K Number
K161214

Validate with FDA (Live)

Date Cleared
2016-06-27

(59 days)

Product Code
Regulation Number
862.3250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Wondfo Amphetamine Urine Test AMP 500 Cup is an immunochromatographic assay for the qualitative determination of Amphetamine in human urine at a Cut-Off concentration of 500 ng/mL. This test is calibrated to d-Amphetamine (calibrator).

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. This test is intended for over-the-counter (OTC) consumer use as the first step in a twostep process to provide consumers with information concerning the presence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing- the second step in the process, is provided in the package labeling.

Wondfo Amphetamine Urine Test AMP 500 DipCard is an immunochromatographic assay for the qualitative determination of Amphetamine in human urine at a Cut-Off concentration of 500 ng/mL. This test is calibrated to d-Amphetamine (calibrator).

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. This test is intended for over-the-counter (OTC) consumer use as the first step in a twostep process to provide consumers with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing- the second step in the process, is provided in the package labeling.

Wondfo Cocaine Urine Test COC 150 Cup is an immunochromatographic assay for the qualitative determination of Benzoylecgonine in human urine at a Cut-Off concentration of 150 ng/mL. This test is callbrated to Benzoylecgonine (calibrator).

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. This test is intended for over-the-counter (OTC) consumer use as the first step in a twostep process to provide consumers with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing- the second step in the process, is provided in the package labeling.

Wondfo Cocaine Urine Test COC 150 DipCard is an immunochromatographic assay for the qualitative determination of Benzoylecgonine in human urine at a Cut-Off concentration of 150 ng/mL. This test is callbrated to Benzoylecgonine (calibrator).

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. This test is intended for over-the-counter (OTC) consumer use as the first step in a twostep process to provide consumers with information concerning the presence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing- the second step in the process, is provided in the package labeling.

Wondfo Methamphetamine Urine Test MET 500 Cup is an immunochromatographic assay for the qualitative determination of Methamphetamine in human urine at a Cut-Off concentration of 500 ng/mL. This test is calibrated to d-Methamphetamine (calibrator).

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. This test is intended for over-the-counter (OTC) consumer use as the first step in a twostep process to provide consumers with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing- the second step in the process, is provided in the package labeling.

Wondfo Amphetamine Urine Test MET 500 DipCard is an immunochromatographic assay for the qualitative determination of Methamphetamine in human urine at a Cut-Off concentration of 500 ng/mL. This test is calibrated to d-Methamphetamine (calibrator).

The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. This test is intended for over-the-counter (OTC) consumer use as the first step in a twostep process to provide consumers with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing- the second step in the process, is provided in the package labeling.

Device Description

WONDFO Urine Test devices are immunochromatographic assays for cocaine, amphetamine and Methamphetamine. Each assay test is a lateral flow, one step system for the qualitative detection of Benzoylecgonine, or D-amphetamine or D-methamphetamine (target analyte) in human urine. The product is a single-use in vitro diagnostic device, which comes in the form of: DipCards, or Cups. It contains a Test Device (in one of the two formats), a package insert and a urine cup. Each test device is sealed with a desiccant in an aluminum pouch.

AI/ML Overview

The document provides information on the acceptance criteria and performance of the Wondfo Amphetamine/Cocaine/Methamphetamine Urine Tests.

Here's the breakdown:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" in a table format with a direct comparison to performance against those criteria. Instead, it presents performance characteristics (precision, cut-off verification, interference, specificity, effects of specific gravity and pH, agreement with GC/MS, and lay-user performance) which implicitly define the device's acceptable performance. The performance demonstrated is the acceptance criteria met.

Implicit Acceptance Criteria and Reported Device Performance (Summary derived from the text):

Performance CharacteristicAcceptance Criteria (Implied)Reported Device Performance
PrecisionConsistent results across concentrations and lots.Amphetamine (AMP) Dip Card & Cup: For all three lots and both formats: -100%, -75%, -50%, -25% Cut-off: 50-/0+ (Negative) observed for all tests.+25%, +50%, +75%, +100% Cut-off: 50+/0- (Positive) observed for all tests.At Cut-off: 43+/7- or 44+/6- (Mix of Positive/Negative) observed, indicating expected behavior around the cut-off. Cocaine (COC) Dip Card & Cup: Similar results to AMP for all three lots and both formats, demonstrating expected performance across the concentration range and at the cut-off. (Methamphetamine (MET) precision data reported in K122961, not detailed here.)
Cut-off VerificationAccurate classification at and around defined cut-off levels.All samples at and above +50% cut-off were positive. All samples at and below -50% cut-off were negative for Amphetamine, Cocaine, and Methamphetamine.
InterferenceNo interference from common physiological/pathological substances.Numerous common substances (listed in the document, pages 10-12) showed no interference at 100 µg/mL. Targeted drugs at 25% above cut-off levels were correctly detected in the presence of these substances.
SpecificityProper reactivity to target analytes and appropriate cross-reactivity.AMP: D-Amphetamine: 100% reactivity at 500 ng/mL. Cross-reactivity observed with L-Amphetamine (2%), D,L-Amphetamine (33%), Phentermine (33%), Hydroxyamphetamine (6%), MDA (20%). No reaction with D/L-Methamphetamine, Ephedrine, MDE, MDMA at 100,000 ng/mL.COC: Benzoylecgonine: 100% reactivity at 150 ng/mL. Cross-reactivity with Cocaine HCl (40%), Cocaethylene (2.4%), Ecgonine (0.9%), Norcocaine (0.3%). MET: D-Methamphetamine: 100% reactivity at 500 ng/mL. 100% reactivity with MDEA, MDA, D/L-Methamphetamine. Cross-reactivity with L-Methamphetamine (5%), L-Amphetamine (1.3%).
Urine Specific Gravity & pH EffectsConsistent results across a range of urine specific gravity and pH.Urine samples with specific gravity 1.000-1.035 and pH 4-9 spiked at +/- 25% cut-off levels showed all positive results for samples at and above +25% cut-off and all negative results for samples at and below -25% cut-off. No differences were observed across formats.
Agreement with GC/MS (Method Comparison)High agreement with GC/MS for both negative and positive samples.AMP (Dip Card & Cup): Across 3 viewers, 0 false positives for negative and low negative samples. For near cut-off negative, 1-2 false positives were reported by viewers. For near cut-off positive and high positive samples, viewers consistently reported positive results with minimal discordance (e.g., 2 samples with GC/MS at 480 or 481 ng/mL were called positive by some viewers, demonstrating sensitivity close to cut-off). COC (Dip Card & Cup): Similar to AMP, with 0 false positives for negative and low negative. Few false positives for near cut-off negative (1-2). High agreement for near cut-off positive and high positive. Discordant results mainly around the cut-off (e.g., GC/MS 145 or 146 ng/mL called positive by viewers).(Methamphetamine (MET) data reported in K122961, not detailed here.)
Lay-User PerformanceHigh percentage of correct results by untrained users.AMP (Cup & Dip Card): 100% correct for -100%, -75%, -50%, +50%, +75% cut-off concentrations. 90-95% correct for -25% and +25% cut-off concentrations (expected variability around cut-off). COC (Cup & Dip Card): 100% correct for -100%, -75%, -50%, +50%, +75% cut-off concentrations. 90% correct for -25% and +25% cut-off concentrations. MET (Cup & Dip Card): 100% correct for -100%, -75%, -50%, +50%, +75% cut-off concentrations. 90% correct for -25% and +25% cut-off concentrations. All lay users found instructions easy to follow, with a Flesch-Kincaid Grade Level of 7.

The study demonstrates that the Wondfo Urine Test devices perform as expected for qualitative drug detection around specified cut-off levels, show appropriate specificity, are not significantly affected by common interfering substances or urine properties, and can be used accurately by lay-users. This collective performance data indicates that the devices meet the implied acceptance criteria for their intended use as preliminary, over-the-counter drug screening tests.


Study Details:

2. Sample Size and Data Provenance

  • Test Set Sample Size:

    • Precision Studies: For each drug concentration, three lots of devices were tested, with two runs per day for 25 days. The number of individual tests per concentration point for each lot is not explicitly stated as a single total, but 50 measurements (e.g., 50-/0+ or 50+/0-) are reported for each concentration for each lot. This implies 50 tests per concentration per lot for each of "Lot 1, Lot 2, Lot 3". Given 9 concentration points per drug, that's 9 * 50 = 450 tests per lot, and 3 lots means 1350 tests per drug for each device format (Cup/DipCard).
    • Cut-off Verification: 150 samples were used in total, equally distributed across 5 concentrations (-50%, -25%, cut-off, +25%, +50%). Tested using three different lots of each device by three different operators. (This means 150 samples * 3 lots * 3 operators = 1350 individual test results reported for this section across all drugs and device formats).
    • Interference & Specificity Studies: Not explicitly stated as a total N, but substances were added to drug-free urine and target-drug urine (25% above cut-off) and tested using "three batches of each device for all formats."
    • Method Comparison (Agreement with GC/MS): 80 unaltered clinical samples (40 negative, 40 positive) were used for each drug (Amphetamine, Cocaine, Methamphetamine) for each device format (Cup/DipCard).
    • Lay-user Study: 280 lay persons for Amphetamine devices, 280 for Cocaine devices, and 280 for Methamphetamine devices. Each lay person tested one blind-labeled sample. For each drug, 7 distinct concentration levels with 20 samples per level were used. This means 7 * 20 = 140 samples were prepared for each drug per device format (e.g., AMP Cup), and these samples were then distributed to the 280 lay users (presumably 140 for the Cup and 140 for the DipCard, but this is not explicitly stated for all, though AMP and COC show 140 samples in total distributed across the 7 concentrations for both Cup and DipCard results tables).
  • Data Provenance:

    • Country of Origin: Wondfo Biotech Co., Ltd. is located in Guangzhou, P.R. China (page 4). The studies were likely conducted internally or by associated labs.
    • Retrospective or Prospective: Not explicitly stated, but the "clinical samples" for the method comparison were "unaltered," which often implies prospectively collected real-world samples. The precision and cut-off studies involved "spiking drug in negative samples" which are laboratory-prepared. The lay-user study used prepared samples and was a controlled prospective study.

3. Number of Experts and Qualifications for Ground Truth

  • Method Comparison (Agreement with GC/MS):

    • The ground truth was established by GC/MS (Gas Chromatography/Mass Spectrometry), which is the preferred confirmatory method for drug testing, considered the "gold standard."
    • No human "experts" were listed as establishing ground truth for the test set; the GC/MS analytical results served as the objective gold standard.
    • Three "laboratory assistants" were involved in running the Wondfo devices, not in establishing the ground truth. Their qualifications are not provided.
  • Precision, Cut-off, Interference, Specificity, Specific Gravity/pH: The ground truth for these studies was established by the precise preparation of samples with known drug concentrations (confirmed by GC/MS where relevant) and the analytical methods used to prepare and characterize the samples.

  • Lay-user Study: The ground truth for the lay-user study was established by GC/MS for the prepared urine samples.

4. Adjudication Method for the Test Set

  • Method Comparison: No formal "adjudication method" among human readers is described for the test sets. The Wondfo device results (read by the laboratory assistants/viewers) were directly compared to the GC/MS ground truth. Discordant results are noted individually for each viewer against the GC/MS result.
  • Precision/Cut-off: Not applicable, as results were based on predefined expectations for concentrations.
  • Lay-user Study: Not applicable, as each lay person read their assigned sample independently against the GC/MS-confirmed concentration.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No MRMC comparative effectiveness study was explicitly performed or reported in this document to assess how much human readers improve with AI (or rather, with the device) versus without assistance.
  • The studies were designed to evaluate the performance of the device itself and the ability of lay users to correctly interpret device results compared to a gold standard (GC/MS).
  • The "Method Comparison" section involved three "viewers" (laboratory assistants) independently reading the device results and comparing them to GC/MS. This is a multi-reader study, but it's evaluating the device's accuracy when read by trained personnel, not necessarily the improvement of those readers.
  • The "Lay-user study" is also a multi-reader study (280 lay persons), but it's assessing the device's ease of use and accuracy in untrained hands, not an improvement over unassisted human performance.

6. Standalone Performance Study

  • Yes, a standalone performance study was done. The core of the document describes the standalone performance of the Wondfo Urine Test device itself.
    • The "Precision," "Cut-off," "Interference," and "Specificity" sections evaluate the analytical performance of the device in a laboratory setting when operated by laboratory personnel, without explicit human-in-the-loop decision-making beyond visual interpretation of a test line.
    • The "Method Comparison" study explicitly compares the device's qualitative results (interpreted visually by laboratory assistants) against the GC/MS gold standard. This is a direct measure of the device's standalone performance in a simulated clinical context.

7. Type of Ground Truth Used

  • The primary ground truth used for all performance studies (precision, cut-off, interference, specificity, method comparison, lay-user study) for drug concentrations was GC/MS (Gas Chromatography/Mass Spectrometry). GC/MS is a highly accurate analytical chemistry technique considered the definitive "confirmatory method" for drug identification and quantification in urine.
  • For the lay-user study, the ground truth for ease of use was collected via user surveys.

8. Sample Size for the Training Set

  • Not Applicable. This device is an immunochromatographic assay (lateral flow immunoassay), not a machine learning or AI-based device that typically requires a "training set" in the computational sense. The device's performance is based on its chemical and biological design, not a trained algorithm.

9. How the Ground Truth for the Training Set Was Established

  • Not Applicable, as there is no "training set" for this type of device. The ground truth for the validation/test samples was established via GC/MS.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 27, 2016

GUANGZHOU WONDFO BIOTECH CO., LTD. C/O JOE SHIA 504 E DIAMOND AVE., SUITE I GAITHERSBURG MD 20877

Re: K161214

Trade/Device Name: Wondfo Amphetamine Urine Test AMP 500 (Cup, Dipcard), Wondfo Cocaine Urine Test COC 150(Cup. Dipcard). Wondfo Methamphetamine Urine Test MET 500 (Cup. Dipcard) Regulation Number: 21 CFR 862.3250 Regulation Name: Cocaine and cocaine metabolite test system Regulatory Class: II Product Code: DIO, DKZ, DJC Dated: April 21, 2016 Received: April 29, 2016

Dear Mr. Shia:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Courtney H. Lias -S

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K161214

Device Name Wondfo Amphetamine Urine Test AMP 500 (Cup, DipCard) Wondfo Cocaine Urine Test COC 150 (Cup, DipCard)

Wondfo Methamphetamine Urine Test MET 500 (Cup, DipCard)

Indications for Use (Describe)

Wondfo Amphetamine Urine Test AMP 500 Cup is an immunochromatographic assay for the qualitative determination of Amphetamine in human urine at a Cut-Off concentration of 500 ng/mL. This test is calibrated to d-Amphetamine (calibrator).

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. This test is intended for over-the-counter (OTC) consumer use as the first step in a twostep process to provide consumers with information concerning the presence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing- the second step in the process, is provided in the package labeling.

Wondfo Amphetamine Urine Test AMP 500 DipCard is an immunochromatographic assay for the qualitative determination of Amphetamine in human urine at a Cut-Off concentration of 500 ng/mL. This test is calibrated to d-Amphetamine (calibrator).

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. This test is intended for over-the-counter (OTC) consumer use as the first step in a twostep process to provide consumers with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing- the second step in the process, is provided in the package labeling.

Wondfo Cocaine Urine Test COC 150 Cup is an immunochromatographic assay for the qualitative determination of Benzoylecgonine in human urine at a Cut-Off concentration of 150 ng/mL. This test is callbrated to Benzoylecgonine (calibrator).

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. This test is intended for over-the-counter (OTC) consumer use as the first step in a twostep process to provide consumers with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing- the second step in the process, is provided in the package labeling.

Wondfo Cocaine Urine Test COC 150 DipCard is an immunochromatographic assay for the qualitative determination of Benzoylecgonine in human urine at a Cut-Off concentration of 150 ng/mL. This test is callbrated to Benzoylecgonine (calibrator).

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. This test is intended for over-the-counter (OTC) consumer use as the first step in a twostep process to provide consumers with information concerning the presence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing- the second step in the process, is provided in the package labeling.

{3}------------------------------------------------

Wondfo Methamphetamine Urine Test MET 500 Cup is an immunochromatographic assay for the qualitative determination of Methamphetamine in human urine at a Cut-Off concentration of 500 ng/mL. This test is calibrated to d-Methamphetamine (calibrator).

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. This test is intended for over-the-counter (OTC) consumer use as the first step in a twostep process to provide consumers with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing- the second step in the process, is provided in the package labeling.

Wondfo Amphetamine Urine Test MET 500 DipCard is an immunochromatographic assay for the qualitative determination of Methamphetamine in human urine at a Cut-Off concentration of 500 ng/mL. This test is calibrated to d-Methamphetamine (calibrator).

The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. This test is intended for over-the-counter (OTC) consumer use as the first step in a twostep process to provide consumers with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing- the second step in the process, is provided in the package labeling.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

    1. Date: June 8, 2016
  1. Submitter: Wondfo Biotech Co., Ltd. South China University of Technology Guangzhou, P.R. China 510641
    1. Contact person: Joe Shia LSI International Inc. 504 East Diamond Ave., Suite I Gaithersburg, MD 20877 Telephone: 240-505-7880 Fax: 301-916-6213 Email:shiajl@yahoo.com
    1. Device Name: WONDFO Urine Test Amphetamine (Cup, DipCard) WONDFO Urine Test Cocaine (Cup, DipCard) WONDFO Urine Test Methamphetamine (Cup, DipCard)
Classification. Class II
ProductCodeCFR #Panel
DIO21 CFR, 862.3250 Cocaine Test SystemToxicology
DKZ21 CFR, 862.3100 Amphetamine Test SystemToxicology
DJC21 CFR, 862.3610 Methamphetamine Test SystemToxicology

Classification: Class II

    1. Predicate Devices:
      K122809 Advin Multi-Drug Screen Test
    1. Intended Use
      WONDFO Urine Test Amphetamine Cup is a rapid test for the qualitative detection of Amphetamine in human urine at a cutoff concentration of 500 ng/mL. This test is calibrated to d-Amphetamine (calibrator).

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

{5}------------------------------------------------

For in vitro diagnostic use only. This test is intended for over-the-counter (OTC) consumer use as the first step in a two-step process to provide consumers with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing- the second step in the process, is provided in the package labeling.

WONDFO Urine Test Amphetamine DipCard is a rapid test for the qualitative detection of Amphetamine in human urine at a cutoff concentration of 500 ng/mL. This test is calibrated to d-Amphetamine (calibrator).

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. This test is intended for over-the-counter (OTC) consumer use as the first step in a two-step process to provide consumers with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing- the second step in the process, is provided in the package labeling.

WONDFO Urine Test Cocaine Cup is a rapid test for the qualitative detection of Benzoylecgonine in human urine at a cutoff concentration of 150 ng/ml. This test is calibrated to Benzoylecgonine (calibrator).

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. This test is intended for over-the-counter (OTC) consumer use as the first step in a two-step process to provide consumers with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing- the second step in the process, is provided in the package labeling.

WONDFO Urine Test Cocaine DipCard is a rapid test for the qualitative detection of Benzoylecgonine in human urine at a cutoff concentration of 150 ng/ml. This test is calibrated to Benzoylecgonine (calibrator).

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. This test is intended for over-the-counter (OTC) consumer use as the first step in a two-step process to provide consumers with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing- the second step in the process, is provided in the package labeling.

{6}------------------------------------------------

WONDFO Urine Test Methamphetamine Cup is a rapid test for the qualitative detection of Methamphetamine in human urine at a cutoff concentration of 500 ng/mL. This test is calibrated to d-Methamphetamine (calibrator).

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. This test is intended for over-the-counter (OTC) consumer use as the first step in a two-step process to provide consumers with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing- the second step in the process, is provided in the package labeling.

WONDFO Urine Test Methamphetamine DipCard is a rapid test for the qualitative detection of Methamphetamine in human urine at a cutoff concentration of 500 ng/mL. This test is calibrated to d-Methamphetamine (calibrator).

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. This test is intended for over-the-counter (OTC) consumer use as the first step in a two-step process to provide consumers with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing- the second step in the process, is provided in the package labeling.

    1. Device Description
      WONDFO Urine Test devices are immunochromatographic assays for cocaine, amphetamine and Methamphetamine. Each assay test is a lateral flow, one step system for the qualitative detection of Benzoylecgonine, or D-amphetamine or D-methamphetamine (target analyte) in human urine. The product is a single-use in vitro diagnostic device, which comes in the form of: DipCards, or Cups. It contains a Test Device (in one of the two formats), a package insert and a urine cup. Each test device is sealed with a desiccant in an aluminum pouch.
    1. Substantial Equivalence Information
      A summary comparison of features of the WONDFO Urine Test and the predicate device is provided in Table 1, Table 2 & Table 3.
Table 1: Features Comparison of WONDFO Cocaine Test and the Predicate Device
ItemDevicePredicate -K122809
Indication(s)for UseFor the qualitative determination ofBenzoylecgonine in human urine.Same

{7}------------------------------------------------

ItemDevicePredicate -K122809
CalibratorBenzoylecgonineSame
MethodologyCompetitive binding, lateral flowimmunochromatographic assays based onthe principle of antigen antibodyimmunochemistry.Same
Specimen TypeHuman UrineSame
Cut-Off Values150 ng/mLSame
IntendedPopulationFor over-the-counter and prescriptionuses.Same
ConfigurationsCup, Dip CardCup, Dip Card,Cassette

Table 2: Features Comparison of WONDFO Amphetamine Test and the Predicate Device

ItemDevicePredicate - K122809
Indication(s)for UseFor the qualitative determination ofD-amphetamine in human urine.Same
CalibratorD-amphetamineSame
MethodologyCompetitive binding, lateral flowimmunochromatographic assays basedon the principle of antigen antibodyimmunochemistry.Same
Specimen TypeHuman UrineSame
Cut-Off Values500 ng/mLSame
IntendedPopulationFor over-the-counter and prescriptionuses.Same
ConfigurationsCup, Dip CardCup, Dip Card,Cassette

Table 3: Features Comparison of WONDFO Methamphetamine Test and the Predicate Device

ItemDevicePredicate - K122809
Indication(s)for UseFor the qualitative determination ofD-methamphetamine in human urine.Same
CalibratorD-methamphetamineSame

{8}------------------------------------------------

ItemDevicePredicate - K122809
MethodologyCompetitive binding, lateral flowimmunochromatographic assays basedon the principle of antigen antibodyimmunochemistry.Same
Specimen TypeHuman UrineSame
Cut-Off Values500 ng/mLSame
IntendedPopulationFor over-the-counter and prescriptionuses.Same
ConfigurationsCup, Dip CardCup, Dip Card,Cassette

9. Test Principle

WONDFO Urine Tests are rapid tests for the qualitative detection of Benzoylecgonine, or D-amphetamine or D-methamphetamine in urine samples. Each assay test is a lateral flow chromatographic immunoassay. During testing, a urine specimen migrates upward by capillary action. If target drugs are present in the urine specimen below its cut-off concentration, it will not saturate the binding sites of its specific antibody (monoclonal mouse antibody) coated on the particles. The antibody-coated particles will then be captured by immobilized drug-conjugate and a visible colored line will show up in the test line region. The colored line will not form in the test line region if the target drug level exceeds its cutoff-concentration because it will saturate all the binding sites of the antibody coated on the particles. A band should form in the control region of the devices regardless of the presence of drug or metabolite in the sample.

10. Performance Characteristics

    1. Analytical Performance
    • a. Precision

Precision studies were carried out for samples with concentrations of -100% cut-off, -50% cut-off, -25% cut-off, at the cut-off, +25% cut-off, +75% cut-off, +75% cut-off and +100% cut-off. These samples were prepared by spiking drug in negative samples. Each drug concentration was confirmed by GC/MS. All sample aliquots were blinded labeled and randomized. For each concentration, tests were performed two runs per day for 25 days. The results obtained are summarized in the following tables:

DrugResult-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Lot 150-/0+50-/0+50-/0+50-/0+43+/7-50+/0-50+/0-50+/0-50+/0-

Amphetamine (AMP) Dip Card Format

{9}------------------------------------------------

Result-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
DrugLot 250-/0+50-/0+50-/0+50-/0+44+/6-50+/0-50+/0-50+/0-50+/0-
DrugLot 350-/0+50-/0+50-/0+50-/0+44+/6-50+/0-50+/0-50+/0-50+/0-
Amphetamine (AMP) Cup Format
Result-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
DrugLot 150-/0+50-/0+50-/0+50-/0+44+/6-50+/0-50+/0-50+/0-50+/0-
DrugLot 250-/0+50-/0+50-/0+50-/0+43+/7-50+/0-50+/0-50+/0-50+/0-
DrugLot 350-/0+50-/0+50-/0+50-/0+44+/6-50+/0-50+/0-50+/0-50+/0-

Cocaine (COC) Dip Card Format

DrugResult-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Lot 150-/0+50-/0+50-/0+50-/0+43+/7-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+44+/6-50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+43+/7-50+/0-50+/0-50+/0-50+/0-

COC Cup Format

DrugResult-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Lot 150-/0+50-/0+50-/0+50-/0+44+/6-50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+43+/7-50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+43+/7-50+/0-50+/0-50+/0-50+/0-

The Methamphetamine precision data was reported in K122961.

  • b. Linearity
    Not applicable, these are visually read devices.

  • c. Stability

{10}------------------------------------------------

The devices are stable at 4-30°C (39-86°F) for 24 months based on the accelerated stability study at 50°C. Control materials are not provided with the device. The labeling provides information on how to obtain control materials.

  • d. Cut-off
    A total of 150 samples equally distributed at concentrations of -50% cut-off; -25% cut-off; cut-off; +25% cut-off; +50% cut-off were tested using three different lots of each device by three different operators. Results were all positive at and above +50% cut-off and all negative at and below -50% cut-off for Amphetamine, Cocaine and Methamphetamine (K122961). The following cut-off values for the test devices have been verified.
TestCalibratorCut-off (ng/mL)
WONDFO Urine Test AmphetamineCup (DipCard)D-amphetamine500
WONDFO Urine Test Cocaine Cup(DipCard)Benzoylecogonine150
WONDFO Urine TestMethamphetamine Cup (DipCard)D-methamphetamine500

e. Interference

Potential interfering substances found in human urine of physiological or pathological conditions were added to drug-free urine and target drugs urine with concentration at 25% above cut-off levels. These urine samples were tested using three batches of each device for all formats.

Compounds that showed no interference at a concentration of 100µg/mL are summarized in the following tables. There were no differences observed for different formats.

4-AcetamidophenolL-EphedrineOxycodone
Acetophenetidin(-) Y EphedrineOxymetazoline
N-AcetylprocainamideErythromycinPapaverine
Acetylsalicylic acidß-EstradiolPenicillin-G
AminopyrineEstrone-3-sulfatePentazocaine
AmitryptylineEthyl aminobenzoatePentobarbital
AmobarbitalFenfluraminePerphenazine
AmoxicillinFenoprofenPhencyclidine
AmpicillinFurosemidePhenelzine
Ascorbic acidGentisic acidPhenobarbital
AspartameHemoglobinPhetoin
AtropineHydralazineL-Phenylephrine
Benzilic acidHydrochlorothiazidePhenylpropanolamine

AMP

{11}------------------------------------------------

Benzoic acidHydrocodonePrednisolone
BenzoylecgonineHydrocortisonePrednisone
BilirubinO-Hydroxyhippuric acidProcaine
Brompheniramine3-HydroxytyraminePromazine
CaffeineIbuprofenPromethazine
CannabidiolImipramineD,L-Propanolol
Cannabinol(-) IsoproterenolD-Propoxyphene
ChloralhydrateIsoxsuprineQuinidine
ChloramphenicolKetamineQuinine
ChlordiazepoxideKetoprofenRanitidine
ChlorothiazideLabetalolSalicylic acid
(±)-ChlorpheniramineLevorphanolSecobarbital
ChlorpromazineLoperamideSulfamethazine
ChloroquineMaprotilineSulindac
CholesterolMeperidineTemazepam
ClomipramineMeprobamateTetracycline
ClonidineMethadoneTetrahydrocortisone
Cocaine hydrochlorideMethylphenidateTetrahydrozoline
CodeineMorphine-3-DglucuronideΔ9-THC-COOH
CortisoneNalidixic acidThebaine
(-) CotinineNaloxoneThiamine
CreatinineNaltrexoneThioridazine
DeoxycorticosteroneNaproxenD,L-Thyroxine
DextromethorphanNiacinamideTolbutamide
DiazepamNifedipineTriamterene
DiclofenacNorcodeinTrifluoperazine
DiflunisalNorethindroneTrimethoprim
DigoxinD-NorpropoxypheneTrimipramine
DiphenhydramineNoscapineTryptamine
DoxylamineD,L-OctopamineD, L-Tyrosine
Ecgonine hydrochlorideOxalic acidUric acid
Ecgonine methylesterOxazepamVerapamil
(1R,2S)-(-)-EphedrineOxolinic acidZomepirac

COC

AcetaminophenEstrone-3-sulfateOxymetazoline
AcetophenetidinEthyl-p-aminobenzoatePapaverine
N-AcetylprocainamideFenoprofenPenicillin-G
Acetylsalicylic acidFurosemidePentobarbital
AminopyrineGentisic acidPerphenazine
AmitryptylineHemoglobinPhencyclidine
AmobarbitalHydralazinePhenelzine
AmoxicillinHydrochlorothiazidePhenobarbital
AmpicillinHydrocodonePhentermine
L-Ascorbic acidHydrocortisoneL-Phenylephrine
DL-Amphetamine SulfateO-Hydroxyhippuric acidβ-Phenylethylamine
Apomorphinep-HydroxymethamphetaminePhenylpropanolamine
Aspartame3-HydroxytyraminePrednisolone
AtropineIbuprofenPrednisone
Benzilic acidImipramineProcaine
Benzoic acidIproniazidPromazine
Benzphetamine(±) - IsoproterenolPromethazine
(±) -BrompheniramineIsoxsuprineDL-Propranolol
CaffeineKetamineD-Propoxyphene
CannabidiolKetoprofenD-Pseudoephedrine
CannabinolLabetalolQuinidine
ChloralhydrateLevorphanolQuinine
ChloramphenicolLoperamideRanitidine
ChlordiazepoxideMaprotilineSalicylic acid
ChlorothiazideMeperidineSecobarbital
(±) -ChlorpheniramineMeprobamateSerotonin
ChlorpromazineMethadoneSulfamethazine
ChloroquineMethoxyphenamineSulindac
Cholesterol(±) -3,4-MethylenedioxyamphetamineTemazepam
ClomipramineMethylene-dioxymethamphetamineTetracycline
ClonidineMorphine-3-B-D glucuronideTetrahydrocortisone3-(β-D glucuronide)
CodeineMorphine SulfateTetrahydrozoline
CortisoneNalidixic acidThebaine
(-) CotinineNaloxoneThiamine
CreatinineNaltrexoneThioridazine
DeoxycorticosteroneNaproxenDL-Tyrosine
DextromethorphanNiacinamideTolbutamide
DiazepamNifedipineTriamterene
DiclofenacNorcodeinTrifluoperazine
DiflunisalNorethindroneTrimethoprim
DigoxinD-NorpropoxypheneTrimipramine
DiphenhydramineNoscapineTryptamine
DoxylamineDL-OctopamineDL-Tryptophan
Ecgonine methylesterOxalic acidTyramine
(-) - Ψ-EphedrineOxazepamUric acid
ErythromycinOxolinic acidVerapamil
ß-EstradiolOxycodoneZomepirac

{12}------------------------------------------------

{13}------------------------------------------------

MET: The MET interference data was reported in K122961.

  • f. Specificity
    To test the specificity, drug metabolites and other components that are likely to interfere in urine samples were tested using three batches of each device for all formats. The obtained lowest detectable concentration was used to calculate the cross-reactivity. There were no differences observed for different formats.
AMPResultReactivity
(D- Amphetamine, Cut-off=500 ng/mL)Positive at 500 ng/mL100%
L - Amphetamine250002%
D,L - Amphetamine150033%
Phentermine150033%
Hydroxyamphetamine80006%
Methylenedioxyamphetamine (MDA)250020%
d-MethamphetamineNegative at 100000<1%
1-MethamphetamineNegative at 100000<1%
ephedrineNegative at 100000<1%
Methylenedioxyethylamphetamine (MDE)Negative at 100000<1%
3,4-methylenedioxy-methamphetamine(MDMA)Negative at 100000<1%
COCResultReactivity
(Benzoylecogonine, Cut-off=150 ng/mL)Positive at 150 ng/mL100%
Cocaine HClPositive at 375 ng/mL40%
CocaethylenePositive at 6250 ng/mL2.4%
EcgoninePositive at 16000 ng/mL0.9%
NorcocainePositive at 50000 ng/mL0.3%
METResultReactivity
(D- Methamphetamine, Cut-off=500 ng/mL)Positive at 500 ng/mL100%
(+/-)3,4-Methylenedioxy-n-ethylamphetamine(MDEA)Positive at 500 ng/mL100%
(+/-) 3,4-Methylenedioxyamphetamine (MDA)Positive at 500 ng/mL100%
d/l-MethamphetaminePositive at 500 ng/mL100%

{14}------------------------------------------------

l-MethamphetaminePositive at10,000 ng/mL5%
l-AmphetaminePositive at 37,500 ng/mL1.3%

MET: Additional MET interference data was reported in K122961.

  • g. Effect of Urine Specific Gravity and Urine pH
    To investigate the effect of urine specific gravity and urine pH, urine samples with of 1.000 to 1.035 specific gravity or urine samples with pH 4 to 9 were spiked with target drugs at 25% below and 25% above cut-off levels. These samples were tested using three batches of each device for all formats. Results were all positive for samples at and above +25% cut-off and all negative for samples at and below -25% Cut-Off. There were no differences observed for different formats.

    1. Comparison Studies
      The method comparison studies for the WONDFO Urine Test (Amphetamine, Cocaine and Methamphetamine) were performed in-house with three different laboratory assistants for each format of the device. Operators ran 80 (40 negative and 40 positive) unaltered clinical samples for each drug. The samples were blind labeled and compared to GC/MS results. The results are presented in the tables below:
AMP
Dip CardformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)NearCutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan +50%)
Viewer APositive0023010
Negative10171100
Viewer BPositive0023010
Negative10171100
Viewer CPositive0013010
Negative10171200
Discordant Results of AMP Dip Card
ViewerSample NumberGC/MS ResultDipcard FormatViewer Results
Viewer AAMP5062480Positive
Viewer AAMP5065481Positive
Viewer BAMP5062480Positive
Viewer CAMP5062480Positive

{15}------------------------------------------------

ViewerSample NumberGC/MS ResultDipcard FormatViewer Results
Viewer BAMP5218365Positive

AMP

CupformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)NearCutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan +50%)
Viewer APositive0023010
Negative10171100
Viewer BPositive0013010
Negative10171200
Viewer CPositive0023010
Negative10171100

Discordant Results of AMP Cup

ViewerSample NumberGC/MS ResultCup FormatViewer Results
Viewer AAMP5062480Positive
Viewer AAMP5065481Positive
Viewer BAMP5062480Positive
Viewer CAMP5062480Positive
Viewer CAMP5065481Positive

COC

{16}------------------------------------------------

Dip CardformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)NearCutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan +50%)
Viewer APositive0023010
Negative10181000
Viewer BPositive0023010
Negative10181000
Viewer CPositive0012911
Negative10181100

Discordant Results of COC DipCard

ViewerSample NumberGC/MS ResultDipCard FormatViewer Results
Viewer ACOC1215145Positive
Viewer ACOC1217146Positive
Viewer BCOC1215145Positive
Viewer BCOC1217146Positive
Viewer CCOC1217146Positive

COC

CupformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)NearCutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan +50%)
Viewer APositive0023010
Negative10181000
Viewer BPositive0013010
Negative10181100
Viewer CPositive0023010
Negative10181000

Discordant Results of COC Cup

{17}------------------------------------------------

ViewerSample NumberGC/MS ResultCup FormatViewer Results
Viewer ACOC1215145Positive
Viewer BCOC1215145Positive
Viewer CCOC1215145Positive
Viewer CCOC1217146Positive
Viewer ACOC1217146Negative

The Methamphetamine method comparison data was reported in K122961.

Lay-user study

A lay user study was performed at three intended user sites with 280 lay persons testing the amphetamine devices, 280 lay persons testing the cocaine devices and 280 lay persons testing the Methamphetamine devices. They had diverse educational and professional backgrounds and ranged in age from 21 to > 50 years. Urine samples were prepared at the following concentrations; negative, +/-75%, +/-50%, +/-25% of the cutoff by spiking drugs into drug free-pooled urine specimens. The concentrations of the samples were confirmed by GC/MS. Each sample was aliquoted into individual containers and blind-labeled. Each participant was provided with the package insert, 1 blind labeled sample and a device. The results are summarized below.

% of CutoffNumberofsamplesd-AmphetamineConcentration by GC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff20126.3020100
-50% Cutoff20248.6020100
-25% Cutoff20377.111995.0
+25% Cutoff20625.818290.0
+50% Cutoff20753.2200100
+75% Cutoff20874.3200100

Comparison between GC/MS and Lay Person Results AMP Cup

Comparison between GC/MS and Lay Person Results AMP Dip Card

% of CutoffNumberofsamplesd-AmphetamineConcentration by GC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results
No. ofNo. of

{18}------------------------------------------------

PositiveNegative(%)
-100% Cutoff200020100
-75% Cutoff20126.3020100
-50% Cutoff20248.6020100
-25% Cutoff20377.121890.0
+25% Cutoff20625.818290.0
+50% Cutoff20753.2200100
+75% Cutoff20874.3200100

Comparison between GC/MS and Lay Person Results COC Cup

NumberofsamplesBenzoylecgonineConcentration by GC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
% of CutoffNo. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff2038.1020100
-50% Cutoff2075.6020100
-25% Cutoff20113.421890.0
+25% Cutoff20186.718290.0
+50% Cutoff20226.3200100
+75% Cutoff20259.8200100

Comparison between GC/MS and Lay Person Results COC Dip Card

% of CutoffNumberofsamplesBenzoylecgonineConcentration by GC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff2038.1020100
-50% Cutoff2075.6020100
-25% Cutoff20113.421890.0
+25% Cutoff20186.718290.0
+50% Cutoff20226.3200100
+75% Cutoff20259.8200100

Comparison between GC/MS and Lay Person Results MET Cup

% of CutoffNumberofsamplesMethamphetamineConcentration by GC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
No. ofPositiveNo. ofNegative

{19}------------------------------------------------

-100% Cutoff200020100
-75% Cutoff20124.8020100
-50% Cutoff20251.3020100
-25% Cutoff20377.621890.0
+25% Cutoff20625.918290.0
+50% Cutoff20748.3200100
+75% Cutoff20878.1200100

Comparison between GC/MS and Lay Person Results MET Dip Card

% of CutoffNumberofsamplesMethamphetamineConcentration by GC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff20124.8020100
-50% Cutoff20251.3020100
-25% Cutoff20377.621890.0
+25% Cutoff20625.918290.0
+50% Cutoff20748.3200100
+75% Cutoff20878.1200100

Lay-users were also given surveys on the ease of understanding the package insert instructions. All lay users indicated that the device instructions can be easily followed. A Flesch-Kincaid reading analysis was performed on each package insert and the scores revealed a reading Grade Level of 7.

    1. Clinical Studies
      Not applicable.

11.Conclusion

Based on the test principle and acceptable performance characteristics including precision, cut-off, interference, specificity and method comparison of the devices, it's concluded that the WONDFO Urine Test devices are substantially equivalent to the predicate.

§ 862.3250 Cocaine and cocaine metabolite test system.

(a)
Identification. A cocaine and cocaine metabolite test system is a device intended to measure cocaine and a cocaine metabolite (benzoylecgonine) in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.(b)
Classification. Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).