K Number
K093104

Validate with FDA (Live)

Manufacturer
Date Cleared
2010-06-25

(267 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Medyssey Co. Ltd., Zenius™ Spinal System, a posterior spinal fixation device, indicated for skeletally mature patients receiving fusion by autogenous bone graft with removal of the implants after the attainment of a solid fusion and is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis).

Device Description

The Medyssey Co. Ltd., Zenius ™ Spinal System is a top-loading posterior spinal fixation system which consists of pedicle screws, rods, set screws, and a transverse (cross) linking mechanism. The Medyssey Co. Ltd., Zenius ™ Spinal System implant components are fabricated from titanium alloy (Ti-6AI-4V ELI) that conforms to ASTM F 136. Various sizes of these implants are available. Specialized instruments made from surgical instrument grade stainless steel are available for the application and removal of the Medyssey Co. Ltd., Zenius ™ Spinal System implants.

AI/ML Overview

This document describes a 510(k) premarket notification for a spinal fixation system, not an AI/ML powered medical device. Therefore much of the requested information, such as sample sizes for test and training sets, number of experts for ground truth, adjudication methods, and MRMC studies, is not applicable to this submission type.

Here’s an analysis based on the provided text, focusing on the available information:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (Implied)Reported Device Performance
Substantial Equivalence to Predicate DevicesThe Medyssey Co. Ltd., Zenius™ Spinal System was found to be "substantially equivalent" to the U&I Corporation OPTIMA™ Spinal System and Medyssey's Co. Ltd., NOVEL™ Spinal System.
Mechanical Performance (Static & Fatigue per ASTM F 1717-04)Mechanical testing (Static Compression Bending, Static Tension Bending, and Static Torsional Testing) was performed per ASTM F 1717-04. The subject device "performed substantially equal or better in the above mentioned testing to that of the predicated devices."
Material CompatibilityImplants are fabricated from titanium alloy (Ti-6Al-4V ELI) that conforms to ASTM F 136.
Indications for UseThe device meets the stated Indications for Use, which are identical to predicate devices: "for skeletally mature patients receiving fusion by autogenous bone graft... as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis)."
Equivalence in Principles of Operation, Labeling, and RegulationPrinciples of operation, indications, contraindications, warnings, and precautions within labeling are similar to predicate devices. The device is classified as Class II (88.3390) under Spinal Pedicle Screw (MNI) and Spondylolisthesis Spinal Fixation Device System (MNH) 21 CFR § 888.3070.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not applicable. The device is a physical medical implant (spinal fixation system), and its performance was evaluated through mechanical testing, not clinical data or a test set in the context of AI/ML. The provenance of the data is the mechanical testing conducted according to ASTM standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable. Ground truth, in the AI/ML sense, is not relevant for this type of device submission. Performance was established through standardized mechanical tests.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable, as there was no test set requiring human adjudication for ground truth.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC study was done, nor would it be relevant for this type of physical implant device. The submission explicitly states: "Clinical tests: No clinical tests conducted on either the subject system nor the predicate systems."

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable. This is a mechanical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For this device, the "ground truth" for proving performance was based on:

  • Adherence to Material Standards: Conformance to ASTM F 136 for titanium alloy.
  • Mechanical Testing Standards: Performance against established benchmarks and predicate devices in static and fatigue testing protocols as defined by ASTM F 1717-04.
  • Predicated Device Equivalence: The performance of existing, legally marketed devices (U&I Corporation OPTIMA™ Spinal System and Medyssey's Co. Ltd., NOVEL™ Spinal System) served as the benchmark for "substantial equivalence."

8. The sample size for the training set

This information is not applicable, as there was no training set for an AI/ML algorithm.

9. How the ground truth for the training set was established

This information is not applicable, as there was no training set for an AI/ML algorithm.

{0}------------------------------------------------

510(k) SUMMARY

JUN 2 5 2010

510 (K) Premarket Notification: Medyssey Co. Ltd, Zenius™ Spinal System

1. Submitter/Sponsor:

Medyssey Co. Ltd. Patrick D. Moore, Official US Correspondent 722-3, 4F. Science Tower, Jihaeng-dong, Dongducheon-city, Gyeonggi-do, Korea

Contact person:

Patrick D. Moore Official US Correspondent Medyssey Co. Ltd. 6170 South 380 West, Suite 200. Murray, Utah. 84107 Tel. 801-266-4811, Fax. 801-266-4363; E-Mail: pdmoore@jemospine.com

Date Prepared:

June 25th, 2010

2. Device Name:

Classification Name:Pedicle Screw Spinal Fixation System
Common/Generic Name:Pedicle Screw Spinal System
Trade Name:Medyssey Co. Ltd, Zenius TM Spinal System

3. Device Classification(s):

Class II (88.3390) following Orthopedic and Rehabilitation Device Advisory Review, for the requested indications:

  • Spinal Pedicle Screw (MNI) 21 CFR § 888.3070 .
  • Spondylolisthesis Spinal Fixation Device System (MNH) 21 CFR § . 888.3070

4. Predicate Device:

U&I Corporation, OPTIMA™ Spinal System – MNI, MNH, KWQ -- (K020279 and K024098) Medyssev Co. Ltd., NOVEL™ Spinal System - MNH, MNI - (K081153)

5. Device Description:

The Medyssey Co. Ltd., Zenius ™ Spinal System is a top-loading posterior spinal fixation system which consists of pedicle screws, rods, set screws, and a transverse (cross) linking mechanism. The Medyssey Co. Ltd., Zenius ™ Spinal System implant components are fabricated from titanium alloy (Ti-6AI-4V ELI) that conforms to ASTM F 136. Various sizes of these implants are available.

Specialized instruments made from surgical instrument grade stainless steel are available for the application and removal of the Medyssey Co. Ltd., Zenius ™ Spinal System implants.

{1}------------------------------------------------

6. Indications for Use:

The Medyssey Co. Ltd., Zenius™ Spinal System, a posterior spinal fixation device, indicated for skeletally mature patients receiving fusion by autogenous bone graft with removal of the implants after the attainment of a solid fusion and is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis).

  1. Comparison with predicate device: The Medyssey Co. Ltd., Zenius ™ Spinal System is substantially equivalent to the currently marketed U&I Opinal Oystem is oubotantially of Sistem and Medyssey's Co. Ltd., Novel™ Spinal System. When considered for posterior applications, the Medyssey Co. Ltd., Zenius TM Spinal System, the Medyssey's Co. Ltd., Novel™ Spinal System Eld., Londo - Opinal Oys., OPTIMA™ Spinal System worst case constructs consist of a universal housing containing a pre-assembled pedicle screw and set screw. All three systems use vertical rods which are both placed into the housing. The set screws are subsequently tightened onto the rod, providing a completed implant assembly.

The principles of operation for the subject Medyssey Co. Ltd., Zenius™ Spinal System device, and the cited predicate technologies are similar. That is, each of these products employs similar indications for use, contraindications for use, warnings and precautions within labeling. The principles of operation of the subject device are directly equivalent to those of the cited predicates cleared by the Agency and currently being marketed.

Both the subject device and the predicate device were evaluated/tested per established requirements. The mechanical testing for both the predicate and the subject included static and fatigue testing performed per ASTM F 1717-04. Testing included: Static Compression Bending; Static Tension Bending; and Static Torsional Testing.

The subject device performed substantially equal or better in the above mentioned testing to that of the predicated devices.

Clinical tests: No clinical tests conducted on either the subject system nor the predicate systems.

Conclusion: The subject device was evaluated against the predicate devices for all performance, safety & effectiveness requirements and found as substantially equivalent to the predicate devices.

Page 2 of 2

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around it. Inside the circle is a stylized image of an eagle or bird-like figure, represented by three curved lines that suggest wings and a head.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

JUN 2 5 2010

Medyssey Co., Ltd. % Mr. Patrick.D. Moore 6170 South 380 West, Suite 200 Murray, Utah 84107

Re: K093104

Trade/Device Name: Medyssey Co., Ltd. Zenius™ Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNI, MNH Dated: June 21, 2010 Received: June 21, 2010

Dear Mr. Moore:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that vour device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{3}------------------------------------------------

Page 2 - Mr. Patrick D. Moore

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please n you about open . fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Barbara Bruchid

Mark N. Melker Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known): K093104

Device Name: Medyssey Co. Ltd., Zenius™ Spinal System

Indications for Use: The Medyssey Co. Ltd., Zenius™ Spinal System, a posterior spinal fixation device, indicated for skeletally mature patients receiving fusion by autogenous bone graft with removal of the implants after the attainment of a solid fusion and is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis).

(Per 21 CFR801.109) Prescription Use______________________________________________________________________________________________________________________________________________________________ × OR Over-the-Counter Use (Optional Format 1-2-96)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Joutish

(Division Sign-Of) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K093104

Page 1 of 1