K Number
K170168

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-09-14

(238 days)

Product Code
Regulation Number
870.2300
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Masimo Rad-97 and Accessories are indicated for hospital-type facilities, mobile, and home environments.

The Masimo Rad-97 and Accessories can communicate with network systems for supplemental remote viewing and alarming (e.g., at a central station).

The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRa). The Masimo Rad-97 and Accessories are indicated for use with adult, pediatic, and neonatal patients during both no motion conditions, and for patients who are well or poorly perfused. In addition, the Masimo Rad-97 and Accessories are indicated to provide the continuous non-invasive monitoring data obtained from the Masimo Rad-97 and Accessories for functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) to multi-parameter devices for the display on those devices.

The Masimo Rad-97 and Accessories are not intended to be used as the sole basis for making diagnosis or treatment decisions related to suspected carbon monoxide poisoning; it is intended to be used in conjunction with additional methods of assessing clinical signs and symptoms.

The optional Nomoline Capnography product family is intended to other medical backboard devices for monitoring of breath rate and CO2. The Nomoline Capnography product family is intended to a patient breathing circuit for monitoring of inspired gases during anesthesia, recovery and respiratory care. The environment is the operating suite, intensive care unit and patient population is adult, pediatric and infant patients.

The optional non-invasive blood pressure (NIBP) module is indicated for the noninvasive measurement of arterial blood pressure. The NIBP module is designed to measure blood pressure for patient population described in the following table:

Patient PopulationApproximate Age Range
Newborn (neonate)Birth to 1 month of age
Infant1 month to 2 years of age
Child2 to 12 years of age
Adolescent12-21 years of age
Adult21 years of age and older
Device Description

The subject device, Masimo Rad-97 System and Accessories (Rad-97) product family, features a touchscreen display that continuously displays numeric values for the measured monitoring parameters. The Rad-97 product family can be operated on AC power or internally rechargeable battery.

The Rad-97 product family comprises the same measurement technologies as cleared in the predicate, Root (with connected external Radical-7 and capnography (ISA) and internal NIBP modules), which includes the Masimo Rainbow SET technology, capnography technology and noninvasive blood pressure (NIBP) technology. These technologies enable the Rad-97 product family to provide noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), Perfusion Index (PI), Pleth Variability Index (PVI), carboxyhemoglobin (SpCO), methemoglobin (SpMet), total hemoglobin (SpHb), oxygen content (SpOC), acoustic respiration rate (RRa) and/or optional capnography parameters or optional noninvasive blood pressure (NIBP) parameters.

The Rad-97 product family is available with different measurement parameter configurations, ranging from a fully loaded configuration to a simplified configuration with pulse oximetry parameters only. In a fully loaded configuration, the Rad-97 model includes all parameters provided by the Masimo Rainbow SET technology including SpO2, PR, PI, PVI, SpCO, SpMet, SpHb, SpOC and RRa. Additionally, this fully loaded version can be optionally available with either NIBP or capnography technology.

The Rad-9 model, an instrument model within the Rad-97 product family, is an embodiment with a simplified configuration. The Rad-9 model includes the Masimo SET technology (a subset of Masimo Rainbow SET technology), which provides pulse oximetry parameters of SpO2, PR, PI and PVI. The Rad-9 model can be optionally available with NIBP technology.

Same as the predicate, the Rad-97 product family includes inout/output interfaces for connection to external devices. Furthermore, same as the predicate, the Rad-97 product family can communicate through wired/wireless connection with networked systems such as Patient SafetyNet (K071047) and/or hospital electronic medical/health record (EMR) systems.

The RD SET Disposable Sensors, listed with other cleared accessories in Section 11.7, are the same as the currently marketed product. The only change is the added labeling information regarding the sensors' performance in terms of limits of agreement (LOA) as defined by Bland-Altman.

AI/ML Overview

Here's an analysis of the provided text to extract the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

ParameterAcceptance Criteria (Requirement)Reported Device Performance (Result)
SpO2, no motion (Upper 95% Limits of Agreement)< 3%3.15%
SpO2, no motion (Lower 95% Limits of Agreement)>= -3%-2.21%

Note: The reported performance for "Upper 95% LOA" (3.15%) indicates that the device did not meet the specified acceptance criterion of "< 3%". The reported performance for "Lower 95% LOA" (-2.21%) indicates that the device met the specified acceptance criterion of ">= -3%".

The document also lists other accuracy specifications for various parameters, but it does not explicitly state an "acceptance criteria" alongside them in the same format as the LOA for SpO2. These are presented as general "Accuracy (ARMS)" specifications.

Accuracy (ARMS) Specifications (not explicitly presented as acceptance criteria vs. performance in the provided text, but rather device specifications):

ParameterAccuracy Specification (ARMS)*Patient Population
SpO2, no motion60-80%, 3%; 70-100%, 2%Adults/pediatrics/infants
SpO2, no motion70-100%, 3%Neonates
SpO2, motion70-100%, 3%All
SpO2, low perfusion70-100%, 2%All
Pulse rate, no motion25-240 bpm, 3 bpmAll
Pulse rate, motion25-240 bpm, 5 bpmAll
Pulse rate, low perfusion25-240 bpm, 3 bpmAll
RRa4-70 breaths per minute, 1 breath per minuteAdults, Pediatics
SpCO1-40%, 3%Adults/pediatrics/infants
SpMet1-15%, 1%All
SpHb8-17 g/dL, 1 g/dLAdults/pediatrics
CO2 (Single dry gasses)0-15 volume % +0.2 volume% +2% of reading-
CO2 (All conditions)0.3 kPa + 4% of reading-
RR (Respiration rate)0-150 breaths/min + 1 breaths/min-
NIBP0-300 mmHg, +3 mmHg-
  • ARMS defined as a statistical calculation of the difference between device measurements fell within +/- ARMs of the reference measurements in a controlled study.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not explicitly stated as a number. The study was conducted on "healthy adult volunteers."
  • Data Provenance: Retrospective or Prospective is not explicitly stated, but clinical studies are generally prospective. The data is from "healthy adult volunteers," implying a clinical study setting. Country of origin is not specified, but the Masimo Corporation is based in Irvine, CA, USA, suggesting the study likely took place in the USA or under its regulations.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • This information is not provided in the text. The ground truth for SpO2 was established by "blood measurements from a laboratory CO-Oximeter," which is an instrument, not human experts.

4. Adjudication Method for the Test Set

  • Not applicable. The ground truth was established by an instrument (laboratory CO-Oximeter) comparing directly to the device measurements; there was no human adjudication process involved for establishing the ground truth.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, an MRMC comparative effectiveness study was not done. This type of study typically involves human readers interpreting cases with and without AI assistance. This document describes the performance of a medical device (oximeter) and its various physiological monitoring capabilities, not an AI-powered diagnostic imaging tool that would typically undergo MRMC studies.

6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance)

  • Yes, a standalone performance study was done. The clinical study described "evaluate the sensor's performance for no motion condition, in the range of 70% to 100% in comparison to blood measurements from a laboratory CO-Oximeter." This evaluates the device directly against a reference standard, which is a standalone performance assessment.

7. Type of Ground Truth Used

  • The ground truth used for SpO2 was instrumental measurement / reference standard. Specifically, it was "blood measurements from a laboratory CO-Oximeter."

8. Sample Size for the Training Set

  • This information is not provided in the text. The document describes a clinical study for validating the device's accuracy, not for training a new algorithm. The device incorporates "Masimo Rainbow SET technology," which is a pre-existing technology, and the study is for the performance of the integrated product and its sensors.

9. How the Ground Truth for the Training Set Was Established

  • This information is not provided in the text, as no specific training set or new algorithm development is detailed. The technology itself (Masimo Rainbow SET) is described as utilizing principles of spectrophotometry and photoplethysmography, implying established scientific principles rather than a continuously trained machine learning model.

{0}------------------------------------------------

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 14, 2017

Masimo Corporation Marguerite Thomlinson Sr. Director, Regulatory Affairs 52 Discoverv Irvine, California 92618

Re: K170168

Trade/Device Name: Masimo Rad-97 and Accessories Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MWI, JKS, CCK, BZQ, DQA, DPZ, DXN, FLL Dated: August 11, 2017 Received: August 14, 2017

Dear Marguerite Thomlinson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

M & Hillelien

for _ Bram D. Zuckerman, M.D.
Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K170168

Device Name

Masimo Rad-97 and Accessories

Indications for Use (Describe)

The Masimo Rad-97 and Accessories are indicated for hospital-type facilities, mobile, and home environments.

The Masimo Rad-97 and Accessories can communicate with network systems for supplemental remote viewing and alarming (e.g., at a central station).

The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRa). The Masimo Rad-97 and Accessories are indicated for use with adult, pediatic, and neonatal patients during both no motion conditions, and for patients who are well or poorly perfused. In addition, the Masimo Rad-97 and Accessories are indicated to provide the continuous non-invasive monitoring data obtained from the Masimo Rad-97 and Accessories for functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) to multi-parameter devices for the display on those devices.

The Masimo Rad-97 and Accessories are not intended to be used as the sole basis for making diagnosis or treatment decisions related to suspected carbon monoxide poisoning; it is intended to be used in conjunction with additional methods of assessing clinical signs and symptoms.

The optional Nomoline Capnography product family is intended to other medical backboard devices for monitoring of breath rate and CO2. The Nomoline Capnography product family is intended to a patient breathing circuit for monitoring of inspired gases during anesthesia, recovery and respiratory care. The environment is the operating suite, intensive care unit and patient population is adult, pediatric and infant patients.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including thetime to review instructions, search existing data sources, gather and maintain the data needed and completeand review the collection of information. Send comments regarding this burden estimate or any other aspectof this information collection, including suggestions for reducing this burden, to:
Department of Health and Human ServicesFood and Druq AdministrationOffice of Chief Information OfficerPaperwork Reduction Act (PRA) StaffPRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection ofinformation unless it displays a currently valid OMB number."

{3}------------------------------------------------

(Masimo Rad-97 and Accessories, continued from page 1)

Indications for Use

The optional non-invasive blood pressure (NIBP) module is indicated for the noninvasive measurement of arterial blood pressure. The NIBP module is designed to measure blood pressure for patient population described in the following table:

Patient PopulationApproximate Age Range
Newborn (neonate)Birth to 1 month of age
Infant1 month to 2 years of age
Child2 to 12 years of age
Adolescent12-21 years of age
Adult21 years of age and older

{4}------------------------------------------------

Submitter and Address ofManufacturing Facility:Masimo Corporation52 DiscoveryIrvine, CA 92618Phone: (949) 297-7000FAX: (949) 297-7592
Submitter:Marguerite Thomlinson
Date:August 11, 2017
Official CorrespondentMarguerite ThomlinsonSenior Director, Regulatory AffairsMasimo CorporationPhone: (949) 297-7683Mthomlinson@Masimo.com
Trade Name:Masimo Rad-97/Rad-9 and Accessories
Common Name:Oximeter
Primary ClassificationRegulation/ Product Code:21 CFR 878.2300, Class II/MWI
Secondary ClassificationRegulation/ Product Code:21 CFR 862.3220, Class II/JKS21 CFR 880.2910, Class II/CCK21 CFR 868.2375, Class II/BZQ21 CFR 870.2700, Class II/DQA21 CFR 870.2710, Class II/DPZ21 CFR 870.1130, Class II/DXN21 CFR 880.2910, Class II/FLL
Establishment RegistrationNumber:2031172
Reason for PremarketNotification:New Device – Masimo Rad-97/Rad-9 and Accessories
Predicate Device:K151644- Masimo Root Monitoring System andAccessories
Performance StandardsNo performance standards for the above device have beenpromulgated pursuant to Section 514.

{5}------------------------------------------------

Device Description

The subject device, Masimo Rad-97 System and Accessories (Rad-97) product family, features a touchscreen display that continuously displays numeric values for the measured monitoring parameters. The Rad-97 product family can be operated on AC power or internally rechargeable battery.

The Rad-97 product family comprises the same measurement technologies as cleared in the predicate, Root (with connected external Radical-7 and capnography (ISA) and internal NIBP modules), which includes the Masimo Rainbow SET technology, capnography technology and noninvasive blood pressure (NIBP) technology. These technologies enable the Rad-97 product family to provide noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), Perfusion Index (PI), Pleth Variability Index (PVI), carboxyhemoglobin (SpCO), methemoglobin (SpMet), total hemoglobin (SpHb), oxygen content (SpOC), acoustic respiration rate (RRa) and/or optional capnography parameters or optional noninvasive blood pressure (NIBP) parameters.

The Rad-97 product family is available with different measurement parameter configurations, ranging from a fully loaded configuration to a simplified configuration with pulse oximetry parameters only. In a fully loaded configuration, the Rad-97 model includes all parameters provided by the Masimo Rainbow SET technology including SpO2, PR, PI, PVI, SpCO, SpMet, SpHb, SpOC and RRa. Additionally, this fully loaded version can be optionally available with either NIBP or capnography technology.

The Rad-9 model, an instrument model within the Rad-97 product family, is an embodiment with a simplified configuration. The Rad-9 model includes the Masimo SET technology (a subset of Masimo Rainbow SET technology), which provides pulse oximetry parameters of SpO2, PR, PI and PVI. The Rad-9 model can be optionally available with NIBP technology.

Same as the predicate, the Rad-97 product family includes inout/output interfaces for connection to external devices. Furthermore, same as the predicate, the Rad-97 product family can communicate through wired/wireless connection with networked systems such as Patient SafetyNet (K071047) and/or hospital electronic medical/health record (EMR) systems.

The RD SET Disposable Sensors, listed with other cleared accessories in Section 11.7, are the same as the currently marketed product. The only change is the added labeling

{6}------------------------------------------------

information regarding the sensors' performance in terms of limits of agreement (LOA) as defined by Bland-Altman1

The table below provides the specifications for a fully loaded configuration of the Rad-97 instrument model with all parameters from the Masimo Rainbow SET technology and optional parameters from the Nomoline Capnography technology or optional parameters from the NIBP technology. The Rad-9 model, a simplified configuration of the Rad-97 product family, has the same instrument specifications but only includes pulse oximetry parameters from the Masimo SET technology, a subset of the Masimo Rainbow SET technology included in the Rad-97 model.

FeatureRad-97 Specification
Display
Display typeTouchscreen, Color LCD (Backlit Active Matrix TFT LCD)
Functional Oxygen Saturation (SpO2): 0-100%
Pulse Rate (PR): 25-240 beats per minute (bpm)
Perfusion Index (PI): 0.02-20%
Pleth Variability Index (PVI): 0-100%
Respiration Rate (RRa): 0-70 breaths per minute
Carboxyhemoglobin Saturation (SpCO): 0-99%
Methemoglobin Saturation (SpMet): 0-99.9%
Measurement rangeTotal Hemoglobin (SpHb): 0-25g/dL; 0-250g/L(0-15.52 mmol/dL)
Oxygen Concentration (SpOC): 0-35ml of O2/dL
Carbon Dioxide (CO2); 0 vol%-15 vol %
End Tidal CO2 (EtCO2): 0-25 %, 0-32.5 kPa, 0-244 mmHg
Fractional Inspired CO2 (FiCO2): 0-25 %, 0-32.5 kPa, 0-244 mmHg
Respiratory Rate (RR): 0-150 breaths/min
Adult, Systolic: 40-260 mmHg, Diastolic: 20-200 mmHg, MAP 26-220mmHg
Pediatric, Systolic: 40-230 mmHg, Diastolic: 20-160 mmHg, MAP 26-183 mmHg
Neonatal, Systolic: 40-130 mmHg, Diastolic: 20-100 mmHg, MAP 26-110mmHg
SpO2: 1%
PR: 1 bpm
RRa: 1breath per minute
Display resolutionSpCO: 1%
SpMet: 0.1%
SpHb: 0.1 g/dL
SpOC: 0.1 ml/dL

1 Bland and Altman. Agreement between methods of measurement with multiple observations per individual. Journal of Biopharmaceutical Statistics (2007) vol. 17 pp. 571-582 [Note: This paper was cited by Guidance for Industry and Food and Drug Administration Staff - Pulse Oximeters - Premarket Notification Submissions 510(k)s, March 4, 2013]

{7}------------------------------------------------

FeatureRad-97 Specification
Accuracy (ARMS)*Masimo Rainbow SET/ Masimo SET Parameters[accuracy range , accuracy, patient population]
60-80%, 3%, adults/pediatrics/infants
70-100%, 2%, adults/pediatrics/infants
SpO2, no motion70-100%, 3%, neonates
Upper 95% Limits of Agreement, 3% **
Lower 95% Limits of Agreement, -3%**
SpO2, motion70-100%, 3%, adults/pediatrics/infants/neonates
SpO2, low perfusion70-100%, 2%, adults/pediatrics/infants/neonates
Pulse rate, no motion25-240 bpm, 3 bpm adults/pediatrics/infants/neonates
Pulse rate, motion25-240 bpm, 5 bpm, adults/pediatrics/infants/neonates
Pulse rate, low perfusion25-240 bpm, 3 bpm, adults/pediatrics/infants/neonates
RRa4-70 breaths per minute, 1 breath per minute for adults4-70 breaths per minute, 1 breath per minute for pediatrics
SpCO1-40%, 3%, adults/pediatrics/infants
SpMet1-15%, 1%, adults/pediatrics/infants/neonates
SpHb8-17 g/dL, 1 g/dL, adults/pediatrics
AccuracyOptional Capnography or Optional NIBP Parameters
Single dry gasses at 22+ 5°C and 1013 + hPa:
CO20-15 volume % +0.2 volume% +2% or readingAll conditions: 0.3 kPa + 4% of reading
RR (Respiration rate)0-150 breaths/min + 1 breaths/min
NIBP0-300 mmHg, +3 mmHg
Displays/ Indicators
Data displayedSpO2, PR, SPCO, SpMet, SpHb, Rra, PI, PVI, SpOC, NIBP, CO2, pleth waveform, alarmstatus, status messages, sensor status, Signal IQ, APOD, FastSat
AlarmAlarm limits (audible and visual), 3D alarm, alarm silence, sensor condition alarms, systemfailure alarms, low battery alarms
SpO2: 2-4, 4-6, 8, 10, 12, 14 and 16 seconds;
Rra: Trending, No Averaging, Fast, Medium, or Slow
Averaging timePVI: Short or Long;PI: Short or Long;SpHb: Short, Medium, or Long
Sensitivity modeNormal, APOD, and MAX
Sleep study mode10 seconds
Electrical
AC powerInput voltage: 100-240 VAC, 47-63 Hz; input power: 60 VA
Internal battery powerRechargeable lithium ion battery
Input/Output Interface
M-20 connector
Patient applied part connectionNomoline Capnography input connectorNIBP Nib
Analog outputNurse call output
USB interfaceSatShare module and external devices (e.g. barcode reader/scanner) connection
Wired (Ethernet) and wireless (Wi-Wired/wireless connection with networked Patient SafetyNet and/or EMR systemsWireless connection with presence tag, Kite, external devices (e.g. Caregiverthermometer)
Fi/Bluetooth) interface
Camera (MIPI camera interface)Surveillance/conferencing capabilities with Patient SafetyNet
AudioMicrophone supporting video/audio modes for surveillance/conferencing capabilities withPatient SafetyNet
Mechanical
Enclosure MaterialThermoplastic

{8}------------------------------------------------

FeatureRad-97 Specification
Dimensions9×4×6.5 inch (22.9×10.2×16.5cm)
Weight< 1.36 kg. (3.0 lbs.) without Capnography and NIBP
Environmental
Operating Temperature0 to 35 °C (32 to 95 °F)
Storage/Transport Temperature-20 to 60 °C (-4 to 140 °F)
Operating Humidity10 to 95%, non-condensing
Storage/Transport Humidity10 to 90%, non-condensing
Operating Atmospheric Pressure540 to 1,060 mbar (540 to 1060 hPa)
Compliance
Electrical Safety/EMCIEC 60601 compliant
Type of ProtectionClass 1, when on AC power; internally powered, when on battery power
Degree of ProtectionDefibrillation proof, BF-applied part
Protection against harm fromliquid ingressIPX1, protection vertically falling water dropsIPX2, protection against falling water drops when enclosure is tilted at 15 degrees
Mode of OperationContinuous operation
  • Accuracy in term of root mean square (ARMS) is a statistical calculation of the difference between device measurements fell within +/- ARMs of the reference measurements in a controlled study.

** Upper and lower limits of agreement specifications have been validated in clinical study on healthy adult volunteers, with RD SET Disposable Sensors and Masimo Rainbow SET technology.

Intended Use/Indications for Use

The Masimo Rad-97 and Accessories are indicated for hospitals, hospital-type facilities, mobile, and home environments.

The Masimo Rad-97 and Accessories can communicate with network systems for supplemental remote viewing and alarming (e.g., at a central station).

The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRa). The Masimo Rad-97 and Accessories are indicated for use with adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused. In addition, the Masimo Rad-97 and Accessories are indicated to provide the continuous non-invasive monitoring data obtained from the Masimo Rad-97 and Accessories for functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) to multi-parameter devices for the display on those devices.

The Masimo Rad-97 and Accessories are not intended to be used as the sole basis for making diagnosis or treatment decisions related to suspected carbon monoxide poisoning; it is intended to be used in conjunction with additional methods of assessing clinical signs and symptoms.

{9}------------------------------------------------

The optional Nomoline Capnography product family is intended to be connected to other medical backboard devices for monitoring of breath rate and CO2. The Nomoline Capnography product family is intended to be connected to a patient breathing circuit for monitoring of inspired/expired gases during anesthesia, recovery and respiratory care. The environment is the operating suite, intensive care unit and patient room. The intent population is adult, pediatric and infant patients.

The optional non-invasive blood pressure (NIBP) module is indicated for the noninvasive measurement of arterial blood pressure. The NIBP module is designed to measure blood pressure for patient population described in the following table:

Patient PopulationApproximate Age Range
Newborn (neonate)Birth to 1 month of age
Infant1 month to 2 years of age
Child2 to 12 years of age
Adolescent12-21 years of age
Adult21 years of age and older

Technological Characteristics

Principle of Operation – Masimo Rainbow SET/ Masimo SET Technology

Optical Measurements (SpO2, Pulse Rate, SpCO, SpMet, SpHb, PI, PVI, SpOC) -The Masimo Rainbow SET technology utilizes the general principles of pulse oximetry for noninvasive optical measurements of SpO2, pulse rate, SpCO, SpMet and SpHb measurements. The perfusion index (PI), pleth variability index (PVI) and oxygen (SpOC) values are derived from these measurements. While the Masimo Rainbow SET technology includes all the above optical measurements, a subset of that technology, the Masimo SET technology, includes pulse oximetry measurements for SpO2, pulse rate, PI and PVI.

The Masimo Rainbow SET/ Masimo SET technology is governed by the following principles:

  • Oxyhemoglobin (oxygenated blood) and deoxyhemoglobin (non-oxygenated blood), carboxyhemoglobin (blood with carbon monoxide content), methemoglobin (blood with oxidized hemoglobin) and blood plasma constituents differ in their absorption of visible and infrared light (using spectrophotometry).

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white line going through it, followed by the word "Masimo" in black bold letters. To the right of the logo, the words "MASIMO CORPORATION" are written in gray, followed by the address "52 Discovery, Irvine, CA 92618".

  • The amount of arterial blood in tissue changes with pulse ● (photoplethysmography). Therefore, the amount of light absorbed by the varying quantities of arterial blood changes as well.
    The Masimo Rainbow SET technology uses a multi-wavelength sensor to distinguish between oxygenated blood, deoxygenated blood, blood with carbon monoxide, oxidized blood and blood plasma.

Acoustic Measurement (RRa) - The Masimo Rainbow Acoustic Monitoring (RAM) technology uses acoustic signals for respiration rate (RRa) measurements. RRa measures a patient's respiration rate based on airflow sounds generated in the upper airway. RRa translates airflow sounds generated in the upper airway to an electrical signal that can be processed to produce a respiration rate, measured as breaths per minute.

Principle of Operation - Capnography

The measurement of CO2 is based on the fact that different gases absorb infrared light at specific wavelengths. The analysis of respiratory gases is therefore performed by continuously measuring the infrared light absorption in the gas flow through an infrared spectrometer.

The infrared spectrometer uses a proprietary broadband infrared radiation source to transmit light through the gas sample. Before reaching the gas sample, the light path is intersected by narrowband optical filters that only let through light corresponding to selected wavelength peaks of the measured gases. At the other end of the light path, a sensor detects the portion of the light that is not absorbed by the gas. The amplitude of the detector output is an inverse function of the gas concentration. Thus, at a concentration of zero, the amplitude is at its maximum.

If the gas sample is a mixture of several components that absorb light at the same wavelength, such as a mixture of two anesthetic agents, the absorbed radiation will be the sum of the absorption of the agents. To determine the concentration of each of the individual gases, several filters have to be used. The spectrometer contains up to nine different narrowband filters to facilitate simultaneous measurement of different gases or a mixture of two gases.

Principle of Operation – Noninvasive Blood Pressure (NIBP)

The oscillometric method of blood pressure measurement is a non-invasive method that monitors the amplitude of cuff pressure changes during cuff deflation to

{11}------------------------------------------------

determine arterial blood pressure. The cuff pressure is first elevated above the patient systolic blood pressure level and the cuff begins to deflate at a certain rate. The initial rise in amplitude of these pressure fluctuations during cuff deflation corresponds closely to the systolic blood pressure. As the cuff is further deflated, these pressure fluctuations increase in amplitude until a peak is reached which is usually referred to as the mean arterial pressure (MAP). As cuff deflation continues, the diastolic pressure can be determined based upon the rapidly diminishing amplitude of the pressure fluctuations. Thus systolic, MAP and diastolic blood pressures can be accurately obtained by supervising the pressure fluctuations while controlling the cuff deflation rate.

Mechanism of Action for Achieving the Intended Effect

The device is turned "on" for its operation.

One end of a patient applied part (e.g. patient cable and SpO2 sensor, capnography sampling line) is connected to the subject device, and the other end of the patient applied part is connected to the patient. Physiological signals collected via patient applied part are sent to the subject device for processing. The subject device then displays the processed signals or measurements.

Data input devices (e.g. barcode scanner) can be used to enter patient identification information to associate the monitor to a patient. Data originating from a third party device can be displayed and transferred through the subject device onto a compatible surveillance monitoring system such as Patient SafetyNet (K071047) and/or EMR systems.

The optional microphone and camera can be used with compatible audio/video conferencing system. When configured with the optional Kite software, the display on the subject device can be re-displayed in near real time on a secondary display.

Once use is complete, the user then turns the power "off" in the device.

Summary of Technological Characteristics of Subject Device Compared to Predicate

The subject device, Rad-97 product family, and the primary predicate device, Root, have the following key similarities:

  • both have the same intended use as patient monitoring device;

{12}------------------------------------------------

  • both have the same principle of operation, mechanism of action for Masimo and performance specifications for Masimo Rainbow SET, capnography and NIBP technologies:
  • both have same input/output interfaces (e.g. USB, patient-applied part interface, wired/wireless connection) that allows connection with external devices and with networked systems;
  • both are compatible with the same applied patient parts (e.g. sensors, blood pressure cuffs):
  • o both have the same SatShare feature

The subject device, Rad-97 product family, and the primary predicate device, Root, have the following key differences:

  • the subject device, with Masimo Rainbow SET and NIBP technologies, is intended to ● be used in healthcare and home environments: whereas the predicate is only intended to be used in healthcare environments [Note, the Radical-7 module for the predicate is the same as the Radical-7 (K110028) as a standalone device indicated for home and hospital use]
  • the subject device comprises Masimo Rainbow SET, capnography and NIBP technologies; whereas the predicate comprises Sedline and body temperature technologies in addition to those included in the subject device; and
  • the subject device does not have docking and MOC-9 interfaces that are included ● with the predicate.

Non-clinical Testing

The following tests, as applicable, were performed for the qualification of the subject device, Rad-97 product family, in accordance with the requirements of the design control regulations and established quality assurance processes to demonstrate substantial equivalence with its predicates:

  • Electrical safety testing per IEC-60601-1
  • EMC testing per IEC-60601-1-2 ●
  • Usability testing per FDA Human Factors and Usability Draft Guidance
  • o Software verification, including integration testing with host monitors, per FDA Software Guidance
  • Mechanical testing per ISTA-2A and MIL-STD 810E
  • Environmental testing per IEC-60601-1 ●

{13}------------------------------------------------

Clinical Testing

To obtain the limits of agreement (LOA) information for the RD SET Disposable Sensors, a clinical study was performed in accordance with ISO-80601-2-61. The study was done on healthy adult volunteers to evaluate the sensor's performance for no motion condition, in the range of 70% to 100% in comparison to blood measurements from a laboratory CO-Oximeter. Using the Bland-Altman method, the mean difference between two methods of measurements (the 'bias') and 95% LOA as the mean difference, two standard deviations, (2 SD) for more precisely (1.96 SD)] was calculated. In this case, the 95% LOA calculation was between the noninvasive arterial oxygen (SpO2) measurements using RD SET Sensors and invasive measurements (blood draws). See the table below for a summary of the clinical results.

LOA (Results)LOA (Requirement)
Upper 95% LOA3.15%< 3%
Lower 95% LOA-2.21%> =3%

Conclusion

The results of the testing demonstrate that all requirements and performance specifications were satisfied and that the subject device is substantially equivalent to its predicate.

§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).

(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).