K Number
K163402

Validate with FDA (Live)

Device Name
NLZ Motor System
Manufacturer
Date Cleared
2017-08-31

(269 days)

Product Code
Regulation Number
872.4200
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NLZ Motor System is intended for use by dental professionals in the performance of dental restoration, prophylaxis and endodontic procedures.

The NLZ Endo is intended for use by dental professionals in the performance of dental endodontic procedures.

Device Description

NLZ Motor Systems are available in two models;

NLZ Motor Set, which consists of the Main Unit, the Control Unit and the Motor.

NLZ E Motor Set, which consists of the Main Unit, the Control Unit, the Motor, and the Contra Angle Handpiece: NLZ Endo. The NLZ Endo is available only for NLZ E Motor Set and is not avialable for the NLZ Motor Set. The NLZ Endo is not equipped with any cooling system because it is a speed-reducing/low-speed handpiece.

The Control Unit drives an electric micromotor, and turns on off or regulates the speed of an electric motor by its foot pedal of the dental unit. This product can be connected to dental units currently in use to add on a brushless electric micromotor with an LED light.

AI/ML Overview

The provided text describes Nakanishi Inc.'s NLZ Motor System and its substantial equivalence to predicate devices, focusing on various performance aspects. However, the document does not present acceptance criteria in a table format with specific quantitative targets and corresponding device performance measurements. Instead, it discusses compliance with general standards and the satisfactory outcome of various tests.

Here's an analysis based on the available information, addressing as many of your points as possible:

1. Table of acceptance criteria and the reported device performance:

The document does not explicitly provide a table of acceptance criteria with numerical performance targets and reported values for the NLZ Motor System. Instead, it states that the device was compliant with various ISO and IEC standards. The "performance" is generally described as "met all requirements" or "do not raise concern."

However, specific comparisons are made with predicate devices, which can infer performance expectations. For instance:

FeatureAcceptance Criteria (Inferred from standards/predicates)Reported Device Performance (NLZ Motor System)
BiocompatibilityConformity to ANSI/AAMI/ISO 10993-5:2009, ISO 10993-10:2010No biocompatibility concern raised
Mechanical PerformanceConformity to ISO 14457:2012All samples compliant with ISO 14457:2012
Electrical SafetyConformity to AAMI / ANSI/ES60601-1:2005/(R)2012 and A1:2012,, C1:2009/(R)2012 and A2:2010/(R)2012Test results indicate no concern
Electromagnetic CompatibilityConformity to IEC 60601-1-2:2014Test results indicate no concern
Overheating Risk ReductionProper Contra-Check judge and motor speed control for Contra-Restriction and Contra-SAFE. Reduced overheating/burn events.Verification test confirmed proper function and no new concern. Analyzed temperature increase for returned handpieces to confirm reduction of events.
Reprocessing ValidationConformity to AAMI TIR 12:2010, AAMI TIR 30:2011, and FDA Reprocessing guidanceMet all requirements of the standards
Range of Rotation Speed100 - 40,000 rpm (Similar to predicate)100 - 40,000 rpm
Rotating DirectionForward and Reverse (Similar to predicate)Forward and Reverse
Light Intensity>10,000 lx (for LED light)>10,000 lx (for LED light)
NLZ Endo EccentricityNot explicitly stated for predicate in table, but performance compliant with ISO 14457.0.018 mm
NLZ Endo Bur ExtractionNot explicitly stated for predicate in table, but performance compliant with ISO 14457.>45N
NLZ Endo Max Bur Length25 mm (Predicate)46 mm

2. Sample size used for the test set and the data provenance:

  • Sample size: Not specified. The document mentions "All samples were compliant with ISO 14457:2012" and "This testing is verification test for software by putting excessive load to the Motor." However, the exact number of devices tested for each criterion is not provided.
  • Data provenance: Bench testing was performed internally by Nakanishi Inc. ("in-house standard," "bench testing was performed in order to verify conformity," "reprocessing validation was performed"). The document does not indicate the country of origin of the data beyond the company's location (Japan) and general compliance with international standards (ISO, IEC, AAMI, ANSI). The studies were retrospective as they describe testing conducted before submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not provided in the document. The testing described is primarily technical bench testing and compliance with standards, not studies involving human experts establishing ground truth for diagnostic or clinical accuracy.

4. Adjudication method for the test set:

This information is not provided as the studies are technical performance and compliance tests rather than clinical evaluations requiring adjudication of subjective outcomes.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

No MRMC comparative effectiveness study was done or is mentioned. This device is a dental motor system, not an AI-assisted diagnostic tool.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

Not applicable. This device does not involve an algorithm for diagnostic or clinical interpretation that would operate in a standalone manner. The "overheating risk reduction functions" are automated features within the device, whose performance was verified through bench testing.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

For the performance testing, the "ground truth" was defined by:

  • International standards: ISO 14457, ANSI/AAMI/ISO 10993, IEC 60601-1-2, AAMI / ANSI/ES60601-1, AAMI TIR 12, AAMI TIR 30, and FDA Reprocessing guidance.
  • In-house standards: For the overheating risk reduction features.
  • Predicate devices: The performance characteristics of the NLZ Motor System were compared to those of the predicate devices for establishing substantial equivalence.

8. The sample size for the training set:

Not applicable. This device does not utilize a "training set" in the context of machine learning or AI.

9. How the ground truth for the training set was established:

Not applicable. As above, there is no mention of a training set.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized symbol that resembles three human profiles facing to the right, stacked on top of each other, with flowing lines beneath them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 31, 2017

Nakanishi Inc. % Izumi Maruo Senior Consultant MIC International Corp. 4-1-17 Hongo Bunkyo-ku, Tokyo 1130033 JAPAN

Re: K163402

Trade/Device Name: NLZ Motor System Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: Class I Product Code: EBW Dated: July 21, 2017 Received: August 1, 2017

Dear Izumi Maruo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{1}------------------------------------------------

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Andrew I. Steen -S

for Lori A. Wiggins, MPT, CLT Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K163402

Device Name NLZ Motor System

Indications for Use (Describe)

The NLZ Motor System is intended for use by dental professionals in the performance of dental restoration, prophylaxis and endodontic procedures.

The NLZ Endo is intended for use by dental professionals in the performance of dental endodontic procedures.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image features the logo of NSK, a company specializing in bearings and motion control technology. The logo consists of the letters 'NSK' in a bold, sans-serif typeface. The letters are white and are set against a solid blue background, creating a strong contrast that makes the logo easily recognizable.

K163402

510(k) Summary

  • a. Owner/Company name, address NAKANISHI INC. 700 Shimohinata, Kanuma-shi, Tochigi, 322-8666, Japan

b. Contact

Kimihiko Satoh General Manager, Regulatory Affairs Department NAKANISHI INC. 700 Shimohinata, Kanuma-shi, Tochigi, 322-8666, Japan

Phone:+81-289-64-4325
Fax:+81-289-62-6665
Email:k-satoh@nsk-nakanishi.co.jp

c. Application Correspondent

Izumi Maruo Senior Consultant MIC International 4-1-17 Hongo, Bunkyo-ku Tokyo, 113-0033, Japan

Phone:+81-3-3818-8577
Fax:+81-3-3818-8573
Email:maruo@mici.co.jp

d. Date prepared

July 21, 2017

e. Name of device

Trade Name: Classification Name: Regulation Name: Regulatory Class: Product Code:

NLZ Motor System Controller, Foot, Handpiece and Cord Dental handpiece and accessories Class I EBW

Image /page/3/Picture/16 description: The image shows a hand holding a blue object that resembles a pen or stylus. The hand is white and appears to be wearing a glove. The blue object is angled across the hand, and the words "Powerful Partners" are written in a small font below the hand. The logo is simple and modern, and the text suggests a business or professional partnership.

{4}------------------------------------------------

f. Predicate devices and reference device

Predicate device
510(k):K090937
Trade name:E-STATIS 40 Electric Motor System
Company name:SciCAN
Product code:EBW

The E-STATIS 40 Electric Motor System (K090937) does not include a handpiece. Therefore, the predicate device for the contra-angle handpiece, NLZ Endo is shown below.

Predicate device for the NLZ Endo
510(k):K150750
Trade name:Contra-Angle Handpiece :Endo 6:1
Company name:Sirona Dental Systems GmbH
Product code:EGS

This submission uses following reference device for the reciprocating mode of the NLZ Motor System. The E-STATIS 40 Electric Motor System (K090937) is not equipped with that mode.

Reference Device
510(k):K103653
Trade name:e3TM Torque Control Motor
Company name:DENTSPLY International Inc.
Product code:EBW
510(k):K103027
Trade name:ELECTROtorque TLC 4893 with INTRAmatic KL 702
Company name:Kaltenbach & Voigt GmbH
Product code:EBW, EKX

g. Description of the device

NLZ Motor Systems are available in two models;

NLZ Motor Set, which consists of the Main Unit, the Control Unit and the Motor.

NLZ E Motor Set, which consists of the Main Unit, the Control Unit, the Motor, and the Contra Angle Handpiece: NLZ Endo. The NLZ Endo is available only for NLZ E Motor Set and is not avialable for the NLZ Motor Set. The NLZ Endo is not equipped with any cooling system because it is a speed-reducing/low-speed handpiece.

The Control Unit drives an electric micromotor, and turns on off or regulates the speed of an electric motor by its foot pedal of the dental unit. This product can be connected to dental

{5}------------------------------------------------

units currently in use to add on a brushless electric micromotor with an LED light.

h. Indications for Use

The NLZ Motor System is intended for use by dental professionals in the performance of dental restoration, prophylaxis and endodontic procedures.

The NLZ Endo is intended for use by dental professionals in the performance of dental endodontic procedures.

Statement of substantial equivalence i.

The Indications for Use of the NLZ Motor System is identical to the E-STATIS 40 Electric Motor System (K090937).

The NLZ Motor System and the E-STATIS 40 Electric Motor System (K090937) have the similar functions and characteristics regarding the range of rotating speed, rotation direction, and auto reverse/auto stop.

The NLZ Motor System is equipped with similar reciprocating mode to the reference device.

Following table is comparison between the subject, the predicate and the reference devices.

{6}------------------------------------------------

Subject DevicePredicate DeviceReference DeviceReference Device
NLZ Motor SystemE-STATIS 40 ElectricMotor System (K090937)ELECTROtorque TLC4893 (K103027)e3TM Torque Control Motor(K103653)
IndicationsforUseThe NLZ Motor System isThe E-STATIS type 40The ELECTROtorgue TLCThe e3 Torque Control Motor
intended for use by dentaldental system is intended4893 is intended to convertis a medical device designed
professionalsinthefor use by dentalpneumatic output from afor use by dentists for use
performanceofdentalprofessionals in thedental treatment center towith dental root canal
restoration,prophylaxisperformance of dentalelectrical energy to driveinstruments in continuous
andendodonticrestoration, prophylaxis andthe INTRAmatic KL 702rotation with torque control or
procedures.endodontic procedures.motor for operation ofin reciprocating movement.
electrically driven
[Indications for Use fordental handpieces. These
NLZ Endo]devices are designed for
The NLZ Endo is intendeduse by a trained
for useby dentalprofessional in the field of
professionalsin thegeneral dentistry.
performanceof
endodontic procedures.
DriveElectronic-micromotorElectronic-micromotorElectronic-micromotorElectronic-micromotor
ComponentsPower supply, Motorcontroller, Electricmicromotor, andPower supply, Motorcontroller, and ElectricPower supply, Motorcontroller, and ElectricPower supply, Motorcontroller, Foot pedal, Electricmicromotor and Contra-angle
Contra-angle handpiecemicromotormicromotorhandpiece (K972436)
(NLZ Endo, NLZ E Motor
Set ONLY)
LightLEDLight bulbLEDNo light
- Light intensity>10,000[lx]UnknownUnknownNA

Table 1. Comparison table for outline of the motor system

{7}------------------------------------------------

Range of rotation speed (rpm)100 - 40,000 rpm100 - 40,000 rpm100 - 40,000 rpm250 - 1,000 rpm
Rotating directionForward and ReverseForward and ReverseForward and ReverseForward and Reverse
Standard operation modeYesYesYesNo
-Overheating risk reduction featuresYes*NoYes*NA
Reciprocating modeYes (NLZ E Motor Set ONLY)NoNoYes
-Intended fileWaveOneNANAWave One
-6:1 gear ratio handpiece designed for rotary and reciprocating modes is provided with the MotorYesNANAYes
- Preset parameters for the reciprocating modeYes (User cannot alter the parameters)NANAYes (User cannot alter the parameters)
Rotary Endo ModeYes (NLZ E Motor Set ONLY)YesYesYes
- Auto reverse/Auto Stop functionYesYesYesYes
MotoroverheatingpreventionfunctionYesUnknownYesUnknown
Maximum air/water pressure
-Coolant airsupply0.40 MPa0.40 MPa0.5 MPaNo coolant circuit
-Spray watersupply0.25 MPa0.25 MPa0.25 MPaNo spray circuit
-Spray airsupply0.25 MPa0.20 MPa0.20 MPaNo spray circuit
SterilizationSterilized by user(steam sterilization)Sterilized by user(steam sterilization)Sterilized by user(steam sterilization)The contra-angle handpieceonly is sterilized by user
AvailableHandpiece typeE-type (ISO3964)E-type (ISO3964)ISO3964 conformedattachmentE-type (ISO3964)

{8}------------------------------------------------

  • For detail comparison regarding overheating risk reduction features, please see Table 2.

{9}------------------------------------------------

Based on Table 1, the NLZ E Motor Set provides "Reciprocating Endo Mode" and "Rotary Endo Mode" in addition to standard operation mode (in the NLZ Motor System, named "General Application Mode"). The Reciprocating Endo Mode of the NLZ E Motor Set is designed for use with WaveOne file designed for unequal counterclockwise (150°) and clockwise (30°) rotation. In order to perform reciprocating endo treatment safely, we provide the NLZ Endo handpiece which was verified for use with the WaveOne and presetting parameters of this treatment for the NLZ Endo with the WaveOne on the NLZ E Motor Set. User cannot alter those presetting parameters. The reciprocating mode of the reference device, e3 M Torque Control Motor (K103653) has identical features to the Reciprocating Endo Mode, which means;

Both modes are designed for use with WaveOne file 1.

  1. Both modes provide automatic unequal counterclockwise (150°) and clockwise (30°) rotation

  2. Both modes provide a 6:1 handpiece which was verified for use with the WaveOne and presetting parameters for using the providing handpiece combined with the WaveOne file. 4. User cannot alter the presetting parameters

The Rotary Endo Mode of the NLZ E Motor Set provides the continuous one direction rotation for use with low-speed handpieces and three modes when the set torque reached. The values of rotation speed and torque can be adjusted. The features for "Endodontic Mode" of the E-STATIS 40 Electric Motor System (K090937) are identical to those for this mode.

Regarding standard operation mode, the provided rotation direction and range of rotation speed of the subject device are identical to those of the E-STATIS 40 Electric Motor System (K090937). Although the NLZ Motor System is equipped with the overheating risk reduction functions in order to reduce patient burns caused by overheated handpiece, the predicate device E-STATIS 40 Electric Motor System (K090937) is not equipped with such function. Therefore, such functions of the subject device are compared with a reference device, ELECTROtorque TLC 4893 (K103027).

{10}------------------------------------------------

Table 2. Comparison of overheating risk reduction functions
OverheatingriskreductionfunctionNLZ Motor System (K163402/S001)ELECTROtorque4893 (K103027)TLC
Detection ofdefectivehandpiecesContra-SAFESAFEdrive
- DetectionofoverheatingIndirectly by measuring load for thehandpiece and the duration of timeIndirectly by continuousmonitoring of the idlingproperties of thehandpiece during its use.
- FeaturesThis function automatically stops motorrotation or controls the speed when higherand continued load of motor current aredetected.If the protective functionis triggered, theSAFEdrive initiallyreduces the motor speedand then stops the motoraltogether if the excessiveload persists
Judge anddisplay ofthehandpiecestatusContra-CheckThe device does not havethis function
- Features1) A manual feature for indirect detectionof excessive friction by measuring currentconsumption of the Motor2) Providing the status of handpieceregarding potential overheating as follows;● OK: The check result is acceptable● OIL: Needs a maintenance such aslubricating with oil● NG: There is a seriousmalfunction● NG blinking: Damaged handpieceNA
Motor speedcontrolfunction afterContra-CheckContra-RestrictionNA
-FeaturesThis function restricts motor rotation whenhandpiece is used for judge other than OKthrough estimating temperature increase bythe Control Unit. The motor is sloweddown and stopped after speed control.NA

Table 2. Comparison of overheating risk reduction functions

{11}------------------------------------------------

● If NG blinking is displayed, the
motor is stopped and cannot be
activated.
If NG is displayed, the motor is
stopped approximately 10 seconds
after continuous use.
If Oil is displayed, the motor is
stopped approximately 20 seconds
after continuous use.

In addition, the subject device and the ELECTROtorque TLC 4893 (K103027) are equipped with the Motor overheat prevention function. The motor overheat prevention function indirectly detects overheat of the Motor itself in order to protect the Motor from failure of the coils. The purpose of design for this function is different from handpiece overheating risk reduction features described above.

This function of the subject device estimates temperature increase of the Motor by the motor current and rotation speed. The ELECTROtorque TLC 4893 (K103027) detects overload.

Regarding the handpiece, the comparison table is shown below;

NLZ EndoContra-AngleHandpiece :Endo 6:1 (K150750 )
Design featuresContra-Angle, Push-ButtonChuck, Motor coupling, Withoutlight and cooling systemContra-Angle, Push-ButtonChuck, Motor coupling, Withoutlight and cooling system
Gear ratio6:1 Reduction6:1 Reduction
Maximum motor/drivespeed6,000 rpm40,000 rpm
MaterialsStainless SteelStainless Steel
Motor couplingCompliant with ISO 3964Compliant with ISO 3964
Source of powerElectronic-micromotorNLZ E Motor Set OnlyElectronic-micromotor
DimensionHead Diameter: 8.6mmHead Height: 11.3mmLength: 81.9mmHead Diameter: 8.7mmHead Height: 11.4mmLength: 92.3mm
Weight66 g70 g
PerformanceCompliant with ISO 14457Compliant with ISO 14457
-Eccentricity0.018 mmUnknown
-Force for bur extraction>45NUnknown
SterilizationSterilized by user(steam sterilization)Sterilized by user(steam sterilization)
LubricantPANA SPRAY Plus(510(k) number is K131014)Sirona T1 spray(510(k)number is K113674)

Table 3. Comparison table for the Contra-angle handpiece

{12}------------------------------------------------

BurContra-angle bursISO1797-1 Type1Contra-angle bursISO1797-1 Type1
-Maximum bur length46 mm25 mm

Both the NLZ Endo and the Contra-Angle Handpiece :Endo 6:1 (K150750) are used for dental endodontic procedures. The main design features and gear ratio of the NLZ Endo are similar to those of the Contra-Angle Handpiece : Endo 6:1 (K150750). Although there are some differences between the NLZ Endo and the Contra-Angle Handpiece :Endo 6:1 (K150750), the performance characteristics of both devices are compliant with ISO 14457.

For the NLZ Motor System, performance testing including biocompatibility testing, Electric safety and EMC testing, bench testing was performed in order to demonstrate substantial equivalence to the predicates. Also, reprocessing validation was performed because both the Motor and the NLZ Endo are required cleaning and sterilization after use by user. Cleaning, sterilization, and drying time recommended in the instruction were validated using AAMI TIR 12:2010, AAMI TIR 30:2011, and FDA Reprocessing guidance. The NLZ Motor System met all requirements of the standards. Because the NLZ Motor System is equipped with the overheating risk reduction functions with which the E-STATIS 40 Electric Motor System (K090937) is not equipped, those functions are compared with the functions of the ELECTROtorque TLC 4893 (K103027). The Contra-SAFE function of the subject device has similar features to the SAFEdrive function of the ELECTROtorque TLC 4893 (K103027). The ELECTROtorque TLC 4893 (K103027) does not equipped the functions such as the Contra-Check and Contra-Restriction of the subject device. The performance testing for the Contra-Check and Contra-Restriction does not raise any new concern. Therefore, the NLZ Motor system demonstrated substantial equivalence to the predicates.

Biocompatibility j.

Following biocompatibility test was performed using the Motor and the NLZ Endo:

  • Cytotoxicity in accordance with ANSI/AAMI/ISO 10993-5:2009
  • Sensitization in accordance with ISO 10993-10:2010
  • Irritation in accordance with ISO 10993-10:2010

In the biocompatibility testing reports, no biocompatibility concern was raised.

k. Bench Testing

The bench tests were performed in order to verify conformity to ISO 14457:2012.

All samples were compliant with ISO 14457:2012.

Also, electromagnetic compatibility testing in accordance with IEC 60601-1-2:2014 and electrical safety testing in accordance with AAMI / ANSI/ES60601-1:2005/(R)2012 and A1:2012,, C1:2009/(R)2012 and A2:2010/(R)2012 were performed. Those test results indicate that the NLZ Motor System does not raise concern.

{13}------------------------------------------------

Performance testing for the overheating risk reductions was performed in accordance with in-house standard. This testing is verification test for software by putting excessive load to the Motor in order to confirm proper Contra-Check judge and motor speed control for Contra-Restriction and Contra-SAFE. The test results indicate that those safety functions do not raise any new concern. In addition, we analyzed temperature increase for returned handpieces due to overheating/burn event and confirmed that the overheating/burn event should be reduced by the overheating risk reduction features.

No animal or clinical testing was performed.

l. Conclusion

The indications for use of the NLZ Motor System is identical to the E-STATIS 40 Electric Motor System (K090937). Although there are some differences in characteristics between the NLZ Motor System and the predicates, a number of performance testing indicated that the NLZ Motor System met the requirement of the standards. Based on above result, we conclude the NLZ Motor System is substantially equivalent to the predicates and does not raise any new safety and effectiveness concern.

§ 872.4200 Dental handpiece and accessories.

(a)
Identification. A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.(b)
Classification. Class I.