The Proficient™ Posterior Cervical Spine System is intended to immobilize and stabilize the spine as an adjunct to fusion for cervical (C2-C7) and thoracic (T1-T3) spinal segments that have been affected by the following acute or chronic instabilities: traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g. pseudoarthrosis); tumors involving the cervical/thoracic spine; and degenerative diseases, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies and degenerative disease of the facets with instability. The Proficient™ Posterior Cervical Spine System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time in patients with advanced stage tumors involving the cervical spine whose life expectancy is insufficient to permit achievement of fusion. In order to achieve additional levels of fixation, the Proficient™ Posterior Cervical Spine System may be connected to the CapSure® Spine System or the Sniper® Spine System using the Proficient™ transition rods.
The Proficient™ Posterior Cervical Spine System consists of a selection of non-sterile, single use polyaxial screws, set screws, rods, and cross-connector components manufactured from titanium (ASTM F136 and ASTM F67) and cobalt chrome alloys (ASTM F1537 and ASTM F1058). The surgeon attaches the rod, screw, and cross connectors to the cervicothoracic region of the spine in order to stabilize the spine during fusion of vertebral bodies.
This document is a 510(k) premarket notification for the Proficient™ Posterior Cervical Spine System. It describes a medical device, not an AI/ML-driven device or study with acceptance criteria related to a diagnostic or predictive algorithm. Therefore, the requested information about acceptance criteria, study details, sample sizes, ground truth establishment, expert qualifications, adjudication methods, and MRMC studies is not applicable to this document.
The performance data listed in Section 8 of the 510(k) summary are for mechanical testing of the spinal implant device, not for the performance of an AI system. These tests are:
- Static and dynamic compression bending (per ASTM F1717)
- Static and dynamic torsion (per ASTM F1717)
- Axial grip (per ASTM 1798)
These are standard engineering tests to evaluate the physical properties and durability of the spinal system and are used to demonstrate its substantial equivalence to predicate devices, rather than meeting acceptance criteria for an AI system's performance.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other, with flowing lines extending downward from the bottom profile.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 7, 2017
Spine Wave, Inc. Ms. Sanja Jahr Regulatory Affairs Specialist Three Enterprise Drive, Suite 210 Shelton, Connecticut 06484
Re: K162639
Trade/Device Name: Proficient™ Posterior Cervical Spine System Regulatory Class: Unclassified Product Code: NKG Dated: January 5, 2017 Received: January 6, 2017
Dear Ms. Jahr:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K162639
Device Name
Proficient™ Posterior Cervical Spine System
Indications for Use (Describe)
The Proficient™ Posterior Cervical Spine System is intended to immobilize and stabilize the spine as an adjunct to fusion for cervical (C2-C7) and thoracic (T1-T3) spinal segments that have been affected by the following acute or chronic instabilities: traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g. pseudoarthrosis); tumors involving the cervical/thoracic spine; and degenerative diseases, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies and degenerative disease of the facets with instability. The Proficient™ Cervical Spine System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time in patients with advanced stage tumors involving the cervical spine whose life expectancy is insufficient to permit achievement of fusion. In order to achieve additional levels of fixation, the Proficient™ Posterior Cervical Spine System may be connected to the CapSure® Spine System or the Sniper® Spine System using the Proficient™ transition rods.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the words "Spine Wave" in a bold, sans-serif font. The words are arranged horizontally, with "Spine" on the left and "Wave" on the right. A blue wave design is located under the word "Wave" and extends to the left.
510(k) Summary Proficient™ Posterior Cervical Spine System
1. Submitter Information
| Submitter: | Spine Wave, Inc. |
|---|---|
| Address: | Three Enterprise DriveSuite 210Shelton, CT 06484 |
| Telephone: | 203-712-1870 |
| Telefax: | 203-944-9493 |
| Contact: | Sanja Jahr |
| Date Prepared: | January 5, 2016 |
2. Device Information
| Trade Name: | Proficient™ Posterior Cervical Spine System |
|---|---|
| Common Name: | Posterior Cervico-Thoracic Fixation system |
| Classification: | Unclassified, Pre-Amendment |
| Classification Name: | Orthosis, Cervical Pedicle Screw Spinal Fixation |
| Product Code: | NKG |
3. Purpose of Submission
The purpose of this submission is to gain clearance for a new posterior cervico-thoracic fixation system.
4. Predicate Device Information
The Proficient™ Posterior Cervical Spine System described in this submission is substantially equivalent to the following predicates:
| Primary Predicate Device | Manufacturer | 510(k) No. |
|---|---|---|
| Synapse™ OCT System | DePuy Synthes, Inc. | K142838 |
| Reference Predicate Devices | Manufacturer | 510(k) No. |
| neon3 TM | Ulrich Medical | K150650 |
| Zero-PTM VA | Synthes Spine, Inc. | K112068 |
| Mountaineer® OCT Spinal System | Depuy Synthes Inc. | K151885 |
| Virage® OCT Spinal Fixation System | Zimmer Spine, Inc | K153631 |
{4}------------------------------------------------
5. Device Description
The Proficient™ Posterior Cervical Spine System consists of a selection of non-sterile, single use polyaxial screws, set screws, rods, and cross-connector components manufactured from titanium (ASTM F136 and ASTM F67) and cobalt chrome alloys (ASTM F1537 and ASTM F1058). The surgeon attaches the rod, screw, and cross connectors to the cervicothoracic region of the spine in order to stabilize the spine during fusion of vertebral bodies.
6. Indications for Use
The Proficient™ Posterior Cervical Spine System is intended to immobilize and stabilize the spine as an adjunct to fusion for cervical (C2-C7) and thoracic (T1-T3) spinal segments that have been affected by the following acute or chronic instabilities: traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g. pseudoarthrosis); tumors involving the cervical/thoracic spine; and degenerative diseases, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies and degenerative disease of the facets with instability. The Proficient™ Posterior Cervical Spine System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time in patients with advanced stage tumors involving the cervical spine whose life expectancy is insufficient to permit achievement of fusion. In order to achieve additional levels of fixation, the Proficient™ Posterior Cervical Spine System may be connected to the CapSure® Spine System or the Sniper® Spine System using the Proficient™ transition rods.
7. Comparison of Technological Characteristics
The substantial equivalence of the Proficient™ Posterior Cervical Spine System to the predicates is demonstrated by similarity in indications for use, materials and performance.
8. Performance Data
Spine Wave performed the following testing to demonstrate the Proficient™ Posterior Cervical Spine System is substantially equivalent to its predicate:
- Static and dynamic compression bending (per ASTM F1717)
- Static and dynamic torsion (per ASTM F1717) ●
- Axial grip (per ASTM 1798) o
9. Conclusion
Based on the indications for use, technological characteristics, performance testing and comparison to the predicates, the Proficient™ Posterior Cervical Spine System has been shown to be substantially equivalent to the predicate devices identified in this submission and does not present any new issues of safety or effectiveness.
N/A