(324 days)
The Certa Dose™ PD Epinephrine 1 mg/mL IM/SC Syringe is intended for healthcare professionals to deliver epinephrine 1 mg/mL via intramuscular or subcutaneous injection for pediatric patients.
The Certa Dose™ PD Epinephrine 1 mg/mL IM/SC Syringe is comprised of (A) a 3/10mL Syringe with an attached 29 Gauge x ½ inch Needle, and (B) a plastic overlay which provides barrel markings. The items on the overlay include (1) a zero line, (2) 0.01 mL unit graduation markings, (3) secondary graduation markings at 0, 0.05, 0.10, 0.15, 0.20, 0.25, 0.30 mL, (4) color bars, and (5) the name of the drug for which the syringe is indicated, epinephrine 1 mg/mL.
The Certa Dose™ PD Epinephrine 1 mg/mL IM/SC Syringe is a pre-filled syringe designed for healthcare professionals to deliver epinephrine to pediatric patients.
Acceptance Criteria and Reported Device Performance
| Acceptance Criteria / Performance Claim | Reported Device Performance (from the document) | Study Supporting Performance |
|---|---|---|
| Drug-Device Compatibility | Short contact times do not adversely impact drug quality. | Drug-device compatibility studies |
| Syringe Accuracy | Markings are applied correctly and remain accurate. | Syringe accuracy testing after sterilization and accelerated aging |
| Human Factors / Usability | The device's use is validated. | Human factors study in accordance with "Guidance for Industry and Food and Drug Administration Staff: Applying Human Factors and Usability Engineering to Medical Devices." dated Feb. 3, 2016. |
| Biocompatibility | All materials of the syringe are acceptable per ISO10993-1: 2009/(R) 2013 for Cytotoxicity, Irritation, Sensitization, Systemic Toxicity, and Hemacompatibility. | Biocompatibility testing per ISO10993-1: 2009/(R) 2013 |
Study Information (Based on the Provided Text)
The provided text focuses on demonstrating substantial equivalence to predicate devices and describes the rationale for design choices and conducted tests rather than a standalone clinical study with detailed statistical results. Therefore, many of the requested details about a "study" in the traditional sense of a diagnostic device's performance are not explicitly available. However, based on the information provided, we can infer the following:
- 1. Sample size used for the test set and the data provenance: This information is not provided in the document for any of the described tests. The studies mentioned (drug-device compatibility, syringe accuracy, biocompatibility, human factors) are engineering and laboratory tests, not clinical studies with patient data.
- 2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: This information is not provided as these were not studies involving expert consensus on patient data.
- 3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: This information is not provided as these were not studies involving expert consensus on patient data.
- 4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: An MRMC comparative effectiveness study was not performed. The device is a syringe and does not involve AI or human readers for interpretation. The human factors study would assess usability, not diagnostic accuracy improvement.
- 5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: The concept of "standalone performance" typically applies to AI algorithms. This device is a physical syringe, so this question is not applicable. The performance tests focus on the physical and material integrity and accuracy of the syringe itself.
- 6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Drug-Device Compatibility: The ground truth would be the expected chemical stability and quality of epinephrine.
- Syringe Accuracy: The ground truth would be established by metrological standards for volume measurement.
- Biocompatibility: The ground truth is defined by the ISO10993-1 standard.
- Human Factors Study: The "ground truth" would be the identified usability issues and user errors during tasks performed by healthcare professionals, evaluated against predefined success criteria.
- 7. The sample size for the training set: This information is not applicable as this is not an AI/machine learning device.
- 8. How the ground truth for the training set was established: This information is not applicable as this is not an AI/machine learning device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 19, 2017
Certa Dose, Inc. % James Boiani Epstein Becker & Green, PC 1227 25th St NW, Suite 700 Washington, District of Columbia 20037
Re: K160589
Trade/Device Name: Certa Dose™ PD Epinephrine 1mg/mL IM/SC Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: POX Dated: December 19, 2016 Received: December 21, 2016
Dear James Boiani:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/8 description: The image contains a signature and a name. The signature is a complex, looping design in black ink. To the right of the signature, the name "Tina Kiang-S" is written in a clear, sans-serif font. The name is stacked vertically, with "Tina" on top and "Kiang-S" below.
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K160589
Device Name
Certa Dose™ PD Epinephrine 1 mg/mL IM/SC Syringe
Indications for Use (Describe)
The Certa Dose™ PD Epinephrine 1 mg/mL IM/SC Syringe is intended for healthcare professionals to deliver epinephrine 1 mg/mL via intramuscular or subcutaneous injection for pediatric patients.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary K160589
I. SPONSOR AND CORRESPONDENT
Sponsor: Certa Dose. Inc. 573 Race Street Denver, CO 80206
Submission Correspondent: James A. Boiani, MS, JD Epstein Becker & Green, P.C. 1227 25th St., NW Suite 700 Washington, DC 20037
Date Prepared: January 19, 2016
II. DEVICE
Name of Device: Certa Dose™ PD Epinephrine 1 mg/mL IM/SC Syringe Classification Name: Piston Syringe, 21 CFR 880.5860 Regulatory Class: II Product Code: PQX, Epinephrine Syringe
III. PREDICATE DEVICES
Primary: Monoject 3/10mL U-100 Insulin Syringe w/ Attached 29 Gauge x 1/2" Needle, K991758 Secondary: Terumo® 31G ThinPro Insulin Syringe, K071630 Reference Device: UniTox® Syringe, K123710 Reference Device: BD Flu+ Syringe, K091377
DEVICE DESCRIPTION IV.
The Certa Dose™ PD Epinephrine 1 mg/mL IM/SC Syringe is comprised of (A) a 3/10mL Syringe with an attached 29 Gauge x ½ inch Needle, and (B) a plastic overlay which provides barrel markings. The items on the overlay include (1) a zero line, (2) 0.01 mL unit graduation markings, (3) secondary graduation markings at 0, 0.05, 0.10, 0.15, 0.20, 0.25, 0.30 mL, (4) color bars, and (5) the name of the drug for which the syringe is indicated, epinephrine 1 mg/mL.
V. INDICATIONS FOR USE
{4}------------------------------------------------
The Certa Dose™ PD Epinephrine 1 mg/mL IM/SC Syringe is intended for healthcare professionals to deliver epinephrine 1 mg/mL via intramuscular or subcutaneous injection for pediatric patients.
SUBSTANTIAL EQUIVALENCE TO PREDICATE DEVICES VI.
Certa Dose™ PD Epinephrine 1 mg/mL IM/SC Syringe is substantially equivalent to the predicates as it has
(1) the same intended use, and
(2) the same technological characteristics as the predicates, or where it has different technological characteristics, they do not adversely affect the device safety or effectiveness relative to the predicates, and do not raise different questions of safety or effectiveness.
1. The Certa Dose™ PD Epinephrine 1 mg/mL IM/SC Syringe Has the Same Intended Use as the Predicates and Other Piston Syringes
The intended use of the device is identical to the predicates: the delivery of fluid. The indication differs from that of the predicates only with respect to the specific medication for which it is indicated. However, FDA has cleared multiple drug-specific syringes within the piston syringe device classification, as discussed above. Further, the decision-making factors that FDA provides in its guidance for evaluating whether a specific indication falls within a general intended use support the conclusion that the intended use of the Certa Dose™ PD Epinephrine 1 mg/mL IM/SC Syringe is the same as those of the predicates.
2. The Differences in Technology from the Primary Predicate Do Not Adversely Affect Safety or Effectiveness, or Raise New Questions of Safety or Effectiveness
The Certa Dose™ PD Epinephrine 1 mg/mL IM/SC Syringe is identical to the materials, design, and packaging of the primary predicate, except for the differences summarized below:
| ATTRIBUTE | Certa Dose™ PD Epinephrine1mg/mL IM/SC Syringe | Primary Predicate |
|---|---|---|
| 510(k) Number | K160586 | K991758 |
| a. Drug and DeviceCompatibility | Epinephrine 1 mg/mL | U-100 Insulin |
| b. Unit Graduations | 0.01 mL | 0.5 Units of Insulin (“UI”) |
| c. Secondary Graduations | 0, 0.05, 0.10, 0.15, 0.20, 0.25,0.30 mL | 5, 10, 15, 20, 25, 30 UI. |
| d. Color Zones | Color zones that aid as aconfirmatory check on volumeto be administered | None |
| e. Application of BarrelMarkings | Applied Using Plastic Overlay | Applied Directly to Barrel |
4 FDA Guidance for Industry: General/Specific Intended Use (1998).
{5}------------------------------------------------
In addition, the Certa Dose™ PD Epinephrine 1 mg/mL IM/SC Syringe is identical to the Secondary Predicate with respect to the use for pediatric populations, with the following exception:
| ATTRIBUTE | Certa Dose™ PDEpinephrine 1 mg/mLIM/SC Syringe | Secondary Predicate |
|---|---|---|
| 510(k) Number | K160589 | K071630 |
| f. Design for IndicatedPediatric Population | Syringe for epinephrine 1mg/mL with 1/2" 29 GaugeUltrafine Needle | Syringe for U-100 insulin with3/8" 31 Gauge UltrafineNeedle |
Each of these attributes (a through f) is addressed below.
- The design of the Certa Dose™ PD Epinephrine 1 mg/mL IM/SC Syringe is appropriate a. for delivering epinephrine. The syringe components that come into contract with the drug are standard nonreactive materials used in syringe construction – clarified radiation grade polypropylene (barrel material), synthetic isoprene (plunger tip material), silicone (lubricant), and stainless steel (needle) - and raise no concerns with respect to delivery of liquid medication, such as insulin (with the predicate devices) or epinephrine (with the submitted device), in light of the nature of materials (nonreactive) and indication for use in drug delivery (resulting in virtually no contact time epinephrine). Drug contact studies were also conducted to confirm that short contact times do not adversely impact drug quality.
- Graduation unit markings on drug-specific syringes are chosen based on what resolution b. of markings would provide the level of specificity commonly required for the drug in question. The 0.01 mL unit graduations markings were chosen to provide the level of specificity that may be commonly required for measuring epinephrine 1 mg/mL volumes. Similar adjustments have been made in the past for other drugs (see predicates and reference devices above).
- c. Secondary graduations markings of 0, 0.05, 0.10, 0.15, 0.20, 0.25, 0.30 mL are evenly spaced graduations that are consistent with those observed with the even spacing of the predicate svringes.
- The color zones (bars of color) on the syringe provide an aid in confirming the drawn d. volume of fluid. Also, the colors being used correspond to the universal color system pioneered by the Broselow-Luten system which has been widely adopted in the professional use pediatric patient setting for which the syringe is intended. A description of the correlation between colors, weights, and volumes is provided below.
{6}------------------------------------------------
Use the Certa Dose Syringe color zones that alian with standard color coding for pediatric weight ranges visual aid to help confirm if calculated (0.01 mg/kg) and measured volumes (mL) of Epinephrine 1 mg/mL are with the standard dosing recommendations for anaphylaxis in pediatric patients under 30 kg (66 lbs),
| Standard PediatricColor Coding System for Weight | Certa Dose™ PD Epinephrine Syringe:Volume and mg Epinephrine | |||
|---|---|---|---|---|
| Color | Weight Range | Color | Volume Range (mL) | Amt. of Epinephrine (mg) |
| Grey | 3-5 kg | Grey | 0.00-0.05 | 0.00-0.05 |
| Pink | 6-7 kg | Pink | 0.05-0.07 | 0.05-0.07 |
| Red | 8-9 kg | Red | 0.07-0.09 | 0.07-0.09 |
| Purple | 10-11 kg | Purple | 0.09-0.11 | 0.09-0.11 |
| Yellow | 12-14 kg | Yellow | 0.11-0.14 | 0.11-0.14 |
| White | 15-18 kg | White | 0.14-0.18 | 0.14-0.18 |
| Blue | 19-23 kg | Blue | 0.18-0.23 | 0.18-0.23 |
| Orange | 24-29 kg | Orange | 0.23-0.29 | 0.23-0.29 |
| Green | 30-36 kg | Green | 0.29-0.30 | 0.29-0.30 |
Certa Dose™ PD Epinephrine Syringe Graduations a
Image /page/6/Figure/3 description: The image shows a color-coded dosage scale for use with Epinephrine 1mg/mL only. The scale ranges from 0.05 to 0.30 mL (milliliter), with each dosage level represented by a different color. The text "SINGLE USE ONLY Certa Dose™ PD Syringe" is printed above the scale, and the text "CERTADOSE™ V1" is printed to the right of the scale.
- The barrel markings are clear, stable (e.g., fade-resistant), and accurately reflect volumes. e. Specifications regarding application of the markings will ensure that the markings are correctly applied and provide for accuracy. Syringe accuracy testing after sterilization and accelerated aging are used to verify that the markings are applied correctly and remain accurate.
- f. The Certa Dose device is mechanically identical to the Monoject Syringe, which has been used by the general population, including children, for several years. The syringe is also similar to the secondary predicate, the Terumo® 31G ThinPro Insulin Syringe with respect to product design, only differing slightly with respect to markings and the use of a slightly larger needle gauge - 29 gauge Ultrafine needle for the Certa Dose™ PD Epinephrine 1 mg/mL IM/SC Syringe (and the Primary Predicate) versus 31 gauge needle for the Secondary Predicate - and slightly longer length - 1/2" versus 3/8". Both gauges are considered "ultrafine."
The questions with regard to evaluating the safety and effectiveness of the device design for the indicated patient population require answering the same questions regarding syringe design. In this case, the syringe is specifically designed to deliver epinephrine 1 mg/mL and employs graduation markings and color zones (which are used for pediatric patients) that are appropriate for pediatric patient application.
{7}------------------------------------------------
Based on the foregoing, the differences with respect to design do not raise new questions of safety or effectiveness, and the information above supports that the Certa Dose™ PD Epinephrine 1 mg/mL IM/SC Syringe is substantially equivalent to its predicates.
VII. PERFORMANCE TESTING
The following performance testing was conducted to support the substantial equivalence of the subject device:
- Drug-device compatibility studies were conducted to confirm that short contact times do ● not adversely impact drug quality.
- Syringe accuracy testing to verify that the markings are applied correctly and remain accurate.
VIII. HUMAN FACTORS STUDY
A human factors study in accordance with "Guidance for Industry and Food and Drug Administration Staff: Applying Human Factors and Usability Engineering to Medical Devices." dated Feb. 3, 2016 to further validate the Certa Dose™ PD Epinephrine 1 mg/mL IM/SC Syringe's use.
BIOCOMPATIBILITY IX.
Biocompatibility testing for all the materials of the syringe has been performed and found acceptable per ISO10993-1: 2009/(R) 2013. The testing includes the acceptable tests for Cytotoxicity, Irritation, Sensitization, Systemic Toxicity and Hemacompatibility as per the individual associated parts of the ISO10993 series.
X. CONCLUSIONS
The information provided supports the substantial equivalence of Certa Dose™ PD Epinephrine 1 mg/mL IM/SC Syringe to the predicates.
§ 880.5860 Piston syringe.
(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).