K Number
K160182

Validate with FDA (Live)

Device Name
LOGIQ S7 Expert
Manufacturer
Date Cleared
2016-02-26

(30 days)

Product Code
Regulation Number
892.1550
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is intended for use by a qualified physician for ultrasound evaluation of Obstetrics (Fetal); Abdominal (includes renal, GYN/Pelvic); Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Cardiac (adult and pediatric); Peripheral Vasculo-skeletal Conventional and Superficial; Urology (including prostate); Transrectal: Transvaginal; Transesophageal and Intraoperative (abdominal, thoracic, vascular).

Device Description

The LOGIQ S7 Expert is a full featured, general purpose diagnostic ultrasound system which consists of a mobile console approximately 530mm (Caster), 520mm (Keyboard), 565mm (Monitor) wide, 865mm deep and 1760mm high that provides digital acquisition, processing and display capability. The user interface includes a computer keyboard, specialized controls, 10.1-inch LCD touch screen and color 23-inch LCD widescreen monitor.

AI/ML Overview

The provided document is a 510(k) Premarket Notification Submission for the GE LOGIQ S7 Expert diagnostic ultrasound system. It outlines the device's indications for use and states its substantial equivalence to previously cleared devices. It explicitly mentions that clinical studies were NOT required to support substantial equivalence for this submission. Therefore, the document does not contain information about acceptance criteria for device performance based on a clinical study, nor does it present the results of such a study.

The submission focuses instead on demonstrating substantial equivalence through:

  • Comparison to Predicate Devices: Showing that the LOGIQ S7 Expert shares the same fundamental scientific technology, clinical intended use, imaging modes, and similar capabilities (measurements, digital image capture, reviewing/reporting) as its predicate devices (K141261 LOGIQ S7 Expert, K152195 LOGIQ S8, K141639 Voluson S8, K143452 Logiq P9).
  • Safety and Performance Testing: Confirming compliance with various safety standards related to acoustic output, biocompatibility, cleaning/disinfection, thermal, electrical, electromagnetic, and mechanical safety.
  • Software Features Migration: Listing new software features (STIC, OmniView, Cardiac AFI, Email2MMS) that were previously cleared on other GE ultrasound systems.
  • Transducer Equivalence: Confirming that new transducers (C1-6-D, C3-10-D, P8D, BE9CS-D) are already cleared on predicate devices (LOGIQ S8), and a new S2-5-D probe is an incremental improvement to an existing one (3Sp-D) and deemed equivalent.

Since no clinical study was conducted as part of this submission, the following information cannot be extracted from the provided text:

  1. A table of acceptance criteria and the reported device performance: Not available as no clinical study was performed for this 510(k).
  2. Sample size used for the test set and the data provenance: Not applicable as no clinical test set was used.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
  4. Adjudication method: Not applicable.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. The device is a diagnostic ultrasound system, not explicitly described as having AI features that would necessitate an MRMC study for improved reader performance. Even if it had such features, no clinical study was performed.
  6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable.
  7. The type of ground truth used: Not applicable.
  8. The sample size for the training set: Not applicable.
  9. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract symbol that appears to be three stylized human profiles facing to the right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 26, 2016

GE HEALTHCARE BRYAN BEHN SR. REGULATORY AFFAIRS MANAGER 9900 INNOVATION DR. WAUWATOSA WI 53226

Re: K160182

Trade/Device Name: Logiq S7 Expert Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: January 20, 2016 Received: January 27, 2016

Dear Mr. Behn:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D.'Hara

For

Robert Ochs Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

1601

Device Name LOGIQ S7 Expert

Indications for Use (Describe)

The device is intended for use by a qualified physician for ultrasound evaluation of Obstetrics (Fetal); Abdominal (includes renal, GYN/Pelvic); Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Cardiac (adult and pediatric); Peripheral Vasculo-skeletal Conventional and Superficial; Urology (including prostate); Transrectal: Transvaginal; Transesophageal and Intraoperative (abdominal, thoracic, vascular).

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular border. The letters and the border are both in a blue color. The logo is simple and recognizable, representing the company's brand identity.

Indications for Use Forms

The following forms represent indications with clinical applications and exam types along with the modes of operation for the LOGIQ S7 Expert system and for all of its probe/mode combinations. Combinations identified by "P" represent those previously cleared with the unmodified LOGIQ S7 Expert (K14261) or on another GE ultrasound system in the case of the probes. "N" is new in this submission.

The subject modification does not alter the previously cleared system level indications and clinical applications.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular border. The letters and the border are both a light blue color, set against a white background. The logo is simple, yet recognizable, representing the multinational conglomerate corporation.

GE Healthcare 510(k) Premarket Notification Submission

Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert Ultrasound System

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationAnatomy/Region of InterestBMPWDopplerCWDopplerColorDopplerColor MDopplerPowerDopplerCombinedModes*HarmonicImagingCodedPulseOther[Notes]
Ophthalmic
Fetal / Obstetrics[7]PPPPPPPPPP5,6
Abdominal[1]PPPPPPPPPP3,5,6
PediatricPPPPPPPPPP3,5,6
Small Organ[2]PPPPPPPPPP3,5,6
Neonatal CephalicPPPPPPPPPP5,6
Adult CephalicPPPPPPPPPP5,6
Cardiac Adult & PediatricPPPPPPPPPP3,5,6
Peripheral VascularPPPPPPPPPP3,5,6
Musculo-skeletal ConventionalPPPPPPPPPP3,5,6
Musculo-skeletal SuperficialPPPPPPPPPP3,5,6
Other[4]PPPPPPPPPP3,5,6
Exam Type, Means of Access
TransesophagealPPPPPPPPPP
Transrectal [8]PPPPPPPPPP3,5,6
TransvaginalPPPPPPPPPP3,5,6
TransuretheralPPPPPPPPPP3,5,6
Intraoperative [8]PPPPPPPPPP3,5,6
Intraoperative Neurological
Intravascular
Laparoscopic

N = new indication; P = previously cleared by FDA

[1] Abdominal includes Renal, GYN/Pelvic. Notes:

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development, GYN

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle has decorative flourishes around its perimeter, giving it a classic and recognizable appearance. The logo is simple, yet distinctive, and is widely recognized as the symbol of the General Electric company.

GE Healthcare 510(k) Premarket Notification Submission

Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with C1-5-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Mode of Operation
Clinical ApplicationAnatomy/Region of InterestBMPWDopplerCWDopplerColorDopplerColor MDopplerPowerDopplerCombinedModes*HarmonicImagingCodedPulseOther(Notes)
Ophthalmic
Fetal / Obstetrics[7]PPPPPPPPPP5,6
Abdominal[1]PPPPPPPPPP3,5,6
PediatricPPPPPPPPPP3,5,6
Small Organ[2]
Neonatal Cephalic
Adult Cephalic
Cardiac Adult & Pediatric
Peripheral VascularPPPPPPPPPP3,5,6
Musculo-skeletal Conventional
Musculo-skeletal Superficial
Other[4]PPPPPPPPPP3,5,6
Exam Type, Means of Access
Transesophageal
Transrectal[8]
Transvaginal
Transuretheral
Intraoperative[8]
Intraoperative Neurological
Intravascular
Laparoscopic

N = new indication; P = previously cleared by FDA

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development, GYN

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle has decorative flourishes around its perimeter, giving it a classic and recognizable appearance. The logo is simple, yet distinctive, and is widely recognized as the symbol of the General Electric company.

Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with 9L-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationAnatomy/Region of InterestBMPWDopplerCWDopplerColorDopplerColor MDopplerPowerDopplerCombinedModes*HarmonicImagingCodedPulseOther(Notes)
Ophthalmic
Fetal / Obstetrics[7]PPPPPPPPP5,6
Abdominal[1]PPPPPPPPP3,5,6
PediatricPPPPPPPPP3,5,6
Small Organ[2]PPPPPPPPP3,5,6
Neonatal Cephalic
Adult Cephalic
Cardiac Adult & Pediatric
Peripheral VascularPPPPPPPPP3,5,6
Musculo-skeletal ConventionalPPPPPPPPP3,5,6
Musculo-skeletal SuperficialPPPPPPPPP3,5,6
Other[4]
Exam Type, Means of Access
Transesophageal
Transrectal[8]
Transvaginal
Transuretheral
Intraoperative[8]
Intraoperative Neurological
Intravascular
Laparoscopic

N = new indication; P = previously cleared by FDA

[1] Abdominal includes Renal, GYN/Pelvic. Notes:

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development, GYN

(8) Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle has decorative swirls around the perimeter, giving it a classic and recognizable appearance. The logo is simple, yet distinctive, representing the brand identity of General Electric.

GE Healthcare 510(k) Premarket Notification Submission

Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with ML6-15 Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Mode of Operation
Clinical ApplicationAnatomy/Region of InterestBMPWDopplerCWDopplerColorDopplerColor MDopplerPowerDopplerCombinedModes*HarmonicImagingCodedPulseOther(Notes)
Ophthalmic
Fetal / Obstetrics[7]
Abdominal[1]
PediatricPPPPPPPPP3,5,6
Small Organ[2]PPPPPPPPP3,5,6
Neonatal CephalicPPPPPPPPP3,5,6
Adult Cephalic
Cardiac Adult & Pediatric
Peripheral VascularPPPPPPPPP3,5,6
Musculo-skeletal ConventionalPPPPPPPPP3,5,6
Musculo-skeletal SuperficialPPPPPPPPP3,5,6
Other[4]
Exam Type, Means of Access
Transesophageal
Transrectal[8]
Transvaginal
Transuretheral
Intraoperative[8]
Intraoperative Neurological
Intravascular
Laparoscopic

N = new indication; P = previously cleared by FDA

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development, GYN

(8) Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a circular border. The color of the logo is blue, and the background is white.

GE Healthcare 510(k) Premarket Notification Submission

Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with IC5-9-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Mode of Operation
Clinical ApplicationAnatomy/Region of InterestBMPWDopplerCWDopplerColorDopplerColor MDopplerPowerDopplerCombinedModes*HarmonicImagingCodedPulseOther(Notes)
Ophthalmic
Fetal / Obstetrics[7]PPPPPPPPP5,6
Abdominal[1]PPPPPPPPP
Pediatric
Small Organ[2]
Neonatal Cephalic
Adult Cephalic
Cardiac Adult & Pediatric
Peripheral Vascular
Musculo-skeletal Conventional
Musculo-skeletal Superficial
Other[4]PPPPPPPPP3,5,6
Exam Type, Means of Access
Transesophageal
Transrectal[8]PPPPPPPPP3,5,6
TransvaginalPPPPPPPPP3.5,6
Transuretheral
Intraoperative[8]
Intraoperative Neurological
Intravascular
Laparoscopic

N = new indication; P = previously cleared by FDA

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development, GYN

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle has decorative swirls around the perimeter, giving it a classic and recognizable appearance. The logo is simple, yet distinctive, representing the brand identity of General Electric.

Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with 3CRF-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Mode of Operation
Clinical ApplicationAnatomy/Region of InterestBMPW DopplerCW DopplerColor DopplerColor M DopplerPower DopplerCombined Modes*Harmonic ImagingCoded PulseOther (Notes)
Ophthalmic
Fetal / Obstetrics[7]PPPPPPPPP5,6
Abdominal[1]PPPPPPPPP5,6
PediatricPPPPPPPPP5,6
Small Organ [2]
Neonatal Cephalic
Adult Cephalic
Cardiac Adult & Pediatric
Peripheral Vascular
Musculo-skeletal Conventional
Musculo-skeletal Superficial
Other[4]PPPPPPPPP5,6
Exam Type, Means of Access
Transesophageal
Transrectal[8]
Transvaginal
Transuretheral
Intraoperative[8]
Intraoperative Neurological
Intravascular
Laparoscopic

N = new indication; P = previously cleared by FDA

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development, GYN

(8) Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle has decorative flourishes around its perimeter, giving it a classic and recognizable appearance. The logo is simple, yet distinctive, and is widely recognized as the symbol of the General Electric company.

GE Healthcare 510(k) Premarket Notification Submission

Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with L8-18i-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Mode of Operation
Clinical ApplicationAnatomy/Region of InterestBMPW DopplerCW DopplerColor DopplerColor M DopplerPower DopplerCombined Modes*Harmonic ImagingCoded PulseOther (Notes)
Ophthalmic
Fetal / Obstetrics[7]
Abdominal[1]
PediatricPPPPPPPPP5,6
Small Organ[2]PPPPPPPPP
Neonatal CephalicPPPPPPPPP
Adult Cephalic
Cardiac Adult & Pediatric
Peripheral VascularPPPPPPPPP3,5,6
Musculo-skeletal ConventionalPPPPPPPPP3,5,6
Musculo-skeletal SuperficialPPPPPPPPP3,5,6
Other[4]
Exam Type, Means of Access
Transesophageal
Transrectal[8]
Transvaginal
Transuretheral
Intraoperative[8]PPPPPPPPP3,5,6
Intraoperative Neurological
Intravascular
Laparoscopic

N = new indication; P = previously cleared by FDA

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development, GYN

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized script, enclosed within a circular shape. The circle is outlined with a thin blue line, and there are small, white, leaf-like shapes evenly spaced around the inside of the circle.

Diagnostic Ultrasound Indications for Use Form

GE LOGIQ S7 Expert with S4-10-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Mode of Operation
Clinical ApplicationAnatomy/Region of InterestBMPWDopplerCWDopplerColorDopplerColor MDopplerPowerDopplerCombinedModes*HarmonicImagingCodedPulseOther[Notes]
Ophthalmic
Fetal / Obstetrics[7]PPPPPPPPPP5
Abdominal[1]PPPPPPPPPP5
PediatricPPPPPPPPPP5
Small Organ[2]PPPPPPPPPP5
Neonatal CephalicPPPPPPPPPP5
Adult Cephalic
Cardiac Adult & PediatricPPPPPPPPPP
Peripheral Vascular
Musculo-skeletal Conventional
Musculo-skeletal Superficial
Other[4]
Exam Type, Means of Access
Transesophageal
Transrectal[8]
Transvaginal
Transuretheral
Intraoperative[8]
Intraoperative Neurological
Intravascular
Laparoscopic

N = new indication; P = previously cleared by FDA

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development; GYN

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle has decorative swirls around the perimeter, giving it a classic and recognizable appearance. The logo is simple, yet distinctive, representing the brand identity of General Electric.

Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with P2D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Mode of Operation
Clinical ApplicationAnatomy/Region of InterestBMPWDopplerCWDopplerColorDopplerColor MDopplerPowerDopplerCombinedModes*HarmonicImagingCodedPulseOther[Notes]
Ophthalmic
Fetal / Obstetrics[7]
Abdominal[1]
Pediatric
Small Organ[2]
Neonatal Cephalic
Adult CephalicPP
Cardiac Adult & PediatricPP
Peripheral VascularPP
Musculo-skeletal Conventional
Musculo-skeletal Superficial
Other[4]
Exam Type, Means of Access
Transesophageal
Transrectal[8]
Transvaginal
Transuretheral
Intraoperative[8]
Intraoperative Neurological
Intravascular
Laparoscopic

N = new indication; P = previously cleared by FDA

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development, GYN

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle has decorative swirls around the perimeter, giving it a classic and recognizable appearance. The logo is simple, yet distinctive, representing the brand identity of General Electric.

Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with P6D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Mode of Operation
Clinical ApplicationAnatomy/Region of InterestBMPW DopplerCW DopplerColor DopplerColor M DopplerPower DopplerCombined Modes*Harmonic ImagingCoded PulseOther [Notes]
Ophthalmic
Fetal / Obstetrics[7]
Abdominal[1]
Pediatric
Small Organ[2]
Neonatal Cephalic
Adult CephalicPP
Cardiac Adult & PediatricPP
Peripheral VascularPP
Musculo-skeletal Conventional
Musculo-skeletal Superficial
Other[4]
Exam Type, Means of Access
Transesophageal
Transrectal[8]
Transvaginal
Transuretheral
Intraoperative[8]
Intraoperative Neurological
Intravascular
Laparoscopic

N = new indication; P = previously cleared by FDA

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development; GYN

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The letters are white, creating a contrast against the blue background.

GE Healthcare 510(k) Premarket Notification Submission

Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with RAB4-8-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Mode of Operation
Clinical ApplicationAnatomy/Region of InterestBMPWDopplerCWDopplerColorDopplerColor MDopplerPowerDopplerCombinedModes*HarmonicImagingCodedPulseOther(Notes)
Ophthalmic
Fetal / Obstetrics[7]PPPPPPPPP5,6
Abdominal[1]PPPPPPPPP5,6
PediatricPPPPPPPPP5,6
Small Organ[2]
Neonatal Cephalic
Adult Cephalic
Cardiac Adult & Pediatric
Peripheral Vascular
Musculo-skeletal ConventionalPPPPPPPPP5,6
Musculo-skeletal Superficial
Other[4]PPPPPPPPP5,6
Exam Type, Means of Access
Transesophageal
Transrectal[8]
Transvaginal
Transuretheral
Intraoperative[8]
Intraoperative Neurological
Intravascular
Laparoscopic

N = new indication; P = previously cleared by FDA

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development, GYN

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle has decorative flourishes around its perimeter, giving it a classic and recognizable appearance. The logo is simple, yet distinctive, and is widely recognized as the symbol of the General Electric company.

GE Healthcare 510(k) Premarket Notification Submission

Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with 11L-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Mode of Operation
Clinical ApplicationAnatomy/Region of InterestBMPWDopplerCWDopplerColorDopplerColor MDopplerPowerDopplerCombinedModes*HarmonicImagingCodedPulseOther(Notes)
Ophthalmic
Fetal / Obstetrics[7]
Abdominal[1]PPPPPPPPPP3,5,6
PediatricPPPPPPPPPP3,5,6
Small Organ[2]PPPPPPPPPP3,5,6
Neonatal Cephalic
Adult Cephalic
Cardiac Adult & Pediatric
Peripheral VascularPPPPPPPPPP3,5,6
Musculo-skeletal ConventionalPPPPPPPPPP3,5,6
Musculo-skeletal SuperficialPPPPPPPPPP3,5,6
Other[4]
Exam Type, Means of Access
Transesophageal
Transrectal[8]
Transvaginal
Transuretheral
Intraoperative[8]
Intraoperative Neurological
Intravascular
Laparoscopic

N = new indication; P = previously cleared by FDA

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development, GYN

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle has decorative flourishes around its perimeter, giving it a classic and recognizable appearance. The logo is simple, yet distinctive, and is widely recognized as the symbol of the General Electric company.

Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with 8C Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationAnatomy/Region of InterestBMPWDopplerCWDopplerColorDopplerColor MDopplerPowerDopplerCombinedModes*HarmonicImagingCodedPulseOther(Notes)
Ophthalmic
Fetal / Obstetrics[7]
Abdominal[1]PPPPPPPPP5
PediatricPPPPPPPPP5
Small Organ[2]PPPPPPPPP5
Neonatal CephalicPPPPPPPPP5
Adult Cephalic
Cardiac Adult & Pediatric
Peripheral Vascular
Musculo-skeletal Conventional
Musculo-skeletal Superficial
Other[4]
Exam Type, Means of Access
Transesophageal
Transrectal[8]
Transvaginal
Transuretheral
Intraoperative[8]
Intraoperative Neurological
Intravascular
Laparoscopic

N = new indication; P = previously cleared by FDA

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development, GYN

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{17}------------------------------------------------

Image /page/17/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle has decorative swirls around the perimeter, giving it a classic and recognizable appearance. The logo is simple, yet distinctive, representing the brand identity of General Electric.

Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with 3Sp-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Mode of Operation
Clinical ApplicationAnatomy/Region of InterestBMPWDopplerCWDopplerColorDopplerColor MDopplerPowerDopplerCombinedModes*HarmonicImagingCodedPulseOther(Notes)
Ophthalmic
Fetal / Obstetrics[7]PPPPPPPPPP5,6
Abdominal[1]PPPPPPPPPP5,6
PediatricPPPPPPPPPP5,6
Small Organ[2]
Neonatal Cephalic
Adult CephalicPPPPPPPPPP
Cardiac Adult & PediatricPPPPPPPPPP
Peripheral Vascular
Musculo-skeletal Conventional
Musculo-skeletal Superficial
Other[4]
Exam Type, Means of Access
Transesophageal
Transrectal[8]
Transvaginal
Transuretheral
Intraoperative[8]
Intraoperative Neurological
Intravascular
Laparoscopic

N = new indication; P = previously cleared by FDA

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development, GYN

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{18}------------------------------------------------

Image /page/18/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a circular border. The color of the logo is blue, and the background is white.

Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with 6S-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Mode of Operation
Clinical ApplicationAnatomy/Region of InterestBMPW DopplerCW DopplerColor DopplerColor M DopplerPower DopplerCombined Modes*Harmonic ImagingCoded PulseOther (Notes)
Ophthalmic
Fetal / Obstetrics[7]PPPPPPPPPP
Abdominal[1]PPPPPPPPPP
PediatricPPPPPPPPPP
Small Organ[2]
Neonatal CephalicPPPPPPPPPPP
Adult Cephalic
Cardiac Adult & PediatricPPPPPPPPPPP
Peripheral Vascular
Musculo-skeletal Conventional
Musculo-skeletal Superficial
Other[4]
Exam Type, Means of Access
Transesophageal
Transrectal[8]
Transvaginal
Transuretheral
Intraoperative[8]
Intraoperative Neurological
Intravascular
Laparoscopic

N = new indication; P = previously cleared by FDA

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development, GYN

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{19}------------------------------------------------

Image /page/19/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The letters are white, creating a contrast against the blue background.

GE Healthcare 510(k) Premarket Notification Submission

Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with RIC5-9-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Mode of Operation
Clinical ApplicationAnatomy/Region of InterestBMPWDopplerCWDopplerColorDopplerColor MDopplerPowerDopplerCombinedModes*HarmonicImagingCodedPulseOther[Notes]
Ophthalmic
Fetal / Obstetrics[7]PPPPPPPPPP5,6
Abdominal[1]PPPPPPPPPP
Pediatric
Small Organ[2]
Neonatal Cephalic
Adult Cephalic
Cardiac Adult & Pediatric
Peripheral Vascular
Musculo-skeletal Conventional
Musculo-skeletal Superficial
Other[4]PPPPPPPPPP3,5,6
Exam Type, Means of Access
Transesophageal
Transrectal[8]PPPPPPPPPP3,5,6
TransvaginalPPPPPPPPPP3,5,6
Transuretheral
Intraoperative[8]
Intraoperative Neurological
Intravascular
Laparoscopic

N = new indication; P = previously cleared by FDA

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development, GYN

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{20}------------------------------------------------

Image /page/20/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a circular border. The color of the logo is blue, and the background is white.

GE Healthcare 510(k) Premarket Notification Submission

Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with 10C-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Mode of Operation
Clinical ApplicationAnatomy/Region of InterestBMPWDopplerCWDopplerColorDopplerColor MDopplerPowerDopplerCombinedModes*HarmonicImagingCodedPulseOther(Notes)
Ophthalmic
Fetal / Obstetrics[7]
Abdominal[1]
PediatricPPPPPPPP5,6
Small Organ[2]
Neonatal CephalicPPPPPPPP5,6
Adult Cephalic
Cardiac Adult & PediatricPPPPPPPP5,6
Peripheral VascularPPPPPPPP5,6
Musculo-skeletal ConventionalPPPPPPPP5,6
Musculo-skeletal SuperficialPPPPPPPP5,6
Other[4]
Exam Type, Means of Access
Transesophageal
Transrectal[8]
Transvaginal
Transuretheral
Intraoperative[8]
Intraoperative Neurological
Intravascular
Laparoscopic

N = new indication; P = previously cleared by FDA

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development, GYN

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{21}------------------------------------------------

Image /page/21/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle has decorative flourishes around its perimeter, giving it a classic and recognizable appearance. The logo is simple, yet distinctive, and is widely recognized as the symbol of the General Electric company.

Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with 6Tc-RS Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Mode of Operation
Clinical ApplicationAnatomy/Region of InterestBMPWDopplerCWDopplerColorDopplerColor MDopplerPowerDopplerCombinedModes*HarmonicImagingCodedPulseOther[Notes]
Ophthalmic
Fetal / Obstetrics[7]
Abdominal[1]
Pediatric
Small Organ[2]
Neonatal Cephalic
Adult Cephalic
Cardiac AdultPPPPPPPPPP
Peripheral Vascular
Musculo-skeletal Conventional
Musculo-skeletal Superficial
Other[4]
Exam Type, Means of Access
TransesophagealPPPPPPPPPP
Transrectal[8]
Transvaginal
Transuretheral
Intraoperative[8]
Intraoperative Neurological
Intravascular
Laparoscopic

N = new indication; P = previously cleared by FDA

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development, GYN

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{22}------------------------------------------------

Image /page/22/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle has decorative flourishes around its perimeter, giving it a classic and recognizable appearance. The logo is simple, yet distinctive, and is widely recognized as the symbol of the General Electric company.

GE Healthcare 510(k) Premarket Notification Submission

Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with BE9CS Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Mode of Operation
Clinical ApplicationAnatomy/Region of InterestBMPWDopplerCWDopplerColorDopplerColor MDopplerPowerDopplerCombinedModes*HarmonicImagingCodedPulseOther(Notes)
Ophthalmic
Fetal / Obstetrics[7]
Abdominal[1]PPPPPPPP3, 5, 6
PediatricPPPPPPPP3, 5, 6
Small Organ[2]
Neonatal Cephalic
Adult Cephalic
Cardiac Adult & Pediatric
Peripheral Vascular
Musculo-skeletal Conventional
Musculo-skeletal Superficial
Other[4]PPPPPPPP3, 5, 6
Exam Type, Means of Access
Transesophageal
Transrectal[8]PPPPPPPP3, 5, 6
TransvaginalPPPPPPPP3, 5, 6
Transuretheral
Intraoperative[8]
Intraoperative Neurological
Intravascular
Laparoscopic

N = new indication; P = previously cleared by FDA

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development, GYN

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{23}------------------------------------------------

Image /page/23/Picture/0 description: The image shows the General Electric (GE) logo, which consists of the letters 'GE' in a stylized script, enclosed within a circular border. The logo is colored in a light blue hue. The 'GE' letters are intertwined and have a flowing, cursive appearance. The circular border has decorative flourishes, adding to the logo's overall design.

GE Healthcare 510(k) Premarket Notification Submission

Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with L3-12-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Mode of Operation
Clinical ApplicationAnatomy/Region of InterestBMPWDopplerCWDopplerColorDopplerColor MDopplerPowerDopplerCombinedModes*HarmonicImagingCodedPulseOther(Notes)
Ophthalmic
Fetal / Obstetrics[7]PPPPPPPPP3, 5, 6
Abdominal[1]PPPPPPPPP3, 5, 6
PediatricPPPPPPPPP3, 5, 6
Small Organ[2]PPPPPPPPP3, 5, 6
Neonatal CephalicNNNNNNNNN3, 5, 6
Adult Cephalic
Cardiac Adult & Pediatric
Peripheral VascularPPPPPPPPP3, 5, 6
Musculo-skeletal ConventionalPPPPPPPPP3, 5, 6
Musculo-skeletal SuperficialPPPPPPPPP3, 5, 6
Other[4]
Exam Type, Means of Access
Transesophageal
Transrectal
Transvaginal
Transuretheral
Intraoperative[8]
Intraoperative Neurological
Intravascular
Laparoscopic

N = new indication; P = previously cleared by FDA

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development, GYN

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{24}------------------------------------------------

Image /page/24/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle has decorative flourishes around its perimeter, giving it a classic and recognizable appearance. The logo is simple, yet distinctive, and is widely recognized as the symbol of the General Electric company.

Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with S1-4-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Mode of Operation
Clinical ApplicationAnatomy/Region of InterestBMPW DopplerCW DopplerColor DopplerColor M DopplerPower DopplerCombined Modes*Harmonic ImagingCoded PulseOther [Notes]
Ophthalmic
Fetal / Obstetrics[7]PPPPPPPPPP5, 6
Abdominal [1]PPPPPPPPPP5, 6
PediatricPPPPPPPPPP5, 6
Small Organ [2]
Neonatal Cephalic
Adult CephalicPPPPPPPPPP5, 6
Cardiac Adult & PediatricPPPPPPPPPP5, 6
Peripheral Vascular
Musculo-skeletal Conventional
Musculo-skeletal Superficial
Other[4]
Exam Type, Means of Access
Transesophageal
Transrectal
Transvaginal
Transuretheral
Intraoperative[8]
Intraoperative Neurological
Intravascular
Laparoscopic

N = new indication; P = previously cleared by FDA

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development, GYN

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{25}------------------------------------------------

Image /page/25/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The letters are white, creating a contrast against the blue background.

T

Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with C1-6-D Transducer

Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationAnatomy/Region of InterestMode of OperationOther[Notes]
BMPWDopplerCWDopplerColorDopplerColor MDopplerPowerDopplerCombinedModes*HarmonicImagingCodedPulse°
Ophthalmic
Fetal/Obstetrics[7]P*P*P*P*P*P*P*P*P*P*P*3, 5, 6
Abdominal[1]P*P*P*P*P*P*P*P*P*P*P*3, 5, 6
PediatricP*P*P*P*P*P*P*P*P*P*P*3, 5, 6
Small Organ[2]
Neonatal Cephalic
Adult Cephalic
Cardiac Adult & Pediatric
Peripheral VascularP*P*P*P*P*P*P*P*P*P*P*P*3, 5, 6
Musculo-skeletal Conventional
Musculo-skeletal Superficial
Other[4]P*P*P*P*P*P*P*P*P*P*P*3, 5, 6
Exam Type, Means of Access
Transesophageal
Transrectal
Transvaginal
Transurethral
Intraoperative[8]
Intraoperative Neurological
Intravascular
Laparoscopic

N = new indication; P* = previously cleared by FDA on LOGIQ S8 K152195)

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development, GYN

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{26}------------------------------------------------

Image /page/26/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The blue color is a medium shade, and the letters are white.

Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with C3-10-D Transducer

Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:

Mode of Operation
Clinical ApplicationAnatomy/Region of InterestBMPWDopplerCWDopplerColorDopplerColor MDopplerPowerDopplerCombinedModes*HarmonicImagingCodedPulse*Other[Notes]
Ophthalmic
Fetal/Obstetrics[7]
Abdominal[1]
Pediatricp*p*p*p*p*p*p*p*5
Small Organ [2]
Neonatal Cephalicp*p*p*p*p*p*p*p*5
Adult Cephalic
Cardiac Adult & Pediatric
Peripheral Vascularp*p*p*p*p*p*p*p*5
Musculo-skeletal Conventional
Musculo-skeletal Superficial
Other[4]
Exam Type, Means of Access
Transesophageal
Transrectal
Transvaginal
Transurethral
Intraoperative[8]
Intraoperative Neurological
Intravascular
Laparoscopic

N = new indication; P* = previously cleared by FDA on LOGIQ S8 K152195)

[1] Abdominal includes Renal, GYN/Pelvic. Notes:

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development, GYN

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{27}------------------------------------------------

Image /page/27/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The letters are white, creating a contrast against the blue background.

Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with BE9CS-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Mode of Operation
Clinical ApplicationAnatomy/Region of InterestBMPW DopplerCW DopplerColor DopplerColor M DopplerPower DopplerCombined Modes*Harmonic ImagingCoded PulseOther (Notes)
Ophthalmic
Fetal / Obstetrics[7]
Abdominal[1]P*P*P*P*P*P*P*P*3, 5, 6
PediatricP*P*P*P*P*P*P*P*3, 5, 6
Small Organ[2]
Neonatal Cephalic
Adult Cephalic
Cardiac Adult & Pediatric
Peripheral Vascular
Musculo-skeletal Conventional
Musculo-skeletal Superficial
Other[4]P*P*P*P*P*P*P*P*3, 5, 6
Exam Type, Means of Access
Transesophageal
Transrectal[8]P*P*P*P*P*P*P*P*3, 5, 6
TransvaginalP*P*P*P*P*P*P*P*3, 5, 6
Transuretheral
Intraoperative[8]
Intraoperative Neurological
Intravascular
Laparoscopic

N = new indication; P* = previously cleared by FDA on LOGIQ S8 K152195)

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development, GYN

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{28}------------------------------------------------

Image /page/28/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle has decorative flourishes around its perimeter, giving it a classic and recognizable appearance. The logo is simple, yet distinctive, and is widely recognized as the symbol of the General Electric company.

Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with P8D Transducer

Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows:

Mode of Operation
Clinical ApplicationBMPW DopplerCW DopplerColor DopplerColor M DopplerPower DopplerCombined Modes*Harmonic ImagingCoded Pulse*Other [Notes]
Anatomy/Region of Interest
Ophthalmic
Fetal/Obstetrics[7]
Abdominal[1]
Pediatric
Small Organ [2]
Neonatal Cephalic
Adult Cephalic
Cardiac Adult
Cardiac Pediatric
Peripheral VascularP*
Musculo-skeletal Conventional
Musculo-skeletal Superficial
Other[4]
Exam Type, Means of Access
Transesophageal
Transrectal
Transvaginal
Transurethral
Intraoperative[8]
Intraoperative Neurological
Intravascular
Laparoscopic

N = new indication; P* = previously cleared by FDA on LOGIQ S8 K152195)

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development, GYN

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{29}------------------------------------------------

Image /page/29/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle has decorative swirls around the perimeter, giving it a classic and recognizable appearance. The logo is simple, yet distinctive, representing the brand identity of General Electric.

GE Healthcare 510(k) Premarket Notification Submission

Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with S2-5-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Mode of Operation
Clinical ApplicationAnatomy/Region of InterestBMPWDopplerCWDopplerColorDopplerColor MDopplerPowerDopplerCombined Modes*HarmonicImagingCodedPulseOther(Notes)
Ophthalmic
Fetal / Obstetrics[7]
Abdominal[1]NNNNNNNNNN5, 6
Pediatric
Small Organ[2]
Neonatal Cephalic
Adult CephalicNNNNNNNNNN5, 6
Cardiac Adult & PediatricNNNNNNNNNN5, 6
Peripheral Vascular
Musculo-skeletal Conventional
Musculo-skeletal Superficial
Other[4]
Exam Type, Means of Access
Transesophageal
Transrectal
Transvaginal
Transuretheral
Intraoperative[8]
Intraoperative Neurological
Intravascular
Laparoscopic

N = new indication; P = previously cleared by FDA

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development, GYN

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

System provides real-time 3D and 4D acquisition when used with special 4D probes

(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{30}------------------------------------------------

Image /page/30/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular shape. The logo is colored in a light blue hue.

510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

Date:January 20, 2016
Submitter:GE Healthcare9900 Innovation DrWauwatosa, WI 53226
Manufacturer:GE Ultrasound Korea, Ltd.9, Sunhwan-ro 214 beon-gil, Jungwon-gu,Seongnam-si, Gyeonggi-do, Republic of Korea
Primary Contact Person:Bryan BehnSr. Regulatory Affairs ManagerGE HealthcareT:(414)721-4214F:(414)918-8275
Secondary ContactPerson:Qingmeng ChenRegulatory Affairs SpecialistGE HealthcareGE Ultrasound Korea, Ltd.T: +82 31 740 6307F: +81-42-6272-2462
Device:Trade Name:LOGIQ S7 Expert
Common/Usual Name:LOGIQ S7 Expert
Classification Names:Class II
Product Code:Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550 90-IYMUltrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYODiagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX
Predicate Device(s):K141261 LOGIQ S7 Expert Diagnostic Ultrasound SystemK152195 LOGIQ S8 Diagnostic Ultrasound SystemK141639 Voluson S8 Diagnostic Ultrasound SystemK143452 Logiq P9 Diagnostic Ultrasound System
Device Description:The LOGIQ S7 Expert is a full featured, general purposediagnostic ultrasound system which consists of a mobileconsole approximately 530mm (Caster), 520mm (Keyboard),565mm (Monitor) wide, 865mm deep and 1760mm high thatprovides digital acquisition, processing and display capability.The user interface includes a computer keyboard, specialized

{31}------------------------------------------------

Image /page/31/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a circular design with the letters "GE" intertwined in the center. The letters and the surrounding circle are a light blue color. The logo is simple and recognizable, representing the company's brand.

controls, 10.1-inch LCD touch screen and color 23-inch LCD widescreen monitor.

The device is intended for use by a qualified physician for Intended Use: ultrasound evaluation of Obstetrics (Fetal); Abdominal (includes renal, GYN/Pelvic); Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transrectal; Transvaginal; Transesophageal and Intraoperative (abdominal, thoracic, vascular).

The LOGIQ S7 Expert employs the same fundamental scientific Technology: technology as its predicate devices

Determination of Substantial Equivalence:

Comparison to Predicate Devices

The following is an overview of the differences between the proposed LOGIO S7 Expert and the predicate LOGIO S7 Expert (K141261). The proposed LOGIQ S7 Expert has had several features and probes migrated to it currently cleared on other GE Ultrasound systems.

  • The systems are all intended for diagnostic ultrasound o imaging and fluid flow analysis.
  • o The LOGIQ S7 Expert and predicate LOGIQ S7 Expert system have the same clinical intended use.
  • o The LOGIO S7 Expert and predicate LOGIO S7 Expert system have the same imaging modes.
  • The systems are manufactured with materials which ● have been evaluated and found to be safe for the intended use of the device.
  • The systems have acoustic power levels which are below the applicable FDA limits.
  • The LOGIO S7 Expert and predicate LOGIO S7 Expert ● system have similar capability in terms of performing measurements, capturing digital images, reviewing and reporting studies.
  • The LOGIQ S7 and predicate systems have been ● designed in compliance with approved electrical and physical safety standards.
  • The probes in LOGIQ S7 Expert and predicate LOGIQ S7 Expert system are identical except for C1-6-D, C3-10-D, P8D, BE9CS-D which are the same on predicate LOGIO S8 (K152195); the S2-5-D probe is new and an

{32}------------------------------------------------

Image /page/32/Picture/0 description: The image shows the General Electric (GE) logo, which is a blue circle with the letters "GE" in a stylized script inside. The letters are also blue, and the circle has a white border. The logo is simple and recognizable, and it is associated with a well-known company.

incremental improvement to the existing 3Sp-D probe and is equivalent.

  • New software features added to the proposed LOGIQ 0 S7 Expert: STIC and OmniView are the same features cleared on predicate Voluson S8 device (K141639); Cardiac AFI is the same feature cleared on predicate Logiq S8 (K152195); Email2MMS (smart phone solution) is the feature cleared on predicate Logiq P9 (K143452).

Summary of Non-Clinical Tests:

The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic, and mechanical safety, and has been found to conform to applicable medical device safety standards. LOGIQ S7 Expert and its applications comply with voluntary standards:

    1. AAMI/ANSI ES60601-1, Medical Electrical Equipment – Part 1: General Requirements for Safety
    1. IEC60601-1-2,Medical Electrical Equipment Part 1-2: General Requirements for Safety – Collateral Standard: Electromagnetic Compatibility Requirements and Tests
    1. IEC60601-2-37, Medical Electrical Equipment Part 2-37: Particular Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment
    1. NEMA UD 3, Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment
    1. ISO10993-1, Biological Evaluation of Medical Devices- Part 1: Evaluation and Testing
    1. NEMA UD 2, Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment
    1. ISO14971, Application of risk management to medical devices: Second edition
    1. NEMA Digital Imaging and Communications in

{33}------------------------------------------------

Image /page/33/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular frame. The color of the logo is a light blue. The logo is simple and recognizable, representing the brand identity of General Electric.

Medicine (DICOM) Set. (Radiology)

The following quality assurance measures are applied to the development of the system:

  • Risk Analysis
  • Requirements Reviews ●
  • Design Reviews
  • Testing on unit level (Module verification) ●
  • . Integration testing (System verification)
  • Final Acceptance Testing (Validation) ●
  • o Performance testing (Verification)
  • Safety testing (Verification) o

Transducer materials and other patient contact materials are biocompatible.

Summary of Clinical Tests:

The subject of this premarket submission, LOGIQ S7 Expert, did not require clinical studies to support substantial equivalence.

  • GE Healthcare considers the LOGIQ S7 Expert to be as safe, as Conclusion: effective, and performance is substantially equivalent to the predicate device(s).

N/A