K Number
K170196

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-05-17

(114 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Measure blood pressure (systolic and diastolic) and pulse rate.

Device Description

UB-1100BLE UltraConnect digital blood pressure monitor has the same design as the predicate devices with an inflatable cuff which is wrapped around the patient's wrist. After the user starts the blood pressure measurement, the cuff is inflated automatically by an internal pump. The systolic and diastolic blood pressures are also determined by oscillometric method. The deflation rate is controlled by an internal exhaust valve. There is a quick exhaust mechanism so that the cuff pressure can be completely released immediately. There is a maximum pressure safety setting at 299 mmHg. UB-1100BLE UltraConnect digital blood pressure monitor will not inflate the cuff higher than 299 mmHg. UB-1100BLE will turn on an irregular heartbeat indicator if an irregular heartbeat was detected during the measurement process. At the end of the measurement, the systolic and diastolic pressures with pulse rate are shown on the OLED and transmitted via Bluetooth wireless module to the connected app. The cuff is also deflated automatically to 0 mmHg at the same time. If the monitor receives no further action from the user for 1 minute, it will automatically turn off by itself.

AI/ML Overview

This document is a 510(k) premarket notification for a blood pressure monitor, not an AI/ML medical device. Therefore, much of the requested information regarding AI acceptance criteria, ground truth, expert adjudication, and MRMC studies is not applicable.

However, I can extract information related to the device's performance and the general testing methodology for this type of medical device as presented in the document.

Acceptance Criteria and Device Performance (based on the provided specifications for a blood pressure monitor):

The acceptance criteria for a non-invasive blood pressure measurement system like the UB-1100BLE are typically defined by recognized standards such as AAMI/ANSI/IEC 80601-2-30.

Acceptance Criteria CategorySpecific Acceptance Criteria (from UB-543/UA-767PBT which UB-1100BLE matches)Reported Device Performance (for UB-1100BLE, stated as "No change from UB-543" or "No change from UA-767PBT")
Measurement Accuracy (BP)+/- 3mmHg (for UB-543) or +/- 3mmHg or +/- 2% of measured value, whichever is greater (for UA-767PBT)No change from UB-543 (therefore, +/- 3mmHg)
Measurement Accuracy (Pulse)+/- 5% (pulse)No change from UB-543 (therefore, +/- 5% pulse)
Measurement Range (BP)0 - 299mmHg (for UB-543)No change from UB-543
Measurement Range (Pulse)40 - 180 beats per minute (for UB-543)No change from UB-543
Minimum Display Resolution1 mmHgNo change
Maximum Pressure SafetyWill not inflate cuff higher than 299 mmHgThe device will not inflate the cuff higher than 299 mmHg.

Study Proving Device Meets Acceptance Criteria:

The document states: "There is no modification requiring conducting clinical tests." This indicates that the device's performance was not proven through a new clinical study. Instead, substantial equivalence was demonstrated based on its similarities to previously cleared predicate devices and adherence to relevant standards.

Detailed Breakdown of Information as applicable to this medical device submission:

  1. A table of acceptance criteria and the reported device performance:
    (See table above)

  2. Sample size used for the test set and the data provenance:

    • Sample Size: Not applicable as no new clinical test set was used to prove the device's performance. The submission relies on the established performance of its predicate devices and non-clinical testing for equivalence.
    • Data Provenance: Not applicable for a new clinical test set. The predicate devices' performance data would have originated from their respective clinical trials and validation studies, the details of which are not in this document. The current submission focuses on non-clinical tests (e.g., electrical safety, EMC, reliability) to demonstrate that the modifications did not alter the fundamental performance.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):
    Not applicable, as this is a blood pressure monitor and not an AI/ML diagnostic imaging device that requires expert human interpretation for ground truth.

  4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
    Not applicable, as there was no test set requiring adjudication in the context of a new clinical study.

  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    Not applicable. This device is a standalone blood pressure monitor and does not involve AI assistance for human readers/clinicians, nor does it conduct diagnostic tasks in the way an imaging AI would.

  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    The device itself operates as a standalone blood pressure measurement system (algorithm only for generating BP readings) in the sense that it provides measurements directly. However, the context of this question usually refers to AI algorithms performing a diagnostic task. In that sense, it's not directly comparable. The performance (accuracy of BP and pulse readings) is what defines its standalone performance, and that's assessed against physical standards and other measurement devices rather than human interpretation.

  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
    For blood pressure monitors, the "ground truth" is typically established by comparative measurements against a reference standard or invasively measured blood pressure, or through clinical validation according to protocols like those outlined in the AAMI/ANSI/IEC 80601-2-30 standard. It's not based on expert consensus or pathology in the typical sense of AI/ML diagnostic tools.

  8. The sample size for the training set:
    Not applicable. This device does not use machine learning in a way that requires a "training set" in the context of AI/ML model development. It uses an oscillometric method, which is a classical signal processing technique.

  9. How the ground truth for the training set was established:
    Not applicable, as there is no training set for an AI/ML model.

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a symbol featuring three stylized human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 17, 2017

A&D Company, Ltd. Jerry Wang Director Of Engineering 1756 Automation Parkwav San Jose, California 95131

Re: K170196

Trade/Device Name: A&D Medical UB-1100BLE UltraConnect Digital Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: April 14, 2017 Received: April 18, 2017

Dear Jerry Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

M.A. Willemen

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K170196

Device Name

UB-1100BLE UltraConnect Digital Blood Pressure Monitor

Indications for Use (Describe)

Measure blood pressure (systolic and diastolic) and pulse rate.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This summary of 510(k) safety and effective information is being submitted in accordance with the requirement of SMDA and 21 CFR 807.92.

1. Date Prepared

February 23, 2016

2. Submitter's Information

A&D Engineering, Inc. Mr. Jerry Wang 1756 Automation Parkway, San Jose, CA 95131 Tel: 408-518-5113 Fax: 408-635-2313 Email: jwang@andonline.com

3. Device Information

Proprietary Name:

Regulation Number: Regulation Name: Regulatory Class: Product Code:

A&D Medical UB-1100BLE UltraConnect Digital Blood Pressure Monitor 21 CFR 870.1130 Noninvasive Blood Pressure Measurement System Class II DXN

4. Information for he 510(k) Cleared Devices (Predicate Devices)

  • A&D Medical UB-543 Digital Blood Pressure Monitor with 510(k) number K141179
  • A&D Medical UA-767PBT Digital Blood Pressure Monitor with 510(k) number . K043217

Both predicate devices are designed and manufactured by the same company and facilities as the modified device.

ട. Indications for Use

Measure blood pressure (systolic and diastolic) and pulse rate

6. Intended Use

A&D Medical UB-1100BLE UltraConnect digital blood pressure monitor is designed for adult at home or clinic use. It measures systolic, diastolic, and pulse rate. UB-1100BLE uses the oscillometric method to determine blood pressure. If the user has a massive body movement or irregular heartbeats, a heart symbol will be shown on the app display. At the end of the measurement, the blood pressure results are compared with JNC VII hypertension classification. The rating is also displayed. UB-1100BLE has Bluetooth® low energy wireless (BLE) communication capability with Smart devices. UB-1100BLE fits wrist size from 13.5 to 21.5 cm.

Device Description – Technological and Operational Characteristics Comparison 7. UB-1100BLE UltraConnect digital blood pressure monitor has the same design as the predicate devices with an inflatable cuff which is wrapped around the patient's wrist. After the user starts the blood pressure measurement, the cuff is inflated automatically by an internal pump. The

{4}------------------------------------------------

systolic and diastolic blood pressures are also determined by oscillometric method. The deflation rate is controlled by an internal exhaust valve. There is a quick exhaust mechanism so that the cuff pressure can be completely released immediately. There is a maximum pressure safety setting at 299 mmHg. UB-1100BLE UltraConnect digital blood pressure monitor will not inflate the cuff higher than 299 mmHg. UB-1100BLE will turn on an irregular heartbeat indicator if an irregular heartbeat was detected during the measurement process. At the end of the measurement, the systolic and diastolic pressures with pulse rate are shown on the OLED and transmitted via Bluetooth wireless module to the connected app. The cuff is also deflated automatically to 0 mmHg at the same time. If the monitor receives no further action from the user for 1 minute, it will automatically turn off by itself.

8. Summary of Substantial Equivalence

Modifications made from the predicate devices:

  • · Changed the plastic molds to have a new appearance.
  • Change LCD display to OLED display.
  • · Change battery type from alkaline to Li-ion.
  • · Increase number of users from 2 to six users.
  • · Increase the memory size to 100 per user except in the visitor mode.
  • · Change pressure sensor from static electricity capacity type to semiconductor type
Predicate DevicesModified Device
ModelUA-767PBTUB-543UB-1100BLE
Measurement MethodOscillometric MethodOscillometric MethodNo change
Measurement RangeBP : 20 - 280mmHgPulse : 40 - 200beats perminuteBP : 0 - 299mmHgPulse : 40 - 180beatsper minuteNo change from UB-543
Measurement AccuracyBP : +/- 3mmHg or +/-2% of measured value,whichever is greaterPulse :+/- 5% (pulse)BP : +/- 3mmHgPulse :+/- 5% (pulse)No change from UB-543
Minimum Display Resolution1 mmHg1 mmHgNo change
Pressure sensorStatic electricity capacitytype sensorStatic electricitycapacity type sensorSemiconducting typesensor
Pressurization SourceAutomatic internal pumpAutomatic internalpumpNo change
Cuff Deflation MethodAutomatic constant speedmechanical exhaust valveSolenoid valve forrapid exhaustNo change from UB-543
Display TypeLCD displayLCD displayOLED display
Cuff Attachment MethodBy plastic hose connectedto monitorIntegrated (hoseless)No change from UB-543
IHBEnabledEnabledNo change
%IHBNoNoEnabled by app display.
Power Source6V DC, 4x1.5V AAbatteries or AC adaptoras an option6V DC, 4x1.5V AAbatteries or ACadaptor as an option3.7V Li-ion batteriesor AC adaptor as anoption
ProtectionagainstelectricalshockClass I, Type BClass I, Type BNo change
Battery Life6 months with onemeasurement dairy6 months with onemeasurement dairy3 months with onemeasurement daily
OperatingEnvironment50°F(10°C) to104°F(40°C)at less than 85%RH50°F(10°C) to104°F(40°C)15%RH to 85%RHNo change
StorageEnvironment-4°F(-20°C) to140°F(60°C)at less than 85%RH-4°F(-20°C) to140°F(60°C)10%RH to 95%RHNo change
DataMemory SizeLast 40 measurements ofsystolic, diastolic, andpulse rateLast 60 measurementseach for user 1 anduser 2.Last 100 measurementsfor each of the five user
Dimensions163.7(H)x111(W)x66.7(L) mm56(W)x88(H)x18(D)mm40(W)x126(H)x26(D) mm
Weight320g without batteries104g without batteries205g without batteries
Arm or WristSize13 to 45 cm (Arm)13.5cm to 21.5cm(Wrist)No change from UB-543
Clock(Time/Date)Yes (By App)YesNo change from UA-767PBT
AM/PMNoYesNo change from UB-543
PressureIndicatorYesYesNo change
WirelessRadioConnectivityBluetooth wirelessNoNo change from UA-767PBT
Field serviceNot allowedNot allowedNo change
AutomaticZero at

Product Specification Comparison Table:

{5}------------------------------------------------

{6}------------------------------------------------

BloodPressureReadingClassificationCriteriaUSA JNC VIIWHO ClassificationCriteriaNo change from UA-767PBT
---------------------------------------------------------------------------------------------------------------------------------------

Key Features Comparison

ParameterPredicate Devices(UB-543 & UA-767PBT)Modified Device(UB-1100BLE)
Field serviceNot allowedNo change – the same
Automatic zero after“START” operationYesNo change – the same
Manual zero adjustmentNot allowedNo change – the same
CalibrationNot allowed in the fieldNo change – the same
SterilizationNot neededNo change – the same

9. Discussion of standards used in the design verification and design validation

A&D Medical follows FDA recognized consensus standards and guidance documents in our medical device development and manufacturing processes. The following standards were used to for the design verification and valuation of UB-1100BLE UltraConnect digital blood pressure monitor. These standards include three major groups. First group is the general quality system requirements. Second group is the special requirements for CDRH (870.1130). The third group is related to software and app life cycle processes requirements.

  • AAMI/ANSI/ISO 14971:2007/(R) 2010 (Corrected 4 October 2007) Medical Devices - Applications Of Risk Management To Medical Devices (FDA Recognized Number 5-70)
  • IEC 60601-1-11 Edition 1.0 2010-04 Medical Electrical Equipment - Part 1-11: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Requirements For Medical Electrical Equipment And Medical Electrical Systems Used In The Home Healthcare Environment [Including: Technical Corrigendum 1 (2011)] (FDA Recognized Number 19-6)
  • . IEC 60601-1-2 Edition 3: 2007-03 Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests (FDA Recognized Number 19-1)
  • . AAMI/ANSI/IEC 60601-1-2:2007/(R)2012 Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests (Edition 3) (FDA Recognized Number 19-2)
  • . AAMI/ANSI/IEC 80601-2-30:2009 & A1:2013 Medical Electrical Equipment - Part 2-30: Particular Requirements For The Basic Safety And Essential Performance Of Automated Noninvasive Sphygmomanometers (FDA Recognized Number 3-130)

{7}------------------------------------------------

  • AAMI/ANSI/IEC 62304:2006 Medical Device Software - Software Life Cycle Processes (FDA Recognized Number 13-32)

10. Substantial Equivalence Conclusion:

Non-clinical Test Summary:

  • Safety Tests: IEC 60601-11 Edition 1.0 2010-04 ●
  • EMC Tests: IEC 60601-1-2 Edition 3: 2007-03
  • ASMI/ANSI/IEC 80601-2-30:2009 & A1:2013 ● Reliability Tests:
  • Risk Assessment: ISO 14971:2012 ●
  • Software Assessment: IEC Medical Life Cycle Process & FDA "Guidance for the ● Content of Premarket Submissions for Software contained in Medical Devices", November 2005.
  • Bluetooth Tests: FCC Part 15 Subpart C : 2014

There is no modification requiring conducting clinical tests.

UB-1100BLE UltraConnect digital blood pressure monitor has the following similarities to the predicate devices, UB-543 and UA-767PBT digital blood pressure monitors, which previously received the 510(k) clearance.

  • Same intended use. ●
  • Same Indications For Use
  • Same oscillometric method to determine the blood pressure & pulse rate
  • Same inflation method - automatic internal pump
  • . Same fast safety deflation method - solenoid controlled exhaust valve
  • Same materials, no new materials used
  • . Same manufacturing processes at the same facility
  • Same wrist size.

As a conclusion. UB-1100BLE UltraConnect digital blood pressure monitor as described in its labeling and comparison analysis has not changed as a result of the modifications. The fundamental scientific technology of the modified device has not changed, either. There is no significant difference that affects the safety or effectiveness of the modified device as compared to the predicate devices.

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).