K Number
K163713

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-10-20

(294 days)

Product Code
Regulation Number
866.3830
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sofia Lyme FIA employs immunofluorescence for the rapid differential detection of human IgM and IgG antibodies to Borrelia burgdorferi from serum and plasma specimens from patients suspected of B. burgdorferi infection. This qualitative test is intended for use as an aid in the diagnosis of Lyme disease. A negative result does not preclude infection with B. burgdorferi. All positive results for IgM and/or IgG should be further tested by a corresponding second-tier western blot assay. Test results are to be used in conjunction with information obtained from the patient's clinical evaluation and other diagnostic procedures.

The Sofia Lyme Control Set is intended for use as assayed quality control materials to verify the performance of the Sofia Lyme FIA test system.

Device Description

The Sofia Lyme FIA is an immunofluorescence-based, lateral flow assay for detection of IgM and/or IgG antibodies to Borrelia burgdorferi in patient specimens. Reagents for the assay are ready-to-use and provided in the kit.

The assay uses a bidirectional test strip format to detect both IgM and IgG antibodies to B. burgdorferi. One side of the test strip detects IgM antibodies to B. burgdorferi and the other side of the test strip detects IgG antibodies to B. burgdorferi.

To perform the test, the patient specimen (30 uL serum/plasma) is added to a pre-filled Reagent Vial. 100 uL of the diluted sample is then dispensed into the round sample well located near the center of the Test Cassette. The Test Cassette is loaded into Sofia for an automatically defined development time (WALK AWAY Mode) or pre-incubated on the bench top prior to loading into Sofia (READ NOW Mode). Sofia scans the test strip, analyzes the fluorescent signal, and then displays two (2) test results: IgM (positive, negative or invalid) and IgG (positive, negative or invalid). As an option, the results are automatically printed on the integrated printer.

Each Sofia Lyme FIA kit will contain one Positive and one Negative Control—each provided in separate dropper bottles. The external controls will be provided separately as well in a Sofia Lyme Control Set. The Positive and Negative QC controls are formulated with patient Lyme IgM and IgG positive serum that are diluted into 1X PBS and 0.3% Microcide is added to the solution as an antimicrobial. The External Controls will be tested by adding 2 drops to the test cassette.

AI/ML Overview

The provided text describes the 510(k) summary for the Sofia Lyme FIA and Sofia Lyme Control Set, which employs immunofluorescence for the rapid differential detection of human IgM and IgG antibodies to Borrelia burgdorferi. The device is intended as an aid in the diagnosis of Lyme disease.

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state numerical "acceptance criteria" for the performance studies in the format of a predefined threshold. Instead, it describes various studies undertaken and concludes that these studies "demonstrated the substantial equivalence" to the predicate devices (Vidas Lyme IgG and Vidas Lyme IgM) and that there was "good performance," "comparable performance," and "very good" specificity.

Performance CharacteristicAcceptance Criteria (Implicit from "Conclusion" and study descriptions)Reported Device Performance
Overall PerformanceDemonstrate substantial equivalence to predicate devices (Vidas Lyme IgM and Vidas Lyme IgG) in terms of safety and effectiveness."These studies demonstrated the substantial equivalence of the Sofia Lyme FIA to the Vidas Lyme IgG and Vidas Lyme IgM tests. Such studies are a critical element in establishing the fundamental safety and effectiveness of the product and its appropriateness for commercial distribution."
Analytical StudiesGood performance in Limit of Detection, cross-reactivity, interfering substances, specimen handling, operating temperature, and various flex studies."All studies demonstrated good performance."
PrecisionTotal precision results for IgM and IgG not significantly different within-run, within-day, between day and total for various sample concentrations (negative, high negative, low positive, moderate positive)."The total precision results for IgM and IgG were not significantly different within-run, within-day, between day and total when tested with the negative (C0), high negative (C5), low positive (C95) and moderate positive (2-3X LOD) samples."
CDC Reference PanelGood overall agreement for Clinical Status of IgG and IgM samples between Sofia Lyme FIA and Western Blot."When testing the CDC reference panel, the overall agreement for Clinical Status of IgG and IgM samples between Sofia Lyme FIA and Western Blot was good."
Clinical SensitivityComparable performance to a 510(k) cleared test for well-characterized clinically or culture-confirmed Lyme disease samples."The clinical sensitivity was compared to a 510(k) cleared test and the performance was comparable."
Clinical SpecificityVery good specificity using serum from asymptomatic (healthy, normal) populations in endemic and non-endemic regions."The overall specificity of the Sofia Lyme FIA was very good."
Method ComparisonComparable performance to Vidas Lyme IgG and Vidas Lyme IgM tests when testing prospectively collected serum specimens from subjects suspected of Lyme disease."This study demonstrated that Sofia Lyme FIA has comparable performance to the Vidas Lyme IgG and Vidas Lyme IgM tests when testing prospectively collected serum specimens from subjects suspected of having Lyme disease."
ReproducibilityAccurate results obtained by operators across laboratories with a panel of test samples at various concentrations."The operators and laboratories obtained accurate results with the Sofia Lyme FIA."

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the exact sample sizes for each test set in the "Performance Data" section.

  • For the Clinical Sensitivity Study, it used "well-characterized clinically or culture confirmed Lyme disease samples."
  • For the Specificity Study, it used "serum obtained from asymptomatic (healthy, normal) populations in both endemic and non-endemic regions."
  • For the Method Comparison, it used "prospectively collected serum specimens from subjects suspected of having Lyme disease."

The provenance of the data (country of origin, retrospective/prospective) is partially mentioned:

  • The Specificity Study included data from both endemic and non-endemic regions, suggesting a diverse geographical origin for these samples.
  • The Method Comparison study used "prospectively collected serum specimens." No explicit mention of the country(ies) of origin is given beyond "endemic and non-endemic regions."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not specify the "number of experts" or their "qualifications" involved in establishing ground truth for the test sets.

  • For the CDC Reference Panel, the ground truth is implicitly established by the CDC and Western Blot results, which are established methods for Lyme disease confirmation.
  • For the Clinical Sensitivity Study, samples were "well-characterized clinically or culture confirmed Lyme disease samples." This implies a clinical diagnosis and/or laboratory confirmation (e.g., culture), but it doesn't detail the number or qualifications of clinicians/pathologists involved.

4. Adjudication Method for the Test Set

The document does not explicitly describe an "adjudication method" (e.g., 2+1, 3+1, none) for resolving discrepancies in ground truth establishment for the test set. Ground truth appears to be based on established clinical/laboratory practices (clinical diagnosis, culture, Western Blot).

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

No Multi Reader Multi Case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance is mentioned. The device, Sofia Lyme FIA, is an automated diagnostic test that provides a reading directly, not an AI-assisted interpretation for human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, the studies described are standalone performance studies of the Sofia Lyme FIA device. The device uses immunofluorescence and an automated analyzer to produce results ("Sofia scans the test strip, analyzes the fluorescent signal, and then displays two (2) test results"). There is no mention of a human-in-the-loop for interpreting the device's primary output. The results are then to be "used in conjunction with information obtained from the patient's clinical evaluation and other diagnostic procedures," indicating that while the device is standalone, its results are part of a broader diagnostic process.

7. The Type of Ground Truth Used

The ground truth used for validation includes:

  • Established Reference Methods: Western Blot (for the CDC reference panel).
  • Clinical Diagnosis/Culture Confirmation: For the clinical sensitivity study, "well-characterized clinically or culture confirmed Lyme disease samples" were used.

8. The Sample Size for the Training Set

The document does not provide information regarding a separate "training set" or its sample size. This type of regulatory submission typically focuses on validation and clinical performance of the finalized device, rather than the developmental training phase of an algorithm.

9. How the Ground Truth for the Training Set was Established

Since no specific "training set" is mentioned or detailed, there is no information provided on how its ground truth was established.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 20, 2017

QUIDEL CORPORATION JENNIFER RIAL DIRECTOR, REGULATORY AFFAIRS 12544 HIGH BLUFF DRIVE, SUITE 200 SAN DIEGO CA 92130

Re: K163713

Trade/Device Name: Sofia Lyme Fia. Sofia Lyme Control Set Regulation Number: 21 CFR 866.3830 Regulation Name: Treponema pallidum treponemal test reagents Regulatory Class: II Product Code: LSR Dated: May 17, 2017 Received: May 19, 2017

Dear Ms. Rial:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Kristian M. Roth -S

For:

Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K163713

Device Name Sofia Lyme FIA Sofia Lyme Control Set

Indications for Use (Describe)

The Sofia Lyme FIA employs immunofluorescence for the rapid differential detection of human IgM and IgG antibodies to Borrelia burgdorferi from serum and plasma specimens from patients suspected of B. burgdorferi infection. This qualitative test is intended for use as an aid in the disease. A negative result does not preclude infection with B. burgdorferi. All positive results for IgG should be further tested by a corresponding second-tier western blot assay. Test results are to be used in conjunction with information obtained from the patient's clinical evaluation and other diagnostic procedures.

The Sofia Lyme Control Set is intended for use as assayed quality control materials to verify the performance of the Sofia Lyme FIA test system.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for QuidelOrtho Corporation. The logo features a square with rounded corners, divided into two sections. The top section is a gradient of colors, including red, orange, yellow, green, blue, and purple. The bottom section is a smaller square that is blue and purple. Below the square is the word "QUIDEL" in a sans-serif font.

5. 510(K) SUMMARY

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for QuidelOrtho Corporation. The logo features a square with a smaller square cut out of the bottom right corner. The larger square is filled with a gradient of colors, including red, orange, yellow, green, blue, and purple. The smaller square is filled with a gradient of blue and purple. Below the square is the word "QUIDEL" in a sans-serif font.

5.1. Submitter

Quidel Corporation 12544 High Bluff Drive, Suite 200 San Diego, California 92130 Telephone: 858-552-7910 Fax: 858-646-8045

5.2. Submission Contact

Jennifer S. Rial

5.3. Date Prepared/Updated

October 18, 2017

5.4. Proprietary and Established Names

Sofia Lyme FIA Sofia Lyme Control Set

5.5. Common Name

Lyme IgG and Lyme IgM test and assayed external controls

5.6. Regulatory Information

Product CodeClassificationRegulatory SectionPanel
LSRII21 CFR 866.3830Immunology andMicrobiology Devices

5.7. Predicate Device

Vidas Lyme IgG and Vidas Lyme IgM

5.8. Device Description

The Sofia Lyme FIA is an immunofluorescence-based, lateral flow assay for detection of IgM and/or IgG antibodies to Borrelia burgdorferi in patient specimens. Reagents for the assay are ready-to-use and provided in the kit.

The assay uses a bidirectional test strip format to detect both IgM and IgG antibodies to B. burgdorferi. One side of the test strip detects IgM antibodies to B. burgdorferi and the other side of the test strip detects IgG antibodies to B. burgdorferi.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for QuidelOrtho Corporation. The logo features a square with a gradient of colors, including red, orange, yellow, green, blue, and purple. Inside the square is a stylized letter "Q" in white. Below the square is the word "QUIDEL" in a simple, sans-serif font.

To perform the test, the patient specimen (30 uL serum/plasma) is added to a pre-filled Reagent Vial. 100 uL of the diluted sample is then dispensed into the round sample well located near the center of the Test Cassette. The Test Cassette is loaded into Sofia for an automatically defined development time (WALK AWAY Mode) or pre-incubated on the bench top prior to loading into Sofia (READ NOW Mode). Sofia scans the test strip, analyzes the fluorescent signal, and then displays two (2) test results: IgM (positive, negative or invalid) and IgG (positive, negative or invalid). As an option, the results are automatically printed on the integrated printer.

Each Sofia Lyme FIA kit will contain one Positive and one Negative Control—each provided in separate dropper bottles. The external controls will be provided separately as well in a Sofia Lyme Control Set. The Positive and Negative QC controls are formulated with patient Lyme IgM and IgG positive serum that are diluted into 1X PBS and 0.3% Microcide is added to the solution as an antimicrobial. The External Controls will be tested by adding 2 drops to the test cassette.

5.9. Intended Use

Sofia Lyme FIA:

The Sofia Lyme FIA employs immunofluorescence for the rapid differential detection of human IgM and IgG antibodies to Borrelia burgdorferi from serum and plasma specimens from patients suspected of B. burgdorferi infection. This qualitative test is intended for use as an aid in the diagnosis of Lyme disease. A negative result does not preclude infection with B. burgdorferi. All positive results for IgM and/or IgG should be further tested by a corresponding second-tier western blot assay. Test results are to be used in conjunction with information obtained from the patient's clinical evaluation and other diagnostic procedures.

Sofia Lyme Control Set: The Sofia Lyme Control Set is intended for use as assayed quality control materials to verify the performance of the Sofia Lyme FIA test system.

5.10. Substantial Equivalence Information:

  • Predicate Device Name: Vidas Lyme IgM and Vidas Lyme IgG 1.
    1. Predicate 510(k) Numbers: K122979 and K122986

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for QuidelOrtho Corporation. The logo features a square with a smaller square cut out of the lower right corner. The larger square is filled with a mosaic of colors, including yellow, orange, red, purple, blue, and green. Below the square is the word "QUIDEL" in a simple, sans-serif font.

ItemProposed DevicePredicate IgM DevicePredicate IgG Device
FeaturesSofia Lyme FIA withSofiaBiomerieux Vidas LymeIgMBiomerieux Vidas LymeIgG
Intended UseSofia Lyme FIA:The Sofia Lyme FIAemploysimmunofluorescence forthe rapid differentialdetection of human IgMand IgG antibodies toBorrelia burgdorferi fromserum and plasmaspecimens from patientssuspected of B.burgdorferi infection. Thisqualitative test is intendedfor use as an aid in thediagnosis of Lyme disease.A negative result does notpreclude infection with B.burgdorferi. All positiveresults for IgM and/or IgGshould be further tested bya corresponding second-tier western blot assay.Test results are to be usedin conjunction withinformation obtained fromthe patient's clinicalevaluation and otherdiagnostic procedures.Sofia Lyme Control Set:The Sofia Lyme ControlSet is intended for use asassayed quality controlmaterials to verify theperformance of the SofiaLyme FIA test system.The VIDAS Lyme IgMassay is an automatedqualitative enzymeimmunoassay intended foruse on the instruments of theVIDAS family in thepresumptive detection ofhuman IgM antibodies toBorrelia burdorferi inhuman serum or plasma. Itshould be used to testpatients with a historyand/or symptoms ofinfection with B.burgdorferi. All VIDASLyme IgM positivespecimens should be furthertested with a Western BlotIgG assay to obtainsupportive evidence ofinfection with B.burgdorferi.The VIDAS Lyme IgGassay is an automatedqualitative enzymeimmunoassay intended foruse on the instruments of theVIDAS family in thepresumptive detection ofhuman IgG antibodies toBorrelia burdorferi inhuman serum or plasma. Itshould be used to testpatients with a historyand/or symptoms ofinfection with B.burgdorferi. All VIDASLyme IgG positivespecimens should be furthertested with a Western BlotIgG assay to obtainsupportive evidence ofinfection with B.burgdorferi.
InstrumentSofiaVIDAS and miniVIDASVIDAS and miniVIDAS
AnalyteHuman IgM and IgGantibodies against B.burgdorferi proteinsHuman IgM antibodiesagainst B. burgdorferiproteinsHuman IgG antibodiesagainst B. burgdorferiproteins
AutomatedAnalysisYesYesYes
ItemProposed DevicePredicate IgM DevicePredicate IgG Device
FeaturesSofia Lyme FIA with SofiaBiomerieux Vidas Lyme IgMBiomerieux Vidas Lyme IgG
Read ResultsRead results on instrument screen or print with optional printerResult report is printedResult report is printed
Read Result Time10 Minutes27 minutes27 minutes
Specimen TypesHuman serum and plasmaHuman serum and plasmaHuman serum and plasma
QualitativeYesYesYes
Test PrincipleImmunofluorescence DeviceImmunofluorescence DeviceImmunofluorescence Device
FormatLateral-flow Bi-directional Test CassetteEnzyme-linked fluorescent assay (ELFA)Enzyme-linked fluorescent assay (ELFA)
Antibodies UsedMonoclonal anti-human IgG and polyclonal anti-human IgMAnti-human IgM antibodiesAnti-human IgG antibodies
Antigens UsedRecombinant Proteins and synthetic peptides of B. burgdorferiRecombinant proteins of B. burgdorferiRecombinant proteins of B. burgdorferi
Detection MethodPolystyrene microparticles dyed with Europium chelateAlkaline phosphatase/4-MUPAlkaline phosphatase/4-MUP
StorageRoom Temperature (15-30°C)2-8°C2-8°C
Running Buffer SolutionOne pre-filled vial containing PBSSample diluent and wash bufferSample diluent and wash buffer
Quality Control FeaturesBuilt-in features include:• Built-in procedural control zone scanned by the analyzer to determine whether adequate flow occurred on the IgG side of the assay.• Built-in reference control line scanned by the analyzer to determine whether adequate flow occurred on the IgM• One positive and one negative control are included and must be tested after opening a new kit to monitor reagent failure.• One positive and one negative control are included and must be tested after opening a new kit to monitor reagent failure.
ItemProposed DevicePredicate IgM DevicePredicate IgG Device
FeaturesSofia Lyme FIA withSofiaBiomerieux Vidas LymeIgMBiomerieux Vidas LymeIgG
side of the assay.• Analyzer preventsused or expiredcartridge from beingread by the reader• Cassette must beproperly inserted

Comparison with Predicate

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Quidel. The logo features a square with rounded corners, filled with a gradient of colors from red to yellow to green to blue. Inside the square is a smaller white square, with a portion of the colored square visible in the lower right corner. Below the square is the word "QUIDEL" in a simple, sans-serif font.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image features the logo for Quidel. The logo consists of a square with rounded corners, filled with a gradient of colors ranging from yellow and orange to red, purple, and blue. A smaller, solid white square is positioned in the upper left corner of the larger square, creating a visual break. Below the square is the word "QUIDEL" in a simple, sans-serif font.

5.11. Performance Data

Numerous studies were undertaken to document the performance characteristics of Sofia Lyme FIA and the Sofia Lyme Control Set, as well as to compare the performance between Sofia Lyme and the Vidas IgG and IgM tests. The studies included the following:

  • a. Various Analytical Studies
    Analytical studies included Limit of Detection, cross-reactivity, interfering substances, specimen collection and handling, operating temperature, and various flex studies. All studies demonstrated good performance.

  • b. Precision
    This study evaluated the within laboratory precision / repeatability of the Sofia Lyme FIA using serum samples. The total precision results for IgM and IgG were not significantly different within-run, within-day, between day and total when tested with the negative (C0), high negative (C5), low positive (C95) and moderate positive (2-3X LOD) samples.

  • CDC Reference Panel C.
    When testing the CDC reference panel, the overall agreement for Clinical Status of IgG and IgM samples between Sofia Lyme FIA and Western Blot was good.

  • Sensitivity Study d.
    This study evaluated the clinical sensitivity of the Sofia Lyme FIA device using wellcharacterized clinically or culture confirmed Lyme disease samples. The clinical sensitivity was compared to a 510(k) cleared test and the performance was comparable.

  • Specificity Study e.
    This study evaluated the analytical specificity of the Sofia Lyme FIA using serum obtained from asymptomatic (healthy, normal) populations in both endemic and nonendemic regions. The overall specificity of the Sofia Lyme FIA was very good.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for QuidelOrtho Corporation. The logo features a square with rounded corners, divided into four quadrants. The top and left quadrants are filled with a gradient of colors, including green, yellow, orange, and red. The bottom right quadrant is filled with shades of blue and purple. Below the square is the word "QUIDEL" in a modern, sans-serif font.

  • f. Method Comparison
    This study demonstrated that Sofia Lyme FIA has comparable performance to the Vidas Lyme IgG and Vidas Lyme IgM tests when testing prospectively collected serum specimens from subjects suspected of having Lyme disease.

  • g. Reproducibility
    This study demonstrated intra- and inter-laboratory reproducibility with a panel of test samples at various concentrations of IgM and IgG antibodies to B. burgdorferi. The operators and laboratories obtained accurate results with the Sofia Lyme FIA.

5.12. Conclusion

These studies demonstrated the substantial equivalence of the Sofia Lyme FIA to the Vidas Lyme IgG and Vidas Lyme IgM tests. Such studies are a critical element in establishing the fundamental safety and effectiveness of the product and its appropriateness for commercial distribution.

§ 866.3830

Treponema pallidum treponemal test reagents.(a)
Identification. Treponema pallidum treponemal test reagents are devices that consist of the antigens, antisera and all control reagents (standardized reagents with which test results are compared) which are derived from treponemal sources and that are used in the fluorescent treponemal antibody absorption test (FTA-ABS), theTreponema pallidum immobilization test (T.P.I.), and other treponemal tests used to identify antibodies toTreponema pallidum directly from infecting treponemal organisms in serum. The identification aids in the diagnosis of syphilis caused by bacteria belonging to the genusTreponema and provides epidemiological information on syphilis.(b)
Classification. Class II (performance standards).