K Number
K131928

Validate with FDA (Live)

Date Cleared
2014-03-14

(260 days)

Product Code
Regulation Number
862.1690
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

FREND™ TSH is designed for in vitro DIAGNOSTIC USE ONLY for the quantitative measurement of Thyroid Stimulating Hormone (thyrotropin or TSH) in human serum and lithium heparin plasma using the FREND™ system.

FREND™ TSH is indicated for use in clinical laboratories upon prescription by the attending physician as an aid to clinicians in the diagnosis of thyroid disease.

Device Description

The FREND™ System is a bench top fluorescence reader containing a simple computerized touchscreen user interface, easily manipulated to order tests, display results and operate the mechanical functions of the instrument. The specimen is added with a transfer pipette to the sample inlet of a single use cartridge by the operator, allowing the appropriate volume of sample (35 uL) to be delivered into the FREND™ TSH cartridge. The cartridge is then placed into the FREND™ system, which is programmed to begin analysis once the sample has reacted with the reagents. All reactions occur in the self-contained plastic cartridge and the reading is done in the cartridge as well. The FREND™ system has a slot that accepts the FREND™ TSH test cartridge containing the reagents and sample, and is programmed to analyze the test when the sample has fully reacted with the on-board in-cartridge reagents. Cartridges are loaded manually one by the operator. The reaction and analysis time is approximately 5 minutes. TSH quantitation is based on the ratio of fluorescence detected by the FREND™ System at the FREND™ TSH test and reference windows in the plastic cartridge compared to a standard curve stored in the TSH Code Chip that is included with each box of the device. A higher ratio of fluorescence is indicative of a higher TSH concentration. In other words, the magnitude of the fluorescent ratio is directly proportional to the amount of TSH in the sample.

AI/ML Overview

The FREND™ TSH device is an in vitro diagnostic test for the quantitative measurement of TSH in human serum and lithium heparin plasma. The study aimed to demonstrate its substantial equivalence to a legally marketed predicate device, the TOSOH ST AIA-PACK TSH (K972586), to support its 510(k) premarket notification.

1. Table of Acceptance Criteria and Reported Device Performance

Performance CharacteristicAcceptance Criteria (Implied by equivalence to predicate or clinical guidelines)Reported Device Performance (FREND™ TSH)
Accuracy (Linearity)Linearity over the measurement interval with non-linearity less than allowable (20% up to 1 mIU/L, then 10%).Linear from <0.06 mIU/L to 25.54 mIU/L. Slope = 0.977, y-Intercept = 0.17, Sw = 0.28, r = 0.999. Non-linearity less than allowable.
Accuracy (Comparability to Predicate)Difference in concentration between test and predicate device less than allowable difference (CLSI EP09-A2-IR, Section 7). Equivalent performance.Slope of 0.951 (95% CI: 0.940; 0.962), y-Intercept of 0.0266 (95% CI: -0.0258; 0.0790) for concentrations from 0.09 ~ 24.96 mIU/L. Strong correlation (R-squared = 0.9849). Thyroid status equivalent in 98.4% of cases.
PrecisionAcceptable repeatability, between-run, between-day, and within-laboratory precision throughout the measurement range.SD (mIU/L) and CV% reported for four TSH concentration levels across repeatability, between-run, between-day, and within-laboratory. E.g., for 0.496 mIU/L, within-laboratory CV% was 9.6%; for 23.763 mIU/L, within-laboratory CV% was 4.1%.
Specificity (Cross-Reactivity)No significant cross-reactivity with structurally similar hormones (LH, FSH, hCG) at specific concentrations.Insignificant interference (e.g., hCG at 200,000 mIU/mL resulted in 2 x 10^-8 % cross-reactivity at 0.49 mIU/L TSH).
Analytical Sensitivity (Limit of Detection - LoD)Similar to predicate device's sensitivity (0.03 mIU/L).Determined to be 0.06 mIU/L (using CLSI EP17-A).
Interference (Endogenous)Recovery within 90-110% of expected TSH concentration.Average TSH recovery: Hemoglobin (98.8%), Bilirubin (100.0%), Total Protein (100.4%), Triglyceride (101.6%), HAMA at 52.5 ng/mL (93.3% - higher concentrations showed interference), RF at 53.8 IU/mL (91.4% - higher concentrations showed interference).
Interference (Pharmaceutical)No significant interference (<10%) from common medications at tested concentrations.No significant interference found from tested drugs (Acetaminophen, Diltiazem, Erythromycin, Verapamil) at specified concentrations.
High Dose Hook EffectNo high dose hook effect within the tested range.No High Dose Hook effect seen in samples with TSH concentrations up to 2500 mIU/L.
Reagent StabilityStability meets storage and handling conditions.Cartridges good for at least one year refrigerated; six months at room temperature (22-28°C).
Matrix Study (Serum vs. Plasma)FREND™ TSH can be measured equally well in serum and lithium heparin plasma.Linear regression of serum (x) vs. lithium heparin plasma (y) yielded y = 0.995x - 0.320 (r = 0.992, r^2 = 0.985).
Reference IntervalsEstablish a single adult reference interval; no significant racial or gender differences.FREND™ TSH Reference Interval: 0.49 ~ 3.82 mIU/L (based on 385 healthy adults, 18-71 years old). Similar to predicate (0.47-4.09 mIU/L).

2. Sample Size Used for the Test Set and Data Provenance

  • Accuracy (Comparability Analysis): 438 serum samples. Data provenance is not explicitly stated as country of origin, but described as "obtained from subjects both apparently normal and with thyroid conditions as well as other undisclosed diseases and conditions." These samples were stored at -70°C for less than one year. This appears to be retrospective, using banked samples.
  • Accuracy (Linearity): Specimens from a high TSH concentration pool were diluted, implying a controlled laboratory experiment rather than patient samples.
  • Precision: Four clinical samples were used.
  • Specificity: Not specified, but involved samples with various concentrations of interferents.
  • Analytical Sensitivity: Not specified, but involved using the CLSI EP17-A protocol with appropriate samples for LoD determination.
  • Interference (Endogenous & Pharmaceutical): Not specified, but involved adding known concentrations of interferents to samples.
  • Matrix Study: 40 sample pairs (serum and lithium heparin plasma aliquots from the same individuals).
  • Reference Intervals: 385 apparently healthy ambulatory adults (195 males and 191 females, ages 18-71). Data provenance is not explicitly stated but implies a clinical study.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This document describes the validation of an in vitro diagnostic (IVD) device for quantitative measurement of TSH. The "ground truth" for such devices typically refers to the true concentration of the analyte (TSH) in samples. This is established through:

  • Predicate Device Comparison: The predicate device itself (TOSOH ST AIA-PACK TSH) serves as a gold standard or a highly reliable comparator. The agreement with this device forms a crucial part of the "ground truth" for demonstrating equivalence.
  • Reference Methods/Materials: For linearity, sensitivity, and interference studies, samples with known or precisely determined TSH concentrations are used, often prepared based on international standards or reference materials.
  • Clinical Characterization: For the reference interval study, "apparently healthy" individuals were selected, but the "truth" here is their health status, not a TSH concentration value. The TSH values themselves are then used to establish the reference range.

Therefore, the concept of "experts" to establish ground truth in the same way it applies to image-based diagnostic AI, for example, is not directly applicable here. The "truth" is based on established laboratory methods, validated predicate devices, and reference materials.

4. Adjudication Method for the Test Set

Not applicable in the context of an in vitro diagnostic immunoassay. The comparison is objective, based on quantitative measurements.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging or interpretation tasks where human readers make subjective judgments. For this quantitative immunoassay, the comparison is directly between numerical output of the device and a predicate device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the studies described are standalone performance evaluations of the FREND™ TSH device. The FREND™ system automatically performs the analysis once the sample is loaded into the cartridge by the operator, and provides a quantitative TSH result. The performance characteristics (accuracy, precision, sensitivity, etc.) are evaluated based on the device's output itself, without immediate human interpretation or modification of its quantitative result. The device is intended for use in clinical laboratories by "qualified technologists," but the performance metrics are intrinsic to the device's analytical capabilities.

7. The Type of Ground Truth Used

  • Comparability: The results obtained from the predicate device (TOSOH ST AIA-PACK TSH) were used as the primary ground truth for the comparability study.
  • Analytical Performance (Linearity, LoD, Specificity, Interference): Ground truth was established using known concentrations of TSH, dilution series from high concentration pools, and samples spiked with known amounts of interfering substances or cross-reactants. These are typically based on certified reference materials or established laboratory methods.
  • Reference Intervals: The TSH values from a cohort of apparently healthy individuals were used to statistically derive the reference interval.

8. The Sample Size for the Training Set

The document does not explicitly mention a "training set" in the context of machine learning, as this is an immunoassay device, not an AI/ML algorithm that undergoes explicit training in the same manner. The studies described are primarily for verification and validation of the device's analytical performance against established standards and a predicate device.

9. How the Ground Truth for the Training Set Was Established

As explained above, the concept of a "training set" for an AI/ML algorithm does not directly apply here. The device's underlying "knowledge" (e.g., standard curve, reaction kinetics) is established during its design and manufacturing based on chemical principles and calibration using reference standards. The studies described are validation studies to ensure this "knowledge" translates into accurate and reliable performance.

{0}------------------------------------------------

510(k) Summary: FREND™ TSH

MAR 1 4 2014

510(k) Number:

K131928:

Summary Preparation Date:

March 13, 2014

Submitted by:

NanoEnTek, Inc. 12F, 5 Digital-ro, 26-gil, Guro-gu Seoul, 152-740, Korea

Jimmy Chen Owner/Operator jim@nanoentek.com

Contact:

Judith E Loebel Director, Clinical and Regulatory Affairs DOCRO, Inc. (CRO assigned as contact with the FDA) 1 Jacks Hill Road Suite 1 A&B Oxford, CT 06478 USA

Proprietary Names:

FREND™ TSH

Common Names: Quantitative TSH Immunoassay

Regulatory Information:

Regulation section: 862.1690 Radioimmunoassay, Thyroid Stimulating Hormone

Classification: Class II .

{1}------------------------------------------------

Panel: Chemistry and Toxicology (DCTD))

Product Code(s): JLW Radioimmunoassay, Thyroid Stimulating Hormone

Other codes used by predicate devices: None

Predicate Devices: TOSOH ST AIA-PACK TSH (K972586)

Indications for Use:

510(k) Number: K131928

Device Name: FREND™ TSH

Indications for Use:

FREND™ TSH is designed for in vitro DIAGNOSTIC USE ONLY for the quantitative measurement of Thyroid Stimulating Hormone (thyrotropin or TSH) in human serum and lithium heparin plasma using the FREND™ system.

FREND™ TSH is indicated for use in clinical laboratories upon prescription by the attending physician as an aid to clinicians in the diagnosis of thyroid disease.

Technological Characteristics:

There are no technological characteristics of FREND™ TSH that are different or with which the agency is unfamiliar. FREND™ TSH, performed using the FREND™ system. is an in vitro diagnostic test for the quantitative measurement of TSH in serum and lithium heparin plasma samples. The FREND™TSH is performed completely within the single-use plastic cartridge once serum or plasma has been placed in the sample well and is for use upon prescription by the attending physician in clinical laboratories with qualified technologists. FREND™ TSH on the FREND™ system has been shown to yield TSH results that are equivalent to those obtained with other previously cleared FDA in vitro diagnostic devices measuring TSH quantitatively in serum and/or plasma.

The FREND™ System is a bench top fluorescence reader containing a simple computerized touchscreen user interface, easily manipulated to order tests, display results and operate the mechanical functions of the instrument. The specimen is

{2}------------------------------------------------

added with a transfer pipette to the sample inlet of a single use cartridge by the operator, allowing the appropriate volume of sample (35 uL) to be delivered into the FREND™ TSH cartridge. The cartridge is then placed into the FREND™ system, which is programmed to begin analysis once the sample has reacted with the reagents. All reactions occur in the self-contained plastic cartridge and the reading is done in the cartridge as well. The FREND™ system has a slot that accepts the FREND™ TSH test cartridge containing the reagents and sample, and is programmed to analyze the test when the sample has fully reacted with the on-board in-cartridge reagents. Cartridges are loaded manually one by the operator. The reaction and analysis time is approximately 5 minutes. TSH quantitation is based on the ratio of fluorescence detected by the FREND™ System at the FREND™ TSH test and reference windows in the plastic cartridge compared to a standard curve stored in the TSH Code Chip that is included with each box of the device. A higher ratio of fluorescence is indicative of a higher TSH concentration. In other words, the magnitude of the fluorescent ratio is directly proportional to the amount of TSH in the sample. A high-level schematic and process diagram of the FREND™ system are included in the User Manual. Results of the test are displayed on the screen and can be printed on an optional printer.

Expected Values

As with every clinical diagnostic test, a reference interval corresponding to the characteristics of the population being tested should be determined by each laboratory. Historically, it has been shown that there are neither racial differences nor gender differences in the reference interval for TSH so creating a single adult reference interval is reasonable and justified per literature.

During a clinical study run to support the FREND™ TSH substantial equivalence to a marketed product with the same indication, TSH measurements were performed on the serum of 385 apparently healthy ambulatory adults (195 males and 191 females ages 18 - 71) who stated they had no known thyroid conditions. All samples were assayed in singlicate on the FREND™ TSH and the predicate device. A single value, determined as an outlier in both the test and the predicate devices, was removed from the data set after which a non-parametric reference interval encompassing the central 95% of the results was determined. Male and female results were separately analyzed with no significant difference found in the calculated reference intervals. Therefore, a single reference interval has been determined.

FREND™ TSH Reference Interval: 0.49 ~ 3.82 mlU/L

With 90% confidence limits the following ranges were estimated:

Lower Reference Interval Range0.42 ~ 0.59 mIU/L
Upper Reference Interval Range3.24 ~ 4.30 mlU/L

{3}------------------------------------------------

As in all in vitro diagnostic testing, a TSH result generated using the FREND™ TSH on the FREND™ system should be interpreted in the light of other clinical findings and diagnostic procedures. Any TSH results not correlating with the clinical condition should be repeated and other testing performed to clarify the situation.

Performance Characteristics

Performance characteristics were evaluated for FREND™ TSH as follows:

1) Accuracy

1a) Dilution Linearity - Specimens from a high TSH concentration pool were diluted with a TSH depleted serum pool according to the CLSI EP06-A document. Linearity was demonstrated from <0.06 mIU/L to 25.54 mIU/L (Slope = 0.977, y-Intercept = 0.17, Sw = 0.28, r = 0.999). Correlation with the expected linearity was excellent showing less than the allowable non-linearity. Performance requirement was verified over the measurement interval.

FREND™ TSH Linearity
% LevelRep. 1Rep. 2Rep. 3MeanLinear Fit
00.010.010.030.0170.173
102.852.782.782.8032.714
205.615.375.475.4835.255
308.477.808.538.2677.796
409.609.4310.629.88310.337
5012.213.1712.5112.62712.878
6015.6315.6414.2715.18015.418
7018.7217.4317.6317.92717.959
8021.8820.0920.3720.78020.500
9022.3823.3723.6823.14323.041
10025.9125.3225.4025.54325.582

FREND™ TSH is linear from 0.06 ~ 25.0 mlU/L

Figure 1

Non-linearity is less than allowable non-linearity: 20% up to 1 mlU/L then 10%.

4

{4}------------------------------------------------

Comparative Analysis - 438 serum samples obtained from subjects both 1b) apparently normal and with thyroid conditions as well as other undisclosed diseases and conditions and stored at -70° C under monitored conditions for less than one year, were analyzed using both the FREND™ TSH and another commercially available TSH fluorescent immunoassay. Results generated using the FREND™ TSH on the FREND™ System (y) were compared to those obtained using a previously FDA cleared TSH assay (x). Results of this study are shown below:

Slope: 0.951 (95% Cl: 0.940; 0.962) v-Intercept: 0.0266 (95% CI: -0.0258; 0.0790) Sylx: 0.475 Number of Samples: 438 Measuring Range: 0.09 ~ 24.96 mIU/L

Comparability using CLSI guideline EP09-A2-IR, Section 7, shows that the difference in concentration between that measured by the test device and that expected (predicate device) is less than the allowable difference and that the two methods compare favorably.

Image /page/4/Figure/3 description: This image is a scatter plot with a linear regression line. The plot contains 438 samples, and the regression equation is y = 0.9509x + 0.0267, with an R-squared value of 0.9849. The x and y axes range from 0.00 to 30.00, and the data points are clustered closely around the regression line, indicating a strong positive correlation.

Figure 2 Comparability to Predicate Device

{5}------------------------------------------------

2) Precision

Precision was determined as described in the CLSI protocol EP05-A2. Four clinical samples were assayed in replicates of two at two separate times per day for twenty days using a single lot of FREND™ TSH cartridge. The findings follow showing repeatability, between-run, between-day and withinlaboratory precision data.

Mean TSHRepeatabilityBetween-runBetween-dayWithin-laboratory
SampleConc. (mIU/L)SD (mIU/L)CV%SD (mIU/L)CV%SD (mIU/L)CV%SD (mIU/L)CV%
10.4960.0438.60.0122.40.0173.40.0479.6
25.9480.3535.90.0821.40.0310.50.3646.1
311.9890.5554.60.3753.10.1561.30.6885.7
423.7630.8463.60.4782.00.0000.00.9724.1

Figure 3 FREND™ Single Site Single Lot Precision

3) Specificity

The a- subunits of luteinizing hormone (LH), follicle stimulating hormone (FSH), human chorionic qonadotropin (hCG) and thvroid stimulating hormone (TSH) are all very similar though their ß- subunits are not. Because of the structural similarities, the specificity of the FREND™ TSH must be examined in the presence of large amounts of these possible cross-reactants. The following substances were evaluated for potential cross-reactivity with the FREND™ TSH at the concentrations indicated below in Figure 4. Testing was done according to the CLSI protocol EP07-A. No significant cross-reactivity was found.

{6}------------------------------------------------

SampleTSHConc.(mIU/L)InterferentMaterialadded% CrossReactivity
0.49hCG200,000mIU/mL2 x $10^{-8}$
0.55LH500mIU/mL2 x $10^{-4}$
0.55FSH500mIU/mL-5 x $10^{-6}$
6.22hCG200,000mIU/mL3 x $10^{-7}$
6.06LH500mIU/mL3 x $10^{-5}$
6.06FSH500mIU/mL1 x $10^{-4}$

Figure 4 Specificity of FREND™ TSH

4) Analytical Sensitivity

The Limit of Detection (LoD) for the FREND™ TSH was measured using the CLSI EP17-A protocol. The analytical sensitivity of the FREND™ TSH was determined to be 0.06 mlU/L.

5) Interference

Interference is defined in this instance as recovery outside of 10% of the known specimen mean concentration. In other words, recovery from 90% to 110% of the expected TSH is considered acceptable performance. The interference studies were performed using the recommendations in the CLSI EP07-A protocol. Results were:

Endogenous Interferents

  • Added hemoglobin (up to 500 mg/dL) does not interfere with the assay. . Average TSH recovery was 98.8%.
  • Added conjugated bilirubin (up to 20 mg/dL) does not interfere with the assay. . Average recovery was 100.0%.
  • . Added total protein up to 12.0 g/dL does not interfere with the assay. Average recovery was 100.4%.
  • . Added triglyceride up to 3 g/dL does not interfere with this assay. Average recovery was 101.6%.

{7}------------------------------------------------

  • Added HAMA to 52.5 ng/mL did not interfere with the assay. Average recovery . was 93.3%. Higher concentrations of HAMA exhibited interference> 10%.
  • RF at 53.8 IU/mL did not interfere with the assay. Average recovery was . 91.4%. Higher concentration of RF did show interference > 10%.

Pharmaceutical Interferents

The following common medications were tested for interference with the FREND™ TSH. The testing showed no significant interference (<10%) from the tested drugs at the listed concentration that would affect the interpretation of a TSH result as assayed on the FREND™ TSH:

Tested DrugConcentration Tested (µM)
Acetaminophen1324
Diltiazem15
Erythromycin81.6
Verapamil4.4

Figure 5

Interference from Common Medications

6) High Dose Hook Effect Testing (Prozone Detection)

No High Dose Hook effect was seen in samples with a TSH concentration as high as 2500 mIU/L.

7) Reagent Stability Studies

Reagent stability studies based on procedures and criteria in the NanoEnTek quality system showed that the cartridges for FREND™ TSH are good for at least one year from the date of manufacturer if stored refrigerated appropriately as directed. Stability of the cartridges at room temperature (22~28°C) was found to be six months.

8) Matrix Study

TSH concentrations in 40 sample pairs, each with serum and lithium heparin plasma aliquots, were measured using the FREND™ TSH. Linear regression analysis of serum results (x) compared to lithium heparin plasma results (y) yielded the following equation (r = 0.992, i2 = 0.985), indicating FREND TSH can be measured equally well in serum and lithium heparin plasma:

$$\mathbf{y} = \mathbf{0.995x} - \mathbf{0.320}$$

{8}------------------------------------------------

CLIA '88 - Complexity Categorization.

NanoEnTek has a previous reagent cartridge assay cleared by 510(k) in the US and this has been assigned a moderate complexity classification. The predicate TSH device is also marketed as a MODERATE complexity device. The logical assumption is that the FREND™ TSH does not require a high complexity classification and there is a valid argument because of the ease of use for assigning the FREND™ TSH a MODERATE COMPLEXITY categorization.

Substantial Equivalence

NanoEnTek has developed the FREND™ TSH for the FREND™ system and completed the necessary analytical and clinical validation studies to demonstrate the performance characteristics of the test for use as supporting data for this 510(k) premarket notification and for its claim of substantial equivalence.

In conclusion, the performance study results summarized below support the claim of substantial equivalence of the FREND™ TSH to the Predicate Device. Detailed data is filed in the 510(k) documentation.

Figure 6 Comparative Characteristics: FREND™ TSH and Predicate Device

DevicePredicate
ItemsFREND™ TSHST AIA-PACKTSH
510(k) NumberK131928K972586
Regulations862.1690 Radioimmunoassay, Thyroid Stimulating HormoneSame
Product CodeJLWSame
Device ClassIISame
Intended UseQuantitative measurement of TSHSame
Indications for UseTo quantitatively measure TSH in serum, heparin plasmaSame
Warnings andPrecautionsFor use in clinical laboratories upon prescription by the attendingphysicianSame
Contra-indicationsShould not be used to measure TSH in patients who havereceived therapeutic doses of mouse monoclonal antibodytherapeutics. Not to be used for newborn screening.Same
Similar AssaySensitivity (LoD)Measured at 0.06 mIU/LMeasured at0.03 mIU/L

Similarities

{9}------------------------------------------------

ItemsDevicePredicate
FREND™ TSHST AIA-PACKTSH
Interference fromDrugsNo interference found in the drugs that were tested at theconcentrations they were tested.Same
Test VesselDisposable single-use reaction vesselSame
Sample TypeSerum and lithium heparin plasmaSame
Sample PrepPrepare serum/plasma from whole bloodSame
Quality controlInternal procedural/instrument quality controls; External QC atnormal and elevated levels.Same
Interpretation ofResultsComparing fluorescence for sample against a standardcalibration curveSame
Reaction TypeAntibody/antigen complexesSame
Type of TestFluorescent immunoassay detecting TSHSame
Sample Carry-overNone detected - single use cartridgeSame- singleuse cup
EndogenousInterferenceNone by Bilirubin, Triglyceride, Cholesterol, Total Proteindetected in testing performedSame
End UsersTechnologists working in a clinical laboratorySame
ResultComparabilityResults across the range of 0.09 – 24.96 mIU/L equivalentbetween ST AIA-PACK TSH and FREND™ TSHSame
Reference Int.0.49 – 3.82 mIU/L0.47-4.09mIU/L

Figure 7

Comparative Characteristics: FREND™ TSH and Predicate Device

Differences

ItemsDevicePredicate
FREND™ TSHST AIA-PACK PA
Random Access/Degree of AutomationNo random access, mainly manualmanipulationRandom access, semi-automated
Test ThroughputSingle Test 5 minutes to result.Single test 18 minutes;50 tests 68 minutes.
Calibration StandardWHO International Standard Hormone,hormone for immunoassay, NIBSC code:81/565WHO 2nd InternationalReference Preparation 80/558(1983)

{10}------------------------------------------------

Figure 8 - Summary

Performance Validation Studies Supporting Substantial Equivalence

No.StudySummary of Results
1.Analytical Sensitivity/Limit of Detection(LoD)FRENDTM TSH Limit of Detection (LoD) wasdetermined to be 0.06 mIU/L. Sensitivity on thepredicate device is listed as being 0.03 mIU/L.Similar sensitivity limits.
2.Interfering SubstancesThe FRENDTM TSH was tested for interferencewith endogenous materials such as hemoglobin,bilirubin, triglycerides, and total protein, at levelsequivalent to those tested by the predicatedevice. Neither assay showed clinically significantinterference.
3.SpecificitySpecificity testing with biological substances withsimilar structures (LH, FSH, βHCG) wasperformed and showed insignificant interferencewith both the test and predicate devices.
4.Carry-overBoth the test system and the predicate systemuse disposable pipette tips and an individualreaction vessel to prevent possible carryover.
5.Reference IntervalsIndependent Reference Intervals were calculatedfor both the test and predicate devices using thesame pool of apparently healthy subjects. TheReference Interval for both methods wasdetermined to be approximately 0.5 – 4.0 mIU/L
6.Comparability438 samples were analyzed on both devices –the TOSO ST AIA-PACKTM TSH and theFRENDTM TSH. Descriptive statistics for thecomparability show equivalent performance.Thyroid status was equivalent between methods98.4% of the time in the study.
7.PrecisionPrecision Material testing demonstratedacceptable performance throughout themeasurement range of the assay for both test andpredicate devices.

Overall Conclusions from the Document

The information and data in this 510(k) document demonstrate that the FREND™ TSH is an accurate, reliable test that correlates well with current cleared methods for the quantitation of serum and lithium heparin plasma

{11}------------------------------------------------

TSH. The contents of this submission demonstrate that the FREND™ TSH on the FREND™ system is substantially equivalent to its predicate device and, therefore, safe and effective for its intended use, measuring quantitatively TSH in human serum and lithium heparin plasma.

Substantial Equivalence Conclusions

The FREND™ TSH is as safe and effective as the "Predicate Device", the TOSOH ST AIA-PACK™ TSH assay. FREND™ TSH has a similar Intended Use and Indications for Use: the quantitative measurement of TSH in serum and heparinized plasma, similar technological and performance characteristics, and principles of operation as its predicate device. The differences between the FREND™ TSH and its predicate device raise no new issues of safety or effectiveness. Performance data, analytical and clinical, demonstrate that the FREND™ TSH is as safe and effective as the "Predicate Device". Thus, the FREND™ TSH must be found to be substantially equivalent to the TOSOH ST AIA-PACK TSH.

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an eagle or bird in flight, composed of three curved lines.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 14, 2014

NANOEN TEK USA, INC. C/O JUDITH LOEBEL DOCRO 1 JACKS HILL ROAD, SUITES A & B OXFORD CT 06478

Re: K131928

Trade/Device Name: FRENDTM TSH Regulation Number: 21 CFR 862.1690 Regulation Name: Thyroid stimulating hormone test system Regulatory Class: II Product Code: JLW Dated: February 4, 2014 Received: February 7, 2014

Dear Ms. Loebel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading,

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{13}------------------------------------------------

Page 2-Ms. Loebel

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resources/or You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Courtney H. Lias -S

Courtney H. Lias. Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{14}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.

510(k) Number (if known) K131928

Device Name FREND™ TSH

Indications for Use (Describe)

FREND™ TSH is designed for in viro DIAGNOSTIC USE ONLY for the quantitative measurement of Thyroid Stimulaing Hormone (thyrotropin or TSH) in human serum and lithium heparin plasma using the FREND™ system.

FREND™ TSH is indicated for use in clinical laboratories upon prescription as an aid to clinicians in the diagnosis of thyroid discase.

. •

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

中国际娱乐官方下载 FOR FOR FOR FOR FDA USE ONLY SE MARK TO THE :: Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Image /page/14/Picture/13 description: The image shows the text "Yung W.Ochan -S" in a bold, sans-serif font. The text is horizontally oriented and appears to be a title or heading. The letters are black against a white background, creating a high contrast.

FORM FDA 3881 (1/14)

{15}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 862.1690 Thyroid stimulating hormone test system.

(a)
Identification. A thyroid stimulating hormone test system is a device intended to measure thyroid stimulating hormone, also known as thyrotrophin and thyrotrophic hormone, in serum and plasma. Measurements of thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders.(b)
Classification. Class II.