K Number
K161843

Validate with FDA (Live)

Date Cleared
2016-09-21

(78 days)

Product Code
Regulation Number
892.1550
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Diaqnostic Ultrasound Systems Aplio i900 Model TUS-AI900, Aplio i800 Model TUS-AI800 and Aplio i700 Model TUS-AI700 are indicated for the visualization of structures, and dynamic processes with the human body usinq ultrasound and to provide image information for diaqnosis in the following clinical applications: fetal, abdominal, intra-operative (abdominal), pediatric, small orqans, trans-vaqinal, trans-rectal, neonatal cephalic, adult cephalic, cardiac (both adult and pediatric), peripheral vascular, transesophageal, musculo-skeletal (both conventional and superficial) and laparoscopic.

Device Description

The Aplio i900 Model TUS-Al900, Aplio i800 Model TUS-Al800 and Aplio i700 Model TUS-Al700, V2.0 are mobile diagnostic ultrasound systems. These systems are Track 3 devices that employ a wide array of probes including flat linear array, convex linear array, and sector array with frequency ranges between approximately 2 MHz to 20 MHz.

AI/ML Overview

This document is a 510(k) Pre-market Notification for a diagnostic ultrasound system (Aplio i900/i800/i700, V2.0). It focuses primarily on demonstrating substantial equivalence to a predicate device and safety, rather than providing detailed acceptance criteria and study results for specific performance metrics that a standalone AI/device would typically have.

Here's an analysis based on the provided text, addressing your points where possible:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of quantitative acceptance criteria alongside corresponding test results. Instead, it describes performance in qualitative terms (e.g., "expected," "can be obtained," "acceptable," "improves workflow," "within specified range").

Feature/TestAcceptance Criteria (Implied/Qualitative)Reported Device Performance (Qualitative)
Sensor 3DProvide expected 3D images, measure accuracy, hardness information, and frequency dispersion of shear wave function.Sensor 3D provides the expected 3D images of phantom structures, measurement accuracy, hardness information and the frequency dispersion of the target of the 3D image with the shear wave function.
Attenuation ImagingObtain accurate quantitative attenuation coefficient results and display a color map of spatial distribution.Accurate quantitative attenuation coefficient results can be obtained using Attenuation Imaging and that a color map is displayed to show the spatial distribution of attenuation coefficient. (Confirmed in-vivo on volunteer livers with color map and numerical results).
Tissue Intensity Analysis (NLV)Visualize distribution of homogeneous and heterogeneous areas and display an acceptable color map.NLV can be used to visualize the distribution of homogeneous and heterogeneous areas of various phantoms by displaying a color map. (Confirmed on volunteer livers that NLV displays acceptable color map images and mean NLV values, in-vivo).
Fusion Auto TrackEnable automatic fusion of real-time ultrasound images to previously acquired CT or MR data sets using OmniTRAX.Fusion Auto Track enables automatic fusion of real-time ultrasound images to previously acquired CT or MR data sets by using the OmniTRAX Active Patient Tracker.
i Auto Volume MeasurementImprove workflow and maintain measurement accuracy within a specified range compared to the predicate device.i Auto Volume Measurement improves workflow using volume transducers compared with the predicate device and that the measurement accuracy is within the specified range.
CHI (Contrast Harmonic Imaging)Visualize nonlinear signals (2nd harmonic) from contrast medium, quantify peak intensity, and time-to-peak.CHI can visualize the nonlinear signal including 2nd harmonic from the Contrast medium, the quantification of the peak intensity, the time to reach the peak intensity.
Shear Wave Dispersion (SWD)Visualize frequency dependency of shear wave speed (Phase velocity).Shear Wave Dispersion (SWD) can visualize a frequency dependency of the shear wave speed (Phase velocity).
Shadow GlassDisplay tissue and tissue with flow in a transparent manner.Shadow Glass displays tissue and tissue with flow in a transparent manner.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify exact sample sizes for the test sets. It mentions "various phantom studies" and "representative clinical images of volunteer livers" (plural, suggesting more than one, but no specific number).
The provenance for clinical data is described as "volunteer livers" (in-vivo), but no country of origin is specified. The studies appear to be prospective for the clinical images used for verification.

3. Number of Experts Used to Establish Ground Truth and Qualifications

The document does not explicitly state the number or qualifications of experts used to establish ground truth for the test set.

4. Adjudication Method

The document does not mention any adjudication method used for the test set.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study is mentioned. The focus is on demonstrating the functionality and capabilities of the new features.

6. Standalone (Algorithm Only) Performance

The device is a diagnostic ultrasound system. The performance tests described (phantom studies, clinical images) inherently evaluate the algorithms as part of the overall system performance. There is no specific mention of "algorithm-only" performance metrics separate from the device's integrated operation. The new features incorporate advanced processing, and their performance is assessed directly through imaging and measurement tasks.

7. Type of Ground Truth Used

  • Phantom studies: Physical phantoms with known properties (e.g., structures, hardness, attenuation coefficients) were used as ground truth.
  • Clinical images: For the in-vivo evaluations, it is implied that the clinical images from "volunteer livers" served as the basis for assessing the qualitative aspects of the new features (e.g., whether the color maps were "acceptable," whether the expected features were visualized). However, there is no explicit mention of an external, independent ground truth (e.g., pathology, other imaging modalities) being used for these volunteer studies, beyond the visual and numerical output of the device itself being deemed acceptable by the evaluators.

8. Sample Size for the Training Set

The document describes pre-market notification for a diagnostic ultrasound system, not a device primarily driven by machine learning with distinct training and test sets in the AI sense. It discusses "new features" and "improvements to previously cleared features." Therefore, the concept of a "training set" in the context of AI models is not applicable or mentioned in this document. The system's development likely involved engineering, signal processing, and iterative refinement, not explicit AI model training.

9. How the Ground Truth for the Training Set Was Established

As explained in point 8, the concept of a "training set" for AI is not explicitly relevant to the descriptions in this document. Therefore, no information is provided on how ground truth for a training set was established.

In summary:

This 510(k) submission primarily focuses on establishing substantial equivalence for an ultrasound system with new and improved features. It describes functional performance verification through bench and clinical assessments, but it does not provide the detailed, quantitative acceptance criteria, explicit ground truth methodologies, or extensive statistical study results that would typically be expected for a standalone AI/ML-based medical device. The "acceptance criteria" are implied qualitative expectations for the performance of the new features, and the "study" is the verification and validation testing, predominantly bench testing with phantoms and limited clinical image acquisition in volunteers.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 21, 2016

Toshiba Medical Systems Corporation % Orlando Tadeo, Jr. Manager, Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive TUSTIN CA 92780

Re: K161843

Trade/Device Name: Aplio i900/i800/i700 Diagnostic Ultrasound System, V2.0 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: September 14, 2016 Received: September 15, 2016

Dear Mr. Tadeo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D.'Hara

For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K161843

Device Name

Aplio i900/i800/i700 Diagnostic Ultrasound System, V2.0

Indications for Use (Describe)

The Diaqnostic Ultrasound Systems Aplio i900 Model TUS-AI900, Aplio i800 Model TUS-AI800 and Aplio i700 Model TUS-AI700 are indicated for the visualization of structures, and dynamic processes with the human body usinq ultrasound and to provide image information for diaqnosis in the following clinical applications: fetal, abdominal, intra-operative (abdominal), pediatric, small orqans, trans-vaqinal, trans-rectal, neonatal cephalic, adult cephalic, cardiac (both adult and pediatric), peripheral vascular, transesophageal, musculo-skeletal (both conventional and superficial) and laparoscopic.

Type of Use (Select one or both, as applicable)

ك Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinica
---------------
Clinical ApplicationMode of OperationOther [Note]
Specific(Tracks 3)BMPWDCWDPrecision ImagingCombined (Specify)Color DopplerApli PureMicro PureBEAMPowerTDIElastographySMI(ADF)Shear wave4D3D Color(Volume color)STICSTIC ColorSmart 3DFusionSmart Navigation2D WMT
Ophthalmic
FetalNNNNN2NNNNNNNNNNNNNNN6,8
AbdominalNNNNN2,3NNNNNNNNNNNNNNN4,5,6,7,8
Intra-operative (Abdominal)NNNNN2NNNNNNNNNNNNNNNN
Intra-operative (Neuro)
LaparoscopicNNNNN2NNNNNNNNNNNNNNN
PediatricNNNNN2,3NNNNNNNNNNNNNNN6,8
Small Organ (Note 1)NNNNN2NNNNNNNNNNNNNNN6
Neonatal CephalicNNNNN3NNNNN
Adult CephalicNNNNN3N
Trans-rectalNNNNN2NNNNNNNNNNNNNNN8
Trans-vaginalNNNNN2NNNNNNNNNNNNNNN8
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)NNNNN2NNNNNNNNNNNNNNN6
Musculo-skeletal (Superficial)NNNNN2NNNNNNNNNNNNNNN6
Intravascular
Other (Specify)
Cardiac AdultNNNNN3NNN7
Cardiac PediatricNNNNN3NNN7
Intravascular (Cardiac)
Trans-esoph. (Cardiac)NNNNN3NN
Intra-cardiac
Other (Specify)
Peripheral vesselNNNNN2NNNNNNNNNNNNNNN6
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: N/A Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

Note 4 ATI

Note 5 Tissue Intensity Analysis

Note 6 Sensor3D

Note 7 CHI (Per FDA approved contrast agent prescribing information)

{4}------------------------------------------------

Transducer: PSI-30BX

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther [Note]
Specific(Tracks 3)BMPWDCWDColor DopplerPrecision ImagingCombined (Specify)Apli PureMicro PureBEAMPowerTDIElastographySMI(ADF)Shear wave4D(Volume color)3D Color(Volume color)STICSTIC ColorSmart 3DFusionSmart Navigation2D WMT
Ophthalmic
Fetal
AbdominalNNNNN3NNNN
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
PediatricNNNNN3NNNN
Small Organ (Note 1)
Neonatal CephalicNNNNN3NN
Adult CephalicNNNNN3N
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac AdultNNNNN3NNN7
Cardiac PediatricNNNNN3NNN7
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: N/A Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

Note 4 ATI

Note 5 Tissue Intensity Analysis

Note 6 Sensor3D

Note 7 CHI (Per FDA approved contrast agent prescribing information)

{5}------------------------------------------------

Transducer: PSI-70BT

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther [Note]
Specific(Tracks 3)BMPWDCWDColor DopplerPrecision ImagingCombined (Specify)Apli PureMicro PureBEAMPowerTDIElastographyShear waveSMI(ADF)4D(Volume color)3D Color(Volume color)STICSTIC ColorSmart 3DFusionSmart Navigation2D WMT
Ophthalmic
Fetal
AbdominalNNNNN3 NNNN
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
PediatricNNNNN3 NNNN
Small Organ (Note 1)
Neonatal CephalicNNNNN3 NNN
Adult CephalicNNNNN3N
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac AdultNNNNN3 NNNN7
Cardiac PediatricNNNNN3 NNNN7
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: N/A Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

Note 4 ATI

Note 5 Tissue Intensity Analysis

Note 6 Sensor3D

Note 7 CHI (Per FDA approved contrast agent prescribing information)

{6}------------------------------------------------

Transducer:_ PST-25BT

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther [Note]
Specific(Tracks 3)BMPWDCWDColor DopplerPrecision ImagingCombined (Specify)Micro PureApli PureBEAMPowerTDISMI(ADF)Shear wave4D(Volume color)3D ColorSTICSTIC ColorSmart 3DFusionSmart Navigation2D WMT
Ophthalmic
Fetal
AbdominalPPPPP3NPN
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPP3NPN
Small Organ (Note 1)
Neonatal CephalicPPPPP3NP
Adult CephalicPPPPP3P
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac AdultPPPPP3NPP
Cardiac PediatricPPPPP3NPP
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K151451 Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

Note 4 ATI

Note 5 Tissue Intensity Analysis

Note 6 Sensor3D

Note 7 CHI (Per FDA approved contrast agent prescribing information)

{7}------------------------------------------------

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther[Note]
Specific(Tracks 3)BMPWDCWDColor DopplerCombined (Specify)Precision ImagingMicro PureBEAMPowerTDIElastographyShear waveSMI(ADF)4D(Volume color)3D ColorSTICSTIC ColorSmart 3DFusionSmart Navigation2D WMT
Ophthalmic
Fetal
AbdominalPPPPP3NPN7(N)
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPP3NPN
Small Organ (Note 1)
Neonatal CephalicPPPPP3P
Adult CephalicPPPPP3P
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac AdultPPPPP3PP7
Cardiac PediatricPPPPP3PP7
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K151451 Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

Note 4 ATI

Note 5 Tissue Intensity Analysis

Note 6 Sensor3D

Note 7 CHI (Per FDA approved contrast agent prescribing information)

{8}------------------------------------------------

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationSpecific(Tracks 3)Mode of OperationBMPWDCWDColor DopplerCombined (Specify)Precision ImagingApli PureMicro PureBEAMPowerElastographyTDISMI(ADF)Shear wave4D3D Color(Volume color)STICSTIC ColorSmart 3DFusionSmart Navigation2D WMTOther[Note]
Ophthalmic
Fetal
AbdominalPPPPP3NP
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPPP3NP
Small Organ (Note 1)
Neonatal CephalicPPPPP3P
Adult CephalicPPPPP3P
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac AdultPPPPP3PP
Cardiac PediatricPPPPP3PP
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K151451 Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD

Note 4 ATI

Note 5 Tissue Intensity Analysis

Note 6 Sensor3D

Note 7 CHI (Per FDA approved contrast agent prescribing information)

{9}------------------------------------------------

Transducer: _ PVI-475BT

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther[Note]
Specific(Tracks 3)BMPWDCWDColor DopplerCombined (Specify)Precision ImagingApli PureMicro PureBEAMPowerTDIElastographySMI(ADF)Shear wave4D3D Color(Volume color)STICSTIC ColorSmart 3DFusionSmart Navigation2D WMT
Ophthalmic
FetalNNNN2NNNNN
AbdominalNNNN2NNNNNNNN6,7
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
PediatricNNNN2NNNNNNN6
Small Organ (Note 1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: N/A Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

Note 4 ATI

Note 5 Tissue Intensity Analysis

Note 6 Sensor3D

Note 7 CHI (Per FDA approved contrast agent prescribing information)

{10}------------------------------------------------

Transducer: _ PVI-475BX

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationSpecific(Tracks 3)Mode of OperationBMPWDCWDColor DopplerPrecision ImagingCombined (Specify)Apli PureMicro PureBEAMPowerTDIElastographySMI(ADF)Shear wave4D(Volume color)3D ColorSTICSTIC ColorSmart 3DFusionSmart Navigation2D WMTOther[Note]
Ophthalmic
FetalNNNNN2NNNNN
AbdominalNNNNN2NNNNNNNNNN4,5,6,7
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
PediatricNNNNN2NNNNNNN6
Small Organ (Note 1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: N/A Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M: B/PWD: BDF/PWD: BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

Note 4 ATI

Note 5 Tissue Intensity Analysis

Note 6 Sensor3D

Note 7 CHI (Per FDA approved contrast agent prescribing information)

{11}------------------------------------------------

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther[Note]
Specific(Tracks 3)BMPWDCWDColor DopplerCombined (Specify)Precision ImagingApli PureMicro PureBEAMPowerTDIElastographyShear waveSMI(ADF)4D(Volume color)3D Color(Volume color)STICSTIC ColorSmart 3DFusionSmart Navigation2D WMT
Ophthalmic
FetalPPPP2PPPNP
AbdominalPPPP2PPPPPPPPPP4(N), 5(N), 6(N), 7(N)
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPP2PPPPPPP6(N)
Small Organ (Note 1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K151451 Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD

Note 4 ATI

Note 5 Tissue Intensity Analysis

Note 6 Sensor3D

Note 7 CHI (Per FDA approved contrast agent prescribing information)

{12}------------------------------------------------

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther
Specific(Tracks 3)BMPWDCWDColor DopplerCombined (Specify)Precision ImagingApli PureMicro PureBEAMPowerTDIElastographySMI(ADF)Shear wave4D3D Color(Volume color)STICSTIC ColorSmart 3DFusionSmart Navigation2D WMT[Note]
Ophthalmic
FetalPPPP2PPPNP6(N)
AbdominalPPPP2PPPPPPPPPP4(N),5(N),6(N),7(N)
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPP2PPPPPPPP6(N)
Small Organ (Note 1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K151451 Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

Note 4 ATI

Note 5 Tissue Intensity Analysis

Note 6 Sensor3D

Note 7 CHI (Per FDA approved contrast agent prescribing information)

{13}------------------------------------------------

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationSpecific(Tracks 3)BMPWDCWDColor DopplerPrecision ImagingCombined (Specify)Apli PureMicro PureBEAMPowerTDIElastographySMI(ADF)Shear wave4D3D Color(Volume color)STICSTIC ColorSmart 3DFusionSmart Navigation2D WMTOther[Note]
Ophthalmic
FetalPPPP2PPPNPP6(N)
AbdominalPPPP2PPPPPPPP6(N),7(N)
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPP2PPPPPPP6(N)
Small Organ (Note 1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K151451 Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

Note 4 ATI

Note 5 Tissue Intensity Analysis

Note 6 Sensor3D

Note 7 CHI (Per FDA approved contrast agent prescribing information)

{14}------------------------------------------------

Transducer:___PVT-674BT

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther[Note]
Specific(Tracks 3)BMPWDCWDColor DopplerCombined (Specify)Precision ImagingApli PureMicro PureBEAMPowerTDIElastographySMI(ADF)Shear wave4D3D Color(Volume color)STICSTIC ColorSmart 3DFusionSmart Navigation2D WMT
Ophthalmic
FetalPPPP2PPPPP
AbdominalPPPP2PPPPP7(N)
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPP2PPPPP
Small Organ (Note 1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K151451 Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

Note 4 ATI

Note 5 Tissue Intensity Analysis

Note 6 Sensor3D

Note 7 CHI (Per FDA approved contrast agent prescribing information)

{15}------------------------------------------------

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther[Note]
Specific(Tracks 3)BMPWDCWDColor DopplerCombined (Specify)Precision ImagingApli PureMicro PureBEAMPowerTDIElastographySMI(ADF)Shear wave4D(Volume color)3D ColorSTICSTIC ColorSmart 3DSmart Navigation2D WMTFusionOther[Note]
Ophthalmic
FetalPPPP2PPPPNPPPP8(N)
AbdominalPPPP2PPPPNPPPP8(N)
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPP2PPPPNPPPP8(N)
Small Organ (Note 1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K151451 Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD

Note 4 ATI

Note 5 Tissue Intensity Analysis

Note 6 Sensor3D

Note 7 CHI (Per FDA approved contrast agent prescribing information)

{16}------------------------------------------------

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther[Note]
Specific(Tracks 3)BMPWDCWDColor DopplerCombined (Specify)Precision ImagingApli PureMicro PureBEAMPowerTDIElastographySMI(ADF)Shear wave4D(Volume color)3D ColorSTICSTIC ColorSmart 3DFusionSmart Navigation2D WMT
Ophthalmic
Fetal
AbdominalNNNN2NNN7(N)
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Note 1)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPP2PPPPPPP8(N)
Trans-vaginalPPPP2PPPPPPP8(N)
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K151451 Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

Note 4 ATI

Note 5 Tissue Intensity Analysis

Note 6 Sensor3D

Note 7 CHI (Per FDA approved contrast agent prescribing information)

{17}------------------------------------------------

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther[Note]
Specific(Tracks 3)BMPWDCWDCombined (Specify)Precision ImagingMicro PureApli PureBEAMPowerTDIElastographyShear waveSMI(ADF)4D3D Color(Volume color)STICSTIC ColorSmart 3DFusionSmart Navigation2D WMT
Ophthalmic
Fetal
AbdominalPPPP2PPPPNP
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
PediatricPPPP2PPPPNP
Small Organ (Note 1)
Neonatal CephalicPPPP2PPPPP
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K151451 Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

  • Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
    Note 4 ATI

  • Note 5 Tissue Intensity Analysis

  • Note 6 Sensor3D

  • Note 7 CHI (Per FDA approved contrast agent prescribing information)

{18}------------------------------------------------

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther[Note]
Specific(Tracks 3)BMPWDCWDColor DopplerPrecision ImagingCombined (Specify)Micro PureApli PureBEAMPowerTDIElastographyShear waveSMI(ADF)4D(Volume color)3D Color(Volume color)STICSTIC ColorSmart 3DFusionSmart Navigation2D WMT
Ophthalmic
Fetal
AbdominalPPPP2PPPPP7(N)
Intra-operative (Abdominal)PPPP2PPPPP
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Note 1)NNNN2NNNNN
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K151451 Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

Note 4 ATI

Note 5 Tissue Intensity Analysis

Note 6 Sensor3D

Note 7 CHI (Per FDA approved contrast agent prescribing information)

{19}------------------------------------------------

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther[Note]
Specific(Tracks 3)BMPWDCWDColor DopplerPrecision ImagingCombined (Specify)Micro PureApli PureBEAMPowerTDIElastographyShear waveSMI(ADF)4D3D Color(Volume color)STICSTIC ColorSmart 3DFusionSmart Navigation2D WMT
Ophthalmic
Fetal
AbdominalPPPP2PPPNP7(N)
Intra-operative (Abdominal)PPPP2PPPP
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Note 1)PPPP2PPPNP
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K151451 Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

Note 4 ATI

Note 5 Tissue Intensity Analysis

Note 6 Sensor3D

Note 7 CHI (Per FDA approved contrast agent prescribing information)

{20}------------------------------------------------

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther[Note]
Specific(Tracks 3)BMPWDCWDColor DopplerCombined (Specify)Precision ImagingMicro PureApli PureBEAMPowerTDIElastographyShear waveSMI(ADF)4D(Volume color)3D ColorSTICSTIC ColorSmart 3DFusionSmart Navigation2D WMT
Ophthalmic
Fetal
AbdominalPPPP2PPPNP
Intra-operative (Abdominal)PPPP2PPPP
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Note 1)PPPP2PPPNP
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K151451 Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

  • Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
    Note 4 ATI

  • Note 5 Tissue Intensity Analysis

  • Note 6 Sensor3D

  • Note 7 CHI (Per FDA approved contrast agent prescribing information)

{21}------------------------------------------------

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther[Note]
Specific(Tracks 3)BMPWDCWDColor DopplerPrecision ImagingCombined (Specify)Apli PureMicro PureBEAMPowerTDIElastographySMI(ADF)Shear wave4D(Volume color)3D ColorSTICSTIC ColorSmart 3DFusionSmart Navigation2D WMT
Ophthalmic
Fetal
AbdominalNNNN2NNNN7(N)
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Note 1)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPP2PPPPPPP6(N)
Trans-vaginalPPPP2PPPPPPP6(N)
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K151451 Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

Note 4 ATI

Note 5 Tissue Intensity Analysis

Note 6 Sensor3D

Note 7 CHI (Per FDA approved contrast agent prescribing information)

{22}------------------------------------------------

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther[Note]
Specific(Tracks 3)BMPWDCWDColor DopplerCombined (Specify)Precision ImagingApli PureMicro PureBEAMPowerTDIElastographySMI(ADF)Shear wave4D3D Color(Volume color)STICSTIC ColorSmart 3DFusionSmart Navigation2D WMT
Ophthalmic
Fetal
AbdominalNNNNN2NNNNN7(N)
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Note 1)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPP2PPPPPPPPP6(N)
Trans-vaginalPPPPP2PPPPPPPPP6(N)
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K151451 Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

Note 4 ATI

Note 5 Tissue Intensity Analysis

Note 6 Sensor3D

Note 7 CHI (Per FDA approved contrast agent prescribing information)

{23}------------------------------------------------

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationSpecific(Tracks 3)BMPWDCWDCombined (Specify)Color DopplerPrecision ImagingApli PureMicro PureBEAMPowerTDIElastographySMI(ADF)Shear wave4D(Volume color)3D ColorSTICSTIC ColorSmart 3DFusionSmart Navigation2D WMTOther[Note]
Ophthalmic
Fetal
AbdominalNNNN2NNN7(N)
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Note 1)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPP2PPPNPPP6(N)
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K151451 Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

Note 4 ATI

Note 5 Tissue Intensity Analysis

Note 6 Sensor3D

Note 7 CHI (Per FDA approved contrast agent prescribing information)

{24}------------------------------------------------

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther[Note]
Specific(Tracks 3)BMPWDCWDCombined (Specify)Precision ImagingMicro PureApli PureBEAMPowerTDIElastographyShear waveSMI(ADF)4D(Volume color)3D ColorSTICSTIC ColorSmart 3DFusionSmart Navigation2D WMT
Ophthalmic
Fetal
AbdominalNNNN2NNN7
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Note 1)NNNN2NNNNNNNNNNNN6
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)NNNN2NNNNNNNNNN6
Musculo-skeletal (Superficial)NNNN2NNNNNNNNNN6
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vesselNNNN2NNNNNNNNNN6
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: N/A Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M: B/PWD: BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

Note 4 ATI

Note 5 Tissue Intensity Analysis

Note 6 Sensor3D

Note 7 CHI (Per FDA approved contrast agent prescribing information)

{25}------------------------------------------------

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther[Note]
Specific(Tracks 3)BMPWDCWDColor DopplerPrecision ImagingCombined (Specify)Apli PureMicro PureBEAMPowerTDIElastographyShear waveSMI(ADF)4D(Volume color)3D ColorSTICSTIC ColorSmart 3DFusionSmart Navigation2D WMT
Ophthalmic
Fetal
Abdominal
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Note 1)NNNN2NNNNNNNNN6
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)NNNN2NNNNNNNNN6
Musculo-skeletal (Superficial)NNNN2NNNNNNNNN6
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vesselNNNN2NNNNNNNNN
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: N/A Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

Note 4 ATI

Note 5 Tissue Intensity Analysis

Note 6 Sensor3D

Note 7 CHI (Per FDA approved contrast agent prescribing information)

{26}------------------------------------------------

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther[Note]
Specific(Tracks 3)BMPWDCWDCombined (Specify)Color DopplerPrecision ImagingApli PureMicro PureBEAMPowerTDIElastographySMI(ADF)Shear wave4D(Volume color)3D ColorSTICSTIC ColorSmart 3DFusionSmart Navigation2D WMT
Ophthalmic
Fetal
AbdominalNNNN2NNN7(N)
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Note 1)PPPP2PPPPPP
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)PPPP2PPPPPP
Musculo-skeletal (Superficial)PPPP2PPPPPP
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vesselPPPP2PPPPPP
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K151451 Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

Note 4 ATI

Note 5 Tissue Intensity Analysis

Note 6 Sensor3D

Note 7 CHI (Per FDA approved contrast agent prescribing information)

{27}------------------------------------------------

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther[Note]
Specific(Tracks 3)BMPWDCWDCombined (Specify)Precision ImagingApli PureMicro PureBEAMPowerTDIElastographySMI(ADF)Shear wave4D(Volume color)3D ColorSTICSTIC ColorSmart 3DFusionSmart Navigation2D WMT
Ophthalmic
Fetal
AbdominalNNNN2NNN7(N)
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Note 1)PPPP2PPPPPP
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)PPPP2PPPPPP
Musculo-skeletal (Superficial)PPPP2PPPPPP
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vesselPPPP2PPPPPN
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K151451 Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

Note 4 ATI

Note 5 Tissue Intensity Analysis

Note 6 Sensor3D

Note 7 CHI (Per FDA approved contrast agent prescribing information)

{28}------------------------------------------------

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther[Note]
Specific(Tracks 3)BMPWDCWDCombined (Specify)Precision ImagingApli PureMicro PureBEAMPowerTDIElastographySMI(ADF)Shear wave4D(Volume color)3D ColorSTICSTIC ColorSmart 3DFusionSmart Navigation2D WMT
Ophthalmic
Fetal
AbdominalNNNN2NNN7(N)
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Note 1)PPPP2PPPPPPPPPP6(N)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)PPPP2PPPPPPPPP6(N)
Musculo-skeletal (Superficial)PPPP2PPPPPPPPP6(N)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vesselPPPP2PPPPPPPPP6(N)
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K151451 Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

Note 4 ATI

Note 5 Tissue Intensity Analysis

Note 6 Sensor3D

Note 7 CHI (Per FDA approved contrast agent prescribing information)

{29}------------------------------------------------

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Specific(Tracks 3)BMawaന്നുംolor Dopplepəuiquvaine (September 1999)นเชียนป์ นับเรเวอ.ərin qalında və qalınmışdır. Bu və bir və qalınmışdır. Bu və bir və qalınmışdır. Bu və bir və qalınmışdır. Bu və bir və qalınmışdır. Bu və bir və qalınmışdır. Bu və bir və bMicro PurceMA.BBPowerIATវិជ្ជប់ខ្លះ រដ្ឋបានខ្លួនខ្លួនខ្មែរ ទ្រ[ACA)IMЭЛЕМ МЕРИand and on anyon asSTICSTIC ColoDE TIEWuoisnDIAERIAEN TIRUITIMM dOther[Note]
Ophthalmic
Fetal
AbdominalNNNN2NNN7(N)
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Note 1)PbPP2bPPPPPP
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)PbbP2PbPPPPP
Musculo-skeletal (Superficial)PPPP2PPPPPPP
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vesselPPPP2bPPPPPP
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K151451 Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

Note 4 ATI

Note 5 Tissue Intensity Analysis

Note 6 Sensor3D

Note 7 CHI (Per FDA approved contrast agent prescribing information)

{30}------------------------------------------------

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
Specific(Tracks 3)BMPWDCWDColor DopplerCombined (Specify)Precision ImagingMicro PureApli PureBEAMPowerElastographyTDIShear waveSMI(ADF)4D(Volume color)3D ColorSTICSTIC ColorSmart 3DFusionSmart Navigation2D WMTOther[Note]
Ophthalmic
Fetal
Abdominal
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Note 1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)PPPPP3PP
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K151451 Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

Note 4 ATI

Note 5 Tissue Intensity Analysis

Note 6 Sensor3D

Note 7 CHI (Per FDA approved contrast agent prescribing information)

{31}------------------------------------------------

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther [Note]
Specific(Tracks 3)BMPWDCWDCombined (Specify)Precision ImagingMicro PureApli PureBEAMPowerTDIElastographyShear waveSMI(ADF)4D(Volume color)3D ColorSTICSTIC ColorSmart 3DFusionSmart Navigation2D WMTOther[Note]
Ophthalmic
Fetal
Abdominal
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Note 1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)NNNNN3NN
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: N/A Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

Note 4 ATI

Note 5 Tissue Intensity Analysis

Note 6 Sensor3D

Note 7 CHI (Per FDA approved contrast agent prescribing information)

{32}------------------------------------------------

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther[Note]
Specific(Tracks 3)BMPWDCWDColor DopplerPrecision ImagingCombined (Specify)Apli PureMicro PureBEAMPowerTDIElastographySMI(ADF)Shear wave4D(Volume color)3D Color(Volume color)STICSTIC ColorSmart 3DFusionSmart Navigation2D WMT
Ophthalmic
Fetal
AbdominalNNNN2NNN7(N)
Intra-operative (Abdominal)
Intra-operative (Neuro)
LaparoscopicPPPP2PPPPPP
Pediatric
Small Organ (Note 1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K151451 Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

Note 4 ATI

Note 5 Tissue Intensity Analysis

Note 6 Sensor3D

Note 7 CHI (Per FDA approved contrast agent prescribing information)

{33}------------------------------------------------

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther[Note]
Specific(Tracks 3)BMPWDCWDPrecision ImagingCombined (Specify)Micro PureApli PureBEAMPowerTDIElastographySMI(ADF)Shear wave4D(Volume color)3D Color(Volume color)STICSTIC ColorSmart 3DFusionSmart Navigation2D WMT
Ophthalmic
Fetal
Abdominal
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
PediatricN
Small Organ (Note 1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac AdultP
Cardiac PediatricP
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vesselP
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K151451 Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

Note 4 ATI

Note 5 Tissue Intensity Analysis

Note 6 Sensor3D

Note 7 CHI (Per FDA approved contrast agent prescribing information)

{34}------------------------------------------------

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther
Specific(Tracks 3)BMPWDCWDColor DopplerPrecision ImagingCombined (Specify)Micro PureApli PureBEAMPowerTDIElastographySMI(ADF)Shear wave4D(Volume color)3D ColorSTICSTIC ColorSmart 3DFusionSmart Navigation2D WMTOther[Note]
Ophthalmic
Fetal
Abdominal
Intra-operative (Abdominal)
Intra-operative (Neuro)
Laparoscopic
PediatricN
Small Organ (Note 1)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal(Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Specify)
Cardiac AdultP
Cardiac PediatricP
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
Peripheral vesselP
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K151451 Prescription Use Only (Per 21 CFR 801.109)

Note 1 Small organ includes thyroid, breast and testicle

Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD

Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD

Note 4 ATI

Note 5 Tissue Intensity Analysis

Note 6 Sensor3D

Note 7 CHI (Per FDA approved contrast agent prescribing information)

{35}------------------------------------------------

Image /page/35/Picture/0 description: The image contains the words "TOSHIBA MEDICAL" in a bold, red font. The word "TOSHIBA" is stacked on top of the word "MEDICAL". The text appears to be a logo or branding element.

510(k) SUMMARY

    1. SUBMITTER'S NAME: Toshiba Medical Systems Corporation 1385 Shimoishigami Otawara-shi, Tochigi-ken, Japan 324-8550
  • OFFICIAL CORRESPONDENT 2. Akinori Hatanaka
    1. ESTABLISHMENT REGISTRATION: 9614698

CONTACT PERSON: 4.

Orlando Tadeo, Jr. Manager, Regulatory Affairs Toshiba America Medical Systems, Inc 2441 Michelle Drive Tustin, CA 92780 (714) 669-7459

    1. Date Prepared:
      September 16, 2016

6. TRADE NAME(S):

Aplio i900/i800/i700 Diagnostic Ultrasound System, V2.0

7. COMMON NAME:

System, Diagnostic Ultrasound

DEVICE CLASSIFICATION: 8.

Class II

Ultrasonic Pulsed Doppler Imaging System – Product Code: 90-IYN [per 21 CFR 892.1550] Ultrasonic Pulsed Echo Imaging System – Product Code: 90-IYO [per 21 CFR 892.1560] Diagnostic Ultrasonic Transducer - Product Code: 90-ITX [per 21 CFR 892.1570]

{36}------------------------------------------------

9. PREDICATE DEVICE:

ProductMarketed by510(k) NumberClearance Date
Aplio 500/400/300Diagnostic Ultrasound System V6.0Toshiba AmericaMedical SystemsK151451July 7, 2015
Vivid E9 Diagnostic UltrasoundSystemGE HealthcareK131514July 12, 2013

10. REASON FOR SUBMISSION:

New device

11. DEVICE DESCRIPTION:

The Aplio i900 Model TUS-Al900, Aplio i800 Model TUS-Al800 and Aplio i700 Model TUS-Al700, V2.0 are mobile diagnostic ultrasound systems. These systems are Track 3 devices that employ a wide array of probes including flat linear array, convex linear array, and sector array with frequency ranges between approximately 2 MHz to 20 MHz.

12. INDICATIONS FOR USE:

The Diagnostic Ultrasound Systems Aplio i900 Model TUS-Al900, Aplio i800 Model TUS-Al800 and Aplio i700 Model TUS-AI700 are indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diagnosis in the following clinical applications: fetal, abdominal, intra-operative (abdominal), pediatric, small organs, trans-vaginal, trans-rectal, neonatal cephalic, adult cephalic, cardiac (both adult and pediatric), peripheral vascular, transesophageal, musculoskeletal (both conventional and superficial) and laparoscopic.

13. SUBSTANTIAL EQUIVALENCE:

This device is substantially equivalent to the Aplio 500/400/300 V6.0 Diagnostic Ultrasound System. K151451. marketed by Toshiba America Medical Systems. The Aplio i900 Model TUS-Al900, Aplio i800 Model TUS-Al800 and Aplio i700 Model TUS-Al700, V2.0 function in a manner similar to and is intended for the same use as the predicate device. The subject device is a compact diagnostic ultrasound system by implementing latest technologies.

Aplio 500/400/300V6.0Aplio i900/i800/i700V2.0Comment
510(K) Control NumberK151451N/A (Subject Device)
Sensor 3DNoYesNew feature
Shadow GlassNoYesNew feature
Attenuation ImagingNoYesNew feature
Shear Wave Dispersion(SWD)NoYesNew feature
Tissue Intensity Analysis(Normalized Local Variance(NLV)/Ratio)NoYesNew feature
CHI (Contrast HarmonicImaging) for LiverNoYesNew feature

{37}------------------------------------------------

Aplio 500/400/300V6.0Aplio i900/i800/i700V2.0Comment
510(K) Control NumberK151451N/A (Subject Device)
Auto Volume MeasurementNoYesNew feature
Fusion Auto TrackNoYesNew feature
PSI-30BX transducerNoYesNew transducer
PSI-70BT transducerNoYesNew transducer
PVI-475BT transducerNoYesNew transducer
PVI-475BX transducerNoYesNew transducer
PLI-1205BX transducerNoYesNew transducer
PLI-2004BX transducerNoYesNew transducer
Vivid E9 DiagnosticUltrasound SystemAplio i900/i800/i700V2.0Comment
510(K) Control NumberK131514N/A (Subject Device)
Sensor 3D-MeasurementsDistance/Angle/Area/VolumeYes(Tru3D(G))YesNew feature
Registration with OmniTRAXActive Patient Tracker (CIVCOMedical Instruments Co., Inc)- CT image- MR imageYesYes(Fusion Auto Track)New feature*CIVCO 510(k)clearancenumbersCT: K092619MR: K143396

Previously cleared features being implemented to the subject device:

Feature510(k) ClearanceComments
Shear Wave ElastographyPreviously cleared under K151451No change to basic function
ElastographyPreviously cleared under K151451No change to basic function
SMIPreviously cleared under K151451No change to basic function
Smart FusionPreviously cleared under K151451No change to basic function
Smart NavigationPreviously cleared under K151451No change to basic function
Auto-NTPreviously cleared under K151451No change to basic function
MultiCast Beam FormerPreviously cleared under K140729Improvement to basic function (Histogram)
Tissue Intensity Analysis (Ratio)Previously cleared under K991710Improvement to basic function (Multiple focus)

{38}------------------------------------------------

14. SAFETY:

The device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with the applicable parts of the IEC60601-1 (2005). IEC 60601-2-37 (2007). IEC 62304 (2006). AIUM RTD2-2004 Output Display and ISO 10993-1 standards.

15. TESTING

Risk Analysis. Verification/Validation testing conducted through bench testing which are included in this submission demonstrates that the requirements for the features have been met.

Performance Testing - Bench

Various phantom studies were conducted to demonstrate that the new features and improvements to previously cleared features being implemented to the subject device performed as intended. The studies concluded the following:

  • Sensor 3D provides the expected 3D images of phantom structures, measurement accuracy, hardness information and the frequency dispersion of the target of the 3D image with the shear wave function
  • Accurate quantitative attenuation coefficient results can be obtained using Attenuation lmaging and that a color map is displayed to show the spatial distribution of attenuation coefficient
  • NLV can be used to visualize the distribution of homogeneous and heterogeneous areas of various phantoms by displaying a color map
  • Fusion Auto Track enables automatic fusion of real-time ultrasound images to previously acquired CT or MR data sets by using the OmniTRAX Active Patient Tracker
  • i Auto Volume Measurement improves workflow using volume transducers compared with the predicate device and that the measurement accuracy is within the specified range
  • CHI can visualize the nonlinear signal including 2nd harmonic from the Contrast medium, the quantification of the peak intensity, the time to reach the peak intensity
  • Shear Wave Dispersion (SWD) can visualize a frequency dependency of the shear wave speed (Phase velocity)
  • Shadow Glass displays tissue and tissue with flow in a transparent manner

Performance Testing - Clinical Images

Clinical evaluations were conducted to demonstrate that the subject device performed as expected. In one study representative clinical images of volunteer livers were obtained and it was confirmed that Attenuation Imaging can display a color map of attenuation coefficient as well as numerical results in-vivo. In another study representative clinical images of volunteer livers were obtained and using NLV, the study confirmed that NLV displays acceptable color map images and mean NLV values, in-vivo.

Software Documentation for a Moderate Level of Concern, per the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005, is also included as part of this submission.

{39}------------------------------------------------

Additionally, testing of this device was conducted in accordance with the applicable standards published by the International Electrotechnical Commission (IEC) for Medical Devices.

16. CONCLUSION

The Aplio i900 Model TUS-Al900, Aplio i800 Model TUS-Al800 and Aplio i700 Model TUS-Al700, V2.0 is substantially equivalent to the predicate devices. The subject devices function in a manner similar to and is intended for the same use as the predicate devices, as described in the labeling. Based upon the bench testing, acquisition of representative clinical images, successful completion of software validation, application of risk management and design controls, it is concluded that this device is safe and effective for its intended use.

N/A