Intended Use

The A&D Medical UB-543 family digital blood pressure monitor is intended for used by adults with 12 years older to measure the systolic and diastolic blood pressure and pulse rate.

Device Description

UB-543 uses an inflatable cuff which is wrapped around the patient's wrist. After the user pushes the "START" button, cuff is inflated automatically by an internal pump once the condition of the CPG (correct position guidance) algorithm is met. Measurement starts and the systolic and diastolic blood pressures are determined by oscillometric method. There is a quick exhaust mechanism so that the cuff pressure can be completely released urgently. There is a maximum pressure safety setting at 300mmHg. UB-543 will not inflate the cuff higher than 300mmHg. UB-543 will turn on an irregular heartbeat indicator if an irregular heartbeat was detected during the measurement process. At the end of the measurement, the systolic and diastolic pressures with pulse rate are shown on the LCD and stored in the device memory. The cuff is also deflated automatically to 0 mmHg at the same time.

AI/ML Overview

The provided document is a 510(k) summary for a digital blood pressure monitor. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed study proving the device meets specific acceptance criteria in a standalone performance study.

Therefore, many of the requested details about acceptance criteria studies, such as sample size for test sets, data provenance, expert ground truth, adjudication methods, MRMC studies, and standalone performance metrics, are not available in this document. The document describes changes from predicate devices and emphasizes that the fundamental scientific technology and intended use have not changed, implying that the established performance of the predicate devices should apply.

Here's a summary of the information that can be extracted from the provided text, along with a statement where information is not available:

1. Table of Acceptance Criteria and Reported Device Performance

ParameterAcceptance Criteria (from predicate)Reported Device Performance (Modified Device UB-543)
Accuracy (BP)+/- 3mmHg or +/- 2% of measured value, whichever is greater+/- 3mmHg
Accuracy (Pulse)+/- 5%+/- 5%
Measurement Range (BP)20 to 280 mmHg0 to 299 mmHg
Measurement Range (Pulse)40 to 200 pulse/min40 to 200 pulse/min
Maximum Pressure Safety SettingUp to 300mmHg (implied)300mmHg
Inflation MechanismAutomatic internal pumpAutomatic internal pump
Deflation MethodQuick exhaust valveQuick exhaust valve

Note: The accuracy criteria for UB-543 are provided as +/- 3mmHg and +/- 5% for pulse. For the predicate, the BP accuracy is +/- 3mmHg or +/- 2%, whichever is greater. While the UB-543 states +/- 3mmHg, it doesn't explicitly state the "whichever is greater" clause, but given the substantial equivalence claim, it's highly likely to adhere to the same underlying standard implying consistency with the predicate. The measurement range for BP is slightly expanded for UB-543 (0-299mmHg vs 20-280mmHg).

2. Sample size used for the test set and the data provenance

  • Sample Size: Not specified in this document.
  • Data Provenance: Not specified in this document. Given that it's a 510(k) for a modified device claiming substantial equivalence, it's possible that clinical data was not required for the modifications, and performance was inferred from the predicate device and internal testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Number of Experts: Not specified.
  • Qualifications of Experts: Not specified.

4. Adjudication method for the test set

  • Not specified.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This is a blood pressure monitor, not an AI-assisted diagnostic imaging device for human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Yes, implicitly. Blood pressure monitors are standalone devices where the algorithm (oscillometric method) determines the blood pressure and pulse rate without human intervention beyond initiating the measurement. The document states: "Measurement starts and the systolic and diastolic blood pressures are determined by oscillometric method."

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Not explicitly stated, but for blood pressure monitors, the "ground truth" or reference standard for accuracy is typically direct intra-arterial measurement or a validated auscultatory method performed by trained personnel according to established protocols (e.g., AAMI, ISO standards). The document refers to "Accuracy: BP : +/- 3mmHg" which indicates a comparison against a reference.

8. The sample size for the training set

  • Not applicable/specified. This device uses a traditional oscillometric algorithm, not a machine learning model that requires a "training set" in the conventional sense.

9. How the ground truth for the training set was established

  • Not applicable/specified (see #8).

{0}------------------------------------------------

K141179 Page 1/3

510(k) Summary

This summary of 510(k) safety and effective information is being submitted in accordance with the requirement of SMDA and 21 CFR 807.92.

    1. Date Prepared May 27, 2014

2. Submitter's Information

A&D Engineering, Inc. Mr. Jerry Wang 1756 Automation Parkway, San Jose, CA 95131 Tel: 408-518-5113 Fax: 408-635-2313 Email: jwang@andonline.com

3. Device Information

Proprietary Name: A&D Medical UB-543Digital Blood Pressure Monitor Common/Usual Name: Blood Pressure Monitor Classification name: Non-invasive blood pressure measurement System 21 CFR 870-1130, Class II, 74DXN,

Predicate Devices 4.

A&D Model UB-511 & UB-512 Digital Blood Pressure Monitors with 510(k) number K042967. Predicate devices are designed and manufactured by the same company and facilities as the modified devices UB-543.

Device Description - Technological and Operaional Characteristics ડ.

UB-543 uses an inflatable cuff which is wrapped around the patient's wrist. After the user pushes the "START" button, cuff is inflated automatically by an internal pump once the condition of the CPG (correct position guidance) algorithm is met. Measurement starts and the systolic and diastolic blood pressures are determined by oscillometric method. There is a quick exhaust mechanism so that the cuff pressure can be completely released urgently. There is a maximum pressure safety setting at 300mmHg. UB-543 will not inflate the cuff higher than 300mmHg. UB-543 will turn on an irregular

heartbeat indicator if an irregular heartbeat was detected during the measurement process.

{1}------------------------------------------------

At the end of the measurement, the systolic and diastolic pressures with pulse rate are shown on the LCD and stored in the device memory. The cuff is also deflated automatically to 0 mmHg at the same time.

6. Intended Use

The A&D Medical UB-543 family digital blood pressure monitor is intended for used by adults with 12 years older to measure the systolic and diastolic blood pressure and pulse rate.

The intended user and the indication for use of A&D Medial UB-543 as described in the labeling are the same as their predicated devices, UB-511 & UB-512.

7. Summary of Substantial Equivalence

Modifications made from the predicate devices:

  • Change the plastic molds so UB-543 has a different appearance .
  • Increase LCD display size .
  • Increase memory size ●
  • Add time and date function .
  • Add correct position guideance (CPG) indicator .

Product Specification Comparison

ParameterPredicate Devices(UB-511 & UB-512)Modified Devices(UB-543)
Power source2 AAA size batteriesNo change - the same
Battery Life4 months with daily measurementNo change - the same
BP MeasurementMethodOscillometric MethodNo change - the same
MeasurementRangeBP : 20 to 280 mmHgPulse : 40 to 200 pulse/minBP : 0 to 299 mmHgPulse : 40 to 200 pulse/min
AccuracyBP : +/- 3mmHg or +/- 2% ofmeasured value, whichever isgreaterPulse : +/- 5 % (pulse)BP : +/- 3mmHgPulse : +/- 5 %
PressurizationSourceAutomatic internal pumpNo change - the same
Cuff DeflationMethodQuick exhaust valveNo change - the same
Display Type &SizeLiquid crystal display with 3.8 cm(1.5") x 2.3 cm (0.9")Liquid crystal display with6.0 cm (2.4") x 4.0 cm (1.6")
IHB (IrregularHeartbeatsDetection)More than +/-25% to the meaninterval of all pulse intervalsNo change - the same

002_510(k) Summary_UB-543 (revised).doc-1/JW

{2}------------------------------------------------

OperatingEnvironment50°F (10°C) to 104°F (40°C)15 %RH to 85% RHNo change - the same
StorageEnvironment14°F (-20°C) to 140°F (60°C)10 %RH to 95% RHNo change - the same
Pressure IndicatorUSA JNC VII guidelineNo change - the same
Data Memory Sizewith Time & DateLast 30 measurementsLast 60 measurements eachfor user 1 and user 2
Pressure SensordesignStatic electricity capacity typeNo change - the same
Number of userUA-511 : oneUB-512 : twoThe same as UB-512
Dimensions in mm58 (H) x 156 (W) x 145 (L)88 (H) x 56 (W) x 18 (D)
Weight82g without batteries103g without batteries
Wrist Size5.1 inches (13.0 cm) to 15.2 inches(38.6 cm)No change - the same

Key Features Comparison

ParameterPredicate Devices(UB-511 & UB-512)Modified Devices(UB-543)
Field serviceNot allowedNo Change - the same
Automatic zero at "START"YesNo Change - the same
Manual zero adjustmentNot allowedNo Change - the same
CalibrationNot allowed in the fieldNo Change - the same
SterilizationNot neededNo Change - the same

Substantial Equivalence Conclusion:

UB-543 Digital blood pressure monitors have the following similarities to the predicate devices, UB-411 and UB-412 digital blood pressure monitors, which previously received the 510(k) clearance.

  • Same Indication of Use .
  • Same intended use .
  • Same oscillometric method to determine the blood pressure & pulse rate .
  • Same inflation method -- automatic internal pump �
  • Same deflation method fast exhaust valve ●
  • ◆ Same materials, no new materials used
  • Same manufacturing processes at the same facility .

As a conclusion, the intended use of the modified device, UB-543 as described in its labeling, has not changed as a result of the modifications. The fundamental scientific technology of the modified device, UB-543, has not changed, either. There is no significant difference that affects the safety or effectiveness of the modified device as compared to the predicate devices.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center WO66-G609 Silver Spring MD 20993-0002

A & D Co., Ltd. c/o Dr. Jerry Wang Director of Engineering 1756 Automation Parkway San Jose, California 95131

Re: K141179

Trade/Device Name: A&D Medical UB-543 Digital Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Non-Invasive Blood Pressure Monitor Regulatory Class: Class II (two) Product Code: DXN Dated: May 29, 2014 Received: May 30, 2014

June 27, 2014

Dear Dr. Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply

{4}------------------------------------------------

Page 2 - Dr. Jerry Wang

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 OIR and Part 809; medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 OIR and Part 809, please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Image /page/4/Picture/9 description: The image shows the text "Sincerely yours," followed by a logo that appears to say "FDA". There is a signature over the logo. The signature is illegible, but it appears to say "for".

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): _

Device Name: A&D Medical UB-543 Digital Blood Pressure Monitor

Indications For Use:

Measure blood pressure (systolic and diastolic) and pulse rate.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use × (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Image /page/5/Picture/11 description: The image shows a logo with the letters FDA in the center. To the right of the logo, there is a date and time stamp that reads 2014.06.27 5:43:16 04:00.

Page 1 of _ 1 __

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).