K Number
K162344

Validate with FDA (Live)

Device Name
INNOLUX RIC
Date Cleared
2016-12-20

(120 days)

Product Code
Regulation Number
892.1680
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Wireless/Wired INNOLUX RIC flat panel detector is intended to capture for display radiographic images of human anatomy. It is intended for use in general projection radiographic applications including pediatric and neonatal exams wherever conventional film/screen or CR systems may be used. The INNOLUX RIC is not intended for mammography, fluoroscopy, tomography, and angiography applications.

Device Description

The INNOLUX RIC Flat Panel Detector is a portable digital detector that integrated with an operating PC and an X-ray generator to acquire and digitize x-ray exposures from, standard radiographic systems and transfer for radiography diagnostic. The INNOLUX RIC is designed to be used in any environment that would typically use a radiographic cassette for examinations of adults, pediatrics and neonates. The detector models support both wireless and wired data communication. Detectors can be placed in a wall bucky for upright exams, a table bucky for recumbent exams, or removed from the bucky for non-grid or free cassette exams.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the INNOLUX RIC flat panel detector. The document asserts the device's substantial equivalence to predicate devices based on non-clinical performance data.

Here's an analysis of the provided information concerning acceptance criteria and study details:

1. A table of acceptance criteria and the reported device performance

The document provides a comparison table between the subject device (INNOLUX RIC) and its predicate device (K142003, FDR D-EVO II Flat Panel Detector System) for various image quality and physical characteristics. The "acceptance criteria" appear to be implicit in the "Same as K142003" statements and the direct numerical comparisons, implying that the INNOLUX RIC's performance should be equivalent to or within acceptable tolerance of the predicate device.

CharacteristicAcceptance Criteria (Implied from Predicate)Reported Device Performance (INNOLUX RIC)
ScintillatorGd2O2S:Tb for -SE models; CsI:Tl for -SE modelsRIC 35G, RIC 43G Gd2O2S:Tb; RIC 35C, RIC 43C CsI:Tl (Same as K142003)
X-ray ConversionIndirect conversion (a-Si)Indirect conversion (a-Si) (Same as K142003)
Detector CordWired / WirelessWired / Wireless (Same as K142003)
Detector WeightDR-ID1201SE,DR-ID1211SE : Approx. 5.8 lbs; DR-ID1202SE,DR-ID1212SE : Approx. 7.1 lbsRIC 35G, RIC 35C: Approx. 6.5 lbs; RIC 43G, RIC 43C: Approx. 8.0 lbs
Exposure Size/Active Area (inch)13.8x16.8 for -SE models; 16.8x16.7 for -SE modelsRIC 35G, RIC 35C: 13.8x16.8; RIC 43G, RIC 43C: 16.8x16.7 (Same as K142003)
Dimensions (inch)18.1(W) x 15.1(D) x 0.6(H) for -SE models; 18.1(W) x 18.1(D) x 0.6(H) for -SE modelsRIC 35G, RIC 35C: 18.1(W) x 15.1(D) x 0.6(H); RIC 43G, RIC 43C: 18.1(W) x 18.1(D) x 0.6(H) (Same as K142003)
Pixel Size150 μm150 μm (Same as K142003)
Acquisition Bit Depth16 bit16 bit (Same as K142003)
DQE (RQA5, 1 lp/mm, 1mR) – detector alone, without tabletopDR-ID1201SE,DR-ID1202SE: 30% (±10%); DR-ID1211SE,DR-ID1212SE: 54% (Csl) (±10%)RIC 35G,RIC 43G: 31% (±10%); RIC 35C,RIC 43C: 54% (Csl) (±10%)
MTF (RQA5, 2 lp/mm)DR-ID1201SE,DR-ID1202SE: 42% (High mode) (±10%); DR-ID1211SE,DR-ID1212SE: 54% (High mode) (±10%)RIC 35G,RIC 43G: 42% (±10%); RIC 35C,RIC 43C: 54% (±10%)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document states, "No clinical study has been performed. The substantial equivalence has been demonstrated by non-clinical studies." Therefore, there is no clinical test set, sample size, or data provenance information related to human subjects. The non-clinical studies would involve physical devices.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Since no clinical study was performed, no experts were used to establish ground truth for a clinical test set.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable, as no clinical study was performed.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a flat panel detector, not an AI-powered diagnostic tool, and no clinical human reader study was conducted.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

The document specifies "non-clinical studies" were performed to evaluate the image quality and other characteristics of the device. This implies rigorous testing of the device itself (including its underlying software/algorithms for image acquisition and processing) in a standalone manner, without human interpretation for diagnostic purposes. The image quality evaluation concluded that the INNOLUX RIC detectors are "substantially equivalent to that of the predicate device." While not explicitly called "standalone algorithm performance," it aligns with evaluating the device's inherent technical capabilities.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the non-clinical studies, the "ground truth" would be established by objective physical measurements and comparisons against established engineering and performance standards (e.g., DQE, MTF measurements, dimensional specifications) and the performance of the predicate device. This is not clinical ground truth.

8. The sample size for the training set

Not applicable. This device is a hardware component (flat panel detector) with associated software for image acquisition and processing, not a machine learning or AI algorithm that requires a separate "training set" in the conventional sense. The "training" in this context would refer to the engineering design and optimization process.

9. How the ground truth for the training set was established

Not applicable, as there is no training set in the AI sense. The "ground truth" for the device's design and engineering would be based on engineering specifications, physical laws, and performance targets derived from the predicate device and relevant industry standards.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 20, 2016

INNOLUX Corporation % Mr. Andy Kuo Director No. 3, Sec. 1, Huansi Rd. Tainan City, Southern Taiwan Science Park, 74147 TAIWAN (R.O.C.)

Re: K162344 Trade/Device Name: INNOLUX RIC Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: November 23, 2016 Received: November 29, 2016

Dear Mr. Kuo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Robert Ochs

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K162344

Device Name

INNOLUX RIC

Indications for Use (Describe)

The Wireless/Wired INNOLUX RIC flat panel detector is intended to capture for display radiographic images of human anatomy. It is intended for use in general projection radiographic applications including pediatric and neonatal exams wherever conventional film/screen or CR systems may be used. The INNOLUX RIC is not intended for mammography, fluoroscopy, tomography, and angiography applications.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Innolux. The logo has the company name in blue and green letters on the top line, with the last letter in red. Below the company name are four Chinese characters in black.

510(k) Summary

INNOLUX RIC

Date: November 23, 2016

Submitter's Information:

INNOLUX Corporation No.3, Sec. 1, Huansi Rd., Tainan City, Southern Taiwan Science Park, 74147, Taiwan (R.O.C.)

Contact Person:

Name:Andy Kuo
Title:Director
Telephone:+886-37-586000 #64330
Facsimile:+886-37-586000 #65992

Identification of the Device:

Proprietary/Trade Name: INNOLUX RIC Classification Name: Stationary x-ray system 21 CFR 892.1680 Regulations Number: Product Codes: MQB Device Class: Class II Review Panel: Radiology Common Name: Flat Panel Digital Detector System

Identification of the Legally Marketed Device:

FDR D-EVO II Flat Panel Detector System. K142003 cleared 10/21/2014 VIVIX-S WIRELESS, K122865 cleared 02/01/2013

l. DEVICE DESCRIPTION

The INNOLUX RIC Flat Panel Detector is a portable digital detector that integrated with an operating PC and an X-ray generator to acquire and digitize x-ray exposures from, standard radiographic systems and transfer for radiography diagnostic. The INNOLUX RIC is designed to be used in any environment that would typically use a radiographic cassette for examinations of adults, pediatrics and neonates. The detector models support both wireless and wired data communication. Detectors can be placed in a wall bucky for upright exams, a table bucky for recumbent exams, or removed from the bucky for non-grid or free cassette exams.

II. INDICATIONS FOR USE

The Wireless/Wired INNOLUX RIC flat panel detector is intended to capture for display radiographic images of human anatomy. It is intended for use in general projection radiographic applications including pediatric and neonatal exams wherever conventional film/screen or CR systems may be used. The INNOLUX RIC is not intended for mammography, fluoroscopy, tomography, and angiography applications.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for Innolux. The word "Innolux" is written in large, bold, sans-serif font, with each letter in a different color. Below the word "Innolux" is the company's name in Chinese characters.

SUBSTANTIAL EQUIVALENCE lll.

INNOLUX RIC is substantially equivalent to the following legally marketed device.

Legally Marketed Device510(k) #Clearance Date
VIVIX-S WIRELESSK12286502/01/2013
FDR D-EVO II Flat Panel Detector SystemK14200310/21/2014

INNOLUX RIC and FDR D-EVOII / VIVIX-S WIRELESS are portable digital detector systems that are used to acquire x-ray exposures. The INNOLUX RIC has the same Indications for Use, and very similar functional and technical requirements as the currently-cleared predicate device, K122865 / K142003. The most detector characteristics remain unchanged for INNOLUX RIC, and the image quality is substantially equivalent to the predicate device. The design made for the INNOLUX RIC have been successfully tested and validated as summarized below.

Predicate Comparison Table:

Subject/Proposed DevicePredicate Device K142003,DR-ID1201SE, DR-ID1202SEDR-ID1211SE, DR-ID1212SE
Indications for UseRIC 35G,RIC 43GRIC 35C,RIC 43CThe Wireless/Wired INNOLUX RIC flat paneldetector is intended to capture for displayradiographic images of human anatomy. It isintended for use in general projection radiographicapplications including pediatric and neonatalexams wherever conventional film/screen or CRsystems may be used. The INNOLUX RIC is notintended for mammography, fluoroscopy,tomography, and angiography applications.The Wireless/Wired FDR D-EVO II flat paneldetector system is intended to capture for displayradiographic images of human anatomy. It isintended for use in general projection radiographicapplications including pediatric and neonatalexams wherever conventional film/screen or CRsystems may be used. The FDR D-EVO II is notintended for mammography, fluoroscopy,tomography, and angiography applications.
ScintillatorSame as K142003RIC 35G, RIC 43G Gd2O2S:TbRIC 35C, RIC 43C CsI:TlDR-ID1201SE, DR-ID1202SE Gd2O2S:TbDR-ID1211SE, DR-ID1212SE CsI:Tl
X-rayConversionSame as K142003Indirect conversion (a-Si)
Detector CordSame as K142003Wired / Wireless
Detector WeightRIC 35G, RIC 35C:Approx. 6.5 lbsRIC 43G, RIC 43C:Approx. 8.0 lbsDR-ID1201SE,DR-ID1211SE : Approx.5.8 lbsDR-ID1202SE,DR-ID1212SE : Approx.7.1 lbs
Exposuresize/Active Area(inch)Same as K142003RIC 35G, RIC 35C:13.8x16.8RIC 43G ,RIC 43C:16.8x16.7DR-ID1201SE, DR-ID1211SE :13.8x16.8DR-ID1202SE, DR-ID1212SE :16.8x16.7
Dimensions(inch)(Detectorexterior)Same as K142003RIC 35G, RIC 35C:18.1(W) x 15.1(D) x 0.6 (H)RIC 43G, RIC 43C:18.1(W) x 18.1(D) x 0.6 (H)DR-ID1201SE, DR-ID1211SE:18.1(W) x 15.1(D) x 0.6 (H)DR-ID1202SE, DR-ID1212SE:18.1(W) x 18.1(D) x 0.6 (H)
Pixel SizeSame as K142003150 μm
Acquisition BitDepthSame as K14200316 bit
DQE (RQA5, 1Ip/mm, 1mR) –detector alone,withouttabletopRIC 35G,RIC 43G31%Measurement tolerance (±10%)RIC 35C,RIC 43C54% (Csl)Measurement tolerance (±10%)DR-ID1201SE,DR-ID1202SE30%Measurement tolerance (±10%)DR-ID1211SE,DR-ID1212SE54% (Csl)Measurement tolerance (±10%)

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Innolux. The logo is in blue, red, and green. Below the English name is the Chinese name for the company in black.

Subject/Proposed DevicePredicate Device K142003,
RIC 35G,RIC 43GDR-ID1201SE,DR-ID1202SE
RIC 35C,RIC 43CDR-ID1211SE,DR-ID1212SE
MTF (RQA5, 2lp/mm)RIC 35G,RIC 43G42%Measurement tolerance (±10%)DR-ID1201SE,DR-ID1202SE42%(High mode)Measurement tolerance (±10%)
RIC 35C,RIC 43C54%Measurement tolerance (±10%)DR-ID1211SE,DR-ID1212SE54%(High mode)Measurement tolerance (±10%)
*StandardConfigurationComponentsSystem requiresone or more RIC detectors, one or two powersupply unit(s), and one control unit*.System requiresone FDX Console, one or more D-EVO detectors,one or two power supply unit(s),and one control unit*.
* In order to display an image, it requires theimage processing unit to meet the interfacespecification of RICMC.*Depending on configuration, the control unitsoftware can be installed in the FDX ConsolePC.
*WorkstationN/AWorkstation is not included in the system.Same as K122865FDX Console Version 8.0 and above

*The Subject device and another predicate. K122865 are not including image processing unit (console).

IV. SUMMARY OF STUDIES

Non-clinical Performance Data: INNOLUX RIC conforms to the voluntary standards such as AAMI/ANSI ES60601-1, IEC 60601-1, IEC 60601-1-2, IEC 62304, IEC 62366. In addition, the FDA's Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices (issued on August 6, 1999) was followed to describe the detector characteristics, and Radio Frequency Wireless Technology in Medical Devices (issued August 14, 2013) was followed to test. As required by the risk analysis, necessary verification and validation activities were performed including software testing, and the results were satisfactory. Furthermore, the image quality evaluation confirmed that the image quality of the INNOLUX RIC detectors is substantially equivalent to that of the predicate device.

Clinical Performance Data: No clinical study has been performed. The substantial equivalence has been demonstrated by non-clinical studies.

V. CONCLUSION

Based upon the supporting data summarized above, we concluded the INNOLUX RIC is as safe and effective as the legally marketed device FDR D-EVO II (K142003) and VIVI-S WIRELESS (K122865), and do not raise different questions of safety and effectiveness than K132509 and K122865.

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.