[Notes] | |
| Anatomy/Region of Interest | ||||||||||||
| Ophthalmic | ||||||||||||
| Fetal/Obstetrics[7] | ||||||||||||
| Abdominal[1] | N | [11, 12, 13] | ||||||||||
| Pediatric | ||||||||||||
| Small Organ[2] | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Peripheral Vascular | ||||||||||||
| Musculo-skeletal Conventional | ||||||||||||
| Musculo-skeletal Superficial | ||||||||||||
| Other[4] | ||||||||||||
| Exam Type, Means of Access | ||||||||||||
| Transesophageal | ||||||||||||
| Transrectal | ||||||||||||
| Transvaginal | ||||||||||||
| Transurethral | ||||||||||||
| Intraoperative[8] | ||||||||||||
| Intraoperative Neurological | ||||||||||||
| Intravascular | ||||||||||||
| Laparoscopic |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development, GYN
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[9] Volume navigation
[10] Shear wave elastography
[11] FibroScan elastography
[12] Controlled Attenuation Parameter at 3.5MHz
[13] A Mode
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
System provides real-time 3D and 4D acquisition when used with special 4D probes
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OlR)
{39}------------------------------------------------
Image /page/39/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized, cursive font. The letters are enclosed within a circular shape, and the entire logo is presented in a light blue color.
GE Healthcare 510(k) Premarket Notification Submission
Section 5: 510(k) Summary
LOGIQ 88
{40}------------------------------------------------
Image /page/40/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. The circle is surrounded by a decorative, swirling pattern, also in blue. The logo is simple, recognizable, and associated with a well-known multinational corporation.
GE Healthcare 510(k) Premarket Notification Submission
510(k) Summary
| In accordance with 21 CFR 807.92 the following summary of information is provided: | |
|---|---|
| Date: | February 13, 2017 |
| Submitter: | GE Healthcare9900 Innovation DrWauwatosa, WI 53226 |
| Manufacturer: | GE Ultrasound Korea, Ltd.9, Sunhwan-ro 214 beon-gil, Jungwon-gu,Seongnam-si, Gyeonggi-do, Republic of Korea |
| Primary Contact Person: | Bryan BehnRegulatory Affairs DirectorGE HealthcareT:(414)721-4214F:(414)918-8275 |
| Alternate Contact Person: | Qingmeng ChenRegulatory Affairs LeaderGE Ultrasound Korea, Ltd.T: +86-28-6272-2040F: +81-42-6272-2462 |
| Device: | Trade Name: |
| LOGIQ S8 | |
| Common/Usual Name: | LOGIQ S8 |
| Classification Names: | Class II |
| Product Code: | Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550 90-IYNUltrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYODiagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX |
| Primary Predicate Device: | K152195 LOGIQ S8 Diagnostic Ultrasound System |
| Reference Predicate Device(s): | K163077 LOGIQ E9 Diagnostic Ultrasound SystemK150087 VIVID E90/E95 Diagnostic Ultrasound SystemK160524 FibroScan 530 Compact Ultrasonic Pulsed EchoImaging SystemK160184 VOLUSON S6/S8/S10 Diagnostic Ultrasound SystemK161047 LOGIQ P7/P9 Diagnostic Ultrasound SystemK160182 LOGIQ S7 EXPERT Diagnostic Ultrasound System |
| Device Description: | The LOGIQ S8 is a full featured, general purpose diagnosticultrasound system which consists of a mobile consoleapproximately 520 mm wide (keyboard), 865 mm deep and 1760 |
| Intended Use: | mm high that provides digital acquisition, processing and display capability. The user interface includes a computer keyboard, specialized controls, 10.1-inch LCD touch screen and color widescreen monitor.The device is intended for use by a qualified physician for ultrasound evaluation of Fetal/Obstetrics; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transrectal; Transvaginal; Transesophageal and Intraoperative (abdominal, thoracic, vascular). |
| Technology: | The LOGIQ S8 employs the same fundamental scientific technology as its predicate device(s). |
| Determination ofSubstantial Equivalence: | Comparison to PredicatesThe following is an overview of the differences between the proposed LOGIQ S8 and the predicate LOGIQ S8 (K152195). The proposed LOGIQ S8 has had several features and probes migrated to it currently cleared on other GE Ultrasound systems. The systems are all intended for diagnostic ultrasound imaging and fluid flow analysis. The LOGIQ S8 and predicate LOGIQ S8 systems have the same clinical intended use. The LOGIQ S8 and predicate LOGIQ S8 systems have the same imaging modes except the new LOGIQ S8 adds Fibroscan elastography and CAP (Controlled Attenuation Parameter) which is substantially equivalent to FibroScan (K160524). The systems are manufactured with materials which have been evaluated and found to be safe for the intended use of the device. The systems have acoustic power levels which are below the applicable FDA limits. The LOGIQ S8 and predicate LOGIQ S8 systems have similar capability in terms of performing measurements, capturing digital images, reviewing and reporting studies. The LOGIQ S8 and predicate systems have been designed in compliance with approved electrical and physical safety standards. The following probes and features have been migrated from GE ultrasound systems LOGIQ E9 (K163077): C1-6VN-D, C2-9VN-D, C2-7VN-D probe, and Contrast parametric imaging. |
{41}------------------------------------------------
Image /page/41/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized, cursive font, enclosed within a circular border. The color of the logo is a light blue.
{42}------------------------------------------------
Image /page/42/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters 'GE' in a stylized script, enclosed within a circular border. The border has decorative swirls or flourishes around the letters, giving it a classic and recognizable appearance. The logo is presented in a blue color scheme.
- The following probe has been migrated from the VIVID E90/E95 (K150087): M5Sc-D probe.
- The following features and probes have been migrated from FibroScan (K160524): FibroScan Elastography, CAP (Controlled Attenuation Parameter), FibroScan M* probe and FiroScan XL+ probe.
- The following features have been migrated from the VOLUSON S6/S8/S10 (K160184) STIC, and Omniview with VCI.
- The following features have been migrated from the ● LOGIQ S7 EXPERT (K160182) Email to MMS and Strain elastography on BE9CS-D.
- The following features have been migrated from the ● LOGIQ P7/P9 (K161047) MyTrainer* and Advanced battery.
Summary of Non-Clinical Tests:
The device has been evaluated for acoustic output. biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic and mechanical safety, and has been found to conform with applicable medical device safety standards. The LOGIO S8 and its applications comply with voluntary standards:
-
- AAMI/ANSI ES60601-1, Medical Electrical Equipment - Part 1: General Requirements for Safety
-
- IEC60601-1-2. Medical Electrical Equipment Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility Requirements and Tests
-
- IEC60601-2-37, Medical Electrical Equipment -Part 2-37: Particular Requirements for Basic Safety and Essential Performance of Ultrasonic Medical Diagnostic and Monitoring Equipment
-
- ISO10993-1. Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing
-
- ISO14971. Application of risk management to medical devices
-
- NEMA UD 2, Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment
{43}------------------------------------------------
Image /page/43/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. The circle is surrounded by a swirling pattern that resembles water or wind. The logo is simple and recognizable, and it is often used to represent the company's brand.
-
- NEMA UD 3, Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment
-
- NEMA, Digital Imaging and Communications in Medicine (DICOM) Set. (Radiology)
The following quality assurance measures are applied to the development of the system:
- Risk Analysis
- Requirements Reviews ●
- Design Reviews ●
- Testing on unit level (Module verification)
- Integration testing (System verification) ●
- Final Acceptance Testing (Validation)
- Performance testing (Verification)
- Safety testing (Verification) .
Transducer materials and other patient contact materials are biocompatible.
Summary of Clinical Tests:
The subject of this premarket submission, LOGIQ S8, did not require clinical studies to support substantial equivalence.
- Conclusion: GE Healthcare considers the LOGIQ S8 to be as safe, as effective, and performance is substantially equivalent to the predicate device(s).
N/A