K Number
K163560

Validate with FDA (Live)

Date Cleared
2017-07-12

(205 days)

Product Code
Regulation Number
870.1200
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Reprocessed Acunav Diagnostic Ultrasound Catheter is intracardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult patients.

The reprocessed device is not indicated for use with pediatric patients.

Device Description

The Reprocessed AcuNav Diagnostic Ultrasound Catheters distal end has an ultrasound transducer providing 2-D imaging. A steering mechanism controls the image plane orientation through four-way articulation of the tip.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a Reprocessed AcuNav Diagnostic Ultrasound Catheter. This document focuses on demonstrating the substantial equivalence of the reprocessed device to legally marketed predicate devices, rather than establishing acceptance criteria or proving performance of a novel AI/ML-driven device.

Therefore, the information required to answer your questions about acceptance criteria, device performance, sample sizes for training/test sets, expert adjudication, MRMC studies, standalone performance, and ground truth establishment is not present in this document.

This document describes a medical device reprocessing process and the testing performed to ensure the reprocessed device's safety and effectiveness are equivalent to new devices. The "Functional and Safety Testing" section highlights the types of tests conducted, but it doesn't provide specific quantitative acceptance criteria or detailed results of those tests.

In summary, this document does not contain the information requested in your prompt regarding acceptance criteria and study results for an AI/ML-driven device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with three heads in profile, facing right. The eagle is positioned within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 12, 2017

Innovative Health, LLC. Rafal Chudzik VP. R&D and Operations 1435 North Hayden Road Suite 100 Scottsdale, Arizona 85257

Re: K163560

Trade/Device Name: Reprocessed AcuNav Diagnostic Ultrasound Catheter (see Enclosed Model List) Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: OWO Dated: July 7, 2017 Received: July 10, 2017

Dear Rafal Chudzik:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

M.A. Wilhelmsson
for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Table 1: K163560 List of Models in Scope

ltem NumberDescriptionSheathUsableLength(cm)FrenchSizeSystem Compatibility
10135936AcuNav Diagnostic UltrasoundCatheter908FAcuson/Siemens
10135910AcuNav Diagnostic UltrasoundCatheter908FGE Vivid i & Vivid qUltrasound System

The item numbers included in the scope of this submission are as follows:

Table 1: Devices in Scope

{3}------------------------------------------------

Indications for Use

510(k) Number (if known)

K163560

Device Name

Reprocessed AcuNav Diagnostic Ultrasound Catheter

Indications for Use (Describe)

The Reprocessed Acunav Diagnostic Ultrasound Catheter is intracardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult patients.

The reprocessed device is not indicated for use with pediatric patients.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

SECTION 5: 510(k) SUMMARY

As required by 21 CFR 807.92(c)

Submitter's Name and Address:

Innovative Health, LLC. 1435 N. Hayden Road, Suite 100 Scottsdale, AZ 85257

Contact Name and Information:

Rafal Chudzik Innovative Health, LLC. VP. R&D and Operations (480) 525-6006 (office) (844) 965-9359 (fax) rchudzik@innovative-health.com

Date prepared:

December 16, 2016

Device Information:

Trade/Proprietary Name:Reprocessed AcuNav Diagnostic Ultrasound Catheters
Common Name:Diagnostic Ultrasound Catheters
Classification Name:Reprocessed Intravascular Ultrasound Catheter
Classification Number:Class II, 21 CFR 870.1200
Product Code:OWQ

Predicate Device:

510(k) Number510(k) TitleManufacturer
K153090Reprocessed AcuNav Diagnostic UltrasoundCathetersInnovative Health, LLC.
K071234AcuNav Diagnostic Ultrasound Catheter 8Fand 10FSiemens Medical Solutions

Device Description:

The Reprocessed AcuNav Diagnostic Ultrasound Catheters distal end has an ultrasound transducer providing 2-D imaging. A steering mechanism controls the image plane orientation through four-way articulation of the tip.

ItemNumberDescriptionSheathUsableLength(cm)FrenchSizeSystemCompatibility
10135936AcuNav DiagnosticUltrasound Catheter908FAcuson/Siemens
10135910AcuNav DiagnosticUltrasound Catheter908FGE Vivid i & Vivid qUltrasound System

{5}------------------------------------------------

Indications for Use:

The Reprocessed AcuNav Diagnostic Ultrasound Catheter is intended for intracardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the hearts of adult patients.

The reprocessed device is not indicated for use with pediatric patients.

Technological Characteristics:

The purpose, design, materials, function, and intended use of the Reprocessed Diagnostic Ultrasound Catheters are identical to the predicate devices. There are no changes to the claims. clinical applications, performance specifications, or method of operation. In addition, Innovative Health's reprocessing of the Diagnostic Ultrasound Catheter includes removal of visible soil and decontamination. Each device is inspected and function tested prior to packaging and labeling.

Functional and Safety Testing:

Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of Reprocessed Diagnostic Ultrasound Catheters. This included the following:

  • Biocompatibility ●
  • Cleaning Validation ●
  • Sterilization Validation
  • Functional testing
    • . Visual Inspection
    • . Dimensional Verification
    • . Ultrasound Transducer Testing
    • . Simulated Use
    • Mechanical Characteristics
  • . Electrical Safety Testing
    • . Dielectric and Current Leakage
  • . Packaging Validation

The Reprocessed Diagnostic Ultrasound Catheters are reprocessed no more than three (3) times. Each device is marked and tracked. After the device has reached the maximum number of reprocessing cycles, the device is rejected from further reprocessing. Reprocessing is performed only by Innovative Health. Innovative Health restricts its reprocessing to exclude devices previously reprocessed by other reprocessors.

Conclusion:

Innovative Health concludes that the Reprocessed Diagnostic Ultrasound Catheter is as safe and effective as the predicate devices described herein.

§ 870.1200 Diagnostic intravascular catheter.

(a)
Identification. An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.(b)
Classification. Class II (performance standards).