K Number
K161295

Validate with FDA (Live)

Device Name
MiniBox +
Date Cleared
2016-10-21

(165 days)

Product Code
Regulation Number
868.1880
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PulmOne MiniBox+ is intended to measure lung function in adult and pediatric patients, 5 years and older, while at rest (including spirometry, lung volumes and diffusing capacity). The PulmOne MiniBox+ is to be used by a physician, respiratory therapist, or technician in a hospital or clinic setting

Device Description

The PulmOne MiniBox+ is intended to measure lung function in adult and pediatric patients while at rest (including spirometry, lung volumes and diffusing capacity). The PulmOne MiniBox+ is to be used by either a physician, respiratory therapist, or technician. A cleared single-use, disposable viral-bacterial filter separates the patient from the internal components of the device.

The MiniBox+ measures all common spirometric measurements as well as relative and absolute lung volumes and diffusing capacity.

The MiniBox+ is equipped with a diffusing capacity module which is embedded into the MiniBox+ enclosure and located in the upper end of the device. This feature allows easy measurement of the volumes and concentrations (CO and CH4) of inspired and expired breath, of the test gas mixture and patient sample respectively, making possible the calculation of the diffusion capacity [DLco]. MiniBox + DLco [diffusing capacity factor of the lung for carbon monoxide (CO)] module uses the single breath concept and enables the measurements by setting and controlling the maneuver stages as defined in ATS/ETS DLco guidelines.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the PulmOne MiniBox+ device, based on the provided document:

Acceptance Criteria and Device Performance

The document describes the performance and technical specifications of the MiniBox+ and compares them to predicate devices. The MiniBox+ is a pulmonary function data calculator, and its performance is evaluated against established standards for spirometry, lung volumes, and diffusing capacity measurements.

Table of Acceptance Criteria and Reported Device Performance

ParameterAcceptance Criteria (from Predicate/Standards)Reported Device Performance (MiniBox+)Notes
SpirometryIdentical to MiniBoxPFT™ 2.0 (K141793)
Flow range+/- 16 L/s (MiniBoxPFT™ 2.0)+/- 16 L/sSame as predicate.
Volume accuracy+/- 3% or 50mL (MiniBoxPFT™ 2.0)+/- 3% or 50mLSame as predicate.
Flow accuracy+/- 5% or 200mL/s (MiniBoxPFT™ 2.0)+/- 5% or 200mL/sSame as predicate.
Dynamic resistance at 14 L/s< 0.7 cmH2O/(L/s) (MiniBoxPFT™ 2.0)< 0.7 cmH2O/(L/s)Same as predicate.
DLco (Diffusing Capacity)Equivalent to ZAN LUNG-FUNCTION LAB (K052328)
CO Analyzer Measurement range0 to 3000 ppm (ZAN 500)0 to 3000 ppmIdentical.
CO Analyzer Accuracy< 1% (ZAN 500)< 1%Identical.
CO Analyzer Neutral point drift< 2% per week (ZAN 500)< 2% per weekIdentical.
CO Analyzer Linearity< 1% (ZAN 500)< 1%Identical.
CH4 Analyzer Measurement range0 to 3000 ppm (ZAN 500)0 to 3000 ppmIdentical.
CH4 Analyzer Accuracy1% (ZAN 500)2.5%The document states this difference is "Equivalent" because DLco is computed by the ratio of CH4-in/CH4-out, and linearity is more important. Linearity was tested and verified by the OEM manufacturer and through extensive tests versus a DLco gas simulator. Human measurements and comparison to the ZAN 500 found no adverse impact on substantial equivalence.
CH4 Analyzer Neural point drift< 2% per week (ZAN 500)< 2% per weekIdentical.
CH4 Analyzer Linearity< 1% (ZAN 500)< 1%Identical.
Test gas mixture0.3% Carbon monoxide (CO), 0.3% Methane (CH4), 21% Oxygen (O2) with balance Nitrogen (N2) (ZAN 500)0.3% Carbon monoxide (CO), 0.3% Methane (CH4), 21% Oxygen (O2) with balance Nitrogen (N2)Identical.
Demand ValveOEM by MSA (ZAN 500)OEM by MSAEquivalent (same component).
Lung Volume MeasurementIdentical to MiniBoxPFT™ 2.0 (K141793)
Measurements (Mouth pressure and flow)Mouth pressure and mouth flow (MiniBoxPFT™ 2.0)Mouth pressure and mouth flowIdentical.
Calculations (TLC, TGV (FRC), RV, VC, IC and ERV)TLC, TGV (FRC), RV, VC, IC and ERV (MiniBoxPFT™ 2.0)TLC, TGV (FRC), RV, VC, IC and ERVIdentical.
Air Container15L elliptic container (MiniBoxPFT™ 2.0)15L elliptic containerIdentical.

Study Information

  1. Sample size used for the test set and the data provenance:

    • DLco human comparative testing: Not specified in the provided text.
    • DLco human reproducibility testing: Not specified in the provided text.
    • No other "test set" sample sizes or data provenance (e.g., country of origin, retrospective/prospective) are mentioned for non-clinical or clinical studies in this document. The focus of the non-clinical testing appears to be on simulator and comparative studies rather than a large clinical test set.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not specified. The document mentions "DLco human comparative testing" and "DLco human reproducibility testing," but does not detail how ground truth was established for these or the number/qualifications of experts involved.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not specified.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC study was mentioned. The device is a "Pulmonary-Function Data Calculator," and the documentation does not describe an AI component that would assist human readers in interpreting images or data in an MRMC setting. It's for direct measurement and calculation.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • The document implies standalone performance testing for the device's measurement accuracy, particularly for DLco using a simulator ("DLco performance verification against simulator [Hans Rudolf]"). However, the "human comparative testing" and "human reproducibility testing" suggest interaction with humans, but not necessarily in an AI-assistive role. The core function is measurement and calculation, which is inherently "standalone" in its data generation, regardless of human interaction for the test.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the DLco gas analyzer, ground truth was established through OEM manufacturer verification (linearity) and testing against a DLco gas simulator [Hans Rudolf, Inc.].
    • For human measurements, the ground truth was implied to be established through internal repeatability measurements and comparison against the predicate device, ZAN Function Lab [ZAN-500]. This suggests that the ZAN 500's measurements served as a de-facto ground truth for comparative purposes.
  7. The sample size for the training set:

    • Not applicable as this device is a measurement and calculation device, not a machine learning model that requires a training set in the conventional sense. Its "training" or calibration would involve ensuring its sensors and algorithms accurately reflect physical properties.
  8. How the ground truth for the training set was established:

    • Not applicable for the same reasons as above. The internal algorithms are based on established scientific principles of pulmonary function testing (e.g., ATS/ETS guidelines, Boyle's law for lung volume calculation) and calibrated against known standards and other validated devices.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 21, 2016

PulmOne Advanced Medical Devices, Ltd % Dr. Susan Alpert Principal SFA Consulting, LLC 200 Park Avenue, Unit 111 Minneapolis, Minnesota 55415

Re: K161295

Trade/Device Name: MiniBox + Regulation Number: 21 CFR 868.1880 Regulation Name: Pulmonary-Function Data Calculator Regulatory Class: Class II Product Code: BZC, BZG Dated: September 20, 2016 Received: September 22, 2016

Dear Dr. Susan Alpert:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K161295

Device Name MiniBox+

Indications for Use (Describe)

The PulmOne MiniBox+ is intended to measure lung function in adult and pediatric patients, 5 years and older, while at rest (including spirometry, lung volumes and diffusing capacity). The PulmOne MiniBox+ is to be used by a physician, respiratory therapist, or technician in a hospital or clinic setting

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------


5 SECTION 5 – 510(K) SUMMARY

5.1 Administrative Information

Date:15-October-2016
Submitter:PulmOne Advanced Medical Devices, Ltd.14 HaMelacha Street, Ra'ananna 4366107 , ISRAELTel: +972 77 5100 938Fax: +972 77 5100 939
Official Correspondent:Avi Lazar, CEO
Trade Name:MiniBox+
Classification Name:Classification Number:Product Code:Device Class:Pulmonary-function data calculator21 CFR 868.1880BZC, BZGClass II
Predicate Devices:Primary Predicate:PulmOne MiniBoxPFT™ 2.0510(k) Number -K141793Secondary Predicate:ZAN LUNG-FUNCTION LAB510(k) Number - K052328
Application Correspondent:Dr. Susan AlpertSFA Consulting, LLC200 Park Ave., unit 111Minneapolis, MN 55415Phone- 612-202-7019drsusanalpert@gmail.com

{4}------------------------------------------------

5.2 DEVICE DESCRIPTION

The PulmOne MiniBox+ is intended to measure lung function in adult and pediatric patients while at rest (including spirometry, lung volumes and diffusing capacity). The PulmOne MiniBox+ is to be used by either a physician, respiratory therapist, or technician. A cleared single-use, disposable viral-bacterial filter separates the patient from the internal components of the device.

The MiniBox+ measures all common spirometric measurements as well as relative and absolute lung volumes and diffusing capacity.

The MiniBox+ is equipped with a diffusing capacity module which is embedded into the MiniBox+ enclosure and located in the upper end of the device. This feature allows easy measurement of the volumes and concentrations (CO and CH4) of inspired and expired breath, of the test gas mixture and patient sample respectively, making possible the calculation of the diffusion capacity [DLco]. MiniBox + DLco [diffusing capacity factor of the lung for carbon monoxide (CO)] module uses the single breath concept and enables the measurements by setting and controlling the maneuver stages as defined in ATS/ETS DLco guidelines.

5.3 INTENDED USE AND INDICATIONS FOR USE

The PulmOne MiniBox+ is intended to measure lung function in adult and pediatric patients, 5 years and older, while at rest (including spirometry, lung volumes and diffusing capacity). The PulmOne MiniBox+ is to be used by a physician, respiratory therapist, or technician in a hospital or clinic setting.

5.4 Summary of Technical Characteristics

The MiniBox+ has been updated from the predicate version K141793. There have been no changes to the general intended use of the device to measure lung function or to the fundamental scientific technology. The Indications for use were updated to include the diffusing capacity module which is substantially equivalent to the one in ZAN 500 [K052328] and both utilize the same technologies and methods of operations. In addition, the software, firmware and hardware for the MiniBox+ in this submission has been updated to introduce minor software, firmware and hardware modifications to support this change.

{5}------------------------------------------------

5.4.1 Summary table of Comparison

SpecificationMiniBox+MiniBoxPFT™ 2.0ZAN LUNG-FUNCTION LAB(ZAN 500 BodyPlethysmograph)Differences discussion
510(k) NumberProposed DeviceK141793K052328NA
Product Code /BZC- Calculator, PulmonaryBZC- Calculator, PulmonaryBZC- Calculator, PulmonaryIdentical
ClassificationFunction DataFunction DataFunction Data
BZG- Spirometer, DiagnosticClass IIBZG- Spirometer, DiagnosticClass IIBZG- Spirometer, DiagnosticClass II
Intended Use andThe PulmOne MiniBox+ isThe PulmOne MiniBoxPFT™ZAN Lung-Function Lab mayEquivalent.
Indications forintended to measure lung2.0 is intended to measuremeasure or monitor pulmonaryMiniBox + shares the same
Usefunction in adult andlung function in adult andfunction in adult and pediatricintended use as the cleared
pediatric patients, 5 yearspediatric patients while atsubjects during exerciseMiniBoxPFT™ 2.0 device with
and older, while at restrest (including spirometry(including diagnosis, training,addition of diffusing capacity
(including spirometry, lungand lung volumes). Theand stress testing) or while atmodule. diffusing capacity
volumes and diffusingPulmOne MiniBoxPFT™ 2.0restmodule added to Minibox + is
capacity). The PulmOneis to be used by either a(including spirometry, airwayincluded with ZAN 500
MiniBox+ is to be used by aphysician, respiratorystrength and compliance,intended use
physician, respiratorytherapist, or technician.diffusion capacity, body
therapist, or technician in ahospital or clinic settingplethysmography, nutritionalassessment, and indirect
cardiac output) for diagnosis,
rehabilitation, performance
enhancement, and other
related activities
TargetAdult and Pediatric, 5 yearsAdult and PediatricAdult and PediatricIdentical to MiniBoxPFT™ 2.0
Populationand older
Standards /General medical deviceGeneral medical deviceGeneral medical device safety:Equivalent.
Performancesafety: IEC 60601-safety:EN 60601-1 Medical electricalMiniBox + complies with the
guideline1:2005+A1:2012 [3.1 Ed. ]IEC 60601-1:2005 (3rd Ed)safety: ElectromagneticIEC 60601-1 [3.1Ed] and IEC60601-1-2[4th Ed]
complianceMedical electrical safety:ElectromagneticElectromagneticcompatibility: EN 60601-1-2
compatibility: IEC 60601-1-compatibility:
2:2014 [4th Ed]IEC 60601-1-2:2007 (3rd Ed)
- ATS/ETS :2005,-Performance per ATS 2005
Standardization of lungStandards
function testing- ISO 26782 compliance to
-- ISO 26782 compliance tosection 7.2, 7.4-7.6 & 7.8-7.9
section 7.2, 7.4-7.6 & 7.8-7.9
PatientPatient direct contact withPatient direct contact withSingle-use, disposable FDA-Identical to MiniBoxPFT™ 2.0
Contacting Partsthe cleared disposable viralthe cleared disposable viralcleared viral bacterial filter or a
and bacterial filter, K051712and bacterial filter, K051712mouthpiece. Single-use,
User InterfaceLCD display with touch panel;LCD display with touch panel;disposable nose clipsLCD display with keyboard andIdentical to MiniBoxPFT™ 2.0
and reportstop table devicetop table devicemouse
Physical21 x 10.5 x 4.5 cm21 x 10.5 x 4.5 cmDimensions: 71 x 87 x 174 cmEquivalent.
DimensionWeight: ~10 kgWeight: ~300grVolume: 980 L Weight: 145 kgAdditional Weight due toembedded DLco module
DIFFUSING CAPACITY [Gas Analyzer]
-Technology: InfraredNot Applicable-Technology: InfraredIdentical
CO Analyzerabsorptionabsorption
SpecificationMiniBox+MiniBoxPFT™ 2.0ZAN LUNG-FUNCTION LAB(ZAN 500 BodyPlethysmograph)Differences discussion
-Measurement range: 0 to3000 ppm-Accuracy: <1%-Neutral point drift: <2% perweek-Linearity: <1%-Measurement range: 0 to 3000ppm-Accuracy: <1%-Neutral point drift: <2% perweek-Linearity: <1%
CH4 analyzerTechnology: Infraredabsorption-Measurement range: 0 to3000 ppm-Accuracy: 2.5%-Neutral point drift: <2% perweek-Linearity: <1%Not Applicable-Technology: Infraredabsorption-Measurement range: 0 to 3000ppm-Accuracy: 1%-Neutral point drift: <2% perweek-Linearity: <1%Equivalent.Since DLco is computed bythe ratio of CH4-in/CH4-out,the more important GasAnalyzer parameter is –linearity. That linearity wastested and verified by theOEM manufacturer andthrough extensive testsversus the DLco gas simulator[Hans Rudolf, Inc. The GasAnalyzer accuracy was alsoestablished by humanmeasurements, based oninternal repeatabilitymeasurements as well asversus the predicate device -the ZAN Function Lab [ZAN-500] concluded that MiniBox+didn't adversely impact thesubstantial equivalence toZAN 500 , the predicatedevice
Test gasmixture of 0.3% Carbonmonoxide (CO), 0.3%Methane (CH4),21% Oxygen (O2) withbalance Nitrogen (N2).Not Applicablemixture of 0.3% Carbonmonoxide (CO), 0.3% Methane(CH4),21% Oxygen (O2) with balanceNitrogen (N2).Identical
Demand ValveOEM by MSANot ApplicableOEM by MSAEquivalentSame component as includedwith ZAN 500
Calculation andDisplayedparametersBHT , DLco, VA,VI, CO , CH4and KCONot ApplicableBHT , DLco, VA,VI, CO , CH4 andKCOIdentical
LUNG VOLUME MEASUREMENT
MeasurementsMouth pressure and mouthflowMouth pressure and mouthflowMouth pressure and mouthflowIdentical
CalculationsTLC, TGV (FRC), RV, VC, IC andERVTLC, TGV (FRC), RV, VC, IC andERVTLC, FRC, RV, IC, VC, TV, IRV,ERVIdentical
Principle ofOperation1) Pressure and flow aremeasured at the mouth2) Spirometry is measuredwith the handheldspirometer.3) Proprietary equation isused to calculate total lung1) Pressure and flow aremeasured at the mouth2) Spirometry is measuredwith the handheldspirometer.3) Proprietary equation isused to calculate total lung1) Pressure and flow aremeasured at the mouth2) Pressure is measured in thebody box3) Equation based on Boyle'slaw is used to calculate totallung capacityIdentical to MiniBoxPFT™ 2.0
SpecificationMiniBox+MiniBoxPFT™ 2.0ZAN LUNG-FUNCTION LAB(ZAN 500 BodyPlethysmograph)Differences discussion
capacitycapacity
Method ofOperation1) Patient seated atdesk/table1) Patient seated atdesk/table1) Patient seated in body box 2)30 sec shallow panting (1 Hz)3) 1-5 second shutter closure4) Inspiration to TLC5) Slow expiration to RVIdentical to MiniBoxPFT™ 2.0
2) 1-2 minutes normalbreathing2) 1-2 minutes normalbreathing
3) 100 ms shutter closure perbreath3) 100 ms shutter closure perbreath
4) Inspiration to TLC4) Inspiration to TLC
5) Slow expiration to RV5) Slow expiration to RV
Air Container15L elliptic container15L elliptic containerBody Box ChamberIdentical to MiniBoxPFT™ 2.0
SPIROMETRY MEASUREMENT
MeasurementsMouth flowMouth flowMouth flowIdentical
FEF25-75, FEV3, FEV3/FVC,FEF25-75, FEV3, FEV3/FVC,Flow-volume loop, FVC, FEV1,FEV1/FVC%, PEF, FEF25%,FEF50%, FEF75%, PIFIdentical to MiniBoxPFT™ 2.0
CalculatedParametersFET, FEF 25, FEF50, FEF75,Evol, FIVC, FIV1, FIVC1/FIVC,PIF, IC, SVC, ERV, TV, IVC,EVC, MVV, MVV rateFET, FEF 25, FEF50, FEF75,Evol, FIVC, FIV1, FIVC1/FIVC,PIF, IC, SVC, ERV, TV, IVC,EVC, MVV, MVV rate
DisplayedParametersFEV1, FVC, FEV6, FEV1/FVC,FEV1/SVC, FEV1/VC,EFV1/FEV6, PEF, FEF25-75,FEV3, FEV3/FVC, FET, FEF 25,FEF50, FEF75, Evol, FIVC,FIV1, FIVC1/FIVC, PIF, IC, SVC,ERV, TV, IVC, EVC, MVV andMVV RateFEV1, FVC, FEV6, FEV1/FVC,FEV1/SVC, FEV1/VC,EFV1/FEV6, PEF, FEF25-75,FEV3, FEV3/FVC, FET, FEF 25,FEF50, FEF75, Evol, FIVC,FIV1, FIVC1/FIVC, PIF, IC, SVC,ERV, TV, IVC, EVC, MVV andMVV RateFlow-volume loop, FVC, FEV1,FEV1/FVC%, PEF, FEF25%,FEF50%, FEF75%, PIFIdentical to MiniBoxPFT™ 2.0
TechnicalSpecificationsFlow range: +/- 16 L/sVolume accuracy: +/- 3% or50mLFlow accuracy: +/- 5% or 200mL/sDynamic resistance at 14 L/s:<0.7 cmH2O/(L/s)Dimension: 210x105x45 mmWeight: 300 gramsFlow range: +/- 16 L/sVolume accuracy: +/- 3% or50mLFlow accuracy: +/- 5% or 200mL/sDynamic resistance at 14 L/s:<0.7 cmH2O/(L/s)Dimension: 210x105x45 mmWeight: 300 gramsFlow range: +/- 20 L/s Volumeaccuracy: <2% Flow accuracy:+/- 2% Dynamic resistance at14 L/s:<0.8 cmH2O/(L/s) Dimensions:100x50x50 mm Weight: 57gramsIdentical to MiniBoxPFT™ 2.0

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

5.5 UTILIZATION OF STANDARDS AND GUIDANCE'S:

The MiniBox+ met the following standards and guidance's:

IEC 60601-1:2005+A1:2012 - Medical electrical equipment - Part 1: General requirements for basic safety and essential performance

IEC 60601-1-2:2014 - Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests

ASTM D4169-14, Standard Practice for Performance Testing of Shipping Containers and Systems

ASTM D4332-14, Standard practice for conditioning containers, packages, or packaging components for testing

ISO 26782:2009, to section 7.2, 7.4-7.6 & 7.8-7.9

ATS/ETS :2005, Standardization of lung function testing

5.6 Summary of Non-Clinical Performance Testing

Summary of Non-Clinical Tests:

The MiniBox+ has been thoroughly tested through verification of Specifications and validation, including software validation. The following verification and validation testing were applied to the development of the system:

DLco performance verification against simulator [Hans Rudolf],DLco human comparative testing and DLco human reproducibility testing.

{9}------------------------------------------------

5.7 Summary of Clinical Performance Data

No clinical study was conducted to support this application.

5.8 CONCLUSIONS

Based on its underlying technology and bench tests performed, the PulmOne MiniBox+ is substantially equivalent to the predicate devices.

§ 868.1880 Pulmonary-function data calculator.

(a)
Identification. A pulmonary-function data calculator is a device used to calculate pulmonary-function values based on actual physical data obtained during pulmonary-function testing.(b)
Classification. Class II (performance standards).