(90 days)
The Wireless/Wired FDR D-EVO II flat panel detector system is intended to capture for display radiographic images of human anatomy. It is intended for use in general projection radiographic applications including pediatric and neonatal exams wherever conventional film/screen or CR systems may be used. The FDR D-EVO II is not intended for mammography, fluoroscopy, tomography, and angiography applications.
Fujifilm's FDR D-EVO II Flat Panel Detector System (DR-ID1200) is a portable digital detector system that interfaces with, and acquires and digitizes x-ray exposures from, standard radiographic systems. The FDR D-EVO II is designed to be used in any environment that would typically use a radiographic cassette for examinations of adults, pediatrics and neonates. The detector models support both wireless and wired/tethered data communication between the detector and the system. Detectors can be placed in a wall bucky for upright exams, a table bucky for recumbent exams, or removed from the bucky for non-grid or free cassette exams.
While maintaining Fujifilm's unique Irradiated Side Sampling (ISS) design delivering high image quality. FDR D-EVO II offers the new and improved 1200 series flat panel detectors with upgraded wireless feature, memory exposure mode, and extended image readout feature. Additionally, the 1200 series detectors are equipped with several changes: Rounded-edge design for easy handling, image compression algorithm for faster image transfer, improved internal circuit design for electronic noise reduction, new LED design for easy detector identification, extra protection against ingress of water, and antibacterial coating designed to maintain cleanness. Not only the 1200 series detectors' robust design enhances load-bearing characteristics, but also the powersaving designs combined with the new battery packs and charger improve battery performance. The weight of 1200 series detectors is less than that of 600 series.
The provided document
"FUJIFILM Medical Systems U.S.A., Inc." (K142003)
is a 510(k) premarket notification for the "FDR D-EVO II Flat Panel Detector System (DR-ID1200)".
This document states that no clinical study was performed to demonstrate that the device meets acceptance criteria. Instead, substantial equivalence was demonstrated by non-clinical studies. This means there is no data on human clinical performance provided in this document to populate the requested table or answer several of the questions, as they pertain to clinical studies.
However, the document does mention non-clinical performance data and states that "the image quality evaluation confirmed that the image quality of the FDR D-EVO II system using new 1200 series detectors is substantially equivalent to that of the predicate device." This suggests that image quality was a key acceptance criterion and was evaluated through non-clinical means.
Here's a breakdown of the requested information based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Non-Clinical) | Reported Device Performance (Non-Clinical) |
|---|---|
| Compliance with voluntary standards (AAMI/ANSI ES60601-1, IEC 60601-1, IEC 60601-1-2, IEC 62304, IEC 62366, IEC 62494-1, DICOM, FDA Guidance for 510(k) for Solid State X-ray Imaging Devices, Radio Frequency Wireless Technology in Medical Devices) | Device conforms to all listed standards and guidance documents. |
| Software changes (memory exposure mode, extended image readout, image compression, battery power-saving designs) evaluated per FDA Guidance for Content of Premarket Submissions for Software Contained in Medical Devices (moderate level of concern). | Successfully evaluated and verified/validated. |
| Load-bearing characteristics | Tested and passed. |
| Enhanced protection against ingress of water | Tested and passed (increased IPX level). |
| Internal circuit design change for noise reduction | Demonstrated through EMC emission testing per IEC60601-1-2 with satisfactory results. |
| Antibacterial coating safety and effectiveness | Demonstrated through ISO 10993 testing and JIS Z 2801 (equivalent to ISO 22196) testing. |
| Image Quality | Image quality is substantially equivalent to the predicate device (FDR D-EVO DR-ID600, K132509). MTF and DQE measurements are very similar. |
2. Sample size used for the test set and the data provenance:
- Not applicable for clinical data: "No clinical study has been performed."
- For non-clinical tests: The document does not specify sample sizes for the various non-clinical engineering and performance tests (e.g., load-bearing, water ingress, EMC, antibacterial coating). The data provenance is internal to Fujifilm's testing and validation processes.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):
- Not applicable for clinical data: "No clinical study has been performed." For non-clinical tests, "ground truth" as referred to in a clinical context (expert consensus on disease presence) is not relevant. Engineering standards and specifications served as the "ground truth" for non-clinical performance.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable for clinical data: "No clinical study has been performed."
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC comparative effectiveness study was not done. The device itself is a flat panel detector (hardware), not an AI-assisted diagnostic tool. No clinical studies were performed.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- No standalone clinical performance study was done. The device is a digital X-ray detector; it does not operate as an "algorithm only" device in a diagnostic sense. Its standalone performance was evaluated through non-clinical means, primarily engineering tests and comparisons of technical imaging characteristics (MTF, DQE) to a predicate device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For non-clinical studies: The "ground truth" for the non-clinical tests was established by adherence to recognized industry standards (e.g., AAMI/ANSI ES60601-1, IEC series, ISO 10993, JIS Z 2801, DICOM), FDA guidance documents, and internal design specifications/requirements verified through engineering testing.
- For clinical studies: Not applicable, as no clinical studies were performed.
8. The sample size for the training set:
- Not applicable: This device is a flat panel detector; it is not described as an AI/machine learning device that would require a "training set" for an algorithm.
9. How the ground truth for the training set was established:
- Not applicable: As the device does not employ machine learning/AI requiring a training set, this question is not relevant.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with a wavy line below them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 21, 2014
FUJIFILM Medical Systems U.S.A., Inc. % Ms. Katherine Choi Regulatory Affairs Lead 419 West Avenue STAMFORD CT 06902
Re: K142003
Trade/Device Name: FDR D-EVO II Flat Panel Detector System Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MOB Dated: July 22, 2014 Received: July 23, 2014
Dear Ms. Choi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Smh.f)
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Section 4
Indications for Use (IFU) Statement
FDR D-EVO II Flat Panel Detector System (DR-ID1200)
This Section contains:
Indications for Use Statement
{3}------------------------------------------------
Page Intentionally Blank
{4}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known)
Device Name
FDR D-EVO II Flat Panel Detector System (DR-ID1200)
Indications for Use (Describe)
The Wireless/Wired FDR D-EVO II flat panel detector system is intended to capture for display radiographic images of human anatomy. It is intended for use in general projection radiographic applications including pediatric and neonatal exams wherever conventional film/screen or CR systems may be used. The FDR D-EVO II is not intended for mammography, fluoroscopy, tomography, and angiography applications.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Page Intentionally Blank
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Fujifilm Medical Systems, USA. The logo is in black and white, with the exception of a red square in the upper right corner of the "J" in Fujifilm. The text "Fujifilm Medical Systems, USA" is located below the main logo in a smaller font size. The logo is simple and modern, and it is likely used on the company's website, marketing materials, and products.
510(k) Summary
FDR D-EVO II Flat Panel Detector System (DR-ID1200)
Date: September 18, 2014
Submitter's Information:
FUJIFILM Medical Systems U.S.A., Inc. 419 West Avenue Stamford, CT, 06902, USA
Contact Person:
| Name: | Katherine Y. Choi, RAC |
|---|---|
| Title: | Regulatory Affairs Lead |
| Telephone: | (203) 602-3568 |
| Facsimile: | (203) 602-3785 |
Identification of the Device:
| Proprietary/Trade Name: | FDR D-EVO II Flat Panel Detector System(DR-ID1200) |
|---|---|
| Classification Name: | Stationary x-ray system |
| Regulations Number: | 21 CFR 892.1680 |
| Product Codes: | 90 MQB |
| Device Class: | Class II |
| Review Panel: | Radiology |
| Common Name: | Flat Panel Digital Detector System |
ldentification of the Legally Marketed Device:
FDR D-EVO Flat Panel Detector System (DR-ID600), K132509 cleared 11/25/2013
l. DEVICE DESCRIPTION
Fujifilm's FDR D-EVO II Flat Panel Detector System (DR-ID1200) is a portable digital detector system that interfaces with, and acquires and digitizes x-ray exposures from, standard radiographic systems. The FDR D-EVO II is designed to be used in any environment that would typically use a radiographic cassette for examinations of adults, pediatrics and neonates. The detector models support both wireless and wired/tethered data communication between the detector and the system. Detectors can be placed in a wall bucky for upright exams, a table bucky for recumbent exams, or removed from the bucky for non-grid or free cassette exams.
While maintaining Fujifilm's unique Irradiated Side Sampling (ISS) design delivering high image quality. FDR D-EVO II offers the new and improved 1200 series flat panel detectors with upgraded wireless feature, memory exposure mode, and extended image readout feature. Additionally, the 1200 series detectors are equipped with several changes: Rounded-edge design for easy handling, image compression algorithm for faster image transfer, improved internal circuit design for electronic noise reduction, new
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the Fujifilm Medical Systems, USA logo. The logo is black, except for a red square on top of the letter 'I' in Fujifilm. The text 'Fujifilm Medical Systems, USA' is located below the Fujifilm logo.
LED design for easy detector identification, extra protection against ingress of water, and antibacterial coating designed to maintain cleanness. Not only the 1200 series detectors' robust design enhances load-bearing characteristics, but also the powersaving designs combined with the new battery packs and charger improve battery performance. The weight of 1200 series detectors is less than that of 600 series.
= INDICATIONS FOR USE
The Wireless/Wired FDR D-EVO II flat panel detector system is intended to capture for display radiographic images of human anatomy. It is intended for use in general projection radiographic applications including pediatric and neonatal exams wherever conventional film/screen or CR systems may be used. The FDR D-EVO II is not intended for mammography, fluoroscopy, tomography, and angiography applications,
. SUMMARY OF TECHNOLOGICAL CHARACTERISTICS
Fujifilm's FDR D-EVO II FPD System (DR-ID1200) has the same Indications for Use as the predicate device (K132509). FDR D-EVO II detectors' scintillator materials (GOS or Csl), indirect conversion method (a-Si), Fujifilm's unique ISS technology, readout properties, 150um pixel pitch, and 16 bit-depth remain the same as the predicate device (K132509). The MTF and DQE measurements are very similar between FDR D-EVO II and the predicate device (K132509). While both FDR D-EVO II and the predicate device (K132509) are equipped with wireless detectors, wireless feature in FDR D-EVO II has been improved by expanding operating frequency options, and adding new wireless components. FDR D-EVO II's system configurations are very similar to the predicate (K132509), but a later version of FDX Console will support FDR D-EVO II. Additionally, active area size and pixel matrix of the FDR D-EVO II detectors are slightly different from the predicate device (K132509). FDR D-EVO II detectors weight less, but can withstand heavier loads when compared to the predicate device (K132509). Other differences includes, as aforementioned, new memory exposure mode, extended image readout feature (up to 10 sec), image compression algorithm, increased IPX level, antibacterial coating, and power-saving designs, all of which have been successfully tested and validated.
IV. SUBSTANTIAL EQUIVALENCE
Fujifilm FDR D-EVO II FPD System (DR-ID1200) is substantially equivalent to the following legally marketed device.
| Legally Marketed Device | 510(k) # | Clearance Date |
|---|---|---|
| FDR D-EVO Flat Panel Detector System (DR-ID600) | K132509 | 11/25/2013 |
Both FDR D-EVO II (DR-ID1200) and FDR D-EVO (DR-ID600) are portable digital detector systems that are used to acquire x-ray exposures. The new FDR D-EVO II has the same Indications for Use, and very similar functional and technical requirements as the currently-cleared predicate device. K132509. The most detector characteristics remain unchanged for FDR D-EVO II, and the image quality is substantially equivalent to the predicate device. The design modifications made for the FDR D-EVO II have been successfully tested and validated as summarized below.
> SUMMARY OF STUDIES
Non-clinical Performance Data: FDR D-EVO II FPD System (DR-ID1200) conforms to the voluntary standards such as AAMI/ANSI ES60601-1, IEC 60601-1, IEC 60601-1-2, IEC 62304, IEC 62366, IEC 62494-1 and DICOM. In addition, the FDA's Guidance for
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the Fujifilm Medical Systems, USA logo. The logo is composed of the word "FUJIFILM" in bold, black letters, with a red square above the "I". Below the main wordmark, the text "Fujifilm Medical Systems, USA" is written in a smaller, black font. The logo is simple and modern, and it is likely used to represent the company's brand.
the Submission of 510(k)'s for Solid State X-ray Imaging Devices (issued on August 6, 1999) was followed to describe the detector characteristics, and Radio Frequency Wireless Technology in Medical Devices (issued on Auqust 14, 2013) was followed to test the changes in wireless features. As required by the risk analysis, necessary verification and validation activities were performed. For example, software changes related to design modifications such as memory exposure mode, extended image readout feature, image compression algorithm, and battery power-saving designs were successfully evaluated according to the FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (issued on May 11, 2005) based on a moderate level of concern. Load-bearing characteristics and enhanced protection against ingress of water were tested and passed. The internal circuit design change to achieve noise reduction was demonstrated through EMC emission testing per IEC60601-1-2, and the results were satisfactory. Antibacterial coating's safety and effectiveness were demonstrated through ISO 10993 testing, and JIS Z 2801 (equivalent to ISO 22196) testing. Furthermore, the image quality evaluation confirmed that the image quality of the FDR D-EVO II system using new 1200 series detectors is substantially equivalent to that of the predicate device.
Clinical Performance Data: No clinical study has been performed. The substantial equivalence has been demonstrated by non-clinical studies.
VI. CONCLUSION
Based upon the supporting data summarized above, we concluded the next generation FDR D-EVO II Flat Panel Detector System (DR-ID1200) is as safe and effective as the legally marketed device DR-ID600 (K132509), and do not raise different questions of safety and effectiveness than K132509.
N/A