(154 days)
The powder free nitrile patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
The proposed device, Powder Free Nitrile Patient Examination Gloves is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
The proposed device is Powder Free Nitrile Patient Examination Gloves, and includes variations of different size and color. The colors of proposed device is Blue.
The proposed device is not provided as sterilized The proposed device is made of Nitrile.
The provided text describes the acceptance criteria and the study results for the Powder Free Nitrile Patient Examination Gloves, Blue. Here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance:
The acceptance criteria are established by various ASTM and ISO standards. The reported device performance generally indicates compliance with these standards.
| Acceptance Criterion (Standard) | Reported Device Performance |
|---|---|
| Physical Properties Before Aging (ASTM D6319) | |
| - Tensile Strength | 15 MPa |
| - Ultimate Elongation | 500 % min |
| Physical Properties After Aging (ASTM D6319) | |
| - Tensile Strength | 14 MPa |
| - Ultimate Elongation | 400 % min |
| Freedom from Holes (ASTM D5151) | Free from holes when tested in accordance with ASTM D5151 |
| Powder Content (ASTM D6124) | Max. 0.32 mg per glove |
| Biocompatibility (ISO 10993-10) | |
| - Irritation | Under the conditions of the study, not an irritant |
| - Sensitization | Under conditions of the study, not a sensitizer |
| Device Dimensions (ASTM D6319) | |
| - Length, mm (XS, S, M, L, XL) | min (specific values for XS-XL not provided for proposed, but for predicate) |
| - Width, mm (XS, S, M, L, XL) | 70, 80, 95, 110, 120 (± 10) |
| - Thickness, mm: Finger | 0.10 (± 0.03) |
| - Thickness, mm: Palm | 0.08 (± 0.03) |
| - Thickness, mm: Cuff | 0.06 (± 0.03) |
| Sampling Procedures for Inspection by Attributes (ISO 2859-1:1999) | Complies with standard |
2. Sample Size Used for the Test Set and the Data Provenance:
The document broadly states that "Bench tests were conducted to verify that the proposed device met all design specifications as compared to the predicate device."
- Sample Size: The exact sample size for each test is not specified in the provided document. The ISO 2859-1:1999 standard is cited for "Sampling Procedures for Inspection by Attributes - Part I: Sampling Plans Indexed by Acceptable Quality Level (AQL) for Lot-by-Lot Inspection," implying that sampling was done according to this standard, but the specific AQL levels or sample sizes used are not detailed.
- Data Provenance: The document implies the data is prospective as it describes tests conducted to verify the proposed device. The manufacturer is "HEBEI TITANS MEDICAL SUPPLY TECHNOLOGY GROUP CO., LTD." in China, suggesting the tests were conducted in or overseen by this Chinese entity.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:
This information is not applicable to this type of device (patient examination gloves). The "ground truth" for glove performance is established through objective physical, mechanical, and biological testing against recognized standards, not through expert human interpretation of results in the way an AI medical imaging device would require.
4. Adjudication Method for the Test Set:
This is not applicable to this type of device. Adjudication methods like 2+1 or 3+1 are typically used for establishing ground truth in expert-based assessments (e.g., medical image interpretation), which is not relevant here. The tests conducted are objective measurements against established standards.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance:
This is not applicable. An MRMC study is relevant for AI-assisted diagnostic devices where human readers interpret medical images or data. The device in question is a physical medical glove.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:
This is not applicable. "Standalone performance" refers to the performance of an algorithm or AI without human input in a diagnostic or interpretive context. The device is a physical glove, not an algorithm.
7. The Type of Ground Truth Used:
The ground truth used for proving the device meets acceptance criteria is based on objective measurement against established industry standards, specifically:
- ASTM Standards: ASTM D6319-10 (physical properties), ASTM D5151-06 (freedom from holes), ASTM D6124-06 (residual powder).
- ISO Standards: ISO 2859-1:1999 (sampling procedures), ISO 10993-10:2010 (biocompatibility: irritation and skin sensitization).
- Performance Metrics: Quantifiable measurements such as tensile strength (MPa), ultimate elongation (%), powder content (mg/glove), dimensions (mm), and qualitative assessments (e.g., "not an irritant," "free from holes").
8. The Sample Size for the Training Set:
This is not applicable as there is no "training set" in the context of physical medical devices like gloves. Training sets are used for machine learning models.
9. How the Ground Truth for the Training Set Was Established:
This is not applicable as there is no training set.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked one behind the other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Hebei Titans Medical Supply Technology Group.,ltd. % Ray Wang Official Correspondent Beijing Believe Tech. Service Co., Ltd. 5-1206, Build 332, Dafangju, No.25 Banbidian Rd., Li Yuan, Tongzhou District Beijing, BeiJing 101121 CN
Re: K163553
Trade/Device Name: Powder Free Nitrile Patient Examination Gloves, Blue Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: April 27, 2017 Received: May 1, 2017
Dear Ray Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K163553
Device Name
Powder Free Nitrile Patient Examination Gloves, Blue
Indications for Use (Describe)
The powder free nitrile patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
Type of Use (Select one or both, as applicable)
] Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
EF
PSC Publishing Services (301) 443-6740
{3}------------------------------------------------
510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) Number: K163553
-
- Date of Preparation: 2017/05/16
-
- Sponsor
HEBEI TITANS MEDICAL SUPPLY TECHNOLOGY GROUP CO., LTD. EAST OF JINGQIANG COMMUNITY, DAQING STREET, NANGONG, XINGTAI, HEBEI PROVINCE, CHINA
Contact Person: Mr. ShaoZhang Nan Tel: +86-0319-7295820 Fax: +86-0319-7295801 Email: nanshaozhang@163.com
-
Submission Correspondent 3. Mr. Ray Wang
Beijing Believe Tech. Service Co., Ltd. Email: Ray.Wang@believe-med.com -
- Proposed Device Identification
Trade Name: Powder Free Nitrile Patient Examination Gloves, Blue Device Name: NITRILE Patient Examination Gloves (Powder Free) Common Name: Patient Examination Gloves
- Proposed Device Identification
Classification: I Product Code: LZA Regulation Number: 21 CFR 880.6250 Review Panel: General Hospital
Indication for Use Statement:
The powder free nitrile patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
{4}------------------------------------------------
-
Predicate Device Identification 5.
510(k) Number: K131440 Product Name: Powder free Patient Examination Gloves, Blue Color Manufacturer: Hebei HongSen Plastics Technology Co., Ltd. -
- Device Description
The proposed device, Powder Free Nitrile Patient Examination Gloves is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
- Device Description
The proposed device is Powder Free Nitrile Patient Examination Gloves, and includes variations of different size and color. The colors of proposed device is Blue.
The proposed device is not provided as sterilized The proposed device is made of Nitrile.
-
- Non-Clinical Test Conclusion
Bench tests were conducted to verify that the proposed device met all design specifications as compared to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- Non-Clinical Test Conclusion
ASTM D6319-10 (Reapproved 2015), Standard Specification for Nitrile Examination Gloves for Medical Application.
ASTM D 5151-06 (Reapproved 2011), Standard Test Method for Detection of Holes in Medical Gloves.
ASTM D6124-06 (Reaffirmation 2011), Standard Test Method for Residual Powder on Medical Gloves.
ISO 2859-1:1999, "Sampling Procedures for Inspection by Attributes - Part I: Sampling Plans Indexed by Acceptable Quality Level (AQL) for Lot-by-Lot Inspection.
ISO 10993-10: 2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.
{5}------------------------------------------------
8. Substantially Equivalent Comparison Conclusion
| Table III-1 General Comparison | |||
|---|---|---|---|
| ITEM | Proposed DevicePowder Free Nitrile Patient ExaminationGloves, Blue | Predicate Device (K131440) Powder freePatient Examination Gloves,Blue Color | Remark |
| Product Code | LZA | LZA | Same |
| Regulation No. | 21 CFR 880.6250 | 21 CFR 880.6250 | Same |
| Class | I | I | Same |
| Indication forUse | The powder free nitrile patient examinationglove is a disposable device intended formedical purposes that is worn on theexaminer's hands to prevent contaminationbetween patient and examiner. | The Titan powder free nitrile patientexamination glove, blue color, is adisposable device intended for medicalpurposes that is worn on the examiner'shands or finger to prevent contaminationbetween patient and examiner. It has bluecolor and is sold as non-sterile. | Same |
| Powdered orPowered free | Powdered free | Powdered free | Same |
| Design Feature | ambidextrous, smooth | ambidextrous, smooth | Same |
| LabelingInformation | Single-use indication, powder free, devicename, glove size and quantity, NitrileExamination Gloves, Non-Sterile | Single-use indication, powder free, devicename, glove size and quantity, NitrileExamination Gloves, Non-Sterile | Same |
Table III-1 General Comparison
Table III-2 Device Dimensions Comparison
| Proposed DevicePowder Free Nitrile Patient ExaminationGloves, Blue | Designation | Size | Tolerance | |||||
|---|---|---|---|---|---|---|---|---|
| Length, mm | XS | S | M | L | XL | min | ||
| Width, mm | 70 | 80 | 95 | 110 | 120 | $\pm$ 10 | ||
| Thickness, mm: | ||||||||
| Finger | 0.10 | $\pm$ 0.03 | ||||||
| Palm | 0.08 | $\pm$ 0.03 | ||||||
| Cuff | 0.06 | $\pm$ 0.03 | ||||||
| Predicate Device (K131440)Powder free Patient ExaminationGloves, Blue ColorRemark | Designation | XS | S | M | L | XL | XXL | Tolerance |
| Different 1 |
{6}------------------------------------------------
Different 1 analysis:
The proposed device has different size specification to the predicate device, and all specific size specifications for the predicate are not available for comparison, but the predicate and proposed device both meet the specifications of ASTM D 6319.
| ITEM | Proposed Device | Predicate Device (K131440) | Remark | |
|---|---|---|---|---|
| Colorant | Powder Free Nitrile PatientExamination Gloves, Blue | Powder free Patient ExaminationGloves, Blue Color | Same | |
| PhysicalPropertiesBeforeAging | TensileStrength | 15 MPa, | Comply with ASTM D6319 | Same |
| UltimateElongation | 500 % min | Comply with ASTM D6319 | Same | |
| PhysicalPropertiesAfterAging | TensileStrength | 14 MPa | Comply with ASTM D6319 | |
| UltimateElongation | 400 % min | Comply with ASTM D6319 | ||
| Comply with ASTM D6319 | Comply with ASTM D6319 | Comply with ASTM D6319 | Same | |
| Freedom from Holes | Be free from holes when tested inaccordance with ASTM D5151 | Be free from holes when testedin accordance with ASTMD5151 | Same | |
| Powder Content | Max. 0.32 mg per glove | Meet the requirements of ASTM6319 | Same |
Table III-3 Performance Comparison
Table III-4Safety Comparison
| ITEM | Proposed Device PowderFree Nitrile PatientExamination Gloves, Blue | Predicate Device (K131440)Powder free Patient ExaminationGloves, Blue Color | Remark |
|---|---|---|---|
| Material | Nitrile | Nitrile | Same |
| Biocompatibility | IrritationUnder the conditions of the study,not an irritant | Comply with ISO 10993-10 | Same |
| SensitizationUnder conditions of the study, not asensitizer. | |||
| Label and Labeling | Meets recommendations as outlinedin the Glove Guidance Manual. | Meets recommendations as outlinedin the Glove Guidance Manual. | Same |
Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.