K Number
K160978

Validate with FDA (Live)

Date Cleared
2016-07-08

(92 days)

Product Code
Regulation Number
880.5860
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LITe® BIO Delivery System is intended to deliver autograft, allograft, or synthetic bone grafting materials to all orthopedic surgical sites.

Device Description

The LITe® BIO Delivery System is a manually operated hand-held device used to facilitate delivery of bone graft material to all orthopedic surgical sites. The LITe® BIO Delivery System consists of a delivery cannula, a delivery plunger, a delivery plunger tip, a loading syringe, and a delivery handle. The cannula is filled with the physician's choice of bone graft material and is then delivered to the intended surgical site.

AI/ML Overview

The provided text describes the 510(k) summary for the LITe® BIO Delivery System, a medical device. This document is a premarket notification for a Class II piston syringe intended to deliver bone grafting materials.

The document does not describe a study involving an AI/Machine Learning device. It refers to a physical medical device for delivering bone graft material. Therefore, the requested information regarding acceptance criteria, performance metrics, sample sizes for test/training sets, ground truth establishment, expert adjudication, or MRMC studies for an AI/ML device is not applicable to this document.

The "Summary of Performance Data" section lists several bench tests conducted to demonstrate the device's performance, but these are engineering performance criteria for a physical device, not an AI/ML algorithm.

Summary of Performance Data Provided:

The document states:
"Testing was conducted to demonstrate performance as intended and included the following bench testing:

  • Functional testing (delivery of graft material through cannula)
  • Life cycle testing
  • Cantilever bending of cannula
  • Radial compression of cannula
  • Retention strength test of cannula
  • Connection strength test of cannula to loading funnel"

This information aims to demonstrate that the LITe® BIO Delivery System is substantially equivalent to other currently marketed bone graft delivery systems and that its technological differences do not raise any new safety or effectiveness issues.

Therefore, I cannot provide the requested information about acceptance criteria and studies for an AI/ML device based on this document.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features an abstract design of a stylized eagle or bird-like figure with three heads in profile, representing the department's focus on health, human services, and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the emblem.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 8, 2016

Stryker Corporation Ms. Meriam Gabera Senior Regulatory Affairs Specialist 2 Pearl Court Allendale, New Jersey 07401

Re: K160978

Trade/Device Name: Lite Bio Delivery System Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF Dated: April 6, 2016 Received: April 7, 2016

Dear Ms. Gabera:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Jennifer R. Stevenson -A

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K160978

Device Name LITe® BIO Delivery System

Indications for Use (Describe)

The LITe® BIO Delivery System is intended to deliver autograft, allograft, or synthetic bone grafting materials to all orthopedic surgical sites.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary as required by 21CFR§807.92(c): LITe® BIO Delivery System
SubmitterStryker Spine2 Pearl CourtAllendale, NJ 07401
Contact PersonMeriam Gabera, MSSenior Regulatory Affairs SpecialistPhone: 201-749-8043Fax: 201-831-3000E-mail: meriam.gabera@stryker.com
Date PreparedJuly 1, 2016
Trade NameLITe® BIO Delivery System
Common NameSyringe, piston
Proposed ClassClass II
Classification Name,CodificationPiston syringe, 21 CFR § 880.5860
Product CodesFMF
Predicate DevicesPrimary Predicate:• Arthrex Mixing and Delivery System - K121124Additional Predicates:• Biomet Graft Delivery Syringes - K140710• NovaBone MIS Cartridge Delivery System - K112773Reference Devices:• Imbibe Bone Marrow Aspiration Syringe – K023074• Aero-AL Lumbar Cage System – K133328• Aero-C Cervical Cage System – K152532
Device DescriptionThe LITe® BIO Delivery System is a manually operated hand-held deviceused to facilitate delivery of bone graft material to all orthopedicsurgical sites. The LITe® BIO Delivery System consists of a deliverycannula, a delivery plunger, a delivery plunger tip, a loading syringe, anda delivery handle. The cannula is filled with the physician's choice ofbone graft material and is then delivered to the intended surgical site.
Indications for UseThe LITe® BIO Delivery System is intended to deliver autograft, allograft,or synthetic bone graft materials to all orthopaedic surgical sites.
Summary ofTechnologicalCharacteristicsThe LITe® BIO Delivery System shares similar materials, design, andfundamental scientific technologies as the predicate devices. Anassessment of the technological differences was completed and thefollowing differences were identified:• The design of the primary predicate system allows formixing/preparing the bone graft mixture within the device asopposed to using an external surgical dish for preparing thebone graft material.
while the subject device is supplied with both sterile, single usedevices and non-sterile, reusable instruments.
The comparison of the LITe® BIO Delivery System against the primarypredicate system demonstrated that the identified design differences donot impact the functionality or performance of the subject system,supporting the claim of substantial equivalence.
Summary ofPerformance DataTesting was conducted to demonstrate performance as intended andincluded the following bench testing:•Functional testing (delivery of graft material through cannula)•Life cycle testing•Cantilever bending of cannula•Radial compression of cannula•Retention strength test of cannula•Connection strength test of cannula to loading funnelThe testing verifies that the subject device is substantially equivalent toother currently marketed bone graft delivery systems and anytechnological differences do not raise any new issues regarding safetyand effectiveness.
ConclusionBased upon a comparison of intended use, technological characteristics,and device performance in the non-clinical testing listed above, theLITe® BIO Delivery System has demonstrated substantial equivalence tothe identified predicate device systems.

{4}------------------------------------------------

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).