K Number
K161570

Validate with FDA (Live)

Date Cleared
2016-08-29

(83 days)

Product Code
Regulation Number
878.4780
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Theia NPWT Foam Wound Dressing Kit is intended to be used in conjunction with the Simex Negative Pressure Wound Therapy Pumps (K113291) for the application of negative pressure wound therapy to the wound. When used in conjunction with the Simex Negative Pressure Wound Therapy Pumps, the Theia NPWT Foam Wound Dressing Kit is indicated for patients who would benefit from a suction device, particularly as the device may promote wound healing by removal of excess exudates, infectious material and tissue debris.

The Theia NPWT Foam Wound Dressing Kit is appropriate for use on the following wounds;

  • Pressure Ulcers
  • · Diabetic/Neuropathic Ulcers
  • Venous Insufficiency Ulcers
  • · Traumatic Wounds
  • · Post-Operative and Dehisced Surgical Wounds
  • · Skin Flap and Grafts
Device Description

The Theia NPWT Foam Wound Dressing Kit without the pump includes a foam dressing composed of a reticulated flexible polyether based polyurethane hydrophobic foam material, an occlusive drape and silicon suction dome with negative pressure tubing. Theia NPWT Foam Wound Dressing Kits are available in three sizes; 1) small, 2) medium and 3) large.

AI/ML Overview

The provided text is a 510(k) Summary for the Theia NPWT Foam Wound Dressing Kit. It describes the device and its claimed substantial equivalence to a predicate device, the UNI NPWT Foam Wound Dressing Kit (K133333).

Based on the document, the following can be extracted:

  1. A table of acceptance criteria and the reported device performance

The document does not specify quantitative acceptance criteria or a direct performance table for the Theia NPWT Foam Wound Dressing Kit against such criteria. Instead, it relies on substantial equivalence to a predicate device. The "Technical Characteristics: Feature Comparison Chart" (page 4) lists features of both the proposed device and the predicate device, showing identical characteristics across all listed categories. This implies that the 'acceptance criteria' are met by demonstrating that the device is identical in all relevant technical characteristics to the legally marketed predicate device.

Feature / Acceptance Criteria (implied by predicate)Theia NPWT Foam Wound Dressing Kit Performance (Reported)
Product Code: OMPOMP
Technology/FunctionIdentical to predicate
Foam Dressing MaterialFlexible Polyether Polyurethane Foam
HydrophobicYes
Dome Assembly ComponentsDome, Skirt, and Tubing
Dome MaterialThermoplastic Elastomer
Skirt MaterialPolyurethane Medical Tape with Adhesive Backing
Tubing MaterialPVC
Occlusive Drape MaterialSemipermeable, polyurethane (polymeric) transparent film
Foam Dressing Dimensions (Small)10 x 8 x 3 cm
Foam Dressing Dimensions (Medium)20 x 13 x 3 cm
Foam Dressing Dimensions (Large)25 x 16 x 3 cm
Used in conjunction with NPWT pumpsFor use with Simex NPWT Pumps (K113291)
NPWT pump provided with Dressing KitNo
Provided SterileYes
Single Use OnlyYes
BiocompatibleYes
  1. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document explicitly states: "No clinical study was conducted" (page 5). Therefore, there is no test set, no sample size, and no data provenance from clinical data for this specific device. The device's safety and effectiveness are established through non-clinical testing and comparison to a predicate device. The non-clinical testing for the Theia NPWT Foam Wound Dressing Kit is stated to be "identical to the testing conducted in K133333 for the predicate device." However, the results or details of that non-clinical testing (including sample sizes, if applicable) are not provided in this document.

  1. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable, as no clinical study with a test set was conducted.

  1. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable, as no clinical study with a test set was conducted.

  1. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a medical device (wound dressing kit) and not an AI-powered diagnostic or interpretive tool that would involve human readers or AI assistance.

  1. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a medical device (wound dressing kit) and not an algorithm.

  1. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

Given that no clinical study was conducted, there is no ground truth established from clinical data for this device. The basis for acceptance is substantial equivalence to a legally marketed predicate device, meaning the 'ground truth' for its performance and safety is implicitly derived from the predicate's regulatory clearance and presumed safe and effective use. The non-clinical testing for the predicate is stated to be the basis for this device.

  1. The sample size for the training set

Not applicable, as no clinical study was conducted and there is no AI component requiring a training set.

  1. How the ground truth for the training set was established

Not applicable, as no clinical study was conducted and there is no AI component requiring a training set.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 29, 2016

LC Medical Concepts, Inc. Ms. Dana M. Ledgerwood Chief Executive Officer P.O. Box 502 Penfield, New York 14526

Re: K161570

Trade/Device Name: Theia NPWT Foam Wound Dressing Kit Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: OMP Dated: May 1, 2016 Received: June 7, 2016

Dear Ms. Ledgerwood:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

David Krause -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K161570

Device Name Theia NPWT Foam Wound Dressing Kit

Indications for Use (Describe)

The Theia NPWT Foam Wound Dressing Kit is intended to be used in conjunction with the Simex Negative Pressure Wound Therapy Pumps (K113291) for the application of negative pressure wound therapy to the wound. When used in conjunction with the Simex Negative Pressure Wound Therapy Pumps, the Theia NPWT Foam Wound Dressing Kit is indicated for patients who would benefit from a suction device, particularly as the device may promote wound healing by removal of excess exudates, infectious material and tissue debris.

The Theia NPWT Foam Wound Dressing Kit is appropriate for use on the following wounds;

  • Pressure Ulcers
  • · Diabetic/Neuropathic Ulcers
  • Venous Insufficiency Ulcers
  • · Traumatic Wounds
  • · Post-Operative and Dehisced Surgical Wounds
  • · Skin Flap and Grafts

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over The Counter Use (21 CFR 801 Subpart C)
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

[X] Prescription Use (Part 21 CFR 801 Subbart DI ounter Use (21 CFR 801 Subbart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Date Prepared:August 24, 2016
Sponsor:LC Medical Concepts, Inc.PO Box 502Penfield, NY 14526
510(k) Contact:Ms. Dana LedgerwoodPO Box 502Penfield, NY 14526Telephone: 585-203-7652e-mail: dm@lcmedicalconcepts.com
Trade Name:Theia NPWT Foam Wound Dressing Kit
Common Name:Foam Dressing Kit
Classification:Powered Suction Pump21 CFR 878.4780Class II
Product Code:OMP - Pump, Portable, Aspiration (manual or powered)
Panel:General & Plastic Surgery
Predicate Device:UNI NPWT Foam Dressing Kit (K133333)

Indications For Use:

The Theia NPWT Foam Dressing Kit is intended to be used in conjunction with the Simex Negative Pressure Wound Therapy Pumps (K113291) for the application of negative pressure wound therapy to the wound. When used in conjunction with the Simex Negative Pressure Wound Therapy Pumps, the Theia NPWT Foam Dressing Kit is indicated for patients who would benefit from a suction device, particularly as the device may promote wound healing by removal of excess exudates, infectious material and tissue debris.

The Theia NPWT Foam Dressing Kit is appropriate for use on the following wounds:

  • Pressure Ulcers ●
  • Diabetic/Neuropathic Ulcers
  • Venous Insufficiency Ulcers
  • . Traumatic Wounds
  • Post-Operative and Dehisced Surgical Wounds .
  • . Skin Flap and Grafts

Device Description:

The Theia NPWT Foam Wound Dressing Kit without the pump includes a foam dressing composed of a reticulated flexible polyether based polyurethane hydrophobic foam material, an

{4}------------------------------------------------

occlusive drape and silicon suction dome with negative pressure tubing. Theia NPWT Foam Wound Dressing Kits are available in three sizes; 1) small, 2) medium and 3) large.

Technical Characteristics:

Technical Characteristics:Feature Comparison Chart
FeatureTheia NPWT Foam Wound Dressing KitUNI NPWT Foam Wound Dressing KitK133333
Indications For UseThe Theia NPWT Foam Wound Dressing Kit isintended to be used in conjunction with the SimexNegative Pressure Wound Therapy Pumps (K113291)for the application of negative pressure wound therapyto the wound. When used in conjunction with theSimex Negative Pressure Wound Therapy Pumps, theTheia NPWT Foam Wound Dressing Kit is indicatedfor patients who would benefit from a suction device,particularly as the device may promote wound healingby removal of excess exudates, infectious material andtissue debris.The Theia NPWT Foam Wound Dressing Kit isappropriate for use on the following wounds:Pressure Ulcers Diabetic/Neuropathic Ulcers Venous Insufficiency Ulcers Traumatic Wounds Post - Operative and Dehisced SurgicalWounds Skin Flap and GraftsThe UNI NPWT Foam Dressing Kit is intended tobe used in conjunction with the Simex NegativePressure Wound Therapy Pumps (K113291) for theapplication of pressure wound therapy to thewound. When used in conjunction with the SimexNegative Pressure Wound Therapy Pumps, theUNI NPWT Foam Dressing Kit is indicated forpatients who would benefit from a suction device,particularly as the device may promote woundhealing by removal of excess exudates, infectiousmaterial and tissue debris.The UNI NPWT Foam Dressing Kit is appropriatefor use on the following wounds:Pressure Ulcers Diabetic/Neuropathic Ulcers Venous Insufficiency Ulcers Traumatic Wounds Post - Operative and Dehisced SurgicalWounds Skin Flap and Grafts
Product CodeOMPOMP
TechnologyThe foam wound dressing functions as thedressing material used to pack the wound.The dome aids in removal of fluids/exudate fromthe wound to the collection canister of the NPWTsuction pump.The tubing as part of the dome assembly aids inthe removal of fluids/exudates from the wound tothe collection canister while maintaining a sealedapplication.The occlusive drape is a semipermeable,transparent sheet applied over the foam dressingto cover the peri wound area and ensure a properseal.The foam wound dressing functions as thedressing material used to pack the wound.The dome aids in removal of fluids/exudatefrom the wound to the collection canister ofthe NPWT suction pump.The tubing as part of the dome assembly aidsin the removal of fluids/exudates from thewound to the collection canister whilemaintaining a sealed application.The occlusive drape is a semipermeable,transparent sheet applied over the foamdressing to cover the peri wound area andensure a proper seal.
Foam DressingMaterialFlexible Polyether Polyurethane FoamFlexible Polyether Polyurethane Foam
HydrophobicYesYes
Dome Assembly:Consists of Dome, Skirt and TubingConsists of Dome, Skirt and Tubing
Dome Material: Thermoplastic ElastomerDome Material: Thermoplastic Elastomer
Skirt Material: Polyurethane Medical Tape withAdhesive BackingSkirt Material: Polyurethane Medical Tape withAdhesive Backing
Tubing Material: PVCTubing Material: PVC
Occlusive DrapeSemipermeable, polyurethane (polymeric) transparentfilmSemipermeable, polyurethane (polymeric)transparent film
Foam DressingDimensions:Small 10 x 8 x 3 cmSmall 10 x 8 x 3 cm
Medium 20 x 13 x 3 cmMedium 20 x 13 x 3 cm
Large 25 x 16 x 3 cmLarge 25 x 16 x 3 cm
Used in conjunctionwith NPWT pumpsFor use with Simex NPWT Pumps (K113291)For use with Simex NPWT Pumps (K113291)
NPWT pumpprovided withNoNo
Dressing Kit
Provided SterileYesYes
Single Use OnlyYesYes
BiocompatibleYesYes

{5}------------------------------------------------

Non-clinical Testing:

The non - clinical testing for the Theia NPWT Foam Wound Dressing Kit is identical to the testing conducted in K133333 for the predicate device.

Clinical Performance Data:

No clinical study was conducted.

Substantial Equivalence:

The Theia NPWT Foam Wound Dressing Kit is substantially equivalent to the currently marketed UNI NPWT Foam Wound Dressing Kit, (K133333, Blue Ocean Medical Products, LLC). Both of these NPWT foam wound dressing kits are intended to be used in conjunction with the Simex Negative Pressure Wound Therapy Pumps (K113291) for the application of negative pressure wound therapy to the wound. Both the Theia NPWT Foam Wound Dressing and the UNI NPWT Foam Wound Dressing Kits are indicated for patients who would benefit from a suction device, particularly as the device may promote wound healing by removal of excess exudates, infectious material and tissue debris.

Both the Theia NWPT Foam Wound Dressing Kits and the predicate dressing kits contain the same components, namely a foam wound dressing, a dome assembly consisting of a dome port, skirt and tubing, and an occlusive drape. Each of the kit components in the Theia NPWT dressing kits have the same design and are manufactured of the same materials as the UNI NPWT dressing kit components. The Theia NPWT Foam Wound Dressing kits are provided in Small, Medium and Large, determined by the dimensions of the foam wound dressing contained in the kit, the same sizes provided in the predicate, UNI NPWT Foam Wound Dressing kits.

Conclusion:

The Theia Foam Wound Dressing Kits are substantially equivalent to the currently marketed UNI NPWT Foam Wound Dressing Kit (K133333) having the same indications for use, technological characteristics, design and materials, and do not raise new issues of safety and effectiveness.

§ 878.4780 Powered suction pump.

(a)
Identification. A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.(b)
Classification. Class II.