Search Results
Found 3 results
510(k) Data Aggregation
(104 days)
The Diaqnostic Ultrasound Systems Aplio i900 Model TUS-AI900, Aplio i800 Model TUS-AI800, Aplio i700 Model TUS-AI700 and Aplio i600 Model TUS-AI600 are indicated for the visualization of structures, and dynamic processes with the human body usinq ultrasound and to provide image information for diaqnosis in the followinq clinical applications: fetal, abdominal, intra-operative (abdominal), pediatric, small organs, trans-vaginal, trans-rectal, neonatal cephalic, adult cephalic, cardiac (both adult and pediatric), peripheral vascular, transesophageal, musculo-skeletal (both conventional and superficial) and laparoscopic.
The Aplio i900 Model TUS-AI900, Aplio i800 Model TUS-AI800, Aplio i700 Model TUS-AI700 and Aplio i600 Model TUS-A1600, V2.4 are mobile diagnostic ultrasound systems. These systems are Track 3 devices that employ a wide array of probes including flat linear array, convex linear array, and sector array with frequency ranges between approximately 2 MHz to 20 MHz.
This document describes the Toshiba Medical Systems Corporation Aplio i900/i800/i700/i600 Diagnostic Ultrasound System, V2.4, a modified device, and compares it to a predicate device (Aplio i900/i800/i700, V2.1, K163702).
Here's a breakdown of the acceptance criteria and study information provided:
1. Table of Acceptance Criteria and Reported Device Performance
The device is substantially equivalent to its predicate device and offers several new features and improvements. Acceptance criteria are typically defined by demonstrating substantial equivalence to a legally marketed predicate device and meeting performance specifications through testing.
Specific performance assessments for new and improved features include:
Feature/Metric | Acceptance Criteria (Demonstrated Equivalency/Performance) | Reported Device Performance |
---|---|---|
3D Wall Motion Tracking (LV Analysis) | Equivalency in volume measurement (EDV/ESV/EF) and strain measurement (Area Change Ratio, Global Longitudinal Strain, Global Circumference Strain, Global Radial Strain) compared to predicate devices. | Demonstrated equivalency in volume and strain measurements. The implemented 3D WMT features enabled cardiac wall trace, local wall motion tracking, wall motion information analysis and display, cardiac volume measurement, and cardiac function analysis and display of three analysis features simultaneously. |
3D Wall Motion Tracking (LA Analysis) | Equivalency in volume measurement (EDV/ESV or Vmax/Vmin) and strain measurement (Global Area Change Ratio, Global Longitudinal Strain and Global Circumference Strain) compared to predicate devices. | Demonstrated equivalency in volume and strain measurements. The implemented 3D WMT features enabled cardiac wall trace, local wall motion tracking, wall motion information analysis and display, cardiac volume measurement, and cardiac function analysis and display of three analysis features simultaneously. |
3D Wall Motion Tracking (RV Analysis) | Equivalency in strain measurement (Area Change Ratio, Global Longitudinal Strain, Global Circumference Strain) and a Correlation coefficient (r) >= 0.90 with n=11 for volume measurement (EDV/ESV/EF) to the predicate function (Vivid E9, K131514 with GE EchoPAC, K120221). | Demonstrated equivalency in strain measurements and a correlation coefficient (r) >= 0.90 for volume measurement. The implemented 3D WMT features enabled cardiac wall trace, local wall motion tracking, wall motion information analysis and display, cardiac volume measurement, and cardiac function analysis and display of three analysis features simultaneously. |
Auto-EF | Demonstrated reduction in operation time for tracing (full-assist function) and appropriate clinical measurements. | Full-assist function for tracing showed reduced operation time compared to the predicate device to obtain EDV/ESV/EF and Global Longitudinal Strain. Clinical evaluation demonstrated appropriate clinical measurements. |
3D ACM | Function correctly measures flow volume and meets performance specifications. | Pulsating flow phantom study demonstrated that flow volume measurement met performance specifications as expected. |
Z-Score Measurement | Function correctly calculates and provides the correct Z-score for cardiac structures (Aortic valve, Pulmonary valve) based on pre-obtained values of femur length (FL), biparietal diameter (BPD), or gestational age (GA) using fetal echocardiography. | Bench study using previously acquired fetal ultrasound data demonstrated correct Z-score calculation for cardiac structures based on provided parameters. |
MPI Measurement | Function correctly measures velocity at an ROI to provide a graph of time variation of the velocity TIC, correctly measures time duration between two points on TIC, and correctly calculates MPI from these durations. | Bench study using Doppler phantom demonstrated correct velocity measurement and TIC generation, correct measurement of time duration, and correct MPI calculation. |
Mitral Valve Analysis (MVA) | Function demonstrates basic function and performance in vivo regarding visualization of mitral valve anatomy, surrounding structures, spatial relationship, and components (annulus, leaflets, segmentation line, commissures). Acceptance also includes image quality, measurement, workflow improvement, and interoperator variability. Measurement items to meet specified criteria compared to CT images. | Side-by-side comparison of 70 patients showed substantial equivalency in basic function, performance, visualization, image quality, measurement, workflow improvement, and interoperator variability. Cardiac phantom study confirmed MVA software met specified criteria for measurement items compared to CT image measurements. |
4D Imaging for 2D Array Transducers | Function correctly depicts shapes and flow within a phantom and performs as intended in clinical images on volunteers. | Bench testing with a phantom demonstrated correct depiction of shapes and flow. Clinical images on volunteers confirmed the function performs as intended. |
Improvements to Existing Features | Improvements to Slice Thickness Control, Auto Volume Measurement, High Frame Rate CHI, 4D CHI, 4D ADF/SMI, iSMI, ECG Sync Acquisition (Shear Wave Elastography), ECG Sync Construction (Sensor 3D), Smart Fusion, 2D WMT (2D Array transducer use), Shear Wave Elastography, and Shear Wave Dispersion Map meet specifications and perform as intended. | Studies demonstrated that these improvements met specifications and performed as intended through phantom and volunteer studies. |
2. Sample Size Used for the Test Set and Data Provenance
- 3D Wall Motion Tracking (RV Analysis): n=11 for the correlation coefficient for volume measurement.
- Mitral Valve Analysis (MVA): Seventy (70) patients were included in a side-by-side comparison study. A cardiac phantom was also used.
- Auto-EF: A group of three volunteers for workflow improvement assessment. A clinical evaluation was also conducted (sample size not specified).
- Z-Score Measurement: Previously acquired fetal ultrasound data. (Sample size not specified)
- 4D Imaging for 2D Array transducers: Phantom studies and volunteer studies (number of volunteers not specified).
- Improvements to existing features: Phantom and volunteer studies (sample sizes not specified).
The data provenance is not explicitly stated as "country of origin" or "retrospective/prospective" for most studies. However, the mention of "previously acquired fetal ultrasound data" for Z-Score implies retrospective data for that specific test. The mention of "volunteers" and "patients" suggests prospective data for other tests.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not provided in the given text.
4. Adjudication Method for the Test Set
This information is not provided in the given text.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
The document does not explicitly describe an MRMC comparative effectiveness study where human readers improve with AI assistance. The testing focuses on the device's performance against predicate devices or pre-defined performance metrics, but not on human-AI collaboration.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
The studies conducted for Auto-EF, 3D ACM, Z-Score Measurement, and MPI Measurement appear to be standalone tests of the algorithm's performance against defined criteria or phantoms. While Auto-EF also involved an assessment of workflow improvement with volunteers, the core performance checks for these features seem to be standalone.
7. The Type of Ground Truth Used
- 3D Wall Motion Tracking (RV Analysis): Comparison to the predicate function (Vivid E9, K131514 with GE EchoPAC, K120221).
- Auto-EF: "Appropriate clinical measurements" as expected.
- 3D ACM: Performance specifications for flow volume measurement using a pulsating flow phantom.
- Z-Score Measurement: Pre-obtained values of femur length (FL), biparietal diameter (BPD), or gestational age (GA) using fetal echocardiography.
- MPI Measurement: Correct measurement of velocity and time durations in a Doppler phantom, and correct MPI calculation according to predefined methods.
- Mitral Valve Analysis (MVA):
- In vivo evaluation: "Basic function and performance" in visualization, measurements, workflow improvement, and interoperator variability using 70 patients.
- Phantom study: Comparison of MVA software output measurements to equivalent parameters obtained from a Computed Tomography (CT) system.
- 4D Imaging for 2D Array transducers: Correct depiction of shapes and flow in a phantom, and performance "as intended" in clinical images on volunteers.
8. The Sample Size for the Training Set
This information is not provided in the given text. The document describes verification and validation testing, but not details about model training.
9. How the Ground Truth for the Training Set was Established
This information is not provided in the given text, as training set details are absent.
Ask a specific question about this device
(31 days)
The device is intended for use by a qualified physician for ultrasound evaluation of Fetal; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transrectal; Transvaginal; Transesophageal and Intraoperative (abdominal, thoracic and vascular).
The LOGIQ E9 is a full featured, general purpose diagnostic ultrasound system which consists of a mobile console approximately 58 cm wide, 87cm deep and 130 cm high that provides digital acquisition, processing and display capability. The user interface includes a computer keyboard, specialized controls, 10-inch LCD touch screen and color wide screen image display.
The provided text is a 510(k) Premarket Notification Submission for the GE LOGIQ E9, LOGIQ E9 XDclear 2.0 Diagnostic Ultrasound System. This document focuses on establishing substantial equivalence to previously cleared devices rather than presenting a new clinical study with acceptance criteria and device performance. Therefore, much of the requested information regarding acceptance criteria and performance studies in the traditional sense of a clinical trial for a novel device is not directly available in this document.
However, I can extract information related to the device's intended use, comparison to predicates, and the non-clinical tests performed.
Important Note: The acceptance criteria and performance data requested in the prompt typically refer to a clinical study demonstrating the efficacy and safety of a novel device. As this submission is for substantial equivalence to existing devices (K142160 GE LOGIQ E9, K131527 GE LOGIQ S8, K131514 GE Vivid E9), the "acceptance criteria" here are met by demonstrating that the new device shares the same intended use, imaging capabilities, technological characteristics, and safety and effectiveness as the predicate devices, and complies with applicable safety standards. There is no new clinical study presented that would define novel acceptance criteria and then report performance against them.
Here's the information based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
As this is a substantial equivalence submission, the "acceptance criteria" are implied by the features and performance of the predicate devices and compliance with relevant standards. The "reported device performance" is the demonstration that the LOGIQ E9, LOGIQ E9 XDclear 2.0 system performs equivalently to the predicates for its intended uses and complies with safety standards.
Acceptance Criteria (Implied by Substantial Equivalence and Standards) | Reported Device Performance (Claimed in Submission) |
---|---|
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body for specified clinical applications (Fetal, Abdominal, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Cardiac, Peripheral Vascular, Musculo-skeletal, Urology, Transrectal, Transvaginal, Transesophageal, Intraoperative). | The device is stated to have the same clinical intended use as the predicate LOGIQ E9. The individual Intended Use Forms for various transducers (e.g., 3CRF, 3CRF-D, 6S-D, 6Tc, 9L-D, 11L-D, C1-5-D, IC5-9-D, L8-18i-D, M4S-D, M5S-D, M6C-D, ML6-15-D, P2D, P6D, RAB2-5-D, RAB4-8-D, RIC5-9-D, RNA5-9-D, RSP6-16-D, S1-5, S4-10, S4-10-D, C2-9-D, RAB6-D, S1-5-D, C1-6-D, C2-9VN-D, C1-6VN-D, C2-7-D, C2-7VN-D, L3-9i-D, C3-10-D, C2-6b-D) detail the specific applications and modes (B, M, PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler, Combined Modes, Harmonic Imaging, Coded Pulse, 3D/4D Imaging, Needle Guidance, Volume Navigation, Elastography Imaging, Shear Wave Elastography). |
Imaging Capabilities: Similar imaging modes and capabilities as predicate devices. | The LOGIQ E9 and predicate LOGIQ E9 systems have the same imaging modes. The LOGIQ E9 and predicate LOGIQ E9 systems have similar capability in terms of performing measurements, capturing digital images, reviewing, and reporting studies. |
Technological Characteristics: Employ the same fundamental scientific technology. | The LOGIQ E9 employs the same fundamental scientific technology as its predicate device(s). New transducers (C2-7-D, C2-7VN-D, C3-10-D, L3-9i-D) are described as "incremental improvements to the existing transducers and are equivalent." Two new software features (AutoEF and AFI) have been migrated from a predicate (Vivid E9 K131514). |
Safety and Effectiveness: Compliance with safety standards and acoustic power limits. | The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness, as well as thermal, electrical, electromagnetic, and mechanical safety. It has been found to conform with applicable medical device safety standards (AAMI/ANSI ES60601-1, IEC60601-1-2, IEC60601-2-37, NEMA UD 3, ISO10993-1, NEMA UD 2, ISO14971). Acoustic power levels are below applicable FDA limits. The system has been designed in compliance with approved electrical and physical safety standards. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This document explicitly states: "The subject of this premarket submission, LOGIQ E9, did not require clinical studies to support substantial equivalence." Therefore, there is no test set, sample size, or data provenance from a new clinical study provided in this submission for device performance evaluation. The substantial equivalence relies on demonstrating that the new device operates similarly and safely as the predicate devices, which were previously cleared.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable, as no new clinical study (and thus no "test set" and "ground truth" to establish) was required for this substantial equivalence submission.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable, as no new clinical study was conducted.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a diagnostic ultrasound system, not an AI-assisted diagnostic tool in the sense of image interpretation. The "AutoEF and AFI" mentioned are software features but there's no MRMC study detailed here to show improved human reader performance.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable, as this is a diagnostic ultrasound imaging system, not a standalone algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Not applicable, as no new clinical study was required for this substantial equivalence submission. The "ground truth" for the predicate devices' clearance would have been established through methods appropriate for diagnostic ultrasound systems at the time of their original clearance (e.g., comparison to other imaging modalities, clinical correlation).
8. The sample size for the training set
Not applicable. This document describes a medical device, not a machine learning algorithm that requires a training set.
9. How the ground truth for the training set was established
Not applicable, as there is no training set for an algorithm described in this submission.
Ask a specific question about this device
(48 days)
The device is a general-purpose ultrasound system, specialized for use in cardiac imaging. It is intended for use by, or under the direction of a qualified physician for ultrasound imaging and analysis of Fetal/Obstetrics; Abdominal (including renal and GYN); Pediatric; Small Organ (breast, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular (PV); Musculo-skeletal Conventional; Urology (including prostate); Transesophagea1; Transrectal (TR); Transvaginal (TV); and Intraoperative (abdominal, thoracic, & vascular).
GE Vivid E80 / E90 / E95 is a Track 3 diagnostic ultrasound system, which is primarily intended for cardiac imaging and analysis, but which also includes vascular and general radiology applications. The Vivid E80 / E90 / E95 incorporates a variety of electronic array transducers operating in linear, curved linear, sector/phased array or matrix array format, including two dedicated CW transducers and several real time 3D transducers. It consists of a mobile console with keyboard control panel; color LCD/TFT touch panel, Alternative OLED or LCD color video display and optional image storage and printing devices. It provides high performance ultrasound imaging and analysis and has comprehensive networking and DICOM capability.
The provided document is a 510(k) Premarket Notification for the GE Vivid E80/E90/E95 Diagnostic Ultrasound System. This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, rather than proving the device meets specific performance criteria through new clinical studies. Therefore, much of the requested information regarding acceptance criteria, specific study design details, and ground truth establishment is not explicitly available in this document.
However, I can extract the following information based on the document's content:
1. Table of Acceptance Criteria and Reported Device Performance:
The document states that the GE Vivid E80/E90/E95 system is substantially equivalent to its predicate devices concerning indications for use, imaging capabilities, technological characteristics, safety, and effectiveness. This "substantial equivalence" acts as the acceptance criterion. The reported device performance is that it meets this criterion by being comparable to the predicate devices.
Acceptance Criterion (Established by Predicate Devices) | Reported Device Performance (GE Vivid E80/E90/E95) |
---|---|
Indications & Clinical Applications: Equivalent to predicate (e.g., Fetal/Obstetrics, Abdominal, Cardiac, Peripheral Vascular, etc.) | Meets: Device has identical indications and clinical applications to the primary predicate, Vivid E9 (K131514). |
Technology: Employs same fundamental scientific technology as predicate. | Meets: Employs the same fundamental scientific technology as its predicate devices. |
Transducer Arrays: Capable of matrix phased, phased, linear, curved, and pencil arrays. | Meets: Supports all specified transducer array types, equivalent to predicate. |
Ultrasound Modes: Capable of B, M, PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler, combined modes, Harmonic, Coded Pulse, Real-time 3D & Multi-plane. | Meets: Supports all specified ultrasound modes, equivalent to predicate. |
Processing & Display Features: Equivalent image mapping, filtering, enhancement, TGC, TVI, SI/SRI, TSI, Pulsatile-Flow, B-Flow, Extended FOV, Tissue Tracking, Real-time 3D, Real-time 3D Color Flow, BFI, Spatial Compounding, Slice View, Laser-Lines, CTO, AFI, Multiplane AFI, LCD/OLED display, StereoVision, Stress, HDLive, Virtual Apex, 4D Auto AVQ. | Meets: Incorporates equivalent processing and display features, with OLED display as an alternative to LCD (predicate only had LCD). |
Standards Compliance: Designed and verified to meet harmonized standards for Electrical Safety, EMC, Biocompatibility, Software Life Cycle Management, Risk Management, and Usability. Track 3 compliance (within FDA limits). | Meets: Designed and verified to meet all listed harmonized standards. Is Track 3 compliant. |
Safety and Effectiveness: Demonstrated to be as safe and as effective as predicate devices. | Meets: Considered as safe and as effective as the predicate devices. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document primarily relies on comparisons to predicate devices and adherence to recognized standards. It explicitly states: "The subject of this premarket submission, Vivid E80 / E90 / E95, did not require clinical studies to support substantial equivalence."
Therefore, there is no specific test set or data provenance details relating to clinical performance provided in this document as new clinical studies were not deemed necessary for this 510(k) submission. The evidence is based on demonstrating the new device performs equivalently to existing cleared devices.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Since new clinical studies were not required, there is no information on experts establishing ground truth for a test set. The predicate devices would have undergone their own validation for their initial clearance.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable, as no new clinical test set was required.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a diagnostic ultrasound system, and the submission is for showing substantial equivalence to predicate ultrasound systems, not for evaluating AI assistance or human reader improvement with AI.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a diagnostic ultrasound system that functions with a human operator for image acquisition and interpretation.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Not applicable, as no new clinical studies requiring external ground truth were conducted for this 510(k). The "ground truth" for demonstrating substantial equivalence is the established performance and safety profiles of the predicate devices.
8. The sample size for the training set
Not applicable, as this is related to a new clinical study or AI development, neither of which were required for this 510(k) submission.
9. How the ground truth for the training set was established
Not applicable, for the same reasons as above. The submission focuses on demonstrating the device's technical and functional equivalence to already cleared devices.
Ask a specific question about this device
Page 1 of 1