K Number
K170219

Validate with FDA (Live)

Date Cleared
2017-07-20

(176 days)

Product Code
Regulation Number
880.2910
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Microlife Digital Infrared Ear Thermometer, Model IR1DR1-1 device, is an electronic clinical thermometer using an infrared sensor to detect body temperature from the auditory canal in the neonatal, pediatric and adult population used in the home setting.

Device Description

The Microlife Digital Infrared Ear Thermometer, Model IR1DR1-1 is an electronic thermometer using an infrared sensor (thermopile) to detect body temperature from the auditory canal. Its operation is based on measuring the natural thermal radiation emanating from the tympanic membrane and the adjacent surfaces.

Mode of operation:

The unit measures the infrared energy emitting from the middle ear and the surrounding tissue. This energy is absorbed by lenses and converted into temperature values.

The Microlife Digital Infrared Ear thermometer, Model IR1DR1-1, consists of the following parts: a) Thermopile Sensor b) Application-Specific Integrated Circuit d) Lens e) LCD and Backlight f) 2 Keys (Start key, O/I key) g) 1 battery 3.0V

AI/ML Overview

Here's an analysis of the provided text regarding the Microlife Digital Infrared Ear Thermometer, Model IR1DR1-1, based on your requested criteria:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the device's accuracy are based on the ASTM E1965-98 (2009) standard for infrared thermometers. The reported performance for accuracy matches these criteria.

Acceptance Criteria (from ASTM E1965-98 (2009))Reported Device Performance (Blackbody Test)
<35.5°C ±0.3°C<35.5°C ±0.3°C
35.5°C-42.0°C ±0.2°C35.5°C-42.0°C ±0.2°C
>42.0°C ±0.3°C>42.0°C ±0.3°C

Other acceptance criteria implicitly met by the "Passed all testing requirements" statements for the listed standards (e.g., electrical safety, EMC, biocompatibility) are:

  • Electrical Safety: Complies with IEC 60601-1: 2005 and A1:2012, ANSI/AAMI ES60601-1:2005/(R2012) and A1:2012, C1:2009/(R)2012 and A2:2010/(R2012), IEC 60601-1-11:2010.
  • Electromagnetic Compatibility (EMC): Complies with IEC 60601-1-2: 2014.
  • Biocompatibility: Complies with ISO 14971: 2007 (risk management), AAMI/ANSI/ISO 10993-1: 2010, AAMI/ANSI/ISO 10993-5: 2010, AAMI/ANSI/ISO 10993-10: 2010, AAMI/ANSI/ISO 10993-12: 2012 (biocompatibility standards).
  • Software Verification and Validation: Complies with FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices dated May 11, 2005.

2. Sample Size Used for the Test Set and Data Provenance

The document mentions "Controlled human clinical studies were conducted in accordance with ASTM E1965-98, IEC80601-2-56 Test Report." However, specific sample sizes (number of subjects/measurements) for the clinical test set are not provided in the document. The data provenance (e.g., country of origin, retrospective or prospective) is also not explicitly stated, though the nature of "Controlled human clinical studies" implies a prospective design.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the document. For a clinical thermometer, the "ground truth" (reference temperature) is typically established by another calibrated reference thermometer, not by human experts adjudicating readings.

4. Adjudication Method for the Test Set

This information is not applicable in the context of a clinical thermometer's performance evaluation, as "adjudication" typically refers to the resolution of disagreements among expert reviewers for image-based diagnostics. The ground truth for temperature measurement is an objective reading from a reference device.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers interpret output (e.g., images), which is not the case for a digital ear thermometer. The comparison is against a reference temperature.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the device performance was evaluated in a standalone manner. The "Accuracy Tested (blackbody)" refers to testing the device in a controlled environment against known temperature sources without human interpretation influencing the measurement itself. The clinical studies also evaluate the device's ability to measure temperature in humans without human intervention in the device's measurement process, though a human operates the device.

7. The Type of Ground Truth Used

For the accuracy testing, the ground truth was established using blackbody radiators (controlled temperature sources) for in-vitro testing. For clinical studies, the ground truth would typically be established by simultaneously taking readings with a calibrated reference thermometer – most likely a traditional clinical thermometer known for high accuracy or another accepted medical standard. The document states "Clinical data was presented evaluating clinical uncertainty and clinical repeatability," which implies comparison against a reference.

8. The Sample Size for the Training Set

The document describes the device as a physical thermometer using an infrared sensor and an Application-Specific Integrated Circuit (ASIC). It does not mention any machine learning or AI algorithms that would require a "training set" in the conventional sense. The device's operation is based on physical principles of infrared measurement and conversion to temperature values. Therefore, a "training set" as understood in AI/ML is not applicable.

9. How the Ground Truth for the Training Set Was Established

As explained in point 8, a "training set" is not applicable for this device.

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The caduceus is a symbol of medicine and healing. The logo also includes the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement around the caduceus symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 20, 2017

Microlife Intellectual Property GmbH % Susan Goldstein-Falk MDI Consultants. Inc. 55 Northern Boulevard Suite 200 Great Neck, New York 11021

Re: K170219

Trade/Device Name: Microlife Digital Infrared Ear Thermometer, Model IR1DR1-1 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: June 1, 2017 Received: June 1, 2017

Dear Ms. Goldstein-Falk:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Tara A. Ryan -S

for Lori Wiggins Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K170219

Device Name

Microlife Digital Infrared Ear Thermometer, Model IR1DR1-1

Indications for Use (Describe)

The Microlife Digital Infrared Ear Thermometer, Model IR 1 device, is an electronic clinical thermometer using an infrared sensor to detect body temperature from the auditory canal in the neonatal, pediation used in the home setting.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
☐ Research Use (Part 21 CFR 201. Subject to Review)☒ Gene Therapy Studies (21 CFR 201. Subject to Review)
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

__ Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K170219 510(K) SUMMARY

Manufacturer's Name:Microlife Intellectual Property GmbH, SwitzerlandEspenstrasse 1399443 Widnau / Switzerland
Corresponding Official:Mr. Gerhard FrickVice President of Technical and Service
Telephone Number:Microlife Intellectual Property GmbH, Switzerland+41 79 216 0070
Email:gerhard.frick@microlife.ch
Preparation Date:July 13, 2017
Trade Name:Microlife Digital Infrared Ear Thermometer, Model IR1DR1-
Common or Usual Name:Clinical Electronic Thermometer
Regulation Name:Regulation Number:Product Code:Device Class:Clinical Electronic Thermometer21 CFR 880.2910FLLClass II
Primary Predicate Device:K034023 Microlife Digital Infrared Ear thermometer, ModelIR1DE1-1

Device Description:

The Microlife Digital Infrared Ear Thermometer, Model IR1DR1-1 is an electronic thermometer using an infrared sensor (thermopile) to detect body temperature from the auditory canal. Its operation is based on measuring the natural thermal radiation emanating from the tympanic membrane and the adjacent surfaces.

Mode of operation:

The unit measures the infrared energy emitting from the middle ear and the surrounding tissue. This energy is absorbed by lenses and converted into temperature values.

The Microlife Digital Infrared Ear thermometer, Model IR1DR1-1, consists of the following parts: a) Thermopile Sensor b) Application-Specific Integrated Circuit d) Lens

{4}------------------------------------------------

e) LCD and Backlight f) 2 Keys (Start key, O/I key) g) 1 battery 3.0V

The new Model IR1DR1-1 has the same intended use and temperature measurement fundamental algorithm as the predicate device 510(k) K034023 Microlife Model IR1DE1-1.

Indications For Use

The Microlife Digital Infrared Ear Thermometer, Model IR1DR1-1 device, is an electronic clinical thermometer using an infrared sensor to detect body temperature from the auditory canal in the neonatal, pediatric and adult population used in the home setting.

Substantial Equivalence Discussion

Both devices use infrared technology to measure and monitor the body temperature by the site of ear canal.

Microlife Digital Infrared Ear Thermometer IR1DR1-1 (cover free series products) has been compared to the cleared "Microlife Digital Infrared Ear Thermometer IR1DE1-1" (510(k) K#034023) as a reference for substantial equivalence. A table comparing the two devices is provided as follow:

AttributeSubject Device –K170219: Microlife EarThermometer 1R1DR1-1Predicate Device - K034023: Microlife EarThermometer 1R1DE1-1Comparison
ThermometerType:Infrared ThermometerInfrared ThermometerSame
Intended useThe device is an electronicclinical thermometer usingan infrared sensor to detectbody temperature from theauditory canal in theneonatal, pediatric andadult population used inthe home setting.The device is an electronicclinical thermometer usingan infrared sensor to detectbody temperature from theauditory canal in theneonatal, pediatric and adultpopulation used in the homesetting.Same
Indications for useThe Microlife DigitalInfrared Ear Thermometer,Model IR1DR1-1 device,is an electronic clinicalthermometer using aninfrared sensor to detectbody temperature from theauditory canal in theThe device is an electronicclinical thermometer usingan infrared sensor todetect body temperaturefrom the auditory canal inthe neonatal, pediatric andadult population used in thehome setting.Same
neonatal, pediatric andadult population used inthe home setting.
Thermometer Measurements
DeviceMeasurementTechnologyInfraredSame-
MeasurementlocationEar CanalSame-
Probe coverneeded?NoSame-
MeasurementRange32.0- 42.2°C(89.6 -108.0 °F)0 - 100°C(32.0 - 212.0°F)Different
DisplayResolution0.1°C or 0.1°FSame-
Accuracy Tested(blackbody)<35.5°C ±0.3°C35.5°C-42.0°C ±0.2°C>42.0°C ±0.3°CSame-
PositioningIndicationYesNoDifferent
ElevatedTemperatureAlarmYes (10 short beeps whenmeasured temperature isgreater than 37.5 °C)Same-
Operating/Storage Environment
OperatingConditions10.0°C-40°C(50.0°F~104.0°F)with relative humidity 95%5°C-40°C (41°F~104°F)with relative humidity 95%Different
Storage Conditions-25°C to 55 °C/-13°F~131°Fwith relative humidity 95%Same-
Device Internal Components
Display typeLCDSame-
MemoryOne temperature memoryTwelve temperaturesmemoryDifferent
Sensor TypeTPS23B sensorSame-
Lens TypeTransparentSame-
Probe Head Dimensions
Probe head : TipWidth7.3 mmSame-
Probe Head:Thickness at 5 mmheight8.56 mm8.35 mmDifferent
Probe Head:Thickness at 1010.32 mm10.22 mmDifferent

{5}------------------------------------------------

{6}------------------------------------------------

ItemTest NameResults
--------------------------
mm height
Integrated Circuit Model and Features
Power SupplyVoltage2.2-3.6 V2.4- 5.5 VDifferent
RAM Space512 Bytes256 BytesDifferent

The indications for use statement and the intended use of the subject device is identical to the predicate device. The differences in the device are:

    1. Measurement temperature range has been narrowed to more physiologically relevant temperatures. This does not introduce any new risk to the device.
    1. Addition of the Position Indication Feature, which has been validated and verified through performance testing and clinical testing.
    1. The integrated circuit model and its features have changed. These changes have been verified through performance testing.
    1. The thermometer's memory can recall the last temperature measurement, which has been decreased from predicate which recalled the last 12 temperature measurements.
    1. The Probe head and the rest of the device is dimensionally larger than the predicate device. This does not change how the device operates. The size difference was validated through clinical studies.
    1. The operating condition has been narrowed. .

Based on the aforementioned modifications to the subject device, the subject device does not raise different types of safety and effectiveness questions when compared to the predicate device.

Performance TestingTesting information demonstrating performance of the Microlife Digital Infrared Ear Thermometer, Model IR1DR1-1 in the intended environment of use is supported by testing that was conducted in accordance with Guidance on the Content of Premarket Notification [510(K)] Submissions for Clinical Electronic Thermometers.

The following testing was conducted to prove safety and effectiveness as well as substantial equivalence to the predicate devices:

{7}------------------------------------------------

1IEC 60601-1: 2005 and A1:2012Passed all testing requirements
2ANSI/AAMI ES60601-1:2005/(R2012)and A1:2012, C1:2009/(R)2012 andA2:2010/(R2012)Passed all testing requirements
3IEC 60601-1-2: 2014Passed all testing requirements
4ASTM E 1965-98 (2009)Passed all testing requirements
5ISO 14971: 2007Passed all testing requirements
6AAMI/ANSI/ISO 10993-1: 2010Passed all testing requirements
7AAMI/ANSI/ISO 10993-5: 2010Passed all testing requirements
8AAMI/ANSI/ISO 10993-10: 2010Passed all testing requirements
9AAMI/ANSI/ISO 10993-12: 2012Passed all testing requirements
10AAMI/ANSI/ISO 80606-2-56: 2009Passed all testing requirements
11IEC 60601-1-11:2010Passed all testing requirements

Performance Testing Summary:

Measurement accuracy of the subject thermometer was tested per and complies with the following standard:

Specification for IR thermometers : ASTM E1965-98 (2009)

Electrical and Electromagnetic Compatibility testing were conducted in accordance with the following standards:

Electrical Safety: IEC 60601-1, IEC 60601-1-11 EMC: IEC 60601-1-2

Software Verification and Validation: per the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices dated May 11, 2005

Clinical Tests

Controlled human clinical studies were conducted in accordance with ASTM E1965-98, IEC80601-2-56 Test Report using the Microlife Digital Infrared Ear Thermometer r Model IR1DR1-1. Clinical data was presented evaluating clinical uncertainty and clinical repeatability per clinical validation for Microlife IR1DR1-1.

Conclusions

The modifications to the subject device include the addition of the position indication feature, the measurement reading range, the operating conditions, the integrated circuit system, temperature memory, and the physical dimensions of the device. There are no changes to the basic measurement technology. The modifications do not raise new or different questions of safety and effectiveness and are supported by non-clinical and clinical testing.

The Microlife Digital Infrared Ear Thermometer IR1DR1-1 is substantially equivalent to the Microlife Digital Infrared Ear Thermometer IR1DE1-1, cleared under K034023.

§ 880.2910 Clinical electronic thermometer.

(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.