K Number
K170310

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-03-30

(57 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The implants of the Alians Clavicle range are dedicated to the fixation of fractures, mal-unions and osteotomies of the clavicle in adults.

Device Description

The Alians Clavicle range consists of plates and screws designed for fixation of fractures, mal-unions, non-unions and osteotomies of the clavicle in adults.

The implants of the Alians Clavicle range will be provided non sterile for sterilization by health care professionals prior to use or provided sterile by gamma sterilization. The instruments of the Alians Clavicle range will be provided non sterile for sterilization by health care professionals prior to use. Single use kits (Initial C) contain implants and instruments provided sterile by gamma sterilization. The Initial C kits contain implants (1 plate + screws), and instrumentation.

AI/ML Overview

The provided document is a 510(k) summary for the "Alians Clavicle range" device. This type of regulatory submission is for medical devices that are substantially equivalent to legally marketed predicate devices, not for novel AI/software devices requiring the kind of extensive performance criteria and studies outlined in your request.

Therefore, many of the requested categories (e.g., sample size for test set, number of experts for ground truth, MRMC studies, standalone algorithm performance, training set details) are not applicable to this submission. This document describes a traditional orthopaedic medical device (bone fixation plates and screws), not an AI/ML algorithm.

However, I can extract the acceptance criteria and study information that is relevant to this type of device and present it in your requested format.


Acceptance Criteria and Device Performance for Alians Clavicle Range

Acceptance Criteria CategoryDescription & Performance
Material EquivalenceThe device uses Titanium alloy Ti-6Al-4V, which conforms to ASTM F 136 and/or ISO 5832-3. This is considered substantially equivalent to the materials of the predicate devices.
Mechanical PerformanceAcceptance Criteria: The device must demonstrate mechanical safety and performance equivalent to or better than the predicate device. Reported Performance: Engineering analyses (non-clinical testing) were conducted in comparison with the Clavicle Locking Plating System of Newclip Technics (K100944). The analyses showed that the Alians Clavicle range is as strong or stronger than the predicate.
Biological Safety (LAL)Acceptance Criteria: The device must meet the endotoxin limit. Reported Performance: LAL testing was performed to demonstrate that the subject device meets the endotoxin limit of <20 EU/device.
Indications for Use (IFU)Acceptance Criteria: The IFU must be substantially equivalent to the predicate devices. Reported Performance: The Alians Clavicle range is indicated for the fixation of fractures, mal-unions, non-unions and osteotomies of the clavicle in adults, which aligns with the IFU of the predicate devices.
Design and Principles of OperationAcceptance Criteria: Design and operating principles must be substantially equivalent to the predicate devices. Reported Performance: The device consists of plates and screws for clavicle fixation, with principles of operation, technology, and design similar to the predicate devices.

Study Information:

  1. Sample size used for the test set and the data provenance: Not applicable. Performance was demonstrated through non-clinical (engineering) testing, not clinical studies involving patient data or a "test set" in the context of an AI/ML algorithm.

  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No "ground truth" in this sense was established as it's not an AI/ML device.

  3. Adjudication method for the test set: Not applicable.

  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: No. This is a physical medical device, not an AI/ML diagnostic tool. "No clinical studies were performed."

  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: No. This is a physical medical device.

  6. The type of ground truth used: For the mechanical performance, the "ground truth" was established through engineering analyses comparing the new device against the known and accepted performance characteristics of legally marketed predicate devices, measured using standardized tests. For biological safety, industry-standard LAL testing was used.

  7. The sample size for the training set: Not applicable. This is not an AI/ML device.

  8. How the ground truth for the training set was established: Not applicable. This is not an AI/ML device.

{0}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

NEWCLIP TECHNICS Mr. J.D. Webb The OrthoMedix Group, Inc. 1001 Oakwood Blvd. Round Rock, Texas 78681

March 30, 2017

Re: K170310

Trade/Device Name: Alians Clavicle range Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: January 26, 2017 Received: February 1, 2017

Dear Mr. Webb:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K170310

Device Name

Alians Clavicle range

Indications for Use (Describe)

The implants of the Alians Clavicle range are dedicated to the fixation of fractures, mal-unions and osteotomies of the clavicle in adults.

Type of Use (Select one or both, as applicable)

ا Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for NEWCLIP-TECHNICS. The logo consists of the letters "nct" inside of a circle, with the "t" in "nct" being a different color. Below the circle is the text "NEWCLIP-TECHNICS".

510 (k) Summary for the ALIANS CLAVICLE range 4.

In accordance with 21 CFR 807.92 of the Federal Code of Regulations, the following 510(k) Summary is submitted for the Alians Clavicle range.

Summary preparation date: January 24, 2017

1. Submitter:Contact Person:
NEWCLIP TECHNICSP.A. de la Lande Saint Martin45 rue des GarottièresF-44115 Haute-Goulaine - FranceTelephone: (33) 2 28 21 37 12J.D. WebbThe OrthoMedix Group, Inc.1001 Oakwood BlvdRound Rock, TX 78681Telephone: 512-388-0199
2. Trade name:Alians Clavicle range
Common Name:Fracture fixation plates and screws
Product code:HRS - Plate, Fixation, BoneHWC - Screw, Fixation, Bone
Classification Name:Single/multiple component metallicbone fixation appliances andaccessories. (21 CFR part. 888.3030)Smooth or threaded metallic bonefixation fastener. (21 CFR part.888.3040)

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Newclip-Technics. The logo consists of the letters "nct" in a circle, with a red arrow pointing upwards replacing the right side of the letter "t". Below the circle, the words "NEWCLIP - TECHNICS" are written in a sans-serif font.

3. Primary predicate:

  • Clavicle Locking Plating System of Newclip Technics (K100944) .

Additional predicate:

  • Congruent Bone Plate System of Acumed (K102998) .

Reference Device:

  • Activ Ankle Locking plate System of Newclip Technics (K143061) .
4. Description of the device:The Alians Clavicle range consists of platesand screws designed for fixation offractures, mal-unions, non-unions andosteotomies of the clavicle in adults.The implants of the Alians Clavicle rangewill be provided non sterile for sterilizationby health care professionals prior to use orprovided sterile by gamma sterilization.The instruments of the Alians Claviclerange will be provided non sterile forsterilization by health care professionalsprior to use.Single use kits (Initial C) contain implantsand instruments provided sterile by gammasterilization. The Initial C kits containimplants (1 plate + screws), andinstrumentation.
Materials:Titanium alloy Ti-6Al-4V (conform to ASTMF 136 and/or ISO 5832-3).
Function:The implants of Alians Clavicle range areindicated for fixation of fractures, mal-unions, non-unions and osteotomies of theclavicle in adults.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a logo for a company called "NEWCLIP-TECHNICS". The logo consists of a circle with the letters "nct" inside. The "t" has an upward-pointing arrow as part of the letter. Below the circle, the words "NEWCLIP-TECHNICS" are written in a sans-serif font.

5. Substantial equivalence claimed to predicate devices:

The Alians Clavicle range is substantially equivalent to the predicate devices in terms of intended use, design, materials used, mechanical safety and performance.

6. Indications for Use:

The implants of the Alians Clavicle range are dedicated to the fixation of fractures, mal-unions, non-unions and osteotomies of the clavicle in adults.

7. Non-clinical Test Summary:

  • . Engineering analyses have been conducted in comparison with Clavicle Locking Plating System of Newclip Technics (K100944).
  • LAL testing was performed to demonstrate that the subject device meets • the endotoxin limit of <20 EU/device.

The analyses show that the Alians Clavicle range is as strong or stronger than the predicate.

8. Clinical Test Summary:

No clinical studies were performed.

9. Conclusions Non-clinical and Clinical:

Newclip considers the Alians Clavicle range to be equivalent to the predicate devices listed above. This conclusion is based upon the device's similarities in principles of operation, technology, materials, and indications for use.

N/A