OnSite-IV
K161800 · Pharmc, LLC · LHI · Dec 6, 2016 · General Hospital
Device Facts
| Record ID | K161800 |
| Device Name | OnSite-IV |
| Applicant | Pharmc, LLC |
| Product Code | LHI · General Hospital |
| Decision Date | Dec 6, 2016 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 880.5440 |
| Device Class | Class 2 |
Intended Use
The OnSite-IV™ is a single use sterile device indicated for the preparation and / or transfer of drugs from a standard pharmaceutical vial into an elastomeric pump or IV bag.
Device Story
OnSite-IV is a sterile, single-use, manually operated device; integrates a syringe barrel, stopper, and plunger with a vial spike. Used by healthcare professionals in clinical settings to reconstitute or mix drugs from standard 20mm pharmaceutical vials. Operation: user connects device to vial; manually agitates to mix; aspirates prepared drug into syringe barrel; transfers drug into IV bag or elastomeric pump. Features bacterial retentive air-venting filter. Replaces multi-component systems with an integrated design. Benefits include simplified drug preparation and transfer process, reducing potential for contamination or handling errors.
Clinical Evidence
Bench testing only. Testing included ISO 7886-1 (syringes), ISO 594-1/2 (Luer connections), ASTM F2101 (bacterial retentive filter), and USP <788> (particulate matter). Biocompatibility testing performed per ISO 10993-1, including cytotoxicity, sensitization, irritation, acute systemic toxicity, hemolysis, and pyrogenicity. Performance verified after sterilization, simulated transport, and accelerated aging.
Technological Characteristics
Integrated syringe/vial spike device. Materials: plastic injection molded components. Features: elliptical syringe barrel, elastomeric stopper (acts as valve), Luer lock connector, bacterial retentive air-venting filter. Manual operation. Sterilization: VDmax 25 method (ANSI/AAMI/ISO 11137-2). Standards: ISO 7886-1, ISO 594-1/2, ISO 10993, ASTM F756, ASTM F2101, USP <788>, USP <151>.
Indications for Use
Indicated for healthcare professionals (physicians, nurses, pharmacists) for the preparation and/or transfer of drugs from a standard pharmaceutical vial into an elastomeric pump or IV bag.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
- E-Z-Link (K133097)
- IV Fluid Transfer Pin (K925401)
Related Devices
- K240940 — Vial2Bag Advanced® 20mm Admixture Device · West Pharmaceutical Services, Inc. · May 3, 2024
- K201415 — Vial2Bag Advanced 20mm Admixture Device · West Pharmaceutical Services Az, Inc. · Oct 21, 2020
- K230988 — Vial2Bag Advanced® 13mm Admixture Device · West Pharma. Services IL, Ltd. · Dec 1, 2023
- K170095 — Vial2Bag Direct Connect · Medimop Medical Project , Ltd. · Mar 8, 2017
- K212525 — Advcare Vial Direct to Bag (VDB) Needle-free Admixture Devices · Advcare Medical · Aug 19, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 6, 2016
PharmaC, LLC % Sigi Caron Medtech Consultants. Inc. 20370 Skyhawk Lane Topanga, California 90290
Re: K161800
Trade/Device Name: OnSite-IVTM Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: LHI Dated: November 7, 2016 Received: November 8, 2016
Dear Sigi Caron:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/8 description: The image contains a signature and the name "Tina Kiang". The signature is a complex, looping design to the left of the name. Below the name is the character string "-s".
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161800
Device Name OnSite-IVTM
Indications for Use (Describe)
The OnSite-IV™ is a single use sterile device for the preparation and / or transfer of drugs from a standard pharmaceutical vial into an elastomeric pump or IV bag.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(K) SUMMARY - K161800
This 510(k) summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| DATE PREPARED | December 6, 2016 |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| APPLICANT | PharmaC, LLC.<br>6655 North Vista Lane<br>Jackson, WY 83001<br>Tel: (307) 690-3648 |
| OFFICIAL CORRESPONDENT | Sigi Caron<br>Vice President, Regulatory and Clinical Affairs, PharmaC,<br>LLC.<br>Principal, MedTech Consultants, Inc.<br>20370 Skyhawk Lane<br>Topanga, CA 90290<br>sigi@medtechconsultants.com<br>Tel: (310) 455-3473<br>Fax: (888) 295-1535 |
| TRADE NAME | OnSite-IVTM |
| COMMON NAME | Vial Adapter / Reconstitution Device |
| DEVICE CLASSIFICATION | Regulation Number - 21 CFR §880.5440<br>Regulation Name - Intravascular administration set<br>Product Code LHI, Class-II |
| PREDICATE DEVICE | E-Z-Link - K133097<br>IV Fluid Transfer Pin - K925401 |
### DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:
The OnSite-IV™ device is a sterile, single use device that, through the use of an integrated vial spike, facilitates creation of a sterile fluid path to transfer liquid from the syringe body to a standard pharmaceutical vial with a 20 mm cap containing a drug, in order to mix or reconstitute the drug and aspirate and transfer the prepared drug back into the syringe body for delivery into an IV bag or elastomeric pump. The device is intended to be used by health care professionals (HCPs) such as physicians, nurses and pharmacists in a clinical setting.
### INDICATIONS FOR USE:
The OnSite-IV™ is a single use sterile device indicated for the preparation and / or transfer of drugs from a standard pharmaceutical vial into an elastomeric pump or IV bag.
{4}------------------------------------------------
# SUBSTANTIALLY EQUIVALENT TO:
The OnSite-IV™ device is substantially equivalent in intended use and technological features to the E-Z Link device (K133097) and the IV Fluid Transfer Pin (K925401).
# TECHNICAL CHARACTERISTICS:
The design and handling characteristics of the OnSite-IV are based on a standard injection syringe. The OnSite-IV consists of three main components: Syringe Barrel, Stopper, and Plunger with integrated vial spike (IV Fluid Transfer Pin - K925401). The Syringe Barrel has an elliptical shape with a Luer lock connector and is printed with graduation markers to indicate volume. The Syringe Barrel also has an arrow marker to indicate rotational position. The Stopper is an elastomeric sealing material that has an elliptical shape to prevent rotation within the Syringe Barrel. Rotation of the Stopper is prevented to allow the stopper to act as a valve. The Stopper has a fluid pathway and features that create a valve at the interface between the Plunger and Stopper. The Plunger has an internal fluid pathway along its length and an arrow marker to indicate rotational position. The integrated vial spike on the Plunger utilizes a bacterial retentive air-venting filter.
The OnSite-IV™ device shares the same design features as the predicate devices and does not introduce any new technological characteristics. The indications for use are similar. Both are intended for use to mix drugs in a standard pharmaceutical vial and to transfer the prepared drug back into the syringe. The OnSite-IV™ is used to transfer prepared drug into an IV bag or an elastomeric pump. This additional capability does not impact the performance of the OnSite-IVTM. The following table compares the feature of the proposed device and its predicates.
| ELEMENT | ONSITE-IV™ | E-Z-LINK<br>K133097 | IV FLUID<br>TRANSFER PIN<br>K925401 |
|-----------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The OnSite-IV™ is a<br>sterile, single use<br>device indicated for<br>the preparation and /<br>or transfer of drugs<br>from a standard<br>pharmaceutical vial<br>into an elastomeric<br>pump or IV bag. | Single use, sterile<br>device for<br>preparation or drugs<br>in a standard vial<br>using liquid from a<br>standard luer<br>syringe to transfer<br>the prepared drug<br>back into the<br>syringe for delivery. | Vented with<br>security clip for<br>preparing and<br>dispensing<br>diluent from<br>standard 20 mm<br>rubber-<br>stoppered vials |
| Design Features | | | |
| Has vial adapter / vial<br>access component | yes | yes | yes |
| Has a syringe component | yes | yes | No syringe<br>component |
| ELEMENT | ONSITE-IVTM | E-Z-LINK<br>K133097 | IV FLUID<br>TRANSFER PIN<br>K925401 |
| Syringe and vial adapter<br>integrated into one<br>component | yes | No, 2 component<br>system | No. Only vial<br>adapter |
| Needleless access to vial | yes | No, contains plastic<br>and stainless steel<br>cannulas | yes |
| Sterile, biocompatible<br>fluid path | yes | yes | yes |
| Assembled from plastic<br>injection molded<br>components | yes | yes | yes |
| Single use, sterile | yes | yes | yes |
| <b>Principles of Operation</b> | | | |
| Manually operated | yes | yes | yes |
| • mechanically<br>connected to a drug<br>vial | yes | yes | yes |
| • transfers liquid<br>manually using a<br>syringe plunger rod | yes | yes | No. Not<br>supplied with<br>syringe |
| • mixing / drug<br>reconstitution<br>achieved through<br>manual agitation of<br>the vial while<br>connected to device | yes | yes | yes |
| • mixed drug is<br>manually aspirated<br>back into the syringe<br>barrel using the<br>syringe plunger rod | yes | yes | No. Not<br>supplied with<br>syringe |
{5}------------------------------------------------
# SUMMARY OF NONCLINICAL TESTING:
Design verification & validation testing was conducted per FDA recognized standards and additional device specific performance requirements. Testing per ISO 7886-1:1993 - Sterile hypodermic syringes for single use -- Part 1: Syringes for manual use included:
- Limits for Acidity or Alkalinity ●
- . Limits for Extractable Metals
- Tolerance on Graduated Capacity ●
{6}------------------------------------------------
- Dead Space ●
- Liquid Leakage at Syringe Piston under Compression
- Air Leakage at Syringe Piston under Compression ●
- Air Leakage past Syringe Piston during Aspiration
- Fiducial Line
- . Fit of Piston in Barrel
Sterilization of the device was validated in accordance with the VDmax 25 method described by ANSI/AAMI/ISO 11137-2:2013 to establish a 106 sterility assurance level (SAL).
Biocompatibility testing was conducted based on the nature and duration of contact, the construction of the device, and both the ISO 10993-1 recommendations and the FDA Guidance for Use of ISO10993-1:
- Cytotoxicity Study using the ISO Elution Method (ISO 10993-5) ●
- Maximization Sensitization (ISO 10993-10)
- Intracutaneous Reactivity Irritation Test in Rabbits (ISO 10993 Part 10)
- Acute Systemic Toxicity Test in Mice (ISO 10993-11)
- Hemolysis, Extract Method (ASTM F 756)
- Hemolysis, Direct Method (ASTM F 756)
- Materials Mediated Pyrogen Test in Rabbits (ISO 10993-11, USP <151>)
Additional performance testing of the device included:
- Luer Connection Performance (ISO 594-1, ISO 594-2) ●
- Particulate Matter (USP <788>)
- . Performance of the Bacterial Retentive Filter (ASTM F2101)
Further performance-based testing to internal requirements was also conducted after sterilization, simulated transportation conditioning and accelerated aging:
- Visual Inspection ●
- Device Retained Volume After Use
- Drug Mixing
- Force to Operate Valve ●
- Graduated Marking Durability
All features provided for the OnSite-IV™ have been verified to operate as specified. Testing confirms that the OnSite-IV™ can be used according to its intended use and in an equivalent manner to the predicate devices.
{7}------------------------------------------------
# CONCLUSIONS - BASIS FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE:
The OnSite-IV™ is substantially equivalent to the listed predicate devices (E-Z Link device, K133097 and IV Fluid Transfer Pin - K925401) with respect to its indications for use (intended use) and technical characteristics. The differences between the E-Z Link and the OnSite-IV™ are that the E-Z link uses plastic and stainless steel cannulas to access the drug vial, and the syringe and adapter are provided as separate components. The OnSite-IV™, on the other hand, utilizes an integrated universal vial spike adapter (one integrated device, not two components). Testing per FDA recognized consensus standards, other standards and device specific bench testing demonstrate that the OnSite-IV™ functions as intended in the specific use conditions. The information and data provided in this 510(k) submission identifies that the subject device is substantially equivalent to the predicate devices.