K Number
K162754

Validate with FDA (Live)

Date Cleared
2017-01-12

(104 days)

Product Code
Regulation Number
862.1825
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The FREND™ Vitamin D test is a rapid indirect competitive fluorescent immunoassay designed for the quantitative measurement of 25-Hydroxy vitamin D and related hydroxylated metabolites in human serum and plasma (K;EDTA, lithium-heparin and citrate) specimens using the FREND™ System, and the FREND™ AP System. Measurements of total 25-hydroxy vitamin D and related hydroxylated metabolites are used to aid in the assessment of vitamin D sufficiency.

The FREND™ Vitamin D microfluidic flow cartridge is designed for use in the FREND™ System fluorescent immunoassay reader, and the FREND™ AP System. The FREND™ Vitamin D Test System is intended for use in clinical laboratories. For in vitro diagnostic use only. The test is not intended for use in point-of-care settings.

Device Description

The FREND™ Vitamin D Test is a rapid fluorescence immunoassay designed to quantify the concentration of total 25-hydroxy (OH) vitamin D and related hydroxylated metabolites in human serum and plasma (K3 EDTA, lithium-heparin and citrate) samples using the FREND™ System.

The FREND™ Vitamin D Test System is a competitive immunoassay with gold microparticles labeled with Vitamin D-specific monoclonal anti-Vitamin D-antibody (mouse), Vitamin D-biotin labeled with fluorescence nanoparticles and fluorescence detection by the FREND™ System.

The FREND™ Vitamin D Cartridge is a single-use disposable rapid lateral flow fluorescence immunoassay designed to quantify the concentration of total 25-OH vitamin D and related hydroxylated metabolites in human serum and plasma (K3 EDTA, lithiumheparin and citrate) samples. Each test cartridge contains reagents to perform one test and bears a bar code with test type and lot information. Each FREND™ Vitamin D Cartridge contains a test zone and a reference zone (internal control). The fluorescence measured in the test and reference zones corresponds to the amount of fluorescent conjugates of Vitamin D-biotin that is bound to each zone. Cartridges are provided in individually sealed pouches and must be stored between 2 and 8 °C.

In addition to the QC Code Chip there is a Vitamin D Code Chip. Lot information and calibration data are loaded into the software via the Vitamin D Code Chip provided with each reagent kit. Up to three different Vitamin D Code Chip lots can be loaded and used at the same time. Each reagent cartridge bears a barcode that the FREND™ System uses to identify the test type and lot number and links it to the appropriate analytical program and calibration information. The FREND™ System will not output results if the Vitamin D Code Chip lot does not match the lot of the cartridge used.

The Vitamin D Pretreatment Tube is a small single-use disposable plastic tube that contains Vitamin D gold nano-particle coniugation antibody. The sample is diluted in the dilution tube containing perfluorohexanoic acid and transferred to the pretreatment tube prior to being inserted into the FREND™ AP System to be mixed and then added to the Vitamin D Test Cartridge. The test cartridge is then inserted into the FREND™ System to be analyzed.

The FREND™ AP System is an "Advanced Preparing" device that performs automated pre-analytical steps of mixing, timed heating, and pipetting of the sample from the Gold Antibody Pretreatment tube into the FREND Vitamin D test cartridge.

The FREND™ System utilizes microfluidic technology and detects immune-complexes bound to Vitamin D. The test cartridge is placed on the warming platform of the FREND™ AP System. A 35 µL patient sample is added to the dilution tube and 70 µL of diluted sample is pipetted into the pretreatment tube containing monoclonal anti-Vitamin D antibody conjugated with gold nanoparticles(Step 1). The tube is then loaded into the FREND™ AP System (Step 2). The tube is mixed and loaded into the test cartridge automatically. In the AP System (Step 3), mixing, reaction, sample addition, hydration of the Vitamin D-biotin fluorescent bead conjugate, and migration along the cartridge channel takes approximately 15 minutes, after which the cartridge is read on the FREND™ System. During migration, the bound Vitamin D in the sample and the fluorescent bead conjugates of Vitamin D-biotin compete to form antigen-antibody complex in the test zone. Unbound Vitamin D-biotin fluorescent conjugates flow through and bind to the anti-Vitamin D antibody that is fixed on the surface in the reference zone. The fluorescent signals in the test and reference zones are measured, typically within 4 minutes.

Vitamin D quantification is based upon the ratio of the intensity of the test and reference zones. A lower ratio of fluorescence is indicative of a higher Vitamin D concentration; in other words, the magnitude of the fluorescent ratio is inversely proportional to the amount of Vitamin D in the sample.

The measuring range of the FREND™ Vitamin D Test System is 13.0 to 96.0 ng/mL. Results are determined via a lot-specific calibration curve which is generated by the manufacturer using a four-point calibration determined from values averaged from 6 replicates at each level. The established curve is uploaded to the FREND™ System via the Vitamin D Code-chip and is valid until the lot expiration date.

The FREND™ System is a portable, automated FREND™ cartridge reader. The FREND™ System is based on quantitative immunoassay technology capable of quantifying single or multiple analytes by measuring laser-induced fluorescence in a single-use disposable reagent cartridge. The test cartridge is a disposable plastic device that contains a port or opening (inlet) where the sample is applied. Once the sample is applied, it will mix with the reagents and travel towards the detection area via capillary action.

The FREND™ cartridge utilizes microfluidics lateral flow technology where the analyte of interest in the sample forms immune complexes while moving through the fluidics pathway in the cartridge. The concentration of the analyte of interest in an unknown sample is calculated using the ratio of the fluorescent intensity of the test zone and the reference zone.

The FREND™ System is a bench top fluorescence reader containing a touch screen user interface. The System has a slot that accepts the FREND™ Vitamin D Test Cartridge (which contains the reagents and sample), and is programmed to analyze the Test when the sample has fully reacted with the on-board in-cartridge reagents. Results of the test are displayed on the screen and can be printed on an optional printer.

The FREND™ System software controls the graphical user interface, communication with hardware, database management and data analysis. The software also controls the functions of the mechanical components including the motor, laser, printer control and acquisition of data from the sensor. The user can set the time and date and enter patient ID through the graphic user interface. The user cannot make any changes to the software.

The FREND™ Vitamin D Test System includes the following in the reagent kit:

  • . 20 FREND™ Vitamin D Test Cartridges
  • 20 FREND™ Sample Dilution Tubes .
  • 20 FREND™ Vitamin D Gold Antibody Pretreatment Tubes .
  • 30 Disposable pipette tips
  • 1 FREND™ Vitamin D Code Chip .
  • . 1 FREND™ Vitamin D Package Insert

One cartridge contains:

  • . Monoclonal mouse anti- Vitamin D 1.6 ± 0.16 na
  • . 25-hydroxyvitamin D 136 ± 13.6 ng
  • . Fluorescence particles 2.4 ± 0.24 µg

One sample dilution tube contains:

  • . Perfluorohexanoic acid 9.5 ± 0.95 µg
    One Gold antibody pretreatment tube contains:

  • . Gold nano-particle conjugation antibody 7.0 ± 0.7 µg
    The FREND™ System (previously cleared in K124056 (FREND™ PSA), K131928 (FREND™ TSH), K152422 (FREND™ Free T4), and K153577 (FREND™ Testosterone) is not provided with the kit but it is required for the FREND™ Vitamin D Test cartridge.

AI/ML Overview

Here's an analysis of the provided text to extract the requested information about acceptance criteria and the study that proves the device meets them:

1. A table of acceptance criteria and the reported device performance

The document doesn't explicitly state "acceptance criteria" in a codified format. However, based on the performance characteristics reported, we can infer the acceptance criteria implied by the study outcomes. For instance, the interference study specifies a recovery range of 90% to 110% for "no interference," which acts as an acceptance criterion. Similarly, the precision study reports specific CV% values, which are implicitly acceptable. For the method comparison, the slope, intercept, and correlation coefficient achieved are the performance metrics, implying that these values were considered acceptable for substantial equivalence.

Performance MetricAcceptance Criteria (Inferred)Reported Device Performance
Precision/Reproducibility(Implicitly acceptable CV%)Repeatability (CV%)
Sample Pool 1 (20.1 ng/mL)8.8%
Sample Pool 2 (50.6 ng/mL)4.5%
Sample Pool 3 (80.5 ng/mL)5.6%
Within-laboratory Precision (CV%)Within-laboratory (CV%)
Sample Pool 1 (20.1 ng/mL)9.1%
Sample Pool 2 (50.6 ng/mL)5.3%
Sample Pool 3 (80.5 ng/mL)5.9%
Linearity/Reportable RangeLinear across defined rangeDemonstrated across 13.0 ng/mL to 96.0 ng/mL
Limit of Quantitation (LoQ)LoQ ≤ 13.0 ng/mL9.9 ng/mL
Analytical SensitivityClaimed sensitivity is metClaimed at 13.0 ng/mL
Interference (Recovery %)90% to 110% of expected valueAll tested interferents (Hemoglobin, Bilirubin, Triglyceride, Cholesterol, Total protein, Biotin, HAMA, Rheumatoid Factor) within 90% to 110% recovery.
Cross-Reactivity(Implicitly low, except for known)Low for Vitamin D2, D3, 1,25-(OH)2-D2, 1,25-(OH)2-D3, 3-epi-25-(OH) D3, 25-(OH) D2, 25-(OH) D3. Paricalcitol showed 14.7-15.8% cross-reactivity.
Method Comparison (Slope)(Implicitly close to 1)1.069
Method Comparison (Y-intercept)(Implicitly close to 0)-0.182
Method Comparison (Correlation R)(Implicitly high, close to 1)0.971
Matrix ComparisonEquivalent results across matricesEquivalent results for serum, K3EDTA, lithium-heparin, and sodium citrate plasma.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Precision/Reproducibility: Not explicitly stated as "test set," but 3 serum pools were used. Each pool was assayed in duplicate twice per day for 20 days, totaling 80 measurements per level.
  • Linearity: 9 levels of serum samples tested in quadruplicate.
  • Limit of Quantitation (LoQ): Not specified (CLSI EP17-A2 protocol was used).
  • Interference Studies: Triplicate determinations for two concentrations of Vitamin D for each interferent tested.
  • Cross-Reactivity Studies: Triplicate measurements for each cross-reactant at two Vitamin D concentrations.
  • Method Comparison: 133 de-identified leftover samples. The provenance is implied to be from a CLIA-certified hospital laboratory (presumably in the US, given FDA submission). The samples were retrospective ("de-identified leftover samples").
  • Matrix Comparison: 36 paired samples.
  • Expected Values/Reference Range: 300 serum specimens from apparently healthy individuals residing in north, central, and southern regions of the US. Collected during summer and winter months. This is a prospective or cross-sectional study.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This device is an in-vitro diagnostic (IVD) assay for measuring 25-hydroxy vitamin D. The "ground truth" for such devices is typically established by reference methods or gold-standard assays, not by expert human readers.

  • For the method comparison study, the reference method was the Abbott ARCHITECT i 25-OH Vitamin D Assay (K110619) run on the Abbott ARCHITECT i System. This predicate device served as the "ground truth" for comparison.
  • For traceability, calibrators were confirmed by measurement on the same predicate device (ARCHITECT i 25-OH Vitamin D assay).

Therefore, there were no human experts establishing ground truth in the way a medical imaging study might use radiologists. The ground truth was based on a previously cleared and accepted analytical method.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. As this is an IVD assay where performance is compared against an established analytical method (predicate device), there is no human adjudication process involved in establishing ground truth for the test set as would be found in medical image analysis.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. The FREND™ Vitamin D Test System is an automated in-vitro diagnostic device. It does not involve human readers interpreting results in a way that an MRMC study comparing human performance with and without AI assistance would be relevant. The device provides a quantitative measurement directly.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, this entire submission focuses on the standalone performance of the FREND™ Vitamin D Test System. It is an automated system designed for use in clinical laboratories, which means it provides quantitative results directly from the sample without subjective human interpretation in the loop. The "FREND™ System" and "FREND™ AP System" are instrument-only components that analyze the cartridge.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The ground truth for the performance studies was predominantly based on comparison to a legally marketed predicate device, specifically the Abbott ARCHITECT i 25-OH Vitamin D Assay (K110619). For studies like linearity, precision, LoQ, interference, and cross-reactivity, the ground truth is defined by the known concentrations of analytes in the samples used (e.g., spiked samples, reference materials).

8. The sample size for the training set

The document does not explicitly mention a "training set" in the context of machine learning or AI models. This device is a fluorescent immunoassay, and its "training" or calibration is handled differently:

  • The device uses a lot-specific calibration curve generated by the manufacturer.
  • This curve is determined from values averaged from 6 replicates at each of four calibration levels.
  • This calibration information is uploaded to the FREND™ System via a Vitamin D Code-chip for each reagent lot.

Therefore, the concept of a large "training set" of patient data as understood in AI/ML is not directly applicable here. The calibration is a much smaller, controlled process.

9. How the ground truth for the training set was established

As noted above, "training set" in the AI/ML sense is not directly applicable. For the device's calibration (which serves a similar function to initial "training" for an assay):

  • The calibrators are prepared gravimetrically. This means their concentrations are established by precisely weighing the components.
  • These gravimetrically prepared calibrators are then confirmed by measurement on the ARCHITECT i 25-OH Vitamin D assay (K110619), the predicate device. This ensures they align with an established method.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized image of an eagle with three heads in profile, facing right. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter of the circle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

NANOENTEK USA, INC. C/O MAŬREEN GARNER NEW WORLD REGULATORY SOLUTIONS, INC 1983 HAZELWOOD ROAD TOMS RIVER NJ 08753

January 12, 2017

Re: K162754

Trade/Device Name: FREND™ Vitamin D Test System Regulation Number: 21 CFR 862.1825 Regulation Name: Vitamin D test system Regulatory Class: II Product Code: MRG Dated: November 23, 2016 Received: November 28, 2016

Dear Maureen Garner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Katherine Serrano -S

For:

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K162754

Device Name FREND™ Vitamin D Test System

Indications for Use (Describe)

The FREND™ Vitamin D test is a rapid indirect competitive fluorescent immunoassay designed for the quantitative measurement of 25-Hydroxy vitamin D and related hydroxylated metabolites in human serum and plasma (K;EDTA, lithium-heparin and citrate) specimens using the FREND™ System, and the FREND™ AP System. Measurements of total 25-hydroxy vitamin D and related hydroxylated metabolites are used to aid in the assessment of vitamin D sufficiency.

The FREND™ Vitamin D microfluidic flow cartridge is designed for use in the FREND™ System fluorescent immunoassay reader, and the FREND™ AP System. The FREND™ Vitamin D Test System is intended for use in clinical laboratories. For in vitro diagnostic use only. The test is not intended for use in point-of-care settings.

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

As required by the Safe Medical Devices Act (SMDA) of 1990 and in accordance with 21 CFR §807.92, a 510(k) summary is provided for K162754.

A. Applicant

Company Name:NanoEnTek, Inc.
Address:12F 5 Digital-ro, 26-gil, Guro-guSeoul 152-740, KOREA
Contact Person:Sunmi Han
Phone Number:+82-2-6220-7940
Facsimile Number:+82-2-6220-7721

B. 510(k) Preparer Information (Contact Person)

Company Name:New World Regulatory Solutions, Inc
Address:P.O. Box 5374
Toms River, New Jersey 08754, USA
Contact Person:Maureen Garner
Phone Number:(732) 779-7422
Facsimile Number:(732) 270-4829
Email:NWRSinc@gmail.com

Date Prepared:

January 6, 2017

C. Purpose for Submission:

New Analyte on FREND™ System

D. Measurand:

Vitamin D

E. Type of Test:

Quantitative, Fluorescence Immunoassay

F. Proprietary and Established Device Name:

FREND™ Vitamin D Test System

Regulatory Information: G.

Test Kit
Classification Name:Vitamin D Test System
Classification:Class II
Generic Name:Competitive Immunoassay, Vitamin D
Proprietary Name:FRENDTM Vitamin D
(reagent, cartridge assay)
Regulation Number:21 CFR §862.1825
Product Code:MRG
PanelClinical Chemistry (75)

Confidential and Proprietary Copyright © 2016 NanoEnTek

{4}------------------------------------------------

H. Intended Use:

    1. Intended use(s): See indications for use below:
    1. Indication(s) for use:

The FREND™ Vitamin D Test System is a rapid indirect competitive fluorescent immunoassay designed for the quantitative measurement of 25-hydroxy vitamin D and related hydroxylated metabolites in human serum and plasma (KsEDTA, lithiumheparin and citrate) specimens using the FREND™ System, and the FREND™ AP System. Measurements of total 25-hydroxy vitamin D and related hydroxylated metabolites are used to aid in the assessment of vitamin D sufficiency.

The FREND™ Vitamin D microfluidic flow cartridge is designed for use in the FREND™ System fluorescent immunoassay reader, and the FREND™ AP System. The FREND™ Vitamin D Test System is intended for use in clinical laboratories. For in vitro diagnostic use only. The test is not intended for use in point-of-care settings.

    1. Special conditions for use statement(s): For prescription use only.
    1. Special instrument requirements: NanoEnTek FREND™ System and FREND™ AP System

l. Device Description

The FREND™ Vitamin D Test is a rapid fluorescence immunoassay designed to quantify the concentration of total 25-hydroxy (OH) vitamin D and related hydroxylated metabolites in human serum and plasma (K3 EDTA, lithium-heparin and citrate) samples using the FREND™ System.

The FREND™ Vitamin D Test System is a competitive immunoassay with gold microparticles labeled with Vitamin D-specific monoclonal anti-Vitamin D-antibody (mouse), Vitamin D-biotin labeled with fluorescence nanoparticles and fluorescence detection by the FREND™ System.

The FREND™ Vitamin D Cartridge is a single-use disposable rapid lateral flow fluorescence immunoassay designed to quantify the concentration of total 25-OH vitamin D and related hydroxylated metabolites in human serum and plasma (K3 EDTA, lithiumheparin and citrate) samples. Each test cartridge contains reagents to perform one test and bears a bar code with test type and lot information. Each FREND™ Vitamin D Cartridge contains a test zone and a reference zone (internal control). The fluorescence measured in the test and reference zones corresponds to the amount of fluorescent conjugates of Vitamin D-biotin that is bound to each zone. Cartridges are provided in individually sealed pouches and must be stored between 2 and 8 °C.

In addition to the QC Code Chip there is a Vitamin D Code Chip. Lot information and calibration data are loaded into the software via the Vitamin D Code Chip provided with each reagent kit. Up to three different Vitamin D Code Chip lots can be loaded and used at the same time. Each reagent cartridge bears a barcode that the FREND™ System uses to identify the test type and lot number and links it to the appropriate analytical

Confidential and Proprietary Copyright © 2016 NanoEnTek

{5}------------------------------------------------

program and calibration information. The FREND™ System will not output results if the Vitamin D Code Chip lot does not match the lot of the cartridge used.

The Vitamin D Pretreatment Tube is a small single-use disposable plastic tube that contains Vitamin D gold nano-particle coniugation antibody. The sample is diluted in the dilution tube containing perfluorohexanoic acid and transferred to the pretreatment tube prior to being inserted into the FREND™ AP System to be mixed and then added to the Vitamin D Test Cartridge. The test cartridge is then inserted into the FREND™ System to be analyzed.

The FREND™ AP System is an "Advanced Preparing" device that performs automated pre-analytical steps of mixing, timed heating, and pipetting of the sample from the Gold Antibody Pretreatment tube into the FREND Vitamin D test cartridge.

The FREND™ System utilizes microfluidic technology and detects immune-complexes bound to Vitamin D. The test cartridge is placed on the warming platform of the FREND™ AP System. A 35 µL patient sample is added to the dilution tube and 70 µL of diluted sample is pipetted into the pretreatment tube containing monoclonal anti-Vitamin D antibody conjugated with gold nanoparticles(Step 1). The tube is then loaded into the FREND™ AP System (Step 2). The tube is mixed and loaded into the test cartridge automatically. In the AP System (Step 3), mixing, reaction, sample addition, hydration of the Vitamin D-biotin fluorescent bead conjugate, and migration along the cartridge channel takes approximately 15 minutes, after which the cartridge is read on the FREND™ System. During migration, the bound Vitamin D in the sample and the fluorescent bead conjugates of Vitamin D-biotin compete to form antigen-antibody complex in the test zone. Unbound Vitamin D-biotin fluorescent conjugates flow through and bind to the anti-Vitamin D antibody that is fixed on the surface in the reference zone. The fluorescent signals in the test and reference zones are measured, typically within 4 minutes.

Vitamin D quantification is based upon the ratio of the intensity of the test and reference zones. A lower ratio of fluorescence is indicative of a higher Vitamin D concentration; in other words, the magnitude of the fluorescent ratio is inversely proportional to the amount of Vitamin D in the sample.

The measuring range of the FREND™ Vitamin D Test System is 13.0 to 96.0 ng/mL. Results are determined via a lot-specific calibration curve which is generated by the manufacturer using a four-point calibration determined from values averaged from 6 replicates at each level. The established curve is uploaded to the FREND™ System via the Vitamin D Code-chip and is valid until the lot expiration date.

The FREND™ System is a portable, automated FREND™ cartridge reader. The FREND™ System is based on quantitative immunoassay technology capable of quantifying single or multiple analytes by measuring laser-induced fluorescence in a single-use disposable reagent cartridge. The test cartridge is a disposable plastic device that contains a port or opening (inlet) where the sample is applied. Once the sample is applied, it will mix with the reagents and travel towards the detection area via capillary action.

The FREND™ cartridge utilizes microfluidics lateral flow technology where the analyte of interest in the sample forms immune complexes while moving through the fluidics Confidential and Proprietary Copyright © 2016 NanoEnTek

{6}------------------------------------------------

pathway in the cartridge. The concentration of the analyte of interest in an unknown sample is calculated using the ratio of the fluorescent intensity of the test zone and the reference zone.

The FREND™ System is a bench top fluorescence reader containing a touch screen user interface. The System has a slot that accepts the FREND™ Vitamin D Test Cartridge (which contains the reagents and sample), and is programmed to analyze the Test when the sample has fully reacted with the on-board in-cartridge reagents. Results of the test are displayed on the screen and can be printed on an optional printer.

The FREND™ System software controls the graphical user interface, communication with hardware, database management and data analysis. The software also controls the functions of the mechanical components including the motor, laser, printer control and acquisition of data from the sensor. The user can set the time and date and enter patient ID through the graphic user interface. The user cannot make any changes to the software.

The FREND™ Vitamin D Test System includes the following in the reagent kit:

  • . 20 FREND™ Vitamin D Test Cartridges
  • 20 FREND™ Sample Dilution Tubes .
  • 20 FREND™ Vitamin D Gold Antibody Pretreatment Tubes .
  • 30 Disposable pipette tips
  • 1 FREND™ Vitamin D Code Chip .
  • . 1 FREND™ Vitamin D Package Insert

One cartridge contains:

  • . Monoclonal mouse anti- Vitamin D 1.6 ± 0.16 na
  • . 25-hydroxyvitamin D 136 ± 13.6 ng
  • . Fluorescence particles 2.4 ± 0.24 µg

One sample dilution tube contains:

  • . Perfluorohexanoic acid 9.5 ± 0.95 µg
    One Gold antibody pretreatment tube contains:

  • . Gold nano-particle conjugation antibody 7.0 ± 0.7 µg
    The FREND™ System (previously cleared in K124056 (FREND™ PSA), K131928 (FREND™ TSH), K152422 (FREND™ Free T4), and K153577 (FREND™ Testosterone) is not provided with the kit but it is required for the FREND™ Vitamin D Test cartridge.

{7}------------------------------------------------

Substantial Equivalence Information: J.

An overview of the similarities and differences between the FREND™ Vitamin D Test System and the predicate is provided in the table below:

Similarities
ItemFRENDTM Vitamin D Test SystemAbbott ARCHITECT i 25-OH Vitamin D Assay (K110619)
Intended UseThe FRENDTM Vitamin D test is a rapid indirect competitive fluorescent immunoassay designed for the quantitative measurement of 25-hydroxy vitamin D and related hydroxylated metabolites in human serum and plasma (K3 EDTA, lithium-heparin and citrate) specimens using the FRENDTM System, and the FRENDTM AP System. Measurements of total 25-hydroxy vitamin D and related hydroxylated metabolites are used to aid in the assessment of Vitamin D sufficiency.The FRENDTM Vitamin D microfluidic flow cartridge is designed for use in the FRENDTM System fluorescent immunoassay reader. and the FRENDTM AP System. The FRENDTM Vitamin D Test System is intended for use in clinical laboratories. For in vitro diagnostic use only. The test is not intended for use in point-of-care settings.The ARCHITECT 25-OH Vitamin D assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of 25-hydroxy vitamin D (25-OH Vitamin D) in human serum and plasma. The ARCHITECT 25-OH Vitamin D assay is to be used as an aid in the assessment of Vitamin D sufficiency.The ARCHITECT 25-OH Vitamin D Calibrators are for the calibration of the ARCHITECT i System when used for the quantitative determination of 25-hydroxy Vitamin D (25-OH Vitamin D) in human serum and plasma.The ARCHITECT 25-OH Vitamin D Controls are for the estimation of test precision and the detection of systematic analytical deviations of the ARCHITECT i System when used for the quantitative determination of 25-hydroxy Vitamin D (25-OH Vitamin D) in human serum and plasma.For in vitro diagnostic use only.
Sample TypeHuman serum and K3EDTA, lithium heparin, and citrate plasmaHuman serum, lithium heparin and sodium heparin
Analyte25-OH Vitamin DSame
Type of TestFluorescent immunoassay determination of Vitamin DChemiluminescent immunoassay determination of Vitamin D
Quality ControlInternal procedural/instrument quality controls; commercially-available external positive and negative assay controlsInternal procedural/instrument quality controls; external positive and negative assay controls cleared with test kit
Interpretation of ResultsInversely proportional Interpolation of fluorescence against a standard curveSame
Dynamic Range13.0 ng/mL to 96.0 ng/mLSame

{8}------------------------------------------------

Differences
ItemFREND™ Vitamin DTest SystemAbbott ARCHITECT i 25-OH Vitamin DAssay (K110619)
Sample Size35 µL for the dilution step and 70 µL forrunning the test60 µL (priority) or 150 µL (≤3h) for thefirst test plus 10 µL for each additionaltest from the same test cup
Test CartridgeDisposable single-use cartridgeNo single-use cartridge
RandomAccess /Degree ofAutomationNo random access, manualmanipulationRandom access, semi-automated

K. Performance Characteristics (if/when applicable)

    1. Analytical performance

a. Precision/Reproducibility

A single lot imprecision study was performed at the NanoEnTek laboratory as described in the CLSI protocol EP5-A3. Three serum pools with low, intermediate and high Vitamin D levels were assaved in duplicate twice per day for 20 days (80 total measurements per level). The results are summarized below:

SamplePoolMean Vitamin DConcentration(ng/mL)RepeatabilityBetween-runBetween-dayWithin-laboratory
SDCV%SDCV%SDCV%SDCV%
120.11.768.8%0.462.3%0.221.1%1.839.1%
250. 62.264.5%1.312.6%0.591.2%2.675.3%
380.54.535.6%1.131.4%0.740.9%4.725.9%
FREND™ Vitamin D Single Site Single Lot Precision
  • b. Linearity/assay reportable range:
    Linearity was established according to CLSI-EP6-A using nine levels of serum Vitamin D tested in quadruplicate. Linearity was demonstrated across the measuring range of the FREND™ Vitamin D (13.0 ng/mL ~ 96.0 ng/mL).

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

The calibrators are for factory calibration. The standards/calibrators are prepared gravimetrically and confirmed by measurement on the ARCHITECT i 25-OH Vitamin D assay (K110619). All calibration statistics and information have been electronically stored on the FREND™ Vitamin D Code chip included in each box of FREND™ Vitamin D cartridges. There is no need for calibration by the operator as the calibration information is in each code chip, which is lot-specific.

  • d. Stability
    Real-time stability testing for the Vitamin D reagent kit was performed according to CLSI EP25-A. Evaluation of Stability of In Vitro Diagnostic Reagents. Reagent stability studies based on procedures and criteria in the NanoEnTek quality system showed that the cartridges for FREND™ Vitamin D will meet performance acceptance criteria for one year from the date of manufacturer if stored refrigerated appropriately as directed.

Confidential and Proprietary Copyright © 2016 NanoEnTek

{9}------------------------------------------------

  • e. Detection Limit:
    The Limit of Quantitation (LoQ) for the FREND™ Vitamin D Test System was established at 9.9 ng/mL according to the CLSI EP17-A2 protocol. The analytical sensitivity of the FREND™ Vitamin D Test System is claimed at 13.0 ng/mL.

  • f. Analytical specificity:

Interference Studies

Interference was determined according to CLSI EP07-A2 using triplicate determinations and two concentration of Vitamin D. No interference was specified if recoveries were between 90% to 110% of the expected Vitamin D value. Results are summarized in the table below.

ClassInterferent (Concentration Tested)% RecoveryVitamin DLow% RecoveryVitamin DHigh
EndogenoussubstancesHemoglobin (500 mg/dL)107.098.4
Bilirubin, conjugated (30 mg/dL)95.695.7
Bilirubin, unconjugated (30 mg/dL)100.094.9
Triglyceride (1500 mg/dL)99.694.0
Cholesterol (500 mg/dL)93.096.1
Total protein (12 g/dL)103.498.8
Biotin (6 µg/mL)99.9102.7
HeterophilicAntibodiesHAMA (1000 ng/mL)97.4108.2
Rheumatoid Factor (536 IU/mL)101.496.8

Cross Reactivity

The following substances were evaluated in triplicate for potential cross-reactivity with the FREND™ Vitamin D Test System at two Vitamin D concentrations. Testing was done according to the CLSI protocol EP07-A2. No significant cross-reactivity was found, with the exception of Paricalcitol. Drugs containing Paricalcitol (e.g., Zemplar) will interfere with the assay.

Cross-reactantCross-reactant Concentration (ng/mL)% Cross-reactivity Vitamin D Low% Cross-reactivity Vitamin D High
Vitamin D2 (Ergocalciferol)5000.50.8
Vitamin D3 (Cholecalciferol)5000.50.7
1,25-(OH)2-Vitamin D21002.20.7
1,25-(OH)2-Vitamin D3 (Calcitriol)1006.91.9
3-epi-25-(OH) Vitamin D31002.12.0
25-(OH) Vitamin D22597.297.3
25-(OH) Vitamin D325107.392.2
Paricalcitol2515.814.7

% Cross-reactivity = 100 x ((Measured value - true value)/concentration of interferent)

g. Assay cut-off:

Not applicable

{10}------------------------------------------------

2. Comparison studies

  • a. Method comparison with predicate device:
    Comparison studies with 133 de-identified leftover samples were performed in a CLIA-certified hospital laboratory. The reference method was the Abbott ARCHITECT i 25-OH Vitamin D Assay (K110619) run on the Abbott ARCHITECT i System. The samples spanned the measuring range of the FREND Vitamin D Test System. All samples were assayed using sera split between the applicant device and the reference method.

Results from the FREND™ Vitamin D on the FREND™ System (y) were compared with the reference results (x) by Passing-Bablok regression analysis, giving a slope of 1.069, a y-intercept of -0.182, and a correlation (R) of 0.971.

  • b. Matrix comparison:
    The matrix comparison study was performed at the NanoEnTek laboratory according to CLSI EP14-A3. Vitamin D concentrations from 36 paired samples, covering the analytical measuring range, were measured in serum, K3EDTA, lithium-heparin and sodium citrate plasma samples with the FREND™ Vitamin D Test System, giving equivalent results.

3. Expected values/Reference range:

A study was conducted to establish a Reference Interval for the FREND™ Vitamin D Test System. It is also recommended that each laboratory establish its own reference range, which may be unique to the population it serves depending upon geographical season, patient, dietary, or environmental factors.

The study was conducted based on CLSI C28-A3. A total of 300 serum specimens from apparently healthy individuals residing in north, central and southern regions of the US were collected during summer and winter months and tested in the FREND™ Vitamin D Test System. The reference interval is provided below.

Reference IntervalLower reference limit90% CIUpper reference limit90% CI
<13.0 to 48.4<13.0 to <13.047.7 to 68.0

These limits should be considered as quidelines only. It is important for each laboratory to establish its own reference range, representative of its typical population.

L. Proposed Labeling

The labeling is sufficient and it satisfies the requirements of 21 CFR §809.10.

M. Conclusion

The submitted information in this premarket notification is complete and supports a substantial equivalence determination.

§ 862.1825 Vitamin D test system.

(a)
Identification. A vitamin D test system is a device intended for use in clinical laboratories for the quantitative determination of 25-hydroxyvitamin D (25-OH-D) and other hydroxylated metabolites of vitamin D in serum or plasma to be used in the assessment of vitamin D sufficiency.(b)
Classification. Class II (special controls). Vitamin D test systems must comply with the following special controls:(1) Labeling in conformance with 21 CFR 809.10 and
(2) Compliance with existing standards of the National Committee on Clinical Laboratory Standards.