K Number
K153558

Validate with FDA (Live)

Date Cleared
2016-04-21

(129 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Ventix Suture Anchors are indicated for use in soft tissue reattachment procedures in the shoulder, wrist/hand, ankle/foot, elbow, knee, and hip. Specific indications are:

Shoulder: Bankart repair, SLAP lesion repair, acromio- clavicular separation, rotator cuff repair, capsule repair or capsulolabral reconstruction, biceps tenodesis, deltoid repair, pectoralis major repair.

Wrist/Hand: Scapholunate ligament reconstruction, ulnar/radial collateral ligament reconstruction.

Ankle/Foot: Lateral stabilization, medial stabilization, Achilles tendon repair/reconstruction, hallux valgus reconstruction, mid- and forefoot reconstruction.

Elbow: Ulnar or radial collateral ligament reconstruction, biceps tendon reconstruction, Lateral epicondylitis repair

Knee: Medial collateral ligament repair, lateral ligament repair, posterior oblique ligament repair, joint capsule closure, iliotibial band tenodesis, and patellar ligament/tendon repair, medial patellofemoral ligament (MPFL) repair or reconstruction, quadriceps tendon repair.

Hip: Proximal hamstring repair

Device Description

The Ventix Suture Anchor is a two-piece design consisting of the vented body and the threaded tip preloaded with suture. The anchor will be offered in three sizes: 3.5mm, 4.75mm, and 5.5mm and is manufactured from polyetheretherketone (PEEK). The Ventix Suture Anchor will be loaded onto a disposable inserter and will be available with #2 MaxBraid or 1.5mm ForceFiber OrthoTape suture.

AI/ML Overview

The provided document is a 510(k) premarket notification for the Ventix Suture Anchor. It primarily focuses on demonstrating substantial equivalence to legally marketed predicate devices, rather than establishing acceptance criteria and detailed study results for the device's performance against specific metrics.

Therefore, much of the requested information about acceptance criteria, detailed study results, sample sizes for test and training sets, ground truth establishment, expert qualifications, adjudication methods, and MRMC studies cannot be extracted from this document.

However, I can provide the limited information available regarding performance testing:

1. A table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance
Pull-out strength: Statistically equivalent or greater fixation strength compared to predicate devices.The Ventix Suture Anchors provide statistically equivalent or greater fixation strength to the predicate devices.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided in the document. The document only mentions that non-clinical laboratory testing was performed using the Ventix Suture Anchors and predicate devices for comparison.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable as the document describes non-clinical mechanical testing, not a study involving expert assessment of medical images or patient outcomes.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable as the document describes non-clinical mechanical testing, not a study involving human readers or subjective assessments.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There was no MRMC comparative effectiveness study done. The document explicitly states "Clinical Tests - None provided as a basis for substantial equivalence." The testing described is non-clinical mechanical testing.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable. The device is a physical suture anchor, not an algorithm or AI system. The testing described is mechanical performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the non-clinical pull-out strength testing, the "ground truth" would be the measured mechanical pull-out strength values obtained directly from the testing equipment. This is objective data.

8. The sample size for the training set

This information is not applicable. The device is a physical product, not an AI or machine learning algorithm that requires a training set.

9. How the ground truth for the training set was established

This information is not applicable as there is no training set for a physical device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized design of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a sense of depth and unity. The seal is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 21, 2016

Biomet Manufacturing Corporation Adam Cargill Regulatory Affairs Specialist 56 East Bell Drive P.O. Box 587 Warsaw, Indiana 46581

Re: K153558

Trade/Device Name: Ventix Suture Anchor Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: March 23, 2016 Received: March 24, 2016

Dear Mr. Cargill:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Lori A. Wiggins -S

for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K153558

Device Name Ventix Suture Anchor

Indications for Use (Describe)

The Ventix Suture Anchors are indicated for use in soft tissue reattachment procedures in the shoulder, wrist/hand, ankle/foot, elbow, and knee. Specific indications are:

Shoulder: Bankart repair, SLAP lesion repair, acromio- clavicular separation, rotator cuff repair, capsule repair or capsulolabral reconstruction, biceps tenodesis, deltoid repair, pectoralis major repair.

Wrist/Hand: Scapholunate ligament reconstruction, ulnar/radial collateral ligament reconstruction.

Ankle/Foot: Lateral stabilization, medial stabilization, Achilles tendon repair/reconstruction, hallux valgus reconstruction, mid- and forefoot reconstruction.

Elbow: Ulnar or radial collateral ligament reconstruction, biceps tendon reconstruction, Lateral epicondylitis repair

Knee: Medial collateral ligament repair, lateral ligament repair, posterior oblique ligament repair, joint capsule closure, iliotibial band tenodesis, and patellar ligament/tendon repair, medial patellofemoral ligament (MPFL) repair or reconstruction, quadriceps tendon repair.

Hip: Proximal hamstring repair

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Ventix Suture Anchor | Traditional 510(k)

510(k) Summary

In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the Ventix Suture Anchor 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)s', issued on August 12, 2005.

Sponsor:Biomet Inc.56 East Bell DrivePO Box 587Warsaw, IN 46581Establishment Registration Number: 1825034Phone: (574) 267-6639Fax: (574) 267-8137
Contact:Adam CargillRegulatory Affairs SpecialistBiomet Sports Medicine
Date:December 11, 2015
Subject Device: Trade
Name: Ventix Suture Anchor
Common Name: Soft Tissue Anchor
Classification Name:• MBI - Fastener, Fixation, Nondegradable, Soft Tissue(21 CFR 888.3040)

Legally marketed devices to which substantial equivalence is claimed:

  • . K060693 – Allthread PEEK Suture Anchors
  • . K150768 - JuggerKnot Soft Anchors

Device Description

The Ventix Suture Anchor is a two-piece design consisting of the vented body and the threaded tip preloaded with suture. The anchor will be offered in three sizes: 3.5mm, 4.75mm, and 5.5mm and is manufactured from polyetheretherketone (PEEK). The Ventix Suture Anchor will be loaded onto a disposable inserter and will be available with #2 MaxBraid or 1.5mm ForceFiber OrthoTape suture.

Intended Use and Indications for Use

The Ventix Suture Anchors are indicated for use in soft tissue reattachment procedures in the shoulder, wrist/hand, ankle/foot, elbow, knee, and hip. Specific indications are:

{4}------------------------------------------------

BIOME

Ventix Suture Anchor | Traditional 510(k)

Shoulder: Bankart repair, SLAP lesion repair, acromio- clavicular separation, rotator cuff repair, capsule repair or capsulolabral reconstruction, biceps tenodesis, deltoid repair, pectoralis major repair.

Wrist/Hand: Scapholunate ligament reconstruction, ulnar/radial collateral ligament reconstruction.

Ankle/Foot: Lateral stabilization, medial stabilization, Achilles tendon repair/reconstruction, hallux valgus reconstruction, mid- and forefoot reconstruction.

Elbow: Ulnar or radial collateral ligament reconstruction, biceps tendon reconstruction, Lateral epicondylitis repair

Knee: Medial collateral ligament repair, lateral ligament repair, posterior oblique ligament repair, joint capsule closure, iliotibial band tenodesis, and patellar ligament/tendon repair, medial patellofemoral ligament (MPFL) repair or reconstruction, quadriceps tendon repair.

Hip: Proximal hamstring repair

Summary of Technological Characteristics

The technological characteristics (materials, design, sizing, and indications) of the Ventix Suture Anchors are similar to the predicate devices or other previously cleared devices with the main difference being the Ventix Suture Anchors have a vented implant when compared to the ALLThread PEEK Suture Anchors.

The rationale for substantial equivalence is based on consideration of the following characteristics:

  • . Intended Use: The proposed Ventix Suture Anchors are intended for use in soft tissue reattachment procedures in the shoulder, wrist/hand, ankle/foot, elbow, knee, and hip which is the same as the predicate devices.
  • . Indications for Use: Indications are identical to the predicate devices.
    • . Materials: The proposed Ventix Suture Anchor uses the identical implant material as the predicate device.
    • . Design Features: The proposed Ventix Suture Anchor incorporates similar design features as the predicate device.
  • . Sterilization: The proposed Ventix Suture Anchor is provided sterile via ethylene oxide (EtO), the same sterilization method utilized for the predicate devices.

Summary of Performance Data (Nonclinical and/or Clinical)

  • . Non-Clinical Tests
    Non-clinical laboratory testing was performed to verify the pull-out strength o of the Ventix Suture Anchors by mechanical pullout testing as compared to the ALLThread PEEK Suture Anchors for substantial equivalence and the JuggerKnot Soft Anchors for use in expanded indications. The test results indicate that the Ventix Suture Anchors provide statistically equivalent or greater fixation

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the word "BIOMET" in a simple, sans-serif font. The letters are all capitalized and appear to be a dark gray or black color. The word is presented against a white background, which provides a clear contrast and makes the text easily readable. The overall design is clean and minimalist, focusing on the clarity and legibility of the word itself.

Ventix Suture Anchor | Traditional 510(k)

strength to the predicate devices and would be functional within the intended use.

  • . Clinical Tests
    • o None provided as a basis for substantial equivalence.

Substantial Equivalence Conclusion

The proposed Ventix Suture Anchor has similar intended use, technological characteristics, and mechanical performance as the predicate devices. The performance testing data identified no new risks and substantial equivalence to the legally marketed predicate devices.

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.