K Number
K122979

Validate with FDA (Live)

Device Name
VIDAS LYME IGM
Manufacturer
Date Cleared
2013-06-04

(251 days)

Product Code
Regulation Number
866.3830
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VIDAS Lyme IgM (LYM) assay is an automated qualitative enzyme immunoassay intended for use on the instruments of the VIDAS family in the presumptive detection of human IgM antibodies to Borrelia burgdorferi in human serum (plain or separation gel) or plasma (sodium heparin or lithium heparin). It should be used to test patients with a history and/or symptoms of infection with B. burgdorferi. All VIDAS Lyme IgM positive and equivocal specimens should be further tested with a Western Blot IgM assay to obtain supportive evidence of infection with B. burgdorferi.

Device Description

The VIDAS Lyme IgM assay principle combines a 2-step enzyme immunoassay sandwich method with a final fluorescent detection (ELFA) (see User's Manual).

The Solid Phase Receptacle (SPR®) serves as the solid phase as well as the pipetting device for the assay. Reagents for the assay are ready-to-use and predispensed in the sealed reagent strips.

All of the assay steps are performed automatically by the instrument. The reaction medium is cycled in and out of the SPR several times.

After a preliminary wash step and a sample dilution step, the antibodies to B. burgdorferi present in the specimen will bind to the B. burgdorferi specific recombinant proteins coating the interior of the SPR.

Unbound sample components are washed away. Anti-human IgM antibodies conjugated with alkaline phosphatase, will attach to the immunocomplex bound to the SPR wall.

A final wash step removes unbound conjugate. During the final detection step, the substrate (4-Methyl-umbellifery) phosphate) is cycled in and of the SPR. The conjugate enzyme catalyzes the hydrolysis of this substrate into a fluorescent product (4-Methyl-umbelliferone) the fluorescence of which is measured at 450 nm. The intensity of the fluorescence is proportional to the quantity of anti-B. burgdorferi IgM antibody present in the sample.

At the end of the VIDAS Lyme IgM assay, results are automatically calculated by the instrument. A test value is generated and a report is printed.

AI/ML Overview

The provided text describes the 510(k) summary for the VIDAS® Lyme IgM Assay. Here's a breakdown of the acceptance criteria and study information provided:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of acceptance criteria for performance metrics. Instead, it presents the device's performance results, which are then compared to a predicate device and established Lyme disease testing methodologies. The "acceptance" can be inferred from the statement in Section I, "Conclusion," which states that the results "demonstrate that the VIDAS® Lyme IgM is substantially equivalent to the predicate device."

However, we can extract the key performance metrics reported in relation to the predicate and other ground truth methods:

Performance MetricAcceptance Criteria (Implied / Comparator)Reported Device Performance (VIDAS Lyme IgM)Study/Section
Sensitivity (Stage I Lyme Disease)Compared to Predicate (Platelia™ Lyme IgM)52.1% (62/119)Clinical Testing - Sensitivity
Sensitivity (Stage II Lyme Disease)Compared to Predicate (Platelia™ Lyme IgM)91.9% (57/62)Clinical Testing - Sensitivity
Sensitivity (Stage III Lyme Disease)Compared to Predicate (Platelia™ Lyme IgM)76.2% (16/21)Clinical Testing - Sensitivity
Overall Sensitivity (All Stages)Compared to Predicate (Platelia™ Lyme IgM)66.8% (135/202)Clinical Testing - Sensitivity
First Tier Positive % Agreement (PPA)Compared to Predicate Lyme IgM51.0% (107/210)Clinical Testing - Method Comparison
Negative % Agreement (NPA)Compared to Predicate Lyme IgM88.6% (678/765)Clinical Testing - Method Comparison
Second Tier Positive % Agreement (PPA)Compared to Predicate Lyme IgM + Western Blot IgM80.8% (84/104)Clinical Testing - Second-Tier Testing
Concordance with IgM Western BlotNot explicitly stated as a target, but reported for comparison.49.0% (95/194)Clinical Testing - Second-Tier Testing
Cross-Reactivity (Overall Range)Not explicitly stated, but reported for various infections/diagnoses.Ranges from 4.34% (Mumps) to 35.00% (HIV)Clinical Testing - Cross-Reactivity
Precision (Total CV%)Not explicitly stated, but clinical data providedSample 1: 5.2%; Sample 2: 9.1%; Sample 3: 4.7%; Sample 4: 6.9%; Positive Control: 10.1%; Negative Control: 0.0%Nonclinical Tests - Precision
Reproducibility (Total CV%)Not explicitly stated, but clinical data providedSample 1: 6.3%; Sample 2: 7.8%; Sample 3: 6.2%; Sample 4: 9.4%; Positive Control: 12.1%; Negative Control: 0.0%Nonclinical Tests - Reproducibility

2. Sample Size for the Test Set and Data Provenance

Sensitivity Testing (Retrospective):

  • Sample Size: 202 retrospective samples.
  • Data Provenance: Samples were from patients meeting a case definition of Lyme Disease (LD) and confirmed positive for B. burgdorferi infection. The country of origin is not specified but implied to be regions with LD prevalence due to the context of "clinical testing." These were clearly retrospective samples.

Method Comparison (Prospective):

  • Sample Size: 975 fresh or frozen prospectively collected sera.
  • Data Provenance: Samples were submitted for routine Lyme disease testing from an endemic area of the United States. These were prospective samples.

Analytical Specificity:

  • Sample Size: 100 sera from apparently healthy subjects from an endemic population (New York) and 100 sera from a non-endemic population (Texas).
  • Data Provenance: Healthy subjects from New York (endemic) and Texas (non-endemic). These are likely retrospective collections of healthy individuals.

CDC Reference Panel:

  • Sample Size: 39 samples (5 Normals, 5 < 1 month, 6 1-2 months, 16 3-12 months, 7 > 1 year).
  • Data Provenance: A serum panel obtained from the CDC. The origin of the individual samples within the panel is not detailed but is typically well-characterized retrospective samples.

Cross-Reactivity Testing:

  • Sample Size: Varied per infection/diagnosis, ranging from 20 (HIV) to 270 (Syphilis). Total samples across all categories would be the sum (e.g., 60 ANA, 61 CRP, etc.).
  • Data Provenance: Samples from patients negative for Lyme but positive for various other conditions. The country of origin is not specified but implies diverse patient populations to cover various diseases. These are likely retrospective collections.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

The document does not explicitly state the number of experts or their specific qualifications (e.g., "radiologist with 10 years of experience") used to establish the ground truth for the test set.

However, the ground truth for the Sensitivity Testing refers to "patients meeting a case definition of LD and confirmed positive for B. Burgdorferi infection." This implies a clinical diagnosis, potentially relying on multiple factors and clinical judgment.

For the Second-Tier Testing, the ground truth for confirmation was a "commercially available Lyme IgM Western Blot method." While Western Blot itself is a laboratory test, its interpretation often involves expertise. The document doesn't specify if expert consensus was used for Western Blot interpretation in this study.

4. Adjudication Method for the Test Set

The document does not describe an explicit adjudication method (like 2+1 or 3+1 consensus) for establishing the ground truth or resolving discrepancies between readers/tests.

For the sensitivity and method comparison studies, the performance of the VIDAS assay was compared to either a "case definition" (for sensitivity) or a "predicate Lyme IgM EIA method" and subsequently "Western Blot IgM assay" for confirmation. The document implies that the results of these comparator methods, particularly Western Blot for the second tier, served as the reference standard without detailing a multi-expert adjudication process.

For both the VIDAS Lyme IgM and the predicate test in the sensitivity study, "equivocal results were considered as positive for the evaluation." This is a defined rule for handling equivocal results rather than an adjudication process.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The study focuses on evaluating the performance of an automated laboratory assay (VIDAS Lyme IgM) against a predicate device and established diagnostic algorithms (e.g., 2-tier testing with Western Blot), not on the comparative effectiveness of human readers with and without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, a standalone performance evaluation was done. The VIDAS® Lyme IgM assay is an automated qualitative enzyme immunoassay. The studies (precision, reproducibility, sensitivity, method comparison, analytical specificity, cross-reactivity, CDC reference panel) evaluate the performance of this assay as a device in itself, producing results automatically by the instrument, without a human-in-the-loop component for result generation or primary interpretation in the context of these evaluations. The output is a numerical index and a qualitative result (positive, equivocal, negative).

7. The Type of Ground Truth Used

The types of ground truth used include:

  • Clinical Case Definition/Confirmed Infection: For the sensitivity study, samples were from "patients meeting a case definition of LD and confirmed positive for B. Burgdorferi infection." This implies a combination of clinical symptoms and other diagnostic evidence.
  • Predicate Device/Western Blot IgM Assay: For the method comparison and second-tier testing, the predicate Platelia™ Lyme IgM assay and a "commercially available Lyme IgM Western Blot method" were used as reference standards for comparison and confirmation, respectively.
  • Known Health Status/Other Infection: For analytical specificity and cross-reactivity, subjects were either "apparently healthy" or known to be infected with other specific pathogens/conditions while being negative for Lyme disease.
  • CDC Characterized Serum Panel: A masked serum panel from the CDC, which is typically characterized through extensive clinical and laboratory data, served as another reference.

8. The Sample Size for the Training Set

The document does not provide any information regarding a training set size. This is typical for a 510(k) summary for an immunoassay, which does not employ machine learning or AI models that require specific training sets in the same way. The device's underlying "algorithm" (ELFA technique) is based on biochemical reactions, not data-driven machine learning.

9. How the Ground Truth for the Training Set was Established

Since no training set is mentioned (as the device is not an AI/ML system), the establishment of ground truth for a training set is not applicable to this device.

{0}------------------------------------------------

K122979

510(k) SUMMARY

VIDAS® Lyme IgM Assay

510(k) SUMMARY

JUN 0 4 2013

1

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirement of SMDA 1990 and 21 CFR 807.92.

VIDAS® Lyme IgM

A. Submitter Information

Submitter's Name:bioMérieux SA
Address:Chemin de l'Orme69280 Marcy-l'Etoile - France
Contact Person:Catherine FRITSCH
Phone Number:+33 4 78 87 23 98
Fax Number:+33 4 78 87 20 75
Date of Preparation:August 2012

B. Device Name

Trade Name:VIDAS® Lyme IgM
Common Name:Lyme IgM Assay
Classification Name:21 CFR 866.3830 - Treponema pallidum treponemal test reagents

C. Predicate Device Name

Trade Name: Platelia™ Lyme IgM

D. Device Description

The VIDAS Lyme IgM assay principle combines a 2-step enzyme immunoassay sandwich method with a final fluorescent detection (ELFA) (see User's Manual).

The Solid Phase Receptacle (SPR®) serves as the solid phase as well as the pipetting device for the assay. Reagents for the assay are ready-to-use and predispensed in the sealed reagent strips.

All of the assay steps are performed automatically by the instrument. The reaction medium is cycled in and out of the SPR several times.

After a preliminary wash step and a sample dilution step, the antibodies to B. burgdorferi present in the specimen will bind to the B. burgdorferi specific recombinant proteins coating the interior of the SPR.

Unbound sample components are washed away. Anti-human IgM antibodies conjugated with alkaline phosphatase, will attach to the immunocomplex bound to the SPR wall.

A final wash step removes unbound conjugate. During the final detection step, the substrate (4-Methyl-umbellifery) phosphate) is cycled in and out of the SPR. The conjugate enzyme catalyzes the hydrolysis of this substrate into a fluorescent product (4-Methyl-umbelliferone) the fluorescence of which is measured at 450 nm. The intensity of the fluorescence is proportional to the quantity of anti-B. burgdorferi IgM antibody present in the sample.

At the end of the VIDAS Lyme IgM assay, results are automatically calculated by the instrument. A test value is generated and a report is printed.

{1}------------------------------------------------

E. Intended Use

The VIDAS Lyme IgM (LYM) assay is an automated qualitative enzyme immunoassay intended for use on the instruments of the VIDAS family in the presumptive detection of human IgM antibodies to Borrelia burgdorferi in human serum (plain or separation gel) or plasma (sodium heparin or lithium heparin). It should be used to test patients with a history and/or symptoms of infection with B. burgdorferi. All VIDAS Lyme IgM positive and equivocal specimens should be further tested with a Western Blot IgM assay to obtain supportive evidence of infection with B. burgdorferi.

F. Technological Characteristics Summary

A general comparison of the similarities and differences of the assays is presented in the table below.

ItemVIDAS® Lyme IgM (LYM) AssayPlatelia™ Lyme IgM (K081362)
Intended UseThe VIDAS Lyme IgM (LYM) assayis an automated qualitative enzymeimmunoassay intended for use onthe instruments of the VIDAS familyin the presumptive detection ofhuman IgM antibodies to Borreliaburgdorferi in human serum (plain orseparation gel) or plasma (sodiumheparin or lithium heparin). It shouldbe used to test patients with ahistory and/or symptoms of infectionwith B. burgdorferi. All VIDAS LymeIgM positive and equivocalspecimens should be further testedwith a Western Blot IgM assay toobtain supportive evidence ofinfection with B. burgdorferi.The Platelia™ Lyme IgM Test is aqualitative test intended for use inthe presumptive detection ofhuman IgM antibodies to Borreliaburgdorferi in human serum orplasma (K3 EDTA, sodiumheparin or sodium citrate). TheEIA system should be used to testserum or plasma from patientswith a history and symptoms ofinfection with Borrelia burgdorferi.All positive and equivocalspecimens should be re-testedwith a specific, second-tier testsuch as Western-Blot. Positivesecond-tier results are supportiveevidence of infection with Borreliaburgdorferi. The diagnosis ofLyme disease should be madebased on history and symptoms(such as erythema migrans), andother laboratory data, in additionto the presence of antibodies toBorreliaburgdorferi. Negative results(either first- or second-tier) shouldnot be used to exclude Lymedisease.
SpecimenSerum or plasmaSerum or plasma
AnalyteIgM antibodies to Borrelia burgdorferiIgM antibodies to Borreliaburgdorferi
AutomatedYesNo
AssayTechniqueEnzyme-linked fluorescent assay(ELFA)Enzyme immunoassay (EIA)

{2}------------------------------------------------

G. Nonclinical Tests

A summary of the non-clinical results is presented below.

Precision

For the precision study, 4 serum samples were tested in duplicate in 40 different runs (2 runs per day over 20 days) with 2 reagent lots at 1 site (n = 80). The precision was calculated following the recommendations of the CLS10 document EP5-A2. The total precision data in the table reflect the 80 values generated per sample for Site 1 and takes into account replicate, run, day, calibration, and lot as potential sources of variation. The total precision for controls include within-day, between-days and between-calibration variability and is lot specific.

Panel MemberNMeanIndexWithin-runWithin-dayBetween-daysTotal
SDCV (%)SDCV (%)SDCV (%)SDCV (%)
Sample 1800.190.013.90.012.70.000.00.015.2
Sample 2800.270.014.70.012.80.001.40.029.1
Sample 3800.380.012.60.013.00.000.30.024.7
Sample 4801.310.032.10.021.60.010.90.096.9
Positive Control400.74NANA0.045.30.000.00.0710.1
Negative Control400.00NANA0.000.00.000.00.000.0

Reproducibility

For reproducibility, 4 serum samples were tested in 40 different runs (2 runs per day over 20 days) with 2 reagent lots at 3 sites (n =240). The reproducibility was calculated following the recommendations of the CLSI® document EP5-A2. The total reproducibility data in the table reflects the 240 values generated per sample for all sites and takes into account replicate, run, day, calibration, lot, and site as potential sources of variation. Out of the 240 total values, 2 Low Positives (Sample 3) gave an equivocal value (< 0.32). The total reproducibility for controls include within-day, between-days, between-calibration and between-site variability and is lot specific.

PanelMemberNMeanIndexWithin-runWithin-dayBetween-daysBetween-siteTotal
SDCV(%)SDCV(%)SDCV(%)SDCV(%)SDCV(%)
Sample12400.190.013.50.002.10.001.00.000.00.016.3
Sample22400.260.014.30.012.70.000.70.012.10.027.8

3

{3}------------------------------------------------

510(k) SUMMARY

VIDAS® Lyme IgM Assay

PanelMemberNMeanIndexWithin-runWithin-dayBetween-daysBetween-siteTotal
SDCV(%)SDCV(%)SDCV(%)SDCV(%)SDCV(%)
Sample32400.370.013.10.012.10.000.00.000.00.026.2
Sample42401.260.032.50.021.90.010.40.000.00.129.4
PositiveControl1200.72NANA0.034.60.000.00.000.00.0912.1
NegativeControl1200.00NANA0.000.00.000.00.000.00.000.0

Interfering Substances

Specimen-related interference

Interferences were studied according to the recommendations of CLS1® document EP7-A2. None of the following factors have been found to significantly influence this assay:

  • hemolysis (after spiking samples with hemoglobin: 5 g/L (monomer)), ।
  • t lipemia (after spiking samples with lipids: 30 g/L equivalent in triglycerides),
  • bilirubinemia (after spiking samples with bilirubin: 0.3 g/L), ।
  • human albumin (after spiking samples with albumin up to 60 g/L).

It is recommended not to use samples that are clearly hemolyzed, lipemic or icteric and, if possible, to collect a new sample.

Exogenous Interferents: Following the recommendations of CLSI® document EP7-A2, the potential interferences with 15 commonly used drugs were studied. No interference was observed at the concentration tested.

DrugConcentration testedDrugConcentration tested
Acetylsalicylic acid3.62 mmol/LIbuprofen2425 µmol/L
Amoxicillin206 µmol/LMinocycline4.1 µmol/L
Azithromycin34 µmol/LPenicillin G240 000 U/L
Betamethasone8.31 µmol/LPenicillin Phenoxymethyl30 000 U/L
Ceftriaxone1460 µmol/LPrednisolone8.31 µmol/L
Cefuroxime axetil1416 µmol/LRoxithromycin15.3 µmol/L
Doxycycline hyclate16.1 µmol/LTetracyclines67.5 µmol/L
Erythromycin22.2 µmol/L

{4}------------------------------------------------

H. Clinical Testing

Sensitivity testing

202 retrospective samples from patients meeting a case definition of LD and confirmed positive for B. Burgdorferi infection were run on the VIDAS Lyme IgM assay and the predicate Lyme IgM assay.

For both the VIDAS Lyme IgM and the predicate test, equivocal results were considered as positive for the evaluation. The following results were obtained:

StageNVIDAS Lyme IgM% Sensitivity *Predicate LymeIgM% Sensitivity *Difference inproportions
Stage I(earlylocalized,single lesion)1 – 30 days11952.1 (62/119)95% CI [42.8 –61.3]%54.6 (65/119)95% CI [45.2 –63.8]%-2.5%95% CI [(-15.2)% –(10.2)%]
Stage II(earlydisseminated,multiplelesions)1 – 30 days6291.9 (57/62)95% CI [82.2 –97.3]%91.9 (57/62)95% CI [82.2 –97.3]%0.0%95% CI [(-9.6)% –(9.6)%]
Stage III(latedisseminated)2176.2 (16/21)95% CI [52.8 –91.8]%61.9 (13/21)95% CI [38.4 –81.9]%14.3%95% CI [(-13.3)% –(41.9)%]
All stages20266.8 (135/202)95% CI [59.9 –73.3]%66.8 (135/202)95% CI [59.9 –73.3]%0.0%95% CI [(-9.2)% –(9.2)%]
  • includes positive and equivocal results.

Method Comparison

A prospective study was performed on 975 fresh or frozen prospectively collected sera submitted for routine Lyme disease testing from an endemic area of the United States. Testing was performed in three laboratories.

At each laboratory, the samples were tested in parallel using a commercially available Lyme IgM EIA method (predicate) and the VIDAS Lyme IgM assay. Positive % Agreement (PPA) is calculated for the positives and equivocals together since the 2-tier testing does not make a distinction and calls for both of them to be tested by Western Blot. Combined results from the three sites are shown below:

N = 975Predicate Lyme IgM
VIDAS Lyme IgMPositiveEquivocalNegative
Positive711032
Equivocal151155
Negative5053678
Total13674765
Positive % Agreement95% CI51.0% (107/210)[44.0 - 57.9]%
Negative % Agreement95% CI88.6% (678/765)[86.2 - 90.8]%

5

{5}------------------------------------------------

Second-Tier Testing: In accordance with the CDC recommendations for use of a 2-tier Lyme disease testing scheme, the VIDAS Lyme IgM positive and equivocal results and the predicate Lyme IgM positive and equivocal results were confirmed using a commercially available Lyme IgM Western Blot method.

The percent agreement between VIDAS and predicate Lyme IgM positives (1ª tier PPA) and the percent agreement between VIDAS-predicate-Western Blot IgM positives and Predicate-Western Blot IgM positives (2nd tier PPA) are shown below.

1st Tier+ or ±IgM WesternPos.Neg.
Predicate IgM210104106
VIDAS IgM1949593*
VIDAS IgM and PredicateIgM1078423

*, Western Blot results were not available for 6 of the positive or equivocal samples by VIDAS Lyme IgM assay.

Agreement results:

18 Tier PPA = 51.0% (107/210), 95% Cl=[44.0% - 57.9%] 20d Tier PPA= 80.8% (84/104), 95% Cl=[72.2% - 87.2%]

Concordance with IgM Western Blot:

Predicate device: 49.5% (104/210) VIDAS Lyme IgM: 49.0% (95/194)

Analytical Specificity

100 sera from apparently healthy subjects from an endemic population (New York) and 100 sera from a non-endemic population (Texas) with no known history of Lyme disease were run on the VIDAS Lyme IgM assay and the predicate Lyme IgM assay. The following results were obtained:

VIDASPredicate
Positivity*NegativityPositivity*Negativity
Endemic12.0%88.0%19.0%81.0%
Non-Endemic14.0%86.0%3.0%97.0%
  • Includes positives and equivocals.

CDC Reference Panel

The following information is from a serum panel obtained from the CDC and tested using the VIDAS Lyme IgM kit. The results are presented as a means to convey further information on the performance of this assay with a masked, characterized serum panel. This does not imply an endorsement of the assay by the CDC.

б

{6}------------------------------------------------

VIDAS® Lyme IgM Assay

.

510(k) SUMMARY

TimefromonsetVIDAS Lyme IgMWestern blot IgM
PositiveorequivocalNegativeAgreementwith clinicalstatusPositiveNegativeAgreementwith clinicalstatus
Normals1480.0 %(4/5)05100.0 %(5/5)
< 1 month3260.0 %(3/5)3260.0 %(3/5)
1 - 2months5183.3 %(5/6)5183.3 %(5/6)
3- 12months12475.0 %(12/16)7943.8 %(7/16)
> 1 year3442.9 %(3/7)3442.9 %(3/7)
Total241569.2%(27/39)182159.0 %(23/39)

Cross-Reactivity

Cross-reactivity is based on the study of samples that are negative with the test being evaluated and positive for the potentially interfering disease. The results of the samples tested according to the disease are shown in the table below:

Infection or DiagnosisNVIDAS LymeIgMEquivocal orpositiveresults% Cross-reactivity
Anti Nuclear Antibodies6035.00
C Reactive Protein6146.55
Cytomegalovirus34617.64
Epstein Barr Virus65710.76
Helicobacter pylori143106.99
Hepatitis A Virus1532214.37
Herpes Simplex Virus981515.30
Human Immunodeficiency Virus20735.00
Human Anti-mouse Antibodies3126.45
Leptospirosis2162210.18
Measles38513.15
Mumps4624.34
Rheumatoid Factor6158.19
Rickettsiosis11265.35
Rubella19210.52
Syphilis270259.25
Systemic Lupus Erythematosus2827.14
Toxoplasmosis26519.23
Varicella Zoster Virus5846.89

7

{7}------------------------------------------------

510(k) SUMMARY

8

The effect of Babesiosis, Erhlichiosis and Rocky Mountain spotted fever pathologies on the VIDAS Lyme IgM performance is not known.

l. Conclusion

The results from the nonclinical and clinical studies submitted in this premarket notification are complete and demonstrate that the VIDAS® Lyme IgM is substantially equivalent to the predicate device identified in Item C of this summary.

{8}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the bird symbol.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

bioMerieux SA c/o Catherine FRITSCH Regulatory Affairs Director 5 rue des Aqueducs 69290 Craponne, France

June 4, 2013

Re: K122979

Trade/Device Name: VIDAS® Lyme IgM Regulation Number: 21 CFR 866.3830 Regulation Name: Treponema pallidum treponemal test reagents Regulatory Class: Class II Product Code: LSR Dated: May 6, 2013 Received: May 7, 2013

Dear Ms. FRITSCH:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA) You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{9}------------------------------------------------

Page 2 - Catherine FRITSCH

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Uwe-Scherf -S for

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{10}------------------------------------------------

Indications for Use

510(k) Number (if known): K122979

Device Name: VIDAS® Lyme IgM

Indications For Use:

The VIDAS Lyme IgM (LYM) assay is an automated qualitative enzyme immunoassay intended for use on the instruments of the VIDAS family in the presumptive detection of human IgM antibodies to Borrelia burgdorferi in human serum (plain or separation gel) or plasma (sodium heparin or lithium heparin). It should be used to test patients with a history and/or symptoms of infection with B. burgdorferi. All VIDAS Lyme IgM positive and equivocal specimens should be further tested with a Western Blot IgM assay to obtain supportive evidence of infection with B. burgdorferi.

Prescription Use X · (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

John Hobson-S
---------------
Division Sign-Off
---------------------
2013.06.03 13:35:34
-04'00'
Office of In Vitro Diagnostics and Radiological Health
--------------------------------------------------------
510(k)K122979
-----------------

Concurrence of CDRH; Office of In Vitro Diagnostics and Radiological Health (OIR)

Page 1 of 1

§ 866.3830

Treponema pallidum treponemal test reagents.(a)
Identification. Treponema pallidum treponemal test reagents are devices that consist of the antigens, antisera and all control reagents (standardized reagents with which test results are compared) which are derived from treponemal sources and that are used in the fluorescent treponemal antibody absorption test (FTA-ABS), theTreponema pallidum immobilization test (T.P.I.), and other treponemal tests used to identify antibodies toTreponema pallidum directly from infecting treponemal organisms in serum. The identification aids in the diagnosis of syphilis caused by bacteria belonging to the genusTreponema and provides epidemiological information on syphilis.(b)
Classification. Class II (performance standards).