K Number
K163225

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-08-08

(264 days)

Product Code
Regulation Number
864.6550
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

AFIAS iFOB in conjunction with AFIAS 50 is a fluorescence immunoassay system for qualitative detection of fecal occult blood (FOB) in human fecal samples. AFIAS iFOB is an in vitro diagnostic test used by professional clinical laboratories and clinical reference laboratories for routine physical examination when gastrointestinal bleeding may be suspected. Intended users/operators for AFIAS iFOB is professional medical personnel.

Device Description

AFIAS iFOB in conjunction with AFIAS-50 is a fluorescence immunoassay system for qualitative detection of fecal occult blood (FOB) in human fecal samples.
Components of AFIAS iFOB: AFIAS iFOB consist of a test cartridge, ID chip, sample collection tube contain the extraction buffer, package insert, applicator sticks, collection slide, mailing envelope, sample collection tissues, instruction for use and patient instructions.
The AFIAS iFOB test cartridge contains a test strip; with a nitrocellulose membrane of which, mouse monoclonal anti hemoglobin labeled with fluorescence and anti rabbit IgG labeled fluorescence have been immobilized at the glaze line, mouse monoclonal anti hemoglobin at the test line and rabbit IgG at the control line. Each test cartridge is individually sealed in an aluminum foil pouch containing a desiccant. Twenty-five sealed test cartridges are packed in a box which also contains an ID chip and 25 mailing envelopes which contain a collection slide, applicator sticks and sample collection tissues.
The ID chip contains a memory device that contains encoded calibration data/information for the batch (lot-to-lot) variation. With the ID chip inserted in the designated port, AFIAS-50 reads and utilizes the calibration data regarding the batch/lot under consideration and applies appropriate correction to the conversion formula while computing the test result.
The AFIAS iFOB extraction buffer tube with extraction buffer is sealed with plastic caps. The upper side is capped with a plastic cap without a sampling stick. The bottom side is capped with a plastic cap with a sampling stick. The extraction buffer contains bovine serum albumin (BSA) as a stabilizer, tween 20 as a surfactant and sodium azide in phosphate buffered saline (PBS) as a preservative. Each extraction buffer tube contains 1 mL extraction buffer. Twenty-five pre-filled extraction buffer tubes are packed in a test cartridge box.
Components of AFIAS-50: Power cable, Barcode reader, Thermal printer paper, Collection tube rack holder, Sample tips, Test cartridge magazine, Waste bin, System check cartridge set.

AI/ML Overview

The provided text describes the performance characteristics of the AFIAS iFOB device, including precision, prozone effect, specificity, cross-reactivity, interference, stability studies, assay cut-off, and method comparison with a predicate device.

Here's the information extracted and organized according to your request:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly define "acceptance criteria" in a separate section with specific numerical thresholds for each performance characteristic. Instead, it presents study results and then generally states whether the results "passed acceptance criteria" or demonstrated "acceptable overall percent agreement." For the purpose of this response, I will interpret the reported performance as meeting implicit acceptance criteria due to phrases like "passed acceptance criteria" or "acceptable overall percent agreement."

CharacteristicAcceptance Criteria (Implicit)Reported Device Performance
PrecisionPassed acceptance criteria for repeatability and reproducibility.Repeatability: Overall Percent Agreement: 98.0%, Positive Percent Agreement: 100.0% (92.7% – 100.0%), Negative Percent Agreement: 95.9% (86.3% – 98.9%). Lot-to-Lot Reproducibility: Overall Percent Agreement: 98.3%, Positive Percent Agreement: 99.3% (96.3% – 99.9%), Negative Percent Agreement: 97.3% (93.2 – 98.9%). Between-run Reproducibility: Overall Percent Agreement: 98.0%, Positive Percent Agreement: 99.3% (96.3% – 99.9%), Negative Percent Agreement: 96.6% (92.2% – 98.5%). Between-Device Reproducibility: Overall Percent Agreement: 98.3%, Positive Percent Agreement: 98.7% (95.2% – 99.6%), Negative Percent Agreement: 98.0% (94.1% – 99.3%). Between-site Reproducibility: Overall Percent Agreement: 98.3%, Positive Percent Agreement: 98.6% (95.2% – 99.6%), Negative Percent Agreement: 98.0% (94.2% – 99.3%). Combined Reproducibility: Overall Percent Agreement: 98.2%, Positive Percent Agreement: 99.1% (98.0% – 99.6%), Negative Percent Agreement: 97.3% (95.8% – 98.3%).
Prozone (Hook Effect)Not susceptible to prozone/hook effect up to a specified hemoglobin concentration.Not susceptible to prozone/hook effect up to a hemoglobin concentration of 2000 ng/mL.
Specificity (Hb Variant)Equally sensitive to 'Hemoglobin S' as human hemoglobin.Equally sensitive to 'Hemoglobin S'.
Cross-ReactivityNo significant cross-reactivity with specified animal hemoglobin.No significant cross-reactivity with bovine, chicken, fish, horse, goat, pig, rabbit, and sheep hemoglobin at specified concentrations.
InterferenceNo significant interference with specified biomolecules.No significant interference with ascorbic acid, bilirubin, albumin, myoglobin, glucose, and triglyceride mixture at specified concentrations.
Test Kit StabilityStable for an estimated period at a given temperature range.Stable for an estimated period of 20 months at 4-30°C.
Stool Sample (Slide) StabilityStable for a specified duration at room temperature.Stable up to 30 days when stored at room temperature (25-30°C).
Stool Sample (Buffer Tube) StabilityStable for a specified duration when refrigerated.Stable up to 14 days when stored at 2 to 8°C.
Stool Sample (Cups) StabilityStable for specified durations at various temperatures.Stable for 30 days if stored at -20°C and for 2 days if stored at 37°C.
Controls StabilityShelf-life and open-vial stability for controls at specified conditions.i-CHROMA iFOB controls shelf-life stability was three months and open-vial stability was one month when stored at 2-8°C.
Humidity EffectNo humidity effect on test results up to a specified humidity level.No humidity effect on AFIAS iFOB test results up to 75±5% of humidity.
Assay Cut-offValidated cut-off value matching the specified threshold.Cut-off determined to be 8.0 µg hemoglobin/g stool or 100 ng/mL (hemoglobin in fecal sample mixed with detection buffer). Performance at cut-off: For 100 ng/mL, 57.5% Positive (85.1%-100.0% CI). For values below (85, 90 ng/mL) and above (110, 130, 150 ng/mL) the cut-off, the percent positive/negative results align with expectations.
Method Comparison (vs. Predicate)Acceptable overall percent agreement, positive percent agreement, and negative percent agreement with predicate device.Combined Sites: Overall Percent Agreement: 98.5% (94.8%-99.4% CI), Positive Percent Agreement: 98.2% (94.8%-99.4% CI), Negative Percent Agreement: 98.6% (96.8%-99.4% CI).
Specimen Collection VerificationNo statistical significance in analyses by two sampling methods, with acceptable agreement.Acceptable overall percent agreement as well as positive percent agreement and negative percent agreement. No statistical significance in the analyses by the two sampling methods.

2. Sample Sizes Used for the Test Set and Data Provenance

  • Precision/Reproducibility:
    • Sample Size: Seven fecal hemoglobin concentrations (4 µg Hb/g stool to 80 µg Hb/g stool or equivalent ng/mL). Fourteen replicates were performed for each sample and concentration level per study (repeatability, lot-to-lot, between-run, between-device, between-site, combined). Totaling 98 samples for repeatability and 294 for each reproducibility component, and 1274 for combined reproducibility results shown in the table.
    • Data Provenance: Not explicitly stated, but implies laboratory-controlled spiked fecal samples. Conducted at "one intended use site" for repeatability, and "three intended use sites" for reproducibility. Locations of these sites (e.g., country of origin) are not specified. The study design (spiked samples) suggests a controlled laboratory setting rather than direct patient samples.
  • Prozone (Hook Effect):
    • Sample Size: Twelve hemoglobin concentrations (700 ng/mL to 2000 ng/mL). Twenty aliquots of each sample mixed with extraction buffer.
    • Data Provenance: Spiked stool specimens with human blood (controlled laboratory setting).
  • Specificity & Cross-Reactivity:
    • Sample Size: Not explicitly stated for each test, but involved testing with specific animal hemoglobin concentrations and a 'Hemoglobin S' human variant.
    • Data Provenance: Spiked test samples (controlled laboratory setting).
  • Interference:
    • Sample Size: Not explicitly stated for each test, but involved testing with specific biomolecule concentrations.
    • Data Provenance: Spiked test samples (controlled laboratory setting).
  • Stability Studies (Test Kit, Stool Sample on Slide, Stool Sample in Buffer Tube, Stool Sample in Cups):
    • Sample Size: Nine hemoglobin concentrations (25 ng/mL to 500 ng/mL) for test kit, slide, and buffer tube stability. Twenty-one aliquots of each concentration for slide and buffer tube stability. Six hemoglobin concentrations (50 ng/mL to 150 ng/mL), with 384 specimen cups (64 cups x 6 concentrations x 4 storage temperatures) for stool sample in cups stability.
    • Data Provenance: Spiked Hb-free stool samples with human blood (controlled laboratory setting).
  • Stability of i-CHROMA iFOB Controls:
    • Sample Size: Ten aliquots of each control level. Three lots of i-CHROMA iFOB Controls (negative and positive).
    • Data Provenance: Not entirely clear if these are spiked or internal controls themselves, implying a controlled laboratory evaluation.
  • Humidity Effect Stability Study:
    • Sample Size: Seven hemoglobin concentrations (50 ng/mL to 1,000 ng/mL). 50 AFIAS iFOB test cartridges.
    • Data Provenance: Spiked Hb-free stool samples with human blood (controlled laboratory setting).
  • Assay Cut-off:
    • Sample Size: Seven fecal hemoglobin concentrations (50 ng/mL to 150 ng/mL). Forty aliquots of each concentration.
    • Data Provenance: Spiked stool samples with human blood (controlled laboratory setting).
  • Method Comparison:
    • Sample Size: 522 patient samples (165 positive, 357 negative).
    • Data Provenance: Patient samples. Conducted at "one professional medical laboratory in the U.S. and two international professional medical laboratories." The document does not specify if these were retrospective or prospective, but the phrasing "patient samples" and "method comparison" often implies prospective or a mix.
  • Specimen Collection Verification:
    • Sample Size: Eight hemoglobin concentrations (25 ng/mL to 1,000 ng/mL). Twenty aliquots of each concentration.
    • Data Provenance: Spiked Hb-free fecal samples with human blood (controlled laboratory setting).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • For Precision, Prozone, Specificity, Cross-Reactivity, Interference, Stability, Humidity Effect, Assay Cut-off, and Specimen Collection Verification studies: The "ground truth" was established by creating spiked fecal samples with known hemoglobin concentrations. This means the "expert" here is the precise control over the spiking process and analytical methods to determine the exact hemoglobin levels. Human experts were not involved in establishing this ground truth, rather the experimental design.
  • For Method Comparison: The ground truth for the 522 patient samples was established by the predicate device, i-CHROMA iFOB test (K132167). This study compared the new device's results to the predicate device's results. No explicit mention of human experts defining the "true" positive or negative status of these patient samples beyond the predicate device's output.

4. Adjudication Method for the Test Set

  • For Precision, Prozone, Specificity, Cross-Reactivity, Interference, Stability, Humidity Effect, Assay Cut-off, and Specimen Collection Verification studies: No adjudication method was mentioned as the ground truth was based on known, spiked concentrations or direct comparison to the predicate's established output for comparison.
  • For Method Comparison: No specific adjudication method (like 2+1 or 3+1) is mentioned. The comparison was directly between the AFIAS iFOB and the predicate i-CHROMA iFOB device.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. This device is an in vitro diagnostic (IVD) fluorescence immunoassay system for qualitative detection of fecal occult blood (FOB), which automatically produces a result (positive or negative) without human interpretation of images or complex data where an "AI assistant" would typically improve a human reader's performance. The studies performed were analytical performance studies and a method comparison with a predicate device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Yes, the performance studies described are essentially standalone. The AFIAS iFOB system (device with its instrument AFIAS-50) is an automated system. The performance characteristics (precision, stability, cut-off, method comparison) evaluate the device's ability to accurately detect FOB based on its internal algorithms and fluorescence detection, without human interpretation of the final result. The professional medical personnel are "users/operators," not "interpreters" who then make a diagnostic decision on top of what the device outputs.

7. The Type of Ground Truth Used

  • For most performance studies (Precision, Prozone, Specificity, Cross-Reactivity, Interference, Stability, Humidity Effect, Assay Cut-off, Specimen Collection Verification): The ground truth was spiked samples with known, carefully defined concentrations of human hemoglobin or other target analytes/interferents. This is a form of analytical ground truth.
  • For Method Comparison: The ground truth for the patient samples was established by the predicate device (i-CHROMA iFOB test). This means the new device was compared against a previously cleared and accepted device's results.

8. The Sample Size for the Training Set

The document does not mention a "training set" or machine learning in the context of device development. This type of IVD device (fluorescence immunoassay) typically relies on pre-programmed calibration data and antigen-antibody reactions, not on machine learning algorithms that require explicit training sets. The "ID chip" contains encoded calibration data, which serves a similar function to calibration in traditional assays rather than training in machine learning.

9. How the Ground Truth for the Training Set Was Established

As no "training set" is mentioned or implied for machine learning, this question is not applicable. The device's operation is based on established immunochemistry principles and factory-calibrated parameters transmitted via an "ID chip" for each lot.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 08. 2017

Immunostics, Inc. Antoinette Prusik Director of Regulatory Affairs 1750 Brielle Avenue, A5 Ocean, New Jersey 07712

Re: K163225

Trade/Device Name: AFIAS iFOB with AFIAS-50 Regulation Number: 21 CFR 864.6550 Regulation Name: Occult blood test Regulatory Class: Class II Product Code: OOX Dated: July 6, 2017 Received: July 7, 2017

Dear Ms. Prusik:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Leonthena R. Carrington -S

Lea Carrington Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K163225

Device Name AFIAS iFOB with AFIAS-50

Indications for Use (Describe)

AFIAS iFOB in conjunction with AFIAS 50 is a fluorescence immunoassay system for qualitative detection of fecal occult blood (FOB) in human fecal samples. AFIAS iFOB is an in vitro diagnostic test used by professional clinical laboratories and clinical reference laboratories for routine physical examination when gastrointestinal bleeding may be suspected. Intended users/operators for AFIAS iFOB is professional medical personnel.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows a red icon that resembles the letter 'i' inside a rectangular frame. The 'i' is stylized with a circle at the top and a rectangular body. The frame and the 'i' have a glossy, slightly transparent appearance, suggesting a digital or graphic design element. The background is plain and light, allowing the red icon to stand out.

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21CFR, Section 807.92 (c).

Submitter Information

Submitter nameImmunostics, Inc.
Address1750 Brielle Avenue, A5, Ocean, New Jersey 07712
Phone732-918-0770 / 800-722-7505 (Toll free)
Fax.732-918-0618
Websitewww.immunostics.com
Contact personAntoinette Prusik / Hyung-Ju Oh
e-mailaprusik@immunostics.com / ohj@boditech.co.kr
510(k) Summary preparationdateNovember, 16, 2016

Device Information

Common name of deviceAFIAS iFOB
Trade nameAFIAS iFOB with AFIAS-50
Classification nameAutomated Occult Blood Analyzer
Regulatory information21 CFR 864.6550, Occult blood test
ClassificationClass II for device
Product codeOOX, Automated occult blood analyzer

Predicate Device

Trade Namei-CHROMA iFOB with i-CHROMA Readeri-CHROMA iFOB Controls
K NumberK132167

750 Brielle Ave. • A5 • Ocean, New Jersey • 07712
www.lmmunostics.com

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows a red and white icon. The icon is a red letter "i" inside of a red rectangle. The letter "i" is white, and the rectangle has a glossy finish.

Intended use(s):

AFIAS iFOB in conjunction with AFIAS-50 is a fluorescence immunoassay system for qualitative detection of fecal occult blood (FOB) in human fecal samples. AFIAS iFOB is an in vitro diagnostic test used by professional clinical laboratories and clinical reference laboratories for routine physical examination when gastrointestinal bleeding may be suspected. Intended users/operators for AFIAS iFOB is professional medical personnel.

Device Description

AFIAS iFOB in conjunction with AFIAS-50 is a fluorescence immunoassay system for qualitative detection of fecal occult blood (FOB) in human fecal samples.

    1. Components of AFIAS iFOB:
      AFIAS iFOB consist of a test cartridge, ID chip, sample collection tube contain the extraction buffer, package insert, applicator sticks, collection slide, mailing envelope, sample collection tissues, instruction for use and patient instructions.
  • The AFIAS iFOB test cartridge contains a test strip; with a nitrocellulose membrane ದಿ. of which, mouse monoclonal anti hemoglobin labeled with fluorescence and anti rabbit IgG labeled fluorescence have been immobilized at the glaze line, mouse monoclonal anti hemoglobin at the test line and rabbit IgG at the control line. Each test cartridge is individually sealed in an aluminum foil pouch containing a desiccant. Twenty-five sealed test cartridges are packed in a box which also contains an ID chip and 25 mailing envelopes which contain a collection slide, applicator sticks and sample collection tissues.

  • b. The ID chip contains a memory device that contains encoded calibration data/information for the batch (lot-to-lot) variation. With the ID chip inserted in the designated port, AFIAS-50 reads and utilizes the calibration data regarding the batch/lot under consideration and applies appropriate correction to the conversion formula while computing the test result.

  • The AFIAS iFOB extraction buffer tube with extraction buffer is sealed with plastic C. caps. The upper side is capped with a plastic cap without a sampling stick. The bottom side is capped with a plastic cap with a sampling stick. The extraction buffer contains bovine serum albumin (BSA) as a stabilizer, tween 20 as a surfactant and sodium azide in phosphate buffered saline (PBS) as a preservative. Each extraction buffer tube contains 1 mL extraction buffer. Twenty-five pre-filled extraction buffer tubes are packed in a test cartridge box.

    1. Components of AFIAS-50:
    • a. Power cable

Image /page/4/Picture/13 description: The image shows the word "immunostics, inc." in a stylized font. The word "immunostics" is in a bold, sans-serif font, and the letters are slightly spaced apart. There is a starburst symbol to the left of the word "immunostics". The word "inc." is in a smaller, sans-serif font and is located to the right of the word "immunostics".

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image is a red and white sign with a stylized letter "i" in the center. The "i" is composed of a circle above a rectangle, both in red. The background of the sign is white, and there is a red border around the edge. The sign has a glossy appearance, suggesting it may be made of plastic or another reflective material.

  • b. Barcode reader
  • Thermal printer paper c.
  • d. Collection tube rack holder
  • Sample tips e.
  • Test cartridge magazine f.
  • Waste bin g.
  • h. System check cartridge set
Similarities
ItemDeviceAFIAS iFOB withAFIAS-50, K163225Predicatei-CHROMA iFOB withi-CHROMA Reader, K132167
Intended/Indicationsfor UseAFIAS iFOB in conjunction withAFIAS-50 is a fluorescenceimmunoassay system for qualitativedetection of fecal occult blood(FOB) in human fecal samples.AFIAS iFOB is an in vitrodiagnostic test used by professionalclinical laboratories and clinicalreference laboratories for routinephysical examination whengastrointestinal bleeding may besuspected. Intended users/operatorsfor AFIAS iFOB is professionalmedical personnel.i-CHROMA iFOB inconjunction with i-CHROMAReader is a fluorescenceimmuno-chromatographic assaysystem for qualitative detectionof fecal occult blood (FOB) inhuman fecal samples. i-CHROMA iFOB is an in vitrodiagnostic test used bylaboratories and physicianoffices for routine physicalexamination whengastrointestinal bleeding may besuspected.
Test PrincipleLateral flow chromatographicfluorescence immunoassaySame
Sample TypeHuman feces (Mixed with detectionbuffer)Same
Assay Cut-off8µg hemglobin (Hb)/g of stool or100 ng/mL (Human hemoglobin inhuman fecal sample mixed withdetection buffer)Same
Test DeviceTest Cartridge (Test strip enclosedin a plastic housing)Same
DetectionMechanismScanning/measurement of intensityof fluorescence on the processedsample-loaded test cartridgemembrane.Same

Image /page/5/Picture/10 description: The image shows the logo for "immunostics, inc.". The logo features a stylized sun or starburst symbol to the left of the text. The text is in a serif font and appears to be in a gray color.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows a red and white icon. The icon is a red rectangle with a white border. Inside the rectangle is a white letter "i" with a circle above it. The icon is likely used to represent information or details.

Differences
ItemDeviceAFIAS iFOB withAFIAS-50, K163225Predicatei-CHROMA iFOB withi-CHROMA Reader, K132167
Sampling andSample ProcessingSamples can be collected with acollection slide/applicator stick orsampling stick only (which is part ofthe extraction buffer tube). The fecalsample is delivered into theextraction buffer tube containing thedetection buffer which extracts thehemoglobin in the sample.Sampling is done with the helpof the Sampling Stick/Samplerwhich is part of the samplecollection tube. The fecalsample is delivered into thesample collection tubecontaining the detection bufferwhich extracts the hemoglobinin the sample.

Test Principle

AFIAS iFOB is an immunoassay system based on antigen-antibody reaction and fluorescence technology. When a human fecal sample is mixed with the extraction buffer, human hemoglobin in fecal occult blood (FOB) is extracted from fecal sample. When the extracted sample is loaded into the sample well of the test cartridge, human hemoglobin binds with mouse monoclonal anti hemoglobin-fluorescence located at the glaze line. This complex migrates through the nitrocellulose matrix of the test strip. The fluorochrome-labeled detector antibody-FOB hemoglobin complexes get captured on to the antibodies (mouse monoclonal anti hemoglobin) which have been immobilized at the test line on the test strip. As a result, the fluorochromelabeled complexes of the detector antibody and FOB hemoglobin capture-antibody are accumulated at the test line on the test cartridge membrane. The more hemoglobin in the human fecal sample, the more complexes that will be accumulated at the test line on the test cartridge membrane. The test strip also contains rabbit IgG immobilized at a control line. The control line of AFIAS iFOB test cartridge functions as the built-in procedural control.

AFIAS-50 is a fluorescence-scanning instrument to be used in conjunction with AFIAS iFOB test which are based on antigen-antibody reaction and fluorescence technology (Fluorescence Immuno-Assay). AFIAS iFOB test cartridges and sample mixed extraction buffer tubes are inserted into the designated positions on AFIAS-50 for testing. AFIAS-50 uses a semiconductor diode laser as the excitation light source for illuminating the sample-loaded AFIAS iFOB cartridge(s) inserted in its magazine station; thereby triggering fluorescence from the fluorochrome-labeled complexes of hemoglobin accumulated at the test line on the cartridge membrane as well as the control line. The fluorescent light is collected together with the scattered laser light. Pure fluorescence is filtered from the mixture of the scattered and fluorescent light. Intensity of the fluorescence generated at the 'test line' and 'control line' is scanned and collected by the optical assembly onto a photo sensor and converted into an electric signal which correlates to the intensity of fluorescence and hence to the concentration of FOB hemoglobin in the test sample. The on-board microprocessor computes the FOB hemoglobin

immunostics, inc.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image is a red and white icon. The icon is a red rectangle with a white border. Inside the rectangle is a white "i" with a red circle above it. The "i" is in a sans-serif font.

concentration based on a pre-programmed calibration derived from the "AFIAS iFOB ID Chip" inserted in the ID chip port. The computed and converted result is displayed by AFIAS-50 in a qualitative (positive or negative) manner based on an analytical cut-off of 100 ng/mL or 8.0 µg hemoglobin/g stool.

Performance Characteristics

Precision/Reproducibility

Repeatability was evaluated using a single test kit lot, one intended use site and one operator; whereas reproducibility was conducted across three intended use sites using three test kit lots, three operators, three AFIAS-50 using one run per day for three non-consecutive days. i-CHROMA iFOB Negative and Positive Controls were also tested daily to ensure and confirm the validity of the test results.

For repeatability and reproducibility, Hb-free fecal samples were collected and spiked with human whole blood having known hemoglobin level to achieve the following seven fecal hemoglobin concentrations: 4 µg Hb/g stool, 6.8 µg Hb/g stool, 7.6 µg Hb/g stool, 8.0 µg Hb/g stool, 8.4 µg Hb/g stool, 8.8 µg Hb/g stool, and 80 µg Hb/g stool, that are equivalent to 50 ng/mL, 85 ng/mL, 95 ng/mL, 105 ng/mL, 105 ng/mL, and 1000 ng/mL, respectively. Fourteen replicates were performed for each sample and concentration level. Repeatability and reproducibility results at all test sites passed acceptance criteria.

Type ofPrecision StudyActualResultsAFIASiFOBExpected ResultsOverallPercentAgreementPositivePercentAgreement(95% CI)NegativePercentAgreement(95% CI)
PositiveResultsNegativeResultsTotalResults
RepeatabilityPositiveResults4925198.0%100.0%(92.7% –100.0%)95.9%(86.3% –98.9%)
NegativeResults04747
TotalResults494998
Lot-to-LotReproducibilityPositiveResults147415198.3%99.3%(96.3% –99.9%)97.3%(93.2 –98.9%)
NegativeResults1142143
TotalResults148146294
Between-runReproducibilityPositiveResults147515298.0%99.3%(96.3% –99.9%)96.6%(92.2% –98.5%)
NegativeResults1141142

immunostics, inc.

rielle Ave. · A5 · Ocean, New Jersey · 07712 00) 722-7505 · Fax: (732) 918-0618 www.lmmunostics.com

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image displays the text "Quality Microbiological, Serological and Immunological Reagents." The text is written in a red, serif font. The words are centered on the image and are the only element present.

Type ofPrecision StudyActualResultsAFIASiFOBExpected ResultsPositiveResultsExpected ResultsNegativeResultsExpected ResultsTotalResultsOverallPercentAgreementPositivePercentAgreement(95% CI)NegativePercentAgreement(95% CI)
TotalResults148146294
Between-DeviceReproducibilityPositiveResults146314998.3%98.7%(95.2% –99.6%)98.0%(94.1% –99.3%)
NegativeResults2143145
TotalResults148146294
Between-siteReproducibilityPositiveResults145314898.6%(95.2% –99.6%)98.0%(94.2% –99.3%)
NegativeResults214414698.3%
TotalResults147147294
CombinedReproducibilityPositiveResults63417651
NegativeResults661762398.2%99.1%(98.0% –99.6%)97.3%(95.8% –98.3%)
TotalResults6406341,274

Prozone (Hook Effect)

Susceptibility of the AFIAS iFOB test to prozone effects was evaluated by testing hemoglobinfree stool specimens spiked with human blood of known hemoglobin concentrations to obtain the following concentrations: 700 ng/mL, 800 ng/mL, 1000 ng/mL, 1100 ng/mL, 1200 ng/mL, 1300 ng/mL, 1400 ng/mL, 1500 ng/mL, 1700 ng/mL, 1800 ng/mL, 1900 ng/mL and 2000 ng/mL. Twenty aliquots of each sample mixed with extraction buffer in the specimen collection tubes were prepared and tested. i-CHROMA iFOB Negative and Positive Controls' were tested to ensure and confirm the validity of the test results obtained with AFIAS iFOB.

It was determined that the AFIAS iFOB test is not susceptible to prozone/hook effect up to a hemoglobin concentration of 2000 ng/mL.

Specificity to Human Hemoglobin Variant, Cross-Reactivity, and Interfering Subtances

Specificity to human hemoglobin variant

AFIAS iFOB has found to be equally sensitive to 'Hemoglobin S' as the human hemoglobin variant associated with sickle cell anemia.

immunostics, inc.

750 Brielle Ave. • A5 • Ocean, New Jersey • 07712 (800) 722-7505 · Fax: (732) 918-0618 www.lmmunostics.com

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows a red icon with a white background. The icon is a stylized letter "i" inside a red rectangle. The letter "i" is composed of a vertical line and a circle above it. The icon is simple and easily recognizable.

Cross-Reactivity

There, in test samples, are biomolecules such as bovine hemoglobin, chicken hemoglobin, fish hemoglobin, horse hemoglobin, goat hemoglobin, pig hemoglobin, rabbit hemoglobin and sheep hemoglobin were added to the test samples. AFIAS iFOB test results did not show any significant cross-reactivity with these animal hemoglobin.

Animal speciesConcentration of animal hemoglobin
Bovine2,000 µg/mL
Chicken500 µg/mL
Fish100 µg/mL
Horse500 µg/mL
Goat500 µg/mL
Pig500 µg/mL
Rabbit500 µg/mL
Sheep500 µg/mL

Interference

There, in test samples, are biomolecules such as ascorbic acid, bilirubin, albumin, myoglobin, glucose and triglyceride mixture were added to the test samples at concentrations much higher than their normal physiological levels or according to EP7-A2. AFIAS iFOB test results did not show any significant interference with these biomolecules.

Endogenous substanceConcentration
Ascorbic acid30 µg/mL
Bilirubin200 µg/mL
Albumin60 mg/mL
Myoglobin2,000 μg/mL
Glucose120 mg/dL
Triglyceride mixture500 mg/dL

Stability Studies

i-CHROMA iFOB Negative and Positive Controls were also tested daily to ensure and confirm the validity of the test results of the stability studies.

1. Test Kit Stability (Accelerated)

The accelerated stability study was conducted with 3 lots of AFIAS iFOB test kit (test cassette and sampling tubes). Test kit stability testing was performed by preparing 21 aliquots of spiked stool test samples collected in extraction buffer in AFIAS iFOB test sample collection tubes. The Hb-free stool specimens were spiked with human blood (of known hemoglobin level) to obtain fecal samples containing nine different hemoglobin concentrations: 25 ng/mL, 50 ng/mL, 85 ng/mL, 90 ng/mL, 100 ng/mL, 110 ng/mL, 130 ng/mL, 150 ng/mL and 500 ng/mL. One operator tested the three lots stored at 50°C and test results were collected at time point 0, 1, 2, 4, 8, 12, 16, 20, 24 and 28 week intervals. The

Image /page/9/Picture/12 description: The image shows the logo for "Immunostics, inc.". The logo is in a serif font and is gray in color. To the left of the word "Immunostics" is a symbol that looks like a starburst.

0 Brielle Ave. · A5 · Ocean, New Jersey · 07712 0) 722-7505 · Fax: (732) 918-0618 www.lmmunostics.com

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image is a red and white icon. The icon is a red rectangle with a white border. Inside the rectangle is a white "i" with a circle above it. The "i" and circle are outlined in red.

results showed that the test kit was stable for an estimated period of 20 months at 4-30°C.

    1. Stool Sample Collected by Collection Slide Stability
      Stool samples collected on collection slides stability testing was performed by using one lot of AFIAS iFOB collection slides and spiked Hb-free stool samples with human blood (of known hemoglobin concentration) to obtain fecal samples containing nine different hemoglobin concentrations: 25 ng/mL, 50 ng/mL, 90 ng/mL, 90 ng/mL, 100 ng/mL, 110 ng/mL, 130 ng/mL, 150 ng/mL and 500 ng/mL. Twenty one aliquots of each of the nine concentrations of spiked stool test samples were added to the AFIAS iFOB sample collection slide and analyzed in random order.

To claim 30 days of sample stability when stored at room temperature (25-30°C) each test sample was tested with AFIAS iFOB on day 0 and every 4 days for a total of 36 days from the start of storage. Stool samples collected on AFIAS iFOB collection slides are stable up to 30 days when stored at room temperature.

    1. Stool Sample Collected by Extraction Buffer Tube Stability
      Stool samples collected in extraction buffer tubes stability testing was performed by using one lot of AFIAS iFOB extraction buffer tubes and spiked Hb-free stool samples with human blood (of known hemoglobin concentration) to obtain fecal samples containing nine different hemoglobin concentrations: 25 ng/mL, 50 ng/mL, 90 ng/mL, 90 ng/mL, 100 ng/mL, 110 ng/mL, 130 ng/mL, 150 ng/mL and 500 ng/mL. Twenty one aliquots of each of the nine concentrations of spiked stool test samples were collected by the AFIAS iFOB sampling sticks at 0 day and added to the AFIAS iFOB extraction buffer tubes and analyzed in random order.

To claim 14 days of sample stability when refrigerated each test sample was stored at 2 to 8°C and tested with AFIAS iFOB on day 0 and every 2 days for a total of 18 days from the start of storage. AFIAS iFOB extraction buffer tubes must be brought back to room temperature before testing. Stool samples collected in AFIAS iFOB extraction buffer tubes are stable up to 14 days when stored at 2 to 8°C.

    1. Stool Sample Collected in Sterile Specimen Collection Cups
      Stool samples collected in collection cups stability testing was performed by using one lot of AFIAS iFOB test kits and spiked Hb-free stool samples with human blood (of known hemoglobin concentration) to obtain fecal samples containing nine different hemoglobin concentrations: 50 ng/mL, 75 ng/mL, 90 ng/mL, 110 ng/mL, 125 ng/mL, 150 ng/mL, Three hundred eighty four specimen cups (64 cups x 6 hemoglobin concentrations x 4 storage temperatures) containing the spiked fecal samples were stored -20° C, 4° C, 25° C and 37° C immediately after spiking/preparation. Each test sample was tested with AFIAS iFOB at day 0 and every other day.

Stool test samples collected in sterile specimen collection cups were stable for 30 days if stored at -20° C and for 2 days if stored at 37° C.

Image /page/10/Picture/12 description: The image shows the logo for Immunostics, Inc. The logo features a stylized sunburst to the left of the word "immunostics" in a serif font. The letters are in gray, and the word "inc." appears to the right of "immunostics" in a smaller font size.

) Brielle Ave. · A5 · Ocean. New Jersey · 07712 00) 722-7505 · Fax: (732) 918-061 www.lmmunostics.com

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image shows a red and white icon. The icon is a red square with a white background. Inside the square is a red "i" with a circle on top. The icon is likely used to indicate information or details.

5. Stability of i-CHROMA iFOB Controls

Stability of i-CHROMA iFOB Controls was evaluated by real time stability using ten aliquots of each of the control levels. One lot of AFIAS iFOB test kits and three lots of i-CHROMA iFOB Controls (negative and positive) were evaluated during the shelf-life and open-vial stability studies of the i-CHROMA iFOB Controls.

The shelf-life stability was evaluated by using separate vials (of each lot) of the controls for each periodic testing. The open-vial stability was evaluated by using the same vial of the control for the initial as well as all successive periodic testing. i-CHROMA iFOB controls from each lot were stored at 2℃, 8℃ and 25℃ and tested at time points of initial start of 0 and biweekly up to 16 weeks from the start of the stability study. The results showed that the i-CHROMA iFOB controls shelf-life stability was three months and the open-vial stability was one month when stored at 2-8°C.

6. Humidity Effect Stability Study

Humidity stability was conducted with one lot of AFIAS iFOB test kits and spiked Hb-free stool samples with human blood (of known hemoglobin concentration) to obtain fecal samples containing seven different hemoglobin concentrations: 50 ng/mL, 95 ng/mL, 100 ng/mL, 105 ng/mL, 110 ng/mL and 1,000 ng/mL. 50 AFIAS iFOB test cartridges were stored at a temperature of 25℃ and the humidity conditions in table 2 below and tested at 0, 0.5, 1, 2 and 5 hours intervals. The positive percent for each concentration in each humidity condition was calculated. The test results showed that there is no humidity effect on AFIAS iFOB test results up to 75±5% of humidity.

Humidity Test Conditions

DayDay 1Day 2Day 3Day 4
Humidity$35\pm5%$$55\pm5%$$65\pm5%$$75\pm5%$

Assay cut-off

The cut-off value for the AFIAS iFOB test was validated in house. Fecal test samples were prepared by spiking stool samples with human blood of known hemoglobin concentration, to obtain the following fecal hemoglobin concentrations: 50 ng/mL, 90 ng/mL, 90 ng/mL, 100 ng/mL, 110 ng/mL, 130 ng/mL, and 150 ng/mL. Forty aliquots of each of the seven concentrations of spiked stool test samples were mixed with extraction buffer in AFIAS iFOB extraction buffer tubes and 40 aliquots of each of the seven concentrations of spiked stool test samples were mixed with the predicate (i-CHROMA iFOB) sample collection tubes. Samples were tested in randomized order. Testing was performed side-by-side with the predicate by comparing the test results of the device with that of the predicate. The cut-off was determined to be 8.0 µg hemoglobin/g stool or 100 ng/mL (hemoglobin in fecal sample mixed with detection buffer).

Image /page/11/Picture/11 description: The image shows the logo for "immunostics, inc.". The logo is in a serif font and is gray in color. To the left of the word "immunostics" is a graphic of a starburst.

{12}------------------------------------------------

ConcentrationNAFIAS iFOB TestPercent Positive(95% CI)Percent Negative(95% CI)
50 ng/mL400400%(0% - 8.8%)100.0%(91.2% - 100.0%)
85 ng/mL402385.0%(9.5% - 90.6%)95.0%(86.5% - 99.5%)
90 ng/mL4053512.5%(51.0% - 100.0%)87.5%(85.8% - 99.5%)
100 ng/mL40231757.5%(85.1% - 100.0%)42.5%(74.3% - 99.0%)
110 ng/mL4038295.0%(86.5% - 99.5%)5.0%(9.5% - 90.6%)
130 ng/mL40400100.0%(91.2% - 100.0%)0%(0% - 8.8%)
150 ng/mL40400100.0%(91.2% - 100.0%)0%(0% - 8.8%)

Method Comparison

A method comparison of AFIAS iFOB test with the predicate test, i-CHROMA iFOB test, was conducted by assessing 522 patient samples, 165 of which were positive, and 357 negative using different lots of the proposed device and one lot of the predicate device. The method comparison study was performed at one professional medical laboratory in the U.S. and two international professional medical laboratories by three different operators at (one) at each site. The i-CHROMA iFOB external controls (positive and negative) were run prior to testing. Statistical analysis of site-wide test results as well as combined results showed that AFIAS iFOB test results have acceptable overall percent agreement as well as positive percent agreement and negative percent agreement with i-CHROMA iFOB test results. The method comparison study (see table below) demonstrated that the analytical performance of the AFIAS iFOB test is substantially equivalent to the predicate device.

StudySiteNew TestAFIASiFOBK163225Predicatei -CHROMA iFOB(K132167)OverallPercentAgreementPositivePercentAgreement(95% CI)NegativePercentAgreement(95% CI)
PositiveNegative
Study Site1Positive50199.3%100%(92.9% 100%)99.0%
Negative099(94.6%-
Total5010099.8%)
Study Site2Positive50397.2%96.2%97.7%
Negative2125(87.0%-(93.3%-
Total5212898.9%)99.2%)
Positive62199.0%

immunostics, inc.

0 Brielle Ave. · A5 · Ocean, New Jersey · 07712 (800) 722-7505 ・Fax: (732) 918-0618 www.lmmunostics.com

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows a red rectangle with rounded corners, inside of which is a white letter 'i' with a circle above it. The letter 'i' and the rectangle are both outlined in red. The background of the image is white.

Study Site3Negative112898.4%99.2%
Total63129(91.5%-99.7%)(95.7%-99.9%)
CombinedSitesPositive162598.5%98.2%98.6%
Negative3352(94.8%-99.4%)(96.8%-99.4%)
Total165357(94.8%-99.4%)(96.8%-99.4%)

Specimen Sampling and Handling

To collect human fecal samples for testing, two different sample collection methods are available. AFIAS iFOB uses specific sampling devices (collection slide and sampling stick). Using a collection slide (for use by patients), fecal samples are collected with an applicator stick applied into the windows which are provided inside of the collection slide. After collecting the fecal sample, the sample strip which detaches from the back side of the collection slide will be inserted into the extraction buffer tube. The patient submits the test sample to the hospital/laboratory for testing. The fecal test sample can also be collected using a sampling stick (intended to be used by professional/reference laboratories only) which is attached in the bottom cap of the extraction buffer tube. After collection of the fecal sample, the sampling stick with fecal sample will be inserted into the extraction buffer tube.

Calibration

The "ID Chip" provided with AFIAS iFOB test cartridge(s) contains a memory device which contains encoded calibration data. An ID chip having a lot number matching with that of the test cartridge and the extraction buffer collection tube is inserted into the designated port of the AFIAS-50 which utilizes the calibration data for computing the test result(s). The end user is not required to perform routine calibration before performing the test.

Quality Control

AFIAS iFOB Cartridge

Internal Procedural Control: The Procedural Control is found in the procedural control region of the test cartridge to assure the operator that the test has been properly performed. It confirms sufficient specimen volume and correct proecedure technique. This control does not ensure that the capture antibody is accurately detecting the presence or absence of Hb in the sample.

AFIAS-50

Pre-programmed System Self-Check and System Check of AFIAS-50 using System Check Cartridge & ID Chip are electronic control mechanisms to check whether electronic features of the AFIAS-50 are within specifications.

External control

It is recommended that positive and negative controls be performed to verify proper test performance. External controls are not provided with the test kit. External controls are used to assure the operator that the capture and conjugated antibodies are present and reactive. Controls should be assayed according to the manufacturer's instructions once per kit lot,

Image /page/13/Picture/14 description: The image shows the logo for "immunostics, inc.". The logo is in a serif font and is gray in color. To the left of the text is a graphic of a starburst with a horizontal line through the middle.

{14}------------------------------------------------

i

following the local and state guidelines. If controls do not perform as expected, the test results should not be used.

Specimen Identification

Enter patient identification information manually on the label of the collection slide or extraction buffer tube.

Specimen Collection Verification

Verification that the sample collection methods using a sample collection slide and applicator stick (used by patients and laboratories) versus sampling stick (used by laboratories only) consistently delivers the specified amount of stool required for optimal test performance was performed by using one lot of AFIAS iFOB extraction buffer tubes (with sampling sticks) and collection slides (with applicator stick). Hb-free fecal samples were collected and spiked with human blood (of known hemoglobin concentration) to obtain fecal samples containing eight different hemoglobin concentrations: 25 ng/mL, 50 ng/mL, 100 ng/mL, 100 ng/mL, 120 ng/mL, 150 ng/mL, 500 ng/mL and 1,000 ng/mL. Twenty aliquots of each of the eight concentrations of spiked stool test samples were tested by each sample collection method. i-CHROMA iFOB Negative and Positive Controls were also tested to ensure and confirm the validity of the test results. Statistical analysis of the sample collection study shows that the test results from the two sampling methods have acceptable overall percent agreement as well as positive percent agreement and negative percent agreement. There was no statistical significance in the analyses of the test samples by the two sampling methods.

Image /page/14/Picture/7 description: The image shows the logo for Immunostics, Inc. The logo is in a gray, sans-serif font. To the left of the word "Immunostics" is a symbol that looks like a starburst with a horizontal line through the middle. The word "inc." is to the right of the word "Immunostics."

§ 864.6550 Occult blood test.

(a)
Identification. An occult blood test is a device used to detect occult blood in urine or feces. (Occult blood is blood present in such small quantities that it can be detected only by chemical tests of suspected material, or by microscopic or spectroscopic examination.)(b)
Classification. Class II (special controls). A control intended for use with an occult blood test is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.