K Number
K161099

Validate with FDA (Live)

Date Cleared
2017-02-23

(310 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Powder Free Nitrile Examination Gloves, Black is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

The proposed device is a Powder Free Nitrile Examination Glove. The glove is available in one color: Black. The proposed device is non-sterile. The Powder Free Nitrile Examination Gloves, Black meets all the requirements of ASTM Specification D6319-10 - Standard Specification for Nitrile Examination Gloves for Medical Application.

AI/ML Overview

The document discusses the substantial equivalence of "Powder Free Nitrile Examination Gloves, Black" (K161099) to a legally marketed predicate device. The study described focuses on non-clinical performance data to demonstrate that the device meets established acceptance criteria.

1. Table of Acceptance Criteria and Reported Device Performance:

CharacteristicStandard (Acceptance Criteria)Reported Device Performance
DimensionsASTM D6319-10: Length ≥ 230mm, Width ≥ 95 ± 10mmMeets ASTM D6319-10 Standard requirements: Minimal value for Length: 240 mm, Minimal value for Width: 95mm
Physical PropertiesASTM D6319-10:Meets ASTM D6319-10 Standard requirements:
Tensile Strength (Before Aging)≥ 14 MPaMinimal Value: 19.05 MPa
Tensile Strength (After Aging)≥ 14 MPaMinimal Value: 16.25 MPa
Ultimate Elongation (Before Aging)≥ 500%Minimal Value: 755%
Ultimate Elongation (After Aging)≥ 400%Minimal Value: 625%
ThicknessASTM D6319-10: Palm ≥ 0.05mm, Finger ≥ 0.05mmMeets standard requirements: Palm ≥ 0.09mm, Finger ≥ 0.11mm
BiocompatibilityISO 10993-10:2010 (Irritation and Skin Sensitization)Under the condition of the Study – Not an irritant, Not a sensitizer
Freedom from Pinholes21 CFR 800.20; ASTM D5151-06, AQL 2.5Meets 21 CFR 800.20 and ASTM D5151-06 Standard Requirements, Passes AQL 2.5
Powder ResidualASTM D6124-06: ≤ 2 mg/ gloveMeets ASTM D6124-06 Standard requirements: Minimal Value: 0.45 mg

2. Sample Size Used for the Test Set and Data Provenance:

The document does not explicitly state a specific "test set" sample size in terms of number of gloves tested for each characteristic. However, the data represents non-clinical performance evaluations against established standards like ASTM and ISO. The provenance of the data would be internal testing conducted by the manufacturer, PT. Maja Agung Latexindo, which is located in North Sumatera, Indonesia. The study is retrospective in the sense that the test results are presented as already completed data to demonstrate compliance.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

This information is not provided in the document. The ground truth for the performance of the gloves is established by adherence to recognized international standards (ASTM and ISO), which detail testing methodologies and acceptance criteria. The tests themselves are standardized procedures, so the "experts" would be the technicians or engineers conducting these tests in a laboratory setting, ensuring the methods are followed correctly.

4. Adjudication Method for the Test Set:

This is not applicable as the study involves standardized physical and chemical tests rather than expert review or adjudication of subjective interpretations. The outcomes are objective measurements against predefined limits.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation of medical images or data, which is not the case for examination gloves.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

No, this question is not applicable. The device is a physical product (examination gloves), not an algorithm or AI system.

7. The Type of Ground Truth Used:

The ground truth used for determining the device's performance is objective measurements derived from established international standards (ASTM D6319-10, ASTM D5151-06, ASTM D6124-06, and ISO 10993-10:2010). These standards define the acceptable range for various physical, chemical, and biological properties of examination gloves.

8. The Sample Size for the Training Set:

This question is not applicable. The device is not an AI or machine learning model that requires a training set. The "training" here refers to the manufacturing process and quality control measures to ensure consistent product attributes.

9. How the Ground Truth for the Training Set was Established:

This question is not applicable for the reasons stated above. The "ground truth" for the manufacturing process is established by internal quality control procedures and adherence to the specifications outlined in the referenced standards.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of three human profiles facing to the right, with a design above them that resembles a bird in flight.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 23, 2017

Pt. Maja Agung Latexindo % Emmy Tjoeng Marketing Director Shamrock Manufacturing Co. Inc. 5445 Daniels Street Chino. California 91710

Re: K161099

Trade/Device Name: Powder Free Nitrile Examination Gloves, Black Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZA Dated: January 17, 2017 Received: January 19, 2017

Dear Emmy Tjoeng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Michael J. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K161099

Device Name

Powder Free Nitrile Examination Gloves, Black

Indications for Use (Describe)

The Powder Free Nitrile Examination Gloves, Black is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for PT. MAJA AGUNG LATEXINDO, a manufacturer of latex gloves. The logo consists of a red square with a stylized white "M" inside, followed by the company name in green, bold letters. Below the company name is the text "Manufacturer of Latex Gloves" in a smaller, black font.

Image /page/3/Picture/1 description: The image contains four certification logos. The first logo is the TUV CERT EN ISO 9001 certification mark. The second logo is the TUV NORD ISO 9001:2008 certification mark. The third logo is the TUV NORD ISO 13485:2007 certification mark with the certificate number 44 221 06 336658-001. The fourth logo is the CE marking with the number 0044, indicating compliance with the Medical Device Directive 93/42/EEC Annex V, and the certificate number 44 235 06 336658-001.

"510 (K)" SUMMARY K161099

1.0 Submitter
Name of applicant: Prakash
Address: PT. MAJA AGUNG LATEXINDO
Jl. Utama No. 98, Pujimulyo, Sunggal 20352
North Sumatera - Indonesia
Phone No.: 62-61-8459170
Fax No.: 62-61-8459180
Contact person in U.S.A: Emmy Tjoeng
Fax No.: 909-591-8878
Email: emmyt@smcgloves.com

2.0 Name of the Device

Powder Free Nitrile Examination Gloves, Black
Common Name : Nitrile Exam Gloves
Classification Name : Patient Examination Gloves (21 CFR 880.6250)
Product Code : LZA
Date Prepared : February 23, 2017

3.0 Identification of The Legally Marketed Devices That equivalency is claimed:

Primary Predicate:

Chlorinated Powder Free Nitrile Examination Gloves (Black Color) Company: Worldmed Manufacturing Sdn.Bhd 510(k): K123116 Regulatory Class I Product Code: LZA.

4.0 Description of the Device:

The proposed device is a Powder Free Nitrile Examination Glove. The glove is available in one color: Black. The proposed device is non-sterile. The Powder Free Nitrile Examination Gloves, Black meets all the requirements of ASTM Specification D6319-10 - Standard Specification for Nitrile Examination Gloves for Medical Application.

41

Factory :

Jl. Utama No. 98 Pujimulyo, Sunggal 20352 Deli Serdang, Sumatera Utara - Indonesia, Phone +62-61 - 8459170 (Hunting) Fax: +62-61 - 8459180 E-mail : mal2510@indosat.net.id

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for PT. MAJA AGUNG LATEXINDO, a manufacturer of latex gloves. The logo features a stylized red and blue graphic on the left, with the company name in green text to the right. Below the company name, the text "Manufacturer of Latex Gloves" is written in a smaller, black font.

Image /page/4/Picture/1 description: The image shows four different certification logos. The first logo is the TUV CERT EN ISO 9001 certification. The second logo is the TUV NORD ISO 9001:2008 certification. The third logo is the TUV NORD ISO 13485:2007 certification with the certificate number 44 221 06 336658-001. The fourth logo is the CE Medical Device Directive 93/42/EEC Annex V certification with the number 0044 and certificate number 44 235 06 336658-001.

5.0 Indications for Use:

The Powder Free Nitrile Examination Gloves, Black is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

6.0 Summary of the Technological Characteristics of the Device:

The Powder Free Nitrile Examination Gloves, Black are summarized with the following technological characteristics and is substantially equivalent to the predicate device with regard to physical characteristics, design, product features, and intended use. Both gloves are made with nitrile and meets ASTM D6319-10 - Standard Specification for Nitrile Examination Gloves for Medical Application or equivalent standards.

CHARACTERISTICSSTANDARDSDEVICE PERFORMANCE
DimensionsASTM D6319-10:Length ">–min. 230mmWidth ">–min. 95 ± 10mmMeets ASTM D6319-10Standard requirementsMinimal value for Length:240 mmMinimal value for Width: 95mm
Physical PropertiesASTM D6319-10:Tensile Strength:Before Aging: 14 MPaAfter Aging: 14 MPAUltimate Elongation:Before Aging: min. 500%After Aging: min. 400%Meets ASTM D6319-10Standard requirementsTensile Strength:Before Aging: Minimal Value:19.05 MPaAfter Accelerated Aging:Minimal Value: 16.25 MPaUltimate Elongation:Before Aging Minimal Value:755%After Aging Minimal Value:625%
ThicknessASTM D6319-10Palm ">– min 0.05mmFinger ">– min 0.05mmMeets standard requirementsStandard requirementsPalm ">– min. 0.09mmFinger ">– min. 0.11mm

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for PT. MAJA AGUNG LATEXINDO. The logo consists of two parts: a red abstract graphic on the left and the company name in green on the right. The company name is written in a bold, sans-serif font, with "PT. MAJA AGUNG" on the top line and "LATEXINDO" on the bottom line.

Image /page/5/Picture/1 description: The image shows four different certification logos. The first logo is a TUV CERT EN ISO 9001 certification. The second logo is an ISO 9001:2008 TUV NORD certification. The third logo is an ISO 13485:2007 TUV NORD certification with the certificate number 44 221 06 336658-001. The fourth logo is a CE 0044 Medical Device Directive 93/42/EEC Annex V certification with the certificate number 44 235 06 336658-001.

Manufacturer of Latex Gloves

BiocompatibilityISO 10993-10:2010Biological evaluation ofmedical devices - Part 10:Tests for irritation and skin sensitizationUnder the condition of the Study –Not an irritant
ISO 10993-10:2010Biological evaluation of medicaldevices -Part 10:Tests for irritation and skin sensitizationUnder the condition of theStudy -Not a sensitizer
Freedom fromPinholes21 CFR 800.20; ASTM D5151-06AQL 2.5Meets 21 CFR 800.20 and ASTMD5151 -06 StandardRequirementsPasses AQL 2.5
PowderResidualASTM D6124-06:$\leq$ 2 mg/ gloveMeets ASTM D6124-06Standard requirementsMinimal Value: 0.45 mg

7.0 Substantial Equivalent Based on Assessment of Non-Clinical Performance Data

The performance test data of non-clinical tests that support a determination of substantial equivalence is the same as described in Section 6.0 above and meets the following standards:

ASTM D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application.

ASTM D5151-06 (Reapproved 2011), Standard Test Method for Detection of Holes in Medical Gloves.

ASTM D6124-06 (Reaffirmation 2011), Standard Test Method for Residual Powder on Medical Gloves.

ISO 10993-10: 2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.

43

Factory :

Jl. Utama No. 98 Pujimulyo, Sunggal 20352 Deli Serdang, Sumatera Utara - Indonesia, Phone +62-61 - 8459170 (Hunting) Fax: +62-61 - 8459180 E-mail : mal2510@indosat.net.id

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for PT. MAJA AGUNG LATEXINDO. The logo consists of a stylized red graphic on the left, followed by the company name in green, block letters. The words "PT. MAJA AGUNG" are on the top line, and "LATEXINDO" is on the bottom line.

Image /page/6/Picture/1 description: The image shows four different certification logos. The first logo is the TUV CERT EN ISO 9001 certification. The second logo is the TUV NORD ISO 9001:2008 certification. The third logo is the TUV NORD ISO 13485:2007 certification with the certificate number 44 221 06 336658-001. The fourth logo is the CE 0044 Medical Device Directive 93/42/EEC Annex V certification with the certificate number 44 235 06 336658-001.

Manufacturer of Latex Gloves

8.0 Substantial Equivalence Comparison of K161099

The Powder Free Nitrile Examination Gloves, Black is substantially equivalent to the predicate device with respect to intended use, product features and the technological characteristics. The substantial equivalence comparison is presented in the table below:

CharacteristicandParametersChlorinated Powder FreeNitrile Examination Gloves(Black Color)Bhd. K123116(Predicate)Powder Free NitrileExamination Gloves,Black. K161099(Proposed)SubstantialEquivalence(SE)
Product CodeLZALZASAME
Intended useIntended for medical purposesthat is worn on the examiner'shands or fingers to preventcontamination betweenpatient and examiner.Intended for medicalpurposes that is worn onthe examiner's hand orfinger to preventcontamination betweenpatient and examiner.SAME
Indications forUse StatementThe Chlorinated Powder FreeNitrile Examination Gloves(Black Color) is a single usedisposable device intended formedical purposes that is wornon the hand of healthcare andsimilar personnel to preventcontamination betweenhealthcare personnel and thepatientThe Powder Free NitrileExaminationGloves,Black is a disposabledevice intendedformedical purposes that isworn on the examiner'shand or finger to preventcontaminationbetweenpatient and examiner.SAME
Device MaterialNitrile Synthetic LatexNitrile Synthetic LatexSAME
ColorBlackBlackSAME
AdditivesNo flavor additiveNo flavor additiveSAME
Instruction forUse on LabelingSingle Use OnlySingle Use OnlySAME
ConstructionAmbidextrousAmbidextrousSAME
SterilityNon-sterileNon-sterileSAME
AcceptanceCriteriaASTM D6319-10ASTM D6319-10SAME
DimensionsMeets ASTM D 6319-10Meets ASTM D 6319-10SAME
Thickness at Finger - min.0.05mmThickness at Palm - min.0.05mmOverall Length – min.230mmWidth(±5mm)Size XS - 75mmSize S - 85mmSize M - 95mmSize L - 110mmSize XL - 115mmThickness at Finger - min.0.05mmThickness at Palm - min.0.05mm
PhysicalPropertiesMeets ASTM D6319-10:Before AgingTensile strength -min.14.0MPaUltimate Elongation -min.500%After AgingTensile Strength – min14.0MPaUltimate Elongation -min.400%
BiocompatibilityTesta. Irritation Testsb. SkinSensitizationTestsUnder the conditions ofthe study, the subjectdevice is a non-irritant andnon-sensitizer.(ISO 10993-10 (2010))Under the conditions ofthe study, the subjectdevice is a non-irritantand non-sensitizer.(ISO 10993-10 (2010))SAME
Residual Powder TestMeets ASTM D6124-06:≤ 2 mg/ gloveMeets ASTM D6124-06:≤ 2 mg/ gloveSAME
Freedom from HolesMeets ASTM D5151-06AQL 2.5Meets ASTM D5151-06Passes AQL 2.5SAME

44

Factory :

Jl. Utama No. 98 Pujimulyo, Sunggal 20352 Deli Serdang, Sumatera Utara - Indonesia, Phone +62-61 - 8459170 (Hunting) Fax: +62-61 - 8459180 E-mail : mal2510@indosat.net.id

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the logo for PT. MAJA AGUNG LATEXINDO. The logo consists of two parts: a stylized red graphic on the left and the company name in green on the right. The company name is written in two lines, with "PT. MAJA AGUNG" on the top and "LATEXINDO" on the bottom.

Image /page/7/Picture/1 description: The image shows four certification logos. The first logo is the TUV CERT EN ISO 9001 certification. The second and third logos are TUV NORD ISO 9001:2008 and ISO 13485:2007 certifications with certificate numbers 44 221 06 336658-001. The fourth logo is the CE 0044 Medical Device Directive 93/42/EEC Annex V certification with certificate number 44 235 06 336658-001.

Manufacturer of Latex Gloves

Factory : II. Utama No. 98 Pujimulyo, Sunggal 20352 Deli Serdang, Sumatera Utara - Indonesia, Phone +62-61 - 8459170 (Hunting) Fax: +62-61 - 8459180 E-mail : mal2510@indosat.net.id

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for PT. MAJA AGUNG LATEXINDO, a manufacturer of latex gloves. The logo features a stylized red graphic on the left, followed by the company name in green text. Below the company name is the text "Manufacturer of Latex Gloves".

Image /page/8/Picture/1 description: The image contains four certification logos. The first logo is a TUV CERT EN ISO 9001 certification mark. The second logo is an ISO 9001:2008 certification mark with TUV NORD. The third logo is an ISO 13485:2007 certification mark with TUV NORD and the certificate number 44 221 06 336658-001. The fourth logo is a CE mark with the number 0044, indicating compliance with the Medical Device Directive 93/42/EEC Annex V, and the certificate number 44 235 06 336658-001.

9.0 Conclusion

The Powder Free Nitrile Examination Gloves, Black meet all of the requirements of FDA-recognized consensus standards; ASTM D6319-10, ASTM D5151-06, ASTM D6124-06 and ISO 10993-10:2010 and meet our labeling claims and pinhole acceptable quality level (AQL) as shown above.

There are no significant differences between the subject and predicate devices and the two are similar in terms of intended use, materials, and performance.

The conclusion drawn from the nonclinical tests demonstrate that the device is as safe and as effective and performs as well as the legally marketed device.

Factory : Jl. Utama No. 98 Pujimulyo, Sunggal 20352 Deli Serdang, Sumatera Utara - Indonesia, Phone +62-61 - 8459170 (Hunting) Fax: +62-61 - 8459180 E-mail : mal2510@indosat.net.id

46

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.