K Number
K160013

Validate with FDA (Live)

Device Name
Crosperio OTW
Manufacturer
Date Cleared
2016-04-29

(116 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Crosperio OTW PTA Balloon Dilatation Catheter is intended to dilate stenosis in the iliac, femoral, popliteal, infrapoplited and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for stent post-dilatation in the peripheral vasculature.

Device Description

Crosperio OTW PTA Balloon Dilatation Catheter (Crosperio OTW) is an over-the-wire type of balloon dilatation catheter, which consists of an outer tube, inner tube, balloon, radiopaque markers, manifold, outer reinforcement tube and strain relief. A balloon is attached to the distal end of the catheter, and it can be inflated and deflated using the inflation device connected to the manifold at the proximal end. Crosperio OTW has a guidewire lumen at the distal end of the catheter through which a guidewire can be inserted, and also an opening along the inner tube to the guidewire port of the manifold for the introduction of a guidewire.

The maximum compatible diameter of a guidewire used together with Crosperio OTW in a PTA procedure is 0.014 inches. Additionally, the compatible size of a sheath introducer is 4 Fr. The nominal inflated balloon diameters range from 1.5 mm to 4.0 mm with balloon working lengths of 20 mm to 200 mm. Three different catheter working lengths will be provided: 650, 900 and 1500 mm.

AI/ML Overview

The provided text is a 510(k) summary for the Crosperio OTW PTA Balloon Dilatation Catheter, a medical device. It describes the device, its intended use, comparison to predicate devices, and performance testing conducted to demonstrate substantial equivalence.

Here's an analysis of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

The document states: "The Crosperio OTW met all the predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs." However, it does not explicitly list the quantitative acceptance criteria for each test or the specific reported device performance values in a table format. It lists the types of tests performed.

Types of In Vitro Tests Performed:

  • Dimensional Verification
  • Balloon Preparation, Deployment and Retraction
  • Balloon Rated Burst Pressure
  • Balloon Fatigue (Repeat Balloon Inflations)
  • Balloon Compliance
  • Balloon Inflation and Deflation Time
  • Catheter Bond Strength
  • Flexibility and Kink Test
  • Torque Strength
  • Radiopacity
  • Coating Integrity
  • Particulate Evaluation
  • Balloon Rated Burst Pressure (in Stent)
  • Balloon Fatigue (Repeat Balloon Inflations; in Stent)
  • Shelf Life Testing

Types of Biocompatibility Tests Performed:

  • Cytotoxicity
  • Sensitization
  • Intracutaneous reactivity (irritation)
  • Systemic toxicity (acute)
  • Pyrogenicity
  • Hemocompatibility (thrombogenicity, hemolysis, and immunology)
  • Genotoxicity (bacterial gene mutation assay, in vitro mammalian genotoxicity assay, and in vivo cytogenetics assay)

Reported Device Performance (General Statement):

"The results from these tests demonstrate that the technological characteristics and performance criteria of the Crosperio OTW are adequate for the intended use of the device and that the device can perform in a manner equivalent to devices currently on the market with the same intended use."

2. Sample Size Used for the Test Set and Data Provenance:

The document does not specify the sample size for the test sets used in the in vitro or biocompatibility tests. It also does not mention data provenance (e.g., country of origin, retrospective or prospective) as these were bench and lab tests on the device itself, not clinical data from patients.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts:

This information is not applicable and therefore not provided in the document. The tests performed are engineering and biological evaluations of the device's physical properties and material safety, not diagnostic imaging or clinical performance requiring expert interpretation of results for ground truth.

4. Adjudication Method for the Test Set:

This information is not applicable and therefore not provided. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or studies involving expert assessment of data, not for bench testing of medical devices.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs without AI assistance:

This information is not applicable and therefore not provided. The Crosperio OTW PTA Balloon Dilatation Catheter is a physical medical device, not an AI software/algorithm. Therefore, an MRMC study related to AI assistance for human readers is irrelevant to this submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

This information is not applicable and therefore not provided. As stated above, this is a physical medical device, not an algorithm, so standalone AI performance is not relevant.

7. The type of ground truth used:

The "ground truth" for the performance and biocompatibility tests was established by engineering specifications, recognized standards, guidance documents (like the FDA's "Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters"), and established biological safety testing protocols (like ISO 10993). It is based on objective, measurable criteria inherent to the device's design and manufacturing, rather than expert consensus, pathology, or outcomes data from patients.

8. The sample size for the training set:

This information is not applicable and therefore not provided. Training sets are relevant for machine learning algorithms, not for physical medical devices being evaluated through bench testing.

9. How the ground truth for the training set was established:

This information is not applicable and therefore not provided. As there is no training set for a machine learning algorithm, there is no ground truth establishment method for it.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The text is written in a sans-serif font.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 29, 2016

Kaneka Corporation % Mr. Christopher M. Sloan Principal Consultant Quintiles Consulting 1801 Rockville Pike, Suite 300 Rockville, MD 20852

Re: K160013

Trade/Device Name: Crosperio OTW PTA Balloon Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT Dated: March 29, 2016 Received: March 31, 2016

Dear Mr. Sloan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.

510(k) Number (if known) K160013

Device Name

Crosperio OTW PTA Dilatation Catheter

Indications for Use (Describe)

The Crosperio OTW PTA Balloon Dilatation Catheter is intended to dilate stenosis in the iliac, femoral, popliteal, infrapoplited and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for stent post-dilatation in the peripheral vasculature.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(K) SUMMARY

Crosperio OTW PTA Balloon Dilatation Catheter

510(k) Submitter

Kaneka Corporation Medical Device Division 1-12-32, Akasaka, Minato-ku Tokyo, Japan Contact Person: Toshihiko Motomine Telephone: +81-3-5574-8023 Email:

Official Correspondent

Christopher M. Sloan Principal Consultant Quintiles Consulting 1801 Rockville Pike, Suite 300 Rockville, MD 20852 Phone: (301) 272-3114

Email: chris.sloan@quintiles.com

Date Prepared: March 22, 2016

{5}------------------------------------------------

Subject Device Name:

Trade NameCrosperio OTW
Common or usual namePercutaneous (peripheral) transluminalangioplasty catheter
Classification namePercutaneous catheter [21 CFR 870.1250;product code LIT]
ClassII
Classification PanelCardiovascular (74)

Predicate Devices:

  • Ultraverse® RX PTA Dilatation Catheter or "Ultraverse RX" [K131199 (Bard ● Peripheral Vascular Inc.)]
  • Metacross OTW PTA Balloon Dilatation Catheter or "Metacross OTW" [K152080 ● (Kaneka Corporation)]

Device Description:

Crosperio OTW PTA Balloon Dilatation Catheter (Crosperio OTW) is an over-the-wire type of balloon dilatation catheter, which consists of an outer tube, inner tube, balloon, radiopaque markers, manifold, outer reinforcement tube and strain relief. A balloon is attached to the distal end of the catheter, and it can be inflated and deflated using the inflation device connected to the manifold at the proximal end. Crosperio OTW has a guidewire lumen at the distal end of the catheter through which a guidewire can be inserted, and also an opening along the inner tube to the guidewire port of the manifold for the introduction of a guidewire.

The maximum compatible diameter of a guidewire used together with Crosperio OTW in a PTA procedure is 0.014 inches. Additionally, the compatible size of a sheath introducer is 4 The nominal inflated balloon diameters range from 1.5 mm to 4.0 mm with balloon Fr. working lengths of 20 mm to 200 mm. Three different catheter working lengths will be provided: 650, 900 and 1500 mm.

Indications for Use

The Crosperio OTW is intended to dilate stenosis in the iliac, femoral, ilio-femoral, popliteal, infra-popliteal and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for stent post-dilatation in the peripheral vasculature.

{6}------------------------------------------------

Comparison of Indications for Use to Predicate Devices

The Crosperio OTW has the same intended use (percutaneous transluminal angioplasty) and comparable indications for use as the Ultraverse RX. Also, the Crosperio OTW has the identical intended use and indications for use to the Metacross OTW. Therefore, the subject device, Crosperio OTW, may be considered substantially equivalent to the predicate devices.

Comparison of Technological Characteristics to Predicate Devices

Technological Comparison to Predicate Devices

The Crosperio OTW has the following similarities to the predicate devices:

  • Same intended use (same as Ultraverse RX and Metacross OTW) .
  • Same indications for use (similar to Ultraverse RX and same as Metacross OTW) ●
  • . Same target population (same as Ultraverse RX and Metacross OTW)
  • Same operating principle (same as Ultraverse RX and Metacross OTW) ●
  • Similar materials (similar to Ultraverse RX and Metacross OTW) ●
  • Same fundamental scientific technology (same as Ultraverse RX and Metacross OTW)
  • . Same sterility assurance level and method of sterilization (same as Ultraverse RX and Metacross OTW)

Performance Testing

To demonstrate substantial equivalence of Crosperio OTW to the predicate devices, the technological characteristics and performance criteria were evaluated using the bench testing recommendations outlined in the FDA Guidance Document "Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters'' dated September 8, 2010. The following in vitro tests were performed on the subject device:

Dimensional Verification, Balloon Preparation, Deployment and Retraction, Balloon Rated Burst Pressure, Balloon Fatigue (Repeat Balloon Inflations), Balloon Compliance, Balloon Inflation and Deflation Time, Catheter Bond Strength, Flexibility and Kink Test, Torque Strength, Radiopacity, Coating Integrity, Particulate Evaluation, Balloon Rated Burst Pressure (in Stent), and Balloon Fatigue (Repeat Balloon Inflations; in Stent), and Shelf Life Testing.

The results from these tests demonstrate that the technological characteristics and performance criteria of the Crosperio OTW are adequate for the intended use of the device

{7}------------------------------------------------

and that the device can perform in a manner equivalent to devices currently on the market with the same intended use.

Biocompatibility:

To demonstrate the biological safety of the body-contacting materials and substantial equivalence of the Crosperio OTW to the predicate devices, the following biocompatibility testing was performed in accordance with "Draft Guidance for Industry and Food and Drug Administration Staff; Use of International Standard ISO 10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing'" (dated April 23, 2013):

Cytotoxicity, Sensitization, Intracutaneous reactivity (irritation), Systemic toxicity (acute), Pyrogenicity, Hemocompatibility (thrombogenicity, hemolysis, and immunology), and Genotoxicity (bacterial gene mutation assay, in vitro mammalian genotoxicity assay, and in vivo cytogenetics assay)

The results from these tests demonstrate that the Crosperio OTW is biocompatible for its intended use similar to the predicate devices.

Conclusions:

The Crosperio OTW met all the predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs. The Crosperio OTW PTA Balloon Dilatation Catheter is substantially equivalent to legally marketed predicate devices.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).