K Number
K170238

Validate with FDA (Live)

Device Name
BodyTom Elite
Date Cleared
2017-06-14

(139 days)

Product Code
Regulation Number
892.1750
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NL4000 BodyTom Elite CT system is intended to be used for x-ray computed tomography applications for anatomy that can be imaged in the 85cm aperture.

The CT system is intended to be used for both pediatric and adult imaging and as such has preset dose settings based upon weight and age. The CT images can be obtained either with or without contrast.

BodyTom Elite CT system can be used for low dose lung cancer screening. The screening must be performed in compliance with the approved and established protocols as defined by professional medical societies.

*Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011: 365:395-409) and subsequent literature, for further information.

Device Description

The BodyTom Elite is an improved version of BodyTom computed tomography (CT) system providing enhanced functionality. It still has the same high resolution, multi row, 85cm bore and 60cm field of view. The lightweight translating gantry consists of a rotating disk with a solid state x-ray generator, solid state detector array, collimator, control computer, communications link, power slip-ring, data acquisition system, reconstruction computer, power system, brushless DC servo drive system (disk rotation) and stepper drive system (translation). The power system consists of batteries which provide system power while unplugged from the charging outlet. The system has the necessary safety features such as the emergency stop switch, x-ray indicators, interlocks, patient alignment laser and 110% x-ray timer. The gantry has retractable rotating caster wheels and electrical drive system so the system can be moved easily to different locations.

AI/ML Overview

Here's an analysis of the acceptance criteria and study information provided, structured as requested:


1. Table of Acceptance Criteria and Reported Device Performance

The provided document primarily focuses on demonstrating substantial equivalence to a predicate device rather than explicitly stating acceptance criteria and their corresponding performance values in a direct numerical table for the entire device. However, it does list several image quality parameters that are important for Low Dose CT Lung Cancer Screening (LDCT LCS) and the general use of CT, along with reasons for their inclusion. The performance is stated as meeting all image quality criteria used for testing the predicate device.

Image Quality Parameters Relevant to LDCT LCS and General CT Use (Implicit Acceptance Criteria)

Imaging ParameterReason for Inclusion / Implicit CriterionReported Device Performance
Modulation Transfer Function (MTF)Describes the size of the smallest object that can be seen with large difference in CT value. Needs at least 2mm sampling rate or resolution of at least 5 lp/cm; 8 lp/cm recommended for detecting 4mm objects."The BodyTom Elite scanner successfully demonstrated that it has comparable image quality as the predicate device and meets all the image quality criteria that are used for testing the BodyTom as it passed all QA requirements." (Implies it meets the 5-8 lp/cm recommendation). Explicitly, the Spatial Resolution MTF at 0% is given as 17.5 lp/cm (matches predicate).
Slice ThicknessDetermines the smallest size that can be seen in scan direction. Needs to be thin enough to allow identification of objects of at least 4.0 mm in diameter. ACR recommends under 2.5 mm."The BodyTom Elite scanner successfully demonstrated that it has comparable image quality as the predicate device and meets all the image quality criteria that are used for testing the BodyTom as it passed all QA requirements." (Implies it meets the <2.5mm ACR recommendation).
CT Number (accuracy and uniformity)Helps determine the nature of the nodule; not affected by dose since filter on tube does not change."The BodyTom Elite scanner successfully demonstrated that it has comparable image quality as the predicate device and meets all the image quality criteria that are used for testing the BodyTom as it passed all QA requirements."
CT Number LinearityThe quality of the X-ray beam is the same at high and low dose, so CT linearity should not be affected by low dose scanning."The BodyTom Elite scanner successfully demonstrated that it has comparable image quality as the predicate device and meets all the image quality criteria that are used for testing the BodyTom as it passed all QA requirements."
Image NoiseBackground noise increases with LDCT LCS which affects detectability of lung nodules. Low contrast analysis should show good detectability."Our low contrast analysis show good detectability as the lung nodules are surrounded by air. CNR results are included in Performance Testing (section 18)." (This implies acceptable noise levels enabling good detectability).
Noise Power Spectrum (NPS)Changes in noise texture may impact lung nodule detection capabilities. Will increase with low dose scanning, but NPS is minimally affected."The BodyTom Elite scanner successfully demonstrated that it has comparable image quality as the predicate device and meets all the image quality criteria that are used for testing the BodyTom as it passed all QA requirements." (Implies NPS is minimally affected as expected).
Contrast to Noise (CNR)Important when background noise is close to density of lung nodules. Typically high for lung nodules surrounded by air, even for low dose scanning."CNR results are included in Performance Testing (section 18)." (Explicitly mentions results are provided, implying acceptable performance).
Scanner SpeedACR recommends a maximum of 15 seconds scan and pitches that vary from 0.7 to 1.5."The BodyTom Elite scanner successfully demonstrated that it has comparable image quality as the predicate device and meets all the image quality criteria that are used for testing the BodyTom as it passed all QA requirements." (Implies it meets the ACR speed recommendations).
Safety StandardsCompliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-44, NEMA XR 25, NEMA XR 29, ISO 14971, 21 CFR §1020.30/§1020.33, and FDA Quality System Regulations and ISO 13485.The device "is designed and manufactured to comply with the FDA Quality System Regulations and ISO 13485 requirements" and "is in conformance with all applicable parts of the following FDA Recognized Consensus Standards: [listed standards]". EMC, Electrical, and Mechanical Safety were ETL Tested.

2. Sample Size Used for the Test Set and Data Provenance

The document describes non-clinical testing primarily using phantoms. No human subject data (patients or test sets for diagnostic accuracy) is mentioned for the BodyTom Elite device's specific performance evaluation in this 510(k) summary.

  • Sample Size: Not applicable as no human test sets are used. Phantom studies do not typically operate on "sample size" in the same way clinical studies do; they involve specific measurements on physical phantoms.
  • Data Provenance: Not applicable. The testing was internal to NeuroLogica Corporation ("tested internally", "validated internally").

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • This information is not provided as the evaluation primarily relies on phantom studies and comparison to a predicate device's established performance, rather than a diagnostic accuracy study requiring expert ground truth for interpretation.

4. Adjudication Method (e.g. 2+1, 3+1, none) for the Test Set

  • This information is not provided and is not applicable to the non-clinical, phantom-based testing described.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not reported for the BodyTom Elite device within this 510(k) summary.
  • The document states that "Published literature, clinical trials including the NLST (N Engl J Med 2011; 365:395-409) and governmental review reflect understanding by the medical community that the specific use of LDCT LCS is a subset of the general use of CT rather than a new intended use." This refers to the evidence base for LDCT LCS in general, not a comparative effectiveness study involving the BodyTom Elite device itself.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

  • Yes, the evaluation conducted was a form of standalone performance assessment, focused on the technical imaging parameters of the CT system itself. It assessed the "algorithm only" in the sense that the CT system's reconstruction algorithms and hardware capabilities were measured directly using phantoms, without human interpretation for diagnostic accuracy.
  • The document states: "Specific image quality metrics were compared between the subject device (BodyTom Elite) and the previously cleared reference device of Philips Multislice CT System... The imaging parameters for the subject device were measured using standard IQ phantom and compared to the IQ results from the reference device."

7. Type of Ground Truth Used

  • The ground truth for the image quality measurements was based on physical phantoms with known properties. These phantoms are designed to have quantifiable features (e.g., specific object sizes, densities, spatial frequencies) that allow for objective measurement of CT system performance against established standards and expected values.

8. Sample Size for the Training Set

  • This information is not applicable as the BodyTom Elite is a CT imaging system, not an AI/ML algorithm that requires a "training set" in the context of machine learning. The "LDCT LCS and MAR protocols have been tested internally and can be used with the filtered back projection reconstruction algorithms for BodyTom Elite," suggesting these are pre-defined protocols and features, not learned from a training set.

9. How the Ground Truth for the Training Set Was Established

  • This information is not applicable as the BodyTom Elite is not an AI/ML algorithm requiring a training set with associated ground truth as typically understood in that context.

{0}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines beneath them, resembling a symbol or emblem.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

NeuroLogica Corporation, a Subsidiary of Samsung Electronics Ninad Gujar Director, Regulatory Affairs & Ouality Assurance 14 Electronics Avenue Danvers, Massachusetts 01923

June 14, 2017

Re: K170238

Trade/Device Name: BodyTom Elite Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: JAK Dated: May 8, 2017 Received: May 17, 2017

Dear Ninad Gujar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Michael D. O'Hara
For

Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K170238

Device Name BodyTom Elite

Indications for Use (Describe)

The NL4000 BodyTom Elite CT system is intended to be used for x-ray computed tomography applications for anatomy that can be imaged in the 85cm aperture.

The CT system is intended to be used for both pediatric and as such has preset dose settings based upon weight and age. The CT images can be obtained either with or without contrast.

BodyTom Elite CT system can be used for low dose lung cancer screening must be performed in compliance with the approved and established protocols as defined by professional medical societies.

  • Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirements of 21 CFR § 807.92

Date: January 24, 2017

Submitter:

NeuroLogica Corporation, a subsidiary of Samsung Electronics Co., Ltd 14 Electronics Avenue, Danvers, MA 01923

Contact:

Dr. Ninad Gujar Director, Regulatory Affairs & Quality Assurance Telephone: 978-564-8503 Fax: 978-564-8677 E-mail: ngujar@neurologica.com

Device Name:

Trade Name:BodyTom Elite
Device Model:NL4000
Classification Name:Computed Tomography X-ray System
Product Code:JAK
Device Classification:Class II (per 21 CFR § 892.1750)

Predicate Device:

Trade Name:BodyTom
510(k) number:K102677 cleared 03/24/2011
Manufacturer:NeuroLogica Corporation, a subsidiary of SamsungElectronics Co., Ltd (same as that of proposed device)
Classification Name:Computed Tomography X-ray System
Product Code:JAK
Device Classification:Class II (per 21 CFR § 892.1750)
Reference Device:
Trade Name:Philips Multislice CT System with Low Dose CT Lung
Cancer Screening
510(k) number:K153444 cleared 04/08/2016
Manufacturer:Philips Medical Systems (Cleveland), Inc
Classification Name:Computed Tomography X-ray System

{4}------------------------------------------------

Product Code:JAK
Device Classification:Class II (per 21 CFR § 892.1750)

Device Description:

The BodyTom Elite is an improved version of BodyTom computed tomography (CT) system providing enhanced functionality. It still has the same high resolution, multi row, 85cm bore and 60cm field of view. The lightweight translating gantry consists of a rotating disk with a solid state x-ray generator, solid state detector array, collimator, control computer, communications link, power slip-ring, data acquisition system, reconstruction computer, power system, brushless DC servo drive system (disk rotation) and stepper drive system (translation). The power system consists of batteries which provide system power while unplugged from the charging outlet. The system has the necessary safety features such as the emergency stop switch, x-ray indicators, interlocks, patient alignment laser and 110% x-ray timer. The gantry has retractable rotating caster wheels and electrical drive system so the system can be moved easily to different locations.

Indications for Use:

The BodyTom Elite CT system is intended to be used for x-ray computed tomography applications for anatomy that can be imaged in the 85cm aperture.

The CT system is intended to be used for both pediatric and adult imaging and as such has preset dose settings based upon weight and age. The CT images can be obtained either with or without contrast.

BodyTom Elite CT system can be used for low dose lung cancer screening. The screening must be performed in compliance with the approved and established protocols as defined by professional medical societies.

*Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011: 365:395-409) and subsequent literature, for further information.

Substantial Equivalence:

NeuroLogica has modified the cleared BodyTom (K102677) within our design controls to include few technology improvements and features like automatic exposure control, metal artifact reduction, low dose lung cancer screen protocol, wireless access point and tilt sensor for floor qualification. Both the predicate and reference devices have the same intended use as the new device. The low dose CT lung cancer screening protocol has been previously cleared on Philips Multislice CT System with Low Dose CT Lung Cancer Screening (K153444) and BodyTom Elite has a similar clinical protocol.

{5}------------------------------------------------

BodyTom Elite incorporates the same fundamental operating principles as the existing marketed products BodyTom (K102677) and Philips Multislice CT systems (K153444).

NeuroLogica Corporation believes that the BodyTom Elite system, for its intended use, is of comparable type in design, material, functionality, technology and is substantially equivalent to the cleared predicate device – BodyTom (K102677).

  • . Design: The BodyTom Elite is similar in general design principle to both the above listed CT systems. Specifically, it shares all of the control system designs and features of the BodyTom.
  • . Material: The BodyTom Elite uses the same material as the predicate device (BodyTom) such as solid state detectors, x-ray generator, slip ring, data acquisition ICs, rotational bearing, and motion control systems.

The following technology differences exist between the subject device (BodyTom Elite) and the previously cleared predicate device (BodyTom).

  • . Automatic exposure control: Automatic exposure control (AEC) was not offered on the predicate device (K102677) at the time of 510(k) clearance but introduced as a device modification through a software/firmware upgrade. AEC feature has been verified as part of IEC 60601-2-44, clause 203.102 in BodyTom Elite similar to the predicate device.
  • . Low dose lung scanning protocol: No hardware design changes were required for this feature but specific clinical protocols were generated based on the AAPM guidelines and published clinical literature.
  • . Metal artifact reduction: The streak artifacts around metal leads and screws can be reduced using metal artifact reduction. This has been tested in the CT system and validated internally.
  • = Tilt sensor for floor qualification: The tilt sensor measures the flatness of the scan floor. The tilt sensor function measures the variation of the scan trajectory which is used to determine the flatness of the scan floor. The tilt sensor has been tested on multiple floors as part the validation protocol and does not raise any new safety or effectiveness concerns.
  • Wireless access point: The wireless access point provides better security and . stability of wireless communication in BodyTom Elite by creating a standalone

{6}------------------------------------------------

wireless network for the scanner compared to the existing Ad-Hoc wireless environment. Physically, the wireless access point is part of the workstation and has undergone EMC testing. This wireless access point has been tested internally for connectivity.

  • . Contrast capability: BodyTom (predicate device) had the ability for axial, helical and dynamic scanning for CT Angiography (CTA) and CT Perfusion (CTP) applications, including bolus tracking. The contrast imaging capability is added to the expanded indications for BodyTom Elite (subject device) and additional performance testing data is provided for this purpose. The BodyTom Elite scanner does not control the dose, rate or route of administration for the contrast agent.
  • . NEMA XR-29: BodyTom Elite is XR-29 compliant and as such includes four key features of CT equipment that enable optimization or management of radiation dose delivery - dose structured reporting, CT dose check, AEC and pediatric and adult reference protocols.

The differences noted above raise no new issues of safety or effectiveness based on all testing performed. The below table highlights comparison of technological characteristics with the predicate device.

SpecificationBodyTom Elite(subject of this 510(k))BodyTom(predicate deviceK102677)
Indications for UseThe NL4000 BodyTomsystem is intended to beused for x-ray computedtomography applicationsthat produce crosssectional images foranatomy that can beimaged in the 85cmaperture.The BodyTom system isintended to be used forboth pediatric and adultimaging and as such haspreset dose settings basedThe NL4000 BodyTom isintended to be used for x-ray computedtomography applicationsfor anatomy that can beimaged in the 85cmaperture.
SpecificationBodyTom Elite(subject of this 510(k))BodyTom(predicate deviceK102677)
upon weight and age. TheCT images can beobtained either with orwithout contrast.
BodyTom CT system canbe used for low dose lungcancer screening.
Aperture (cm)8585
Image Field Of View (mm)6060
Detector MaterialSolid State CdWO4Solid State CdWO4
Detector Coverage (mm)4040
Slice Configurations3232
Spatial ResolutionMTF at 0% (lp/cm)17.517.5
X-ray Tube TypeRotating AnodeRotating Anode
Heat Storage (MHU)3.53.5
CoolingOil to AirOil to Air
Tube Current (mA)10 to 30010 to 300
Tube Voltage Range (kV)80 to 14080 to 140
Maximum Rotation Speed (s)1.01.0
Gantry Weight (kg)12001200
SpecificationBodyTom Elite(subject of this 510(k))BodyTom(predicate deviceK102677)
Mobile / StationaryMobileMobile
Battery / Wall powerBatteryBattery
Input VoltageSingle phase110-240 voltsSingle phase110-240 volts
Max Input Power (kVA)3.6 kW3.6 kW
PACS DICOM 3.0YesYes
2D Scout ScanYesYes
Bolus TrackingYesYes
Axial ScanYesYes
Helical ScanYesYes
Dynamic ScanYesYes
2D ViewingYesYes
MPR ViewingYesYes
3D ViewingYesYes
Gantry TiltNoNo
Scan Range (cm)200200
Scan LocalizerLaserLaser
Cardiac/Respiratory GatingNoNo
Patient TableNoNo
SpecificationBodyTom Elite(subject of this 510(k))BodyTom(predicate deviceK102677)
Image recon speed max (imagesper second)1616
X-ray Warning LightsYesYes
110% X-ray TimerYesYes
Emergency StopYesYes
Operator X-ray On SwitchYesYes
Quality Test PhantomYesYes
Quality Test ReportYesYes
X-ray FilterBowtieBowtie
Administrator PrivilegesYesYes
Dose DisplayYesYes
Dose Report / AuditYesYes
Protocol Override ProtectionYesYes
Dose CheckYesYes
Pediatric ProtocolsYesBy Age/weightYesBy Age/weight
Automatic Exposure ControlYesNo
BiocompatibilityN/AN/A
SterilityN/AN/A
Chemical SafetyN/AN/A
SpecificationBodyTom Elite(subject of this 510(k))BodyTom(predicate device K102677)
EM EmissionsETL TestingETL Testing
Electrical Safety (IEC 60601-1)ETL TestingETL Testing
Mechanical Safety (IEC 60601-1)ETL TestingETL Testing
Where UsedMobile, Fixed Radiology,ICU, ED, OR, ProtonTherapy, HDR suite,Clinic, Office(Mobile) Radiology, ICU,ED, OR, Clinic, Office
Anatomical SiteWhole bodyWhole body

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

Specific image quality metrics were compared between the subject device (BodyTom Elite) and the previously cleared reference device of Philips Multislice CT System with Low Dose CT Lung Cancer Screening (K153444), specifically the Philips Brilliance 16 Family (K012009) was used as the point of comparison. The subject device and the reference device have similar technological characteristics and fundamental operating principles.

The LDCT LCS and MAR protocols have been tested internally and can be used with the filtered back projection reconstruction algorithms for BodyTom Elite.

Intended Use Discussion:

In accordance with FDA guidance for Guidance for Industry on General/Specific Intended Use and the presentation FDA/MITA Meeting on Low Dose CT Lung Cancer Screening, the modified indication of use does not alter the intended use for the legally marketed predicate and reference devices.

The specific use of low dose CT (LDCT) lung cancer screening (LCS) was evaluated based on the above criteria:

    1. LDCT LCS does not introduce new risks not normally associated with the general use of the CT device. The risks are fundamentally the same as those associated with the use of CT.
    1. The specific use of LDCT LCS does not impact public health to a significantly greater degree than the general diagnostic use of the CT device.

{11}------------------------------------------------

    1. Published literature, clinical trials including the NLST (N Engl J Med 2011; 365:395-409) and governmental review reflect understanding by the medical community that the specific use of LDCT LCS is a subset of the general use of CT rather than a new intended use.
    1. The performance used to evaluate general use of CT has not changed and can also be applied to the specific use of LDCT LCS.
    1. LDCT LCS does not perform any treatment.
    1. No adjunctive products required for LDCT using CT of the target population.
    1. No device modifications / design changes were required to the general use of the CT device and specific clinical protocols were generated based on the AAPM guidelines and published clinical literature. The qualified CT system continues to comply with the applicable US and international safety and performance standards such as 21 CFR, Chapter I, Subchapter J - Radiological Health, NEMA, DICOM and IEC standards.

NeuroLogica believes that there is more than sufficient scientific and medical evidence in the published literature and coverage decisions of the safety and effectiveness of LDCT LCS when performed with a pre-defined high risk group, according to a pre-defined screening program that includes dose targets.

Non Clinical testing:

All components of the BodyTom Elite CT system that are subject to Federal Diagnostic Equipment Performance Standard and applicable regulations of 21 CFR §1020.30 and §1020.33 are certified to meet those requirements. To minimize electrical, mechanical and radiation hazards. NeuroLogica adheres to recognized and established industry practices.

BodyTom Elite system is designed and manufactured to comply with the FDA Quality System Regulations and ISO 13485 requirements. The device is in conformance with all applicable parts of the following FDA Recognized Consensus Standards:

  • I IEC 60601-1, Medical Electrical Equipment - Part 1: General Requirements for Safety
  • . IEC 60601-1-2, Medical electrical equipment - Part 1-2: General Requirements Basic Safety and Essential Performance - - Collateral Standard: for Electromagnetic Compatibility. Requirements and Test

{12}------------------------------------------------

  • . IEC 60601-1-3, Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment
  • IEC 60601-2-28, Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis
  • . IEC 60601-2-44, Medical Electrical Equipment - Part 2-44: Particular Requirements for Basic Safety and Essential Performance of X-Ray Equipment for Computed Tomography
  • NEMA XR 25: Computed Tomography Dose Check
  • NEMA XR 29: Standard Attributes on CT Equipment Related to Dose . Optimization and Management
  • ISO 14971 Medical devices Application of risk management to medical devices

In addition to conformance to the above harmonized standards, BodyTom Elite quality assurance activities include the following:

  • Risk analysis and mitigation
  • Software verification and validation testing
  • 트 System verification and validation testing
  • I Image quality tests
  • Testing at unit level 트

The software contained in the proposed device has been developed & tested in the software development procedure on IEC 62304, and the FDA guidance for Content of Premarket Submissions for Software Contained in Medical Devices. Software is critical to the operation of the BodyTom Elite CT scanner and a malfunction or design flaw in the software could result in delay in delivery of appropriate medical care. As such, the risk management analysis identified potential hazards which were controlled and mitigated during development of BodyTom Elite. The level of concern for the software is Moderate because a failure, latent flaw, or delayed information could indirectly result in minor injury to the patient. The verification & validation testing ensured the safety and effectiveness of BodyTom Elite.

Image quality metrics such as uniformity, slice width, low and high contrast resolution, radiation metrics, and modulation transfer function were measured utilizing phantom

{13}------------------------------------------------

image quality tests in accordance with the equipment performance standards for diagnostic x-ray systems administered by the FDA. In addition the beam profile and the noise power spectrum were also measured. The BodyTom Elite scanner successfully demonstrated that it has comparable image quality as the predicate device and meets all the image quality criteria that are used for testing the BodyTom as it passed all QA requirements.

The image quality metrics utilized to test general CT use are also applicable for LDCT LCS. The imaging parameters for the subject device were measured using standard IQ phantom and compared to the IQ results from the reference device. The imaging parameters that were tested includes the CT number linearity, slice thickness, uniformity and noise level, the low-contrast resolution and the high contrast resolution. For LDCT LCS, in order to measure nodules of at least 4 mm in diameter, the important parameters are listed below:

ImagingParameterReason for Inclusion
ModulationTransferFunction(MTF)The MTF describes the size of the smallest object that can be seenwith large difference in CT value, e.g., soft tissue-air, soft tissue-bone and bone-air. In order to detect a 4 mm object the scanner needsto have at least 2mm sampling rate or resolution of at least 5 1p/cm.A resolution of at least 8 lp/cm is recommended.
SliceThicknessThe slice thickness determines the smallest size that can be seen inscan direction. The CT number is averaged over the volume coveredby the slice width. The slice need to be wide enough not to have ahigh noise and thin enough to allow the identification of objects of atleast 4.0 mm in diameter. The ACR recommend a slice thicknessunder 2.5 mm.
CT number(accuracy anduniformity)The measured CT number will help determine the nature of thenodule.
CT numberlinearityThe dose does not affect the CT number, since the filter on the tubedoes not change. The quality of the X-ray beam is the same at highand low dose. As such, the CT linearity should not be affected bylow dose scanning.
Image noiseThe background noise increases with LDCT LCS which affects thedetectability of the lung nodules based on their size. Our low contrastanalysis show good detectability as the lung nodules are surroundedby air. CNR results are included in Performance Testing (section 18).

{14}------------------------------------------------

ImagingParameterReason for Inclusion
Noise PowerSpectrum(NPS)Similar to image noise, changes in noise texture may impact lungnodule detection capabilities as the image noise varies within thefield of view. The noise will increase with low dose scanninghowever the NPS is minimally affected.
Contrast toNoise (CNR)The CNR is important when the background noise is close to thedensity of the lung nodules. In general, since lung nodules aresurrounded by air, the CNR is typically high even for low dosescanning.
Scanner SpeedThe ACR recommend a maximum of 15 seconds scan and pitchesthat varies from 0.7 to 1.5.

Conclusion:

Based upon the above considerations, NeuroLogica Corporation, subsidiary of Samsung Electronics, believes that the BodyTom Elite Computed Tomography System is of comparable type in design, material, functionality, technology and is, for its intended use, substantially equivalent to the previous version and the predicate device: BodyTom (K102677).

The proposed modifications to BodyTom do not result in any new safety or effectiveness concerns. The BodyTom Elite CT system performs as well in its intended use as similar devices currently on the market.

§ 892.1750 Computed tomography x-ray system.

(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.