K Number
K043535

Validate with FDA (Live)

Date Cleared
2005-01-06

(15 days)

Product Code
Regulation Number
892.1550
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The HD11 is intended for diagnostic ultrasound imaging and fluid flow analysis. Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body.

Device Description

The HD11 is a diagnostic ultrasound device. It consists of a system console containing the power supply and electronic circuitry required to generate the image, a display screen and a connection to the separate transducers. It is substantially equivalent to other ultrasound systems including the Philips M2540A EnVisor and HDI-5000 series Ultrasound systems with transducers.

The 3D8-4 and L15-7io transducers are substantially equivalent to other Philips sector, linear and endo-cavity ultrasound transducers.

The HD11 system and transducers function in a manner identical to all ultrasound systems and transducers. The system circuitry generates an electronic voltage pulse, which is transmitted to the transducer. In the transducer, a piezo-electric array converts the electronic pulse into an ultrasonic pressure wave. When coupled to the body, the pressure wave transmits through body tissues. The differing acoustic properties of the tissues in the body reflect some of the transmitted energy back to the transducer, where it is converted back to electrical signals and sent back to the system. In the system, advanced signal processing technologies convert the returned signals into images of the tissues. The Doppler functions of this system process the Doppler shift frequencies from the echoes of moving targets (such as blood), to detect and graphically display the Doppler shifts of these tissues as flow.

AI/ML Overview

The provided text describes the Philips Ultrasound HD11 Diagnostic Ultrasound System and its transducers as substantially equivalent to predicate devices. It does not contain information about specific acceptance criteria, study methodologies, or performance metrics beyond broad claims of "safety and effectiveness."

Therefore, I cannot fulfill your request for:

  • A table of acceptance criteria and reported device performance: No specific performance criteria or results are detailed. The document relies on substantial equivalence to predicate devices rather than proving performance against new criteria.
  • Sample size used for the test set and data provenance: No test set information or data provenance is provided.
  • Number of experts used to establish ground truth and their qualifications: Not applicable, as no a ground truth establishment process is described.
  • Adjudication method: Not applicable.
  • MRMC comparative effectiveness study: No such study is mentioned.
  • Standalone performance study: No standalone performance study information is provided.
  • Type of ground truth used: Not applicable, as no ground truth is described.
  • Sample size for the training set: No training set information is provided.
  • How the ground truth for the training set was established: Not applicable.

Summary of Device and Equivalence Claim (as provided in the document):

The Philips Ultrasound HD11 Diagnostic Ultrasound System is a diagnostic ultrasound device designed for imaging and fluid flow analysis. It is claimed to be substantially equivalent in safety and effectiveness to the Philips M2540A EnVisor and HDI-5000 series Ultrasound systems with transducers (predicate devices).

The basis for this substantial equivalence includes:

  • Identical intended uses (diagnostic ultrasonic imaging and fluid flow analysis).
  • Same gray-scale and Doppler capabilities.
  • Essentially the same technologies for imaging, Doppler functions, and signal processing.
  • Acoustic output levels below Track 3 FDA limits.
  • Manufactured under equivalent quality systems.
  • Manufactured with materials of equivalent biosafety.
  • Designed and manufactured to the same electrical and physical safety standards.

The document lists various transducers and their specific indications for use, indicating which applications are "new indications" (N), "previously cleared by FDA" (P), or "added under Appendix E" (E) for different imaging modes (B, M, PWD, CWD, Color Doppler, Combined, Other). The modes listed under "Other" include advanced features like SonoCT, X-Res, Angio, Panoramic, 3-D/4-D Imaging, Directional Angio Imaging, Tissue Doppler Imaging, IMT Measures, iSCAN, 2D Doppler, Color Power Angio, Fetal Heart (Spatial Temporal Image Correlation), and Real Time Biopsy (for specific transducers).

{0}------------------------------------------------

K0435535

JAN - 6 2005

Summary of Safety and Effectiveness 3.2

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 as implemented in 21 C.F.R. §807.92.

The submitter of this premarket notification is:

Steve Singlar Regulatory Submissions Manager Philips Ultrasound MS 0135 3000 Minuteman Road Andover, MA 01810 Tel: (978) 659-2101 Fax: (978) 975-7324

This summary was prepared on November 12, 2004.

The proprietary name of the device is the HD11 Diagnostic Ultrasound System. In combination with the transducers listed in the Indications for use tables are commonly known as a diagnostic . ultrasound system and transducers.

These devices are classified as follows:

90IYNUltrasonic Pulsed Doppler Imaging System
90IYOUltrasonic Pulsed Echo Imaging System
90ITXDiagnostic Ultrasound Transducer

As stated in 21 CFR, parts 892.1550, 892.1560 and 892.1570, each of these generic types of devices have been classified as Class II.

The HD11 is a diagnostic ultrasound device. It consists of a system console containing the power supply and electronic circuitry required to generate the image, a display screen and a connection to the separate transducers. It is substantially equivalent to other ultrasound systems including the Philips M2540A EnVisor and HDI-5000 series Ultrasound systems with transducers.

The 3D8-4 and L15-7io transducers are substantially equivalent to other Philips sector, linear and endo-cavity ultrasound transducers.

{1}------------------------------------------------

The HD11 system and transducers function in a manner identical to all ultrasound systems and transducers. The system circuitry generates an electronic voltage pulse, which is transmitted to the transducer. In the transducer, a piezo-electric array converts the electronic pulse into an ultrasonic pressure wave. When coupled to the body, the pressure wave transmits through body tissues. The differing acoustic properties of the tissues in the body reflect some of the transmitted energy back to the transducer, where it is converted back to electrical signals and sent back to the system. In the system, advanced signal processing technologies convert the returned signals into images of the tissues. The Doppler functions of this system process the Doppler shift frequencies from the echoes of moving targets (such as blood), to detect and graphically display the Doppler shifts of these tissues as flow.

The HD11 is intended for diagnostic ultrasound imaging and fluid flow analysis.

The HD11 is substantially equivalent in safety and effectiveness to the predicates identified above:

  • Both the predicate device and the HD11 are indicated for the diagnostic ultrasonic imaging ବି and fluid flow analysis.
  • . Both the predicate device and the HD11 have the same gray-scale and Doppler capabilities.
    • Both the predicate device and the HD11 use essentially the same technologies for imaging, , ● Doppler functions and signal processing.
    • Both the predicate device and the HD11 have acoustic output levels below the Track 3 . FDA limits.
    • Both the predicate device and the HD11 are manufactured under equivalent quality systems. .
    • Both the predicate device and the HD11 are manufactured of materials with equivalent . biosafety. The materials have been evaluated and found to be safe for this application.
    • Both the predicate device and HD11 are designed and manufactured to the same electrical . and physical safety standards.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wing. The eagle faces right. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN - 6 2005

Philips Ultrasound % Ms. Laura Danielson Responsible Third Party Official TÜV Product Service 1775 Old Highway 8 NEW BRIGHTON MN 55112-1891

Re: K043535

Trade Name: HD11 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: 90 IYN, IYO, and ITX Dated: December 11, 2004 Received: December 22, 2004

Dear Ms. Danielson:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your boouse as substantially equivalentially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Interstate continered problem to may 20, 1978, etc. ssified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to I cloudi i ood, Drug, and Cooments Free Act. The general controls provisions of the Act include the general controls provisions of and listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for I ins determination of substainter equirantee applied in your premarket notification:

Transducer Model Number

S3-1 (21711A) Sector Transducer S4-2 (989605344981) Sector Transducer PA 4-2 (21422A) Sector Transducer

{3}------------------------------------------------

S8-3 (21750A) Sector Transducer S12-4 (21780A) Sector Transducer T6H (21378A) Omni III Transesophagel Transducer S7-2omni (21778A) Omni III Sector Transducer S7-3t (21781A) Minimulti Sector Transducer D2cwc (4000-0947-01) 1.9 MHz non-imaging Pencil Probe D5cwc (4000-0950-01) 5.0 MHZ non-imaging Pencil Probe D2tcd (8500-1860-01) 2.0 MHz non-imaging Pencil Probe C8-5 (8500-1664-01) Curved Linear Transducer C5-2 (21426A) Curved Linear Transducer C9-4 (8500-1658-01) Curved Linear Transducer C8-4y (21437A) Curved Linear Transducer C9-5ec (8500-1655-01) Endocavity Transducer L12-5 (8500-1660-01) Linear Transducer L12-3 (21475A0 Linear Transducer L8-4 (8500-1659-01) Linear Transducer L15-7io (989605341571) Linear Intra-operative Transducer 3D8-4 (989605345331) Curved Linear Transducer 3D6-2 (8500-2023-01) Curved Linear Transducer 3D9-3V (8500-1715-01) Curved Linear Transducer

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.

The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:

{4}------------------------------------------------

Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212.

Sincerely yours,

David A. Logeman

for

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure(s)

{5}------------------------------------------------

Indications for Use Tables 4.3.2

DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM

510(k) Number: K

System: Philips Ultrasound HD11 Diagnostic Ultrasound System

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmicNNNNNNN
Fetal Imaging& OtherFetalNNNNNNN
AbdominalNNNNNNN
Intra-operative(Epicardial/Vascular)NNNNNNN
Intra-operative (Neuro)NNNNNNN
Laparoscopic
PediatricNNNNNNN
Small Organ (Thyroid,Scrotum, Breast)NNNNNNN
Neonatal CephalicNNNNNNN
Adult CephalicNNNNNNN
Trans-rectalNNNNNNN
Trans-vaginalNNNNNNN
Trans-urethral
Trans-esoph. (non-Card.)
Intra-luminal
Other (Gynecological)NNNNNNN
CardiacCardiac AdultNNNNNNN
Cardiac PediatricNNNNNNN
Trans-esoph. (Cardiac)NNNNNNN
Other (Fetal)NNNNNNN
PeripheralVesselPeripheral vesselNNNNNNN
Other (Carotid, I/O)NNNNNNN
Musculo-skel (conventional)NNNNNNN
Musculo-skel (superficial)NNNNNNN

N= new indication; P= previously cleared by FDA; E= added under Appendix E

*Other modes: Angio, Panoranics (Tissue & Contrast), 3-D/4-D Imaging, Directional Angio Imaging, Tissue Doppler Imaging, 4D Fetal Heart, STIC (Spatial Temporal Image Correlation), SonoCT, X-Res, IMT Measures, iSCAN, 2D Doppler, Color Power Angio.

Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color, Dual Mode,

Previous submission: None

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

Daniel A. Seymour

(Division Sign-Off) Division of Reproductive, Abdominal ano Radiological Devices 5) Yk) Number ____

{6}------------------------------------------------

510(k) Number: K

System: Philips Ultrasound HD11 Diagnostic Ultrasound System

S3-1 (21711A) Sector transducer Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
FetalNNNNNNN
AbdominalNNNNNNN
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
Fetal Imaging& OtherPediatricNNNNNNN
Small Organ (Scrotum,Thyroid, Breast)
Neonatal Cephalic
Adult CephalicNNNNNNN
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Intra-luminal
Other (Gynecological)
Cardiac AdultNNNNNNN
CardiacCardiac PediatricNNNNNNN
Trans-esoph. (Cardiac)
Other (Specify)
PeripheralVesselPeripheral vessel
Other (Carotid)
Musculo-skel (conventional)
Musculo-skel (superficial)

new indication; P= previously cleared by FDA; E= added under Appendix E

*Other modes: X-Res, Angio, Panoramic, Harmonics (Tissue & Contrast), 3-D Freehand Imaging, Directional Angio Imaging, Tissue Doppler Imaging, Color Power Angio

Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color, Dual

Previous submission: Cleared in K030455

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

David A. Bergman

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________

{7}------------------------------------------------

510(k) Number: K

System: Philips Ultrasound HD11 Diagnostic Ultrasound System

S4-2 (989605344981) Sector Transducer Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmicNNNNNNN
FetalNNNNNNN
AbdominalNNNNNNN
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
Fetal Imaging& OtherPediatricNNNNNNN
Small Organ (Scrotum,Thyroid, Breast)
Neonatal Cephalic
Adult CephalicNNNNNNN
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Intra-luminal
Other (Gynecological) GYNNNNNNNN
Cardiac AdultNNNNNNN
CardiacCardiac PediatricNNNNNNN
Trans-esoph. (Cardiac)
Other (Specify)
PeripheralVesselPeripheral vesselNNNNNNN
Other (Carotid)
Musculo-skel (conventional)
Musculo-skel (superficial)

N= new indication; P= previously cleared by FDA; E= added under Appendix E

*Other modes: X Res, Angio, Panoramic, Harmonics (Tissue & Contrast), 3-D Imaging, Directional Angio Imaging, Tissue Doppler Imaging

Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color, Dual

Previous submission: Cleared as PA4-2 in K014191

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

Daniel R. Seymore

(Division Sign-Off) Circision of Reproductive, Abdon and Radiological Devices 510(k) Number ___

{8}------------------------------------------------

510(k) Number: K

System: Philips Ultrasound HD11 Diagnostic Ultrasound System

PA 4-2 (21422A) Sector transducer Transducer:

I ransuacer .
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmicNNNNNNN
FetalNNNNNNN
AbdominalNNNNNNN
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
Fetal Imaging& OtherPediatricNNNNNNN
Small Organ (Scrotum,Thyroid, Breast)
Neonatal Cephalic
Adult CephalicNNNNNNN
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Intra-luminal
Other (Gynecological)NNNNNNN
Cardiac AdultNNNNNNN
CardiacCardiac PediatricNNNNNNN
Trans-esoph. (Cardiac)
Other (Specify)
PeripheralVesselPeripheral vessel
Other (Carotid)
Musculo-skel (conventional)
Musculo-skel (superficial)
N= new indication; P= previously cleared by FDA; E= added under Appendix E

11 Thew Indication, 1 - provider, Harmonics (Tissue & Contrast), 3-D Imaging, Directional Angio Imaging, Tissue Doppler Imaging

Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color, Dual

Previous submission: K014191

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) (1 BE. 101 De 1 Center for Devices and Radiological Health, Office of Device Evaluation

Prescription Use (Per 21 CFR 801.109)

David A. Leyson

:

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 5 PORQ Number ________________________________________________________________________________________________________________________________________________________________

{9}------------------------------------------------

510(k) Number: K

System: Philips Ultrasound HD11 Diagnostic Ultrasound System

Transducer: S8-3 (21750A) Sector Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
FetalNNNNNN
AbdominalNNNNNNN
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
Fetal Imaging& OtherPediatricNNNNNNN
Small Organ (Scrotum,Thyroid, Breast)
Neonatal CephalicNNNNNNN
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Intra-luminal
Other (Gynecological)NNNNNNN
Cardiac AdultNNNNNNN
CardiacCardiac PediatricNNNNNNN
Trans-esoph. (Cardiac)
Other (Fetal Heart)NNNNNNN
PeripheralVesselPeripheral vesselNNNNNNN
Other (Carotid)
Musculo-skel (conventional)
Musculo-skel (superficial)

iously cleared by FDA; E= added under Appendix E

*Other modes: X-res, Angio, Panoramic, 3-D Imaging, Directional Angio Imaging, Tissue Doppler Imaging, Color Power Angio

Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color, Dual

Previous submission: Cleared as s8 (21350A) in K014191

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

David A. Larson

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________

{10}------------------------------------------------

510(k) Number: K

System: Philips Ultrasound HD11 Diagnostic Ultrasound System

S12-4 (21780A) Sector transducer Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal
Abdominal
Intra-operative (Cardiac)NNNNNNN
Intra-operative (Neuro)
Laparoscopic
Fetal Imaging& OtherPediatricNNNNNNN
Small Organ (Scrotum,Thyroid, Breast)
Neonatal CephalicNNNNNNN
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Intra-luminal
Other (Gynecological)
CardiacCardiac AdultNNNNNNN
Cardiac PediatricNNNNNNN
Trans-esoph. (Cardiac)
Other (Specify)
PeripheralVesselPeripheral vesselNNNNNNN
Other (Carotid)
Musculo-skel (conventional)
Musculo-skel (superficial)

N= new indication; P= previously cleared by FDA; E= added under Appendix E

*Other modes: X-Res, Angio, Panoramic, 3-D Inaging, Directional Angio Imaging, Tissue Doppler Imaging, Color Power Angio

Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color, Dual

Previous submission: Cleared as s12 (21380A) in K014191, K971116

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

Daniel A. Lyons

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________

{11}------------------------------------------------

510(k) Number: K

System: Philips Ultrasound HD11 Diagnostic Ultrasound System T6H (21378A) Omni III Transesophagel transducer Transducer: Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
Fetal Imaging& OtherPediatric
Small Organ (Scrotum,Thyroid, Breast)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Intra-luminal
Other (Gynecological)
Cardiac AdultNNNNNNN
CardiacCardiac PediatricNNNNNNN
Trans-esoph. (Cardiac)NNNNNNN
Other (Specify)
PeripheralVesselPeripheral vessel
Other (Carotid)
Musculo-skel (conventional)
Musculo-skel (superficial)
N= new indication: P= previously cleared by FDA; E= added under Appendix E

*Other modes: X-Res, Angio, , Harmonics (Tissue & Contrast), 3-D Imaging, Directional Angio Imaging, Tissue Doppler Imaging

Combined modes: Duplex = 2D + Doppler, Triplex = 2D + Doppler + Color, Dual

Previous submission: K014191, K022303

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

David R. Lyson


Division Sign Off

(Division Sign-Off) Division of Reproductive, Abdomina and Radiningical Devices 5 FO(k) Nomber __

{12}------------------------------------------------

510(k) Number: K

System: Philips Ultrasound HD11 Diagnostic Ultrasound System

S7-2omni (21778A) Omni III Sector transducer Transducer:

I ransuler : . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Clinical ApplicationMode of Operation
SpecificBMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
General(Track I Only)(Tracks I & III)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
Fetal Imaging& OtherPediatric
Small Organ (Scrotum,Thyroid, Breast)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Intra-luminal
Other (Gynecological)
Cardiac AdultNNNNNNN
CardiacCardiac PediatricNNNNNNN
Trans-esoph. (Cardiac)NNNNNNN
Other (Specify)
PeripheralVesselPeripheral vessel
Other (Carotid)
Musculo-skel (conventional)
Musculo-skel (superficial)

N= new indication; P= previously cleared by FDA; E= added under Appendix E

  • Other modes: X-Res, Angio, Harmonics (Tissue & Contrast), 3-D Imaging, Directional Angio Imaging, Tissue Doppler Imaging, Dual, Color Power Angio

Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color, Dual

Previous submission: Cleared as 21378A in K022303

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) (1 ESABE DO FOT WEBS and Radiological Health, Office of Device Evaluation

Daniel A. Jessop

(Division Sign-Off) Division of Reproductive, A and Radiological Devices 519(k) Number __

{13}------------------------------------------------

510(k) Number: K

System: Philips Ultrasound HD11 Diagnostic Ultrasound System

Transducer: S7-3t (21781A) Minimulti Sector transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
Fetal Imaging& OtherPediatric
Small Organ (Scrotum,Thyroid, Breast)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Intra-luminal
Other (Gynecological)
Cardiac AdultNNNNNNN
CardiacCardiac PediatricNNNNNNN
Trans-esoph. (Cardiac)NNNNNNN
Other (Specify)
PeripheralVesselPeripheral vessel
Other (Carotid)
Musculo-skel (conventional)
Musculo-skel (superficial)
N= new indication; P= previously cleared by FDA; E= added under Appendix E
*Other modes: X-Res, Angio, Harmonics (Tissue & Contrast), 3-D Imaging, Directional Angio Imaging, Tissue Doppler
Imaging, Color Power Angio.

Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color, Dual

Previous submission: None

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

Daniel A. Seymour

(Division Sign-Off) Division of Reproductive, Abdominal and Pachological Device bil grammamber __

{14}------------------------------------------------

510(k) Number: K

System: Philips Ultrasound HD11 Diagnostic Ultrasound System

D2cwc (4000-0947-01) 1.9MHz non-imaging Pencil Probe Transducer:

I ransuded Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
Fetal Imaging& OtherPediatric
Small Organ (Scrotum,Thyroid, Breast)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Intra-luminal
Other (Gynecological)
Cardiac AdultN
CardiacCardiac PediatricN
Trans-esoph. (Cardiac)
Other (Specify)
PeripheralVesselPeripheral vessel
Other (Carotid)
Musculo-skel (conventional)
Musculo-skel (superficial)
N= new indication; P= previously cleared by FDA; E= added under Appendix E
*Other modes: None
Combined modes: None
Previous submission: Cleared as 21221B in K014191

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) (1 LEASE De of Center for Devices and Radiological Health, Office of Device Evaluation

Daniel A. Seymore

(Division Sign-Off) Division of Reproductive, Abdomin and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________

{15}------------------------------------------------

510(k) Number: K

System: Philips Ultrasound HD11 Diagnostic Ultrasound System

Transducer: D5cwc (4000-0950-01) 5.0 MHz non-imaging Pencil Probe

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: rate

Clinical ApplicationMode of OperationOther*(Specify)
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Scrotum,Thyroid, Breast)
Neonatal Cephalic
Adult Cephalic
Trans-rectal ·
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Intra-luminal
Other (Gynecological)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esoph. (Cardiac)
Other (Specify)
PeripheralVesselPeripheral vesselNN
Other (Carotid)
Musculo-skel (conventional)
Musculo-skel (superficial)
N= new indication; P= previously cleared by FDA; E= added under Appendix E
*Other modes: None
Combined modes: None
Previous submission: Cleared as 21223B in K014191

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

Daniel A. Swanson

(Division Sign-Off) Division of Reproductive, Abdomin and Radiological Devices 510(k) Number ___

{16}------------------------------------------------

510(k) Number: K

System: Philips Ultrasound HD11 Diagnostic Ultrasound System

D2tcd (8500-1860-01) 2.0MHz non-imaging Pencil Probe Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
Fetal Imaging& OtherPediatric
Small Organ (Scrotum,Thyroid, Breast)
Neonatal Cephalic
Adult CephalicN
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Intra-luminal
Other (Gynecological)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esoph. (Cardiac)
Other (Specify)
PeripheralVesselPeripheral vesselN
Other (Carotid)
Musculo-skel (conventional)
Musculo-skel (superficial)
N= new indication; P= previously cleared by FDA; E= added under Appendix E*Other modes: NoneCombined modes: None

| Previous submission: Cleared as D2TC in K030455

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

David R. Ingram

(Division Sign-Off) Division of Reproductive, A and Radiological Devices 51 are spinner ___

{17}------------------------------------------------

510(k) Number: K

System: Philips Ultrasound HD11 Diagnostic Ultrasound System

Transducer: C8-5 (8500-1664-01) Curved Linear Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
FetalNNNNNN
AbdominalNNNNNN
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
Fetal Imaging& OtherPediatricNNNNNN
Small Organ (Scrotum,Thyroid, Breast)
Neonatal CephalicNNNNNN
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Intra-luminal
Other (Gynecological)NNNNNN
Cardiac Adult
CardiacCardiac Pediatric
Trans-esoph. (Cardiac)
Other (Fetal)NNNNNN
PeripheralVesselPeripheral vesselNNNNNN
Other (Carotid)
Musculo-skel (conventional)
Musculo-skel (superficial)

N= new indication; P= previously cleared by FDA; E= added under Appendix E

*Other modes: SonoCT, X-Res, Angio, Panoramic, 3-D Imaging, Directional Angio Imaging, iSCAN, Doppler/2D, Color Power Angio

Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color, Dual

Previous submission: Cleared on K030455

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

David A. Hyman

(Division Sign-Off) Division of Reproductive, Ab and adiological Device 与:Number___

{18}------------------------------------------------

510(k) Number: K

System: Philips Ultrasound HD11 Diagnostic Ultrasound System

C5-2 (21426A) Curved Linear Transducer Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
FetalNNNNNN
AbdominalNNNNNN
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
Fetal Imaging& OtherPediatricNNNNNN
Small Organ (Scrotum,Thyroid, Breast)NNNNNN
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Intra-luminal
Other (Gynecological)NNNNNN
Cardiac Adult
CardiacCardiac Pediatric
Trans-esoph. (Cardiac)
Other (Specify)
PeripheralVesselPeripheral vesselNNNNNN
Other (Carotid)
Musculo-skel (conventional)
Musculo-skel (superficial)

N= new indication; P= previously cleared by FDA; E= added under Appendix E

*Other modes: SonoCT, X-Res, Angio, Panoramic, Harmonics (Tissue & Contrast), 3-D Imaging, Directional Angio Imaging, iSCAN, Doppler/2D, Color Power Angio

Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color, Dual

Previous submission: K014191

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

Daniel A. Layson

(Division Sign-Off) Division of Reproductive, Abo and Radiological Devices 510(k) Number _

{19}------------------------------------------------

510(k) Number: K

System: Philips Ultrasound HD11 Diagnostic Ultrasound System

Transducer: C9-4 (8500-1658-01) Curved Linear Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
FetalNNNNNN
AbdominalNNNNNN
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
Fetal Imaging& OtherPediatricNNNNNN
Small Organ (Scrotum,Thyroid, Breast)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Intra-luminal
Other (Gynecological)NNNNNN
Cardiac Adult
CardiacCardiac Pediatric
Trans-esoph. (Cardiac)
Other (Fetal)NNNNNN
PeripheralVesselPeripheral vesselNNNNNN
Other (Carotid)
Musculo-skel (conventional)
Musculo-skel (superficial)

indication; P= previously cleared by FDA; E= added under Appendix E

*Other modes: SonoCT, X-Res, Angio, Panoramic, 3-D Imaging, Directional Angio Imaging, Color Power Angio Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color, Dual

Previous submission: Cleared on K030455

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

David H. Lyman

(Division Sign-Off)4 Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________

{20}------------------------------------------------

510(k) Number: K

System: Philips Ultrasound HDI Diagnostic Ultrasound System

C8-4v (21437A) Curved Linear Transducer Transducer:

I l'allsuded Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetalNNNNNN
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Scrotum,Thyroid, Breast)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginalNNNNNN
Trans-urethral
Trans-esoph. (non-Card.)
Intra-luminal
Other (Gynecological)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esoph. (Cardiac)
Other (Specify)
PeripheralVesselPeripheral vessel
Other (Carotid)
Musculo-skel (conventional)
Musculo-skel (superficial)
N= new indication; P= previously cleared by FDA; E= added under Appendix E
*Other modes: SonoCT, X-Res, Angio, Panoramic, 3-D Imaging, Directional Angio Imaging, Color Power Angio
Combined modes: Duplex = 2D + Doppler Triplex = 2D + Doppler + Color, Dual

Previous submission: K014191

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) (1 EE210E DO FOT WEES and Radiological Health, Office of Device Evaluation

David A. Legron

(Division Sign-Off) Division of Reproductive, Abd and Radiological Devices 51(X) Number _

{21}------------------------------------------------

510(k) Number: K

System: Philips Ultrasound HDI I Diagnostic Ultrasound System

Transducer: C9-5ec (8500-1655-01) Endocavity Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
FetalNNNNNN
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
Fetal Imaging& OtherPediatric
Small Organ (Scrotum,Thyroid, Breast)
Neonatal Cephalic
Adult Cephalic
Trans-rectalNNNNNN
Trans-vaginalNNNNNN
Trans-urethral
Trans-esoph. (non-Card.)
Intra-luminal
Other (Gynecological)NNNNNN
Cardiac Adult
CardiacCardiac Pediatric
Trans-esoph. (Cardiac)
Other (Specify)
PeripheralVesselPeripheral vessel
Other (Carotid)
Musculo-skel (conventional)
Musculo-skel (superficial)
N= new indication; P= previously cleared by FDA; E= added under Appendix E
*Other modes: SonoCT, X-Res, Angio, Panoramic, 3-D freehand Imaging, Directional Angio Imaging, Color Power Angio
Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color, Dual
Previous submission: Cleared on K030455

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

David R. Lawson

(Division Sign-Off) Division of Reproductive, Abdom and Radiological Devices 510(k) Number

{22}------------------------------------------------

510(k) Number: K

System: Philips Ultrasound HD11 Diagnostic Ultrasound System

L12-5 (8500-1660-01) Linear Transducer Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Intended Use: Diagnostic ultrasound imaging
Clinical ApplicationMode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
Fetal
AbdominalNNNNNN
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
Fetal Imaging& OtherPediatricNNNNNN
Small Organ (Scrotum,Thyroid, Breast)NNNNNN
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Intra-luminal
Other (Gynecological)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esoph. (Cardiac)
Other (Specify)
PeripheralVesselPeripheral vesselNNNNNN
Other (Carotid)NNNNNN
Musculo-skel (conventional)NNNNNN
Musculo-skel (superficial)NNNNNN

N= new indication; P= previously cleared by FDA; E= added under Appendix E

*Other modes: SonoCT, X-Res, Angio, Panoramic, 3-D Imaging, Directional Angio Imaging, IMT measures, Color Power Angio

Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color, Dual

Previous submission: Cleared in K030455

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

David Lehman

(Division Sign-Off) Division of Reproductive, Abdomin and Radiological Devices 510(k) Number _

{23}------------------------------------------------

510(k) Number: K

System: Philips Ultrasound HD11 Diagnostic Ultrasound System

L12-3 (21475A) Linear Transducer Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
SpecificBMPWDCWDColor DopplerCombined (Specify)Other* (Specify)
General(Track I Only)(Tracks I & III)
OphthalmicOphthalmic
Fetal
AbdominalNNNNNN
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
Fetal Imaging& OtherPediatricNNNNNN
Small Organ (Scrotum,Thyroid, Breast)NNNNNN
Neonatal Cephalic
Adult Cephalic.
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Intra-luminal
Other (Gynecological)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esoph. (Cardiac)
Other (Specify)
PeripheralVesselPeripheral vesselNNNNNN
Other (Carotid)NNNNNN
Musculo-skel (conventional)NNNNNN
Musculo-skel (superficial)NNNNNN
N= new indication; P= previously cleared by FDA; E= added under Appendix E
*Other modes: SonoCT, X-Res, Angio, Panoramic, Harmonics (Tissue & Contrast), 3-D Imaging, Directional Angio
Imaging, IMT Measures, Color Power Angio
Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color, Dual

Previous submission: K014191

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

Prescription Use (Per 21 CFR 801.109)

David R. Lehman

1

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number _

{24}------------------------------------------------

510(k) Number: K

System: Philips Ultrasound HD11 Diagnostic Ultrasound System

Transducer: L8-4 (8500-1659-01) Liner Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
Fetal
AbdominalNNNNNN
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
Fetal Imaging& OtherPediatricNNNNNN
Small Organ (Scrotum,Thyroid, Breast)NNNNNN
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Intra-luminal
Other (Gynecological)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esoph. (Cardiac)
Other (Specify)
PeripheralVesselPeripheral vesselNNNNNN
Other (Carotid)NNNNNN
Musculo-skel (conventional)NNNNNN
Musculo-skel (superficial)NNNNNN

N= new indication; P= previously cleared by FDA; E= added under Appendix E

*Other modes: SonoCT, X-Res, Angio, Panoramic, Harmonics (Tissue & Contrast), 3-D Imaging, Directional Angio Imaging, IMT, Color Power Angio

Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color, Dual

Previous submission: Cleared on K030455

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

David R. Larson

(Division Sign-Off) Division of Reproductive. and Radiological Device 510(k) Number _

{25}------------------------------------------------

510(k) Number: K

System: Philips Ultrasound HDI 1 Diagnostic Ultrasound System

L15-7io (989605341571) Linear Intra-operative Transducer Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal
Abdominal
Intra-operative (Cardiac,Vascular)NNNNNN
Intra-operative (Neuro)NNNNNN
Laparoscopic
PediatricNNNNNN
Small Organ (Scrotum,Thyroid, Breast)NNNNNN
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Intra-luminal
Other (Gynecological)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esoph. (Cardiac)
Other (Specify)
PeripheralVesselPeripheral vesselNNNNNN
Other (I/O)NNNNNN
Musculo-skel (conventional)NNNNNN
Musculo-skel (superficial)NNNNNN
N= new indication; P= previously cleared by FDA; E= added under Appendix E
*Other modes: SonoCT, X-Res, Angio, Panoramic, 3-D Imaging, Directional Angio Imaging,
Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color, Dual
Previous submission: Cleared K030455

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

David h. Leppan

(Division Sign-Off) Division of Reproductive, Abdomin and Radiological Devices 510(k) Number .

{26}------------------------------------------------

510(k) Number: K

System: Philips Ultrasound HDI Diagnostic Ultrasound System

3D8-4 (989605345331) Curved Linear Transducer Transducer:

I l'ansuucer. - - - JDG - (7070055 15603) in the how analysis of the human body as follows:

Clinical ApplicationMode of Operation
GeneralSpecificBMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
(Track I Only)(Tracks I & III)
OphthalmicOphthalmic
FetalNNNNNN
AbdominalNNNNNN
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
Fetal Imaging& OtherPediatricNNNNNN
Small Organ (Scrotum,Thyroid, Breast)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Intra-luminal
Other (Gynecological)NNNNNN
Cardiac Adult
CardiacCardiac Pediatric
Trans-esoph. (Cardiac)
Other (Fetal Heart )NNNNNN
PeripheralVesselPeripheral vessel
Other (Carotid)
Musculo-skel (conventional)
Musculo-skel (superficial)
N= new indication; P= previously cleared by FDA; E= added under Appendix E

*Other modes: SonoCT, X-Res, Angio, Panoramic, Harmonics (Tissue), 3-D/4-D Imaging, Directional Angio Imaging, Fetal Heart (Spatial Temporal Image Correlation), Color Power Angio

real Tical (opatial Temporal Intego Colex Triplex = 2D + Doppler + Color, Real Time Biopsy, Dual

Previous submission: None

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) (I EEAOL DO NOT 171215 Devices and Radiological Health, Office of Device Evaluation

David R. Hanson

(Division Sign-Off) Division of Reproductive, Ab and Radiological Devices 510(k) Number _

{27}------------------------------------------------

510(k) Number: K

System: Philips Ultrasound HD11 Diagnostic Ultrasound System

Transducer: 3D6-2 (8500-2032-01) Curved Linear Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Cliniani A

Clinical ApplicationMode of Operation
General(Track I Only)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Specify)Other*(Specify)
OphthalmicOphthalmic
FetalNNNNNN
AbdominalNNNNNN
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
Fetal Imaging& OtherPediatricNNNNNN
Small Organ (Scrotum,Thyroid, Breast)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Intra-luminal
Other (Gynecological)NNNNNN
Cardiac Adult:
Cardiac.Cardiac Pediatric
Trans-esoph. (Cardiac)
Other (Fetal Heart)NNNNNN
PeripheralVesselPeripheral vessel
Other (Carotid)
Musculo-skel (conventional)
Musculo-skel (superficial)
N= new indication; P= previously cleared by FDA; E= added under Appendix E
*Other modes: SonoCT, X-Res, Angio, Panoramic, Harmonics (Tissue), 3-D/4-D Imaging, Directional Angio Imaging,Fetal Heart (Spatial Temporal Image Correlation), Color Power Angio
Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color, Real Time Biopsy, Dual
Previous submission: Cleared on K034003

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

Prescription Use (Per 21 CFR 801.109)

.

David R. Johnson

(Division Sign-Off) Division of Reproductive, Abdomin and Radiological Devices 510(k) Number

{28}------------------------------------------------

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
GeneralSpecificBMPWDCWDColorCombinedOther *
(Track I Only)(Tracks I & III)Doppler(Specify)(Specify)
OphthalmicOphthalmic
FetalNNNNNN
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
Fetal ImagingPediatric
& OtherSmall Organ (Scrotum,
Thyroid, Breast)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginalNNNNNN
Trans-urethral
Trans-esoph. (non-Card.)
Intra-luminal
Other (Gynecological)NNNNNN
CardiacCardiac Adult
Cardiac Pediatric
Trans-esoph. (Cardiac)
Other (Fetal Heart )NNNNNN
PeripheralPeripheral vessel
VesselOther (Carotid)
Musculo-skel (conventional)
Musculo-skel (superficial)
N= new indication; P= previously cleared by FDA; E= added under Appendix E
*Other modes: SonoCT, X-Res, Angio, Panoramic, Harmonics (Tissue), 3-D/4-D Imaging, Directional Angio Imaging,
Fetal Heart (Spatial Temporal Image Correlation), Color Power Angio

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation

David A. Seymann

(Division Sign-Off) Division of Reproductive, Abdon and Radiological Devices 51(){k) Number _

N/A